NEGATIVE POSITIVE. Rev , 06/11. 5 min. Mono Test. For fi ngertip. blood: 1 DROP 1 DROP For serum, whole blood. plasma or. samples.

Similar documents
Mono Test 1 DROP 1 DROP. For fingertip blood: For serum, plasma or whole blood samples in tubes: Rev , 05/09

SAMPLE PROCEDURE 907-9, 06/11

FOR INFORMATIONAL USE ONLY

SAMPLE PROCEDURE , 07/11

PROCEDURE MANUAL. Procedure: CLIA Complexity: Waived for Whole Blood, Moderate for Serum, Plasma. Prepared By Date Adopted Supersedes Procedure #

Rubella Latex Agglutination Test

Infectious Mononucleosis IM Cassette Test RAPU04A830

SAMPLE PROCEDURE , 07/11

QUICK REFERENCE INSTRUCTIONS. THYROCHEK TSH Cassette

Rotavirus Test Kit. Instructions For Use. Format: Cassette Specimen: Fecal Extract Catalog Number: VEL-001-ROTA

CDIA TM Rubella IgG/IgM Rapid Test Kit

See external label 2 C-8 C = C-REACTIVE PROTEIN (CRP) LATEX SLIDE TEST

PROCEDURE. Sekisui Diagnostics OSOM ifob Rapid Test. Title: Procedure #: Institution: Prepared by: Date: Title: Accepted by: Date adopted: Title:

Strep-a-Test Twister Test

Helicobacter pylori Antigen Test

Rapid-VIDITEST. Strep A Card. One step Strep A Card for the detection of Group A Streptococcal antigen from throat swabs or culture.

Hepatitis C Virus (HCV) Antibody Test

Rubella virus IgG ELISA Kit

Human Influenza A (Swine Flu) Rapid test

Rapid-VIDITEST. Strep A Blister. One step Strep A Blister for the detection of Group A Streptococcal antigen from throat swabs or culture.

Rapid-VIDITEST. Astrovirus Card

Doc#: ASI COLOR MONO II TEST

Rapid-VIDITEST. Influenza A

See external label 2 C-8 C 96 tests Chemiluminescence. CMV IgM. Cat # Diluted samples, controls & calibrator 100 µl 30 minutes

Step-by-Step Instructions For OraQuick HCV Rapid Antibody Test

Rapid-VIDITEST Swine Flu

IBD EZ VUE I. PRINCIPLE

Dengue IgG/IgM/NS1 Combo Rapid Test Device

Human Cytomegalovirus IgM ELISA Kit

Treponema Pallidum (TP) Antibody Test

HAV IgG/IgM Rapid Test

See external label 2 C-8 C 96 tests CHEMILUMINESCENCE. CMV IgG. Cat # Step (20-25 C Room temp.) Volume

WHO Prequalification of Diagnostics Programme PUBLIC REPORT

Rapid-VIDITEST. Influenza A+B

DIAGNOSTIC AUTOMATION, INC.

H. pylori IgM CLIA kit

TB LAM Ag Lateral Flow Assay Standard Operating Procedure

Toxoplasma gondii IgM ELISA Kit

Mycoplasma pneumoniae IgG ELISA Kit

Toxoplasma gondii IgM ELISA Kit

CYTOMEGALOVIRUS (CMV) IgM ELISA Kit Protocol

See external label 96 tests HSV 2 IgA. Cat #

Toxoplasma gondii IgM (Toxo IgM)

EBV-EA IgG. Cat # 1415Z. EBV -EA IgG ELISA. ELISA: Enzyme Linked Immunosorbent Assay. ELISA - Indirect; Antigen Coated Plate

Human Cytomegalovirus Virus (CMV) IgG ELISA Kit

ifob Test Rev. P5576-A, 04/16 5. Read the result in 5-10 minutes. RESULTS AFTER 10 MINUTES. Important: DO NOT READ

Rapid-VIDITEST FOB Blister

Malaria Combo Test Kit

CLIA Complexity: WAIVED INTENDED USE SUMMARY AND EXPLANATION

H.Pylori IgG

Herpes Simplex Virus 2 IgM HSV 2 IgM

Rapid-VIDITEST. Rota-Adeno Blister. One Step Rotavirus and Adenovirus Antigen Blister test. Instruction manual

Chlamydia trachomatis (CHLa)Test Kit

H.pylori IgA Cat #

See external label 2 C-8 C Σ=96 tests Cat # EBV-VCA IgA. Cat # EBV -VCA IgA ELISA. ELISA: Enzyme Linked Immunosorbent Assay

ifob Test Rev. P /15 6. Read the result in 5-10 minutes. RESULTS AFTER 10 MINUTES. Important: DO NOT READ

H.Pylori IgG Cat # 1503Z

HAV IgM Rapid Test(Cassette)

H. pylori IgM. Cat # H. pylori IgM ELISA. ELISA: Enzyme Linked Immunosorbent Assay. ELISA - Indirect; Antigen Coated Plate

CHEMILUMINESCENCE ENZYME IMMUNOASSAY (CLIA) Toxoplasma IgG. Cat # (20-25 C Room temp.) Volume

H.Pylori IgM Cat # 1504Z

WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: DPP HIV 1/2 Assay WHO reference number: PQDx

Herpes Simplex Virus 2 IgG HSV 2 IgG

Rapid-VIDITEST Calprotectin

SIV p27 Antigen ELISA Catalog Number:

WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: OraQuick HIV 1/2 Rapid Antibody Test WHO reference number: PQDx

QUICK REFERENCE INSTRUCTIONS For use with Sofia and Sofia 2. Rx only

QUICK REFERENCE INSTRUCTIONS

NEPHROCHECK Calibration Verification Kit Package Insert

WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: OraQuick HIV 1/2 Rapid Antibody Test WHO reference number: PQDx

Helicobacter pylori IgA ELISA Kit

CrAg Lateral Flow Assay Standard Operating Procedure

HSV-1 IgM ELISA. Catalog No (96 Tests) For Research Use Only. Not for use in Diagnostic Procedures.

Measles IgM ELISA Kit

Malaria Pf/pan antigen Rapid Test

Rapid-VIDITEST Shigella dysenteriae

Effective Date: 01/07 Supersedes Revision/Date: 01/02 Revision: 01/07 Date Adopted:

DETERMINE SPECIMEN REQUIREMENTS HIV-1/2 Ag/Ab Combo TESTING PROCEDURE

Rapid-VIDITEST C. difficile Ag (GDH) Card/Blister

Human HBcAb IgM ELISA kit

NEPHROCHECK Liquid Control Kit Package Insert

IMMUNODOT MONO-M TEST

CLIA Complexity: WAIVED

AN IMMUNOASSAY TEST FOR THE QUALITATIVE DETECTION OF STREP A ANTIGEN IN THROAT SWAB SPECIMENS

HBeAg and HBeAg Ab ELISA Kit

EBV-VCA IgA Cat #

NCCLS. Accutest H. pylori WB Test. (Catalog. No. ID-92A225) For Professional and In Vitro Diagnostic Use Only

Unigold Recombigen HIV 1/2 Training for HIV Testing Sites. Updated: February 2018 Cicely Richard Office of HIV/AIDS

EBV-VCA IgM Enzyme Immunoassay

CLIA Complexity: Waived

LeadCare BLOOD LEAD ANALYZER. Quick Reference Guide

Hepatitis A virus IgM ELISA Kit

IgG Antibodies To Toxoplasma Gondii ELISA Kit Protocol

MALARIA P. FALCIPARUM / P. VIVAX

See external label 2 C-8 C Σ=96 tests Cat # 1505Z. MICROWELL ELISA H.Pylori IgA Cat # 1505Z

IVD information *Droppers for the sensitized and control cells. Not for use other than dispensing the sensitized and control cells.

Effective Date: 09/08 Supersedes Revision/Date: Original Revision: 09/08 Date Adopted:

To provide you with necessary knowledge and skills to accurately perform 3 HIV rapid tests and to determine HIV status.

ONE-STEP WHOLE BLOOD H.PYLORI TEST. Product Code: 24529

INSTANT-VIEW H. pylori Rapid Test CLIA WAIVED Test for Whole Blood QUALITATIVE IN-VITRO DIAGNOSTIC TEST FOR EXTERNAL USE ONLY

Transcription:

Mono Test 1 2 1 DROP 1 DROP For serum, plasma or whole blood samples in tubes: For fi ngertip blood: 3 4 5 min POSITIVE NEGATIVE Rev. 3078-0, 06/11

Mono Test CLIA Complexity: Waived for Whole Blood Non-Waived for Serum or Plasma FOR LABORATORY AND PROFESSIONAL IN VITRO DIAGNOSTIC USE ONLY. INTENDED USE The is intended for the qualitative detection of infectious mononucleosis heterophile antibodies in serum, plasma or whole blood as an aid in the diagnosis of infectious mononucleosis. SUMMARY AND EXPLANATION OF TEST The diagnosis of infectious mononucleosis (IM) is suggested on the basis of the clinical symptoms of fever, sore throat and swollen lymph glands. The highest incidence of symptomatic IM occurs during late adolescence (15 24 years of age). Infectious mononucleosis is caused by the Epstein-Barr Virus (EBV). (1,2) The laboratory diagnosis of IM is based on the detection of IM heterophile antibodies. These heterophile antibodies are directed against antigens found in bovine, sheep and horse erythrocytes. The OSOM Mono Test utilizes an extract of bovine erythrocytes to give the required sensitivity and specifi city. PRINCIPLES OF TEST The OSOM Mono Test uses color immunochromatographic dipstick technology with bovine erythrocyte extract coated on the membrane. In the test procedure, serum, plasma or whole blood is mixed with the Diluent. Then the Test Stick is placed in the mixture and the mixture migrates along the membrane. If the specifi c IM heterophile antibody is present in the sample, it will form a complex with the bovine erythrocyte extract conjugated color particles. The complex will then be bound by bovine erythrocyte extract immobilized on the membrane and a visible blue Test Line will appear to indicate a positive result. KIT CONTENTS AND STORAGE 25 Test Sticks in a container 25 Test Tubes 25 Transfer Pipettes 25 Capillary Tubes with 1 Capillary Bulb 1 Diluent (contains buffer with 0.2% sodium azide) 1 Mono Positive Control (contains rabbit anti-beef stroma in tris buffer with 0.2% sodium azide and 0.05% gentamycin sulfate preservatives) 1 Mono Negative Control (contains goat albumin in tris buffer with 0.2% sodium azide) 1 Work Station 1 Directional Insert 2 Additional test sticks have been included in the kit for external QC testing Note: Extra components (tubes, pipettes, capillary tubes and capillary bulb) have been provided for your convenience. Store the Test Sticks and reagents tightly capped at 15 30 C (59 86 F). Do not use the Test Sticks or reagents after their expiration dates. MATERIALS REQUIRED BUT NOT PROVIDED Specimen collection containers. A timer or watch. PRECAUTIONS For in-vitro diagnostic use only. Follow your laboratory safety guidelines in the collection, handling, storage and disposal of patient specimens and all items exposed to patient specimens. The Diluent and Controls contain sodium azide which may react with lead or copper plumbing to form potentially explosive metal azide. Large quantities of water must be used to fl ush discarded Diluent or Controls down a sink. The Capillary Bulb contains dry natural rubber. Do not interchange or mix components from different kit lots. SPECIMEN COLLECTION AND PREPARATION Serum, Plasma, or Whole Blood Sample Obtain specimens by acceptable medical technique. Collect whole blood samples using a tube containing EDTA or heparin as an anticoagulant. Other anticoagulants have not been tested. Serum and plasma specimens may be refrigerated (2 8 C; 36 46 F) and tested within 48 hours; serum and plasma specimens held for longer times should be frozen (below -10 C; 14 F) and tested within 3 months. Test whole blood specimens within 24 hours. Specimens must be at room temperature (15 30 C; 59 86 F) when tested. Fingertip Whole Blood Hold the capillary tube horizontally while collecting the sample. Touch the end of the capillary tube to the drop of blood on the patient s fi nger. Fill the capillary tube completely. Place the small end of the black bulb onto the capillary tube. Place your fi ngertip over the opening in the bulb. Squeeze the bulb to dispense the whole blood sample into the test tube.

QUALITY CONTROL External Quality Control For external QC testing, use the controls provided in the kit. Add one free falling drop of control to the Test Tube and then proceed in the same manner as with a patient sample. Quality Control requirements should be established in accordance with local, state and federal regulations or accreditation requirements. Minimally, Sekisui Diagnostics recommends that positive and negative external controls be run with each new lot and with each new untrained operator. Some commercial controls may contain interfering additives. The use of these controls is not recommended. Internal Quality Controls The OSOM Mono Test provides two levels of internal procedural controls with each test procedure. The red Control Line is an internal positive procedural control. The Test Stick must absorb the proper amount of test material and be working properly for the red Control Line to appear. A clear background is an internal negative procedural control. If the test has been performed correctly and the Test Stick is working properly, the background will clear to give a discernible result. If the red Control Line does not appear, the test may be invalid. If the background does not clear and interferes with the test result, the test may be invalid. Call Sekisui Diagnostics Technical Assistance if you experience either of these problems. LIMITATIONS As with all diagnostic assays, the results obtained by this test yield data that must be used as an adjunct to other information available to the physician. The OSOM Mono Test is a qualitative test for the detection of IM heterophile antibody. A negative result may be obtained from patients at the onset of the disease due to heterophile antibody levels below the sensitivity of this test kit. If symptoms persist or intensify, the test should be repeated. Some segments of the population with acute IM are heterophile antibody negative. (1) EXPECTED VALUES A heterophile antibody response is observed in approximately 80 90% of adults and children with EBV-caused IM. This percentage drops to approximately 50% for children under four years of age. (1) While the incidence of IM refl ects wide seasonal, ethnic and geographical variation, a large epidemiological study noted that the highest incidence of symptomatic IM occurred during late adolescence (15 24 years of age) (2). PERFORMANCE CHARACTERISTICS A total of 439 specimens (183 serum, 176 plasma and 80 whole blood) were evaluated by two clinical labs in a clinical study. Test results of the OSOM Mono Test were compared to results obtained with a commercially available latex particle agglutination test for the qualitative determination of infectious mononucleosis heterophile antibodies. Discrepancies between the results given by the OSOM Mono Test and the latex particle agglutination test were resolved by Epstein-Barr Virus (EBV) specifi c serological assays. In these assays, the specifi c antibodies to the EBV capsid antigen (IgM) and EBV nuclear antigen-1 (IgM and IgG) were determined. Serum Specimens: Whole Blood Specimens: + 74 8* + 30 3* 0 101 0 47 *6 out of 8 tested postitive by EBV testing *1 out of 3 tested postitive by EBV testing Plasma Specimens: + 67 15* 0 94 *8 out of 15 tested postitive by EBV testing All Specimens: + 171 26* 0 242 *15 out of 26 tested postitive by EBV testing When compared to a commercially available latex particle agglutination test for infectious mononucleosis heterophile antibodies, the OSOM Mono Test showed a sensitivity of 100% and a specifi city of 90.3%. The overall agreement was 94.1%. Fifteen of the 26 discrepant samples were determined to be recent or acute EBV infections by EBV serological testing, in which case the sample was considered positive. Including the samples confi rmed positive by EBV serological testing, the overall clinical study specifi city of the OSOM Mono Test is 95.9% and the overall sensitivity is 100%. POL Studies An evaluation of the OSOM Mono Test was conducted at three physicians offi ces or clinical laboratories where testing was performed by personnel with diverse educational backgrounds. Each site tested the randomly coded panel consisting of negative (5), low positive (3) and moderate positive (4) specimens for three days. The results obtained had 99.1% agreement (107/108) with the expected results.

TEST PROCEDURE Absorbent End Result Window Handle End STEP 1 Addition of Specimen For serum, plasma, or whole blood samples in tubes: Use the Transfer Pipette provided and add one drop to the Test Tube. For fi ngertip blood: After fi lling a capillary tube end to end, dispense all of the blood into the Test Tube. STEP 2 Slowly add 1 drop of Diluent to the bottom of the Test Tube. Mix. STEP 3 Remove the Test Stick(s) from the container. Re-cap the container immediately. Place the Absorbent End of the Test Stick into the treated sample. Leave the Test Stick in the Test Tube. 5 min STEP 4 Read results at 5 minutes. Positive results may be read as soon as the red Control Line appears. REFERENCES 1. Lennette, E.T., Epstein-Barr Virus, in Manual of Clinical Microbiology. Balows, A., Hausler, W.J. Jr., Herrmann, K.L., Isenberg, H.D., Shadomy, H.J., Editors, 5th Edition, American Society for Microbiology, Washington D.C., pp 847 852, 1991. 2. Heath, C.W. Jr., Brodsky, A.L., Potolsky, A.l., Infectious Mononucleosis in a General Population. Am. J. Epidemiol., 95:46, 1972. ASSISTANCE For technical assistance, call Sekisui Diagnostics Technical Assistance at (800) 332-1042. MANUFACTURED BY Sekisui Diagnostics, LLC 6659 Top Gun Street San Diego, CA 92121 USA RE-ORDER No.145 (25 Tests) OSOM and the QC Inside logo are registered trademarks of Sekisui Diagnostics, LLC. Licensed under U.S. Patent Nos. 5,714,389; 5,989,921 and 6,485,982 and related non-u.s. patents and patent applications.

INTERPRETATION OF TEST RESULTS Positive A blue Test Line and a red Control Line is a positive result for the detection of infectious mononucleosis heterophile antibody. Note that the blue line can be any shade of blue and can be lighter or darker than the line in the picture. Negative A red Control Line but no blue Test Line is a negative result. No infectious mononucleosis heterophile antibody has been detected. Invalid If after 5 minutes, no red Control Line appears or background color makes reading the red Control Line impossible, the result is invalid. If this occurs, repeat the test on a new Test Stick or call Sekisui Diagnostics Technical Assistance. Notes A blue or red line which appears uneven in color density is considered a valid result. Framingham, MA 01701 Tel: 800-332-1042 www.sekisuidiagnostics.com 2011 Sekisui Diagnostics, LLC All rights reserved.

KEY TO COMPONENT LABELING Use by YYYY-MM Batch code Catalog number Contents suffi cient for <n> tests In vitro diagnostic medical device Temperature limitation Manufacturer/Manufactured by Consult instructions for use Authorized representative in the European Community Caution, consult accompanying documents.