Optimising tracheal intubation success rate using the Airtraq laryngoscope

Similar documents
Comparison of two methods of gum elastic bougie aided endotracheal intubation using Airtraq video laryngoscope

Omiros Chalkeidis*, Georgios Kotsovolis, Apostolos Kalakonas, Maria Filippidou, Christos Triantafyllou, Dimitris Vaikos, Epaminondas Koutsioumpas

RESPIRATION AND THE AIRWAY. Editor s key points

Samal et al: Comparison of the Macintosh and Airtraq Laryngoscope

Is the Airtraq optical laryngoscope effective in tracheal intubation by novice personnel?

Evaluation of the Airtraq Ò and Macintosh laryngoscopes in patients at increased risk for difficult tracheal intubation*

British Journal of Anaesthesia 100 (4): (2008) doi: /bja/aen002 Advance Access publication January 31, 2008

Displacement of the epiglottis during intubation with the Pentax-AWS Airway Scope. Suzuki, Akihiro ; Katsumi, Norifumi ; Honda, Takashi ; Sasakawa,

Comparison of the Berman Intubating Airway and the Williams Airway Intubator for fibreoptic orotracheal intubation in anaesthetised patients.

LMA CTrach TM in patients with anticipated difficult airway: A retrospective study

British Journal of Anaesthesia 102 (1): (2009) doi: /bja/aen342

Cervical spine movement during laryngoscopy using the Airway Scope compared with the Macintosh laryngoscope

Comparison of the Airtraq to the Bonfils Fibroscope for Endotracheal Intubation in a Simulated Difficult Airway

Review Article. Pentax Airway Scope â vs Macintosh laryngoscope for tracheal intubation in adult patients: a systematic review and metaanalysis*

Airway Scope for Emergency lntubations: Usefulness of a New Video-laryngoscope

INFORMED CONSENT FOR publication of the present

British Journal of Anaesthesia 101 (6): (2008) doi: /bja/aen288 Advance Access publication October 3, 2008

A comparison between the GlideScope Ò Video Laryngoscope and direct laryngoscope in paediatric patients with difficult airways a pilot study

Airway Management & Safety Concerns Experience from Bariatric Surgery

A clinical evaluation of the Bonfils Intubation Fibrescope*

The LMA CTrach TM, a new laryngeal mask airway for endotracheal intubation under vision: evaluation in 100 patients

Clinical Study McGrath Video Laryngoscope May Take a Longer Intubation Time Than Macintosh Laryngoscope

Difficult Airway. Department of Anesthesiology University of Colorado Health Sciences Center (prepared by Brenda A. Bucklin, M.D.)

Corresponding Author:Dr Nidhi P Sehgal

The reasons 13/11/ Cost 2. Availability 3. Comparison 4. Complications 5. Knowledge. Pulmonary and critical care medicine (PCCM) fellows.

(ix) Difficult & Failed Intubation Queen Charlotte s Hospital

If you suspect airway problems, get a second opinion before you anaesthetise, not after!

Original Contributions

Dr.Bharghavi.M 2 nd year post graduate Dept of Anaesthesia

Diagnosis & Management of the Difficult Airway

Hemodynamic response to laryngoscopy in ischemic heart disease: Macintosh blade versus GlideScope videolaryngoscope

Effect of cricoid pressure on laryngoscopic view and ease of intubation with Airtraq laryngoscope

LEVITAN S FIBREOPTIC STYLET: BEYOND BARRIERS. - Our Perspective.

Comparison of the Baska â mask with the single-use laryngeal mask airway in low-risk female patients undergoing ambulatory surgery

In 2011 I received an unrestricted lecture honorarium from Ambu I have been loaned equipment by Intavent Direct (Teleflex) and Aircraft Medical As an

The Glidescope Ò system: a clinical assessment of performance

9/20/18. Ear To Sternal Notch. Primary Methods to Rescue & Prevent Failed Intubation. Ear to Sternal Notch Position

Tracheal intubation in patients with rigid collar immobilisation of the cervical spine: a comparison of Airtraq and LMA CTrach TM devices*

Laryngoscopy: Time to broaden our horizon.

How do you use a bougie as an airway adjunct for endotracheal intubation?

Current Anaesthesia & Critical Care

ISPUB.COM. The Video-Intubating Laryngoscope. M Weiss THE LARYNGOSCOPE INTRODUCTION TECHNICAL DESCRIPTION

Video Laryngoscopy and its future in Airway Management

Airway Anatomy. Soft palate. Hard palate. Nasopharynx. Tongue. Oropharynx. Hypopharynx. Thyroid cartilage

Video laryngoscopy in paediatric anaesthesia in South Africa

Initial anatomic investigations of the I-gel airway: a novel supraglottic airway without inflatable cuff*

Research Article Comparison of Direct and Indirect Laryngoscopes in Vomitus and Hematemesis Settings: A Randomized Simulation Trial

Original Article. Summary. Introduction

Clinical Studies. Letters to the Editor Manikin Studies

Force and pressure distribution using Macintosh and GlideScope laryngoscopes in normal and difficult airways: a manikin study

Aiji Sato (Boku) 1*, Kazuya Sobue 2, Eisuke Kako 2, Naoko Tachi 1, Yoko Okumura 1, Mayuko Kanazawa 1, Mayumi Hashimoto 1 and Jun Harada 2

Randomized controlled trial of the A.P. Advance, McGrath, and Macintosh laryngoscopes in normal and difficult intubation scenarios: a manikin study

Comparison between C-MAC video-laryngoscope and Macintosh direct laryngoscope during cervical spine immobilization

Use of the Aintree Intubation Catheter with the Laryngeal Mask Airway and a Fiberoptic Bronchoscope in a Patient with an Unexpected Difficult Airway

Airway/Breathing. Chapter 5

Airway Workshop Lecture. University of Ottawa

How to Predict and Avoid Airway Disasters. Muhammad Umer Ihsan

Educational Session: Evaluation and Management of the Difficult Airway

FACTOR ANALYSIS IN DIFFICULT TRACHEAL INTUBATION: LARYNGOSCOPY-INDUCED AIRWAY OBSTRUCTION

EUROANESTHESIA 2007 Munich, Germany, 9-12 June 2007

British Journal of Anaesthesia 101 (5): (2008) doi: /bja/aen231 Advance Access publication September 9, 2008

Airway/Breathing. Chapter 5

RESPIRATION AND THE AIRWAY Comparison of the performance of four laryngoscopes in a high-fidelity simulator using normal and difficult airway

Haemodynamic response to endotrachial intubation: direct versus video laryngoscopy

Original Article. An assessment of oropharyngeal airway position using a fibreoptic bronchoscope. Summary. Introduction. Methods

Incidence and predictors of difficult nasotracheal intubation with airway scope

The implication of using dominant hand to perform laryngoscopy: an analysis of the laryngoscopic view and blade-tooth distance

Cronicon ANAESTHESIA. Khaled Mohammed Elnaghy 1 and Javier Yuste 2 * Abstract. Keywords: Intubation; Video laryngoscopes; GlideScope; LMA CTrach

Pre anaesthetic evaluation of difficult tracheal intubation

Accident and Emergency Medicine Academic Unit, The Chinese University of Hong Kong, Shatin, Hong Kong

Comparative Evaluation of Performance of Videolaryngoscope vs Fastrach Intubating Laryngeal Mask Airway

ACCORDING to the American Society of Anesthesiologists,

Anesthetic challenges when elective case becomes emergent

Preface... Acknowledgements... Contributors... 1 The Difficult Airway: Definitions and Algorithms The Expected Difficult Airway...

The Superiority Of Mcgrath Videolaryngoscope After Failed Conventional Laryngoscopy

Advanced Airway Management. University of Colorado Medical School Rural Track

RESPIRATION AND THE AIRWAY. Key points. K. B. Greenland 1,2,3 *, M. J. Edwards 1, N. J. Hutton 1, V. J. Challis 4, M. G. Irwin 3 and J. W.

External Laryngeal Manipulation for Better Laryngeal View

Hong Kong Journal of Emergency Medicine. KL Tsui, CY Hung, CW Kam. t p= p=0.004

Original Article. Tracheal intubation with a camera embedded in the tube tip (Vivasight TM ) Summary. Methods

Updates on Airway Management

Cervical Spine Motion

DIFFICULT AIRWAY MANAGMENT. Dr.N.SANTHOSH KUMAR MD ANESTHESIA (2 nd Yr)

The Laryngeal Mask and Other Supraglottic Airways: Application to Clinical Airway Management

Comparison. F Agro immobilization. to DL Trauma/SCI. GS experiences. to other. GS experiences. GS experiences. to DL.

12 Larynx. I - Cartilages. Learning Objectives

A clinical sign to predict difficult tracheal intubation: a prospective study

Video Laryngoscopes Novelty, Mandatory or Morbidity. Dr Brent May FANZCA MBBS Specialist Anaesthetist Retrieval Consultant - Adult Retrieval Victoria

NECK CIRCUMFERENCE TO THYROMENTAL DISTANCE RATIO IN COMPARISON WITH MALLAMPATI TEST: A STUDY TO PREDICT DIFFICULT INTUBATION

Unanticipated difficult tracheal intubation - during routine induction of anaesthesia in an adult patient

A systematic review of the role of videolaryngoscopy in successful orotracheal intubation

PREDICTORS OF DIFFICULT INTUBATION: STUDY IN KASHMIRI POPULATION

MD (Anaesthesiology) Title (Plan of Thesis) (Session )

Orotracheal Intubation

Comparison of sniffing position and simple head extension for visualization of glottis during direct laryngoscopy.

LARYNGOSCOPES/TRACHEASCOPES

Am J Nucl Med Mol Imaging 2018;8(3): /ISSN: /ajnmmi Zhi-Qiang Zhou 1, Xu Zhao 2, Hong-Bing Xiang 1

Endotracheal intubation using semi-rigid optical stylet in simulated difficult airways of high grade modified Cormack and Lehane laryngeal views

Research Article The AirView Study: Comparison of Intubation Conditions and Ease between the Airtraq-AirView and the King Vision

Success of Tracheal Intubation with Intubating Laryngeal Mask Airways

Transcription:

doi:10.1111/j.1365-2044.2008.05757.x APPARATUS Optimising tracheal intubation success rate using the Airtraq laryngoscope G. Dhonneur, 1 W. Abdi, 2 R. Amathieu, 2 S. Ndoko 2 and L. Tual 2 1 Professor, 2 Staff, Department of Anaesthesia and Intensive Care Medicine, Jean Verdier University Hospital of Paris (APHP), 93143 Bondy, France Summary In this study we have used a video-recording, retrospective analysis technique to evaluate the influence of the Airtraq TM laryngoscope manipulations and the resulting changes in position of the glottic opening and inter-arytenoids cleft, on the success rate of tracheal intubation. The video recordings of the internal views of 109 tracheal intubation attempts, in 50 anaesthetised patients were analysed. We demonstrated that successful tracheal intubation using the Airtraq laryngoscope require the glottic opening to be centred in the view, and positioning the inter-arytenoid cleft medially below the horizontal line in the centre of the view. We also demonstrated that repositioning of the Airtraq laryngoscope following a failed tracheal intubation attempt required the performance of a standard series of manoeuvres.... Correspondence to: Prof. Gilles Dhonneur E-mail: gilles.dhonneur@jvr.aphp.fr Accepted: 27 September 2008 The Airtraq TM laryngoscope (Fannin (UK) Ltd, Calcot, Reading, UK) is a novel, single use, optical laryngoscope which has been shown to improve the ease of intubation in patients with normal and difficult airways [1 3]. Although, the Airtraq laryngoscope produces a reduction in the time required for tracheal intubation in most patients with difficult airways, tracheal intubation on the first attempt, may not always be successful [4, 5]. In obese patients a clear view of the glottic opening can rapidly be obtained following insertion. However, tracheal intubation, may fail, requiring repositioning of the Airtraq laryngoscope in the pharynx prior to a further intubation attempt subsequently being successful. The manipulations frequently required are a lowering of the position of the glottis within the view. To investigate this, we analysed the influence of the position of the glottic opening and the inter-arytenoid cleft position in the laryngeal view, on the success rates of tracheal intubation, and described the repositioning of the Airtraq laryngoscope required to succeed, following a failed tracheal intubation attempt. Methods The Ethics Review Board approved this trial and written informed consent was obtained from each patient for filming and recording of the airway management technique used. The study was based on a retrospective analysis of videos recorded in the operating theatre during the airway management of elective patients using the Airtraq laryngoscope. The videos were of internal views of the larynx and external recordings. All airway management techniques were performed in anaesthetised patients, who had received neuromuscular blockade, by senior anaesthetists providing anaesthesia for patients in the morbid obesity and gynaecological units. External films of tracheal intubation were performed by an assistant using a standard video-camera and internal views were automatically recorded using the videocapture system (Vygon, Écouen, France). All recorded films were converted to a similar 20 images per second format and transferred to a computer for image analysis. External and internal video recordings were synchronised. We then analysed the internal recordings of 50 patients tracheal intubations requiring greater than one tracheal intubation attempt. For each video and for each tracheal intubation attempt, we selected the image recorded immediately prior to tracheal tube advancement towards the glottic opening. On each image, the outline of the glottic opening and the exact Journal compilation Ó 2009 The Association of Anaesthetists of Great Britain and Ireland 315

G. Dhonneur et al. Æ Optimising tracheal intubation success rate using the Airtraq laryngoscope Anaesthesia, 2009, 64, pages 315 319 Figure 1 On each image (a) immediately prior to a tracheal intubation attempt, the outlines of the glottic opening and the position of the inter-arytenoids cleft immediately adjacent to posterior glottis were marked (b). position of the inter-arytenoid cleft immediately adjacent to the posterior glottis were traced and saved (Fig. 1). All traces of successful tracheal intubation and of failed intubation attempts were superimposed to identify the position of the glottic opening and interarytenoid cleft in tracheal intubation success and failure respectively. The failed attempts zone (FZ) and the target zone (TZ), were defined as the largest rectangular area, including all glottic opening and inter-arytenoid cleft positions, associated with success or failure of tracheal intubation. External videos of all the tracheal intubation attempts were then reviewed. Manipulations applied to the Airtraq laryngoscope by the anaesthetist resulting in further success of tracheal attempts were analysed. Results The demographic data and tracheal intubation characteristics of 50 obese patients are presented (Table 1). A total Table 1 Demographic data and tracheal intubation characteristics of the patients. Demographic data Age, years 38 (11) Gender: male female, (number of patients) 22 28 Body mass index, kg m )2 41 (9) Mallampati score, 1 4 (number of patients) 1 16, 2 19, 3 15, 4 0 Thyromental distance, mm 65 (7) Mouth aperture, mm 33 (4) Tracheal intubation characteristics Duration, s 39 (17) Percentage of glottic opening visible, % 100 Three tracheal attempts, (number of patients) 5 Four tracheal attempts, (number of patients) 4 of 109 recordings were used for the analysis, this included 59 failures and 50 successes. The failed attempts zone (FZ) and the target zone (TZ) of the glottic opening overlapped. FZ and TZ of the inter-arytenoid cleft did not overlap (Fig. 2). Two efficient techniques were found to be associated with lowering the position of the glottic opening and the inter-arytenoid cleft. The first manoeuvre performed consisted of three movements of the Airtraq laryngoscope tip: downward, backward and upward movements (Fig. 3). Reducing cervical spine extension with in-line head and neck alignment also resulted in lowering the position of the glottic opening and inter-arytenoid cleft in seven non-obese patients. Discussion This study has demonstrated that the position of both the glottic opening and the inter-arytenoid cleft affected the success rate of tracheal intubation using the Airtraq laryngoscope. Most successful tracheal intubation attempts occurred when the glottic opening was centrally placed within the view, and the inter-arytenoid cleft was medially located below the horizontal line in the centre of the view. We have also shown that correcting the position of the Airtraq laryngoscope following a failed tracheal intubation attempt, required the anaesthetist to perform a series of standardised manoeuvres. Improving the factors associated with successful tracheal intubation when using the Airtraq laryngoscope is important for improving the learning process, it is also relevant for daily clinical practice. We found that anaesthetists learning to use the Airtraq laryngoscope, found two manoeuvres were more difficult to master in patients with potentially difficult airways, (i.e. morbidly obese patients). Firstly, placing the Airtraq laryngoscope in the pharynx required significant distal pressure on the tip of the blade to overcome the narrowing at the junction between the oral and pharyngeal spaces. We developed an alternative rotational manoeuvre during pharyngeal insertion of the blade [6] to overcome this problem. Secondly, although most physicians were able to obtain a view of the glottic opening in all patients (including anticipated difficult airway patients), we found that such optimal laryngeal exposure was not always associated with successful tracheal intubation. The manipulations required when the glottic opening was not centred in the view were relatively simple to perform. Most anaesthetists intuitively performed small amplitude rotating movements (in the horizontal plane) to the proximal section of the Airtraq laryngoscope to align the axes of the tube emerging from the lateral channel with that of the glottis. 316 Journal compilation Ó 2009 The Association of Anaesthetists of Great Britain and Ireland

G. Dhonneur et al. Æ Optimising tracheal intubation success rate using the Airtraq laryngoscope Figure 2 (a d) Represent schematic illustrations of the marks superimposed for glottic opening (a, b) and for the inter-arytenoid cleft (c, d) positions, just prior to failed (a, c: n = 59) and successful (b, d: n = 50) tracheal intubation attempts. The failed zone (FZ) and the target zone (TZ) were defined as the largest rectangular area including all glottic opening (GLOO) and interarytenoid cleft (IARYC) positions just prior to failed (a, c: n = 59) and successful (b, d: n = 50) tracheal intubation attempts, respectively. TZ-IARYC TZ-GLOO. Figure 3 Schematic illustration of the triple manoeuvre: down (1)-back (2)-up (3, 4) applied to the tip of the Airtraq laryngoscope blade. We found that failures of tracheal intubation were almost always associated with an abnormally high position of the glottic opening in the laryngoscopic view, suggesting that the distal tip of the blade was not optimally inserted into the pharynx. However, as we have demonstrated, manipulations to lower the vertical position of the glottic opening were not routinely used, unless a failure to intubate the trachea occurred. The reasons for attempting tracheal intubation with the glottic opening in a high position in the view will be discussed. Firstly, we found that tracheal intubation was sometimes possible if the tip of the blade of the Airtraq Journal compilation Ó 2009 The Association of Anaesthetists of Great Britain and Ireland 317

G. Dhonneur et al. Æ Optimising tracheal intubation success rate using the Airtraq laryngoscope Anaesthesia, 2009, 64, pages 315 319 laryngoscope was positioned below the epiglottis. Indeed, the second attempt at tracheal intubation succeeded in three patients even though the tip of the Airtraq laryngoscope was positioned below the epiglottis and the glottic opening remained relatively highly positioned. In contrast, similar features resulted in a failure of the second tracheal intubation attempt in four other patients requiring repositioning of the distal tip of the Airtraq laryngoscope blade in the vallecula before the third tracheal intubation attempt was ultimately successful. In all of the other cases, failure of the first attempt occurred while the epiglottis was visible in the initial view suggesting that the tip of the Airtraq laryngoscope was correctly positioned in the vallecula. For these patients, another parameter, the distance between the glottic opening and the exit of the Airtraq laryngoscope lateral channel, probably affected the success rate of tracheal intubation attempts. The channel design of the Airtraq favours the tracheal tube initially directed downwards [7] before it ascends. A minimal distance between the exit of the Airtraq laryngoscope lateral channel and the glottic opening seems to be important to ensure that the tracheal tube passes over the arytenoids. In the case of abnormal anatomical features or non optimal Airtraq laryngoscope distal tip position, this distance might be too short resulting in failure of the tracheal intubation attempt. Interestingly, the downback-up manoeuvre and the reduction of cervical spine extension invariably moved the glottic opening and lowered its position in the laryngoscopic view. Although we did not make measurements on the traces, we believe that these manipulations of the Airtraq laryngoscope and of the patient s neck [8, 9] probably increased the distance between the exit of the Airtraq laryngoscope lateral channel and the glottic opening due to the effect of gravity on the larynx. Of interest in some morbidly obese patients, we could not relate changes in head and neck position to tracheal intubation success rate, but rather to the effect of the down-back-up manoeuvre [10]. By using the down-back-up manoeuvre and the alignment of the head and neck, the glottic opening and the inter-arytenoid cleft position changed from the FZ to the TZ. Although the glottic opening and the inter-arytenoid cleft positions changed significantly following airway manipulations, the position of both anatomical landmarks have different predictive values on the likelihood of success of tracheal intubation. As seen in Fig. 2, the FZ and the TZ of the glottic opening overlapped vertically, suggesting a poor predictability for success of tracheal intubation. Conversely, we have demonstrated that the FZ and the TZ of the inter-arytenoid cleft position associated with failure and success of tracheal intubation attempts did not overlap indicating that the inter-arytenoid cleft position is a predictor of tracheal intubation success using the Airtraq laryngoscope. Indeed, when the glottic opening was centred in the view in association with the inter-arytenoids cleft being positioned below the horizontal midline, all tracheal intubation attempts were successful. In conclusion, we have demonstrated that the interarytenoid cleft position is the main factors influencing the likely success of tracheal intubation using the Airtraq laryngoscope. Acknowledgement Prodol LtD Company provided the Airtraq laryngoscopes used in this study. Financial support This study was funded by Departmental sources. Conflict of interest The authors have no conflict of interest with the Vygon and Prodol Ltd companies. References 1 Maharaj CH, Costello JF, Harte BH, Laffey JG. Evaluation of the Airtraq and Macintosh laryngoscopes in patients at increased risk for difficult tracheal intubation. Anaesthesia 2008; 63: 182 8. 2 Woollard M, Mannion W, Lighton D, et al. Use of the Airtraq laryngoscope in a model of difficult intubation by prehospital providers not previously trained in laryngoscopy. Anaesthesia 2007; 62: 1061 5. 3 Tong JL. The management of difficult direct laryngoscopy and intubation in a field hospital: an alternative to fibreoptic endoscopy. Journal of the Royal Army Medical Corps 2007; 153: 181 3. 4 Woollard M, Lighton D, Mannion W, et al. Airtraq vs standard laryngoscopy by student paramedics and experienced prehospital laryngoscopists managing a model of difficult intubation. Anaesthesia 2008; 63: 26 31. 5 Maharaj CH, Higgins BD, Harte BH, Laffey JG. Evaluation of intubation using the Airtraq or Macintosh laryngoscope by anaesthetists in easy and simulated difficult laryngoscopy a manikin study. Anaesthesia 2006; 61: 469 77. 6 Dhonneur G, Ndoko SK, Amathieu R, et al. A comparison of two techniques for inserting the Airtraq laryngoscope in morbidly obese patients. Anaesthesia 2007; 62: 774 7. 318 Journal compilation Ó 2009 The Association of Anaesthetists of Great Britain and Ireland

G. Dhonneur et al. Æ Optimising tracheal intubation success rate using the Airtraq laryngoscope 7 Suzuki A, Abe N, Sasakawa T, Kunisawa T, Takahata O, Iwasaki H. Pentax-AWS (Airway Scope) and Airtraq: big difference between two similar devices. Journal of Anesthesia 2008; 22: 191 2. 8 Hirabayashi Y, Fujita A, Seo N, Sugimoto H. A comparison of cervical spine movement during laryngoscopy using the Airtraq or Macintosh laryngoscopes. Anaesthesia 2008; 63: 635 40. 9 Hirabayashi Y, Seo N. In-line head and neck position is preferable for tracheal intubation with the Airtraq laryngoscope compared to the sniffing position. Journal of Anesthesia 2008; 22: 189 90. 10 Ndoko SK, Amathieu R, Tual L, et al. Tracheal intubation of morbidly obese patients: a randomized trial comparing performance of Macintosh and Airtraq laryngoscopes. British Journal of Anaesthesia 2008; 100: 263 8. Journal compilation Ó 2009 The Association of Anaesthetists of Great Britain and Ireland 319