Reusable Instrument Lifespan Manual

Similar documents
Reusable Instrument Lifespan Manual

3.0 mm Cannulated Screw System. Surgical Technique

FastFrame External Fixation System. Damage Control Surgical Technique

Tibial Fixation. with TunneLoc Device. Surgical Technique by Mark J. Albritton, M.D. and Sherwin Ho, M.D.

Zimmer Biomet RF Electrosurgical Instruments. System Brochure

CrossFix II. All-Suture, All-Inside Meniscal Repair System. Surgical Technique

ComposiTCP Anchor with BroadBand Tape

SureLock All-Suture Anchor System

CoAxial Spray Kit. Simplifying the Application of Platelet-Rich Plasma onto Graft Material. Surgical Technique

The AperFix II System

Flexible Fragment Fixation. Surgical Technique

StageOne Knee Cement Spacer Molds. Surgical Technique

Arcos Modular Femoral Revision System. Broach and Calcar Proximal Bodies Surgical Technique

DEVELOPED BY SURGEONS

JuggerLoc Bone-to-Bone System for Ankle Syndesmosis Fixation. Surgical Technique

THE PERSONALIZED KNEE

Zimmer Biomet US Cement and Infection Product Portfolio

Reprocessing Guide. Autoclavable Arthroscope and Hardware Sterilization Tray

RibFix Blu. Thoracic Fixation System

BTB ACL Reconstruction with the ToggleLoc Fixation Device with ZipLoop Technology. Surgical Technique by James R. Andrews, M.D.

Avenir Femoral Hip System

NexGen Trabecular Metal Tibial Tray. Surgical Technique

Gender Solutions Patello-Femoral Joint System

Arcos Interlocking Distal Stem. Surgical Technique Addendum to the Arcos Modular Femoral Revision System

Surgical Technique Guide

Signature Personalized Patient Care

OSS Modular Arthrodesis System. Assembly Guide

Hip Arthroscopy. All Inclusive Brochure

Medial Portal ACL Reconstruction

Correction System. Surgical Technique

Biomet Multi-Axial Correction (MAC) System Correction Atlas

Basic Trocar System. Basic transbuccal instrumentation system for simplified intraoral plate and screw application.

HIP ARTHROSCOPY BROCHURE

N-Force Fixation System. Surgical Technique

Correction System. Surgical Technique

Technique Guide Small Bone Fusion System

LBL-014. Rev

M6-C Artificial Cervical Disc Surgical Instruments

Pinit Plate Small Bone Fusion System Bone Plate & Screw System

INnate. Surgical Technique Guide. Anatomic Fixation in Your Hands. The INnate Implant

Ease of Use. Have the instrument you need when you need it.

Pediatric LCP Plate System. For osteotomies and fracture fixation of the proximal and distal femur.

Technique Manual Technique Manual Rev. A 01/11/2011

Persona. The Personalized Knee. Trabecular Metal Tibia. Surgical Technique

PediLoc 3.5mm and 4.5mm Contour Femur Plate Surgical Technique

Technique Guide. Epoca Revision Instrument Set. For extraction and revision of Epoca shoulder implants.

OSS Orthopedic Salvage System Compress Device. Surgical Technique

Sidus Stem-Free Shoulder. Secure fixation through a bone-sparing design

SI-BONE ifuse Implant System Instruments Hospital Cleaning and Sterilization Instructions 0344

fassier DUVal TelescoPIc IM system s RESCUE SYSTEM 2.0

RAPIDSORB RAPID ReSORBABLe CRANIAL CLAMP

Comprehensive Reverse Shoulder System Augmented Baseplate

Taperloc Complete Hip System. Surgical Technique

Large External Fixator Modular Knee Bridge. Using multi-pin clamps.

Personalized Solutions. Portfolio Brochure

Peanut Growth Control Plating System

JuggerKnot Soft Anchor 1.4/1.5 mm with Percutaneous Instrumentation for Low Profile/Trans-Cuff SLAP Repair

PediLoc 3.5mm and 4.5mm Bowed Femur Plate Surgical Technique

Hip Resurfacing System

PROXIMAL FEMORAL NAIL REMOVAL SET

HP KNEE EXTRACTION Instrumentation. Product Overview

InCoreTM Lapidus System

MatrixNEURO. The next generation cranial plating system.

BIOLOX delta Option Ceramic Femoral Head System

FASSIER DUVAL TELESCOPIC IM SYSTEM S RESCUE SYSTEM 2.0

Fitness Circle. Includes 15 Exercises:

MIS Cemented Tibial Component

INSTRUCTIONS FOR USE AND CARE

Femoral Fixation for ACL Reconstruction. Surgical Protocol by Mark Gittins, D.O.

Zimmer PSI Shoulder Trabecular Metal Reverse Glenoid Base Plate. Surgical Technique

Collinear Reduction Clamp

3.5 mm LCP Tibial Plateau Sets. Configuration options.

Lag Screw Device Intended for symphyseal fracture fixation of the mandible

Lapidus Arthrodesis System Instructions for Use

Visit Linvatec.com today to learn more. Surgical Technique: Sequential Meniscal Running Stitch

StageOne. Shoulder Cement Spacer Molds. Surgical Technique

Sterile-Packaged Large External Fixator Kits. For treatment of long bone and pelvic fractures that require external fixation.

MANUAL FOR USE NO. 5 FLEXIBLE INTRAMEDULLARY REAMER REV ( GB ) non - sterile

Olecranon Locking Plate II

Technique Guide Hip Resurfacing System

JuggerKnot and JuggerKnotless Soft Anchor. All-Inclusive Brochure

Aesculap Implant Systems

Cannulated Angled Blade Plate 3.5 and 4.5, 90.

Cervical Solutions. Trinnect. Hydrated Anterior Cervical Spacer System. Surgical Technique Guide

Vanguard. Total Knee System

AS Medizintechnik GmbH Sattlerstraße Tuttlingen Germany

EXACTECH KNEE. Operative Technique ADDENDUM. Metaphyseal Cones. Surgeon focused. Patient driven. TM

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

Constrained Posterior Stabilized (CPS)

ULTR A SERIES G7-S75

Technique Guide. Locking Attachment Plate. For treatment of periprosthetic fractures.

Smarter Thinking. Simpler Design. Prima Plus. Surgical Manual

a perfect fit INSTRUCTION MANUAL CAMLOG TORQUE WRENCH WITH CONTINUOUS TORQUE ADJUSTMENT Handling Sterilization Care

Implant And Abutment Level Impressions

14 mm x +30 mm Stem Extension

Plant, components and devices for very high and high pressures (7000 bar and over)

Surgical Technique. Proximal Humerus Locking Plate

Button Plate. Reinforcement for transosseous fixations.

Positioning Sleeve Surgical Technique

Transcription:

Reusable Instrument Lifespan Manual

Table of Contents Introduction... 3 Bending... 7 Discoloration... 13 Corrosion... 16 Fracture... 20 Thread Damage... 24 Surface Damage... 27

INTRODUCTION 3

Purpose This manual is intended to assist the user in determining whether a reusable instrument has worn to an extent that it is no longer suitable for use. For additional reusable instrument care instructions, see the Zimmer Instrument Care, Cleaning, Maintenance and Sterilization Instructions (Doc No. 97-5000-170-00). INTRODUCTION Scope This manual provides information applicable to the reusable instruments used to implant hip, knee, shoulder, elbow, ankle, and trauma implants. This information is NOT APPLICABLE to single use instruments (sterile and non-sterile). Glossary Indicator A term used to describe damage to a device that can indicate the instrument is no longer suitable for use. Feedback Refers to any specific visual, auditory, or tactile feedback that serves as an indicator of decision/action. Quick Check An assessment of the instrument s condition to detect the presence of wear. Reprocessing The necessary validated processes including cleaning, disinfection, and sterilization to render a medical device, which has been previously used or contaminated, fit for its intended subsequent single use. Understanding Document Structure This manual describes several types of wear including: i. Bending ii. Discoloration iii. Corrosion iv. Fracture v. Thread Damage vi. Surface Damage 4

Understanding Document Structure (cont.) The indicators identified above are signs of wear and damage on reusable instruments. The manual is separated by the aforementioned indicators. Each indicator section contains: INTRODUCTION Images representing the indicator category to which it belongs. Images are only a representation of possible wear. s of the indicator shown in the image and content specific to the indicator. Quick Check Methods to assist in assessing the instrument for the presence of said indicator. on any reusable instrument specific to the indicator described, not particularly those shown in the images provided, to aid in assessing if the instrument is suitable for use. Symbol legend ü Suitable for Use Not Suitable for Use Inspection/Function Testing While loading instruments into their respective instrument cases after cleaning and prior to sterilization, reference the manual and follow the instructions below. 1. Instruments should be inspected for completeness and function. 2. Inspection includes: a. Checking instruments that form part of a larger assembly or assemble with mating components. b. Checking internal mechanisms such as o-rings, springs, and subcomponents, if the device is intended to be disassembled for proper reprocessing. c. Actuating moving parts such as hinges/joints and moveable features such as handles, ratcheting, couplings, and sliding parts. d. Inspecting for all forms of wear outlined in this manual. 3. Results of assembly, actuation, and extent of all forms of wear should be considered in determining whether an instrument is suitable for use. 4. If the reusable instrument is determined no longer suitable for use or if the suitability for use is still in question after inspecting the instrument and referencing the Reusable Instrument Lifespan Manual, initiate the process to return the instrument(s) to the manufacturer. 5

Returning the Reusable Instrument(s) If an allegation is made related to the identity, quality, durability, reliability, safety, effectiveness, or performance of a device at any other time, please complete and submit a Product Experience Report (CF04001) and return the reusable instrument(s) to the manufacturer for investigation. INTRODUCTION If the instrument to be returned is not a complaint, complete and submit a Worn Instrument Return Form (F-01413 D). The Worn Instrument Return Program is not a replacement for the Complaints Submission Process. Please do not attempt to dispose of the reusable instrument(s) through any other means. 6

BENDING 7

BENDING ü Bent Kinked Crooked Unraveled coils Roll Quick Check Action: Instrument is rolled back and forth on flat surface. Result: If instrument wobbles, wear is present and the instrument must be returned to the manufacturer. Inclined Quick Check Action: Instrument is rolled down an incline. Result: If instrument does not roll freely, wear is present and the instrument must be returned to the manufacturer. Inadequate force applied Inadequate engagement of components Inadequate disengagement of components Inadequate reprocessing Contact unintended soft tissue or bone Inaccurate feedback Inadequate removal of bone, soft tissue, or bone cement to a specific/non-specific shape Inadequate application/removal of bone cement Free End Lift Quick Check Action: Allow instrument to lay flat on a surface. Result: If deflection is observed at the free end, wear is present and the instrument must be returned to the manufacturer. 8

BENDING Bowed Curved Warped Unraveled coils Roll Quick Check Action: Instrument is rolled back and forth on flat surface. Result: If instrument wobbles, wear is present and the instrument must be returned to the manufacturer. Inclined Quick Check Action: Instrument is rolled down an incline. Result: If instrument does not roll freely, wear is present and the instrument must be returned to the manufacturer. Inadequate force applied Inadequate engagement of components Inadequate disengagement of components Inadequate reprocessing Contact unintended soft tissue or bone Inaccurate feedback Inadequate removal of bone, soft tissue, or bone cement to a specific/non-specific shape Inadequate application/removal of bone cement Free End Lift Quick Check Action: Allow instrument to lay flat on a surface. Result: If deflection is observed at the free end, wear is present and the instrument must be returned to the manufacturer. Camber Quick Check Action: Place instrument on a flat surface with either end on a straight edge. Result: If there is deflection in the width direction, wear is present and the instrument must be returned to the manufacturer. 9

BENDING ü Curled cutting edges Inadequate force applied Inadequate engagement of components Inadequate disengagement of components Inadequate reprocessing Contact unintended soft tissue or bone Inaccurate feedback Inadequate removal of bone, soft tissue, or bone cement to a specific/non-specific shape Inadequate application/removal of bone cement 10

BENDING Bent light pipe or cable Light Quick Check Action: Hold end of light pipe or cable against a light source and look into pipe or cable at the other end. Result: If light will no longer travel through the pipe or cable, wear is present and the instrument must be returned to the manufacturer. Inadequate light for guidance Inadequate engagement of components Inadequate disengagement of components 11

ü BENDING Bent, bowed or warped: Cases Bases Trays Lids Brackets or posts Insufficient function of handles, lids, and rotating components Can allow free travel of contents Interference between case components and contents Inadequate reprocessing Inadequate engagement of components Inadequate disengagement of components 12

DISCOLORATION 13

DISCOLORATION Discolored product markings Readability Quick Check Action: Read instrument markings. Result: If instrument markings are illegible or difficult to read, wear is present and the instrument must be returned to the manufacturer. Illegible product markings Inaccurate feedback 14

DISCOLORATION Discolored product markings due to corrosion Pencil Eraser Quick Check Action: Rub a pencil eraser on the affected area. If the discoloration is removed by the eraser, it is a stain. Result: If the discoloration is not removed by the eraser or reveals pitting, wear is present and the instrument must be returned to the manufacturer. Illegible product markings Inaccurate feedback Readability Quick Check Action: Read instrument markings. Result: If instrument markings are illegible or difficult to read, wear is present and the instrument must be returned to the manufacturer. 15

CORROSION 16

CORROSION ü Surface corrosion Pitting Pencil Eraser Quick Check Action: Rub a pencil eraser on the affected area. If the discoloration is removed by the eraser, it is a stain. Result: If the discoloration is not removed by the eraser or reveals pitting, wear is present and the instrument must be returned to the manufacturer. Inaccurate feedback Inadequate reprocessing Inadequate removal of bone, soft tissue, or bone cement to a specific/non-specific shape Inadequate engagement of components Inadequate disengagement of components 17

CORROSION ü Crevice corrosion External/Internal thread corrosion Threading Quick Check Action: Thread the components. Result: If there is difficulty fastening or unfastening components, wear is present and the instrument must be returned to the manufacturer. Pencil Eraser Quick Check Action: Rub a pencil eraser on the affected area. If the discoloration is removed by the eraser, it is a stain. Result: If the discoloration is not removed by the eraser or reveals pitting, wear is present and the instrument must be returned to the manufacturer. Inaccurate feedback Inadequate reprocessing Inadequate engagement of components Inadequate disengagement of components Inadequate removal of bone, soft tissue, or bone cement to a specific/nonspecific shape Excessive torque to engage threads Excessive torque to disengage threads 18

CORROSION Discolored product markings due to corrosion Pencil Eraser Quick Check Action: Rub a pencil eraser on the affected area. Result: If the discoloration is removed by the eraser, it is a stain. If the discoloration is not removed by the eraser or reveals pitting, wear is present and the instrument must be returned to the manufacturer. Illegible product markings Inaccurate feedback Readability Quick Check Action: Read instrument markings. Result: If instrument markings are illegible or difficult to read, wear is present and the instrument must be returned to the manufacturer. 19

FRACTURE 20

FRACTURE Fractured Inadequate reprocessing Inadequate accommodation of hand use Inadequate force applied Inaccurate feedback Inadequate engagement of components Inadequate disengagement of components 21

FRACTURE ü Cracked Inadequate reprocessing Inadequate accommodation of hand use Inadequate force applied Fracture Inaccurate feedback Inadequate engagement of components Inadequate disengagement of components 22

FRACTURE Cracked/Fractured in flexible shaft body Inadequate reprocessing Inadequate accommodation of hand use Inadequate force applied Fracture Inaccurate feedback Inadequate engagement of components Inadequate disengagement of components 23

THREAD DAMAGE 24

THREAD DAMAGE ü Stripped external/internal threads Damaged external/internal threads Threading Quick Check Action: Thread the components. Result: If there is difficulty fastening or unfastening components, wear is present and the instrument must be returned to the manufacturer. Excessive torque to engage threads Excessive torque to disengage threads Inaccurate feedback Inadequate engagement of components Inadequate disengagement of components 25

THREAD DAMAGE ü Crevice corrosion External/Internal thread corrosion Threading Quick Check Action: Thread the components. Result: If there is difficulty fastening or unfastening components, wear is present and the instrument must be returned to the manufacturer. Pencil Eraser Quick Check Action: Rub a pencil eraser on the affected area. If the discoloration is removed by the eraser, it is a stain. Result: If the discoloration is not removed by the eraser or reveals pitting, wear is present and the instrument must be returned to the manufacturer. Inaccurate feedback Inadequate reprocessing Inadequate engagement of components Inadequate disengagement of components Inadequate removal of bone, soft tissue, or bone cement to a specific/nonspecific shape Excessive torque to engage threads Excessive torque to disengage threads 26

SURFACE DAMAGE 27

SURFACE DAMAGE Scratched Damaged surface Fingernail Quick Check Action: Slide fingernail on the surface. Result: If the nail is obstructed/resisted, wear is present and the instrument must be returned to the manufacturer. Inaccurate feedback Inadequate removal of bone, soft tissue, or bone cement to a specific/non-specific shape Contact unintended soft tissue or bone Inadequate accommodation of hand use Inadequate engagement of components Inadequate disengagement of components Inadequate reprocessing Inadequate force applied 28

SURFACE DAMAGE Nicked Dented Chipped Fingernail Quick Check Action: Slide fingernail on the surface. Result: If the nail is obstructed/resisted, wear is present and the instrument must be returned to the manufacturer. Inaccurate feedback Inadequate removal of bone, soft tissue, or bone cement to a specific/non-specific shape Contact unintended soft tissue or bone Inadequate accommodation of hand use Inadequate engagement of components Inadequate disengagement of components Inadequate reprocessing Inadequate force applied 29

SURFACE DAMAGE Burr Fingernail Quick Check Action: Slide fingernail on the surface. Result: If the nail is obstructed/resisted, wear is present and the instrument must be returned to the manufacturer. Inadequate removal of bone, soft tissue, or bone cement to a specific/non-specific shape Contact unintended soft tissue or bone Inadequate accommodation of hand use Inadequate engagement of components Inadequate disengagement of components Inadequate reprocessing 30

SURFACE DAMAGE Gouged Scratched Dented Fingernail Quick Check Action: Slide fingernail on the surface. Result: If the nail is obstructed/resisted, wear is present and the instrument must be returned to the manufacturer. Inaccurate feedback Inadequate accommodation of hand use Inadequate engagement of components Inadequate disengagement of components Inadequate reprocessing 31

SURFACE DAMAGE Loss of coating Inaccurate feedback Inadequate removal of bone, soft tissue, or bone cement to a specific/non-specific shape Inadequate reprocessing Inadequate force applied 32

SURFACE DAMAGE Stripped drivers Rounded edges Inaccurate feedback Inadequate engagement of components Inadequate disengagement of components Inadequate reprocessing Inadequate force applied 33

SURFACE DAMAGE Damaged cutting edges Inadequate removal of soft tissue, bone, or bone cement to a specific/non-specific shape Excessive force to remove soft tissue, bone, or bone cement Inadequate accommodation of soft tissue or bone 34

SURFACE DAMAGE Damaged cutting edges Inadequate removal of soft tissue, bone, or bone cement to a specific/non-specific shape Excessive force to remove soft tissue, bone, or bone cement Inadequate accommodation of soft tissue or bone 35

All content herein is protected by copyright, trademarks and other intellectual property rights, as applicable, owned by or licensed to Zimmer Biomet or its affiliates unless otherwise indicated, and must not be redistributed, duplicated or disclosed, in whole or in part, without the express written consent of Zimmer Biomet. This material is intended for health care professionals, the Zimmer Biomet sales force and Zimmer Biomet employees. Distribution to any other recipient is prohibited. Zimmer Biomet does not practice medicine. The information being presented is of a general nature and does not represent or constitute medical advice or recommendations and is for medical education purposes only. 2017 Zimmer Biomet Authorized Representative Zimmer U.K. Ltd. 9 Lancaster Place South Marston Park Swindon, SN3 4FP UK 1219.1-GLBL-en-REV0417 Legal Manufacturer Zimmer, Inc. 1800 West Center Street Warsaw, Indiana 46580 United States www.zimmerbiomet.com 0086 CE mark is not valid unless there is a CE mark on the product label.