Anatomic factors associated with acute endograft collapse after Gore TAG treatment of thoracic aortic dissection or traumatic rupture

Similar documents
Anatomic Factors Associated with Acute Endograft Collapse after Gore TAG Treatment of Thoracic Aortic Dissection or Traumatic Rupture

Dynamic Cine-CT Angiography for the Evaluation of the Thoracic Aorta; Insight in Dynamic Changes with Implications for Thoracic Endograft Treatment

Influence of Oversizing on Outcome in Thoracic Endovascular Aortic Repair

Challenges with Complex Anatomies Advancing Care in Endovascular Aortic Treatment

Thoracic aortic trauma A.T.O.ABDOOL-CARRIM ACADEMIC HEAD VASCULAR SURGERY DEPARTMENT OF SURGERY UNIVERSITY OF WITWATERSRAND

THE THE MORE MORE NATURAL APPROACH TO OPTIMAL FIT

Indications for use. Contraindications within the United States

Introducing the GORE TAG Conformable Thoracic Stent Graft with ACTIVE CONTROL System

Observations on the Failure of Stent-Grafts in the Aortic Arch.

Endoanchor-assisted TEVAR

Percutaneous Approaches to Aortic Disease in 2018

THE ENDURANT STENT GRAFT IN HOSTILE ANEURYSM NECK ANATOMY

Acute dissections of the descending thoracic aorta (Debakey

Optimal Treatment of Chronic Dissection

Abdominal and thoracic aneurysm repair

Anatomical challenges in EVAR

Increased Flexibility of AneuRx Stent-Graft Reduces Need for Secondary Intervention Following Endovascular Aneurysm Repair

Endovascular Management of Thoracic Aortic Pathology Stéphan Haulon, J Sobocinski, B Maurel, T Martin-Gonzalez, R Spear, A Hertault, R Azzaoui

TriVascular Ovation Prime Abdominal Stent Graft System

EVAR replaced standard repair in most cases. Why?

Subject: Endovascular Stent Grafts for Disorders of the Thoracic Aorta

Access More Patients. Customize Each Seal.

DIFFICULT ACCESS REMAINS A CONTRAINDICATION FOR EVAR APOSTOLOS K. TASSIOPOULOS, MD, FACS PROFESSOR AND CHIEF DIVISION OF VASCULAR SURGERY

I SECTION I. Thoracic aortic aneurysms

Technique and Tips for Complicated AAA Cases with Stent Graft

Performance of the conformable GORE TAG device in Type B aortic dissection from the GORE GREAT real world registry

Chapter 18. Recommended clinical competencies for initiating a program in endovascular repair of the thoracic aorta

Development of Stent Graft. Kato et al. Development of an expandable intra-aortic prothesis for experimental aortic dissection.

When to use standard EVAR with EndoAnchors or CHEVAR in short-neck AAAs LINC ASIA 18

Degeneration of the Neck Post Implementation - a New Era of AAA Stent

Talent Abdominal Stent Graft

Current State of Thoracic Branch Devices and Ongoing Clinical Trials

How to achieve a successful proximal sealing in TEVAR? Pr L Canaud

How to Categorize the Infrarenal Neck Properly? I Van Herzeele Dept. Thoracic and Vascular Surgery, Ghent University, Belgium

Considerations for a Durable Repair

Hostile Neck During EVAR, The Role Of Endoanchores

Secondary interventions after elective thoracic endovascular aortic repair for degenerative aneurysms

Ascending Aorta: The Endovascular Approach

Santi Trimarchi, MD, PhD Vascular Surgeon Thoracic Aortic Research Center, Director IRCCS Policlinico San Donato University of Milan

Ascending Aorta: Is The Endovascular Approach Realistic?

Role of Gender in TEVAR and EVAR results from the GREAT registry

Aortic stents, types, selection, tricks in deployment.

European Experience with a New Thoracic Device. D.Böckler University Hospital Heidelberg Germany

I-Hui Wu, M.D. Ph.D. Clinical Assistant Professor Cardiovascular Surgical Department National Taiwan University Hospital

Tips and techniques for optimal stent graft placement in angulated aneurysm necks

Thoracic Endograft Explant Analysis

GORE TAG Thoracic Endoprosthesis ANNUAL CLINICAL UPDATE SEPTEMBER 2018 Abstract. Section I GORE TAG Device Clinical experience. Section II Conformable

Endovascular aortic stent grafts have forever

Tips and Tricks to Deliver a Stengraft to the Ascending Aorta

Fig 1. Calculation of angulation. A, Three-dimensional reconstruction image. B, Illustration of vector inner product

History of the Powerlink System Design and Clinical Results. Edward B. Diethrich Arizona Heart Hospital Phoenix, AZ

Challenging anatomies demand versatility.

GORE EXCLUDER AAA Endoprosthesis demonstrates long-term durability. Michel Reijnen Rijnstate Hospital Arnhem, The Netherlands

Endologix PowerWeb System EPW?

Type 1a Endoleak in hostile neck anatomies: Endoanchor can fix it! D. Böckler University Hospital Heidelberg, Germany

Patient selection in Hostile Necks and how. to prevent endoleaks a word of caution

Right Choice for Right Angles

How to select FEVAR versus EVAR + endoanchors in short-necked AAAs

Intravascular Ultrasound in the Treatment of Complex Aortic Pathologies. Naixin Kang, M.D. Vascular Surgery Fellow April 26 th, 2018

Bifurcated system Proximal suprarenal stent Modular (aortic main body and two iliac legs) Full thickness woven polyester graft material Fully

Improving Endograft Durability with EndoAnchors

TEVAR FOR! THORACIC AORTIC TRAUMA"

Eccentric stent graft compression: An indicator of insecure proximal fixation of aortic stent graft

From 1996 to 1999, a total of 1,193 patients with

Durable outcomes. Proven performance.

Influence of patient selection and IFU compliance on outcomes following EVAS

Zenith Renu AAA Converter Graft. Device Description Planning and Sizing Deployment Sequence Patient Follow-Up

Neurological Complications of TEVAR. Frank J Criado, MD. Union Memorial-MedStar Health Baltimore, MD USA

Hostile Proximal Neck: A New Conformable EVAR Device

MODERN METHODS FOR TREATING ABDOMINAL ANEURYSMS AND THORACIC AORTIC DISEASE

Challenges. 1. Sizing. 2. Proximal landing zone 3. Distal landing zone 4. Access vessels 5. Spinal cord ischemia 6. Endoleak

RadRx Your Prescription for Accurate Coding & Reimbursement Copyright All Rights Reserved.

Use of covered stents and endograft as a rescue treatment in a patient with a complex form of recurrent aortic coarctation

Malperfusion Syndromes Type B Aortic Dissection with Malperfusion

Treatment of complex thoracic cases Focus on the new Gore Active Control TAG device

Haemodynamically unstable patient with chest trauma

Risks for Retrograde Type-A Dissection After TEVAR

Arch Repair with the Bolton Medical RelayBranch Thoracic Stent-graft system: Multicenter experience

Bilateral use of the Gore IBE device for bilateral CIA aneurysms and a first interim analysis of the prospective Iceberg registry

Experience of endovascular procedures on abdominal and thoracic aorta in CA region

Remodeling of proximal neck angulation after endovascular aneurysm repair

TEVAR for the Ascending Aorta

Advances in Treatment of Traumatic Aortic Transection

ENCORE, a Study to Investigate the Durability of Polymer EVAR with Ovation A Contemporary Review of 1296 Patients

No Disclosure. Aortic Dissection in Japan. This. The Challenge of Acute and Chronic Type B Aortic Dissections with Endovascular Aortic Repair

Mid-term results from ANCHOR: How does this data influence the treatment algorithm for hostile EVAR anatomies

Ovation. Sean Lyden, MD Department Chair, Vascular Surgery Cleveland Clinic

Jean M Panneton, MD Professor of Surgery Program Director Vascular Surgery Chief EVMS. Arch Pathology: The Endovascular Era is here

Are stent-grafts for acute type B dissection durable? Est-ce que les stents graft pour la dissection aigue de type B sont efficaces à moyen terme?

No Neck Is Needed! Treat the Aneurysm Instead! Andrew Holden, MD Associate Professor of Radiology Auckland City Hospital

Optimizing Accuracy of Aortic Stent Grafts in Short Necks

Development of a Branched LSA Endograft & Ascending Aorta Endograft

11/20/2014. Disclosures. Kissing Balloons and Stents. Treatment of Aortoiliac Occlusive Disease. Data on Patency of Kissing Stents.

Analysis of Type IIIb Endoleaks Encountered with Endologix Endografts

Treatment options of late failures of EVAS. Michel Reijnen Rijnstate Arnhem The Netherlands

The Short Proximal AAA Neck

Clinical trial and real-world outcomes of an endovascular iliac aneurysm repair with the GORE Iliac Branch Endoprosthesis (IBE)

From the Western Vascular Society

Advantage of a precurved fenestrated endograft for aortic arch disease: Simplified arch aneurysm treatment in Japan 2010 and 2011

Objective assessment of current stent grafts: which graft for which lesion. Ludovic Canaud, MD, PhD Pierre Alric, MD, PhD Montpellier, France

Transcription:

Anatomic factors associated with acute endograft collapse after Gore TAG treatment of thoracic aortic dissection or traumatic rupture Bart E. Muhs, MD, a Ron Balm, MD, PhD, b Geoffrey H. White, MD, c and Hence J. M. Verhagen, MD, PhD, d New York, NY; Amsterdam and Utrecht, The Netherlands; and Sydney, Australia Objective: The potentially devastating complication of total or near total thoracic endoprosthesis collapse has been described with the TAG device (W. L. Gore & Associates, Flagstaff, Ariz). This rare complication has resulted in a warning to clinicians and speculation about the etiology of this problem. This report evaluates potential causative anatomic factors that may increase the probability of endoprosthesis collapse in patients undergoing endovascular thoracic aneurysm repair (TEVAR). Methods: Preoperative and postoperative computed tomography scans were collected worldwide representing six patients who had experienced radiologically confirmed TAG endoprosthesis collapse. These were compared with a matched cohort of five patients with a TAG endoprosthesis in the same anatomic position in which no collapse occurred. Anatomic variables of aortic arch angulation, apposition, intraluminal lip length, proximal aortic diameter, distal aortic diameter, intragraft aortic diameter, percentage of oversizing, and angle of the proximal endograft to the aortic arch were compared between groups. Differences between groups were determined using the Student t test, with P <.05 considered significant. Results: The two groups (collapse vs no collapse) were evenly matched demographically, and all underwent endoluminal treatment with the TAG device, with no differences in gender, graft position in the aorta, operative indication, or age (P ). Distal sealing zone aortic diameter standard deviation of 18.9 1.7 mm vs 22.7 2.7 mm and minimum aortic diameter within the endograft of 18.6 1.7 mm vs 22.4 3.1 mm predicted collapse (P <.05). Proximal aortic diameter, apposition, intraluminal lip length, aortic arch angle, and angle of proximal endograft to aortic arch did not predict collapse (P ). Conclusion: Thoracic endograft collapse is an exceedingly rare event. In this series, endoprosthesis collapse occurred in patients who were treated outside the manufacturer s instructions for use for minimum required aortic diameter. Although distal aortic diameter and minimum intragraft aortic diameter predicted collapse, other variables may also influence this complication but were not significant owing to potential type II statistical errors. In the future, caution should be exercised when contemplating TEVAR in patients with small (<23 mm) aortic diameters. (J Vasc Surg 2007; 45:655-61.) Endovascular management has emerged during the last 10 years as a valuable treatment modality for thoracic aortic aneurysms that can lessen surgical morbidity and mortality, decrease hospital stay, and provide for improved outcomes in properly selected patients. 1-4 Endovascular thoracic aortic aneurysm repair (TEVAR) has demonstrated excellent short-term and medium-term results for degenerative aneurysms, traumatic aneurysms, and contained traumatic aortic ruptures alike. 5,6 Endovascular procedures From the Division of Vascular Surgery, New York University School of Medicine, a the Department of Vascular Surgery, Academic Medical Center, b the Department of Vascular Surgery, Royal Prince Alfred Hospital, c and the Department of Vascular Surgery, University Medical Center Utrecht. d Competition of interest: Dr Muhs receives a speaking honorarium from W. L. Gore & Associates, and Drs White and Verhagen have consulting agreements with W. L. Gore & Associates. Additional material for this article may be found online at www.jvascsurg. org. Reprint requests: Hence J. M. Verhagen, MD, PhD, Department of Vascular Surgery, G.04.129, University Medical Center Utrecht, PO Box 85500, 3508GA Utrecht, The Netherlands (e-mail: H.Verhagen@umcutrecht.nl). 0741-5214/$32.00 Copyright 2007 by The Society for Vascular Surgery. doi:10.1016/j.jvs.2006.12.023 of the thoracic aorta are not without complications, however. 7-10 TEVAR for the treatment of traumatic aortic rupture is particularly promising given the associated serious comorbidities that accompany these polytrauma patients. 11,12 Endovascular repair of traumatic aortic rupture allows for definitive treatment of the vascular injury without the need for bypass or thoracotomy and reduces the recovery time that is associated with these procedures. 6,13 However, many of the complications associated with TEVAR for aneurysmal disease are uncommon in the endovascular treatment of traumatic aortic rupture. Traumatic patients represent a different disease process than aneurysm patients, with dissimilar anatomy, treatment considerations, and complications. Concerning case reports have described the potentially devastating complication of total or near total acute endoprosthesis collapse or infolding with the TAG system (W. L. Gore & Associates, Flagstaff, Ariz). 14,15 This complication is not unique to the TAG endograft, but has been observed with other brands as well, including the Aneurx (Medtronic, Minneapolis, Minn) and the Zenith (Cook, Bloomington, Ind). This complication has primarily been reported in pa- 655

656 Muhs et al JOURNAL OF VASCULAR SURGERY April 2007 Fig 1. Measurements included (A) aortic arch angulation, (B) angle of endograft extending into the aortic lumen, (C) endograft lip length, (D) confirmation of infolding, presence of motion artifact, endograft distance from the subclavian artery, and aortic diameters perpendicular to the center lumen line at the proximal, intragraft, and distal landing zones. tients treated for traumatic aortic rupture or dissection and rarely in patients treated for thoracic aneurysms. Patients with traumatic rupture or dissection often have relatively normal, small proximal aortas. These anatomic realities prompted us to question whether specific anatomic factors predispose certain patients to endoprosthesis collapse and if a careful analysis of preoperative anatomy might minimize this risk. This rare complication has resulted in a warning to clinicians and speculation about the etiology of this problem. The current the market leader in thoracic endografts and the only device approved by United States Food and Drug Administration (FDA) is the Gore TAG endoprosthesis. The purpose of this report is to evaluate potential causative anatomic factors that may increase the probability of endoprosthesis collapse in patients undergoing endovascular repair of aortic trauma and dissection. METHODS Indication, preoperative sizing, choice of device, and surgical method was left solely to the discretion of the operating team at each institution. Postoperative imaging and any secondary interventions were also entirely the responsibility of the operating surgeon. Preoperative or postoperative computed tomography (CT) scans, or both, were collected from three international sites representing six patients who had experienced radiologically confirmed TAG endoprosthesis collapse between 2004 and 2006. Five patients with collapse had been treated for trauma and one for an aortic dissection.

JOURNAL OF VASCULAR SURGERY Volume 45, Number 4 Muhs et al 657 Fig 2. A-D, Representative images of the collapsed endoprosthesis in four patients are shown. It is known that collapse is not a binary phenomenon. Some degrees of infolding may compromise seal and structural integrity without narrowing the lumen enough to compromise flow. In this report, TAG endoprosthesis collapse was determined solely on CT imaging and not on hemodynamic or clinical parameters. Additional patients with confirmed collapse exist, but were unknown to us or did not have adequate imaging for analysis. The control group, which was matched for demographics, consisted of five patients from a single institution with a traumatic aortic rupture who were treated with a TAG endoprosthesis and who did not show collapse. The control institution also provided two of the patients with a confirmed endoprosthesis collapse. All patients in the study had preoperative sizing and postoperative collapse confirmed through the use of CT scanning. Other imaging was used in conjunction with the CT scans, including transesophageal echo, chest radiography, and angiography, but was not used in the analysis of anatomic factors predicting collapse. The original Digital Imaging and Communication in Medicine (DICOM) data, when available, was provided to an independent imaging company (Medical Metrix Solutions, West Lebanon, NH), which performed the requested measurements (Fig 1). Two CT scans did not have original DICOM data, and measurements on these images were performed manually using commercially available imaging software. Measurements included aortic arch angulation, angle of endograft extending into the aortic lumen, the length of proximal endograft not in apposition to the aortic wall (endograft lip length), presence of motion artifact, endograft distance from the subclavian artery, and aortic diameters perpendicular to the center lumen line at the proximal, intragraft, and distal landing zones (Fig 2). Statistical analysis. A Student t test for unpaired data was used to analyze changes in area and diameters. Significance was assumed at P.05. Data are expressed as mean standard deviation. RESULTS Image quality was considered good-to-excellent in all images, and satisfactory analysis was performed. The two groups (collapse vs no collapse) were evenly matched demographically, with no differences in gender or age (P ). Representative images of the collapsed endoprosthesis are shown (Fig 2). Dynamic cine CT angiography was available in one collapse. This demonstrated video evidence of the enormous forces exerted on stents placed in the aortic arch, the problem of a collapsed graft, and a very small true lumen with a blind-ending false lumen (Video clip, online only). Endovascular salvage consisting of implantation of additional endoprosthesis inside the collapsed TAG (Fig 3) or ballooning (Fig 4) were attempted in four of the six patients who experienced collapse, with varying results. In one patient, a 28-mm 10-cm TAG was deployed first, but

658 Muhs et al JOURNAL OF VASCULAR SURGERY April 2007 Fig 3. In one patient, attempts were made to correct the graft infolding through endovascular salvage with the implantation of additional endoprosthesis inside a collapsed TAG. This attempt was unsuccessful. A, A TAG (28 mm 10 cm) was deployed first, but owing to a proximal type I endoleak, a second TAG (28 mm 15 cm) was deployed slightly more proximally. B and C, Computed tomography (CT) imaging showed clear infolding of the second TAG. D, To treat the infolding, a giant Palmaz stent was deployed. A CT scan 1 day after the placement of the Palmaz stent showed complete infolding of both the second TAG as well as the Palmaz stent. owing to a proximal type I endoleak, a second TAG (28 mm 15 cm) was deployed slightly more proximally. Clear infolding of the second TAG was seen on CT imaging. To treat the infolding, a giant Palmaz stent (Cordis, Miami Lakes, Fla) was deployed. A CT scan 1 day after the placement of the Palmaz stent showed complete infolding of both the second TAG as well as the Palmaz stent. In a second patient, infolding was observed on the postoperative CT scan, and the patient underwent balloon angioplasty. The follow-up CT scan after balloon angioplasty demonstrated correction of the infolding. In one patient, open explant and repair was performed successfully. One patient declined further treatment and is currently asymptomatic. Of the variables compared, distal sealing zone aortic diameter (18.9 1.7 mm vs 22.7 2.7 mm) and minimum aortic diameter within the endograft, defined as the preoperative minimum aortic diameter obtained from the proximal sealing zone to the distal sealing zone taken at 1-m increments (18.6 1.7 mm vs 22.4 3.1 mm), predicted collapse (P.05; Table I). Proximal aortic diameter, apposition, intraluminal lip length, aortic arch angle, and angle of proximal endograft to aortic arch failed to reach statistical significance and therefore did not predict collapse (P ; Table I). No patient who experienced endograft collapse demonstrated radiographic evidence of stent fracture or fabric tear, nor did stent graft migration occur. However, the stent graft from the patient who underwent explantation had two rows of completely fractured stents in the place of infolding (Fig 5). DISCUSSION TEVAR has been a significant advance in the treatment of thoracic aneurysmal disease; however, its use has not

JOURNAL OF VASCULAR SURGERY Volume 45, Number 4 Muhs et al 659 Fig 4. Attempts were made in one patient to correct the graft infolding by using endovascular salvage with additional balloon angioplasty inside a collapsed TAG. This attempt was successful. A, Infolding was observed on the postoperative computed tomography (CT). B, The patient was taken back to the operating room for balloon angioplasty. C, The follow-up CT scan after balloon angioplasty demonstrated correction of the infolding. Table I. Variables predicting collapse Variable P Proximal aortic diameter Distal aortic diameter.02 Smallest aortic diameter (within endograft).03 Intraluminal lip length Arch radius of curvature Lip to arch angle Complete stent graft apposition Coverage of subclavian Percent oversizing been without significant adverse events. In low-risk patients, up to 32% experienced at least one acute ( 30 days) complication. 16 Procedure-related mortality approaches 17% in high-risk patients with thoracic aortic aneurysm treated with an endovascular approach. 7 Specifically regarding the TAG device, spine fractures were a problem with the earlier device. 16 Although resulting in minimal adverse clinical events, in the phase II clinical trial of the first generation TAG device, up to 14% of patients experienced a fracture of the longitudinal spine. 16 As a consequence, Gore redesigned the device, which included removal of the longitudinal spine, wires made thicker and stronger, deployment made faster, and fabric Fig 5. Open explant and repair was performed successfully in one patient. The explanted stent graft demonstrated infolding and two rows of fractured stents. changed to a low permeability material. To our knowledge, no collapses of the TAG device had been reported before these changes. The redesigned current device has encountered this problem, however, specifically in patients treated for aortic rupture or dissection, and is the subject of this report. Although this report was not designed to determine design characteristics leading to collapse, one can speculate that removal of the spine resulted in decreased structural support that may have increased the potential of collapse. A more likely possibility is that as clinicians gain expertise and comfort with TEVAR, the current TAG device is

660 Muhs et al JOURNAL OF VASCULAR SURGERY April 2007 Table II. Anatomic and device variables for all patients Patient Collapse TAG diamete Percent oversizing (%) Liplength Proximal aorta Distal aorta Radius curvature (cm) Smallest aorta 1 Yes 28 135 20 26.2 20.7 2.2 20.7 2 No 34 138 0 28.3 25 6 24.6 3 No 40 167 8 21.6 21 2.2 24 4 No 28 126 13 25 22.3 5.4 22.3 5 No 39 115 0 25.6 25.7 N/A 24 6 Yes 26 130 18 22 20 3.8 20 7 Yes 26 144 0 21 18 3.1 18 N/A, Not applicable. being deployed in clinical situations that are less than optimal. Deployment of the TAG outside of Gore s instructions for use (IFU), such as in small diameter aortas, difficult aortic arches, poor sealing zones, and in high-risk patients, virtually assures increased complication rates when compared with the earlier clinical trials. All of the patients in this series who experienced endoprosthesis collapse were treated outside Gore s IFU for a minimum required aortic diameter of 23 mm. No collapse occurred in patients treated with aortic diameters of 23 mm. Clearly, the physicians choosing to deploy the thoracic endografts outside the IFU did so knowing that these patients were severely injured and unfit for open surgical repair. The main issue is that the grafts used were severely oversized because they are indicated for aneurysm patients. If smaller stent grafts had been available, they would have been used. In our experience and in the experiences described in the literature, most collapses have occurred in patients treated for traumatic aortic trauma or dissection and not in patients with atherosclerotic aneurysms. Although traumatic aortic trauma and thoracic dissection are certainly different disease entities, we believe that they share some anatomic factors that may predispose to acute collapse. Specifically, they are often located in smaller, younger aortas and often extend high into the aortic arch. Collapses occurred when the stent grafts were placed high in the aorta, which always included the distal portion of relatively steeply angulated arches common in young patients, most commonly involved in trauma. Anecdotally, this complication rarely occurs in the longer and wider aortas associated with aneurysmal disease. These two populations of traumatic aortic injury and dissection vs atherosclerotic aneurysmal disease represent two different populations with dissimilar operative indications and postoperative complications. It is very difficult to follow IFU sizing criteria in most dissection patients, where there is often an acute taper or narrowing of the true lumen. Based on this potentially devastating complication, many surgeons are reluctant to use the TAG device in any trauma or dissection case. Although this report assesses anatomic factors that may predict collapse, the unique packaging and release mechanism of the TAG device may worsen the problem of infolding. The device is packaged in a furled configuration with inbuilt infolding. Thus, a small aorta prevents it from unfolding completely. Although distal aortic diameter and minimum intragraft aortic diameter predicted collapse, other variables may have influenced this complication but were not significant in this series owing to potential type II statistical errors related to small sample size. Thoracic endograft collapse is an exceedingly rare event, making it difficult to collect imaging on large patient numbers. However, others have suggested and it seems intuitive that factors such as poor apposition of the endograft to the aortic wall, acute aortic arch angulation, and potentially other anatomic factors may play a causative role in endoprosthesis collapse. Nevertheless, we believe that caution should be exercised when these anatomic factors are present. Dynamic cine CT angiography scanning illustrates the very dynamic interface between blood flow in the lumen and the inner curvature of the graft (Video clip, online only). In this, one observes a TAG in the arch with a significant lip protruding into the aortic lumen. Infolding seems a logical result. Anecdotally, especially in the United States, physicians have implanted aortic extender cuffs to treat traumatic aortic ruptures. These cuffs are very short (3 to 4 cm long), and no infolding has been reported as far as we know. This should not be viewed as an endorsement of this off-label use, however, because there are numerous inherent flaws in using such short, stacked cuffs. Only within the last 10 years have thoracic endografts been introduced into wide scale clinical practice. 17,18 Significant design modifications have already taken place, and it seems likely that further modifications will continue in an effort to minimize complications and improve clinical outcomes. Although this report was not designed to determine the ideal design of future endografts, we can speculate that future designs incorporating increased radial force, better attachment systems, increased flexibility, preformed stent graft curves, and noncircular stent grafts will be forthcoming. CONCLUSION Thoracic endograft collapse is an exceedingly rare event. All of the patients in this series who experienced endoprosthesis collapse were treated outside the manufacturer s IFU for minimum required aortic diameter.

JOURNAL OF VASCULAR SURGERY Volume 45, Number 4 Muhs et al 661 Although distal aortic diameter and minimum intragraft aortic diameter predicted collapse, other variables may also influence this complication but were not significant owing to potential type II statistical errors related to small sample size. In the future, caution should be exercised when contemplating TEVAR with small ( 23 mm) aortic diameters when using the current design TAG endoprosthesis. (Table 2). AUTHOR CONTRIBUTIO Conception and design: BM, HV Analysis and interpretation: BM, HV, RM, GW Data collection: BM, HV, RM, GW Writing the article: BM, HV Critical revision of the article: BM, HV, RM, GW Final approval of the article: BM, HV, RM, GW Statistical analysis: BM, HV Obtained funding: BM Overall responsibility: BM REFERENCES 1. Criado FJ, Clark, Barnatan MF. Stent graft repair in the aortic arch and descending thoracic aorta: a 4-year experience. J Vasc Surg 2002; 36:1121-8. 2. Chabbert V, Otal P, Bouchard L, Soula P, Van TT, Kos X, et al. Midterm outcomes of thoracic aortic stent-grafts: complications and imaging techniques. J Endovasc Ther 2003;10:494-504. 3. Melissano G, Civilini E, Bertoglio L, Setacci F, Chiesa R. Endovascular treatment of aortic arch aneurysms. Eur J Vasc Endovasc Surg 2005;29: 131-8. 4. Leurs LJ, Bell R, Degrieck Y, Thomas S, Hobo R, Lundbom J; EUROSTAR; UK Thoracic Endograft Registry collaborators. Endovascular treatment of thoracic aortic diseases: combined experience from the EUROSTAR and United Kingdom Thoracic Endograft registries. J Vasc Surg 2004;40:670-9; discussion 679-80. 5. Melissano G, Civilini E, de Moura MR, Calliari F, Chiesa R. Single center experience with a new commercially available thoracic endovascular graft. Eur J Vasc Endovasc Surg 2005;29:579-85. 6. Wellons ED, Milner R, Solis M, Levitt A, Rosenthal D. Stent-graft repair of traumatic thoracic aortic disruptions. J Vasc Surg 2004;40: 1095-100. 7. Hansen CJ, Bui H, Donayre CE, Aziz I, Kim B, Kopchok G. Complications of endovascular repair of high-risk and emergent descending thoracic aortic aneurysms and dissections. J Vasc Surg 2004;40:228-34. 8. Kasirajan K, Milner R, Chaikof EL. Late complications of thoracic endografts. J Vasc Surg 2006;43(suppl A):94A-99A. 9. Melissano G, Tshomba Y, Civilini E, Chiesa R. Disappointing results with a new commercially available thoracic endograft. J Vasc Surg 2004;39:124-30. 10. Kasirajan K. Thoracic endografts: procedural steps, technical pitfalls and how to avoid them. Semin Vasc Surg 2006;19:3-10. 11. Balm R, Legemate DA. Traumatic aortic rupture. Br J Surg 2006;93: 1033-4. 12. Hoornweg LL, Dinkelman MK, Goslings JC, Reekers JA, Verhagen HJ, Verhoeven EL, et al. Endovascular management of traumatic ruptures of the thoracic aorta: a retrospective multicenter analysis of 28 cases in The Netherlands. J Vasc Surg 2006;43:1096-102; discussion 1102. 13. Milas ZL, Milner R, Chaikoff E, Wulkan M, Ricketts R. Endograft stenting in the adolescent population for traumatic aortic injuries. J Pediatr Surg 2006;41:e27-30. 14. Steinbauer MG, Stehr A, Pfister K, Herold T, Zorger N, Topel I, et al. Endovascular repair of proximal endograft collapse after treatment for thoracic aortic disease. J Vasc Surg 2006;43:609-12. 15. Idu MM, Reekers JA, Balm R, Ponsen KJ, de Mol BA, Legemate DA. Collapse of a stent-graft following treatment of a traumatic thoracic aortic rupture. J Endovasc Ther 2005;12:503-7. 16. Makaroun MS, Dillavou ED, Kee ST, Sicard G, Chaikof E, Bavaria J, et al. Endovascular treatment of thoracic aortic aneurysms: results of the phase II multicenter trial of the GORE TAG thoracic endoprosthesis. J Vasc Surg 2005;41:1-9. 17. Boudghene F, Sapoval M, Bigot JM, Michel JB. Endovascular graft placement in experimental dissection of the thoracic aorta. J Vasc Interv Radiol 1995;6:501-7. 18. May J, White G, Yu W, Sachinwalla T, McGahan T, Monaghan G. Endoluminal repair of atypical dissecting aneurysm of descending thoracic aorta and fusiform aneurysm of the abdominal aorta. J Vasc Surg 1995;22:167-72. Submitted Oct 7, 2006; accepted Dec 11, 2006. Additional material for this article may be found online at www.jvascsurg.org.