Review Article A Meta-Analysis about the Screening Role of Pulse Oximetry for Congenital Heart Disease

Similar documents
Accuracy of pulse oximetry in screening for congenital heart disease in asymptomatic newborns: a systematic review

IMAGES. in PAEDIATRIC CARDIOLOGY

Title: Oxygen Saturation Screening in Newborns: a Clinical Effectiveness Review

Screening for critical congenital heart defects with pulse oximetry: medical aspects Ewer, Andrew

Pulse oximetry screening for critical congenital heart defects. Where are we and where next?

Effectiveness of pulse oximetry screening for congenital heart disease in asymptomatic new-borns

Positive predictive value of pulse oximetry in the screening of critical congenital heart defects in term neonates

Pulse oximetry screening for critical congenital heart defects in asymptomatic newborn babies: a systematic review and meta-analysis

ARTICLE. Missed Diagnosis of Critical Congenital Heart Disease. Ruey-Kang R. Chang, MD, MPH; Michelle Gurvitz, MD; Sandra Rodriguez, MS

Trust Guideline on Routine Oxygen Saturation Measurement on the New-born (Pulse Oximetry)

Pulse oximetry screening in newborns to enhance detection of critical congenital heart disease

Spectrum and age of presentation of significant congenital heart disease in KwaZulu Natal, South Africa

Pulse oximetry with clinical assessment to screen for congenital heart disease in neonates in China: a prospective study

Pulse oximetry screening in newborns to enhance detection of critical congenital heart disease

Clinical Screening for Congenital Heart Disease at Birth: A Prospective Study in a Community Hospital in Kerala

Pulse Oximetry Screening in Newborns to Enhance the Detection Of Critical Congenital Heart Disease. Frequently Asked Questions

Can echocardiographic screening in the early days of life detect critical congenital heart disease among apparently healthy newborns?

Clinical Study Incidence of Retinopathy of Prematurity in Extremely Premature Infants

Daofang Zhu, Xianming Dou, Liang Tang, Dongdong Tang, Guiyi Liao, Weihua Fang, and Xiansheng Zhang

Research Article Challenges in Assessing Outcomes among Infants of Pregnant HIV-Positive Women Receiving ART in Uganda

CYANOTIC CONGENITAL HEART DISEASES. PRESENTER: DR. Myra M. Koech Pediatric cardiologist MTRH/MU

Research Article Prevalence and Trends of Adult Obesity in the US,

Congenital Heart Disease. CCCHD In WI. Critical Congenital Heart Disease. Why Screen? 4/20/2018. Early Detection = Better Outcomes

Age at Referral and Mortality From Critical Congenital Heart Disease. abstract

Reliability of Echocardiography Measurement of Patent Ductus Arteriosus Minimum Diameter: A Meta-analysis

Conference Paper Antithrombotic Therapy in Patients with Acute Coronary Syndromes: Biological Markers and Personalized Medicine

Management of a child with cyanosis

This guideline is applicable to medical and nursing staff caring for neonates in the West of Scotland.

Research Article Predictions of the Length of Lumbar Puncture Needles

E ven though a thorough physical examination of every

Names of document authors Line Dr David Booth (Clinical Director in Paediatrics) Mr Charles Graham (Head of Technical Cardiology)

Research Article Predictive Factors for Medical Consultation for Sore Throat in Adults with Recurrent Pharyngotonsillitis

Correspondence should be addressed to Alicia McMaster;

Case Report Asymptomatic Pulmonary Vein Stenosis: Hemodynamic Adaptation and Successful Ablation

Clinical Study IVIG Effects on Erythrocyte Sedimentation Rate in Children

Research Article Photovoice: A Novel Approach to Improving Antituberculosis Treatment Adherence in Pune, India

Clinical Study Metastasectomy of Pulmonary Metastases from Osteosarcoma: Prognostic Factors and Indication for Repeat Metastasectomy

Am I Blue? Newborn Pulse Oximetry Screening for Critical Congenital Heart Disease The State of the State

Actionable Patient Safety Solution (APSS) for #7B: FAILURE TO DETECT CRITICAL CONGENITAL HEART DISEASE (CCHD) IN NEWBORNS

Devendra V. Kulkarni, Rahul G. Hegde, Ankit Balani, and Anagha R. Joshi. 2. Case Report. 1. Introduction

RESEARCH ABSTRACT INTRODUCTION

Isolated major aortopulmonary collateral as the sole pulmonary blood supply to an entire lung segment

Congenital Chylothorax

Correspondence should be addressed to Taha Numan Yıkılmaz;

Screening Compliance for Critical Congenital Heart Disease in Texas Newborns

Presentation of congenital heart disease in infancy: implications for routine examination

Research Article Identifying Prognostic Criteria for Survival after Resuscitation Assisted by Extracorporeal Membrane Oxygenation

Nordic pulse oximetry screening - implementation status and proposal for uniform guidelines

Screening for Critical Congenital Heart Disease

Research Article Measures of Adherence and Challenges in Using Glucometer Data in Youth with Type 1 Diabetes: Rethinking the Value of Self-Report

Case Report Tortuous Common Carotid Artery: A Report of Four Cases Observed in Cadaveric Dissections

Research Article Balloon Dilatation of Pediatric Subglottic Laryngeal Stenosis during the Artificial Apneic Pause: Experience in 5 Children

Research Article Clinical and Epidemiological Investigation of TCM Syndromes of Patients with Coronary Heart Disease in China

Critical congenital heart disease screening by pulse oximetry in a neonatal intensive care unit

Research Article Abdominal Aortic Aneurysms and Coronary Artery Disease in a Small Country with High Cardiovascular Burden

Case Report Complete Obstruction of Endotracheal Tube in an Infant with a Retropharyngeal and Anterior Mediastinal Abscess

Research Article Clinical Outcome of a Novel Anti-CD6 Biologic Itolizumab in Patients of Psoriasis with Comorbid Conditions

Case Report Successful Implantation of a Coronary Stent Graft in a Peripheral Vessel

after birth) based on existing clinical practice and to investigate factors associated with late detection.

Case Report Bilateral Distal Femoral Nailing in a Rare Symmetrical Periprosthetic Knee Fracture

Clinical Study Rate of Improvement following Volar Plate Open Reduction and Internal Fixation of Distal Radius Fractures

Case Report Long-Term Outcomes of Balloon Dilation for Acquired Subglottic Stenosis in Children

Case Report Evolution of Skin during Rehabilitation for Elephantiasis Using Intensive Treatment

Research Article The Effect of Elevated Triglycerides on the Onset and Progression of Coronary Artery Disease: A Retrospective Chart Review

Research Article A Selective Ensemble Classification Method Combining Mammography Images with Ultrasound Images for Breast Cancer Diagnosis

Research Article Comparison of Colour Duplex Ultrasound with Computed Tomography to Measure the Maximum Abdominal Aortic Aneurysmal Diameter

Research Article Hb A1c Separation by High Performance Liquid Chromatography in Hemoglobinopathies

1. Introduction. Correspondence should be addressed to Richard J. Drew; Received 23 July 2015; Accepted 15 November 2015

Research Article The Impact of the Menstrual Cycle on Perioperative Bleeding in Vitreoretinal Surgery

Emily J. Hadgkiss, 1 George A. Jelinek, 1,2 Tracey J. Weiland, 1,3 Naresh G. Pereira, 4 Claudia H. Marck, 1 and Dania M.

Department of Internal Medicine, Saitama Citizens Medical Center, Saitama , Japan

Research Article Urinary Catheterization May Not Adversely Impact Quality of Life in Multiple Sclerosis Patients

Systematic Reviews and meta-analyses of Diagnostic Test Accuracy. Mariska Leeflang

Case Report Medial Radial Head Dislocation Associated with a Proximal Olecranon Fracture: A Bado Type V?

William W. Hale III, 1 Quinten A. W. Raaijmakers, 1 Anne van Hoof, 2 and Wim H. J. Meeus 1,3. 1. Introduction

Case Report Arthroscopic Treatment of Septic Arthritis of the Elbow in a 4-Year-Old Girl

Research Article Decreasing Prevalence of Transfusion Transmitted Infection in Indian Scenario

Case Report Intra-Articular Entrapment of the Medial Epicondyle following a Traumatic Fracture Dislocation of the Elbow in an Adult

Correspondence should be addressed to Vitharon Boon-yasidhi;

Research Article Opioid Use Is Not Associated with Incomplete Wireless Capsule Endoscopy for Inpatient or Outpatient Procedures

Research Article Timing of Caffeine Therapy and Neonatal Outcomes in Preterm Infants: A Retrospective Study

These authors contributed equally to this work. These authors also contributed equally to this work. * Methods and Findings

How to Recognize a Suspected Cardiac Defect in the Neonate

11/27/2012. Objectives. What is Critical Congenital Heart Disease?

Baris Beytullah Koc, 1 Martijn Schotanus, 1 Bob Jong, 2 and Pieter Tilman Introduction. 2. Case Presentation

R. J. L. F. Loffeld, 1 P. E. P. Dekkers, 2 and M. Flens Introduction

Case Report Cervical Stimulation in the Treatment of Children with Lymphedema of All Four Extremities: A Case Report and Literature Review

Case Report Uncommon Mixed Type I and II Choledochal Cyst: An Indonesian Experience

Case Report Pediatric Transepiphyseal Seperation and Dislocation of the Femoral Head

Research Article Stroke Risk among Patients with Type 2 Diabetes Mellitus in Zhejiang: A Population-Based Prospective Study in China

Case Report Successful Treatment of Double-Orifice Mitral Stenosis with Percutaneous Balloon Mitral Commissurotomy

Research Article The Need for Improved Management of Painful Diabetic Neuropathy in Primary Care

Tom Eisele, Benedikt M. Muenz, and Grigorios Korosoglou. Department of Cardiology & Vascular Medicine, GRN Hospital Weinheim, Weinheim, Germany

Surveillance report Published: 9 January 2017 nice.org.uk

Case Report A Unique Case of Left Second Supernumerary and Left Third Bifid Intrathoracic Ribs with Block Vertebrae and Hypoplastic Left Lung

Research Article Maintenance of the Results of Stage II Lower Limb Lymphedema Treatment after Normalization of Leg Size

Actionable Patient Safety Solution (APSS) for #7B: FAILURE TO DETECT CRITICAL CONGENITAL HEART DISEASE (CCHD) IN NEWBORNS

Research Article A Meta-Analysis of Randomized Controlled Trials on Acupuncture for Amblyopia

Research Article Validity of Palmar Pallor for Diagnosis of Anemia among Children Aged 6 59 Months in North India

Transcription:

Hindawi BioMed Research International Volume 2017, Article ID 2123918, 6 pages https://doi.org/10.1155/2017/2123918 Review Article A Meta-Analysis about the Screening Role of Pulse Oximetry for Congenital Heart Disease Caiju Du, 1 Dianmei Liu, 2 Guojing Liu, 3 and Huaixin Wang 4 1 Cardiovascular Surgery, Affiliated Hospital of Weifang Medical University, 2428 Yuhe Road, Kuiwen District, Weifang, Shandong Province 261031, China 2 Imaging Center, Affiliated Hospital of Weifang Medical University, 2428 Yuhe Road, Kuiwen District, Weifang, Shandong Province 261031, China 3 Operating Room, Affiliated Hospital of Weifang Medical University, 2428 Yuhe Road, Kuiwen District, Weifang, Shandong Province 261031, China 4 Emergency Department, Yidu Central Affiliated Hospital of Weifang Medical University, 4138 Linglong South Road, Qingzhou, Shandong Province 262550, China Correspondence should be addressed to Dianmei Liu; dianml05@126.com Received 8 June 2017; Accepted 20 November 2017; Published 10 December 2017 Academic Editor: Vladimiro Vida Copyright 2017 Caiju Du et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Objective. The opinions about the application of pulse oximetry in diagnosis of congenital heart disease (CHD) were debatable. We performed this meta-analysis to confirm the diagnostic role of pulse oximetry screening for CHD. Methods. Relevant articles were searched in the databases of Pubmed, Embase, Google Scholar, and Chinese National Knowledge Infrastructure (CNKI) up to April 2017. Data was processed in the MetaDiSc 1.4 software. Pooled sensitivity and specificity with 95% confidence interval (95% CI) were calculated to explain the diagnostic role of pulse oximetry screening for CHD. I 2 50%orp < 0.05 indicated significant heterogeneity. Area under curve (AUC) of summary receiver operating characteristics (SROC) was calculated to assess its diagnostic accuracy. The robustness of overall results was evaluated by sensitivity analysis. Publication bias was evaluated by Deek s funnel plot. Results. 22 eligible articles were selected. Pooled sensitivity and specificity were 0.69 (0.67 0.72) and 0.99 (0.99-0.99), respectively. The corresponding AUC was 0.9407, suggesting high diagnostic accuracy of pulse oximetry screening for CHD. Sensitivity analysis demonstrated that the pooled results were robust. Deek s funnel plot seemed to be symmetrical. Conclusions. Pulse oximetry screening could be used to diagnose CHD. It shows high diagnosis specificity and accuracy. 1. Introduction Congenital heart disease (CHD) is regarded as a main cause of infant death, with an incidence of 8 in every 1000 live births [1]. It needs invasive intervention during the neonatal period and these neonates suffering this disease benefit most from early detection [2]. Prenatal diagnosis just picks up <50% of all cases [3 6]. Routine neonatal inspection fails to detect above than 50% of CHD infants. More than 55% of neonates show no murmur symptom in the nursery, and 82% of them are discharged before diagnosis results are obtained [7]. Early diagnosis of CHD is crucial since the delayed diagnosis results in cardiovascular collapse, cardiac failure, and death, whereas early diagnosis during the first few days of life is difficult. Therefore, an effective screening program for CHD is necessary. In recent years, pulse oximetry has been suggested as adiagnosistoolforchd[8 10].Itisproductiveinthe detection of CHD before discharge and could decrease missed cases to 4% [11, 12]. Some states of United States have made screening of pulse oximetry for CHD mandatory before discharge in hospital. Pulse oximetry can detect mild hypoxemia, which is common feature for many forms of CHD. It could recognize the cases that are not recognized by clinical examination [13]. Since the introduction of pulse oximetry to screen CHD in 1995, many studies have focused on the subject [14 17]. Despite the fact that there were differences in screening time, cut-off values, target lesions, and others

2 BioMed Research International 488 relevant articles were obtained from database 358 articles were excluded after screening titles or abstracts 130 articles for detailed assessment 108 articles were excluded: 71 for no available data 4 for duplicate publication 11 for only SpO 2 level in CHD patients 22 eligible articles were included in the present review Figure 1: Flow chart of articles selection. 22 articles were selected for meta-analysis. among the relevant studies, the opinion is consistent that pulseoximetryscreeningisausefuldiagnosticmethodof CHD. However, existing data is still insufficient to initiate a recommendation for application of pulse oximetry in routine care. This present meta-analysis was performed to confirm the diagnostic role of pulse oximetry for CHD. The obtained results contribute to clinical application of pulse oximetry for diagnosing CHD. 2. Materials and Methods 2.1. Search Strategy. We searched the relevant articles on the databases of Pubmed, Embase, Google Scholar, and Chinese National Knowledge Infrastructure (CNKI) up to April 2017. The following keywords were used: pulse oximetry OR SpO2 AND congenital heart disease OR CHD. The references lists of obtained articles were manually searched for eligible studies. No language restriction was applied. The studies published in abstract were not considered. 2.2. Article Selection. These obtained articles were selected according to inclusion criteria. The criteria were as follows: (a) SpO 2 was assessed with pulse oximetry; (b) SpO 2 was used to detect CHD subjects; (c) true positive (TP), false positive (FP), true negative (TN), and false negative (FN) or other data available for calculating them were reported. The review article, abstract article, and case reports were removed from the present analysis. 2.3. Data Extraction. Two authors were responsible for extracting data. The data included name of first author, year of publication, country, number of patients and healthy controls, screening time, screening limb, TP, FP, TN, and FN. The inconsistent opinion was solved with a discussion with the third author. 2.4. Statistical Analysis. Data was processed in the MetaDiSc 1.4 software. Deek s funnel plot was obtained with Stata 12.0 software. Summary sensitivity and specificity along with 95% confidence interval (95% CI) were adopted to confirm the diagnostic role of pulse oximetry screening for CHD. I 2 50% orp < 0.05 indicated significant heterogeneity. Area under curve (AUC) of summary receiver operating characteristics (SROC) was calculated to assess the diagnostic accuracy of SpO 2.Therobustnessofoverallresultswasevaluatedby sensitivity analysis. Publication bias was evaluated by Deek s funnel plot. 3. Results 3.1. Selection Process of Eligible Articles. Atotalof488relevant articles were obtained after search on the databases. The titles and abstracts were screened and 358 articles were excluded. The remaining 130 articles were provided detailed assessment and 108 articles were excluded for no available data, duplicate publication, and only SpO 2 level in CHD patients. Finally, 22 eligible articles were included in the present meta-analysis [8,10 12,14 31].Thedetailedselectionprocesswasshownin Figure 1. Basic information of each study was listed in Table 1. 3.2. Diagnostic Role of Pulse Oximetry Screening for CHD. Pulse oximetry screening showed high specificity in detecting CHD (specificity: 0.99), while having relatively low sensitivity (0.69) (Figure 2). In the analyses of sensitivity and specificity, we observed significant heterogeneity (sensitivity: p = 0.0000, I 2 = 89.6%; specificity: p = 0.0000, I 2 = 99.9%). The corresponding AUC was 0.9407, suggesting high diagnostic accuracy of pulse oximetry screening for CHD (Figure 3).

BioMed Research International 3 Table 1: Basic information of included studies. Author Year Country Limb Test timing, h Arlettaz et al. 2006 Switzerland Right or left foot 6 12 Bakr and Habib 2005 Egypt Right upper and lower limbs 31.7 (average) de Wahl Granelli et al. 2005 Sweden Right hand and one foot 12 48 Hoke et al. 2002 America Right arm and either leg 24 (average) Koppel et al. 2003 America >24 Richmond et al. 2002 UK One or other foot >2 Rosati et al. 2005 Italy 72 h (median) China Both on the right hand and on either foot 6 72 Riede et al. 2010 Germany Foot 24 72 Mathur et al. 2015 India Right upper limb and either foot 72 (median) Hu et al. 2016 China Right hand and either foot 25 (median) Jones et al. 2016 UK <24 Ozalkaya et al. 2016 Turkey Lower and right upper extremity >24 Jawin et al. 2015 Malaysia Left foot 20 (median) Ewer et al. 2011 UK Right hand and either foot In the first few hours Taksande et al. 2013 India All the four limbs Within the first 4 hours van Niekerk et al. 2016 South Africa 60 (median) de-wahl Granelli et al. 2009 Sweden Right hand and either foot 38 (median) Meberg et al. 2008 Norway Foot <12 Mo et al. 2015 China Right hand and either foot >24 Jia et al. 2016 China Right hand 24 Lu et al. 2016 China Right hand and right foot >24 Note. indicates that the information was not mentioned in the article. Sensitivity (95% CI) Arlettaz et al. 2006 1.00 (0.78 1.00) Bakr and Habib 2005 0.31 (0.09 0.61) de Wahl Granelli et al. 2005 0.77 (0.65 0.87) Hoke et al. 2002 0.75 (0.58 0.88) Koppel et al. 2003 0.60 (0.15 0.95) Richmond et al. 2002 0.89 (0.52 1.00) Rosati et al. 2005 0.67 (0.09 0.99) 0.59 (0.53 0.64) 0.84 (0.77 0.89) Riede et al. 2010 0.78 (0.52 0.94) Mathur et al. 2015 0.95 (0.76 1.00) Hu et al. 2016 0.84 (0.60 0.97) Jones et al. 2016 1.00 (0.16 1.00) Ozalkaya et al. 2016 0.60 (0.26 0.88) Jawin et al. 2015 1.00 (0.48 1.00) Ewer et al. 2011 0.75 (0.53 0.90) Taksande et al. 2013 1.00 (0.59 1.00) van Niekerk et al. 2016 0.50 (0.07 0.93) de-wahl Granelli et al. 2009 0.62 (0.42 0.79) Meberg et al. 2008 0.77 (0.60 0.90) Mo et al. 2015 0.79 (0.75 0.83) Jia et al. 2016 0.52 (0.38 0.65) Lu and Li 2016 0.24 (0.17 0.33) Specificity (95% CI) Arlettaz et al. 2006 1.00 (0.99 1.00) Bakr and Habib 2005 de Wahl Granelli et al. 2005 1.00 (0.98 1.00) Hoke et al. 2002 0.88 (0.87 0.89) Koppel et al. 2003 Richmond et al. 2002 0.99 (0.99-0.99) Rosati et al. 2005 Riede et al. 2010 Mathur et al. 2015 0.52 (0.47 0.57) Hu et al. 2016 0.44 (0.43 0.46) Jones et al. 2016 Ozalkaya et al. 2016 Jawin et al. 2015 Ewer et al. 2011 0.99 (0.99-0.99) Taksande et al. 2013 0.95 (0.94 0.96) van Niekerk et al. 2016 1.00 (0.99 1.00) de-wahl Granelli et al. 2009 Meberg et al. 2008 0.99 (0.99-0.99) Mo et al. 2015 0.86 (0.84 0.87) Jia et al. 2016 0.99 (0.99-0.99) Lu and Li 2016 0.98 (0.97 0.99) 0.2.4.6.8 1 Sensitivity Pooled sensitivity = 0.69 (0.67 to 0.72) Pooled specificity = 0.99 (0.99 to 0.99) 2 = 211.86; df =22(p = 0.0000) 2 = 23149.49; df =22(p = 0.0000) Inconsistency (I 2 ) = 89.6% 0.2.4.6.8 1 Inconsistency (I 2 )=99.9% Specificity Figure 2: Diagnostic sensitivity and specificity of pulse oximetry screening for congenital heart disease (CHD). The sensitivity and specificity were 0.69 (0.67 0.72) and 0.99 (0.99-0.99), respectively. 3.3. Sensitivity Analysis and Publication Bias Detection. Robustness of pooled results was assessed by sensitivity analysis by deleting one study each time. The analysis indicated that the pooled results were robust. Deek s funnel plot was used to assess publication bias. The funnel plot seemed to be symmetrical (Figure 4). 4. Discussion CHD is a series of heterogeneous disorders that catheter intervention or surgery is mandatory to achieve patients survival. Clinical examination shows limitation in detecting all forms of CHD [32, 33]. Heart murmurs, one of the hallmarks

4 BioMed Research International Sensitivity 1.9.8.7.6.5.4.3.2.1 SROC curve Symmetric SROC AU# = 0.9407 SE(AUC) = 0.0246 Q = 0.8784 SE(Q ) = 0.0313 0 0.2.4.6 1 Mpecificity.8 1 Figure 3: Area under curve (AUC) of SROC curve. AUC was 0.9407. Diagnostic odds ratio 1000 100 10 Log Odds Ratio versus 1/sqrt(Effective Sample Size)(Deeks) 1 0.00 0.10 0.20 1/root(ESS) Study Regression line 0.30 Figure 4: Deek s funnel plot. We found no significant publication bias in the present meta-analysis. of CHD, may be misleading or absent due to the reduced ventricular function, prolonged decline of pulmonary vascular resistance, and underlying anatomy. Although prenatal diagnosis is widely applied, a large proportion of CHD neonates are still not diagnosed before being discharged and after birth [34, 35], which may be strengthened by earlier discharge and certain postnatal care [36]. It has been thought that the application of prenatal ultrasound, clinical observation, and physical examination may be sufficient for early diagnosis of CHD [37]. This opinion may be true under specific circumstances; however, the prerequisites possibly do not exist in the majority of hospitals. Thus, a broad consensus that efficient diagnostic tool for CHD is urgently needed has been achieved. Pulse oximetry is an accurate and noninvasive test for quantification of hypoxaemia. The application of this method for diagnosing CHD is based on the theory that undetectable hypoxaemia in clinic exists in most life-threatening cases. Pulse oximetry screening for CHD has gained more attention over the last decade. It has been demonstrated to be costeffective and acceptable to the mothers [38, 39]. The existing protocol of pulse oximetry to detect CHD is restricted to 24 to 48 hours of age for neonates in well infant nursery [40]. Our meta-analysis included as many eligible articles as possible via systematic search. These obtained articles were selected carefully according to inclusion criteria. Moreover, the quality of included articles was high. Besides, the results were based on 22 eligible studies involving both Western and Asian countries. Therefore, our results were reliable. The pooled results suggested that combined sensitivity, specificity, and AUC were 0.69 (0.67 0.72), 0.99 (0.99-0.99), and 0.9407, respectively, which is similar to the previous meta-analysis [41]. Among the included articles, the results showed great differences. In the study by Mathur et al., pulse oximetry readings were taken at admission from 950 neonates and the diagnostic sensitivity, specificity, positive predictive value, and negative predictive value were 95.2%, 52.4%, 9.5%, and 99.5%. The diagnostic specificity was poor. Similarly, Hu et al. reported that diagnostic specificity of pulse oximetry screening for CHD was just 44.22%. Meanwhile, Niekerk et al. reported that the diagnostic sensitivity of pulse oximetry was merely 50%, while the specificity was 99.9%. On the contrary, Arlettaz et al. investigated the contribution of pulse oximetry to the early detection of CHD in newborns and found that the sensitivity and specificity were 100% and 99.7%, respectively. In the study of Jones et al., the estimated sensitivity and specificity were 100% and 99.8% of pulse oximetry screening

BioMed Research International 5 for diagnosing CHD. Thus, the conclusion of our analysis is significant to confirm the diagnostic role of pulse oximetry screening. However, we must acknowledge that there were limitations in the present meta-analysis. First, cut-off value, diagnosis criteria, target location, and test timing of pulse oximetry were inconsistent among the included studies, which might affect the diagnostic accuracy of pulse oximetry screening. And the significant heterogeneity in the present analysis might result from these variances. Besides, the difference in theseverityofchdalsomightinfluencetheaccuracyof pulse oximetry screening. 5. Conclusion In conclusion, pulse oximetry screening may serve as a valuable diagnostic tool with high accuracy for CHD. The diagnostic sensitivity and specificity are 0.69 (0.67 0.72) and 0.99 (0.99-0.99), respectively. Conflicts of Interest The authors declare that there are no conflicts of interest regarding the publication of this paper. Authors Contributions Caiju Du and Dianmei Liu were involved in the conception and design, or analysis and interpretation of the data; Guojing Liu was involved in the interpretation of the data; Huaixin Wang revised the manuscript critically for intellectual content. All the authors approve the final version to be published and agree to be accountable for all aspects of the work. Acknowledgments This work was supported by the Science and Technology Development Plan Project in Medicine and Health Care of Shandong Province under Grant no. 2015WS0044. References [1] R. M. Payne, M. C. Johnson, J. W. Grant, and A. W. Strauss, Toward a molecular understanding of congenital heart disease, Circulation,vol.91,no.2,pp.494 504,1995. [2] R.-K. R. Chang, M. Gurvitz, and S. Rodriguez, Missed diagnosis of critical congenital heart disease, JAMA Pediatrics,vol. 162, no. 10, pp. 969 974, 2008. [3] M. Abu-Harb, J. Wyllie, E. Hey, S. Richmond, and C. Wren, Presentation of obstructive left heart malformations in infancy., Archives of Disease in Childhood - Fetal and Neonatal Edition,vol.71,no.3,pp.F179 F183,1994. [4] C. Bull, Current and potential impact of fetal diagnosis on prevalence and spectrum of serious congenital heart disease at term in the UK, The Lancet, vol.354,no.9186,pp.1242 1247, 1999. [5] Z.-D. Du, N. Roguin, and M. Barak, Clinical and echocardiographic evaluation of neonates with heart murmurs, Acta Paediatrica,vol.86,no.7,pp.752 756,1997. [6] J. Gregory, A. Emslie, J. Wyllie, and C. Wren, Examination for cardiac malformations at six weeks of age, ADC - Fetal and Neonatal Edition,vol.80,no.1,pp.F46 F48,1999. [7]B.Varan,K.Tokel,andG.Yilmaz, Malnutritionandgrowth failure in cyanotic and acyanotic congenital heart disease with and without pulmonary hypertension, Archives of Disease in Childhood,vol.81,no.1,pp.49 52,1999. [8] R. I. Koppel, C. M. Druschel, T. Carter et al., Effectiveness of pulse oximetry screening for congenital heart disease in asymptomatic newborns, Pediatrics, vol. 111, no. 3, pp. 451 455, 2003. [9] J. Reich, S. Miller, and B. Brogdon, The use of pulse oximetry to detect congenital heart disease, ACC Current Journal Review, vol. 12, no. 4, p. 92, 2003. [10] S.Richmond,G.Reay,andM.AbuHarb, Routinepulseoximetryintheasymptomaticnewborn, ADC - Fetal and Neonatal Edition,vol.87,no.2,pp.F83 F88,2002. [11] A. de-wahl Granelli, M. Wennergren, K. Sandberg et al., Impact of pulse oximetry screening on the detection of duct dependent congenital heart disease: a Swedish prospective screening study in 39,821 newborns., BMJ (Clinical research ed.),vol.338,p.a3037,2009. [12] F. T. Riede, C.Wörner, I. Dähnert, A. Möckel, M. Kostelka, and P. Schneider, Effectiveness of neonatal pulse oximetry screening for detection of critical congenital heart disease in daily clinical routine-results from a prospective multicenter study, European Pediatrics,vol.169,no.8,pp.975 981,2010. [13]C.P.F.O Donnell,C.Omar,F.Kamlin,P.G.Davis,J.B. Carlin, and C. J. Morley, Clinical assessment of infant colour at delivery, ADC - Fetal and Neonatal Edition,vol.92,no.6,pp. F465 F467, 2007. [14] R. Arlettaz, A. S. Bauschatz, M. Mönkhoff, B. Essers, and U. Bauersfeld, The contribution of pulse oximetry to the early detection of congenital heart disease in newborns, European Pediatrics, vol. 165, no. 2, pp. 94 98, 2006. [15] A. F. Bakr and H. S. Habib, Combining pulse oximetry and clinical examination in screening for congenital heart disease, Pediatric Cardiology,vol.26,no.6,pp.832 835,2005. [16] T.R.Hoke,P.K.Donohue,P.K.Bawaetal., Oxygensaturation as a screening test for critical congenital heart disease: A preliminary study, Pediatric Cardiology,vol.23,no.4,pp.403 409, 2002. [17] E. Rosati, G. Chitano, L. Dipaola, C. De Felice, and G. Latini, Indications and limitations for a neonatal pulse oximetry screening of critical congenital heart disease, Perinatal Medicine,vol.33,no.5,pp.455 457,2005. [18] A. de Wahl Granelli, M. Mellander, J. Sunnegårdh, K. Sandberg, and I. Ostman-Smith, Screening for duct-dependant congenital heart disease with pulse oximetry: a critical evaluation of strategies to maximize sensitivity., Acta paediatrica (Oslo, Norway : 1992),vol.94,no.11,pp.1590 1596,2005. [19] Q. M. Zhao, X. J. Ma, X. L. Ge et al., Pulse oximetry with clinical assessment to screen for congenital heart disease in neonates in China: a prospective study, Lancet, vol. 384, no. 9945, pp. 747 754, 2014. [20] N. B. Mathur, A. Gupta, and S. Kurien, Pulse oximetry screening to detect cyanotic congenital heart disease in sick neonates in a neonatal intensive care unit, Indian Pediatrics,vol. 52,no.9,pp.769 772,2015. [21] X.-J. Hu, Q.-M. Zhao, X.-J. Ma et al., Pulse oximetry could significantly enhance the early detection of critical congenital heart disease in neonatal intensive care units, Acta Paediatrica, vol. 105, no. 11, pp. e499 e505, 2016.

6 BioMed Research International [22] A. J. Jones, C. Howarth, R. Nicholl, E. Mat-Ali, and R. Knowles, The impact and efficacy of routine pulse oximetry screening for CHD in a local hospital, Cardiology in the Young, vol.26, no.7,pp.1397 1405,2016. [23] E. Özalkaya, A. Akdağ, I. Şen, E. Cömert, and H. Melek Yaren, Early screening for critical congenital heart defects in asymptomatic newborns in Bursa province, The Maternal- Fetal and Neonatal Medicine,vol.29,no.7,pp.1105 1107,2016. [24] V. Jawin, H.-L. Ang, A. Omar, and M.-K. Thong, Beyond critical congenital heart disease: Newborn screening using pulse oximetry for neonatal sepsis and respiratory diseases in a Middle-Income Country, PLoS ONE,vol.10,no.9,ArticleID e0137580, 2015. [25] A. K. Ewer, L. J. Middleton, A. T. Furmston et al., Pulse oximetry screening for congenital heart defects in newborn infants (PulseOx): A test accuracy study, The Lancet, vol. 378, no. 9793, pp. 785 794, 2011. [26] A. M. Taksande, B. Lakhkar, and A. Gadekar, Accuracy of pulse oximetry screening for detecting critical congenital heart disease in the newborns in rural hospital of Central India, Images Paediatr Cardiol,vol.15,no.4,pp.5 10,2013. [27]A.M.vanNiekerk,R.M.Cullis,L.L.Linley,andL.Zühlke, Feasibility of pulse oximetry pre-discharge screening implementation for detecting critical congenital heart lesions in newborns in a secondary-level maternity hospital in the Western Cape, South Africa: The POPSICLe study, South African Medical Journal,vol.106,no.8,pp.817 821,2016. [28] A. Meberg, S. Brügmann-Pieper, R. Due Jr. et al., First Day of Life Pulse Oximetry Screening to Detect Congenital Heart Defects, Pediatrics,vol.152,no.6,pp.761 765,2008. [29]Z.D.Mo,X.L.Guo,andW.J.Wang, Effectivenessofpulse oximetry for the screening of congenital heart disease in newborn infants, Chinese Neonatology, vol. 30, no. 5, pp. 343 346, 2015. [30] A. Q. Jia, J. H. Wu, and A. Y. Guo, Analysis of intervals of pulse oximetry in congenital heart disease screening, JClinPediatr, vol. 34, no. 5, pp. 357 359, 2016. [31] X. Y. Lu and Y. Li, Comparison between pulse oximetry and traditional method in screening congenital heart disease, Chinese Clinical Rational Drug Use,vol.9,no.10A,pp. 152-153, 2016. [32] S. B. Ainsworth, J. P. Wyllie, and C. Wren, Prevalence and clinical significance of cardiac murmurs in neonates, ADC - Fetal and Neonatal Edition,vol.80,no.1,pp.F43 F45,1999. [33] C. Wren, S. Richmond, and L. Donaldson, Presentation of congenital heart disease in infancy: Implications for routine examination, ADC - Fetal and Neonatal Edition, vol. 80, no. 1, pp.f49 F53,1999. [34]R.-K.R.Chang,S.Rodriguez,andT.S.Klitzner, Screening newborns for congenital heart disease with pulse oximetry: Survey of pediatric cardiologists, Pediatric Cardiology, vol.30, no. 1, pp. 20 25, 2009. [35] C. Wren, Z. Reinhardt, and K. Khawaja, Twenty-year trends in diagnosis of life-threatening neonatal cardiovascular malformations, ADC - Fetal and Neonatal Edition, vol. 93, no. 1, pp. F33 F35, 2008. [36] M. Mellander and J. Sunnegårdh, Failure to diagnose critical heart malformations in newborns before discharge - An increasing problem? Acta Paediatrica, vol. 95, no. 4, pp. 407 413, 2006. [37]D.M.Sendelbach,G.L.Jackson,S.S.Lai,D.E.Fixler,E.K. Stehel, and W. D. Engle, Pulse oximetry screening at 4 hours of age to detect critical congenital heart defects, Pediatrics,vol. 122, no. 4, pp. e815 e820, 2008. [38] A. Ewer, A. Furmston, L. Middleton et al., Pulse oximetry as a screening test for congenital heart defects in newborn infants: a test accuracy study with evaluation of acceptability and costeffectiveness., Health Technology Assessment, vol. 16, no. 2, 2012. [39] R. Powell, H. M. Pattison, A. Bhoyar et al., Pulse oximetry screening for congenital heart defects in newborn infants: An evaluation of acceptability to mothers, ADC - Fetal and Neonatal Edition,vol.98,no.1,pp.F59 F63,2013. [40] A.R.Kemper,W.T.Mahle,G.R.Martinetal., Strategiesfor implementing screening for critical congenital heart disease, Pediatrics,vol.128,no.5,pp.e1259 e1267,2011. [41] S. Thangaratinam, K. Brown, J. Zamora, K. S. Khan, and A. K. Ewer, Pulse oximetry screening for critical congenital heart defects in asymptomatic newborn babies: A systematic review and meta-analysis, The Lancet, vol. 379, no. 9835, pp. 2459 2464, 2012.

MEDIATORS of INFLAMMATION The Scientific World Journal Gastroenterology Research and Practice Diabetes Research International Endocrinology Immunology Research Disease Markers Submit your manuscripts at https://www.hindawi.com BioMed Research International PPAR Research Obesity Ophthalmology Evidence-Based Complementary and Alternative Medicine Stem Cells International Oncology Parkinson s Disease Computational and Mathematical Methods in Medicine AIDS Behavioural Neurology Research and Treatment Oxidative Medicine and Cellular Longevity