INSTRUKTION MANUAL NANO NEBULIZER PISTON COMPRESSOR Diagnosis S.A. ul. Gen. W. Andersa 38A Białystok, Poland

Similar documents
Compressor Nebulizer. Model UN-014

HCS70004P, HCS70004G HCS70004C, HCS70004BL HCS70004DP, HCS70004 AEROMIST COLORS NEBULIZER COMPRESSOR KIT. Instruction Manual

PORTABLE COMPRESSOR NEBULIZER Guidebook & Manual Reorder No. 5606

Medescan Nebuliser Med - S600A Instructions Manual

Compressor Nebulizer. Guidebook

Compressor Nebulizer System

Compressor Nebulizer. Guidebook

Compressor Nebulizer Instruction Manual Part No.: 5055

COMPRESSOR NEBULIZER MODEL: JB USER S MANUAL

INSTRUCTION MANUAL. Compressor Nebulizer System NE-C25

A82-LIFE-Dog Nebulizer NBB02

DeVilbiss Pulmo-Aide Compact Compressor/Nebulizer Instruction Guide

DeVilbiss Pulmo-Aide Compressor/Nebulizer Instruction Guide

COMPRESSOR NEBULIZER MODEL: JB USER S MANUAL

Portable Equine Nebuliser System. User Manual

Mini-Mist. Owner s Manual. portable nebulizer. PMM241 Rev.3 12/10

Instruction Manual Aerogen, Inc. Part No. AG-AL1010 Rev. C

Effective Date: August 31, 2006 SUBJECT: CARE AND USE OF NEBULIZER AND INTERMITTENT POSITIVE PRESSURE BREATHING DEVICE

Nebu 100s Series. Instruction Manual

Flex-Point HEATED MUSCLE MASSAGER

IMPORTANT SAFETY INSTRUCTIONS

Phonak PilotOne II. User Guide

This page has been intentionally left blank

MESH NEBULIZER INSTRUCTION MANUAL. Model NE-U22

Instructions for Use REMOTE CONTROL RC-N

Compact Compressor Nebulizer

RC-S. Remote Control INSTRUCTIONS FOR USE

/ INSTRUCTIONS FOR USE

Joint Relaxation Heated Vibration Massager. Knee and Joint Physiotherapy Unit

Care and Use of Nebulisers

PARI INSTRUCTIONS FOR USE COMPRESSOR NEBULIZER SYSTEM. Portable Power Kit (optional)

Nebulizer with Micromesh Technology GT Instruction manual

Nebulizer Compressor Item JB User Manual. Read this manual before operating the nebulizer. Save this manual for future reference.

Comfortmax Thermax Hot/Cold Water Circulation System. Instructions for Use

深圳市百瑞琪仪器设备有限公司. Shenzhen Bi-Rich Medical Devices Co.,Ltd. BR-CN151 Audio Compressor Nebulizer

IMPACT Pro R Instructions and guidance

Charging base guide. A Sonova brand

Chair Massager With Heat Instructions

PilotOne II. User Guide

Custom instruments. Insio primax User Guide. Hearing Systems

Compatible with the following products TENS. Replacement Pads. User Manual. Last revised: V Last revised: V

Nebulizer Compressor Item JB User Manual. Read this manual before operating the nebulizer. Save this manual for future reference.

Invacare Pediatric Bear Nebulizer System

Identification. Eartips. Battery Door. 2-Position Switch. Eartip. Removal Cord. Flexible Neck Cord Protective Case. Filter Tool and Extra ACCU Filters

! Caution (continued)

Motorised Treadmill Model No: JI1625

TOWN OF FAIRFIELD HEALTH DEPARTMENT PUBLIC HEALTH NURSING

INSTALLATION MANUAL. VIDEO Camera, Probe and Lightsource OTOSCOPES.

Nebulizer Compressor Item JB User Manual. Read this manual before operating the nebulizer. Save this manual for future reference.

Getting started with Wireless Fitting

Manufacturer: Phonak AG Laubisrütistrasse 28 CH-8712 Stäfa Switzerland / /A+W Printed in Switzerland

Safety manual for hearing instruments

ER75 Electro-Acoustic Ear Simulator. Operating Manual

TH CROS RIC 19 Transmitter. User Guide

IMPORTANT SAFETY INSTRUCTIONS

In-the-ear hearing system. User Guide

MEET THE DRIVING FORCE IN NEBULIZER THERAPY... Matched Nebulizer Compressor System

INSTRUCTION MANUAL. Compressor Nebulizer NE-C801

Operating Instructions SaHoMa -II mobile ultrasonic nebulizer

Wireless accessories. Remote Control 3.0. Instructions for use

Lily 说明书 尺寸 56*118mm 封面 250G白卡白珠光 烫银 内页 128g双铜纸四色印刷 装订方式 胶装 Tas i, L i! Lily 说明书

Gronk Massager STRONG, DEEP-KNEADING RELIEF FOR HARD-TO-REACH MUSCLES

...Porta-neb matched systems from Respironics (UK)

The Biomet EBI Bone Healing System. Patient Manual

Table of Contents. Introduction Indications For Use Contraindications Warnings Precautions...5

EVER Pharma D-mine Pen Pen injector for Apomorphine 10 mg/ml

InsuPad User Manual. Charger Base The charger base can be connected by the attached USB cable to the power adapter supplied for charging.

TH Premium IF 19. User Guide

ProPocket TM. User Guide

TH Premium RIC LI 19. Guide for Hearing Care Professionals

Alta2 Nera2. designrite INSTRUCTIONS FOR USE. Alta2, Alta, Nera2, Nera, Ria2, Ria

English. Fealite Nasal Pillows System. User Manual

HDM 330/500 Operating instructions Mode d emploi Manual de instrucciones Manual de instruções

For more information visit or contact hearx:

Cordless 3D Shiatsu Massager DEEP RELIEF. WIDE COVERAGE. FEELS LIKE THE HANDS OF A MASSEUSE

Cushion Massager with heat

Cordless 7-in-1 Super Massager WITH INTERCHANGEABLE NODES FOR 7 MAGNIFICENT MASSAGE TECHNIQUES

I N S T R U C T I O N S F O R U S E CIC, MIC, ITC, ITE

OWNER'S MANUAL OxyTrainer

Allergy Reliever. User Manual. AR 1 Series

Body Fat Scales. instruction manual. with body analysis functionality. Model Number BF801-BK SL/BL

Dear HighQ Check System Owner :

Contents Safety Instructions... 4 Package Contents... 6 Notes on Measurements... 7

Unitron remote control guide

SLEEP IMPROVING WRISTBAND. Item No Owner s Guide

Dip Station G

ENGLISH. USER MANUAL Register your product at: multi-action toothbrush

the range USER MANUAL

Active Sport Massager WITH 3 INTERCHANGEABLE MASSAGE NODES

TABLE OF CONTENTS. 2.0 USING MANAGER PREFERENCE 2.1 Using manager preference mode 2.2 Selecting US or metric 2.3 Using CSafe

THE TOTALLY WICKED TORNADO NC CONGRATULATIONS ON PURCHASING YOUR TOTALLY WICKED TORNADO NC ELECTRONIC CIGARETTE KIT

PICTOR USER GUIDE V.1.0.R7

Deep Kneading Shiatsu Massager EXTRA-POWERFUL SHIATSU RELIEF

PROFESSIONAL CARE OXYJET air. powered. Downloaded from manuals search engine. filter. OxyJet

Pure 312 Nx, Pure 13 Nx

Essential oils,sandalwood (cas ) MSDS

estore relief soothing restore relief soothing restore restore relief soothing Instruction Booklet Touch n Tone TM MASSAGER

Remote control 2 guide

Transcription:

INSTRUKTION MANUAL EN NANO NEBULIZER PISTON COMPRESSOR 0197 Diagnosis S.A. ul. Gen. W. Andersa 38A 15-113 Białystok, Poland www.diagnosis.pl 7015

DEAR CUSTOMER Thank you for buying the Diagnostic NANO nebulizer intended for inhalation. Inhalation therapy is an efficient and safe method of treatment of respiratory diseases. The treatment should be started after consultation with your doctor. Before the first use of the nebulizer, carefully read the instructions for use. 1. INTRODUCTION Diagnostic NANO is a compact medical device intended for intermittent use (30 min operation, 30 min break). Enables the medical preparation to immediately and directly reach bronchial tubes and lungs as an inhalant. If proper precautionary measures are taken, the nebulizer will ensure appropriate treatment quality for a number of years. 2. OPERATING DESIGN IThe Diagnostic NANO nebulizer was designed for the treatment of asthma, allergies and other respiratory diseases. The device produces a stream of air led through the air tube to the nebulizer. After the introduction of air into the nebulizer, the medical preparation will be atomized, creating an easily-inhaled mist. The main flow of air into and out of the compressor is presented in Figure 1. We advise you to read this user manual in order to familiarize yourself with the characteristics of the product. The product should not be used for other than the intended purpose. MAIN UNIT (COMPRESSOR) 3. PRODUCT FEATURES Before using the device, make sure that the following elements are included: 1 x main unit (compressor) 1 x nebulizer (spray nozzle) 1x air tube 1x mask for adulst, mask for children 1x mouthpiece 5x air filter air filter air outlet inlet outlet compressor Fig.1 Power cord On/off switch Vents Air tube Nebulizer (spray nozzle) Fig.2 Mouthpiece Mask for adults Filter and filter cover Mask for children DiSTRIBUTION of particles size in comliance with EN 13544-1 Łączna ilość rozpylanych cząsteczek w % SPECIFICATION OF PARTICLES` SIZE ATOMIZED BY THE NEBULIZER Data may vary depending on medicine used. Measurements done by using MALVERN technology and physiological salt solution. Data provided may not be applicable to suspended solids or to high viscosity medicine therefore, in case of using medicine different than physiological salt solution please contact the manufacturer of the medicine used for inhalation Air tube socket Wielkość rozpylanych cząsteczek (µm) 1

AEROSOL THERAPY SYSTEM FOR HOME USE IMPORTANT WARNINGS: failure to comply with the warnings and the contents of this user manual may expose the user to such risks as: deterioration of health, burns, electric shock, recurrent infections, death, fire, environmental pollution. The use of any electrical appliance requires observance of the following fundamental principles: The device contains small parts that may be dangerous if BEFORE USING THE APPLIANCE CAREFULLY READ THIS USER MANUAL swallowed. Do not use the device in combination with accessories Failure to follow the user manual relating to the operation of the device may cause malfunction. Follow the electrical specifications stated on the device; Concerning the issues related to the medication type, dosage and usage schedule, the advice of a doctor must be followed. The intensity of use of the device must always be consulted with a physician. The inhalation of any substances should be carried out with permission of the doctor who decides on doses and usage; Lack of electricity supply, sudden failure or other adverse conditions may cause device malfunctions; for that reason, it is recommended to equip oneself with a device or medication (in accordance with the doctor's recommendations) that could be used interchangeably; Use the device only for its intended purpose (aerosol inhalation) and in the manner described in this manual; Do not allow the device to be used by children or persons with disabilities without supervision; in such cases, special care must be taken during therapy; Do not use the device if the plug or power cord are damaged. Never leave the device, its accessories, or packaging within the reach of children, unauthorized persons, pet animals or insects. If necessary, use adapters or extension cords compliant with the applicable safety standards, paying attention not to exceed the limit of current load and reducing the maximum power indicated on the adapter; Do not use the device with wet or damp hands, do not use it while showering; Never allow the device to be immersed in water or other liquids; do not use the device if it has been accidentally soaked; The device and the power cable should be located away from heat sources; The device is equipped with liquid ingress protection; Do not use the device near flammable objects or explosives; Do not use the device in an environment where sprays have recently been used; ventilate the room before starting therapy; Do not block or clog vents on the device, do not place it on a soft surface such as a bed or couch; Do not insert any objects into the vents; Do not use the device if it emits abnormal sounds; In order to prevent deterioration of the device's functionality, use only original accessories; Accessories should only be used by one person. In the event that the nebulizer is used by other family members, we recommend cleaning the them before each use or purchase additional accessories intended for individual use only. Do not pull the power cord or the appliance itself in order to remove the plug from the electrical outlet. After use, always unplug the power plug. Do not expose the device to atmospheric conditions; store it at room temperature. Do not put the air tube or the power cord on the neck other than those described in the user manual. Do not use the device if it was stored in a polluted environment. RESPONSIBILITY In terms of safety, efficiency and reliability, the responsibility lies with the manufacturer only if: assembly, calibration, repairs or modifications are carried out by authorized persons; electrical installation complies with the applicable standards; the user manual has been complied with. The manufacturer shall not be liable for the improper, incorrect or unreasonable use of the appliance. FILTER REPLACEMENT The filter should be replaced every 3 months or every 300 uses. To replace the air filter, follow these steps: 1. Remove the filter cover. 2. Replace the filter. 3. Reinstall the filter cover. NOTE: If the filter becomes wet, it should be completely dried before use FUSE There is a fuse installed inside the device, which must be replaced in the event of a failure. The replacement operation, as well as other possible repairs, should be carried out by the manufacturer or its authorized personnel ELECTROMAGNETIC COMPATIBILITY The device complies with current standards regarding electromagnetic compatibility and is suitable for use in all buildings, including residential buildings. The radio frequency emissions level of the device is very low and does not cause interference with devices that are located nearby. However, it is recommended that you do not position it on or near other devices. If interferences with other electrical equipment occur, the device should be moved away or plugged into another electrical outlet. Radio communication equipment may affect the operation of the device. Keep it at a distance of at least 3 m from the device. 2

4. PREPARING FOR NEBULIZATION BUDOWA nebulizatora (rozpylacza) To pour the medication into the nebulizer, perform the following steps: 1. Open the medication cup. 2. Fill the medication cup in accordance with the doctor's recommendations. Make sure that the medication does not exceed the volume of 8 ml. 3. Close the medication cup. 4. Attach the mouthpiece or mask. maska pokrywa pojemnika na lek ustnik dysza pojemnik na lek max 6 4 2 NOTE: The mouthpiece is designed for oral inhalation. NOTE: Remember to use the baffle in the nebulizer (in the medication cup). It is necessary for the atomizer to work properly przewód powietrzny Rys.3 5. HOW TO USE THE DEVICE To start the nebulizer, perform the following steps: Connect the plug of the compressor to the mains, but do not connect the nebulizer. Make sure that the air filter is clean and located in the device. Connect the nebulizer using air tube with the main unit (compressor). Turn on the power. Start inhalation in accordance with the doctor's recommendations. Nebulizer operation time: 30 min operation / 30 min break. After 30 minutes of operation, the device should be turned off in order to prevent compressor overheating. Inhalation is completed when there is no more medicament in the cup. After the end of the inhalation, turn off the unit before disconnecting the air tube 6. CLEANING All activities associated with cleaning of the device should be carried out with the plug removed the from the electrical outlet. In order to avoid incorrect operation of the device and inhaling undesirable substances, clean the appliance and the accessories immediately after finishing therapy. Cleaning the appliance should be carried out using a cloth slightly moistened with alcohol. After cleaning the equipment, before using it, wait for the alcohol to evaporate completely. Do not use other liquids or cleaning agents. Do not use cloths that are too wet, since the contact of liquids with electrical devices may lead to operational failure, irreversible damage, as well as can be dangerous to your health. 7. CLEANING THE ACCESSORIES 1. Before the first use and after using the device each time, thoroughly clean the accessories (except the power cord) with lukewarm water and dry with a cloth. They should be subsequently placed in a clean place. During cleaning, make sure to have removed any residues of substances intended for inhalation and then thoroughly dry the accessories. In any event, do not use cleaning and disinfecting agents that may be toxic in contact with the skin or mucous membranes, when swallowed or through inhalation. Any moisture remaining in the air tubes should removed by starting the unit for a few minutes without the nebulizer connected. If the air tube is dirty, it must be replaced. 3

2. In order for the nebulizer to work properly, after each use do the following: After you disconnect the air tube remove the mouthpiece or mask. Open the cap of the medication cup and empty it. Wash your nebulizer under running water or leave it in warm water for 15 minutes. To achieve a better cleaning effect, add some vinegar to water (a mixture of 1/3rd white vinegar, 2/3rds water). NOTE: Do not boil the accessories as this may result in damage. Do not put nebulizer accessories into the dishwasher. Prior to an extended storage period, the accessories should be completely dried. 8. SPARE ACCESSORIES To ensure effective inhalation, proper operation and safety, use only original accessories available for purchase. All the accessories are suitable for direct contact with the skin. They do not contain phthalates. In case of doubt, please contact Diagnosis. Free infoline 800 70 30 11. 9. PERIODIC INSPECTIONS To maintain high effectiveness of the device, it is recommended to replace the medication cup every 12 months (every 6 months if the cup is used 3 times a day). A damaged medication cup should be replaced immediately. 10. TECHNICAL SPECIFICATION Power Supply Power Nebulizer flow Compressor flow Pojemność pojemnika na lek Particle size MMAD Nebulization efficiency Deposit quantity Working time Noise level Operating conditions Storage conditions Dimensions (L x W x H) Device weight Accessories NOTE! 230V AC, 50Hz below 65W 6l 8-10 l 2-8 ml 0,5 do 6 microns 2,44 microns 0,4 ml/min 0,15 ml 30 min operation / 30 min break below 55 dba 10 C do 40 C (50 F do 104 F) 10% to 95% relative humidity -20 C do 70 C (-4 F do 158 F) 10% to 95% relative humidity 167 x 142 x 93 mm 1,51 kg Nebulizer, mask (large), mask (small), 5 x air filter, air tube 150 cm, mouthpiece Specifications may change without prior notice. 4

11. PrOBLEM solving. Problem Nebulizer does not start Too noisy Probable cause The main switch of the compressor is switched off. No air filter in the device. Nebulizer has not been cleaned after previous use. Water drops are formed in the air tube. Solution Turn on the device. Install filter in the unit. Clean the device. Air tube is folded. Remove any bends or knots of the air tube. Filter is clogged. No medication. Replace the filterr. Add the appropriate quantity of drug prescribed by the doctor to the medication cup. Adjust the amount of the drug in the cup to the proper level, connect the air tube to the compressor and turn on the device. The remaining moisture in the air tubing can be removed by starting the unit for a few minutes without connecting the nebulizer. If the air tube is dirty, it must be replaced. Too much medication added. Air tube has not been dried. NOTE: If you experience any other problems with the device, contact Diagnosis. Free hotline: 800 70 30 11 12. EXPLANATION OF SYMBOLS USED: Rev. Important warnings Prąd zmienny Fuse Manufacturer Date of the last review IP21 SN Serial number Catalog Number Produciton year Isolation Class II Urządzenie Typ BF Read the user Protect before moisture WARING! Portable wireless communication equipment may affect MEDICAL ELECTRICAL DEVICES Working in the vicinity (within a distance of 2.8 m) of a cellular phone may cause instability of the stimulator. WARING! Specifications required by the abovementioned standard are available at http://www.diagnosis.pl/norma-en-60601-1-2.html Degree of protection The worn out product should be taken to a waste collection facility. Contains components that are dangerous for the environment. The correct disposal of the device allows to preserve valuable resources and avoid negative effects on health and the environment, which may be threatened by inappropriate handling of waste. Rev. 2016.09.13 przygotowana na podstawie polskiej wersji językowej z dnia 2015.09.04 an 5

NOTES

EN Diagnosis S.A. ul. Gen. W. Andersa 38A, 15-113 Białystok, Poland www.diagnosis.pl diagnosis@diagnosis.pl store stamp and signature of salesperson WARRANTY CARD DEVICE NAME.................................... MODEL........................ SERIAL NUMBER.................................. DATE OF SALE.................. WARRANTY TERMS 1. Diagnosis S.A. grants a warranty: 2 years for DIAGNOSTIC NANOnebulizer (excluding accessories) 12 months for nebulizer accessories Hardware defects revealed during the warranty period will be rectified free of charge within 21 days. The term runs from the date of delivery of the equipment to the service center. 2. The purchaser shall be entitled to replace the equipment for a new one, free of defects, when: the repair has not been made within the time limit set in item 1 an authorized service center found an irreparable manufacturing defect during the warranty period, 4 repairs were effected, and the equipment still shows defects that prevent its use in accordance with its intended purpose. The concept of repair shall not include operations related to equipment check and cleaning. 3. The warranty shall not cover: batteries, products with illegible or damaged serial number, damage due to the operation and storage inconsistent with the user manual, ingress of liquids or foreign bodies, overvoltage of mains, repairs by unauthorized persons and random events. 4. Faulty equipment should be delivered by the buyer to the address of the main service center or one of the Authorized Service Centers (listed in the appendix). 5. The warranty for the sold consumer goods shall not exclude, restrict, or suspend the powers of the buyer resulting from non-conformity of the goods with the contract. 6. The only basis for the warranty rights shall be the warranty card with the date of sale, stamp and signature of the salesperson. If the card is not completed, filled in wrongly, with traces of corrections and entries made by unauthorized persons, illegible as a result of damage - it shall be invalid.