FreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in Glycemic Clamp Protocols

Similar documents
FreeStyle Lite A Blood Glucose Meter That Requires No Coding

Accuracy and Reliability of the Nova StatStrip Glucose Meter for Real-Time Blood Glucose Determinations during Glucose Clamp Studies

Bern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D.

Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors

Antisense Mediated Lowering of Plasma Apolipoprotein C-III by Volanesorsen Improves Dyslipidemia and Insulin Sensitivity in Type 2 Diabetes

Supplementary Appendix

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers

Validity and Reliability of a Glucometer Against Industry Reference Standards

Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System

A New Consensus Error Grid to Evaluate the Clinical Significance of Inaccuracies in the Measurement of Blood Glucose

Comparison of the accuracy of the hemocue glucose analyzer with the Yellow Springs Instrument glucose oxidase analyzer, particularly in hypoglycemia

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System

Relationship between glucose meter error and glycemic control efficacy

GlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data

Clinical Evaluation for. Embrace No Code Blood Glucose Monitoring System

Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System

Use of Continuous Glucose Monitoring in the Detection and Prevention of Hypoglycemia

Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements

Standard Operating Procedures Clinical and Translational Research Center

Clinical Impact of Prandial State, Exercise, and Site Preparation on the Equivalence of Alternative-Site Blood Glucose Testing

Ultra Blue. Test Strips UNISTRIP. Charlotte, NC V2-10/25/2018 TECHNOLOGIES,LLC

Performance Evaluations of Continuous Glucose Monitoring Systems: Precision Absolute Relative Deviation Is Part of the Assessment

Accuracy of the 5-Day FreeStyle Navigator Continuous Glucose Monitoring System. Comparison with frequent laboratory reference measurements

Hypoglycemia Detection and Prediction Using Continuous Glucose Monitoring A Study on Hypoglycemic Clamp Data

Aina Blood Monitoring System

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard

Today s Goals 10/6/2017. New Frontiers in Diabetes Technology. Disclosures

Agreement between Glucose Trends Derived from Three Simultaneously Worn Continuous Glucose Sensors

Determination of Ethanol in Breath and Estimation of Blood Alcohol Concentration with Alcolmeter S-D2

Subjects and Methods 2017, 64 (8),

Diabetes II Insulin pumps; Continuous glucose monitoring system (CGMS) Ernest Asamoah, MD FACE FACP FRCP (Lond)

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

Insulin Tolerance Test Protocol - RNS Endocrinology

Diabetologia 9 Springer-Verlag 1984

Early Patient and Clinician Experiences with Continuous Glucose Monitoring

Evolution of Analytical Performance in Portable Glucose Meters in the Last Decade

Anomaly Detection in the Artificial Pancreas

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System

The Realities of Technology in Type 1 Diabetes

Exenatide Treatment for 6 Months Improves Insulin Sensitivity in Adults With Type 1 Diabetes

Simone Del Favero, PhD, Andrea Facchinetti, PhD, Giovanni Sparacino, PhD, and Claudio Cobelli, PhD

Computing the Surveillance Error Grid Analysis: Procedure and Examples

Report. Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013

Objectives. Glucose Monitoring Today. Tracking Glucose ISO Standards. Continuous Glucose Monitoring & Diabetes Management 3/7/2015

Better Diagnostics for Life

Evaluation of portable blood glucose meters using canine and feline pooled blood samples

Preventing Hypoglycemia Using Predictive Alarm Algorithms and Insulin Pump Suspension

Efficacy/pharmacodynamics: 85 Safety: 89

Cross-Matches for Bioequivalence Evaluation Division using Needle Free Jet Injector (Comfort-In) and conventional Pen type syringe.

The Current Environment of CGM Technologies. Barry H. Ginsberg, M.D., Ph.D.

Closing the loop, squaring the circle: Studies on insulin delivery, glucose monitoring and the artificial pancreas Luijf, Y.M.

Accuracy of the SEVEN Continuous Glucose Monitoring System: Comparison with Frequently Sampled Venous Glucose Measurements

Performance of the Medtronic Sentrino continuous glucose management (CGM) system in the cardiac intensive care unit

Glucagon secretion in relation to insulin sensitivity in healthy subjects

Non-invasive blood glucose measurement by near infrared spectroscopy: Machine drift, time drift and physiological effect

Continuous Glucose Monitoring and the Reality of Metabolic Control in Preschool Children With Type 1 Diabetes

EVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:2003 ACCURACY AND USER EVALUATION REQUIREMENTS

Continuous Glucose Monitoring Considerations for the Development of a Closed-Loop Artificial Pancreas System

Screening of Elevated Glucose Levels in Gingival Crevice Blood Using a Novel, Sensitive Self-Monitoring Device

CareLink. software REPORT REFERENCE GUIDE. Management Software for Diabetes

NICO /14 Nipro Diagnostics, Inc.

Advances in Diabetes Care Technologies

Clinical Implications and Economic Impact of Accuracy Differences among Commercially Available Blood Glucose Monitoring Systems

Dr Sander Zwart MD, FACP, ECNU Arizona-AACE 2016 Annual Meeting

Glycemic Variability:

Statistical Approach of Assessing the Reliability of Glucose Sensors: The GLYCENSIT Procedure

Advances in Diabetes Care Technologies

GD50. Simple. Accurate. Reliable. Test Simply. Live Better. All FORA Diamond Series Blood Glucose Monitoring Systems use FORA Diamond Test Strips.

Report Reference Guide

Accuracy Requirements for a Hypoglycemia Detector: An Analytical Model to Evaluate the Effects of Bias, Precision, and Rate of Glucose Change

NEW TECHNOLOGIES FOR MANAGING DIABETES ANGELA THOMPSON DNP, FNP-C, BC-ADM, CDE, FAANP

Reconstruction of Glucose in Plasma from Interstitial Fluid Continuous Glucose Monitoring Data: Role of Sensor Calibration

Smart & Simple Mobile Solutions.

INSULIN INITIATION AND INTENSIFICATION WITH A FOCUS ON HYPOGLYCEMIA REDUCTION

Subject Index. Breastfeeding, self-monitoring of blood glucose 56

Clinical Performance of the TRUE2go Blood Glucose Monitoring System Exceeds Minimum Criteria for Accuracy Using ISO 15197:2013

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

Performance of three portable blood glucose monitoring devices used in a veterinary application

The Benefits of Continuous Glucose Monitoring and a Glucose Monitoring Schedule in Individuals with Type 1 Diabetes during Recreational Diving

Imaging Core Laboratory Standard Operating Procedure

Since the discovery of insulin, the management of type 1 diabetes

Learning Objectives. At the conclusion of this module, participants should be better able to:

Pumps & Sensors made easy. OPADA ALZOHAILI MD FACE Endocrinology Assistant Professor Wayne State University

Blood Glucose Monitoring System. Owner s Booklet

Advances Towards the Bionic Pancreas.

Diabetes: Definition Pathophysiology Treatment Goals. By Scott Magee, MD, FACE

Advances in Diabetes Care Technologies

Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood Glucose Meter

Norbert Hermanns, PhD 1,2, Beatrix Schumann, MD 2, Bernhard Kulzer, PhD 1,2, and Thomas Haak, MD 1,2. Original Article

Updates in Diabetes Technology

The Commission on Dental Accreditation s

Somatostatin is an endogenous peptide and neurotransmitter

Figure 2.1: Glucose meter

Performance Evaluation of B. Braun Omnitest 5 Blood Glucose Monitoring System for Self-Monitoring of Blood Glucose

Quality Control of Self-Monitoring of Blood Glucose: Why and How?

Accuracy Evaluation of Five Blood Glucose Monitoring Systems: The North American Comparator Trial

Glucose, glucose oxidase/peroxidase method, photometry, diabetes mellitus.

Transcription:

Journal of Diabetes Science and Technology Volume 2, Issue 5, September 2008 Diabetes Technology Society CLINICAL APPLICATIONS FreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in Glycemic Clamp Protocols Ohad, M.D., 1 Sigal Shaklai, M.D., Ph.D., 1 Ephraim Gabis, B.Sc., 2 and Michael A. Pani, M.D. 3 Abstract Objective: We assessed the accuracy of the FreeStyle Mini (FSM) meter for use in glycemic clamp and meal protocols in comparison with the HemoCue Glucose 201 DM Analyzer (HemoCue) and the YSI 2300 STAT Glucose Oxidase Analyzer (YSI). Methods: Seven volunteers with type 2 diabetes mellitus, 35 69 years old, underwent a frequently sampled meal test and a graded hyperglycemic test, on two separate days, with one of the volunteers undergoing each test twice. Samples for glucose measurements were obtained from arterialized venous blood. A total of 420 samples (with glucose levels ranging from 63 to 388 mg/dl) were available for comparison. On average, 10 measurements were available for every 5 mg/dl increment in glucose level in the range of 130 310 mg/dl. Blood glucose measurements were done on each sample with the FSM, HemoCue, and YSI. Results: FreeStyle Mini blood glucose values correlated closely with the YSI readings. Of the FSM measurements, 99.0% were within the Clarke error grid zone A; 51.3%, 84.7%, and 96.2% of the FSM readings were within 5%, 10% and 15% of the YSI values, respectively. The FSM was significantly more accurate than the HemoCue (84.7% vs 76.6% of results within 10% of the YSI results; p =.0038). The mean average relative difference of the FSM (5.8%) was also significantly lower than that of the HemoCue (6.8%; p =.0013) Conclusions: The FSM provides accurate results and constitutes a suitable alternative for bedside blood glucose measurements in experimental procedures, helping to reduce sample size, turnaround time, and cost. J Diabetes Sci Technol 2008;2(5):890-895 Author Affiliations: 1 Institute of Endocrinology, Ch. Sheba Medical Center, Sackler School of Medicine, Tel Aviv University, Israel; 2 Orsense, Nes Ziona, Israel; and 3 Abbott Diabetes Care, Wiesbaden, Germany Abbreviations: (FSM) FreeStyle Mini, (HemoCue) HemoCue Glucose 201 DM Analyzer, (ISO) International Organization for Standardization, (MAD) mean absolute difference, (MARD) mean average relative difference, (SD) standard deviation, (T2DM) type 2 diabetes mellitus, (YSI) Yellow Springs Instruments 2300 STAT Glucose Oxidase Analyzer Keywords: glucose monitor, glycemic clamp, diabetes mellitus Corresponding Author: Ohad, M.D., Institute of Endocrinology, Ch. Sheba Medical Center, Tel Hashomer, Israel 52621; email address ohad.cohen@sheba.health.gov.il 890

Introduction Clamp protocols are commonly used for the study of metabolic dysfunction in type 2 diabetes mellitus (T2DM) and other carbohydrate and lipid disorders. 1 In these protocols, emphasis is put on the accuracy of glucose measurements, as these are critical for the clamping procedure and for the interpretation of data. Glucose measurements are performed at intervals as short as two minutes and usually include 30 60 measurements per single clamp procedure. The resulting need for a rapid and accurate diagnostic tool requiring only small blood samples to minimize blood loss is challenging. Glucose oxidase-based analyzers, such as YSI 2300 STAT (YSI) (Yellow Springs Instruments Incorporated, Yellow Springs, OH), are considered accurate and provide reference measurements for glucose. The use of the YSI as a bed-side tool rather than as a source of reference measurements is impractical, because of the relatively long turnaround intervals and calibration time outs that necessitate the use of more than one analyzer. Some institutes use the Beckman Glucose Analyzer (Fullerton, CA) 2 as a bed-side tool, though it is still cumbersome and needs trained staff for rapid and accurate operation. The accuracy of hand-held devices for glucose measurements, though sufficient for homebased clinical management, is inadequate for clamp procedures. To fulfill the need for accurate and rapid glucose measurements that use minimal blood samples, studies have been conducted to assess the accuracy of hand-held devices in clamp protocols. Brunner et al. 3 found the analytical accuracy of a hand-held meter, as evaluated by variance components analysis, to be suboptimal. More than 3% of measurements deviated by more than +/-20% from the reference value, rendering the home glucose monitor inadequate for clamp procedures. Similar results from a hyperinsulinemic clamp study were reported by Stork et al. 4 with 4% of samples deviating by more than +/-20%. Consequently, the YSI and Beckman analyzers remain the most frequently used devices in clamp protocols. Unlike most other hand-held meters, the FreeStyle Mini [(FSM) Abbott Diabetes Care, Alameda, CA] uses a glucose dehydrogenase assay based on coulometry and requires only a small sample (0.3 µl) to produce highly accurate results across the 20 500 mg/dl range, independent of hematocrit, temperature, and many interfering substances. 5 Given the unmet need in clamp studies for an accurate, rapid, and cost-effective glucose monitoring system that requires only small blood samples, we investigated the suitability of using the FSM in glucose clamp protocols. Materials and Methods A total of 420 arterialized venous blood samples were obtained from seven patients with T2DM during eight extended meal tests and eight stepped hyperglycemic clamp tests. These clamp protocols were part of an extended study of the effects of medications on beta-cell function. All protocols performed at our clinical research center were approved by the center s internal review board and informed consent was obtained from every participant. These tests were conducted as follows: Graded hyperglycemic clamp. After an overnight fast, patients were admitted to the clinical research center and, after resting supine, they had a 20-gauge catheter inserted into an antecubital vein (for the infusion of test substances) and another catheter threaded in a retrograde manner into a wrist vein of the same arm, for blood sampling. The hand was kept in a heated box (55 C) for the duration of the study, to achieve and maintain arterialization of venous blood. The study protocol consisted of three consecutive 30-minute squarewave steps of hyperglycemia (each of 50 mg/dl above the previous level), achieved by a variable infusion of 20% dextrose, followed by an intravenous bolus of 5 g arginine. Blood samples for the determination of plasma glucose, insulin, and C-peptide concentrations were obtained every 2 minutes for the first 10 minutes and every 5 minutes thereafter for the final 20 minutes of each step. Meal test. The patients were admitted to the metabolism unit after an overnight (11- to 12-h) fast. An 18-gauge cannula was inserted in a retrograde manner into a dorsal hand vein to obtain arterialized venous blood samples. A mixed meal (10 kcal/kg, 45% carbohydrate, 15% protein, 40% fat) was consumed within 15 minutes and blood samples were obtained at 0, 5, 10, 15, 20, 30, 40, 50, 60, 75, 90, 120, 150, 180, 210, and 240 minutes. Blood samples were retrieved through a 3 way stopcock. When collecting the blood sample, the first 2 ml were discarded, and immediately thereafter a sample of approximately 3 ml was drawn, 2 ml were put into a prechilled vial containing powdered sodium fluoride and potassium oxalate (BD Vacutainer Plus; Becton, Dickinson, and Company, Franklin Lakes, NJ) and immediately 891

centrifuged for 15 s at 5800 g, separated, and placed on ice; the venous line was then flushed with saline. Subsequently, the glucose level in the plasma was analyzed with the YSI. From the remainder of the sample, a small amount of whole blood was immediately applied from the tip of the syringe and analyzed with the FSM glucose meter and a HemoCue Glucose 201 DM Analyzer (HemoCue AB, Ängelholm, Sweden). All instruments were maintained according to the manufacturers recommendations. Additionally, the YSI was calibrated at the start of every experimental day. Statistical Analysis Values were used for comparisons without corrections. The following statistical analyses were performed using SAS software (SAS Institute Inc., Cary, NC): 1. 2. 3. 4. Two-tailed Student s t-test on mean values from the HemoCue, FSM, and YSI with levels of significance defined as p <.05. Bias measures: Method of residuals: the differences between the YSI plasma glucose concentrations and the blood glucose meter readings (residuals) were calculated and presented as the mean, median, and standard deviation (SD) of the differences. Percentages were calculated for the HemoCue and FSM values within 5%, 10%, and 15%and within the International Organization for Standardization (ISO) requirements of the YSI value. System accuracy plot. Clarke and consensus grids. 6 Data are presented as means ± SD, unless otherwise indicated. As this trial was not devised exclusively to determine the accuracy of the FSM, no power analysis was performed or is presented. Data were collected from successive patients without excluding patients from the analysis. The use of successive data prevents data bias. Results A total of 420 matched glucose values were obtained from five male and two female patients undergoing eight meal tests and eight stepped hyperglycemic clamps (one female underwent two sets of studies). The mean age was 55.8 years (range: 35 69 years), and weight 88.3 ± 15.6 kg. The mean duration of diabetes was five years and the mean hemoglobin A1c was 7.4 ± 0.6%. Mean glucose levels as obtained by the YSI, FSM, and HemoCue were 230.3 ± 63.5 mg/dl, 237.8 ± 68.3 mg/dl and 241.8 ± 70.8 mg/dl, respectively. Glucose measurements ranged from 63 to 388 mg/dl; on average, 10 measurements were available for every 5 mg/dl increment in glucose levels in the range of 130 310mg/dl. The two protocol procedures differed in glucose levels with means for the meal test of 200.2 ± 59.9 mg/dl, 205.7 ± 63.8 mg/dl, and 209.1 ± 65.1 mg/dl for YSI, FSM, and HemoCue, respectively. For the clamp, mean glucose levels were 246.0 ± 59.6 mg/dl, 254.6 ± 64.5 mg/dl, and 258.9 ± 67.7 mg/dl for the YSI, FSM, and HemoCue, respectively. The results from the three glucose analyzers were highly correlated with an r 2 of 0.98 for the FSM vs YSI, 0.96 for the FSM vs HemoCue, and 0.97 for theysi vs HemoCue (p = nonsignificant). Grid analysis for the FSM (Figure 1) showed that 99.5% and 99.0% of the readings were in zone A on consensus and Clarke error grids, respectively. The results for the HemoCue were 99.5% and 98.6%, respectively (Figure 2). Bias measures for the FSM are presented in Table 1. The mean, median, and percent overall differences between the FSM and YSI were 7.5 mg/dl, 7.0 mg/dl, and 3.0%, respectively. The absolute mean, median, and percent differences between the FSM and YSI were 13.0 mg/dl, 11.0 mg/dl, and 5.8%, respectively. The differences during the hyperglycemic clamp were slightly greater than those during the meal tests. Of the FSM readings, 99.0% were Table 1. Bias Measures (FSM vs YSI) Measure Glucose Level Mean SD Median Absolute difference (mg/dl) All results 13.0 10.7 11.0 Absolute percent difference All results 5.8 4.5 4.9 Absolute difference (mg/dl) Clamp 14.0 11.3 12.0 Absolute percent difference Clamp 5.8 4.4 4.9 Absolute difference (mg/dl) Meal 11.2 8.9 8.5 Absolute percent difference Meal 5.8 4.7 4.8 Within 5% Within 10% Within 15% Within 20% (ISO) 215/419 (51.3%) 355/419 (84.7%) 403/419 (96.2%) 415/419 (99.0%) 892

within the ISO specifications. 7 A total of 51.3%, 84.7%, and 96.2% of the FSM glucose values were within 5%, 10%, and 15% of the YSI values, respectively. The absolute mean (15.7 mg/dl), median (13.0 mg/dl) and percent differences (6.8%) were significantly higher for the HemoCue (Table 2). A total of 42.7%, 76.6%, and 94.3% of the HemoCue glucose values fell within 5%, 10%, and 15% of the YSI values, respectively (p <.005 vs FSM) (Table 3). Discussion This study demonstrates the accuracy of the FSM during meal and graded hyperglycemic clamp protocols. These studies are characterized by rapid undulations in glucose levels spanning the range of 63 388 mg/dl, as depicted in Figure 3, a representative graded hyperglycemic clamp. Consensus and Clarke error grids demonstrated the excellent clinical acceptability of the FSM results (Figure 1). However, these performance criteria, pertaining to the Consensus error grid Clarke error grid Figure 1. YSI vs FSM performance (Consensus and Clarke error grids). Table 2. Comparison of Bias Measures (FSM vs HemoCue) Mean absolute difference Figure 2. YSI vs HemoCue performance (Clarke error grids). FS-YSI HemoCue- YSI Measure N Mean SD Mean SD Absolute difference (mg/dl) Absolute percent difference (%) 420 13.0 10.7 15.7 12.5 420 5.8 4.5 6.8 4.8 Median absolute difference FS-YSI HemoCue- YSI Measure N Median Median Absolute difference (mg/dl) Absolute percent difference (%) Paired difference (t-test p value) -2.6 (p <.0001) -0.94 (p =.0013) Wilcoxon Rank Sum p value 420 11.0 13.0.0025 420 4.9 6.1.0023 893

Table 3. Percentage of FSM and HemoCue Measurements Within YSI Measurements (glucose 75mg/dl) System Within 5% Within 10% Within 15% Within 20% (ISO) FreeStyle 215/419 (51.3%) 355/419 (84.7%) 403/419 (96.2%) 415/419 (99.0%) HemoCue 179/419 (42.7%) 321/419 (76.6%) 395/419 (94.3%) 414/419 (98.8%) Fisher s Exact Test 2-sided p-value.0154.0038 NS:.2565 NS: >.99 NS, nonsignificant Glucose (mg/dl) Time Figure 3. Representative plot of a graded hyperglycemic clamp. home-based clinical use of this device might not be appropriate in a clamp protocol, where high analytical accuracy is needed. To assess the analytical accuracy of the FSM on a level appropriate for clamp studies, bias measures were performed. The results of these studies indicated that the FSM was highly accurate, with a mean absolute difference (MAD) and mean average relative difference (MARD) of 13.0 mg/dl and 5.8%, respectively. The accuracy of the FSM was significantly superior to that of the HemoCue analyzer. Similar results for the HemoCue were presented by Trajanoski et al. 8 using a similar protocol but applying a hyperinsulinemic stepped hypoglycemic clamp. Thus, the study by Trajanoski et al. and the present study reflect a wide range of glucose levels, spanning from hypoglycemia to hyperglycemia. Similarly, arterialized venous blood was used and the reference values were obtained by a Beckman Glucose Analyzer II. The better results of the FSM can be due to assay accuracy and/or the fact that our study was conducted mostly in the hyperglycemic range. Whole blood glucose levels (by HemoCue) were not converted to match plasma levels (by YSI), as the MAD was 13.0 for the FSM and 15.7 (higher) for the HemoCue. In a similar study comparing the HemoCue to the YSI, Stork et al. 4 converted the whole blood glucose obtained by the HemoCue to plasma levels obtained by the YSI, by adding 10% to HemoCue whole blood. The differences between the results of the studies using conversion to plasma are not clear, as both studies used arterialized blood, which might lessen the differences between whole blood levels measured by a hand-held glucometer and serum levels measured by the YSI. 9 Finally, correcting the readings would make them only more accurate, but would not change the relationship between the FSM and HemoCue readings. Again, different glucose ranges, with an emphasis on hypoglycemia in other studies, 3,4,8 versus hyperglycemia in this study, might be a contributing factor to the differences between the results. For the FSM, the MAD and MARD of 13.0 mg/dl and 5.8%, as obtained in this study, indicate that this device is acceptable for use in glucose clamping protocols, particularly when rapid information is needed to correspond to rapid alterations in glycemic levels. Our findings concur with those of Brunner et al. 3 ; we found an individual variance in accuracy such that half of the patients had a MARD less than 5% and four patients had a MARD above 5%. Thus, we calibrate FSM readings during clamp studies with YSI measurements performed at different critical steps, for example, during a steady state interval or a step before increasing the glucose level; this culminates in 6-10 YSI measurements during the clamp. Additional advantages of the FSM are the convenience of use, small sample volume, rapid results, and low cost. The protocol is, as described, for stepped hyperglycemic clamp studies; it has been similarly described and used in several studies of beta-cell function, including studies of responses to different medications. This protocol was not devised specifically for assessing the accuracy of the FSM. In conclusion, we have shown that under specific conditions present in graded hyperglycemic clamp protocols, the measurements of the FSM glucose meter are sufficiently accurate to supply valid and useful glycemic data. Calibration with YSI results during the clamp procedure is recommended, as individual factors might impact the accuracy of the FSM measurements. 894

References: 1. 2. 3. 4. 5. DeFronzo RA, Tobin JD, Andres R. Glucose clamp technique: a method for quantifying insulin secretion and resistance. Am J Physiol. 1979;237(3):E214-23. Clarke WL, Anderson S, Farhy L, Breton M, Gonder-Frederick L, Cox D, Kovatchev B. Evaluating the clinical accuracy of two continuous glucose sensors using continuous glucose-error grid analysis. Diabetes Care 2005;28(10):2412-7. Brunner GA, Ellmerer M, Sendlhofer G, Wutte A, Trajanoski Z, Schaupp L, Quehenberger F, Wach P, Krejs GJ, Pieber TR. Validation of home blood glucose meters with respect to clinical and analytical approaches. Diabetes Care. 1998;21(4):585-90. Stork AD, Kemperman H, Erkelens DW, Veneman TF. Comparison of the accuracy of the HemoCue glucose analyzer with the Yellow Springs Instrument glucose oxidase analyzer, particularly in hypoglycemia. Eur J Endocrinol. 2005;153(2):275-81. Feldman B, McGarraugh G, Heller A, Bohannon N, Skyler J, DeLeeuw E, Clarke D. FreeStyle: a small-volume electrochemical glucose sensor for home blood glucose testing. Diabetes Technol Ther. 2000;2(2):221-9. 6. Clarke WL, Cox D, Gonder-Frederick LA, Carter W, Pohl SL. Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care. 1987;10(5):622-8. 7. 8. International Organization for Standardization. Requirements for in vitro blood glucose monitoring systems for self-testing in managing diabetes mellitus. ISO/TC 212/WG 3. In: Draft International Standard ISO/DIS 15197. Geneva, Switzerland: International Organization for Standardization; 2001. Trajanoski Z, Brunner GA, Gfrerer RJ, Wach P, Pieber TR. Accuracy of home blood glucose meters during hypoglycemia. Diabetes Care. 1996;19(12):1412-5. 9. Liu D, Moberg E, Kollind M, Lins PE, Adamson U, Macdonald IA. Arterial, arterialized venous and capillary blood glucose measurements in normal man during hyperinsulinaemic euglycaemia and hypoglycaemia. Diabetologia. 1992;35(3):287-90. 895