Reading centers are important in clinical trials to

Similar documents
Baseline Visual Field Characteristics in the Ocular Hypertension Treatment Study

Idiopathic intracranial hypertension (IIH) is a disorder of

Definitions. Indications for Perimetry. Indications for Perimetry. Purposes of Perimetry. Indications for Perimetry 3/4/2015

STANDARD AUTOMATED PERIMETRY IS A GENERALLY

3/16/2018. Perimetry

21st Century Visual Field Testing

Investigative Ophthalmology & Vision Sciences MSc Course. Glaucoma Module. Visual Field Reliability Indices. David Henson 2014.

CHAPTER 3 HOW TO PERFORM PERIMETRY YOU CAN TRUST

Misleading Statistical Calculations in Faradvanced Glaucomatous Visual Field Loss

Eye Movements, Strabismus, Amblyopia, and Neuro-Ophthalmology

VISUAL FIELDS. Visual Fields. Getting the Terminology Sorted Out 7/27/2018. Speaker: Michael Patrick Coleman, COT & ABOC

Is There an Association of Topical Ocular Hypotensive Medication with Lens Opacification and Decreased Visual Function?

Ocular hypertension is present in approximately 8%

Spontaneous Intraocular Pressure Reduction in Normal-Tension Glaucoma and Associated Clinical Factors

Bilateral Refractive Amblyopia Treatment Study

The Evolution of Fundus Perimetry

Snellen VA 2 IOP 3 Ocular Examination * Only dilate if repeat photos required

HFA3 with SITA Faster Frequently Asked Questions

Visual Fields Shawn L. Cohen, M.D. Part 2 of 4. Definitions / Tables (Part 2 of 2) Static Perimetry (Humphrey, Octopus)

Access to the published version may require journal subscription. Published with permission from: Elsevier

T he retinal ganglion cells of different sizes have distinct

Fluctuations on the Humphrey and Octopus Perimeters

The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation

Present relevant clinical findings of four landmark glaucoma trials OHTS, EMGT, CNTGS and CIGTS.

Science & Technologies

Dr Taha Abdel Monein Labib Professor of Eye Surgery Cairo University.

Conventional automated perimetry measures a subject s

Reason for Unscheduled Visit

CORRELATING OF THE VISUAL FIELD INDEX WITH MEAN DEVIATION AND PATTERN STANDARD DEVIATION IN GLAUCOMA PATIENTS

Expanding your field of vision. Visual Field Analyzers from Carl Zeiss

Characteristics of Frequency-of-Seeing Curves in Normal Subjects, Patients With Suspected Glaucoma, and Patients With Glaucoma

Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma

CLINICAL SCIENCES. The Ocular Hypertension Treatment Study

Visit Type (Check one)

Visual Field Interpretation Anthony B. Litwak, OD, FAAO VA Medical Center Baltimore, MD

Fluctuation of Intraocular Pressure and Glaucoma Progression in the Early Manifest Glaucoma Trial

Targeting Intraocular Pressure in Glaucoma: a Teaching Case Report 1

The Measure of Confidence

PRESCRIBING IN GLAUCOMA: GUIDELINES FOR NZ OPTOMETRISTS

Test-retest variability in visual field testing using frequency doubling technology

Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma

Case Report Optic Disk Pit with Sudden Central Visual Field Scotoma

CLINICAL SCIENCES. Classification of Visual Field Abnormalities in the Ocular Hypertension Treatment Study

Financial Disclosure. Visual Field Interpretation RELIABILITY VISUAL FIELD INTERPRETATION IN GLAUCOMA METHODS OF DATA PRESENTATION

53 year old woman attends your practice for routine exam. She has no past medical history or family history of note.

Evaluation of Sensitivity and Specificity of Spatial Resolution and Humphrey Automated Perimetry in Pseudotumor Cerebri Patients and Normal Subjects

International Journal of Health Sciences and Research ISSN:

MEDICAL POLICY SUBJECT: OCULAR PHOTOSCREENING. POLICY NUMBER: CATEGORY: Technology Assessment

CLINICAL SCIENCES. Steven L. Mansberger, MD; Pamela A. Sample, PhD; Linda Zangwill, PhD; Robert N. Weinreb, MD

Criteria for early CLINICAL STUDY. N Fujimoto 1, N Saeki 2, O Miyauchi 1

MEASURING THE SIZE OF A TREATMENT EFFECT: RELATIVE RISK REDUCTION, ABSOLUTE RISK REDUCTION, AND NUMBER NEEDED TO TREAT

Effect of brimonidine on intraocular pressure in normal tension glaucoma: A short term clinical trial

Ocular Hypotensive Efficacy of Netarsudil Ophthalmic Solution 0.02% Over a 24-Hour Period: A Pilot Study

Intro to Glaucoma/2006

Aerie Pharmaceuticals, Inc. Additional Rhopressa Rocket 1 Analyses and Path Forward

Vision Research & Clinical Trials Havener Eye Institute Jenna Rajczyk Demarcus Williams 13 April 2018

The Utility of the Monocular Trial

PATTERNS OF GLAUCOMATOUS VISUAL FIELD LOSS IN SITA FIELDS AUTOMATICALLY IDENTIFIED USING INDEPENDENT COMPONENT ANALYSIS

The Role of the RNFL in the Diagnosis of Glaucoma

TUMBLING E RESOLUTION PERIMETRY IN GLAUCOMA

MEDICAL POLICY SUBJECT: CORNEAL ULTRASOUND PACHYMETRY. POLICY NUMBER: CATEGORY: Technology Assessment

CLINICAL SCIENCES. (FDP) was designed to emphasize the response characteristics of the parasol

LOCSU Community Services. Glaucoma Repeat Readings & OHT Monitoring Community Service Pathway. Issued by Local Optical Committee Support Unit May 2009

IIH 1.0 L152~ L 33 IIIII_,. AI MICROCOPY RESOLUTION TEST CHARI JRFAU - A RDS-1963-A. doi

Role of Central Corneal Thickness in Circadian Intraocular Pressure Fluctuations among Patients with Primary Open Angle Glaucoma

Clinical Guidance and Monitoring for Change. Cecilia Fenerty MD FRCOphth Manchester Royal Eye Hospital

Optic Disk Pit with Sudden Central Visual Field Scotoma

CLINICAL SCIENCES. Baseline Topographic Optic Disc Measurements Are Associated With the Development of Primary Open-Angle Glaucoma

Quality in Optometry Record Card Keeping.

F requency doubling technology (FDT) perimetry was

A new visual field test in empty sella syndrome: Rarebit perimetry

GLAUCOMA SUMMARY BENCHMARKS FOR PREFERRED PRACTICE PATTERN GUIDELINES

Description of new EyeSuite visual field and trend analysis functions

Comparative study of new imaging technologies for the diagnosis of glaucoma: Protocol Approved by the Ethics Committee

CHAPTER 11 KINETIC PERIMETRY WHAT IS KINETIC PERIMETRY? LIMITATIONS OF STATIC PERIMETRY LOW SPATIAL RESOLUTION

Glaucoma What You Should Know

Ocular Hypertension Treatment Study. Manual of Procedures. Version /17/2015

Correlation of Blue Chromatic Macular Sensitivity with Optic Disc Change in Early Glaucoma Patients

Ultrasound in Emergency Medicine

Goals. Glaucoma PARA PEARL TO DO. Vision Loss with Glaucoma

Central Corneal Thickness-An important variable for prognostication in Primary Open Angle glaucoma; A Kolkata based study in Eastern India

Retinal Nerve Fiber Analysis: the Role It Plays in Assessing Glaucoma Grace Martin, CPOT

CLINICAL SCIENCES. Screening for Glaucoma With Frequency-Doubling Technology and Damato Campimetry

The improvement in perimetric results occurring over

Hidden LED Array Perimeter IVS-201A / IVS-201B. Reliable Accurate Economical. Lis t ic

Glaucoma: Diagnostic Modalities

Perimetry Phobia: Don t fear the field Savory Turman, COMT, CPSS

NIH Public Access Author Manuscript Arch Ophthalmol. Author manuscript; available in PMC 2014 March 26.

NIH Public Access Author Manuscript J AAPOS. Author manuscript; available in PMC 2006 April 25.

RETINAL NERVE FIBER LAYER

Use of a portable head mounted perimetry system to assess bedside visual fields

Use of a portable head mounted perimetry system to assess bedside visual fields

FORUM Glaucoma Workplace from ZEISS Clinical Interpretation Guide

Retinal nerve fiber layer thickness in Indian eyes with optical coherence tomography

Electrodiagnostics Alphabet Soup

Role of central corneal thickness measurement in management of open angle glaucoma and glaucoma suspects in Calabar, Nigeria

Introduction. Paulo de Tarso Ponte Pierre-Filho, 1 Rui Barroso Schimiti, 1 Jose Paulo Cabral de Vasconcellos 1 and Vital Paulino Costa 1,2

OCTOPUS 300 Perimetry you can trust

Scanning Laser Polarimetry in Patients with Acute Attack of Primary Angle Closure

LOCSU Community Services. Glaucoma Repeat Readings & OHT Monitoring Community Service Pathway. Issued by Local Optical Committee Support Unit May 2009

Transcription:

REVIEW ARTICLE Visual Field Quality Control in the Ocular Hypertension Treatment Study (OHTS) John L. Keltner, MD,*w Chris A. Johnson, PhD,z Kimberly E. Cello, BSc,* Shannan E. Bandermann, MA,* Juanjuan Fan, PhD,y Richard A. Levine, PhD,y Michael A. Kass, MD,J and Mae O. Gordon, PhDJ the Ocular Hypertension Treatment Study Group Objective: To report the impact of visual field quality control (QC) procedures on the rates of visual field unreliability, test parameter errors, and visual field defects attributed to testing artifacts in the Ocular Hypertension Treatment Study (OHTS). Methods: OHTS technicians were certified for perimetry and were required to submit 2 sets of visual fields that met study criteria before testing study participants. The OHTS Visual Field Reading Center (VFRC) evaluated 46,777 visual fields completed by 1618 OHTS participants between February 1994 and December 2003. Visual field QC errors, rates of unreliability, and defects attributed to testing artifacts were assessed. The OHTS QC system addressed 3 areas of clinic performance: (1) test parameter errors, (2) patient data errors, and (3) shipment errors. A visual field was classified as unreliable if any of the reliability indices exceeded the 33% limit. Clinical sites were immediately contacted by the VFRC via fax, e-mail, and/ or phone and instructed on how to prevent further testing errors on fields with defects attributed to testing artifacts. Main Outcome Measures: QC errors (test parameter errors) and unreliability rates. Results: A total of 2.4% (1136/ 46,777) of the visual fields were unreliable and 0.23% (107/46,777) had incorrect test parameters. Visual field defects attributed to testing artifacts occurred in approximately 1% (483/46,777) of the visual fields. Received for publication November 28, 2006; accepted March 2, 2007. From the Departments of *Ophthalmology and Vision Science; wneurology and Neurological Surgery, University of California, Davis; ydepartment of Mathematics and Statistics, San Diego State University, San Diego, CA; zdiscoveries in Sight, Devers Eye Institute, Portland, OR; and JDepartment of Ophthalmology and Visual Sciences, Washington Universiity, St Louis, MO. Supported in part by: Grants EY 09307 and EY 09341 from the National Eye Institute and the National Center on Minority Health and Health Disparities, National Institutes of Health, Bethesda, MD; Merck Research Laboratories, White House Station, NJ; and by an unrestricted grant from Research to Prevent Blindness, Inc, New York, NY. For a complete list of authors, refer to the OHTS web site: https:// vrcc.wustl.edu. Reprints: John L. Keltner, MD, Department of Ophthalmology and Vision Science, 4860 Y Street, Suite 2400, Sacramento, CA 95817 (e-mail: jlkeltner@ucdavis.edu). Copyright r 2007 by Lippincott Williams & Wilkins Conclusions: Prompt transmission of visual fields to the VFRC for ongoing and intensive QC monitoring and rapid feedback to technicians helps to reduce the frequency of unreliable visual fields and incorrect testing parameters. Visual field defects attributed to testing artifacts were infrequent in the OHTS. Key Words: mesodermal dysplasia spectrum, oculo-auriculovertebral spectrum, progressive noninfectious anterior vertebral fusion, kyphosis, MRI (J Glaucoma 2007;16:665 669) Reading centers are important in clinical trials to standardize procedures, train and certify technicians, perform quality control (QC) procedures, store data, perform analyses, and assess outcomes in a masked fashion. The Visual Field Reading Center (VFRC) at the University of California, Davis has participated in 2 major multicenter clinical trials. 1 13 Procedures used by the VFRC can reduce the number of unreliable or unusable visual fields and improve the quality of visual field data. 1,3,5,9 The Ocular Hypertension Treatment Study (OHTS) used Humphrey visual fields to assess changes in visual function due to primary open angle glaucoma. Recent papers written by Kass et al 6 and Gordon et al 10 have been published regarding the outcome of the OHTS. The purpose of this article is to report the impact of visual field quality assurance and control procedures on the rates of visual field unreliability, test parameter errors, and visual field defects attributed to testing artifacts. METHODS The OHTS visual field testing protocol included training, certification, and on-going feedback to technicians on each visual field test and also quarterly reports on visual field quality. Quarterly reports were distributed to OHTS clinic investigators, coordinators, Executive Committee members and Data Safety and Monitoring Committee members, reading centers (VFRC and Optic Disc Reading Center), and the National Eye Institute. The reports contained information on overall study and clinic performance from the reading centers and the data coordinating center (completed visits, number of J Glaucoma Volume 16, Number 8, December 2007 665

Keltner et al J Glaucoma Volume 16, Number 8, December 2007 participants, intraocular pressure goal, useable fields, and photographs). Clinic ranking information based on the number of goals met was also included. In addition, the OHTS Data Safety and Monitoring Committee required a more comprehensive biannual report examining the frequency of abnormal visual fields, unreliable visual fields, overall clinic QC performance, and clinic QC ranking. This information provided a direct comparison between OHTS clinical sites. The VFRC and Optic Disc Reading Center also provided more detailed quarterly performance reports with emphasis on data quality and timeliness of data transmission to the Data Coordinating Center. Between February 1994 and December 2003, 246 technicians from 22 clinical sites were certified for OHTS. Technicians were given the VFRC perimetry protocol to review before the certification examination. To become certified for OHTS, each technician was required to successfully pass a telephone certification examination and submit 2 sets of certification visual fields on nonstudy participants to the VFRC. For a participant to meet OHTS visual field entry criteria, 2 out of a maximum of 3 visual field tests (Humphrey full threshold 30-2) for each eye were required to be normal and reliable at baseline. The OHTS perimetry protocol provided explicit instructions on how to configure the Humphrey Field Analyzer s (HFA) s main menu to run the correct visual field test (30-2 full threshold). It was necessary for the technician to manually program the HFA before each test was performed. In August 2003, the OHTS perimetry protocol changed its testing parameter requirements from full threshold 30-2 to SITA Standard 30-2. Visual fields were classified as unreliable if either false-positive errors, falsenegative errors, or fixation losses were greater than 33% (33% fixation losses are used as the limit for OHTS, in contrast to the HFA s default value of 20%). 1 Follow-up visits were scheduled every 6 months from the date of randomization. We recommended that visual fields be sent to the VFRC weekly but required that they be sent within 21 days of testing. The OHTS VFRC protocol required that all abnormal and/or unreliable visual fields be faxed to the VFRC within 7 days of testing. When the VFRC received a visual field that was abnormal and/or unreliable, the site was contacted via fax with instructions for the next follow-up visual field. If a visual field was rendered abnormal owing to a testing artifact (ie, trial lens rim, heavy/droopy lid), the clinical site was contacted via fax, e-mail, and/or telephone and instructed on how to prevent further testing errors. VFRC staff emphasized that during visual field testing, the lenses be moved as close as possible to the eye, heavy/droopy lids be taped, and the participant should be made as comfortable as possible. If the next follow-up visual field (second) was again abnormal due to a testing artifact, the OHTS Coordinating Center was also notified. To prevent the next follow-up visual field (third) from being abnormal due to testing artifact, the VFRC notified technicians (via faxes, e-mail, and telephone) with instructions for the next follow-up fields, emphasizing the importance of properly aligning the patient and/or taping heavy or droopy lids before testing. As a secondary, preventative measure, the VFRC also contacted the OHTS central coordinators. The coordinators were responsible for following up with the VFRC s original request regarding proper testing instructions. Previous papers described the study protocol and baseline clinical characteristics of participants enrolled in the OHTS. 2,3 Study rationale, organization, and procedures are discussed in the OHTS Manual of Procedures. 4 According to OHTS criteria, a follow-up visual field was considered to be abnormal if the glaucoma hemifield test and/or the corrected pattern standard deviation global index were outside normal limits (P<1% and <5%, respectively). Final determination as to whether the visual field abnormality was attributable to primary open angle glaucoma was made by the OHTS Endpoint Committee after review of all optic disc photographs, visual fields, and medical and ocular history of both eyes of the participant. 6 The OHTS visual field QC system addressed 3 areas of clinic performance: (1) test parameter errors, (2) patient data errors, and (3) shipment errors (Table 1). Each visual field was graded on a 100-point scale with regard to whether the protocol was followed in each of these 3 areas. A score was assigned to each QC error, depending on the severity of the violation. On this scale, zero represented a perfect score, with increasing point scores reflecting more severe QC errors. Factors that rendered the data unusable for the study (eg, using the wrong test procedure) had higher point values, whereas those that produced minor errors (eg, incorrect data entry) had lower point totals. RESULTS Between February 1994 and December 2003, 246 certified technicians from 22 clinical sites performed 46,777 visual field tests (7257 eligibility and 39,520 follow-ups) on 1618 participants in the OHTS. Only 2.4% (1136/46,777) of the visual fields were unreliable and only 0.23% (107/46,777) had incorrect test parameters. Fixation loss errors were the most common reliability index outside normal limits; 76% (866/1,136) of the unreliable fields were due to excessive fixation losses. Excessive false-positive errors contributed to 17% (188/ 1136) of the unreliable visual fields, and excessive falsenegative errors contributed to 11% (124/1136) of the unreliable fields (Table 2). Table 3 shows the comparison of OHTS reliability rates with those obtained for previous studies of ocular hypertension or glaucoma. The OHTS attained superior overall reliability (with an unreliability rate of only 2.4%) compared with other published studies. 5 Fixation losses, false-positive and false-negative errors contributed to 1.9%, 0.4%, and 0.3% of the total fields (46,777), respectively. When comparing the prior studies with the findings reported for OHTS, it should be noted that there were 666 r 2007 Lippincott Williams & Wilkins

J Glaucoma Volume 16, Number 8, December 2007 Visual Field Quality Control TABLE 1. Technician QC Errors Errors Point Values Test parameter errors (maximum of 60 points) Program 30-2 not used 60 STATPAC 2 not used 60 Stimulus size III not used 60 Full threshold strategy not used 60 Short-term fluctuation test turned off 60 Foveal threshold test turned off 10 Blind spot check turned off 10 Inappropriate blind spot check size used 10 Central fixation target not used 5 Visual field data not saved on disk 20 Patient data errors (maximum of 30 points) Pupil diameter less than 3 mm or not entered 10 Visual acuity not entered 3 Birth date incorrect 5 Site ID not entered in ID field 3 Site ID incorrect 2 Technician ID not entered in ID field 3 Technician ID incorrect 2 Visit code not entered in ID field 3 Visit code incorrect 2 Dilation information incorrect 2 Patient ID not entered in name field 5 Patient ID incorrect 4 Distance Rx not entered in name field 10 Distance Rx entered incorrectly 3 Rx used not entered 10 Sphere error in patient s Rx 7 A sphere of ± 0.25 D must be dropped for test 3 Cylinder discrepancy in patient s Rx 3 Axis discrepancy in patient s Rx 2 A cylinder less than 1.00 D must be dropped and 3 a spherical equivalent used A cylinder lens must be used for a cylinder of 1.00 D 3 or greater VFRC Shipment errors (maximum of 10 points) Visual field printout not included in shipment 3 Humphrey disk file shipped late (r30 d) 3 Humphrey disk file shipped late (>30 d) 10 Humphrey disk directory not included in shipment 2 Visual field not checked off on Humphrey disk directory 1 Shipment not packaged properly 2 Shipment not addressed properly 2 Humphrey disk not labeled 2 Humphrey disk not labeled correctly 1 Prestudy form not included with visual field 2 Prestudy form not completed correctly 1 This field was not faxed to the VFRC as required 3 This field was not faxed to the VFRC on time 2 TABLE 2. Unreliability and Incorrect Test Parameters in the OHTS Visual Field Status OHTS Unreliable* 2.4% (1136/46,777) Fixation loss errors 76% (866/1136) False-positive errors 17% (188/1136) False-negative errors 11% (124/1136) Incorrect test parameters 0.23% (107/46,777) *Note that more than one reliability index can be outside normal limits on a single visual field examination. As shown in Figure 1, OHTS mean test, patient data, and shipment errors have generally decreased over time (1994 to 2003) with the exception of mean test error points which intermittently fluctuated owing to the introduction of a new perimeter (HFA II second Qrt. 1999) and a change in the testing protocol from full threshold to SITA Standard (third Qrt 2003). Visual field defects attributed to testing artifacts (trial lens rim artifacts and heavy/droopy lids) occurred in approximately 1.03% (483/ 46,777) of the OHTS total visual fields. Trial lens rim artifacts typically produced visual field abnormalities such as peripheral rim, partial peripheral rim, or inferior depression defects; heavy/ droopy lids typically produced abnormalities such as superior depression defects (Table 4). Although we have not reported these types of visual field testing artifacts as QC measures for the OHTS, abnormalities caused by a trial lens rim artifact or ptosis will automatically generate feedback from the VFRC to reduce the probability that they will occur in future tests. DISCUSSION To our knowledge, few clinical trials have used visual field testing protocols with such stringent and timely QC procedures as the OHTS. QC is of the utmost importance for maintaining good reliability rates and ensuring the validity and integrity of visual field data. In addition, QC can reduce the expense of missing data and potential for bias. One of the key elements in QC functions is to adapt procedures to identify problems in a particular trial. The many differences. To our knowledge, QC procedures were not employed in the prior studies, inclusion/exclusion criteria were different, the other studies were from a single center, and patient selection was not as stringent. While we believe that QC procedures represent a major factor for reliable visual fields in the OHTS, other issues (patient selection, standardized protocols, technician training, and certification, etc) are also possible contributors to higher quality visual field information. In addition, we feel that practitioners who employ these techniques will find an improvement in the reliability and accuracy of their visual field test results. TABLE 3. Comparison of OHTS Reliability Rates With Other Reported Rates* Study (Participants) Overall Unreliability Fixation Losses (%) False-positive False-negative OHTS* (OH) 2.4 1.9 0.4 0.3 Bickler-Bluth 15.0 <7.0 <7.0 (OH) 14 Birt (G) 15 25.0 5.0 9.0 *FLs, FPs, and FNs reflect 33% cut-off rate unless otherwise noted; n = 46,777. G indicates glaucomatous; OH, ocular hypertensives. r 2007 Lippincott Williams & Wilkins 667

Keltner et al J Glaucoma Volume 16, Number 8, December 2007 FIGURE 1. QC performance of OHTS clinics includes all processed visual fields performed on OHTS participants from January 1, 1994 to December 31, 2003. best example for the need of a VFRC occurred with the Optic Neuritis Treatment Trial (ONTT). The UC Davis VFRC has been the reading center for the ONTT since 1988. Recent papers written by Keltner et al and Johnson et al have been published regarding the baseline, QC functions, and the 1-year follow-up of the ONTT. 8,9,13 Both the ONTT and the OHTS had similar protocols for visual field testing even though the study samples differed in disease entities, glaucoma/ocular hypertension, and optic neuritis, respectively. The OHTS and the ONTT 10-year follow-up study unreliability rates were both very low at 2.4% (1136/ 46,777) and 6% (43/722), respectively; however, the OHTS unreliability rate was lower than the ONTT. There were many differences in the OHTS and ONTT testing protocols (Table 5). An important QC error occurred in the ONTT 10-year follow-up study. A total of 19% (34/173) of the ONTT visual fields had incorrect test parameter errors, whereas only 0.23% (107/46,777) of the OHTS visual fields had incorrect test parameters. The VFRC analyzed the cause for this high rate of incorrect test parameter errors and determined that the one-time follow-up visit in the 10-year follow-up study lacked ongoing visual field feedback and clinic center ranking. Thus in the 15-year ONTT follow-up study, the VFRC developed a new strategy which included immediate fax requests and a point-value system. The VFRC reviewed each faxed visual field, and if it was determined that the wrong test parameter was used, the clinic would be TABLE 4. Percentage of Visual Field Testing Artifacts in OHTS Total Visual Fields VF Abnormality No. VFs % of VFs (n = 46,777) Peripheral rim 6 0.01 Partial peripheral rim 293 0.63 Superior depression 153 0.33 Inferior depression 31 0.06 Total 483 1.03 immediately notified and the correct visual field was performed. In addition to immediate feedback, each visual field was graded on a 100-point scale. By adapting this similar OHTS strategy to the ONTT QC procedures, incorrect test parameter errors decreased from 19% (34/ 173) in the 10-year ONTT follow-up study to 1.5% (10/ 637) in the 15-year follow-up study. Thus it would appear that with appropriate feedback, testing errors can be reduced by analyzing QC issues and taking the appropriate corrective action. The effects of ptosis on visual field testing can be prevented by taping the participant s lid with surgical tape, whereas trial lens rim artifacts can be prevented by (1) correctly aligning the participant s eye in the center of the trial lens rim(s) during testing and (2) instructing the participant not to lose forehead contact with the perimeter s brow band. As shown in Table 4, visual field loss attributed to testing artifacts occurred in only 1.03% of the total OHTS visual fields. Interestingly, the ONTT 10-year follow-up study also reported that visual field defects attributed to testing artifact occurred in less than 1% (7/722) of its total visual fields. To our knowledge, few previous studies have reported visual field abnormalities likely due to testing artifacts. Thus, we cannot adequately compare the frequencies of these types of artifactual abnormalities with other studies. The QC procedures used in the OHTS have likely proven very effective in producing high quality visual field data; however, there is no control group to completely validate this assumption. As seen in Table 3, the OHTS unreliability rate is superior to other glaucoma studies previously published. Unfortunately, complete details of the QC procedures used in these older articles are not well described. While QC measures used to produce high quality visual field data are essential in clinical trials, we believe that many of the same principles can be used in routine clinical settings. In every ophthalmology practice, it would be advisable for technicians to: (1) read standard 668 r 2007 Lippincott Williams & Wilkins

J Glaucoma Volume 16, Number 8, December 2007 Visual Field Quality Control TABLE 5. Major QC Differences Between Study Protocols Used in OHTS and ONTT Study Parameters OHTS ONTT Visual field entry criteria 2 sets of normal, reliable baseline/qualifying visual fields 1 abnormal visual field consistent with optic neuritis (not required to be reliable) Visual field shipment criteria Weekly Not specified Visual field testing frequency Every 6 mo Once for 180 participants; twice for 90 participants Technician certification Certified for perimetry by the VFRC* via: Certified for perimetry by Jaebw via: (1) Phone examination (1) Phone examination (2) Visual field testing (4 satisfactory fields from (2) No certification fields required both eyes of 2 nonstudy participants) QC ranking Clinics ranked by their QC scores and detailed quarterly reports sent to clinic PIs/coordinators/ techs (Table 3) None Clinic error notification *Visual Field Reading Center, Sacramento, CA. wjaeb Center for Health Research, Tampa, FL. Contacted within 24 h via QC reports, faxes, telephone, and e-mails Contacted within 24 h via telephone and e-mails. However, often all visual field testing completed and sent in batches before feedback was possible instructions to the participant each time before testing (this has been shown to be effective in reducing variability), 16 (2) use the proper refraction, 3 (3) insure proper participant ergonomics to avoid fatigue, (4) properly align the trial lens to the eye to avoid a trial lens rim artifact, and (5) tape the lids if ptosis is present to avoid a superior visual field defect. REFERENCES 1. Keltner JL, Johnson CA, Quigg JM, et al. Confirmation of visual field abnormalities in the ocular hypertension treatment study. Arch Ophthalmol. 2000;118:1187 1194. 2. Gordon MO, Kass MA. The Ocular Hypertension Treatment Study: design and baseline description of the participants. Arch Ophthalmol. 1999;117:573 583. 3. Johnson CA, Keltner JL, Cello KE, et al. Baseline visual field characteristics in the Ocular Hypertension Treatment Study (OHTS). Ophthalmology. 2002;109:432 437. 4. Gordon MO, Kass MA. Ocular Hypertension Treatment Study (OHTS) Manual of Procedures. Washington, DC: National Technical Information Services; 1997. 5. Keltner JL, Johnson CA, Cello KE, et al, Ocular Hypertension Study Group. Visual field technician performance in the Ocular Hypertension Treatment Study (OHTS). Invest Opthalmol Vis Sci. 2001;42:S152. Poster presented at the 2001 NANOS and AAO Meetings. 6. Kass MA, Heuer DK, Higginbotham EJ, et al, for the Ocular Hypertension Treatment Study Group. The Ocular Hypertension Treatment Study. A randomized trial determines that topical ocular hypotensive medication delays or prevents the onset of primary open-angle glaucoma. Arch Ophthalmol. 2002;120:701 713. 7. Keltner JL, Johnson CA, Cello KE, et al, Ocular Hypertension Treatment Study Group. Classification of visual field abnormalities in the Ocular Hypertension Treatment Study. Arch Ophthalmol 2003;121:643 650. 8. Keltner JL, Johnson CA, Spurr JO, et al. Baseline visual field profile of optic neuritis. The experience of the optic neuritis treatment trial. Optic Neuritis Study Group. Arch Ophthalmol. 1993; 111:231 234. 9. Keltner JL, Johnson CA, Beck RW, et al, the Optic Neuritis Study Group. Quality Control Functions of the Visual Field Reading Center (VFRC) for the Optic Neuritis Treatment Trial (ONTT). Controlled Clin Trials. 1993;14:143 159. 10. Gordon MO, Beiser JA, Brandt JD, et al, for the Ocular Hypertension Treatment Study Group. The Ocular Hypertension Treatment Study. Baseline factors that predict the onset of primary open-angle glaucoma. Arch Ophthalmol. 2002;120:714 720. 11. Optic Neuritis Study Group. Visual function more than 10 years after optic neuritis: experience of the Optic Neuritis Treatment Trial. Am J Ophthalmol. 2003;137:77 83. 12. Optic Neuritis Study Group. High and low risk profiles for the development of multiple sclerosis within 10 years after optic neuritis: experience of the Optic Neuritis Treatment Trial. Arch Ophthalmol. 2003;121:944 949. 13. Keltner JL, Johnson CA, Spurr JO, et al, for the Optic Neuritis Study Group. Visual field profile of optic neuritis: one-year followup in the Optic Neuritis Treatment Trial. Arch Ophthalmol. 1994;112:946 953. 14. Bickler-Bluth M, Trick GL, Kolker AE, et al. Assessing the utility of reliability indices for automated visual fields. Testing ocular hypertensives (see comments). Ophthalmol. 1989;96: 616 619. 15. Birt CM, Shin DH, Samudrala V, et al. Analysis of reliability indices from Humphrey visual field tests in an urban glaucoma population. Ophthalmology. 1997;104:1126 1130. 16. Kutzko KE, Brito CF, Wall M. Effect of instructions on conventional automated perimetry. Invest Opthalmol Vis Sci. 2000; 41:2006 2013. r 2007 Lippincott Williams & Wilkins 669