Two-Dimensional HPLC Determination of Water-Soluble Vitamins in a Nutritional Drink

Similar documents
Two-dimensional HPLC determination of water-soluble vitamins in a nutritional drink

Determination of Tigecycline in a Cell Lysate

Determination of Water- and Fat-Soluble Vitamins in Nutritional Supplements by HPLC with UV Detection

Rapid and sensitive UHPLC screening for water soluble vitamins in sports beverages

Rapid and sensitive UHPLC screening of additives in carbonated beverages with a robust organic acid column

Impurity Profiling of Carbamazepine by HPLC/UV

Improved Extraction and Analysis of Hexavalent Chromium from Soil and Water

Determination of Phenolic Compounds in Apple Orchard Soil

Determination of Carbohydrates in Kombucha Using HPAE-PAD

Carl Fisher, Terri Christison, Hua Yang, Monika Verma, and Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA

Detection and Quantification of Inorganic Arsenic in Fruit Juices by Capillary Ion Chromatography with Suppressed Conductivity Detection

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

Doubling the throughput of long chromatographic methods by using a novel Dual LC workflow

Rapid determination of phosphate and citrate in carbonated soft drinks using ion chromatography

Measuring Phytosterols in Health Supplements by LC/MS. Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA

A Rapid UHPLC Method for the Analysis of Biogenic Amines and Metabolites in Microdialysis Samples

Simultaneous determination of water- and fat-soluble vitamins in tablets and energy drinks by using a novel Vanquish Flex Duo system for Dual LC

Determination of Ethyl Sulfate Impurity in Indinavir Sulfate Drug Using Ion Chromatography

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract.

Enhanced LC-MS Sensitivity of Vitamin D Assay by Selection of Appropriate Mobile Phase

Mercury Speciation Determinations in Asian Dietary Supplements

Determination of Inorganic Ions and Organic Acids in Non-Alcoholic Carbonated Beverages

Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D

Determination of Clinically Relevant Compounds using HPLC and Electrochemical Detection with a Boron-Doped Diamond Electrode

ISSN (Print)

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

Rapid and Direct Analysis of Free Phytosterols by Reversed Phase HPLC with Electrochemical Detection

Analysis of Counterfeit Antidiabetic Drugs by UHPLC with the Agilent 1220 Infinity Mobile LC

The Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis

F. Al-Rimawi* Faculty of Science and Technology, Al-Quds University, P.O. Box 20002, East Jerusalem. Abstract

A HIGH PERFORMANCE LIQUID CHROMATOGRAPHIC ASSAY FOR LERCANIDIPINE HYDROCHLORIDE

Analysis of Emulsifiers in Foods by High Pressure Liquid Chromatography and Corona Charged Aerosol Detection

International Journal of Pharma and Bio Sciences DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR THE ESTIMATION OF STRONTIUM RANELATE IN SACHET

Determination of Benzoate in Liquid Food Products by Reagent-Free Ion Chromatography

Development and Validation for Simultaneous Estimation of Sitagliptin and Metformin in Pharmaceutical Dosage Form using RP-HPLC Method

Qualitative and quantitative determination of phenolic antioxidant compounds in red wine and fruit juice with the Agilent 1290 Infinity 2D-LC Solution

J Pharm Sci Bioscientific Res (4): ISSN NO

IJPAR Vol.3 Issue 4 Oct-Dec-2014 Journal Home page:

Analysis of Amino Acids Derived Online Using an Agilent AdvanceBio AAA Column

ISSN: ; CODEN ECJHAO E-Journal of Chemistry 2011, 8(3),

Available online Research Article

Sialic Acid Determination in Glycoproteins: Comparison of Two Liquid Chromatography Methods. Experimental. Introduction. Method 1.

LC-MS/MS Method for the Determination of Tenofovir from Plasma

Development and Validation of a Simultaneous HPLC Method for Quantification of Amlodipine Besylate and Metoprolol Tartrate in Tablets

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

Journal of Chemical and Pharmaceutical Research, 2017, 9(9): Research Article

Development and validation of stability indicating RP-LC method for estimation of calcium dobesilate in pharmaceutical formulations

GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants

The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection

A New HILIC/RP Mixed-Mode Column and Its Applications in Surfactant Analysis

Analysis of Isoflavones with the PerkinElmer Flexar FX-15 UHPLC System Equipped with a PDA Detector

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR ESTIMATION OF LACOSAMIDE IN BULK AND ITS PHARMACEUTICAL FORMULATION

Development, Estimation and Validation of Lisinopril in Bulk and its Pharmaceutical Formulation by HPLC Method

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

Rapid Gradient and Elevated Temperature UHPLC of Flavonoids in Citrus Fruit

Fig.1. Denatonium benzoate (DB) chemical structure

Pelagia Research Library

Journal of Chemical and Pharmaceutical Research

LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine

Analysis of Phospholipids in Natural Samples by Normal Phase HPLC and Corona Charged Aerosol Detection

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

Available online at Scholars Research Library

Pelagia Research Library

A HPAE-PAD Chromatographic Assay for Carbohydrates in Urine as a Measure of Intestinal Permeability

2D-LC as an Automated Desalting Tool for MSD Analysis

Multigenics Chewable

International Journal of Pharma and Bio Sciences

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

Validation of Changes to the USP Assay Method for Ibuprofen Tablets

RP-HPLC METHOD DEVELOPMENT AND VALIDATION FOR THE ESTIMATION OF BACLOFEN IN BULK AND PHARMACEUTICAL DOSAGE FORMS

Scholars Research Library. Der Pharmacia Lettre, 2015, 7 (5):44-49 (

Scholars Research Library. Der Pharmacia Lettre, 2016, 8 (6): (

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS

Robust and Fast Analysis of Tobacco-Specific Nitrosamines by LC-MS/MS

A simple validated RP-HPLC method for quantification of sumatriptan succinate in bulk and pharmaceutical dosage form

CHAPTER 2 SIMULTANEOUS DETRMINATION OF ANASTROZOLE AND TEMOZOLOMIDE TEMOZOLOMIDE CAPSULES 20 MG AND ANASTROZOLE TABLETS 1 MG

ASSAY AND IMPURITY METHOD FOR DURACOR TABLETS BY HPLC

Determination of Plant-Derived Neutral Oligo- and Polysaccharides Using the CarboPac PA200

Quantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD ESTIMATION OF TOLVAPTAN IN BULK PHARMACEUTICAL FORMULATION

Concanavalin A Column for Analysis of Glycoproteins and Their Tryptic Glycopeptides

Pankti M. Shah et al, Asian Journal of Pharmaceutical Technology & Innovation, 04 (17); 2016; 07-16

Determination of Bisphenol A in Milk Powder using a Chromolith HighResolution RP-18 endcapped column

Evaluation of an LC-MS/MS Research Method for the Analysis of 33 Benzodiazepines and their Metabolites

Analytical Method Development for USP Related Compounds in Paclitaxel Using an Agilent Poroshell 120 PFP

METHOD 8316 ACRYLAMIDE, ACRYLONITRILE AND ACROLEIN BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC)

RP-HPLC Analysis of Temozolomide in Pharmaceutical Dosage Forms

Dayana Argoti, Kerry Hassell, Sarah J. Fair, and Joseph Herman Thermo Fisher Scientific, Franklin, MA, USA

Determination of Steviol Glycosides by HPLC with UV and ELS Detections

HPLC to UHPLC Transfer of USP Method for Amlodipine Besylate Using the Agilent 1290 Infinity II LC

METHOD DEVELOPMENT AND VALIDATION BY RP-HPLC FOR ESTIMATION OF ZOLPIDEM TARTARATE

SIMULTANEOUS ESTIMATION OF VALSARTAN AND HYDROCHLOROTHIAZIDE IN TABLETS BY RP-HPLC METHOD

Separation of 2AA-Labeled N-Linked Glycans from Glycoproteins on a High Resolution Mixed-Mode Column

Sanjog Ramdharane 1, Dr. Vinay Gaitonde 2

Selectivity Comparison of Agilent Poroshell 120 Phases in the Separation of Butter Antioxidants

Research Article Simultaneous Estimation of DL-Methionine and Pyridoxine Hydrochloride in Tablet Dosage Form by RP-HPLC

Two-Dimensional HPLC Combined with On-Line SPE for Determination of Sudan Dyes I IV in Chili Oil

Transcription:

Two-Dimensional HPLC Determination of Water-Soluble Vitamins in a Nutritional Drink Dai Zhenyu, Chen Jing, Xu Qun, and Jeffrey Rohrer Thermo Fisher Scientific, Shanghai, People s Republic of China; Thermo Fisher Scientific, Sunnyvale, C, US pplication Note 09 Key Words Food nalysis, Food Quality, cclaim Polardvantage Column, cclaim 0 C8 Column, Hypersil GOLD Phenyl Column Goal To develop an efficient high-performance liquid chromatography (HPLC) method for simple and sensitive determination of watersoluble vitamins in a complex multivitamin/mineral drink. Target analytes are group vitamins, including thiamine (V ), riboflavin (V ), nicotinamide (V ), pantothenic acid (V ), pyridoxine (V ), biotin (V ), and cyanocobalamin (V ); and ascorbic acid (V C ). Introduction Vitamins are a well-known group of compounds that are essential for human health. They can be classified into two main groups: water- and fat-soluble. With the exception of V and V, water-soluble vitamins are not stored in the body. Thus, if one s dietary vitamin intake is insufficient, a vitamin supplement should be added to the diet. The vitamin supplement can be in tablet form, a clear vitamin-enhanced functional drink, vitamin-enhanced milk, or a nontransparent multivitamin/mineral nutritional drink with additions of other substances (e.g., fruit extracts) that make it more complex than clear products. To ensure that these products contain the labeled amounts of vitamins, a number of reliable quality control assays are available. - For a vitamin tablet or a clear functional drink, the analysis is relatively simple and a routine HPLC method (e.g., a C8 column with UV detection) is satisfactory for quantifying the vitamins. Some samples, however, have too many additional components to allow a routine HPLC vitamin-quantification method. Vitamin-enhanced milk and a nontransparent multivitamin/mineral nutritional drink referred to as a multivitamin nutritional drink throughout the rest of this study are two such samples. In addition to vitamins, these samples also supply amino acids, minerals, coenzyme Q0, the compounds contained in grape extracts, and more. These additional compounds interfere with the separation of vitamins, making quantification difficult. Therefore, a simple and sensitive two-dimensional HPLC (D-HPLC) method is needed to quantify vitamins in these complex samples. Equipment and Software Thermo Scientific Dionex UltiMate 000 x Dual HPLC system, including: DGP-00RS Dual-Gradient Rapid Separation Pump System with SRD-00 Integrated Solvent and Degasser Rack WPS-000TRS Rapid Separation Wellplate Sampler, Thermostatted, with a 00 μl sample loop and a 00 μl syringe TCC-000RS Rapid Separation Thermostatted Column Compartment equipped with two p p valves DD-000RS Rapid Separation Diode rray Detector with µl flow cell Mixer for 800 μl Mixing Volume (P/N 00.0) Thermo Scientific Dionex Chromeleon Chromatography Data System (CDS) software version. Thermo Scientific Orion -Star ph enchtop Meter

Reagents and Standards Deionized (DI) water, 8. MΩ-cm resistivity cetonitrile (CH CN) for HPLC (Fisher Scientific P/N C00000) Potassium Phosphate Monobasic (KH PO ) (Fisher Scientific P/N P8-) o-phosphoric cid (H PO ), 8%, (Fisher Scientific P/N 0-00) Products from the National Institute for the Control of Pharmaceutical and iological Products, eijing, China: V V V V V V V V C Preparation of Standard Solutions To prepare water-soluble vitamin standards of V, V, V, V, V, V, and V C, weigh 0 mg of the vitamin powder and add DI water to 0 ml in a volumetric flask to make stock solutions of.0 mg/ml for each vitamin. Make a fresh preparation of V C, due to its limited stability. lso, because of the limited solubility of V in water, decrease the concentration of the V stock solution to 0.0 mg/ml. Weigh mg of V into 00 ml DI water to address the solubility issue. If a 0 ml volumetric flask were used, 0. mg of V would have to be weighed, but that would exceed the precision range of the balance. dd 00 μl of V stock solution, 00 μl each of V and V stock solutions, and 0 μl each of V, V, V, V, and V C stock solutions to a ml sample vial. ring the final volume to ml by adding 0 μl of DI water to make the mixed stock standard solution. In this mixed stock standard solution, the concentration of V will be μg/ml, the concentrations of V and V will be 00 μg/ml, and the concentration of the other vitamins will be 0 μg/ml. For the preparation of mixed working standard solutions for calibration, add the appropriate volume of the mixed stock standard solution into 0 ml glass vials and bring to 0 ml with DI water. See Table for details. Sample Preparation multivitamin nutritional drink supplemented with V, V, V, V, V, V, V, and V C was provided by a customer. The drink also contained some fat-soluble vitamins, minerals, amino acid complex, grape seed extract, coenzyme Q0, and other ingredients added to meet daily nutritional needs. Table lists the sample components. Dilute the drink sample with DI water if necessary and filter through a 0. µm filter. To determine if the sample needs to be diluted, compare the labeled values to the calibration ranges in this study. Store the sample in a brown bottle at C before analysis. Table. Preparation of mixed working standard solutions of water-soluble vitamins. Mixed Working Standard Solution Volume of Mixed Stock Standard Solution for a 0 ml Preparation (ml) 0.0 0. 0..0.0 0 Concentration of Each Vitamin (mg/l) V 0.0 0. 0...0 V, V 0..0.0 0 0 00 V, V, V, V, V C 0. 0..0.0 0 0 Table. Labeled values of the multivitamin nutritional drink. Ingredient mount Per Serving* Ingredient mount Per Serving Total Carbohydrate 9 g Zinc (as gluconate) mg Vitamin 0000 IU Selenium (as L-selenium methione) 00 mcg Vitamin C 000 mg Copper (as gluconate) mg Vitamin D 00 IU Manganese (as gluconate) mg Vitamin E 00 IU Chromium (as amino acid chelate) 00 mcg Vitamin K 00 mcg Potassium (as citrate) 00 mg Vitamin 0 mg Choline (as bitartrate) 0 mg Vitamin 0 mg Inositol 0 mg Vitamin 0 mg oron (as amino acid chelate) mg Vitamin 0 mg mino cid Complex (proprietary formula) mg Vitamin 0 mg Grape Seed Extract mg Vitamin 00 mcg Coenzyme Q-0 mg Vitamin 00 mcg Dimethyl Glycine mg Folate 00 mcg Paba 0 mg Calcium 00 mg Citrus ioflavonoids mg Phosphorus 80 mg Glucolactone 0 mg Iron (as gluconate) mg Plant-Derived Minerals 00 mg Magnesium (as citrate, gluconate) 00 mg * Serving size: fluid oz

Table. Gradient program and valve switching. Time (min) First Dimension (Left Pump) Valve Switching Second Dimension (Right Pump) Flow Rate (ml/min) % ( mm Phosphate uffer, ph.0) % (CH CN) Time (min) Right Valve Position Left Valve Position Time (min) Flow Rate (ml/min) % ( mm Phosphate uffer, ph.0) 0 0.8 00 0 0 0 0.8 00 0 0.8 00 0.9 0.8 00 0. 0.8.0.0.. 0.8 0 0.0 0.8 0.0 80.0. 0.8 00 0.0 0.8 0.0 80.0.89 0 0.8 00 0. 0.8 00 0.8 0 0.8 00 0 8...8..8..0.9. 8. % (CH CN) Conditions First Dimension Columns: For water-soluble vitamins except for V, Thermo Scientific cclaim Polardvantage (P), μm nalytical,. 0 mm (P/N 0) For V, Thermo Scientific Hypersil GOLD Phenyl nalytical HPLC, μm,. 0 mm (P/N 90-0) Mobile Phase:. mm phosphate buffer (dissolve ~. g KH PO in L water and adjust the ph to.0 with H PO ). CH CN Gradient: See Table Flow Rate: 0.8 ml/min Inj. Volume: 0 μl Temperature: C Detection: UV, absorbance at 0 and nm Second Dimension Column: cclaim 0 C8, μm nalytical,. 0 mm (P/N 098) Mobile Phase: Same as used in the First Dimension Flow Rate: 0.8 ml/min Temperature: C Detection: UV absorbance at 0,, 8, and 9 nm Results and Discussion Conventional HPLC Method for the Determination of Vitamins Currently, there is no U.S. Pharmacopeia (USP) method for the separation of a mixture of all eight water-soluble vitamins. USP method for individual vitamins is complicated (i.e., sodium perchlorate, phosphoric acid, dimethyl sulfoxide, acetonitrile, and water are needed for biotin determination) and involves an ion-pairing agent to retain hydrophilic vitamins. Due to the irreversible impact of the ion-pairing agent on column performance, extensive research was conducted to search for methods without ion-pairing agents, and this for a simpler mobile phase. These conditions apply to Figures through.

Recently, acidic or neutral phosphate buffer/organic solvent mobile phases have been used to separate vitamins in an extract of a multivitamin tablet and in vitaminenhanced functional drinks that are less complex than the one investigated here. For complex samples such as certain multivitamin nutritional drinks, the additional supplements can interfere with vitamin separation and, ultimately, their detection by UV absorbance. Figure shows that use of a typical HPLC method for a multivitamin nutritional drink results in the water-soluble vitamin peaks being hard or even impossible to quantify due to the large number of UV-absorbing interfering peaks. Obviously, Peaks cannot be quantified. Peaks 8 can be detected in the sample, but due to all the additional peaks, the vitamin peaks cannot be precisely quantified. Comparison of the UV spectra of the standard and the sample confirmed that Peaks 8 were not pure enough for quantification. The D-HPLC Method Two-dimensional HPLC has been used to achieve efficient separation of complex samples. Not only do two columns with different chemistries provide additional separation power, but use of heart-cutting technology in on-line D-HPLC also simplifies the separation in the second dimension. The work shown here uses D chromatography to determine water-soluble vitamins in a complex sample. Figure shows the configuration of the D-HPLC system used for this study. fter simple sample preparation (filter the sample, then dilute with DI water if necessary), inject the sample into the first dimension and partially separate it using an cclaim P column. The in-line UV detector determines where the water-soluble vitamins elute. n injection of the mixed standard determines the start and end times for each vitamin peak. Use these values to switch the valves. Switch the right valve to the _ position to individually transfer the vitamin peaks to the flow path of the second column. For early eluting vitamin peaks, switch the left valve to the _ position so that these vitamin fractions Column: cclaim P, µm nalytical,. 0 mm Mobile Phase:. mm phosphate buffer (dissolve ~. g KH PO in L water and adjust ph to.0 with H PO ). CH CN Gradient: CH CN, 0 min, 0%;. min, %; min, 80%;. 0 min, 0% Flow Rate: 0.8 ml/min Inj. Volume: 0 µl Temperature: C Detection: UV absorbance at 0 nm Samples:. Standard mixture. Multivitamin nutritional drink Peaks:. V 0 mg/l. V 00. V C 0. V 0. V 0. V. V 0 8. V 00 00 0 0 0 0 0 Figure. standard mixture () and a multivitamin nutritional drink sample () using a conventional HPLC method. are directly transferred to the second column. For late eluting vitamin peaks, switch the left valve to the _ position to put the 0 μl mixer in line. In this configuration, the water mobile phase in the second dimension will dilute the acetonitrile from the first-dimension mobile phase. This enables the second-dimension column to trap the vitamin peaks. Note: When the right valve is switched to the _ position, UV Detector will be connected between the two columns. However, the backpressure of the second-dimension column may exceed the pressure limit of the UV Detector flow cell. Thus, choose the proper particle size and length of the seconddimension column to keep the backpressure below the pressure limit of the flow cell. The pressure limit command in the instrument method can also be set to stop the flow in any situation when backpressure may become too high. 8 Dual Gradient Pump utosampler Column L Mixer R Dual Gradient Pump Waste Waste Detector Column Figure. Configuration of the D-HPLC system.

Development of the D-HPLC Method Valve switching To achieve optimal resolution and peak shape in D-HPLC applications, the critical part of method development is the transfer of first-dimension peaks to the second-dimension column. Ideally, the first-dimension mobile phase being cut and transferred with target peaks to the second dimension will be same as the seconddimension mobile phase at the time of transfer. ut in reality, the two mobile phases usually will have different concentrations of acetonitrile or other organic solvent. Figure, Chromatogram shows that Peaks,, and 8 disappear when they are directly transferred to the second dimension. Obviously, the late eluting peaks from the first column are contained in too high a concentration of acetonitrile to be retained on the second column. The traditional approach to this problem is to provide water to dilute the acetonitrile in the transferred fraction with a tertiary pump so the fraction can be retained on the trap column ahead of the second column. However, the need for an additional pump limits this application. Thermo Scientific pplication Note (N) 0 demonstrated an alternative way to solve this problem. riefly, a 0 μl mixer was configured in line before the eluted fraction was transferred to the second dimension. Then the peak mobile phase was mixed extensively with the second-dimension starting mobile phase (water phase) before the second-dimension separation. In this study, the authors initially applied the N 0 approach for all target peaks. It worked well for late eluting peaks, which had high concentrations of acetonitrile, but the early eluting peaks broadened (Figure, Chromatogram ). fter conducting additional experiments, the authors discovered that early eluting vitamin peaks broadened due to their physiochemical characteristics. These peaks cannot be trapped at the head of the second column, even with a 00% water mobile phase. Thus, the configuration was modified so that early eluting peaks are directly transferred to the second column, whereas late eluting peaks are transferred into the 0 μl mixer before the second column. Figure, Chromatogram C shows that the final configuration works well for all eight water-soluble vitamins. Detection: UV absorbance at 0 nm Valve Position: See Table Configurations:. 0 µl mixer always off-line. 0 µl mixer always in-line C. 0 µl mixer off-line for Peaks, in-line for Peaks 8 Peaks:. V 0 mg/l. V 00. V C 0. V 0. V 0. V. V 0 8. V 00 00 0 00 0 0 C 0 0 0 0 0 Figure. Second-dimension analysis of the standard mixture with three different configurations. Choice of wavelength for the UV detector Vitamins,,, and C were detected at, 8, 9, and nm, respectively, which are the wavelengths of maximum UV absorbance of each. Vitamin was detected at 8 nm, which is close to its maximum UV absorbance. The maximum UV absorbance wavelengths of V and V are below 00 nm; therefore, to minimize noise, both were detected at 0 nm. Vitamin was detected with the nm channel. If desired, a fifth channel of nm can be configured to detect V at its absorbance maximum. 8 8

Column choice and its impact on separation Several combinations of columns were tested. Previous work suggested that good choices for most water-soluble vitamins are an cclaim P column used as the firstdimension column and an cclaim 0 C8 column used as the second-dimension column. 8 Most water-soluble vitamins can be separated with good resolution using this column combination, but V coelutes with an impurity in the drink sample, making quantification of V impossible. The authors searched for another column to use for the first dimension. The Hypersil GOLD Phenyl column worked well for this purpose. Figure shows that V was not detected at nm when an cclaim P column was used in the first dimension. However, when the Hypersil GOLD Phenyl column was used in the first dimension, V was detected (Figure ). The authenticity of the V peak was confirmed by its UV spectrum. Detection: UV absorbance at nm Valve Position: See Table Samples:. Standard mixture. Multivitamin nutritional drink Detection: UV absorbance at nm Valve Position: See Table Samples:. V standard mixture (0 mg/l). Multivitamin nutritional drink sample 00 Peaks:. V 0 mg/l. V 0. V 0. V C 0. V 0. V Peak: 0. V 0 0 0 0 0 Figure. standard mixture () and a multivitamin nutritional drink sample () using an cclaim P column in the first dimension. 0 Figure. V standard () and a multivitamin nutritional drink sample () using a Hypersil GOLD Phenyl column in the first dimension.

Table. Reproducibility of retention time and peak area for water-soluble vitamins. Water-Soluble Vitamin Retention Time RSD Peak rea RSD V 0.0.08 V 0.0 0. V 0. 0. V 0.0 0. V 0.0 0. V 0.0 0.8 V 0.0 0.8 V C 0.. Reproducibility, Linearity, and Detection Limits Prior to sample analysis, reproducibility was estimated by making eight replicate injections of water-soluble vitamins. The RSDs for retention time and peak area are shown in Table. n overlay of the eight injections is shown in Figure. Calibration linearity for the water-soluble vitamins was investigated by making three replicate injections of a mixed standard prepared at five or six different concentrations. The external standard method was used to calculate the calibration curve and quantify these compounds in samples. Table reports the data from the calibration as calculated by Chromeleon CDS software. The authors found linear calibration curves for each vitamin over the ranges evaluated. The singlesided Student s t test method was used for estimating method detection limits (MDL). These data are reported in Table. 00 Detection: UV absorbance at 0 nm Valve Position: See Table Samples: _H, consecutive injections 8 H G F E D C Peaks:. V 0 mg/l. V 00. V C 0. V 0. V 0. V. V 0 8. V 00 0 0 0 0 0 Figure. Chromatogram overlays of eight consecutive injections of a mixture of water-soluble vitamin standards. Sample nalysis Vitamins and were not detected in the multivitamin nutritional drink, possibly due to their low concentration in the sample. Vitamin was surprisingly not detected. Vitamin was fully recovered when spiked into the sample at its labeled value; therefore, the method is capable of determining V. The other water-soluble vitamins were detected close to their labeled values. 8 Table. Calibration data and MDLs for the water-soluble vitamins. Water-Soluble Vitamin Detection Wavelength (nm) Range Regression Equation r MDL V 0. 0 =0.c-0.080 0.998 0.0 V 8 0. 0 =0.8c+0.0 0.9999 0. V 9 0. 0 =0.98c+0. 0.9999 0.8 V 0.0 00 =0.0c+0.0 0.9999 0. V 0. 0 =0.0c+0.0 0.9999 0.0 V 8 0. =0.98c-0.0 0.999 0.0 V 0.0 00 =0.00c 0.999. V C

lthough V C in the working standard solutions is quite unstable, its presence in the multivitamin nutritional drink appeared to be relatively stable. s shown in Figure, V C is a major peak in the sample chromatogram. Good recoveries of water-soluble vitamins in the spiked sample (Table ) provided another positive indicator of method accuracy. This D-HPLC method greatly simplifies the second-dimension chromatogram, as shown in Figure. 0 Detection: UV absorbance at nm Valve Position: See Table Sample: Multivitamin nutritional drink sample Peaks:. V. V. Unknown. V C. V. V Table. nalysis results of water-soluble vitamins in the multivitamin nutritional drink sample. nalyte Labeled (mg/ml)* Detected dded Found Recovery (%) V.0.98 9 V ND 0.9 9 V.. 0 V..0 V 0.0 ND 0.8 8 V.0 0.. 0 V 0.0 ND.0 pplication Note 09 0 0 0 0 Figure. Second dimension of multivitamin nutritional drink sample (00-fold dilution). Conclusion Two-dimensional HPLC simplifies the determination of the vitamin content of a multivitamin nutritional drink, a complex sample. nalysis of this complex sample requires only off-line filtration because the remainder of the sample preparation is automated by the UltiMate 000 x Dual HPLC system and Chromeleon CDS software. References. Dionex (now part of Thermo Scientific) pplication Note : Determination of Water- and Fat-Soluble Vitamins in Nutritional Supplements by HPLC with UV Detection. Sunnyvale, C, 00. [Online] www.dionex. com/en-us/webdocs/88-n-hplc-vitaminsupplements-june00-lpn.pdf (accessed Oct., 0).. Moreno, P.; Salvado, V. Determination of Eight Water- and Fat-Soluble Vitamins in Multivitamin Pharmaceutical Formulations by High-Performance Liquid Chromatography. J. Chromatogr., 000, 80, 0.. Perveen, S.; Yasmin, ; Khan, K.M. Quantitative Simultaneous Estimation of Water Soluble Vitamins, Riboflavin, Pyridoxine, Cyanocobalamin and Folic cid in Nutraceutical Products by HPLC. The Open nalytical Chemistry Journal 009,,. 0 V C Detected *The sample is diluted 000 times before analysis; thus, mg/ml will be detected as µg/ml. ND = not detected.. Dionex (now part of Thermo Scientific) pplication Note : Determination of Water- and Fat-Soluble Vitamins in Functional Waters by HPLC with UV-PD Detection. Sunnyvale, C, 009. [Online] www.dionex. com/en-us/webdocs/98-n-hplc-vitamins- FunctionalWaters-0pril09-LPN.pdf (accessed Oct., 0).. Vitamins/Dietary Supplements, The U.S. Pharmacopeia, NF 0, Washington, DC, 008, pp 9 0.. Dionex (now part of Thermo Scientific) pplication Note 8: Two-Dimensional HPLC Combined with On-Line SPE for Determination of Sudan Dyes I IV in Chili Oil. Sunnyvale, C, 0. [Online] www.dionex. com/en-us/webdocs/-n8-hplc-sudan- Dyes-Chili-Oil-Sept0-LPN99.pdf (accessed Oct., 0).. Thermo Scientific pplication Note 0: Determination of Sudan Dyes I IV in Curry Paste. Sunnyvale, C, 0. [Online] www.dionex.com/en-us/ webdocs/8-n0-lc-d-sudan-dyes-curry- Paste-N0_E.pdf (accessed Oct., 0). 8. Dionex (now part of Thermo Scientific) Technical Note 89: Determination of Water- and Fat-Soluble Vitamins by HPLC. Sunnyvale, C, 00. [Online] www.dionex.com/en-us/webdocs/888-tn89-hplc- WaterFatSolubleVitamins-Oct00-LPN98.pdf (accessed Oct., 0). www.thermoscientific.com/dionex 0 Thermo Fisher Scientific Inc. ll rights reserved. ISO is a trademark of the International Standards Organization. ll other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Fisher Scientific, Sunnyvale, C US is ISO 900 Certified. frica + 0 0 ustralia + 9 00 ustria + 80 8 0 elgium + razil + 0 Canada + 800 0 8 China 800 80 8 (free call domestic) 00 0 8 Denmark + 0 0 Europe Other + 0 0 Finland +8 9 9 000 France + 0 9 8 00 Germany +9 0 08 0 India +9 99 Italy +9 0 90 9 Japan +8 88 Korea +8 0 800 Latin merica + 88 800 Middle East + 0 0 Netherlands + 9 New Zealand + 9 980 00 Norway + 8 8 00 Russia/CIS + 0 0 Singapore + 89 90 Sweden + 8 8 00 Switzerland + 00 Taiwan +88 8 UK/Ireland + US + 800 N0_E 0/S