DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

Similar documents
LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Characteristics Analysis of High Energy External Radiotherapy Beams in Water

INTRODUCTION. Material and Methods

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

Detector comparison for small field output factor measurements with flattening filter free photon beams

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Comparative Study of Absorbed Doses in Different Phantom Materials and Fabrication of a Suitable Phantom

IN VIVO DOSIMETRY MEASUREMENTS FOR BREAST RADIATION TREATMENTS

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS. SB Crowe, E Whittle, C Jones, T Kairn

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

PMP. Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers in Low-Energy Electron Dosimetry. Original Article.

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

Quality assurance in external radiotherapy

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Small field diode dosimetry

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Ion recombination corrections of ionization chambers in flattening filter-free photon radiation

Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

Diode calibration for dose determination in total body irradiation

Analysis of 6 MV Energy Quality File Index using Percentage Depth Dose (PDD) and Tissue Phantom Ratio (TPR) Methods on Linac Siemens and Electa

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams

Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry

Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba.

Report to the 18 th Meeting of the CCRI(I), May 2007 Recent Activities in Measurement Standards and Dosimetry at METAS,

CHARACTERIZATION OF THE ELECTRON BEAM RADIATION FIELD BY CHEMICAL DOSIMETRY

Amendment No. 2. Item No. 2 (Rfx/ Event number )

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}

Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry

The Journey of Cyberknife Commissioning

Application(s) of Alanine

Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor

Protection of the contralateral breast during radiation therapy for breast cancer

Absolute Dosimetry. Versatile solid and water phantoms. Introduction Introduction Introduction Introtro Intro Intro

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results

Patient-specific quality assurance for intracranial cases in robotic radiosurgery system

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

AAPM s TG 51 Protocol for Clinical Reference Dosimetry of High-Energy Photon and Electron Beams

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

Diode Dosimetric Characteristics Assessment for In-Vivo Dosimetry in Radiotherapy Treatment

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy)

Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC)

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy

A comparative study between flattening filter-free beams and flattening filter beams in radiotherapy treatment

Standardization of dosimetry practices for small and large animal irradiation in radiobiological studies

GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

Optimization of the T2 parametric image map calculation in MRI polymer gel dosimetry

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

Learning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Dosimetry and QA of proton and heavier ion beams

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119

ISO INTERNATIONAL STANDARD

Applications of Monte Carlo simulations to radiation dosimetry

The determination of timer error and its role in the administration of specified doses

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

tsmall FIELD DOSE CALIBRATIONS WITH GAFCHROMIC FILM

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system

Commissioning and Radiobiology of the INTRABEAM System

Basic dosimetric data for treatment planning system

Neutron dose evaluation in radiotherapy

IAEA/WHO NETWORK OF SECONDARY STANDARD DOSIMETRY LABORATORIES


IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data

IMRT point dose measurements with a diamond detector

Stefan Gutiérrez Lores 1, Gonzalo Walwyn Salas 1, Daniel Molina Pérez 1, Raudel Campa Menéndez 2.

IORT with mobile linacs: the Italian experience

Transcription:

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National Institute for Laser, Plasma and Radiation Physics, 409 Atomistrilor str., Magurele, Ilfov, E-mail:eugenia_badita@yahoo.com 2 Clinical Hospital "Coltea", Radiotherapy Department, Bucharest, Romania 3 Bucharest of University, Physics Faculty, 407 Atomistilorstr, Magurele, Ilfov The beam quality Q of high energy photon beams produced by a clinical linear accelerator was assessed using the IAEA TRS-398, AAPM's TG-51 and DIN 6800-2 international protocols. The tissue phantom ratio TPR 20,10 is the beam quality Q of high energy photons produced by clinical accelerators and is measured at 10 g cm -2 and 20 g cm -2 depth in a water phantom with a constant source - phantom surface distance (SSD) of 100 cm and a field size of 10 cm x 10 cm at the plane of the chamber. According to AAPM's TG-51 the specified beam quality %dd(10) x is the percentage depth dose at 10 g cm -2 in a water phantom due only to photons. Key words: high energy photons, radiation therapy, tissue phantom ratio, clinical dosimetry. 1. INTRODUCTION The determination of absorbed dose delivered to a patient during radiotherapy treatment, is based on the measurement of absorbed dose to water [1]. The quality of a photon radiation beam is a basic quantity used to enable the commissioning of radiotherapy installations that generate photon beams, and is used in calculating the patient's received radiation dose. Also, in radiotherapy it is most usefully expressed in terms of its penetrating power which is mainly a function of the mean photon energy and may be fully described by its depth dose characteristics in water [2, 3]. The quality of a photon radiation beam based on the tissue-phantom ratio, TPR 20,10 in IAEA TRS-398 Code of Practice and DIN 6800-2. In the AAPM's TG- 51 the quality of photon radiation is specified by the percentage depth dose at 10 g cm -2 in a water phantom, % dd(10) x. Plane-parallel chambers can also be used to determine beam quality but were not used in this paper [4-8]. The tissue-phantom ratio is obtained from the ratio of the absorbed doses on the central axis at depths of 20 g cm -2 and 10 g cm -2 in a water phantom or by

2 measurement of the percentage depth dose (PDD) at depths of 20 g cm -2 and 10 g cm -2. After measuring, the PDD is useful firstly in radiotherapy treatment and secondly to evaluate and investigate the physics of radiation beams as function of field size, photon energy and source - surface phantom distance, SSD. Also, the measurement of absorbed dose is performed using cylindrical ionization chambers and using water or any other equivalent media phantom, which is kept perpendicular on the path of beam [4-8]. The measurements for determination of the photon beam quality were carried out at the Mevatron Primus clinical accelerator belonging to the Department of Radiation Oncology at the "Coltea" Clinical Hospital, Bucharest, Romania, at a constant source-phantom surface distance of 100 cm and a field size of 10 cm x 10 cm at the phantom surface [4]. For a better assessment of the obtained results and for demonstrating a good traceability of the measurements the z-score was calculated [1, 9]. 2. EQUIPMENT AND METHOD For determining the radiation quality of the photon beam at a Mevatron Primus clinical accelerator, a PTW MP3-M water phantom (PTW, Freiburg, Germany) with a scanning range of 50 cm x 50 cm x 40.75 cm was used. The percentage of absorbed dose was determined along the central axis at a source - phantom surface distance (SSD) of 100 cm, using two Semiflex chambers (Type TN 31010, PTW, Freiburg, Germany) with an active volume of 0.125 cc. For better precision, an ionization chamber was connected to the T10005 UNIDOS electrometer while the other was connected to the acquisition system of the Mevatron Primus clinical accelerator. The Mevatron Primus clinical accelerator that belongs to the Department of Radiation Oncology of the "Coltea" Clinical Hospital, Bucharest, Romania is manufactured by Siemens Medical Systems Inc-Oncology Care Group, Germany. This accelerator supplies electron beams and photon beams and has an adjustable field size [8]. The measurements presented in this paper were performed in a 6 MV high energy photon beam, with a field size of 10 cm x 10 cm. The dosimetric equipment that was used for the measurements belongs to Department of Radiation Oncology of the "Coltea" Clinical Hospital and to the Secondary Standard Dosimetry Laboratory at High Energies (STARDOOR) [10]. For the purposes of reference beam dosimetry, the beam quality in accelerator photon beams in IAEA TRS-398 and DIN 6800-2 is specified by the tissue-phantom ratio, TPR 20,10 and in AAPM's TG-51 the beam quality is specified by the percentage depth dose at 10 g cm -2 in a water phantom, % dd(10) x. Both are specific for high-energy photons produced by clinical accelerators with a beam quality Q but %dd(10) x does not include the effects of electron beam contamination

[4-8]. These are the basic procedures used to commission radiotherapy photon beams, and therefore are used in calculating the patient's radiation dose [2]. The method used for determining the TPR 20,10 constitutes in measuring absorbed dose and percentage depth dose at 10 g cm -2 and 20 g cm -2 in water phantom [4, 5, 8]. The quality of the photon beam was determined by utilizing an irradiation geometry that kept constant the source-phantom surface distance at 100 cm and a 10 cm x 10 cm field size at the phantom surface and at the chamber plane [4, 5, 10-12]. The ionization chamber TN 31010 was calibrated with 60 Co source and hence the chamber carried a calibration factor N D,w. Absorbed dose to water was determinate in according to the IAEA TRS-398, AAPM's TG-51 and DIN 6800-2 international protocols [4-7]. D Table 1 Formula used for the determination of absorbed dose to water IAEA TRS-398 AAPM's TG-51 DIN 6800-2 n Co 60 w, Q = ND, w, Q M 0 Q kq, Q D 0 w, Q = ND, w M kq Dw, Q = ( M M 0) N ki i= 1 In particular, TPR 20,10 is independent of the distance from the source and is determinate with the formulae [4, 5]: TPR =.2661 PDD 0.0595 (1) 20,10 1 20, 10 and TPR =.2661 D / D 0.0595 (2) 20,10 1 20 10 where PDD 20,10 /(D 20 /D 10 ) is the ratio of the percentage depth dose/absorbed dose at 20 g cm -2 and 10 g cm -2 depths for a field size of 10 cm x 10 cm at an SSD of 100 cm [4, 5]. The method used for determining the %dd(10) x constitutes in determining the percentage depth at 10 g cm -2 depth in a water phantom and is defined for a field size of 10 cm x 10 cm at the source-phantom surface distance of 100 cm. For lowenergy beams, i.e., for energies below 10 MV with %dd(10) 75% is calculated using the formula: % dd(10) x = % dd(10) (3) where the value of %dd(10) is the measured photon beam percentage depth dose at 10 g cm -2 depth in a 10 cm x 10 cm field on the surface of a water phantom at an SSD of 100 cm [6, 7]. 3

4 3. EXPERIMENTAL SET-UP The gantry was set to upright position and the water tank is set for a constant SSD of 100 cm. The field size was 10 cm x 10 cm at the surface of the water phantom. The distance and field sizes were measured using a laser system. All distances and field sizes were measured rigorously. The ionization chamber was positioned using a detector positioning system at the effective point of measurement. In the photon beam the PDD data was measured along the central axis at a constant SSD of 100 cm and a10 cm x 10 cm square field size [8, 12] and the ionization chamber was moved to different depths [11]. The measurements were performed at 6 MV energy. In order to measure and compare the results, the percentage depth dose was firstly measured using the Semiflex ionization chamber belonging to the STARDOOR Laboratory and then it was measured using the Semiflex ionization chamber belonging to the "Coltea" Hospital. The ionization chamber belonging to the STARDOOR laboratory was connected to the UNIDOS dosimeter and the ionization chamber belonging to the "Coltea" Hospital was connected to the Mevatron Primus accelerator acquisition system. The percentage depth dose was calculated by using the ionization chamber belonging to the STARDOOR laboratory to determine the absorbed dose in water for each point of measurement. The acquisition system of the Mevatron linear accelerator was used to measure the PDD directly [8]. For determining the radiation quality of the photon beam for this experimental set-up, according to IAEA TRS-398 and DIN 6800-2, we will consider the ratio of depth doses on the central axis at 20 g cm -2 and 10 g cm -2 depths, respectively D 20 /D 10 [4, 5]. The results of the measurements according to IAEA TRS-398 and DIN 6800-2 are equivalent. In according to AAPM's TG-51 only the percentage depth dose at 10 g cm -2 will be considered. To avoid experimental and stability errors the quality of the radiation beam was measured in six irradiation stages for the three international protocols. The pressure and temperature were monitored and recorded for each measurement. Correction for leakage and ion-recombination effect was applied and the uncertainty for measuring absorbed dose to water was calculated in conformity with reference standards. 4. RESULTS AND DISCUSSION The percentage depth dose measured with the two ionization chambers is presented in Fig.1. From the figures we can observe that the percentage depth dose measured with both ionization chambers is almost identical [12]. After determining the PDD, the quality of the radiation beam was calculated using the relationships 1

and 3. The determination of correction factors was performed according to each standard [4-6]. 5 a) b) Fig. 1 - The percentage depth dose: a) determined using the ionization chamber of the STARDOOR Laboratory; b) determined using the acquisition system of the "Coltea" Hospital The results of quality of radiation photon beam measured in six irradiation stages for the three international protocols are presented in Table 2 along with the specific correction factors for the beam radiation quality that were deduced for each IAEA TRS-398, AAPM's TG-51 and DIN 6800-2 protocols. The corresponding value was then interpolated for our ionization chamber TN 31010 [7]. Uncertainty for the measurements was calculated in conformity with the specified standards. These values are in the limit of uncertainty given by the manufacturer. Table 2 Photon beam radiation quality obtained at 6 MV at the Mevatron Primus clinical linear accelerator using the three international protocols (IAEA TRS-398, AAPM's TG-51 and DIN 6800-2) Photon IAEA TRS-398 AAPM's TG-51 DIN 6800-2 energy TPR 20,10 k Q,Qo %dd(10) x k Q TPR 20,10 k Q 6 MV 0.673 0.991 67.20 0.990 0.673 0.989 6 MV 0.673 0.991 67.41 0.990 0.673 0.989 6 MV 0.670 0.991 67.29 0.990 0.670 0.989 6 MV 0.673 0.991 66.37 0.992 0.673 0.989 6 MV 0.673 0.991 66.82 0.991 0.673 0.989 6 MV 0.672 0.991 67.49 0.990 0.672 0.989 Assessing the photon beam radiation quality was made using the z-score, determined according to the International Standard ISO 13528 [9]. The values of

6 the z score are presented in Fig. 2 and in Table 3 and are within the range [-2, +2] which proves that the quality system implemented in the STARDOOR laboratory (written procedures, trained personnel, etc.) ensure a high quality level of its services. Fig. 2 - z-score for measurement results at 6 MV photon beam for the quality of radiation photon beam Table 3 z-score for measurement results at 6 MV photon beam for quality photon beam Photon IAEA TRS-398 AAPM's TG-51 DIN 6800-2 energy TPR 20,10 z-score %dd(10) x z-score TPR 20,10 z-score 6 MV 0.673 0.71 67.20 0.39 0.673 0.71 6 MV 0.673 0.71 67.41 1.39 0.673 0.71 6 MV 0.670-1.43 67.29 0.48 0.670-1.43 6 MV 0.673 0.71 66.37-0.99 0.673 0.71 6 MV 0.673 0.71 66.82-0.76 0.673 0.71 6 MV 0.672 0.00 67.49-0.49 0.672 0.00 To compare the result of absorbed dose to water obtained in conformity with the three protocols, the absorbed dose to water was measured at reference depth, z ref = 5 g cm -2. IAEA TRS-398 has been taken as reference. The ratio of absorbed dose to water was obtained by dividing the doses of the protocols by the doses of dosimetry protocol IAEA TRS-398. The results presented in Fig. 3 show that the absorbed dose to water at 5 g cm -2 reference depth obtained following the three

protocols are close to each other. Also, Fig. 3 shows that the AAPM's TG-51 yielded approximately the same value as IAEA TRS-398 whereas DIN 6800-2 gave a minor deviation from the IAEA s protocol. The discrepancies are within acceptable range. The maximum dose ratio discrepancy when referencing AAPM's TG-51 and DIN 6800-2 protocols to the IAEA TRS 398 is between -0.001% to +0.18% [7]. 7 Fig. 3 - Comparison of the discrepancy between absorbed dose ratios 5. CONCLUSION In this article, the measurements for determining the quality of a 6 MV photon radiation beam energy beam have been performed at the Mevatron Primus clinical accelerator of the "Coltea" Clinical Hospital, Department of Radiation Oncology, Bucharest, Romania. The radiation quality was determined following the international protocols: IAEA TRS-398, AAPM's TG-51 and DIN 6800-2. According to IAEA TRS-398 and DIN 6800-2, the photon beam radiation quality refers to the tissue-phantom ratio, TPR 20,10 and in AAPM's TG-51 it is given in terms of percentage depth dose at 10g cm -2 in water phantom, %dd(10) x [4-6]. The photon radiation beam for 6 MV energy obtained with the IAEA TRS- 398 and DIN 6800-2 international protocols are similar, whereas the other measured with the AAPM's 6800-2 protocol is slightly different because the

8 method is different [7]. To show if the values obtained for the radiation quality are in accordance with the manufacturer's date the z score was calculated. The results obtained are within the range [-2; +2] which shows that the values obtained for the radiation quality with the three standards are within the range of the manufacturer's measuring range. Absorbed dose to water values, measured at a reference depth of 5 g cm -2, following the three protocols are similar. To compare the result of absorbed dose to water the IAEA TRS-398 protocol has been taken as reference. The AAPM's TG- 51 protocol yielded approximately the same value as IAEA TRS-398 whereas DIN 6800-2 gave a minor deviation from the IAEA s protocol. Despite the discrepancies the values are within acceptable range. The maximum dose ratio difference between the AAPM TG-51 and DIN 6800-2 protocol with reference to the IAEA TRS-398 is from -0.001% to +0.18%. This comes to demonstrate that the quality system implemented in the STARDOOR laboratory (written procedures, trained personnel, etc.) assures a high level of service quality. Acknowledgements: Work supported by the National Authority for Scientific Research and Innovation, Romania (Nucleus Program PN 0939/0105). REFERENCES 1. E. Badita, C. Vancea, I. Calina, D. Stroe, M. Dumitrache, E. Stancu, F. Scarlat, Long term stability of the performance of a clinical linear accelerator and z-score assessment for absorbed dose to water quantity, Romanian Reports in Physics, 69, No. 1, 606 (2017). 2. L. I. R. Garcia, J. F. Almansa, Technical Note: An algorithm to calculate the tissue phantom ratio from depth dose in radiosurgery, Medical Physics, 38, No. 5, p. 2359-2365 (2011). 3. B. Xhafa, T. Mulaj, G. Hodolli, G. Nafezi, Dose distribution of Photon by Siemens linear accelerator, International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, vol. 3, p. 67-70 (2014). 4. IAEA TRS 398-2004, Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry based on Standards of Absorbed Dose to Water, 23 April 2004. 5. DIN 6800-2, Dosimetry method according to the probe method for photon and Electron Radiation - Part 2: Dosimetry of high-energy photon and photonics Electron radiation with ionization chambers, March 2008. 6. AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams, 4 June 1999. 7. M. M. Hasan, M. A. Islam1, K. C. Paul and g. M. Zakaria, Dosimetric comparison in high energy photon beams for external beam radiotherapy following the protocols IAEA TRS-398, AAPM TG-51 and DIN 6800-2, Bangladesh Journal of Physics, 13, 77-85, June 2013. 8. R. Popa, M. Dumitrache, A. Ciovlica, A comparative study on 6 MeV photon beam percentage depth dose of VARIAN Clinac 2300 C/D, ELEKTA Synergy Platform, and SIEMENS Primus Linacs, Romanian Reports in Physics, vol. 64, No. 4, p. 997-1010 (2012). 9. Statistical methods for use in proficiency testing by interlaboratory comparisons, International standard ISO 13528:2005.

10. F. Scarlat, A. Scarisoreanu, R. Minea, E. Badita, E. Sima, M. Dumitrascu, E. Stancu, C. Vancea, Secondary standard dosimetry laboratory at INFLPR, Optoelectronic and Advanced Materials - Rapid Communication, Vol. 7, No. 7-8, p. 618-624 (2013). 11. International Atomic Energy Agency, Absorbed dose determination in photon and electron beams, IAEA TRS-277, Vienna, 1987. 12. G. Narayanasamy, W. Cruz, N. Papanikolaou, S. Stathakis, Comparison between measured tissue phantom ratio values and calculated from percent depth doses with and without peak scatter correction factor in a 6 MV beam, International Journal of Cancer Therapy and Oncology, vol. 3, No. 2 (2015). 13. E. Stancu, C. Vancea, J. Valenta, J. Zeman, E. Badita, A. Scarisoreanu, Assessment of absorbed dose to water in high energy photon beams using different cylindrical chambers, Romanian Reports in Physics, vol. 67, No. 2, p. 693-699 (2015). 14. E. Stancu, C. Vancea, J. Valenta, J. Zeman, E. Badita, A. Scarisoreanu, Absorbed to water measurements in high energy electron beams using different plane parallel chambers, Romanian Reports in Physics, vol. 67, No. 3, p. 1151-1157 (2015). 9