Reproducibility of Scanning Laser Doppler Flowmetry in the Rat Retina and Optic Nervehead

Similar documents
Real-Time Blood Velocity Measurements in Human Retinal Vein Using the Laser Speckle Phenomenon

N umerous techniques have shown vascular abnormalities

Diffuse luminance flicker increases blood flow in major retinal arteries and veins

LABORATORY SCIENCES. Depth of Penetration of Scanning Laser Doppler Flowmetry in the Primate Optic Nerve. may be anatomically divided

Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma

CLINICAL SCIENCES. The Effect of Ocular Warming on Ocular Circulation in Healthy Humans 1-3 RECENTLY

Retinal Blood Flow Autoregulation after Dynamic Exercise in Healthy Young Subjects

Effects of Oxygen and Carbon Dioxide on Human Retinal Circulation

Scanning Laser Tomography to Evaluate Optic Discs of Normal Eyes

Influence of diffuse luminance flicker on choroidal and optic nerve head blood flow

Comparison of Optic Disc Topography Measured by Retinal Thickness Analyzer with Measurement by Heidelberg Retina Tomograph II

Comparative Flow Velocity of Erythrocytes and Leukocytes in Feline Retinal Capillaries

Neovascular Glaucoma Associated with Cilioretinal Artery Occlusion Combined with Perfused Central Retinal Vein Occlusion

The Heidelberg Retinal Flowmeter (HRF, Heidelberg Engineering,

Inability of a Confocal Scanning Laser Doppler Flowmeter to Measure Choroidal Blood Flow in the Pig Eye

Case Report: Indocyanine Green Dye Leakage from Retinal Artery in Branch Retinal Vein Occlusion

Retinal tissue has an intrinsic ability to maintain a relatively

Correlation of Blue Chromatic Macular Sensitivity with Optic Disc Change in Early Glaucoma Patients

Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection

Corporate Medical Policy

scanning laser ophthalmoscope in rabbits

NERVE FIBER LAYER THICKNESS IN NORMALS AND GLAUCOMA PATIENTS

Advances in OCT Murray Fingeret, OD

NIH Public Access Author Manuscript Br J Ophthalmol. Author manuscript; available in PMC 2010 May 1.

The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation

Laser Speckle and Hydrogen Gas Clearance Measurements of Optic Nerve Circulation in Albino and Pigmented Rabbits With or Without Optic Disc Atrophy

Influence of Myopic Disc Shape on the Diagnostic Precision of the Heidelberg Retina Tomograph

A Patient s Guide to Diabetic Retinopathy

Vitrectomy for Diabetic Cystoid Macular Edema

Long-Term Visual Outcome in Proliferative Diabetic Retinopathy Patients After Panretinal Photocoagulation

Multifocal Electroretinograms in Patients with Branch Retinal Artery Occlusion

Spontaneous Intraocular Pressure Reduction in Normal-Tension Glaucoma and Associated Clinical Factors

This policy is applicable to Commercial Products only. For BlueCHiP for Medicare, see related policy section.

retina and optic nerve head

Ultrahigh Speed Imaging of the Rat Retina Using Ultrahigh Resolution Spectral/Fourier Domain OCT

Laser Doppler Velocimefry Study of the Effect of Pure Oxygen Breathing on Retinal Blood Flow

LABORATORY SCIENCES. Confocal scanning laser Doppler flowmetry

Visualize. Analyze. Personalize. OCT + OCTA

Retinal circulation during a spontaneous rise of intraocular pressure

Evaluation of optic disc blood flow of intraconal orbital tumors using laser speckle flowgraphy.

Morphometric changes of the choriocapillaris and the choroidal vasculature in eyes with advanced glaucomatous changes q

Technologies and Methods for Visualizing the Retina

Role of ocular Doppler ultrasonography in primary open angle glaucoma

Ophthalmology. Ophthalmology Services

Morphological Study of Corneal Endothelium and Corneal Thickness in Pseudoexfoliation Syndrome

T he retinal ganglion cells of different sizes have distinct

Intravenous Nicardipine in Cats Increases Optic Nerve Head But Not Retinal Blood Flow

Mitsuko Yuzawa,* Takako Isomae,* Ryuzaburo Mori,* Hiroyuki Shimada* and Izumi Utsunomiya

Thickness Changes in the Fovea and Peripapillary Retinal Nerve Fiber Layer Depend on the Degree of Myopia

OCT Angiography: The Next Step in Retinal Imaging Jonathan Zelenak D.O.

Reduced response of retinal vessel diameters to flicker stimulation in patients with diabetes

Year 2 MBChB Clinical Skills Session Ophthalmoscopy. Reviewed & ratified by: Mr M Batterbury Consultant Ophthalmologist

Intraocular irrigating solution containing epinephrine usually

RETINAL CONDITIONS RETINAL CONDITIONS

The Measure of Confidence

Use of Scanning Laser Ophthalmoscope Microperimetry in Clinically Significant Macular Edema in Type 2 Diabetes Mellitus

FACING YOUR FUNDIC FEARS: EXAMINATION OF THE OCULAR FUNDUS J. Seth Eaton, VMD, DACVO Cornell University Veterinary Specialists

Effect of Timolol Maleate on the Retinal Circulation of Human Eyes with Ocular Hypertension

OUR EYES & HOW WE SEE

Swept-Source OCT Angiography: SS OCT Angio TM

Measurement of Subfoveal Choroidal Thickness Using Spectral Domain Optical Coherence Tomography

Test-Retest Reliability of the CSV-1000 Contrast Test and Its Relationship to Glaucoma Therapy

Effects of Caffeine on Microcirculation of the Human Ocular Fundus

NIH Public Access Author Manuscript Br J Ophthalmol. Author manuscript; available in PMC 2014 November 01.

Hyperglycemia affects flicker-induced vasodilation in the retina of healthy subjects

Repeatability and Reproducibility of Retinal Blood Flow Measurement Using a Doppler Optical Coherence Tomography Flowmeter in Healthy Subjects

Diffuse luminance flicker increases retinal vessel diameter in humans

Retinal Hemodynamics in Proliferative Diabetic Retinopathy A Laser Doppler Velocimetry Study

GENERAL INFORMATION DIABETIC EYE DISEASE

Review Article. Use of laser speckle flowgraphy in ocular blood flow research. Introduction. Basic technology

ADVANCED DIAGNOSTIC TECHNIQUES

Scrub In. What is the function of vitreous humor? What does the pupil do when exposed to bright light? a. Maintain eye shape and provide color vision

RETINAL NERVE FIBER LAYER

Indocyanine Green Angiographic Findings of Chorioretinal Folds

UNDERSTAND MORE ABOUT UVEITIS UVEITIS

Oishi A, Miyamoto K, Yoshimura N. Etiology of carotid cavernous fistula in Japanese. Jpn J Ophthalmol. 2009;53:40-43.

Follow up of focal narrowing of retinal arterioles in glaucoma

Seiji T. Takagi, Yoshiyuki Kita, Asuka Takeyama, and Goji Tomita. 1. Introduction. 2. Subjects and Methods

Structural examina.on: Imaging

Choroidal Blood Flow Change in Eyes with High Myopia

Morphometric Features of Laminar Pores in Lamina Cribrosa Observed by Scanning Laser Ophthalmoscopy

The ideal tool for early detection and monitoring of AMD.


Ophthalmologic Techniques That Evaluate the Posterior Segment for Glaucoma

Transient Intraocular Pressure Elevation after Trabeculotomy and its Occurrence with Phacoemulsification and Intraocular Lens Implantation

Long-Term Prognosis of Extracapsular Cataract Extraction and Intraocular Lens Implantation in Diabetic Patients

Zinn Haller arterial ring observed by ICG angiography in high myopia

Optic nerve head morphometry in healthy adults using confocal laser scanning tomography

Retinal blood flow reduction after panretinal photocoagulation in Type 2 diabetes mellitus: Doppler optical coherence tomography flowmeter pilot study

* Şükrü Bayraktar, MD, Zerrin Bayraktar, MD, and Ömer Faruk Yilmaz, MD

Automated Detection of Vascular Abnormalities in Diabetic Retinopathy using Morphological Entropic Thresholding with Preprocessing Median Fitter

Role of Nitric Oxide in Optic Nerve Head Blood Flow Regulation during Isometric Exercise in Healthy Humans MATERIALS AND METHODS.

MEDICAL POLICY. Proprietary Information of Excellus Health Plan, Inc. A nonprofit independent licensee of the BlueCross BlueShield Association

Eye Examination Techniques in Horses

Diabetic retinopathy damage to the blood vessels in the retina. Cataract clouding of the eye s lens. Cataracts develop at an earlier age in people

Analysis of Changes in Corneal Shape and Refraction Following Scleral Buckling Surgery

Widefield Retinal Imaging with Auto Fluorescence Technology in the Optometric Practice

Retinal blood flow in type 1 diabetes patients with no or mild diabetic retinopathy during euglycemic clamp

Experience Spectacular Retinal Imaging with the new NIDEK F-10 Digital Ophthalmoscope

ASSESSING THE EYES. Structures. Eyelids Extraocularmuscles Eyelashes Lacrimal glands: Lacrimal ducts Cornea Conjunctiva Sclera Pupils Iris.

Transcription:

Reproducibility of Scanning Laser Doppler Flowmetry in the Rat Retina and Optic Nervehead Akitaka Tsujikawa,* Junichi Kiryu,* Atsushi Nonaka,* Kenji Yamashiro,* Yuichiro Ogura and Yoshihito Honda* *Department of Ophthalmology and Visual Sciences, Kyoto University Graduate School of Medicine, Kyoto, Japan; Department of Ophthalmology, Nagoya City University Medical School, Nagoya, Japan Purpose: The purpose of this study is to confirm the reliability of scanning laser Doppler flowmetry in the rat retina and optic nervehead, and the validity of measuring changes of retinal blood flow in rats while breathing 100% oxygen. Methods: We used a commercially available scanning laser Doppler flowmeter. To ascertain reliability, five consecutive and separate perfusion measurements of 12 eyes of 12 anesthetized pigmented rats were performed. To evaluate the validity of the system, repeated measurements were taken in anesthetized rats breathing room air or 100% oxygen. This series of measurements was repeated three times. Results: The reliability coefficients of volume, flow, and velocity in the optic nervehead and the retina ranged from 0.80 to 0.83 and 0.77 to 0.82, respectively. After the first exposure to oxygen, the measured values of volume, flow, and velocity were reduced by an average of 20.9 24.0%, 21.2 28.2%, and 19.5 24.5%, respectively. After the values returned to the basal condition, the second and third exposures to oxygen yielded measured values that were reduced by the same amounts as at the first exposure. Conclusions: Scanning laser Doppler flowmetry provided relatively good reliability in measurements of blood flow in the rat retina and optic nervehead. This study has indicated the possibility of applying this system to the rat retina. Jpn J Ophthalmol 2000;44:257 262 2000 Japanese Ophthalmological Society Key Words: Blood flow, rat, retina, scanning laser Doppler flowmeter. Introduction Received: May 6, 1999 Correspondence and reprint requests to: Junichi KIRYU, MD, Department of Ophthalmology and Visual Sciences, Kyoto University Graduate School of Medicine, Sakyo-ku, Kyoto 606-8507, Japan Various methods have been developed to measure retinal blood flow in humans, including laser Doppler velocimetry, 1,2 laser Doppler flowmetry, 3 dye dilution technique, 4 blue field entoptic phenomenon, 5 laser speckle phenomenon, 6 and scanning laser angiography. 7 There are, however, many ethical and statistical limitations in using humans for research. Animal methods are another way to investigate the pathogenesis of diseases and innovative therapies. The results from animal studies should help bring about better understanding of the pathological processes in humans. Therefore, it would be valuable to establish one method that could be applied to both humans and animals. The rat is a familiar animal model in ophthalmological experiments. However, few methods are available to evaluate retinal blood flow in both human and rats other than the dye dilution technique. 8 Scanning laser Doppler flowmetry 9 is a relatively new system to evaluate regional blood flow in the retina 10 and the optic nervehead, 11 noninvasively and repeatedly. In this system, laser Doppler flowmetry is applied to a scanning laser system. In many previous studies, relatively good reliability has been obtained in measurements of human retinal blood flow with this method, 10 12 but, at the same time, the limitations of this system have been discussed in many reports. 13 15 So far, scanning laser Doppler flowmetry has been applied to many patients, especially to Jpn J Ophthalmol 44, 257 262 (2000) 2000 Japanese Ophthalmological Society 0021-5155/00/$ see front matter Published by Elsevier Science Inc. PII S0021-5155(99)00226-9

258 Jpn J Ophthalmol Vol 44: 257 262, 2000 glaucoma patients. 16 18 To our knowledge, few reports, however, have described the application of this system to animal research. 19 The purpose of this study is to evaluate the reliability of a commercial scanning laser Doppler flowmeter in the rat retina and optic nervehead and its validity by measuring changes of retinal blood flow in rats when breathing 100% oxygen. Reliability Materials and Methods We used a commercially available scanning laser Doppler flowmeter (Heidelberg Engineering, Heidelberg, Germany). To determine the reliability of scanning laser Doppler flowmetry, we used male pigmented Long-Evans rats (200 250 g). Five consecutive and separate perfusion measurements of 12 eyes of 12 rats were performed. After the animals were anesthetized with intramuscular injection of xylazine hydrochloride (4 mg/kg) and ketamine hydrochloride (10 mg/kg), the pupils were dilated with 0.5% tropicamide and 2.5% phenylephrine hydrochloride eye drops. A contact lens was used to retain corneal clarity throughout the experiment. Each rat was placed on a stereotaxic platform in a prone position. The fundus of a rat was scanned with a scanning laser Doppler flowmeter in a 20 field around the optic nervehead and temporal to the optic disc. To compare several measurements from the same retinal location, we set the region of interest (ROI) squares of 10 10 pixels away from large visible vessels in the retina. Figure 1 shows the 4 ROIs used in this study. All experiments were performed in accordance with the ARVO Statement for the Use of Animals in Ophthalmic and Vision Research. In general, the measured value combines the true mean value, the biological variability component, To evaluate the validity of the system, we used 10 eyes of 10 male pigmented Long-Evans rats. After being anesthetized, their pupils were dilated with the same agents used in the reliability study. While the rats were breathing room air, their arterial blood pressure and heart rate were monitored with a blood pressure analyzer (IITC; Woodland Hill, CA, USA), and the fundus was scanned with a scanning laser Doppler flowmeter in a 20 field. Then the rat was exposed to 100% oxygen over 5 minutes in a closed cage, in which 100% oxygen flowed at a rate of 6 L per minute. Immediately after the rat was removed from the cage, within 10 seconds, the arterial blood pressure and heart rate were measured and the fun- and the error component as indicated in Eq. (1). 10,11 We calculated the reliability coefficient r1 by estimating the variances of the biological component and the error component. The variances of the two components are defined in Eq. (2). The reliability coefficient r1 is defined in Eq. (3). xij = x + ( xi x ) + ( xij xi) = + ai + eij (1) where xij measurement value; x mean of all measurement values; x i mean of measurement values of each eye; mean value; ai biological component; eij measuring error component; I 1,..., n eyes; j 1,...m measurements. ΣΣ( xij x) 2 = mxi ( x) 2 + ΣΣ( xij xi) 2 ST = SA + SE (2) where ST variance of measurement value; SA variance of biological component; SE variance of error component. r1 = SA ( SA + SE) (3) Validity Figure 1. Fundus image of rat retina obtained with use of scanning laser Doppler flowmetry. Regions of interest (ROI) 1 4 represent areas of evaluation of blood flow (10 10 pixels). ROI 1 is on optic nervehead; ROI 2 4 are on retina; ROI 2 is at edge of optic nervehead; ROI 2 and 3 are at a distance of 1 or 3 disc diameters from edge of optic nervehead. All ROI are positioned to spare major retinal vessels.

A. TSUJIKAWA ET AL. 259 SLDF IN THE RAT RETINA Figure 2. Perfusion map of rat retina. Major retinal vessels were identified with bright colors, because blood flow velocity is very fast. Left: cursor is positioned at region of interest (ROI) 2. Right: cursor is positioned at ROI 4. dus was scanned with a scanning laser Doppler flowmeter. After the rat breathed room air for 5 minutes, the same measurements were performed. Each series of measurements was repeated three times. After the experiment, we evaluated the blood flow in the same 4 ROIs mentioned. All values were presented as mean SD. Analysis of variance was used to compare measured values, with post-hoc comparisons tested using Scheffe s F- test. Differences were considered statistically significant at P.05. Reliability Results Figure 2 shows a perfusion map of a rat retina produced using scanning laser Doppler flowmetry. As the blood flow velocity in the major retinal vasculature was very fast, both retinal veins and arteries appear brightly colored. Table 1 shows the reliability coefficients r1 of volume, flow, and velocity for each ROI. The reliability coefficients r1 of volume, flow, and velocity of the optic nervehead (ROI 1), which were 0.80 to 0.83, were good. The reliability coefficients r1 of the retina (ROI 2, 3, and 4) ranged from 0.77 to 0.82. Most of them were relatively low compared with those of the optic nervehead (ROI 1), but some were not. Validity In the basal condition, measured values of volume, flow, and velocity of the optic nervehead (ROI 1) were 20.6 2.6 arbitrary units (AU), 654 124 AU, and 2.19 0.40 AU, respectively. Measured values of volume, flow, and velocity of the retina (ROI 2 4), which were 16.4 21.7 AU, 513 670 AU, and 1.78 2.24 AU, were not uniform (P.0033, P.027, and P.032, respectively). The blood flow of the retina was reduced near the optic disc. Measured values of volume, flow, and velocity of ROI 2 were reduced by 34.0%, 23.5%, and 20.3%, respectively, compared with those of ROI 4, in the normal condition. Table 2 shows the heart rate and the mean arterial blood pressure of the rats during oxygen exposure. There was no significant change throughout the experiments. With oxygen exposure, measured values of volume, flow, and velocity were significantly reduced (P.001) (Figure 3). After the first exposure to oxygen, the average reductions of measured values of volume, flow, and velocity were 20.9 24.0%, 21.2 28.2%, and 19.5 24.5%, respectively. While the rats were breathing room air after the first exposure to 100% oxygen, measured values returned to those of the basal condition (98.6 100.8%). At the second and third exposures to oxygen, measured values were reduced as much as at the first exposure (74.3 81.6%). While the rats were breathing room air, there were no significant changes of measured values compared with those of the basal condition. Discussion In this study, we investigated the reliability and the validity of applying a commercial scanning laser Table 1. Reliability Coefficients r1 of Volume, Flow, and Velocity at Each Measurement Point Region of Interest Volume Flow Velocity 1 0.82 0.83 0.80 2 0.77 0.81 0.81 3 0.80 0.81 0.82 4 0.77 0.81 0.79

260 Jpn J Ophthalmol Vol 44: 257 262, 2000 Table 2. Heart Rate and Mean Arterial Blood Pressure of Rats During Oxygen Exposure Air O 2 Air O 2 Air O 2 Air Heart rate (beats/min) 285 14 290 21 278 14 268 9 290 20 276 19 282 22 MABP (mm Hg)* 97.6 19.7 100.4 13.3 96.0 16.1 90.4 11.7 92.5 14.6 99.5 15.5 90.0 14.6 *MABP mean arterial blood pressure. There were no statistically significant changes in heart rate or MABP throughout experiment. Doppler flowmeter to the rat retina and optic nervehead. In most previous studies, a scanning laser Doppler flowmetry measurement of the human fundus in a 10 field was reported to depict an area of about 0.7 mm 2.7 mm. 17,18,20 In a perfusion map, 1 pixel represented about 10 m. 9,21 Because the axial length of the rat eye is about one-fourth that of the human eye, the rat fundus is visualized at a higher magnification than that of the human eye. After measurements of blood flow, the rats were sacrificed with an anesthetic overdose, and their eyes were enucleated to calibrate areas of ROI on the perfusion map (pixel 2 ) with real areas on the retina ( m 2 ). In this study, in which the rat fundus was scanned with scanning laser Doppler flowmetry in a 20 field, 1 pixel in the perfusion map represented an area of about 5 m. Initially, Michelson and Schmauss 11 reported that the reliability coefficients of scanning laser Doppler flowmetry of the optic nervehead in healthy volunteers were 0.84 0.85. Many subsequent reports have indicated good reliability in applying scanning laser Doppler flowmetry to the juxtapapillary retina 13 and the macula. 12 Michelson et al 10 have reported that the reliability coefficients of the juxtapapillary retina were 0.81 0.83. In this study, reliability coefficients of the optic nervehead and retina were 0.80 0.83 and 0.77 0.82, respectively. Similar to previous investigations in humans, 10,11 measurements of the optic nervehead had a tendency to show higher reliability than those of the retina. While good reliability was demonstrated in applying scanning laser Doppler flowmetry to rats, the reliability coefficients were somewhat lower than those found in humans. Simi- Figure 3. Changes of measurement variables (volume, flow, and velocity) of optic nervehead (region of interest [ROI] 1) and retina (ROI 2 4) during oxygen exposure.

A. TSUJIKAWA ET AL. 261 SLDF IN THE RAT RETINA lar to the finding of Michelson et al, 9,10,18 the reliability coefficient in this study was defined as the ratio of biological variation to the sum of biological variation and error variation. In human research, subjects have many biological variations due to sex, age, and physical characteristics. Conversely, the rats used in this study were so homologous that biological variability was minimal, which could account for the underestimation of the reliability coefficients. In previous studies of healthy volunteers, 12,22 the changes of optic nerve or retinal blood flow during oxygen breathing detected with scanning laser Doppler flowmetry has been described. Langhans et al 22 have reported that blood flow in the optic nervehead and juxtapapillary retina was reduced by 33 37% with oxygen breathing in nonsmoking volunteers. Another study by Strenn et al 12 indicated that the reduction of macular blood flow was 29 33% during 100% oxygen breathing. In our study, the values of flow were reduced by 24% and 23 26% at the optic nervehead and retina, respectively, after oxygen breathing. The changes of the retinal blood flow in rat retina were relatively compatible with the results demonstrated in experiments with humans. 12,22 The reduction of blood flow during oxygen exposure, however, was relatively small, as compared with the values measured by other methods. With use of the dye dilution technique, Takagi et al 8 have reported that retinal blood flow in rats was reduced by 62% with 100% oxygen exposure for 15 minutes. While our results reflect regional blood flow in capillaries, they measured total retinal blood flow. 8 The changes of the total blood flow might not necessarily agree with the changes of the regional blood flow measured by a scanning laser Doppler flowmeter, because there are many anastomoses between retinal arteries and veins or between retinal vessels and choroidal vessels in the rat. Moreover, one limitation of scanning laser Doppler flowmetry may be the large zero off-set, 12 because it cannot be determined in vivo. 15 This off-set can cause an underestimation of the change compared with the basal condition, which could account for the relatively small changes detected with scanning laser Doppler flowmetry during oxygen breathing. This feature would indicate the need for caution in interpreting the results measured by scanning laser Doppler flowmetry. The limitations in applying scanning laser Doppler flowmetry in humans have been discussed. When applying the method to rats, other problems may arise. Because scanning laser Doppler flowmetry scans with a repetition rate of 4,000 Hz, Doppler frequencies up to 2,000 Hz can be detected with reliability. 9,21 Although it has been reported that Doppler broadenings in the normal retina range from 300 Hz to 800 Hz, 3,10 larger Doppler shifts may occur due to retinal diseases or the administration of an agent. It may not be possible to measure retinal blood flow in all cases. Also, media opacity often degrades the images obtained with scanning laser Doppler flowmetry. To avoid drying of the corneas, the anesthetized rats need to wear a contact lens to retain corneal clarity throughout the experiment. Although the human subject may find fixation on one point difficult for more than 2 seconds without eye movement, fixation does not appear to be a problem in using anesthetized rats. However, the eye of the rat is so small that correction of the refractive error is needed. The ability of the scanning laser Doppler flowmeter is such that it can obtain measurements in eyes from 12 diopters to 12 diopters, which is sufficient in adult-sized rats. Dye dilution technique, 4,8 which is the other available method to evaluate blood flow in both the rat and human retina, is too invasive to measure blood flow repeatedly. In contrast, scanning laser Doppler flowmetry can be used to perform repeated measurements within short or long time periods. 9,11 Although there may be limitations in applying scanning laser Doppler flowmetry to the rat retina, 12,15,21 this study has indicated the possibility of applying this system to the rat retina. This work was supported by a Grant-in-Aid for Scientific Research from the Ministry of Education, Science, Sports, and Culture of Japan. References 1. Pakola SJ, Grunwald JE. Effects of oxygen and carbon dioxide on human retinal circulation. Invest Ophthalmol Vis Sci 1993;34:2866 70. 2. Feke GT, Riva CE. Laser Doppler measurements of blood velocity in human retinal vessels. J Opt Soc Am 1978;68:526 31. 3. Riva CE, Harino S, Petrig BL, Shonat RD. Laser Doppler flowmetry in the optic nerve. Exp Eye Res 1992;55:499 506. 4. Riva CE, Feke GT, Ben-Sira I. Fluorescein dye-dilution technique and retinal circulation. Am J Physiol 1978;234:H315 22. 5. Sinclair SH, Loebl M, Riva CE. Blue field entoptic phenomenon in cataract patients. Arch Ophthalmol 1979;97:1092 5. 6. Tamaki Y, Araie M, Tomita K, Nagahara M, Tomidokoro A, Fujii H. Real-time measurement of human optic nerve head and choroid circulation, using the laser speckle phenomenon. Jpn J Ophthalmol 1997;41:49 54. 7. Wolf S, Arend O, Toonen H, Bertram B, Jung F, Reim M. Retinal capillary blood flow measurement with a scanning laser ophthalmoscope. Preliminary results. Ophthalmology 1991;98:996 1000. 8. Takagi C, King GL, Takagi H, Lin YW, Clermont AC,

262 Jpn J Ophthalmol Vol 44: 257 262, 2000 Bursell SE. Endothelin-1 action via endothelin receptors is a primary mechanism modulating retinal circulatory response to hyperoxia. Invest Ophthalmol Vis Sci 1996;37:2099 109. 9. Michelson G, Langhans MJ, Groh MJ. Clinical investigation of the combination of a scanning laser ophthalmoscope and laser Doppler flowmeter. Ger J Ophthalmol 1995;4:342 9. 10. Michelson G, Schmauss B, Langhans MJ, Harazny J, Groh MJ. Principle, validity, and reliability of scanning laser Doppler flowmetry. J Glaucoma 1996;5:99 105. 11. Michelson G, Schmauss B. Two dimensional mapping of the perfusion of the retina and optic nerve head. Br J Ophthalmol 1995;79:1126 32. 12. Strenn K, Menapace R, Rainer G, Findl O, Wolzt M, Schmetterer L. Reproducibility and sensitivity of scanning laser Doppler flowmetry during graded changes in PO. Br J Ophthalmol 1997;81:360 4. 13. Nicolela MT, Hnik P, Schulzer M, Drance SM. Reproducibility of retinal and optic nerve head blood flow measurements with scanning laser Doppler flowmetry. J Glaucoma 1997;6: 157 64. 14. Kagemann L, Harris A, Chung HS, Evans D, Buck S, Martin B. Heidelberg retinal flowmetry: factors affecting blood flow measurement. Br J Ophthalmol 1998;82:131 136. 15. Van Heuven WAJ, Kiel JW, Elliot WR, Harrison JM, Sponsel WE. Evaluation of the Heidelberg retina flowmeter. Invest Ophthalmol Vis Sci 1996;37:S967. 16. Kerr J, Nelson P, O Brien C. A comparison of ocular blood flow in untreated primary open-angle glaucoma and ocular hypertension. Am J Ophthalmol 1998;126:42 51. 17. Nicolela MT, Hnik P, Drance SM. Scanning laser Doppler flowmeter study of retinal and optic disk blood flow in glaucomatous patients. Am J Ophthalmol 1996;122:775 83. 18. Michelson G, Langhans MJ, Groh MJ. Perfusion of the juxtapapillary retina and the neuroretinal rim area in primary open angle glaucoma. J Glaucoma 1996;5:91 8. 19. Okamoto N, Harino S, Kitanishi K, Suzuki M. Evaluation of scanning laser Doppler flowmetry in aneshetized cat. Nippon Ganka Gakkai Zasshi (J Jpn Ophthalmol Soc) 1998;102:587 93. 20. Groh MJ, Michelson G, Langhans MJ, Harazny J. Influence of age on retinal and optic nerve head blood circulation. Ophthalmology 1996;103:529 34. 21. Chauhan BC, Smith FM. Confocal scanning laser Doppler flowmetry: experiments in a model flow system. J Glaucoma 1997;6:237 45. 22. Langhans M, Michelson G, Groh MJ. Effect of breathing 100% oxygen on retinal and optic nerve head capillary blood flow in smokers and non-smokers. Br J Ophthalmol 1997;81: 365 9.