Dose reduction in Hologic Selenia FFDM units through AEC optimization, without compromising diagnostic image quality.

Similar documents
Lifetime risk of radiation-induced cancer from screening mammography

Standardizing mammographic breast compression: Pressure rather than force?

Breast asymmetries in mammography: Management

PGMI classification of screening mammograms prior to interval cancer. Comparison with radiologists' consensus classification.

BI-RADS 3, 4 and 5 lesions on US: Five categories and their diagnostic efficacy and pitfalls in interpretation

64-MDCT imaging of the pancreas: Scan protocol optimisation by different scan delay regimes

Slowly growing malignant nodules and rapidly growing benign nodules: Evaluation of the value of volume doubling time

Digital breast tomosynthesis (DBT) occult breast cancers: clinical, radiological and histopathological features.

Single cold nodule in Graves' disease: benign vs malignant

Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA

Evaluation of BI-RADS 3 lesions in women with a high risk of hereditary breast cancer.

Role of positron emission mammography (PEM) for assessment of axillary lymph node status in patients with breast cancer

Correlation Between BIRADS Classification and Ultrasound -guided Tru-Cut Biopsy Results of Breast Lesions: Retrospective Analysis of 285 Patients

Does the synthesised digital mammography (3D-DM) change the ACR density pattern?

Cognitive target MRI-TRUS fusion biopsies of MRI detected PIRADS 4 and 5 lesions

AFib is the most common cardiac arrhythmia and its prevalence and incidence increases with age (Fuster V. et al. Circulation 2006).

The Role of Radionuclide Lymphoscintigraphy in the Diagnosis of Lymphedema of the Extremities

Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography

The effect of CT dose reduction on performance of a diagnostic task

The effect of CT dose reduction on performance of a diagnostic task

Hyperechoic breast lesions can be malignant.

Purpose. Methods and Materials. Results

Direct half value layer measurements in mammography - is near enough good enough?

Breast ultrasound appearances after Mammotome vacuumassisted

Diffuse high-attenuation within mediastinal lymph nodes on non-enhanced CT scan: Usefulness in the prediction of benignancy

Identification and numbering of lumbar vertebrae using various anatomical landmarks on MRI of lumbosacral spine

BI-RADS classification in breast tomosynthesis. Our experience in breast cancer cases categorized as BI-RADS 0 in digital mammography

Spectrum of findings of sclerosing adenosis at breast MRI.

Artifact in Head CT Images Due to Air Bubbles in X-Ray Tube Oil

Computed tomography for pulmonary embolism: scan assessment of a one-year cohort and estimated cancer risk associated with diagnostic irradiation.

Sonographic and Mammographic Features of Phyllodes Tumours of the Breast: Correlation with Histological Grade

BI-RADS 3 category, a pain in the neck for the radiologist which technique detects more cases?

Diffusion-weighted MRI (DWI) "claw sign" is useful in differentiation of infectious from degenerative Modic I signal changes of the spine

Correlation between lesion type and the additional value of digital breast tomosynthesis

Seemingly isolated greater trochanter fractures do not exist

Multislice CT versus cone beam CT: dosimetric and image quality comparision.

FDG-18 PET/CT - radiation dose and dose-reduction strategy

Radiological features of Legionella Pneumophila Pneumonia

Application of international standards to diagnostic radiology dosimetry

Radiologic Findings of Mucocele-like Tumors of the breast: Can we differentiate pure benign from associated with high risk lesions?

Scientific Exhibit Authors: V. Moustakas, E. Karallas, K. Koutsopoulos ; Rodos/GR, 2

Bolus administration of esmolol allows for safe and effective heart rate control during coronary computed tomography angiography

Basic low - field MR imaging of meniscal injuries in children.

Feasibility of magnetic resonance elastography using myofascial phantom model

Ethanol ablation of benign thyroid cysts and predominantly cystic thyroid nodules: factors that predict outcome.

Radiologic and pathologic correlation of non-mass like breast lesions on US and MRI: Benign, high risk, versus malignant

Radiologic and pathologic correlation of non-mass like breast lesions on US and MRI: Benign, high risk, versus malignant

THI-RADS. US differentiation of thyroid lesions.

THI-RADS. US differentiation of thyroid lesions.

Monophasic versus biphasic contrast application in CT of patients with head and neck tumour

Radiation dose levels assessment in mammography

MR-guided prostatic biopsy at 3T: the role of PI-RADS-score: a histopahologic-radiologic correlation

Triple-negative breast cancer: which typical features can we identify on conventional and MRI imaging?

Long bones manifestations of congenital syphilis

CT Fluoroscopy-Guided vs Multislice CT Biopsy ModeGuided Lung Biopies:a preliminary experience

Computed tomographic dacryocystography as compared with X-ray dacryocystography in patients with dacryostenosis

Computed tomography and Modified RECIST criteria for assessment of response in malignant pleural mesothelioma

Comparison of one-view digital breast tomosynthesis (DBT) and two-view full-field digital mammography (FFDM)

Medical device adverse incident reporting in interventional radiology

Role of ultrasound in the evaluation of the ileocecal valve

Pathologic outcomes of coarse heterogeneous calcifications detected on mammography

Influence of pulsed fluoroscopy and special radiation risk training on the radiation dose in pneumatic reduction of ileocoecal intussusceptions.

Valsalva-manoeuvre or prone belly position for computed tomography (CT) scan when an orbita varix is suspected: a single-case study.

Single ventricle on cardiac MRI

Shear Wave Elastography in diagnostics of supraspinatus tendon.

The imaging evaluation of breast implants

User Guide for Dental and Maxillofacial Cone Beam Computed Tomography (CBCT)

Cavitary lung lesion: Two different diagnosis with similar appearence

Mammographic and ultrasonographic study of patients who underwent breast augmentation through injection of hyaluronic acid (Macrolane )

High density thrombi of pulmonary embolism on precontrast CT scan: Is it dangerous?

Cairo/EG, Khartoum/SD, London/UK Biological effects, Diagnostic procedure, Ultrasound, Mammography, Breast /ecr2015/C-0107

Triple Negative Breast Cancer: Clinical Presentation and Multimodality Imaging Characteristics

Scientific Exhibit Authors:

Ultrasonic evaluation of superior mesenteric vein in cancer of the pancreatic head

MRI BI-RADS: How to make it out?

Intracystic papillary carcinoma of the breast

Comparison of MRI and ultrasound based liver volumetry in iron overload diseases

Quantitative imaging of hepatic cirrhosis on abdominal CT images

Quantifying Dual-energy computed tomography (DECT) in patients with renal calculi using a Toshiba Aquilion One Scanner.

Vacuum-assisted breast biopsy using computer-aided 3.0 T- MRI guidance: diagnostic performance in 173 lesions

Computed tomography for the detection of thumb base osteoarthritis, comparison with digital radiography.

Reliability of the pronator quadratus fat pad sign to predict the severity of distal radius fractures

Improvement in ROC curves of readers with next generation of mammography CAD

The solitary pulmonary nodule: Assessing the success of predicting malignancy

Tissue characterisation, Cancer, Quality assurance /ecr2015/B-0553

"Ultrasound measurements of the lateral ventricles in neonates: A comparison of multiple measurements methods."

How to plan a Zenith AAA stent-graft from a CTA: Basic measurements and concepts explained

Popliteal pterygium syndrome

Using diffusion-tensor imaging and tractography (DTT) to study biological characteristics of glyoma in brain stem for neurosurgical planning

Assessment of renal cell carcinoma by two PET tracer : dual-time-point C-11 methionine and F-18 fluorodeoxyglucose

The Virtual Lung Nodule Clinic

Categorical Classification of Spiculated Mass on Breast MRI

Characterisation of cervical lymph nodes by US and PET-CT

Breast calcification: Management and Pictorial Review

Blinded, independent, central image review in oncology trials: how the study endpoint impacts the adjudication rate

Breast cancer tumor size: Correlation between MRI and histopathology

The "whirl sign". Diagnostic accuracy for intestinal volvulus.

Comparison of Image quality in temporal bone MRI at 3T using 2D selective RF excitation versus a routine SPACE sequence

Low-dose computed tomography (CT) protocol in the screening of patients with social exposure to asbestos

A pictorial review of normal anatomical appearences of Pericardial recesses on multislice Computed Tomography.

Transcription:

Dose reduction in Hologic Selenia FFDM units through AEC optimization, without compromising diagnostic image quality. Poster No.: C-1636 Congress: ECR 2015 Type: Authors: Keywords: DOI: Scientific Exhibit W. CONRADIE, A. Erasmus; Bloemfontein/ZA Breast, Mammography, Dosimetry, Technology assessment, Comparative studies, Dosimetric comparison, Quality assurance 10.1594/ecr2015/C-1636 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply ECR's endorsement, sponsorship or recommendation of the third party, information, product or service. ECR is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold ECR harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies, ppt slideshows and any other multimedia files are not available in the pdf version of presentations. www.myesr.org Page 1 of 13

Aims and objectives Hologic Selenia Full Field Digital Mammography (FFDM) is FDA approved for conventional 2D, 3D tomosynthesis and C-view 2D images. More than 75 digital units have been installed in South Africa since 2007 which include more than 40 units capable of performing 3D tomosynthesis. One of the many reasons for this vendor's popularity is its Automatic Exposure Control (AEC) function which has 3 different modes. In the popular AUTOFILTER mode the AEC automatically selects the filter and kv based on compressed breast thickness and the mas based on calculations after a low dose scout exposure to compensate for breast parenchymal density. 1 The amount of photons to produce an image is proportional to dose and affects the Signal-to-Noise ratio (SNR) and the Contrast-to-Noise ratio (CNR). These ratios in turn affect image quality, contributing to image's smoothness or grainy appearance. In digital mammography there should always be a balance between image dose and diagnostic quality. The purpose of the study was to determine if dose reduction in Hologic Selenia FFDM is achievable through AEC optimization. To compare the diagnostic quality of the current higher- to new lower dose images. Investigate if radiologists have a preference towards higher- or lower dose images in FFDM. Methods and materials A phantom study was done to evaluate AEC optimization using the European Guidelines. 2,3 Using an ionization chamber and 20-70 mm of Perspex, simulating 21-90 mm of compressed breast thickness, the monitor displayed mean glandular dose (MGD) values was confirmed to be the same as the measured values. The CDMAM phantom (Figure 1) was then imaged at different dose settings and simulated breast thicknesses to obtain a Contrast-Detail Curve for our institutions' Selenia. The selected settings were the current AEC setting, 25% more and 25%, 35% and 50% less. After the phantom study a before-and-after clinical trial was done on patients recruited after their routine annual mammography investigation at the institution. Patients were selected based on their compressed breast thickness and parenchymal density but excluded if they were younger than 40 years, had a previous mastectomy or presented with painful breast lesions. Page 2 of 13

After consent, each participant had 1 additional same day lower dose image, craniocaudal (CC) or medio-lateral oblique (MLO) projection, to each breast after AEC optimization. Optimization was achieved by adjusting the AEC compensation to -2, efectively reducing the dose per image by 22-33% depending on the individual breast characteristics. The corresponding paired dose images (routine study and new lower dose CC or MLO image) were then evaluated separately by two mammography radiologists who were blinded to the image dose. Images were firstly evaluated for: Benign or malignant masses Macro- or microcalcifications Architectural distorsion Axillary lympnodes Pectoral muscle detail Secondly, radiologists were asked to indicate their preferred image. Images for this section: Page 3 of 13

Fig. 1: The CDMAM phantom Page 4 of 13

Results The Phantom study proved that our institutions' Selenia complied with international image quality standards. The Contrast-Detail Curve for this Selenia (Figure 2), showed that the contrast achieved in the current AEC setting (black dotted line) was way below the European achievable line, indicating room for AEC optimization. Comparing the curves of the other dose settings, data confirmed that a theoretical dose reduction of 25-35% could be achieved (green triangles and red squares in curve). Twenty one patients were recruited for the clinical trial. All participants were female with an average age of 59 years (Range 40-76 years). For the purpose of the study, participants were divided into 3 groups based on their breast parenchymal density (using the TABAR classification) to ensure comparable numbers in terms of patient age, breast thickness and density (Figure 3-4). Apart from three patients with suspicious BI-RADS 4 and 5 breast lesions, all participants had benign mammography findings. Radiologists found that the malignant lesions were clearly visible in both set of images and that there was no significant difference in the diagnostic quality for the remainder of the breast characteristics (Figure 5). Out of 42 paired images there were a total of six concerns raised by the radiologists (Figure 6). Data analysis showed that radiologists did not have a clear preference to any set of dose images and this remained true when breast density and compressed breast thickness were considered (Figure 7-10). After AEC optimization an AGD reduction of 25% for all CC views and 26.5% for all MLO views was achieved. A breakdown of the dose reduction for each of the parenchymal density groups shows that the low density breast group had the most benefit followed by the denser breast groups (Figure 11). Images for this section: Page 5 of 13

Fig. 2: Contrast-Detail Curve. The European 'acceptable and achievable values' are indicated by the red and green straight lines. Ideally readings should be close to the 'achievable' line to indicate optimal AEC function. Page 6 of 13

Fig. 3: The graph illustrates equal distribution in terms of patient age (y-axis) and the number of patients (x-axis) in the different breast density groups. Fig. 4: Equal distribution of patient numbers (y-axis) in terms of compressed breast thickness (y-axis) for the three breast density groups. Page 7 of 13

Fig. 5: Summary of the radiologists' findings when comparing the higher- to the lower dose images. Fig. 6: Summary of the 6 concerns raised by the radiologists. Note that the same concern was not identified by both radiologists and that most concerns were related to benign breast characteristics and issues regarding breast re-positioning. In 50% of cases these concerns were seen in the original higher dose AEC setting Page 8 of 13

Fig. 7: Summary of the combined and individual radiologists' image preference data. Fig. 8: The graph illustrates that lower dose images were preferred in the lower density breast group and higher dose images more often preferred in the denser breast groups. The latter could be explained by less noise in higher dose images. Page 9 of 13

Fig. 9: Fig 9: In terms of breast thickness it was interesting to see that radiologists more often preferred higher dose images for thinner breasts and low dose images for thicker breasts, irrespective of breast density (Fig 10). This suggests that thicker breasts will not limit the use of lower dose images even in dense breasts. Page 10 of 13

Fig. 10 Page 11 of 13

Fig. 11 Page 12 of 13

Conclusion Our study achieved an average breast dose reduction of 25-26% in digital mammography through AEC optimization. There appear to be no significant difference in the diagnostic quality between the higher and lower dose images. In terms of the radiologists' preference towards image dose, data showed that both breast density and compressed breast thickness influence this descision. Although higher dose images were more often selected in denser breasts, there was a preference for lower dose images in thicker breasts, irrespective of breast density. Personal information Conradie, W.J. (MBChB) Resident/Registrar; Erasmus, A. (B.Med.Sc.Hons) Medical Physicist; Otto S.F. (MMed Rad) Specialist Radiologist. Department of Clinical Imaging Science, Universitas academic Hospital. UFS, Bloemfontein. South Africa References 1. www.hologic.com 2. Young et al. Technical evaluation of the Hologic Selenia full field digital mammography system. NHSBSP Equipment Report 0701. NHSBPS January 2007. 3. Young K.C, Oduko J.M. Technical evaluation of the Hologic Selenia full field digital mammography system with a Tungsten tube. NHSBSP Equipment Report 0701. NHSBPS April 2008. Page 13 of 13