Size Exclusion Chromatography YMC-Pack Diol

Similar documents
YMC Phases for Biochromatography IEX SEC RP NP/HILIC

- intermediate polarity between normal phase silica and other alkyl bonded reversed phases - operates in either normal phase or reversed phase mode

YMC HILIC Columns for Polar Analytes

Bioseparation. From Microanalysis To Plant-Scale Purification PB-0065E

YMC-Actus. Columns for Semi-preparative HPLC. Selectivity Durability Loadability

YMC-Triart C18. Multilayer hybrid material for HPLC ph and temperature stability. Innovative micro-reactor technology

Purification of Insulin with YMC-Triart Prep C8. Increased Output at Reduced Costs

Meteoric Core. Core-Shell Columns for UHPLC & HPLC. Easy Method Optimization! Excellent Chromatography with Fast Separation!

YMC-Triart C18. Multi-layer hybrid material for HPLC ph and temperature stability. Innovative micro-reactor technology

YMC Chiral columns. Polysaccharides NEA(R)(S) CD BR

Ion Exchange Chromatography COSMOGEL IEX Series Columns

YMC*Gel HG-series. High grade silica phases for preparative HPLC. The Selectivity Company

YMC CHIRAL columns. Polysaccharides NEA(R)(S) CD BR

TSKgel Hydrophobic Interaction Chromatography Columns

Agilent Anion-Exchange Media for Proteins - Loading vs Resolution - Effect of Flow Rate and Example Protein Separations

YMC-Triart C18. Multilayer hybrid material for HPLC ph and temperature stability. Innovative micro-reactor technology

InertSustainBio C18. Rapid Separations of Proteins and Peptides

The high efficiency of sub-2 µm UHPLC columns combined with the low pressure and high speed of monolith columns.

176 YMC Chiral Columns

3 Normal-Phase Columns. Reversed-Phase C18 Columns (ODS) Reversed-Phase Columns (Other than ODS) Columns for HTS and LC/MS Analysis.

SeQuant ZIC -HILIC For all who expect more...

SeQuant ZIC -HILIC For all who expect more...

COSMOSIL HPLC Columns

A New HILIC/RP Mixed-Mode Column and Its Applications in Surfactant Analysis

Core-Shell Technology for Proteins and Peptides

Hydrophobic Interaction Chromatography

Analytical and Preparative SFC Columns

LC Columns - Exceed the limit. A premium inert range of LC columns delivering optimal peak shape. ProteCol -P PEEK lined

Tosoh Bioscience llc Distributed by Phenomenex in the USA, Canada, and Puerto Rico only.

Normal Phase Chemistries

Physical Properties. Silica

HC-75 Calcium Form. 305 x 7.8 mm HC-75 Calcium Form (P/N 79436)

Normal Phase Chemistries

DELFIA Tb-DTPA ITC Chelate & Terbium Standard

Separation of Vitamin D and Vitamin D Metabolites on FLARE C18 MM (Mixed Mode) HPLC Column

Eurospher II the logical choice. Physical properties of Eurospher II silica gel: Silica gel: ultra pure, > % Metal content: < 10 ppm

Agilent Technologies Prep LC Columns

InertSustain Phenylhexyl. A New Second Choice Column to Change Chromatographic Behavior

HICHROM. Chromatography Columns and Supplies NEW PRODUCTS. Catalogue 9. Hichrom Limited

Superose 6 Increase columns

Catalogue. Resins and Columns For High Performance Liquid Chromatography

DELFIA Tb-N1 DTA Chelate & Terbium Standard

Normal Phase HPLC Column COSMOSIL SL-II

DELFIA Eu-DTPA ITC Chelate & Europium Standard

ACE. For increased polar retention and alternative selectivity. Alternative selectivity for method development

Separation of Saccharides Using TSKgel Amide-80, a Packing Material for High-Performance Normal Phase Partition Chromatography (1) Table of Contents

size exclusion

Sepax Technologies, Inc.

Overview on YMC Column Hardware and Phases

COSMOSIL HILIC. HPLC Column for Hydrophilic Interaction

Ultimate Core Performance to Maximize Your Investment

HPLC Column for Structural Isomers COSMOSIL Cholester

TSKgel Size Exclusion Chromatography Columns

Separation of Macrocyclic Lactones (Avermectins) on FLARE C18 MM & FLARE C18+ Columns

Reversed-Phase Columns (Other than ODS)

VertiSep TM PRP. HPLC Columns. VertiSep TM PRP HPLC Columns

Characterization of Disulfide Linkages in Proteins by 193 nm Ultraviolet Photodissociation (UVPD) Mass Spectrometry. Supporting Information

Choosing Preparative Columns

Overview on YMC Column Hardware and Phases

Data File. Sephadex ion exchange media. Ion exchange chromatography. Introduction. Sephadex ion exchangers General description

11:00 a.m. EST Telephone Number: Chair Person: Tim Spaeder

14 Separation of the Phosphatidylcholines Using Reverse Phase HPLC

Aeris. Precision Engineered Core- Shell Particles for Ultra-High Resolution BioSeparations. Aeris PEPTIDE. Aeris WIDEPORE

Converting a CHP Method for Insulin to Agilent Poroshell 120 Columns

Charged Surface Hybrid C18 for High Resolution LC and LC/MS Peptide Separations

Fused-Core Particles:

Hypersil BDS and Hypersil Classical HPLC Columns

Separation of Saccharides Using TSKgel Amide-80, a Packing Material for High Performance Normal Phase Partition Chromatography (2)*

Application Note. Separation of clindamycin phosphate and process impurities. Summary. Introduction. Category Pharmaceutical analysis Matrix

APPLICATIONS Improving Intact Biogeneric Protein Separations with Aeris WIDEPORE Core-Shell Columns

Confident, lower-pressure analysis of carbohydrates, alcohols, and organic acids. Agilent Hi-Plex Ligand-Exchange HPLC Columns

C30 ISOMERS HAVE MET THEIR MATCH

LANCE Eu-W1024 ITC Chelate & Europium Standard AD0013 Development grade

Restek Ultra II HPLC Columns

Unequalled durability against water elution. % tr

Chromatogram Search Tool Search by compound name, synonym, CAS # or keyword Pinnacle II Columns

High-Throughput Separations for QA/QC, LC/MS, LC/MS/MS, Combinatorial Chemistry and Rapid Method Development

Sepax Technologies, Inc.

Waters Amide Column Technology For Food Analysis

CE Columns and Accessories

Insulin SEC analysis. Insulin Samples:

Hyper-fast & Super-rugged

GlycanPac AXR-1 Columns

The Raptor HILIC-Si Column

HPLC Analysis of Sugars

Protein and peptide separations,

Phenyl-Hexyl. UHPLC Columns. lternate, complementary selectivity to C18 and C8 bonded phases

COSMOSIL CHiRAL Series

LC Columns with Liquid Separation Cell Technology

Studies on Stationary Phase Selectivity for Solid-Core Particles

ZIDOVUDINE, LAMIVUDINE AND ABACAVIR TABLETS Draft proposal for The International Pharmacopoeia (September 2006)

Sephadex G-25 media and prepacked formats

Application Note. Sensitive analysis of the lactose content of lactose-free labelled products using HPAEC-PAD. Summary.

Protein MultiColor Stable, Low Range

Using Hydrophilic Interaction Chromatography (HILIC) for the Retention of Highly Polar Analytes

SUPPLEMENTARY INFORMATION

Effective use of Pharmacopeia guidelines to reduce cost of chromatographic analysis for Fluticasone propionate

Transcription:

Size Exclusion Chromatography YMC-Pack Diol www.ymc.de

What is special about YMC SEC-Columns? Method development Scalability Reproducibility Cost-effective YMC-Column for SEC: YMC-Pack Diol Analysis of many polymers has an additional complication in that apart from determination of functional groups, etc, such compounds do not vary in chemical composition but their functionality varies as a function of molecular size. Therefore Size Exclusion Chromatography (SEC) can be employed to further characterise polymeric materials. SEC is divided into Gel Permeation Chromatography (GPC) which is performed using organic mobile phases and Gel Filtration Chromatography (GFC) which is carried out in aqueous eluent systems. The silica is derivatised with 1,2-dihydroxypropane and, with the variety of pore size available, allows analyses over a broad range of peptide to large proteins Why modify the phase? YMC-Pack Diol stationary phases are surface modified for a versatile alternative to underivatised silica. Advantages include a shorter equilibration time as well as good selectivity. Due to this short equilibration time, the Diol phase is a good choice for SEC in comparison with underivatised silica. unmodified phase = long equilibration time modified phase = short equilibration time YMC Europe GmbH Tel.: 02064/427-0 Fax: 02064/427-222 Email: info@ymc.de Website: www.ymc.de

Principles of separation Molecules with shapes such as rigid rods, random chains and spheres but with the same molecular weight can behave differently. The principle of separation is based on different the hydrodynamic radius of the molecules in solution. Molecules with a larger radius elute earlier and those with the smallest radius are retained longer. The separation limit can stated such that only those compounds which differ more than 10% in MW can be separated. 1. Injection Intensity Injection 2. Separation Elution Volume V e small molecules = long retention time bigger molecules = short retention time 3. Elution of Fraction Large molecules exit the column more rapidly as they cannot permeate the porous structure of stationary phase. Smaller ones with the lowest hydrodynamic volume elute with longer retention times because they are able to penetrate some or all of the pores of the stationary phase. Molecules of intermediate size elute in an intermediate position. * Application data by courtesy YMC Co., Ltd.

High flexibility Proteins and Peptides* 10 6 1 1. Glutamate dehydrogenase 2. Lactate dehydrogenase 3. Enolase 4. Adenylate kinase 5. Cytochrome c MW 10 5 2 3 Diol-200 Diol-300 Diol-120 4 10 4 5 10 15 20 Column: Eluent: Flow rate: Temperature: Detection: YMC-Pack Diol, 500 x 8.0 mm ID 0.1M KH 2 PO 4 -K 2 HPO 4 (ph 7.0) containing 0.2M NaCl 0.7 ml/min ambient UV at 280 nm Elution Volume. ml YMC-Pack Diol is available in four porosities and is, therefore, suitable for separation and molecular weight determination of a wide range of peptides, proteins, oligonucleotides, carbohydrates and other biopolymers with molecular weights of 10,000 to several hundred thousand. Furthermore, YMC offers a wide range of column dimensions suitable for analytical determinations to preparative separations. Scalability min min YMC guarantees a seamless, reproducible scale up through all particle sizes for all stationary phases min YMC Europe GmbH Tel.: 02064/427-0 Fax: 02064/427-222 Email: info@ymc.de Website: www.ymc.de

Reproducibility YMC-Pack Diol columns are distinguished, not only by their high flexibility and good cost/performance ratio, but also by their high degree of lot-to-lot reproducibility. YMC-Pack Diol-120 Average pore diameter Bonding density Average pore diameter Bonding density Lot Number Lot Number YMC-Pack Diol-200 Average pore diameter Bonding density Average pore diameter Bonding density Lot Number Lot Number YMC-Pack Diol-300 Average pore diameter Bonding density Average pore diameter Bonding density Lot Number Lot Number YMC-Pack Diol Columns YMC-Pack Diol for small proteins for intermediate proteins for large proteins for very large proteins pore size / nm 6 12 20 30 particle size / μm 5 5 5 5 surface area / m 2 g -1 650 330 175 100 recommended ph range 5.0-7.5 5.0-7.5 5.0-7.5 5.0-7.5 * Application data by courtesy YMC Co., Ltd.

YMC applications for SEC For separation of proteins with molecular weights from 10.000 to several 100.000 MW 10 6 10 5 10 4 1 2 3 4 Diol-120 6 5 Diol-200 7 8 9 10 Diol-300 11 12 13 14 15 16 1. IgM 2. Thyroglobulin 3. IgA 4. Fibrinogen 5. g-globulin 6. IgG 7. Transferrin 8. HSA (human serum albumin) 9. a1-antitrypsin 10. Ovalbumin 11. Carbonic anhydrase 12. Trypsin inhibitor 13. Myoglobin 14. a-lactalbumin 15. Ribonuclease A 16. Cytochrome c MW 900,000 670,000 390,000 340,000 158,000 150,000 75,000 66,000 50,000 45,000 30,000 20,100 17,000 14,100 13,700 12,400 large pore size = large MW s 5 7 9 11 13 Elution volume (ml) Column: YMC-Pack Diol, 300 x 8.0 mm ID Eluent: 0.1 M KH2PO4-K2HPO4 (ph 7.0) containing 0.2 M NaCl Flow rate: 0.5 ml/min Temperature: 25 C Detection: UV at 280 nm Column Selection Tool YMC-Pack Diol-60 for MW < 10.000 YMC-Pack Diol-120 for MW 5.000 to 100.000 YMC-Pack Diol-200 for MW 10.000 to several 100.000 YMC-Pack Diol-300 for MW several 10.000 to 1.000.000 For separation of proteins with molecular weights from 10.000 or less 1. Insulin (bovine) 2. Neurotensin 3. Angiotensin-II 4. Glycine MW 10 4 10 3 10 2 1 2 4 5 3 7 9 12 14 15 6 8 11 16 13 17 10 1. Myoglobin (MW: 17,000) 2. Ribonuclease A (MW: 13,700) 3. Cytochrome c (MW: 12,400) 4. Insulin (bovine) (MW: 5,700) 5. Insulin B chain (MW: 3,496) 6. α-mating factor (MW: 1,684) 7. Neurotensin (MW: 1,672) 8. Angiotensin-I (MW: 1,296) 9. CCK-octapeptide (MW: 1,143) 10. Bradykinin (MW: 1,060) 11. Angiotensin-II (MW: 1,040) 12. Oxytocin (MW: 1,007) 13. Angiotensin-III (MW: 931) 14. Met-enkephalin (MW: 573) 15. Leu-enkephalin (MW: 555) 16. Tetraglycine (MW: 246) 17. Glycine (MW: 75) min Column: YMC-Pack Diol (6 nm) 500 x 8.0 mm ID Ordering-No.: DL06S055008WT Eluent: 0.1 M KH 2 PO 4 -K 2 HPO 4 (ph 7) containing 0.2 M NaCl / acetonitrile (70/30) Temperature: ambient Detection: UV at 215 nm 10 15 Elution Volume. ml 20 small pore size = small MW s YMC Europe GmbH Tel.: 02064/427-0 Fax: 02064/427-222 Email: info@ymc.de Website: www.ymc.de

YMC applications for SEC For molecular weight determination of carbohydrates 20 nm 1. P-800 (MW: 853,000) 2. P-50 (MW: 48,000) 3. P-20 (MW: 23,700) 4. P-5 (MW: 5,800) Column: YMC-Pack Diol (20 nm) 500 x 8.0 mm ID Ordering-No.: DL20S055008WT Eluent: water Flow rate: 1.0 ml/min Temperature: ambient Detection: RI min MW 10 6 10 5 10 4 10 3 10 2 1 2 3 4 5 6 7 8 Diol-60 Diol-300 Diol-200 9 10 Diol-120 11 12 13 14 15 10 15 20 25 Elution Volume. ml 1. Pullulan (P-800) (MW: 853,000) 2. Pullulan (P-400) (MW: 380,000) 3. Pullulan (P-200) (MW: 186,000) 4. Pullulan (P-100) (MW: 100,000) 5. Pullulan (P-50) (MW: 48,000) 6. Pullulan (P-20) (MW: 23,700) 7. Pullulan (P-10) (MW: 12,200) 8. Pullulan (P-5) (MW: 5,800) 9. Maltopentadecaose (G 15 ) (MW: 2,448) 10. Maltoundecaose (G 11 ) (MW: 1,800) 11. Maltoheptaose (G 7 ) (MW: 1,152) 12. Maltopentaose (G 5 ) (MW: 852) 13. Maltotriose (G 3 ) (MW: 504) 14. Maltose (G 2 ) (MW: 342) 15. Glucose (G 1 ) (MW: 180) sugar analysis Plasma constituents blood analysis 0.2 AUFS 0.07 AUFS 0.03 AUFS 0.04 AUFS 0.03 AUFS 0.03 AUFS 0.07 AUFS 0.03 AUFS 0.03 AUFS 0.1 AUFS 0.2 AUFS 0.18 AUFS 0.005 AUFS (N080616A) (N080616E) (N080616M) (N080617B) (N080616F) (N080616B) (N080616C) (N080616D) (N080616K) (N080616G) (N080616H) (N080617A) (N080616I) A) Human serum B) IgM (MW 900,000) C) IgA (MW 390,000) D) Fibrinogen (MW 340,000) F) IgG (MW 150,000) G) Transferrin (MW 75,000) H) Albumin (MW 66,000) J) Carbonic anhydrase (MW 30,000) K) Lysozyme (MW 14,300) L) Uric acid (MW 168) M) Creatinine (MW 113) 0 10 20 30 40 50 60 min Column: YMC-Pack Diol-300 + Diol-200, 300 x 8.0 mm ID x 2 Ordering-No.: DL30S053008WT + DL20S053008WT Eluent: 0.1 M KH 2 PO 4 K 2 HPO 4 (ph 7.0) containing 0.2 M NaCl Flow rate: 0.5 ml/min Temperature: Detection: Injection: Sample: ambient (25 C) UV at 280 nm 20 µl (L: 1 µl) A) 100 µl/ml; B-M) 1.0 mg/ml * Application data by courtesy YMC Co., Ltd.

YMC applications for SEC Plasmid pbr322 restriction fragment 1. Plasmid pbr322 cleaved with restriction endonuclease Msp I Column: YMC-Pack Diol-300 + Diol-200, 500 x 8.0 mm ID x 2 Ordering-No.: DL30S055008WT + DL20S055008WT Eluent: 0.1M KH 2 PO 4 -K 2 HPO 4 (ph 7.0) containing 0.2M NaCl Temperature: ambient (26 C) Detection: UV at 260 nm, 0.01 AUFS Injection: 3 µl (0.49 mg/ml) Sample: Plasmid pbr322 cleaved with restriction endonuclease Msp I Plasmid pbr322 restriction and pbr322 Hae III restriction fragment 1. Plasmid pbr322 (4,361 bp) 2. Plasmiid pbr322 Hae III digest (8-587 bp) 1. Plasmiid pbr322 Hae III digest (8-587 bp) 2. Plasmid pbr322 (4,361 bp) SEC vs. IC! Column: YMC-Pack Diol-300 + Diol-200, 500 x 8.0 mm ID x 2 Ordering-No.: DL30S055008WT + DL20S055008WT Eluent: 0.1 M KH 2 PO 4 -K 2 HPO 4 (ph 7.0) containing 0.2 M NaCl Detection: UV at 260 nm Injection: 10 µl Column: YMC-BioPro QA-F (5 µm), 100 x 4.6 mm ID Ordering-No.: QF00S051046WT Eluent: A) 20 mm Tris-HCl (ph 8.1) B) 20 mm Tris-HCl (ph 8.1) containing 0.1 M NaCl 70-85% B (0-20 min), 85% B (20-25 min) Flow rate: 0.5 ml/min Temperature: 35 C Detection: UV at 260 nm Injection: 10 µl YMC Europe GmbH Tel.: 02064/427-0 Fax: 02064/427-222 Email: info@ymc.de Website: www.ymc.de

YMC applications for SEC Peptide fragments from myoglobin Peptide mapping 1. 1-153 (MW 17,000) 2. 1-131 (MW 14,500) 3. 56-153 (MW 10,700) 4. 56-131 (MW 8,200) 5. 1-55 (MW 6,300) 6. 132-153 (MW 2,500) Column: YMC-Pack Diol-120 + Diol-60, 500 x 8.0 mm ID x 2 Ordering-No.: DL12S055008WT + DL06S055008WT Eluent: 0.1 M KH 2 PO 4 -K 2 HPO 4 (ph 7.0) containing 0.2 M NaCl/ acetonitrile (70/30) Detection: UV at 220 nm Injection: 5 µl Sample: Tryptic digest of BSA Proteins in human serum Column: YMC-Pack Diol-120, 500 x 8.0 mm ID Ordering-No.: DL12S055008WT Eluent: 0.1 M KH 2 PO 4 -K 2 HPO 4 (ph 7.0) containing 0.2 M NaCl/ acetonitrile (70/30) Detection: UV at 215 nm, 0.32 AUFS Injection: 20 µl (2.0 mg/ml) Sample: Cyanogen bromide cleavages of horse heart myoglobim. Molecular Weight Marker for proteins, manufactured by Fluka Chemie AG. YMC-Pack Diol-300 + Diol-200 300 x 8.0 mm ID x 2 YMC-Pack Diol-300 300 x 8.0 mm ID Proteins in mouse ascites fluid YMC-Pack Diol-200 300 x 8.0 mm ID Column: YMC-Pack Diol-300 + Diol-200, 300 x 4.6 mm ID x 2 Ordering-No.: DL30S053046WT + DL20S053046WT Eluent: 0.1 M KH 2 PO 4 -K 2 HPO 4 (ph 7.0) Flow rate: 0.17 ml/min Detection: UV at 220 nm Injection: 10 µl (60 times dilution with water) Eluent: 0.1 M KH 2 PO 4 -K 2 HPO 4 (ph 7.0) containing 0.2 M NaCl Flow rate: 0.5 ml/min Detection: UV at 280 nm Injection: 20 µl Sample: Human serum (100 µl/ml) * Application data by courtesy YMC Co., Ltd.

YMC applications for SEC lgg, Fab and Fc fragments Human Immunglobulin 1. Rabbit lgg 2. Mouse lgg FC fragment 3. Mouse lgg Fab fragment 1. Human IgM (k), myeloma protein 2. Normal human IgG Column: YMC-Pack Diol-200, 300 x 8.0 mm ID Ordering-No.: DL20S053008WT Eluent: 0.1 M KH 2 PO 4 -K 2 HPO 4 (ph 6.9) containing 0.2 M NaCl Flow rate: 1.0 ml/min Temperature: ambient (27 C) Detection: UV at 220 nm Injection: 5 µl (0.4, 0.5 mg/ml) Column: YMC-Pack Diol-300, 500 x 8.0 mm ID Ordering-No: DL30S055008WT Eluent: 0.1M NaH 2 PO 4 -Na 2 HPO 4 (ph 6.8) containing 0.1M Na 2 SO 4 Flow rate: 1.0 ml/min Temperature: ambient (24 C) Detection: UV at 280 nm, 0.04 AUFS Injection: 40 µl (0.5 mg/ml) Human lgg (λ), myeloma protein Monoclonal antibody (MAb) 1. Human lgg (λ), myeloma protein MAb: Mouse monoclonal lgg1 (anti-human lgg4) Column: YMC-Pack Diol-300 + Diol-200, 500 x 8.0 mm ID x 2 Ordering-No.: DL30S055008WT + DL20S055008WT Eluent: 0.1 M KH 2 PO 4 -K 2 HPO 4 (ph 7.0) containing 0.2 M NaCl Detection: UV at 260 nm Injection: 10 µl Column: YMC-Pack Diol-200, 300 x 4.6 mm ID Ordering-No.: DL20S053046WT Eluent: 0.1 M KH 2 PO 4 -K 2 HPO 4 (ph 7.0) Flow rate: 0.17 ml/min Detection: UV at 220 nm Injection: 10 µl Sample: a commercially available mouse monoclonal lgg1 (0.05 mg/ml) (purified by DEAE chromatography, containing NaN 3 ) YMC Europe GmbH Tel.: 02064/427-0 Fax: 02064/427-222 Email: info@ymc.de Website: www.ymc.de

YMC applications for SEC Proteins for molecular weight marker 1. Glutamate dehydrogenase (MW 290,000) 2. Lactate dehydrogenase (MW 142,000) 3. Enolase (MW 67,000) 4. Adenylate kinase (MW 32,000) 5. Cytochrome c (MW 12,400) Proteins for molecular weight marker 1. Glutamate dehydrogenase (MW 290,000) 2. Lactate dehydrogenase (MW 142,000) 3. Enolase (MW 67,000) 4. Adenylate kinase (MW 32,000) 5. Cytochrome c (MW 12,400) Column: YMC-Pack Diol-300, 500 x 8.0 mm ID Ordering-No.: DL30S055008WT Eluent: 0.1 M KH 2 PO 4 -K 2 HPO 4 (ph 7.0) containing 0.2 M NaCl Temperature: ambient (26 C) Detection: UV at 280 nm, 0.08 AUFS Injection: 15 µl (100 µl / 1 vial) Sample: MW-Marker (HPLC), manufactured by ORIENTAL YEAST CO., LTD. Column: YMC-Pack Diol-200, 500 x 8.0 mm ID Ordering-No.: DL20S055008WT Eluent: 0.1 M KH 2 PO 4 -K 2 HPO 4 (ph 7.0) containing 0.2 M NaCl Temperature: ambient (26 C) Detection: UV at 280 nm, 0.08 AUFS Injection: 15 µl (100 µl / 1 vial) Sample: MW-Marker (HPLC), manufactured by ORIENTAL YEAST CO., LTD. Proteins for molecular weight marker 1. Glutamate dehydrogenase (MW 290,000) 2. Lactate dehydrogenase (MW 142,000) 3. Enolase (MW 67,000) 4. Adenylate kinase (MW 32,000) 5. Cytochrome c (MW 12,400) Proteins for molecular weight marker 1. Glutamate dehydrogenase (MW 290,000) 2. Lactate dehydrogenase (MW 142,000) 3. Enolase (MW 67,000) 4. Adenylate kinase (MW 32,000) 5. Cytochrome c (MW 12,400) Column: YMC-Pack Diol-200, 300 x 8.0 mm ID Ordering-No.: DL20S053008WT Eluent: 0.1 M KH 2 PO 4 -K 2 HPO 4 (ph 7.0) containing 0.2 M NaCl Temperature: ambient (26 C) Detection: UV at 280 nm, 0.08 AUFS Injection: 15 µl (100 µl / 1 vial) Sample: MW-Marker (HPLC), manufactured by ORIENTAL YEAST CO., LTD. Column: YMC-Pack Diol-120, 500 x 8.0 mm ID Ordering-No.: DL12S055008WT Eluent: 0.1 M KH 2 PO 4 -K 2 HPO 4 (ph 7.0) containing 0.2 M NaCl Temperature: ambient (26 C) Detection: UV at 280 nm, 0.08 AUFS Injection: 15 µl (100 µl / 1 vial) Sample: MW-Marker (HPLC), manufactured by ORIENTAL YEAST CO., LTD. * Application data by courtesy YMC Co., Ltd.

YMC Applications for SEC Proteins for molecular weight marker 1. Glutamat dehydrogenase (MW 290.000) 2. Lactat dehydrogenase (MW 142.000) 3. Enolase (MW 67.000) 4. Adenylat kinase (MW 32.000) 5. Cytochrome c (MW 12.400) good selectivity Column: YMC-Pack Diol-200 + Diol-200, 500 x 8.0 mm ID + 300 x 8.0 mm ID Ordering-No.: DL20S055008WT + DL20S053008WT Eluent: 0.1M KH 2 PO 4 -K 2 HPO 4 (ph 7.0) containing 0.2M NaCl Temperature: ambient (26 C) Detection: UV at 280 nm, 0.08 AUFS Injection: 15 µl (100 µl/1 vial) Sample: MW-Marker (HPLC) manufactured by ORIENTAL YEAST CO., LTD. Proteins for molecular weight marker 1. Glutamate dehydrogenase (MW 290.000) 2. Lactate dehydrogenase (MW 142.000) 3. Enolase (MW 67.000) 4. Adenylate kinase (MW 32.000) 5. Cytochrome c (MW 12.400) Column: YMC-Pack Diol-120 + Diol-300, 500 x 8.0 mm ID x 2 Ordering-No.: DL12S055008WT + DL30S055008WT Eluent: 0.1M KH 2 PO 4 -K 2 HPO 4 (ph 7.0) containing 0.2M NaCl Temperature: ambient (26 C) Detection: UV at 280 nm, 0.08 AUFS Injection: 15 µl (100 µl/1 vial) Sample: MW-Marker (HPLC) manufactured by ORIENTAL YEAST CO., LTD. YMC Europe GmbH Tel.: 02064/427-0 Fax: 02064/427-222 Email: info@ymc.de Website: www.ymc.de

YMC applications for SEC Heparin Heparin sulfat 1. Heparin, Na salt 1. Heparan sulfat, Na salt Column: YMC-Pack Diol-300, 500 x 8.0 mm ID Ordering-No.: DL30S055008WT Eluent: water Flow rate: 1.0 ml/min Temperature: ambient (24 C) Detection: RI, 32 x 10-6 RIU/FS Injection: 20 µl (5.0 mg/ml) Column: YMC-Pack Diol-300, 500 x 8.0 mm ID Ordering-No.: DL30S055008WT Eluent: water Flow rate: 1.0 ml/min Temperature: ambient (24 C) Detection: RI, 32 x 10-6 RIU/FS Injection: 20 µl (5.0 mg/ml) Keratan polysulfat Cyclodextrins 1. Keratan polysulfat, Na salt 1. g-cyclodextrin 2. b-cyclodextrin 3. a-cyclodextrin Column: YMC-Pack Diol-300, 500 x 8.0 mm ID Ordering-No.: DL30S055008WT Eluent: water Flow rate: 1.0 ml/min Temperature: ambient (24 C) Detection: RI 32 x 10-6 RIU/FS Injection: 20 µl (5.0 mg/ml) Column: YMC-Pack Diol 60, 500 x 8.0 mm ID Ordering-No.: DL06S055008WT Eluent: water Flow rate: 1.0 ml/min Temperature: ambient (24 C) Detection: RI 32 x 10-6 RIU/FS Injection: 30 µl (1.67 mg/ml) * Application data by courtesy YMC Co., Ltd.

YMC applications for SEC Glycine oligomers Peptides 1. Pentaglycine 2. Triglycine 3. Glycine 1. Insulin (Bovine) 2. Neurotensin 3. Angiotensin II 4. Glycine Column: YMC-Pack Diol-60, 500 x 8.0 mm ID Ordering-No.: DL06S055008WT Eluent: 0.1M KH 2 PO 4 -K 2 HPO 4 (ph 7.0) / acetonitrile (70/30) Flow rate: 1.0 ml/min Temperature: ambient (24 C) Detection: UV at 215 nm, 0.08 AUFS Injection: 20 µl (0.25 ~ 2.5 mg/ml) Column: YMC-Pack Diol-60, 500 x 8.0 mm ID Ordering-No.: DL06S055008WT Eluent: 0.1M KH 2 PO 4 -K 2 HPO 4 (ph 7.0) / containing 0.2 M NaCl / acetonitrile (70/30) Detection: UV at 215 nm, 0.16 AUFS Injection: 25 µl (0.07 ~ 5.3 mg/ml) Peptides Peptides 1. Insulin (Bovine) 2. Neurotensin 3. Angiotensin II 4. Glycine 1. Insulin (Bovine) 2. Neurotensin 3. Angiotensin II 4. Glycine Column: YMC-Pack Diol-120, 500 x 8.0 mm ID Ordering-No.: DL12S055008WT Eluent: 0.1M KH 2 PO 4 -K 2 HPO 4 (ph 7.0) / containing 0.2 M NaCl / acetonitrile (70/30) Detection: UV at 215 nm, 0.16 AUFS Injection: 25 µl (0.07 ~ 5.3 mg/ml) Column: YMC-Pack Diol-120 + 60, 500 x 8.0 mm ID x 2 Ordering-No.: DL12S055008WT + DL06S055008WT Eluent: 0.1M KH 2 PO 4 -K 2 HPO 4 (ph 7.0) / containing 0.2 M NaCl / acetonitrile (70/30) Detection: UV at 215 nm, 0.16 AUFS Injection: 25 µl (0.07 ~ 5.3 mg/ml) YMC Europe GmbH Tel.: 02064/427-0 Fax: 02064/427-222 Email: info@ymc.de Website: www.ymc.de

Ordering Information YMC-Pack Diol, 6 nm, 5 μm Column Column length (mm) i.d. (mm) 250 300 500 4.6 DL06S052546WT DL06S053046WT n.a. 6.0 DL06S052506WT DL06S053006WT DL06S055006WT 8.0 n.a. DL06S053008WT DL06S055008WT 10.0 DL06S052510WT DL06S053010WT DL06S055010WT YMC-Pack Diol, 12 nm, 5 μm Column Column length (mm) i.d. (mm) 250 300 500 4.6 DL12S052546WT DL12S053046WT n.a. 6.0 DL12S052506WT DL12S053006WT DL12S055006WT 8.0 n.a. DL12S053008WT DL12S055008WT 10.0 DL12S052510WT DL12S053010WT DL12S055010WT YMC-Pack Diol, 20 nm, 5 μm Column Column length (mm) i.d. (mm) 250 300 500 4.6 DL20S052546WT DL20S053046WT n.a. 6.0 DL20S052506WT DL20S053006WT DL20S055006WT 8.0 n.a. DL20S053008WT DL20S055008WT 10.0 DL20S052510WT DL20S053010WT DL20S055010WT YMC-Pack Diol, 30 nm, 5 μm Column Column length (mm) i.d. (mm) 250 300 500 4.6 DL30S052546WT DL30S053046WT n.a. 6.0 DL30S052506WT DL30S053006WT DL30S055006WT 8.0 n.a. DL30S053008WT DL30S055008WT 10.0 DL30S052510WT DL30S053010WT DL30S055010WT Guard Columns are available in different dimensions. For more details please contact us: Phone 02064-427-0 or email info@ymc.de.

YMC SEC columns provide: wide range of applications: Proteins, peptides, carbohydrates and nucleic acid components good cost/performance ratio scalability: from 5 µm to 75 µm minimal secondary interactions Your local distributor: Europe GmbH Schöttmannshof 19 D-46539 Dinslaken Germany TEL. +49(0)2064/427-0, FAX +49(0)2064/427-222 www.ymc.de Co., Ltd. YMC Karasuma-Gojo Bld. 284 Daigo-cho, Karasuma Nisiiru Gojo-dori Shimogyo-ku, Kyoto 600-8106 Japan TEL. +81(0)75-342-4515, FAX +81(0)75-342-4550 www.ymc.co.jp America, Inc. 941 Marcon Boulevard Suite 301 Allentown, PA18109 USA TEL. +1-610-266-8650, FAX +1-610-266-8652 www.ymcamerica.com India Ltd. CX - 07, 3rd Floor, Lobe - 1, Tower - A, The Corenthum, Plot No- A-41, Sector - 62, Noida - 201301 (UP) India. TEL. +91(0)120-4276020 - 25, FAX +91(0)120-4276026 www.ymcindia.com