Extraction of the magnetohydrodynamic blood flow potential from the surface electrocardiogram in magnetic resonance imaging

Similar documents
MR Advance Techniques. Vascular Imaging. Class II

Magnetic Resonance Angiography

MR Advance Techniques. Cardiac Imaging. Class IV

NIH Public Access Author Manuscript Int J Cardiovasc Imaging. Author manuscript; available in PMC 2013 August 01.

Electrocardiography I Laboratory

ECG Electrode Placements for Magnetohydrodynamic Voltage Suppression

Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter Detection

HST-582J/6.555J/16.456J-Biomedical Signal and Image Processing-Spring Laboratory Project 1 The Electrocardiogram

NIH Public Access Author Manuscript Ann Biomed Eng. Author manuscript; available in PMC 2015 December 01.

Removing ECG Artifact from the Surface EMG Signal Using Adaptive Subtraction Technique

ECG. Prepared by: Dr.Fatima Daoud Reference: Guyton and Hall Textbook of Medical Physiology,12 th edition Chapters: 11,12,13

Objectives 8/17/2011. Challenges in Cardiac Imaging. Challenges in Cardiac Imaging. Basic Cardiac MRI Sequences

BASIC CONCEPT OF ECG

Performance of QRS Detection for Cardiac Magnetic Resonance Imaging With a Novel Vectorcardiographic Triggering Method

Non Contrast MRA. Mayil Krishnam. Director, Cardiovascular and Thoracic Imaging University of California, Irvine

CARDIAC MRI. Cardiovascular Disease. Cardiovascular Disease. Cardiovascular Disease. Overview

UK Biobank. Imaging modality Cardiovascular Magnetic Resonance (CMR) Version th Oct 2015

CASE 10. What would the ST segment of this ECG look like? On which leads would you see this ST segment change? What does the T wave represent?

PCA Enhanced Kalman Filter for ECG Denoising

Outline. Electrical Activity of the Human Heart. What is the Heart? The Heart as a Pump. Anatomy of the Heart. The Hard Work

NIH Public Access Author Manuscript Magn Reson Med. Author manuscript; available in PMC 2015 April 01.

Aortic Vessel Wall Imaging Using 3D Phase Sensitive Inversion Recovery in Children and Young Adults

Biomedical Signal Processing

Testing the Accuracy of ECG Captured by Cronovo through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device

Signal Processing of Stress Test ECG Using MATLAB

General Cardiovascular Magnetic Resonance Imaging

PROBLEM SET 3. Assigned: February 19, 2004 Due: February 26, 2004

Fully Refocused Gradient Recalled Echo (FRGRE): Factors Affecting Flow and Motion Sensitivity in Cardiac MRI

MR Advance Techniques. Cardiac Imaging. Class III

ECG Beat Recognition using Principal Components Analysis and Artificial Neural Network

Automated Identification of Minimal Myocardial Motion for Improved Image Quality on MR Angiography at 3 T

Influence of Velocity Encoding and Position of Image Plane in Patients with Aortic Valve Insufficiency Using 2D Phase Contrast MRI

Essentials of Clinical MR, 2 nd edition. 99. MRA Principles and Carotid MRA

Powerline Interference Reduction in ECG Using Combination of MA Method and IIR Notch

ρ = 4(νp)2 Scale -200 to 200 V = m/s Grad = 34 mmhg V = 1.9 m/s Grad = 14 mmhg Types

1Pulse sequences for non CE MRA

Measurement of Respiratory and Cardiac Motion Using a Multi Antenna Continuous Wave Radar Operating in the Near Field

Previous talks. Clinical applications for spiral flow imaging. Clinical applications. Clinical applications. Coronary flow: Motivation

MULTILEAD SIGNAL PREPROCESSING BY LINEAR TRANSFORMATION

: Biomedical Signal Processing

Automatic Detection of Abnormalities in ECG Signals : A MATLAB Study

Applying a Level Set Method for Resolving Physiologic Motions in Free-Breathing and Non-Gated Cardiac MRI

ECG INTERPRETATION MANUAL

Flow Quantification from 2D Phase Contrast MRI in Renal Arteries using Clustering

DETECTION OF HEART ABNORMALITIES USING LABVIEW

SMRT Student Scope Submission

The Electrocardiogram part II. Dr. Adelina Vlad, MD PhD

ELECTROCARDIOGRAPHY (ECG)

INTRODUCTION TO ECG. Dr. Tamara Alqudah

Removal of Baseline Wander from Ecg Signals Using Cosine Window Based Fir Digital Filter

Detection and Classification of QRS and ST segment using WNN

Go With The Flow: Role of 4D Flow Imaging

The Cardiac Cycle Clive M. Baumgarten, Ph.D.

11/18/13 ECG SIGNAL ACQUISITION HARDWARE DESIGN. Origin of Bioelectric Signals

Objectives. CMR Volumetric Analysis 8/25/11. CMR Volumetric Analysis Technique. Cardiac imaging plane acquisition. CMR Volumetric Analysis

ISSN: ISO 9001:2008 Certified International Journal of Engineering and Innovative Technology (IJEIT) Volume 2, Issue 10, April 2013

Introduction to ECG Gary Martin, M.D.

ECG Signal Characterization and Correlation To Heart Abnormalities

Biomedical Signal Processing

Introduction. Cardiac Imaging Modalities MRI. Overview. MRI (Continued) MRI (Continued) Arnaud Bistoquet 12/19/03

Heart Abnormality Detection Technique using PPG Signal

Temporal Analysis and Remote Monitoring of ECG Signal

Local Pulse Wave Velocity Estimation in the Carotids Using Dynamic MR Sequences

Comparison of Different ECG Signals on MATLAB

Clinical Applications

DIFFERENCE-BASED PARAMETER SET FOR LOCAL HEARTBEAT CLASSIFICATION: RANKING OF THE PARAMETERS

- why the T wave is deflected upwards although it's a repolarization wave?

Measurement of Ventricular Volumes and Function: A Comparison of Gated PET and Cardiovascular Magnetic Resonance

New Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor. Cardiothoracic Radiology

Electrocardiography Biomedical Engineering Kaj-Åge Henneberg

SPECTRAL ANALYSIS OF LIFE-THREATENING CARDIAC ARRHYTHMIAS

CFD Challenge: Simulation of Hemodynamics in a Patient-Specific Aortic Coarctation Model

Relationship of Number of Phases per Cardiac Cycle and Accuracy of Measurement of Left Ventricular Volumes, Ejection Fraction, and Mass

An electrocardiogram (ECG) is a recording of the electricity of the heart. Analysis of ECG

How to Learn MRI An Illustrated Workbook

Prospective cardiac motion self-gating

Vital Responder: Real-time Health Monitoring of First- Responders

PERFORMANCE CALCULATION OF WAVELET TRANSFORMS FOR REMOVAL OF BASELINE WANDER FROM ECG

Improvement of cardiac imaging in electrical impedance tomography by means of a new

A Magnetic Resonance Imaging Method for

Cardiac CT - Coronary Calcium Basics Workshop II (Basic)

ECG Acquisition System and its Analysis using MATLAB

Impaired Regional Myocardial Function Detection Using the Standard Inter-Segmental Integration SINE Wave Curve On Magnetic Resonance Imaging

ECG SIGNS OF HYPERTROPHY OF HEART ATRIUMS AND VENTRICLES

Lect.6 Electrical axis and cardiac vector Cardiac vector: net result Vector that occurs during depolarization of the ventricles Figure:

Heart Rate Calculation by Detection of R Peak

3/26/15 HTEC 91. EKG Sign-in Book. The Cardiac Cycle. Parts of the ECG. Waves. Waves. Review of protocol Review of placement of chest leads (V1, V2)

CARDIOVASCULAR PHYSIOLOGY ECG. Dr. Ana-Maria Zagrean

MORPHOLOGICAL CHARACTERIZATION OF ECG SIGNAL ABNORMALITIES: A NEW APPROACH

Electrocardiography negative zero LA/VL RA/VR LL/VF recording electrode exploring electrode Wilson right arm right arm, left arm left arm

The Rate-Adaptive Pacemaker: Developing Simulations and Applying Patient ECG Data

ACR MRI Accreditation: Medical Physicist Role in the Application Process

Assessment of Cardio- & Neurovascular Hemodynamics in the Human Circulatory System using 4D flow MRI

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography

Visualization strategies for major white matter tracts identified by diffusion tensor imaging for intraoperative use

Removal of Baseline wander and detection of QRS complex using wavelets

Self-Gated Cardiac Cine MRI

cardiac imaging planes planning basic cardiac & aortic views for MR

Measurement of Left Ventricular Velocities: Phase Contrast MRI Velocity Mapping Versus Tissue-Doppler-Ultrasound in Healthy Volunteers

FREQUENCY DOMAIN BASED AUTOMATIC EKG ARTIFACT

Transcription:

Med Biol Eng Comput (2008) 46:729 733 DOI 10.1007/s11517-008-0307-1 TECHNICAL NOTE Extraction of the magnetohydrodynamic blood flow potential from the surface electrocardiogram in magnetic resonance imaging Grace M. Nijm Æ Steven Swiryn Æ Andrew C. Larson Æ Alan V. Sahakian Received: 30 July 2007 / Accepted: 16 January 2008 / Published online: 1 February 2008 Ó International Federation for Medical and Biological Engineering 2008 Abstract The magnetohydrodynamic effect generates voltages related to blood flow, which are superimposed on the electrocardiogram (ECG) used for gating during cardiac magnetic resonance imaging (MRI). A method is presented for extracting the magnetohydrodynamic signal from the ECG. The extracted magnetohydrodynamic blood flow potential may be physiologically meaningful due to its relationship to blood flow. Removal of the magnetohydrodynamic voltages from the ECG can potentially lead to improved gating and diagnostically useful ECGs. Keywords Magnetohydrodynamic effect Electrocardiogram Magnetic resonance imaging Blood flow Cardiac Abbreviations MHD Magnetohydrodynamic MRI Magnetic resonance imaging G. M. Nijm S. Swiryn A. C. Larson A. V. Sahakian (&) Department of Electrical Engineering and Computer Science, Northwestern University, 2145 Sheridan Road, Evanston, IL 60208, USA e-mail: sahakian@ece.northwestern.edu G. M. Nijm e-mail: g-nijm@northwestern.edu S. Swiryn HeartCare Midwest, Peoria, IL, USA A. C. Larson Department of Radiology, Northwestern University Medical School, Chicago, IL, USA A. C. Larson A. V. Sahakian Department of Biomedical Engineering, Northwestern University, 2145 Sheridan Road, Evanston, IL 60208, USA ECG PC GRE 1 Introduction Electrocardiogram Phase contrast Gradient echo The magnetohydrodynamic (MHD) potential arises when a conductive fluid such as blood travels through a magnetic field, inducing a voltage perpendicular to both the magnetic field lines and the direction of the fluid flow [2, 11]. The magnitude of this induced voltage is directly related to the magnetic flux density, fluid velocity, and inter-electrode spacing. The induced voltage V may be expressed as: V ¼ Z L 0 u B dl where u is the velocity (m/s), B is the magnetic flux density (T), and L is the distance vector between electrodes (m). The MHD blood flow potential may also be expressed by a simplified equation: V ¼ BLu which makes the assumption that B, L, and u are uniform and all orthogonal to one another. These equations provide a simplified method for calculating the MHD blood flow potential by assuming that flow is in a straight, cylindrical, rigid tube. Maxwell s and Navier Stokes equations are needed to completely describe the generated MHD blood flow potentials; consequently, computational methods are typically used for such calculations. The MHD blood flow potential is of clinical importance because it generates voltages which distort the

730 Med Biol Eng Comput (2008) 46:729 733 electrocardiogram (ECG) during cardiac MRI [10]. The MHD potential arises because of blood flow within a strong static magnetic field. It results in denigration of the diagnostic quality of the ECG, particularly for examination of the S T segment and T wave [9], since these typically correspond to time periods of rapid ejection of blood from the ventricles, and thus, a large voltage contribution from the MHD blood flow potential. The ECG is used for gating, which coordinates the timing of MR image acquisition. Improper triggering can occur because of voltages resulting from the MHD blood flow potentials. Alternative gating methods, such as vectorcardiography and self-gating, have been proposed [1, 3, 5]. While these techniques may result in sufficiently accurate gating, they do not provide a diagnostically useful ECG during MRI since they use means other than the ECG to produce cardiac gating signals. Since the MHD potential is related to blood flow, we hypothesized that it can be extracted from the ECG and characterized as a meaningful physiological signal on its own. Cancellation of the MHD blood flow potential can likely improve gating as well as the diagnostic quality of the ECG. As scanners evolve from 1.5T to 3.0T in clinical practice, distortion caused by the MHD blood flow potential only worsens since its magnitude is directly related to magnetic field strength. Consequently, removal of the MHD blood flow potential from the ECG is becoming increasingly more practically important. 2 Methods The study protocol was approved by our institutional review board, and written informed consent was obtained from each subject. A Siemens 3.0T Trio MRI scanner with a wireless MR-compatible ECG lead system was used for the study. The ECG electrode pair was placed on the chest and oriented in the transverse plane in order to be orthogonal to the magnetic field lines of the static magnetic field. With this arrangement, blood flow through the aortic arch was roughly perpendicular to both the magnetic field lines and the electrode pair. The ECG electrode pair was centered on the fourth intercostal space at the left sternal border with an inter-electrode spacing of 10 cm. Thirty second ECG recordings, sampled at 100 Hz, were taken during breath holds to minimize motion artifacts due to respiration. The ECG recordings were taken with subjects in the supine and prone positions with subjects outside the magnet, inside the magnet without scanning, and inside the magnet during conventional gradient echo (GRE) scans. MATLAB (The MathWorks Inc., Natick, MA) code was written to perform the signal processing tasks. First, baseline wander was removed by a highpass filter with a cutoff frequency of 0.5 Hz. Next, QRS complexes were detected using a modified version of the Pan and Tompkins algorithm [6] in order to identify each PQRST segment, or beat, of the ECG signal. Template matching was applied, which involved finding the mean of beats which had at least a 95% correlation with the median beat [7]. Finally, the MHD blood flow potential was extracted by subtracting the mean beat generated when the subject was outside the magnet from the mean beat generated when the subject was inside the magnet. The MHD blood flow potential was also extracted from the subtraction of the mean beat generated with the subject outside the magnet from the mean beat generated with the subject inside the magnet during a conventional cine GRE scan. Cine GRE sequence parameters used were TR = 5.0 ms, TE = 2.6 ms, FOV = 340 mm, flip-angle = 15, voxel size = 2.3 9 1.8 9 6.0 mm, BW = 450 Hz/pixel. GRE cine was chosen because it is a commonly used during cardiac MRI protocols [8]. The MHD blood flow potentials generated without scanning were compared with those generated while scanning. The extracted MHD blood flow potential was analyzed as a signal, referred to as the MHD signal. Measurements of the morphology and timing of the MHD signals included the maximum amplitude of the negative deflection of the MHD signal, the duration of the MHD negative deflection, and the delay between the onset of the QRS complex and the onset of the MHD negative deflection. A sample MHD signal indicating these measurements is shown in Fig. 1. In addition to the ECG recordings, axial orientation 2D phase contrast (PC) MRI scans were performed Fig. 1 Sample MHD signal with measurements of duration of the MHD negative deflection, maximum amplitude of the MHD negative deflection, and delay between onset of QRS complex and onset of the MHD negative deflection indicated. In the figure, time t = 0 corresponds to the P-wave onset in the original ECGs from which the MHD signal was derived

Med Biol Eng Comput (2008) 46:729 733 731 independently to allow blood flow measurements in the descending aorta near the aortic arch. These scans were performed to compare the timing of the blood flow with that observed on the MHD signal; the time resolution of the PC MRI scans was 50 ms. The sequence parameters used were TR = 66.0 ms, TE = 3.0 ms, FOV = 320 mm, flipangle = 20, voxel size = 2.0 9 1.3 9 5.0 mm, BW = 383 Hz/pixel, velocity encoding = 150 cm/s. Argus postprocessing software was used to draw the region-of-interest within the lumen of the vessel in order to measure mean flow through the cardiac cycle. 2.1 Subject population Data from ten healthy subjects were recorded, but the ECGs for one subject were eliminated from analysis due to poor signal quality attributed to electrode failure. Of the remaining nine subjects, three were female and six were male, ranging in age from 20 to 31 years (mean ± standard deviation, 24 ± 3 years). The subjects average resting heart rate ranged from 46 to 76 beats per minute (bpm) (60 ± 11 bpm). The resting heart rate for each subject did not significantly change during the study. 3 Results 3.1 MHD signal characteristics MHD signals were similar in all subjects in both the supine and prone positions and included a prominent negative deflection during the S-T segment and T wave. Figure 2 shows the MHD signals from all subjects for both the supine and prone positions. The maximum amplitude of the MHD negative deflection ranged from 0.13 to 0.27 mv (mean ± SD, 0.19 ± 0.04 mv) for the supine position compared to 0.14 to 0.30 mv (0.21 ± 0.08 mv) for the prone position. There was no statistically significant difference (NS) in amplitude between the supine and prone positions. The duration of the MHD negative deflection ranged from 320 to 380 ms (340 ± 20 ms) for the supine position compared to 300 to 360 ms (330 ± 20 ms) for the prone position (NS). The delay between the onset of the QRS complex and the onset of the MHD negative deflection ranged from 110 to 190 ms (140 ± 20 ms) for both the supine and prone positions (NS). 3.2 Effect of GRE cine scanning The presence of GRE scanning during ECG signal acquisition had a minimal effect on the resulting MHD signals. Fig. 2 MHD signals for all subjects for one cardiac cycle. a Subjects in the supine position, b Subjects in the prone position. The signals are aligned on the minimum of the primary negative deflection; they are different lengths due to different cardiac cycle lengths among the subject population The maximum amplitude of the MHD negative deflection ranged from 0.13 to 0.28 mv (0.19 ± 0.05 mv) for the supine position and 0.13 to 0.30 mv (0.20 ± 0.06 mv) for the prone position; these values were not statistically significantly different from those obtained without scanning. Neither the duration of the MHD negative deflection nor the delay between the onset of the QRS complex and the onset of the MHD negative deflection differed from those obtained without scanning (NS). In addition, for all subjects the correlation between the two MHD signals (with and without scanning) ranged from 97.8 to 99.4%. 3.3 Comparison with MRI flow measurements MHD signals were compared with standard MRI flow measurements in the descending aorta. The negative deflection following the QRS complex was clearly defined in both MHD and MRI flow signals for all subjects. The

732 Med Biol Eng Comput (2008) 46:729 733 duration of the MRI flow signal negative deflection ranged from 280 to 370 ms (330 ± 40 ms), which was comparable to the MHD negative deflection duration, which ranged from 320 to 380 ms (340 ± 20 ms) for the supine position and 300 to 360 ms (330 ± 20 ms) for the prone position (NS). The minimum of the MRI flow signal lagged behind the minimum of the MHD signal in all subjects; the difference in timing between the minima of these signals ranged from 20 to 90 ms (60 ± 20 ms) for the supine position and 10 to 80 ms (50 ± 20 ms) for the prone position (NS). The MHD signals and MRI flow data for a sample subject are shown in Fig. 3. 4 Discussion The main finding of this work was that the MHD blood flow potential can be extracted from the ECG obtained in the MRI magnet. In theory, the MHD blood flow potential should reflect the net effect of the blood flow occurring at a particular instant and orientation, as registered on the Fig. 3 MHD signal (solid line) and MRI flow (dashed line). a Subject in the supine position, b Subject in the prone position surface of the body. A primary contributor to the MHD blood flow potential observed in this study was likely blood flow in the aortic arch for several reasons. First, the velocity of blood is greatest in the aorta immediately following ejection from the left ventricle. Second, the aorta is the largest diameter blood vessel. Finally, blood flow through the aortic arch is approximately perpendicular to both the electrode pair used in this study and the magnetic field lines of the MRI magnet. The magnitude of the MHD blood flow potential is influenced by the velocity of the fluid and the orientation of flow relative to the electrode pair and the magnetic field lines. Though blood flow in other vessels which are orthogonal to the magnetic field lines and the electrode pair may also contribute to the magnitude of the MHD blood flow potential, the velocity of blood flow is greatest in the aorta; therefore, due to its orientation, flow in the aortic arch should make the most significant contribution to the observed MHD blood flow potential. Since the aortic arch is roughly orthogonal to the magnetic field lines and the electrode pair in both the supine and prone positions, the MHD signals were, as expected, not significantly different when subjects were supine or prone. The observed delay between the onset of the QRS complex and the onset of the MHD deflection agreed with the anticipated lag as a result of the delay between electrical activation and mechanical activation, time for isovolumetric contraction prior to ejection, and time for the flow to propagate into the aortic arch [4]. Scanning during ECG signal acquisition had a minimal effect on the resulting MHD signal. This is reasonable because radiofrequency interference was removed by filtering, and an optimized lead system as well as the use of template matching minimized the switching MR gradient artifacts. The correlation in timing and similarity in morphology between the independently obtained MRI flow data and the MHD signal help to verify that the MHD signal is in fact flow-related. The coarse time resolution of 50 ms for the MRI flow data may explain the small discordance in timing between the two signals. Although MRI flow measurements were used for sake of comparison in this study, a major potential advantage of flow data obtained from the ECG is its superior temporal resolution in comparison to MRI flow data. In this study, the ECG was sampled at 100 Hz while MRI flow measurements were acquired only every 50 ms, or at a rate of 20 Hz. In addition, the negative deflection of the MHD signal begins after and ends before the negative deflection of the MRI flow signal because the MHD signal is generated over a shorter time interval than the MRI flow signal. Blood flow through the aortic arch largely produces the negative deflection of the MHD signal; however, the negative deflection of the MRI flow signal is produced before flow has propagated into the

Med Biol Eng Comput (2008) 46:729 733 733 aortic arch and continues after the bolus of blood has traveled through the aortic arch. The next set of challenges will be to use these methods not only to improve gating but also to record a diagnostic quality ECG during cardiac MRI. Further application of this work has potential to make use of the MHD signal as a physiologically meaningful signal due to its relationship to blood flow. 5 Conclusions The MHD signal can be extracted from the ECG obtained in the MRI magnet. The MHD signal so obtained correlates well in timing with the flow signal acquired from standard MRI flow measurement techniques. Acknowledgments This material is based upon work supported under a National Science Foundation Graduate Research Fellowship, NIH HL079148, and a grant from the Dr. Scholl Foundation. References 1. Crowe ME, Larson AC, Zhang Q, Carr J, White RD, Li D, Simonetti OP (2004) Automated rectilinear self-gated cardiac cine imaging. Magn Reson Med 52:782 788 2. Dimick RN, Hedlund LW, Herfkens RJ, Fram EK, Utz J (1987) Optimizing electrocardiograph electrode placement for cardiacgated magnetic resonance imaging. Invest Radiol 22:17 22 3. Fischer SE, Wickline SA, Lorenz CH (1999) Novel real-time R-wave detection algorithm based on the vectorcardiogram for accurate gated magnetic resonance acquisitions. Magn Reson Med 42:361 370 4. Guyton AC, Hall JE (2000) Textbook of medical physiology, 10th edn. W.B. Saunders, Philadelphia 5. Larson AC, White RD, Laub G, McVeigh ER, Li D, Simonetti OP (2004) Self-gated cardiac cine MRI. Magn Reson Med 51:93 102 6. Pan J, Tompkins WJ (1985) A real-time QRS detection algorithm. IEEE Trans Biomed Eng 32:230 236 7. Slocum J, Byrom E, McCarthy L, Sahakian AV, Swiryn S (1985) Computer-detection of atrioventricular dissociation from surface electrocardiograms during wide QRS complex tachycardias. Circulation 72:1028 1036 8. Sola S, White RD, Desai M (2006) MRI of the heart: Promises fulfilled? Cleve Clin J Med 73:663 670 9. Tenforde TS, Gaffey CT, Moyer BR, Budinger TF (1983) Cardiovascular alterations in Macaca monkeys exposed to stationary magnetic fields: experimental observations and theoretical analysis. Bioelectromag 4:1 9 10. Tenforde TS (2005) Magnetically induced electric fields and currents in the circulatory system. Prog Biophys Mol Biol 87:279 288 11. Togawa T, Okai O, Oshima M (1967) Observation of blood flow e.m.f. in externally applied strong magnetic field by surface electrodes. Med Biol Eng 5:169 170