Analysis of Organic Fertilizers for Nutrients with AAnalyst 800 Atomic Absorption Spectrophotometer

Similar documents
Trace Metal Characterization of Soils Using the Optima 7300 DV ICP-OES

Trace Elemental Characterization of Edible Oils with Graphite Furnace Atomic Absorption Spectrophotometer

Analysis of Micronutrients in Milk by Flame Atomic Absorption Using FAST Flame Sample Automation for Increased Sample Throughput

Determination of available nutrients in soil using the Agilent 4200 MP-AES

Baby Food and Juice Compendium BABY FOOD AND JUICE

Purity Tests for Modified Starches

Application note. Determination of metals in soil by microwave plasma - atomic emission spectrometry (MP-AES) using DTPA extraction.

Application Note novaa 800 D. Determination of Macro and Trace Minerals as well as Toxic Trace Metals in Powdered Milk. Challenge.

Analysis of trace elements in nutraceuticals in compliance with USP chapter <2232> Elemental Contaminants in Dietary Supplements

Admixtures for concrete, mortar and grout Test methods Part 12: Determination of the alkali content of admixtures

US EPA SW-846 Method 6010D using the Thermo Scientific icap 7400 ICP-OES Duo

Determination of metals in industrial wastewaters by microwave plasmaatomic emission spectrometry

Determination of Copper in Green Olives using ICP-OES

SECTIOn 3. Multi-Element ClP Standards for ICP & ICP-MS

Evaluating the performance of the OneNeb Series 2 Nebulizer with PerkinElmer Optima 7/8x00 Series ICP-OES systems

Corn Starch Analysis B-47-1 PHOSPHORUS

MULTI-COMPONENT ANALYSIS OF HEAVY METALS

Determination of Calcium in Serum Samples by AAS Using a Fuel Lean Flame

Journal of Chemical and Pharmaceutical Research

Analysis of Milk for Major and Trace Elements by ICP-MS

High performance graphite furnace tube for determination of lead in blood

Analysis of Common Sweeteners and Additives in Beverages with the PerkinElmer Flexar FX-15 System Equipped with a PDA Detector

Guidelines for Troubleshooting and Maintenance of AA Systems. Presented by Eric Vanclay, Atomic Spectroscopy Consumables Product Manager

Note: The use of different equipment configurations (standard, XI, PlasmaScreen...) will treat different experiment Other services ICP- MS

Performance Characteristics of the Agilent High Matrix Sample Introduction (HMI) Accessory for the 7500 Series ICP-MS. Product Overview.

Research Article Derivative Spectrophotometric Method for Estimation of Metformin Hydrochloride in Bulk Drug and Dosage Form

Routine Analysis of Fortified Foods using the Agilent 7800 ICP-MS

Analysis of Isoflavones with the PerkinElmer Flexar FX-15 UHPLC System Equipped with a PDA Detector

Lead and Cadmium Levels in Some Commercially Available Local and Imported Peach Juice Samples in Alkoms City

Determination of selected micro and macronutrients in sugarcane growing soils at Kakamega North District, Kenya

Nitrophoska. Cereals, fodder beet, horticulture, maize and vegetables. Precise nutrition for superior plant performance

Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinElmer Altus UPLC System with PDA Detection

Estimation Of Copper, Zinc And Lead In Choornas By Flame Atomic Absorption Spectroscopy Krishna veni N 1 *, Meyyanathan S N 1 and Suresh B 2

TECHNICAL BULLETIN METHOD 1: DETERMINATION OF TOTAL DIETARY FIBRE

Understanding a Soil Report

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No.

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE

Food safety is an area of growing concern as a result of pollution,

Analysis of Cannabinoids in Cannabis by UHPLC Using PDA Detection

By Andrew & Erin Oxford, Bethel

Bottom Ash Data Week 37

Matrix Interferences in ICP-MS: Causes, Effects, and Strategies to Reduce or Eliminate Them

Selenium, atomic absorption spectrometric, hydride, automated

Lutein Esters from Tagetes Erecta

02006B 1 vial 02006B 1 vial Store at -20 C. Lyophilized recombinant IL-2

Rat Insulin ELISA. For the quantitative determination of insulin in rat serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.

Journal of Chemical, Biological and Physical Sciences

Bottom Ash Data Week 30

METHOD OF TEST FOR DETERMINATION OF INSOLUBLE RESIDUE OF CARBONATE AGGREGATES

USER SPECIFICATIONS FOR QUINTOLUBRIC 888 Series DESCRIPTION OF THE MOST IMPORTANT PROPERTIES AND THE POSSIBLE VARIATIONS AND TOLERANCES

IJRPC 2011, 1(4) Rohan et al. ISSN: INTERNATIONAL JOURNAL OF RESEARCH IN PHARMACY AND CHEMISTRY

A reliable, practical and easy-to-use Determination of Soil Conductivity and Salt Content with TRIME Probes

Rat C-peptide ELISA. For the quantitative determination of C-peptide in rat serum

Bottom Ash Data Week 17

Supplying Nutrients to Crops

TOP 7 ACTIVITIES OF CEN AND ISO STANDARDIZATION OF METHODS FOR ANALYSIS OF ANIMAL FEEDS

Development and validation of stability indicating RP-LC method for estimation of calcium dobesilate in pharmaceutical formulations

FERTILIZING GREENHOUSE CROPS

Optimization of digestion methods for mineral in spirullina by Inductively Coupled Plasma-Optical Emission Spectrometry (ICP-OES)

GB Translated English of Chinese Standard: GB NATIONAL STANDARD

AVIO 200 ICP-OES CONSUMABLES AND SUPPLIES. Consumables and Supplies Reference Guide

Analysis of Cannabinoids in Hemp Seed Oils by HPLC Using PDA Detection

Bovine Insulin ELISA

Use of Soil and Tissue Testing for Sustainable Crop Nutrient Programs

CORESTA Recommended Method No. 84

Soil Testing Options in High Tunnels. Bruce Hoskins University of Maine anlab.umesci.maine.edu

Multi Analyte Custom Grade Solution

Porcine/Canine Insulin ELISA

Greenhouse Horticulture

CORESTA RECOMMENDED METHOD N 8

BOTANY AND PLANT GROWTH Lesson 9: PLANT NUTRITION. MACRONUTRIENTS Found in air and water carbon C oxygen hydrogen

BRIEFING Assay + + +

Analysis of Cannabinoids in Hemp Seed Oils by HPLC Using PDA Detection

Bottom Ash Data Week 9

Mouse C-peptide ELISA

Effects of FGD-Gypsum, Used-Wallboard and Calcium Sulfate on Corn and Soybean Root Growth

METHOD 8316 ACRYLAMIDE, ACRYLONITRILE AND ACROLEIN BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC)

YaraVita PROCOTE. The colors of yield.

Chapter 1: Overview of soil fertility, plant nutrition, and nutrient management

Mouse C-peptide ELISA

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF

Bottom Ash Data Week 38

App Note. Heavy Metal Contamination of Red Pepper Spices and Hot Sauce

USP <232> and <233> Understanding Your Path to Compliance with the New Elemental Impurity Chapters. Steve Wall Agilent Technologies

ELEMENTAL COMPOSITION IN VARIOUS THALLUS PARTS OF THREE BROWN SEAWEEDS FROM KARACHI COAST

Rat C-peptide ELISA. For the quantitative determination of C-peptide in rat serum. For Research Use Only. Not For Use In Diagnostic Procedures.

It is recommended to use titanium coils. ph meter to control ph of the process. It is advised to use an automatic dosage for the maintenance products.

Fertilization Programming

Rat Proinsulin ELISA

Nitrate/Nitrite Assay Kit Manual Catalog #

Water Soluble Fertilizer for Foliar Application

COMPARATIVE ANALYSIS FOR MACRO AND TRACE ELEMENTS CONTENT IN GOJI BERRIES BETWEEN VARIETIES FROM CHINA AND R. MACEDONIA

FERTILISER AND LIME QUALITY REPORT 2017

Optimization of the Mark VI Flame Atomization System

V. LAB REPORT. PART I. ICP-AES (section IVA)

Mouse Ultrasensitive Insulin ELISA

Elemental Scientific Semiconductor Catalog for Thermo ELEMENT2

FACT SHEET. Understanding Cation Exchange Capacity and % Base Saturation

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE

Supplementary materials for manuscript entitled Development and Analytical

Transcription:

application Note Atomic Absorption Author Praveen Sarojam, Ph.D. PerkinElmer Global Application Center Mumbai, India Analysis of Organic Fertilizers for Nutrients with AAnalyst 800 Atomic Absorption Spectrophotometer Introduction The fertilizer industry helps to ensure that farmers have the nutrients they need to grow enough crops to meet the world's requirements for food, feed, fiber and energy. Nutrients in manufactured fertilizers are in the form that can be absorbed readily by the plants. All of these nutrients exist in nature, but the quantities are not sufficient to meet the needs of our growing, urbanized population. Soils may be naturally low in nutrients, or they may become deficient due to nutrient removal by crops over the years without replenishment or when high-yielding varieties are grown that have higher nutrient requirements than do local varieties. All of the essential nutrients are important but in varying quantities. Macronutrients (N, P, K, Ca, Mg, etc.) are needed by plants in large quantities. The primary nutrients are nitrogen, phosphorus and potassium. Today, sulphur is also considered a key macronutrient. Macronutrients include both primary and secondary nutrients. Micronutrients (or trace elements ) (Fe, Mn, Zn, Cu, Ni, etc.) are required in very small amounts for correct plant growth. They need to be added in small quantities when they are not provided by the soil. Every plant nutrient, whether required in large or small amounts, has a specific role in plant growth and food production. One nutrient cannot be substituted for another. For example, potassium activates more than 60 enzymes (the chemical substances that govern life and play a vital part in carbohydrate and protein synthesis). It improves a plant's water regime and increases tolerance to drought, frost and salinity. Plants that are well supplied with potassium are less affected by disease. Magnesium is the central constituent of chlorophyll, the green pigment in leaves that functions as an acceptor of the energy supplied by the sun: 15-20% of the magnesium in a plant is found in the green

parts. Magnesium is also involved in enzyme reactions related to energy transfer. Fertilizers are needed to obtain high yields because they supply crops with the nutrients the soil lacks. By adding fertilizers, crop yields can often be doubled or even tripled. The UN Food and Agriculture Organization (FAO) Fertilizer Program undertook extensive demonstrations and trials in 40 countries over a period of 25 years. The weighted average increase resulting from the best fertilizer treatment for wheat was about 60%. Fertilizers contribute to efficient use of land and water. They ensure the most effective use of both land and water. Where rainfall is low or crops are irrigated, the yield per unit of water used may be more than doubled and the rooting depth of the crop increased through fertilizer application. The present work demonstrates the ability of the PerkinElmer AAnalyst 800 atomic absorption spectrophotometer to analyze the micronutrients in several organic fertilizers. The results obtained from using the conventional AOAC Method 965.09 using dry ashing with a muffle furnace and EPA Method 3052, the official EPA method for the microwave digestion of siliceous and organic based samples, are compared. Experimental The measurements were performed using the PerkinElmer AAnalyst 800 atomic absorption spectrophotometer (PerkinElmer, Inc., Shelton, CT, USA) (see Figure 1) equipped with WinLab32 for AA Version 6.5 software, which features all the tools to analyze samples, report and archive data and ensure regulatory compliance. The revolutionary doublebeam design of the AAnalyst 800 provides continuous, simultaneous measurement of both sample and reference beams for flame measurements. Utilizing a high-efficiency fiber optic, the sample and reference beams are simultaneously focused on the state-of-art, solid-state detector. This translates into increased signal-integration time (without increasing analysis time) for improved precision, detection limits and reproducibility. The totally corrosion-resistant solid titanium burner head easily installs onto the spray chamber, constructed of an inert polymeric material that provides superior performance for corrosive and high solid matrices. The systems include an eight-lamp turret mount with builtin power supplies, for the automatic alignment of both the hollow cathode lamps and electrodeless discharge lamps. A PerkinElmer corrosion-resistant nebulizer, which can be used for solutions containing hydrofluoric acid, was used for all flame absorption measurements. A single slot air-acetylene 10cm burner head was used for all air-acetylene elements. A Multiwave 3000 Microwave Oven (PerkinElmer/Anton Paar) was used for the microwave-assisted digestion. This is an industrial-type oven which is equipped with various accessories to optimize the sample digestion. The samples were digested using the Rotor 8XF100 that consists of eight 100 ml high pressure vessels/liners made of PTFE-TFM material protected by its individual ceramic jackets. TFM is chemically modified PTFE that has enhanced mechanical properties at high temperatures compared to conventional PTFE. This vessel/liner has a working pressure of 60 bar (870 psi) and operational temperatures up to 260 C. A Pressure/Temperature (P/T) Sensor Accessory was also used for this work. The P/T sensor simultaneously measures temperature and pressure in one reference vessel/ liner. All vessels temperatures were also monitored with the IR Temperature Sensor Accessory. This measures the temperature of the bottom surface of each vessel/liner, thus providing protection in case of over-temperature during the process of digestion. PerkinElmer NIST -traceable calibration standards for atomic spectroscopy were used as the stock standards for preparing working standards. All the working standards were prepared daily in ASTM type I water acidified with 1% Suprapur nitric acid. Working standards were prepared in polypropylene vials (Sarstedt ) on volume-by-volume dilution. Eppendorf (Germany) micropippetes and tips were used for pippetting solutions. Certified Reference Standard for trace metals in river sediment and soil solutions from High Purity Standards (Lot # 0800230 and Lot # 0733733) was used for quality control. Multi-element ICP standard for trace metal ions in 5% HNO 3 from Spex Certiprep, (New Jersey, USA) prepared Figure 1. PerkinElmer AAnalyst 800 atomic absorption spectrophotometer. Figure 2. PerkinElmer Multiwave 3000 microwave digestion system. 2

at midpoint of the calibration curve was used as the quality control check standard. The nitric acid and hydrofluoric acid used were also of Suprapur grade, from Merck in Germany. Sample Preparation Three organic based fertilizer samples, one of them (Sample 3) urea, were collected from the local market. Representative fertilizer samples in fine particles were prepared by grinding them in an agate mortar with its pestle, then passing them through a 20 mesh sieve. The microwave digestion of soil samples was performed in accordance with EPA Method 3052. This method is applicable to the microwave-assisted acid digestion of siliceous matrices, and organic matrices and other complex matrices. This method is provided as a rapid multi-element, microwave-assisted acid digestion prior to analysis protocol so that decisions can be made about the site or material. The goal of this method is total sample decomposition for most matrices with judicious choice of acid combinations. Approximately 0.5 g of the fertilizer sample was accurately weighed directly into the PTFE-TFM digestion vessel/liners. Nine ml of concentrated nitric acid and 3 ml of concentrated hydrofluoric acid were added to this sample. Reagent blanks were also prepared using the same volumes of these two acids. The vessels were sealed and placed into the Rotor 8XF100 for the microwave digestion. After the digestion process, the digestate liquids were transferred to the 50 ml polypropylene autosampler vials and made up in 25 ml volume with the laboratory ASTM type I water. Muffle furnace ashing was done as follows: Approximately 1 g of fertilizer sample was accurately weighed into a 150 ml beaker (Pyrex, or equivalent), charred on a hot plate and ashed for 1 to 3 hours at 500 C in a muffle furnace with its door open to allow free access of air. The residue cake was ground into fine particles for dissolution in 10 ml of concentrated HCl. This solution was boiled and evaporated on a hot plate almost to dryness, while not allowing it to boil dry, then re-dissolved in 20 ml 2 M HCl, boiling gently if necessary. The solution was filtered through fast filter paper into 100 ml volumetric flask. The filter paper and residue were thoroughly washed with water and drained into the volumetric flask. This final solution is analyzed directly, or with dilution with 0.5 M HCl such that its absorbance was within the concentration range of the instrument. If calcium (Ca) is to be determined, the solution was added with enough lanthanum (La) stock solution to make the final dilution to contain 1% La. Results and Discussion The AOAC method for the determination of nutrients in fertilizers has been performed with a laboratory grade muffle furnace. This is a time consuming process for sample digestion, whereas microwave digestion is easier and reliable when compared with conventional sample preparation. Standard calibrations were prepared for all the metal ions of interest with three standard and one blank solution. Accuracy of the developed method was tested by analyzing two certified reference materials in solutions obtained from High Purity Standards. The recoveries were excellent which demonstrate the accuracy and reliability of the developed method. In order to verify the calibration curve generated, a quality control standard prepared from Spex Certiprep was aspirated at midpoint of calibration for all the metal ions. The recovery of QC sample done immediately after calibration and at the end of analysis was excellent (within the 90-110%). To demonstrate the ability of AAnalyst 800 to work in difficult matrices, spike recovery studies were carried out and were found to have excellent results (90-110%). Method detection limits were calculated on the basis of standard deviation of seven replicates of reagent blank (student t value of 3.14 for a confidence interval of 98%). Conclusions A method was developed for the sequential determination of nutrients in fertilizers by using the AAnalyst 800 atomic absorption spectrophotometer. Sensitivity and accuracy in flame AA are directly correlated to the optimization of the burner position as well as the gas flows. Careful optimization of these parameters, which are different for each element, maximizes sensitivity and can significantly reduce or eliminate interferences. The unique computer-controlled, motorized burner system and gas controls allow precise, automated adjustments. Since WinLab32 software allows optimized burner position and gas flows to be stored with an element Figure 3. Samples. 3

method, each element in a sequential multi-element run can be determined using its own optimized parameters. The ability of the PerkinElmer/Anton-Paar Multiwave 3000 microwave digestion system to digest the soil samples in accordance with EPA Method 3052 was demonstrated. The method detection limits obtained indicates that the reliable analysis of micronutrients in difficult matrices such as organic matrix-based fertilizer samples is possible with the AAnalyst 800 atomic absorption spectrophotometer. The analysis of reference materials and excellent spike recovery results gave further proof to the accuracy of the developed method. The low recovery of analytes obtained with muffle furnace ashing is due to analyte loss from evaporation. In the case of microwave digestion we had better recovery, and the time required to complete the decomposition was also shorter. The whole digestion process took one hour to complete, including cooling time. As the Multiwave 3000 system offers complete digestion of organic fertilizers in accordance with EPA Method 3052, it is the preferred technique for the sample preparation for organic based fertilizer samples. Table 1. Experimental Conditions for AAnalyst 800. Element Cu Fe Mn Zn Mg Ca Wavelength (nm) 324.8 248.3 279.5 213.9 285.2 422.7 Slit (nm) 0.7 0.2 0.2 0.7 0.7 0.7 Mode AA AA AA AA AA AA Flame Air-Ac Air-Ac Air-Ac Air-Ac Air-Ac Air-Ac Burner Head 10 cm 10 cm 10 cm 10 cm 10 cm 10 cm Calibration Linear through Linear through Linear through Linear through Linear through Linear through zero zero zero zero zero zero Lamp HCL HCL HCL HCL HCL HCL Lamp current, ma 15 30 20 15 6 10 Standards (mg/l) 1.0, 2.5 & 5.0 1.0, 2.5 & 5.0 0.1, 0.25 & 0.5 0.1, 0.25 & 0.5 0.1, 0.25 & 0.5 1.0, 2.5 & 5.0 Spiked concentration (mg/l) 2.5 2.5 0.25 0.25 0.25 2.5 Read Time (seconds) 3.0 3.0 3.0 3.0 3.0 3.0 Replicates 3 3 3 3 3 3 Air Flow (L/min) 17.0 17.0 17.0 17.0 17.0 17.0 Acetylene flow (L/min) 1.5 1.5 1.5 2.0 1.5 1.3 Table 2. Results of Fertilizer Analysis. Metal Sample 1 Sample 1 Sample 2 Sample 2 Sample 3 Sample 3 (Avg. of (Avg. of (Avg. of (Avg. of (Avg. of (Avg. of Duplicates) Duplicates) Duplicates) Duplicates) Duplicates) Duplicates) MDS Muffle Furnace MDS Muffle Furnace MDS Muffle Furnace Cu 49.25 ug/g 34.0 ug/g 210.49 ug/g 211.4 ug/g 1.40 ug/g 0.83 ug/g Fe 0.534% 48.8 ug/g 2.51% 481.4 ug/g 36.8 ug/g 41.1 ug/g Mg 0.85% 0.57% 1.51% 1.3% 98.23 ug/g 69.8 ug/g Ca 18.7% 16.2% 5.19% 4.2% 300 ug/g 607.7 ug/g Zn 0.014% 92.9 ug/g 0.016% 140.5 ug/g 3.54 ug/g 2.5 ug/g Mn 0.024% 159.3 ug/g 0.048% 512 ug/g 6.57 ug/g 1.32 ug/g Note: MDS = microwave-assisted digestion 4

Table 3. Results of CRM Analysis Soil Solution A, Lot # 0733733. Metal Certified Value (µg/ml) Recovery (%) Cu 0.30 ±0.5% 101.2 Fe 200 ±0.5% 109.4 Mg 70.0 ±0.5% 95.14 Ca 350 ±0.5% 96.0 Zn 1.0 ±0.5% 102.8 Mn 0.10 ±0.5% 105.4 Table 6. Method Detection Limits (MDLs). Metal MDL (µg/g) Cu 0.72 Fe 1.50 Mg 0.08 Ca 1.48 Zn 1.92 Mn 0.27 Table 4. Results of CRM Analysis River Sediment A, Lot # 0800230. Metal Certified Value (µg/g) Recovery (%) Cu 1.0 ±0.5% 104.9 Fe 1200 ±0.5% 114.9 Mg 70.0 ±0.5% 101.7 Ca 300 ±0.5% 109.4 Zn 15.0 ±0.5% 99.88 Mn 8.0 ±0.5% 99.81 Table 7. Microwave Temperature Program. Parameter Optimum condition Temperature 180 ±5.0 C Ramp time 5.5 min Hold time 9.5 min Cool down time 20 min Pressure 50 bar Amount of fertilizer 0.5 g HF 3.0 ml HNO 3 9.0 ml Table 5. Results of QC and Spike Recovery. Sample 1 Sample 2 Sample 3 Metal QC 1 QC 2 Spike Spike Spike (%) (%) Recovery (%) Recovery (%) Recovery (%) Cu 106.0 106.3 109.9 110.7 113.3 Fe 109.1 109.1 112.2 108 115.6 Mg 106.0 105.7 103.1 93.1 105.8 Ca 100.8 101.8 105.2 96.0 106.8 Zn 102.3 101.4 103 99.4 102 Mn 100.2 101.7 101.1 100.9 100.0 References 1. U.S. EPA Method 3052. 2. AOAC Method 965.09. 3. http://www.ces.ncsu.edu/cumberland/fertpage/inorganic. html 4. http://en.wikipedia.org/wiki/fertilizer PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com For a complete listing of our global offices, visit www.perkinelmer.com/contactus Copyright 2009, PerkinElmer, Inc. All rights reserved. PerkinElmer is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 008876_01