Orthodontic bracket slot dimensions as measured from entire bracket series

Similar documents
With judicious treatment planning, the clinical

Introducing the Clarity SL Self-Ligating Appliance System

INDICATIONS. Fixed Appliances are indicated when precise tooth movements are required

Advantages. Fantastic. systems. Together with Unitek Lateral Development Archwires, this system puts you in control to

Advantages. Fantastic. Treatment. Finishes. 3M Self-Ligating Appliances Lateral Development System. Unitek Lateral Development Archwires

Outline the significance of the pre-adjusted Edgewise appliance system and useful bracket variations in orthodontics

The Tip-Edge appliance and

Integrative Orthodontics with the Ribbon Arch By Larry W. White, D.D.S., M.S.D.

Use of a Tip-Edge Stage-1 Wire to Enhance Vertical Control During Straight Wire Treatment: Two Case Reports

Third-Order Torque and Self-Ligating Orthodontic Bracket Type Effects on Sliding Friction

THE RELIABLE AESTHETIC CHOICE

THE MBT VERSATILE+ APPLIANCE SYSTEM

Treatment of Class II, Division 2 Malocclusion with Miniscrew Supported En-Masse Retraction: Is Deepbite Really an Obstacle for Extraction Treatment?

Plus Combines Traditional Twin Bracket Treatment With Self-Ligation Convenience

Bracket Systems. A good bracket is the foundation of a good treatment plan... Go ahead, choose the best. Metal Bracket.

A New Low-Friction Ligation System

Study Model-based Evaluation of Built-in Tip, Torque, and In out Characteristics of a Third-generation Preadjusted Edgewise Appliance

There has been increased use of self-ligating. Friction between various self-ligating brackets and archwire couples during sliding mechanics

Victory Series Active Self-Ligating Brackets. Reliable. and. Effective

Incognito. Appliance System goes Digital. 3Unitek. Now in 3D. The Incognito Appliance System Digital Setup with 3D PDF Documentation

A New Fixed Interarch Device for Class II Correction

HYCON DEVICE: A PRECISE AND CONTROLLED METHOD OF SPACE CLOSURE A CASE REPORT

Sliding mechanics in an extraction case treated with 3M Clarity ADVANCED Ceramic Brackets and the 3M MBT Appliance System.

Victory Series Buccal Tubes Designed for Comfort and Performance. Advanced, Precise. Solutions. for Molars

The Ultimate in Crystal Clear Aesthetics

Virtual Treatment Planning

Experience with Contemporary Tip-Edge plus Technique A Case Report.

SnapLink Re-Convertible Tube

ch00-m3395-plims 8/3/07 21:29 Page i SmartClip TM Self-Ligating appliance System

Arch dimensional changes following orthodontic treatment with extraction of four first premolars

Controlled Space Closure with a Statically Determinate Retraction System

The Rondeau System. Brackets, Bands, Buccal Tubes and Archwires. Orthodontic supplies recommended by Dr. Brock Rondeau

A THESIS SUBMITTED TO THE FACULTY OF THE GRADUATE SCHOOL OF THE UNIVERSITY OF MINNESOTA BY

INCLUDES: OVERVIEW ON CLINICAL SITUATIONS FREQUENTLY ENCOUNTERED IN ORTHODONTIC TREATMENTS MECHANOTHERAPY USED TO RESOLVE THESE SITUATIONS

3M Incognito Appliance System extraction case study.

A comparison of traditional and computer-aided bracket placement methods

Comparative analysis of load/deflection ratios of conventional and heat-activated rectangular NiTi wires

Space Closure Biomechanics Applied Using The MBT System Technique

MBT System as the 3rd Generation Programmed and Preadjusted Appliance System (PPAS) by Masatada Koga, D.D.S., Ph.D

Lab Forms and Communications Precise Indirect Bonding Systems.

Table of Contents Section 6 Table of Contents

Angle Class II, division 2 malocclusion with deep overbite

The management of impacted

Orthodontic Treatment Using The Dental VTO And MBT System

Measurement of Forces and Moments in Three- Dimensional Archwires

essentialstm cutting your consumable costs brackets tubes wires springs instruments T:

Evaluation of methods of archwire ligation on frictional resistance

Uniformed Services University of the Health Sciences. Manuscript/Presentation Approval or Clearance

Downloaded from by guest on 15 November 2018

Characterizing constraining forces in the alignment phase of orthodontic treatment

Clarity. Form meets function: Discover the science behind a versatile self-ligating bracket system. Aesthetic Orthodontic Solutions

Premium Heat Activated Arch Wire

Aesthetic Brackets and Accessories

Nonsurgical Treatment of Adult Open Bite Using Edgewise Appliance Combined with High-Pull Headgear and Class III Elastics

Thakur H et al.applicability of various Mixed Dentition analysis among Sriganganagar School children

Clarity ADVANCED Ceramic Brackets. Advanced Ceramics. Enhanced Beauty.

System Orthodontic Treatment Program By Dr. Richard McLaughlin, Dr. John Bennett and Dr. Hugo Trevisi

A Comparison of Frictional Resistance between Active and Passive Self-ligating Brackets with Conventional Bracket Systems

The Loading And Unloading Properties Of Various Arch Wires As A Function Of Cross Sectional Dimension And Inter Bracket Span Width System

ORTHOdontics SLIDING MECHANICS

An Evaluation of Efficiency and Effectiveness of Self-ligating Bracket Systems: A Prospective Clinical Study

Influence of Attachments and Interproximal Reduction on the Accuracy of Canine Rotation with Invisalign

A study of the frictional characteristics of four commercially available self-ligating bracket systems

Dual Force Cuspid Retractor

Tooth angulation and dental arch perimeter the effect of orthodontic bracket prescription

3D CONTROL HAS ARRIVED

GAC Archwire Collection

Efficient Bonding Protocol for the Insignia Custom Bracket System

Clarity. Esthetic Orthodontic Solutions. 3M Clarity Ultra Self-Ligating Brackets Practice confidently. Smile beautifully.

HYCON Device. Fast, precise and successful space closure made easy! 25 years - around the globe research - design - manufacturing - distribution

CBCT Assessment of Dental and Skeletal Changes Using the Damon versus Conventional (MBT) System

Compromised nonsurgical treatment of a patient with a severe Class III malocclusion

RESEARCH ARTICLE /jp-journals

Correction of a maxillary canine-first premolar transposition using mini-implant anchorage

Details on the new features, enhancements, and problem fixes of a software release June 3, 2016

Comparison of Frictional Resistance in Conventional Brackets with Different Stainless Steel Wires

Anchorage system. tomas / Sets Page 270 tomas / Pins Page 271 Instruments and accessories Page 273 Patient consultation material Page 282

An Anterior Tooth Size Comparison in Unilateral and Bilateral Congenitally Absent Maxillary Lateral Incisors

Class II Correction using Combined Twin Block and Fixed Orthodontic Appliances: A Case Report

Dental Materials Journal 2015; 34(3): Keisuke TOCHIGI, Souichiro ODA and Kazuhito ARAI INTRODUCTION

Developing lingual orthodontics

The effectiveness of laceback ligatures: A randomized controlled clinical trial

Molar anchorage loss has been shown to occur

Simply Great Wire! Great prices.

Unilateral Horizontally Impacted Maxillary Canine and First Premolar Treated with a Double Archwire Technique

Cephalometric Analysis

GREAT LAKES BRACKETS. Treat with Quality. Treat with Confidence.

Brackets

Table of Contents Section 1 Table of Contents

Enhanced Control in the Transverse Dimension using the Unitek MIA Quad Helix System by Dr. Sven G. Wiezorek

Thermal Nitanium. Archwires. Reduced Friction. Accelerated Results.

Archwire Insertion and Disengagement Instruments Technique Guide

TURN CLASS II INTO SIMPLE CLASS I PATIENTS.

The effect of buccal lingual slot dimension size on third-order torque response

The Science Behind The System Clinical Abstracts, Volume 2

Mollenhauer Aligning Auxiliary for Bodily Alignment of Blocked-out Lateral Incisors in Preadjusted Edgewise Appliance Therapy

Instability of tooth alignment and occlusal relationships

Correction of Class II Division 2 Malocclusion by Fixed Functional Class II Corrector Appliance: Case Report

Effect of passive self-ligating bracket placement on the posterior teeth on reduction of frictional force in sliding mechanics

Honing Damon System Mechanics for the Ultimate in Efficiency and Excellence Jeff Kozlowski, DDS East Lyme, CT

Transcription:

Original Article Orthodontic bracket slot dimensions as measured from entire bracket series Paul Brown a ; Warren Wagner b ; Hyden Choi c ABSTRACT Objective: To measure the slot dimensions of an entire series of metal orthodontic brackets. Materials and Methods: Ten bracket series approximating five complete sets of brackets each were imaged and measured. Descriptive statistics were generated. Results: Slot dimension varied significantly from series to series as well as within the series themselves. About one-third of the brackets would not accommodate a full-size wire, and 15% to 20% are 0.001 inches or larger than the nominal advertised size. Conclusion: The clinician is unlikely to have on hand complete sets (upper and lower 5-5) of ideal brackets and should both expect and be able to be accommodate tooth movement through wire bending in three planes of space to overcome any bracket deficiencies. (Angle Orthod. 2015;85:678 682.) KEY WORDS: Brackets; Mechanics INTRODUCTION Contemporary orthodontists have a variety of methods to move teeth. When traditional brackets are the treatment of choice, there are dozens of brands with an almost endless variety of wings, slots, ligation features, and first-, second-, and third-order prescriptions. They can be mixed and matched in either the 0.018-inch or 0.022-inch size, depending on preference. Despite the near universal use of brackets in orthodontic treatment, there are few scientific studies of the bracket slot s tolerances, which are especially important when using a preadjusted appliance. The brackets and tubes for an edgewise appliance must be precisely manufactured so that the internal slot dimensions are accurate to at least 1 mil (1 mil50.001 inches). 1 Bennett stated, Oversize slots undermine the whole basis of preadjusted edgewise, which is intended to minimize wire bending. 2 Bennett further stated that there should be a reasonable expectation that the appliance accuracy a Retired, Midway, Utah. b Associate Professor, Department of Restorative Dentistry, University of Detroit Mercy School of Dentistry, Detroit, Mich. c Private practice, Detroit, Mich. Corresponding author: Dr Paul Brown, c/o Dr Warren Wagner, Department of Restorative Dentistry, UDM School of Dentistry, 2700 Martin Luther King Jr Boulevard, Detroit, MI 48208-2576 Accepted: August 2014. Submitted: April 2014. Published Online: September 29, 2014 G 2015 by The EH Angle Education and Research Foundation, Inc. should be exactly as claimed in advertising brochures. While theoretical tooth positions or prescription values may be debated, few have questioned whether a specific bracket or bracket series is even capable of moving the tooth to the desired position. Dentists appear to be relying more on the advice of peers and the opinions of experts, as scientific data are considered valuable only 42% of the time when selecting dental products. 3 Past bracket studies that are frequently cited that deal with bracket slot dimension and shape were limited to either relatively small sample sizes and/or the inclusion of just a single tooth-specific bracket (ie, upper left central) from a bracket series. This may or may not be representative of all the other individual tooth brackets in the series. In a metal injection molded bracket series, for example, each specific tooth is manufactured from a separate and unique mold. Cash et al. 4 measured the slots of five upper left central 0.022-inch brackets from 11 bracket series representing six different manufacturers. They reported that all bracket systems were oversized with a range of 5% to 24%. Four systems had parallel slot walls, five systems converged from the base to the top of the slot, and two systems showed diverging slot walls. Kusy and Whitley 5 measured three brackets (unspecified as per tooth) from 24 unique bracket series from eight different companies. 2 Their sample included brackets manufactured from four different materials and included three different slot sizes 0.018, 0.0185, and 0.022 inches. They reported that 15% of the brackets were smaller than reported and slot sizes 678 DOI: 10.2319/042814-307.1

ORTHODONTIC BRACKET SLOT DIMENSIONS 679 Table 1. Bracket Series and Their Characteristics n Method of Manufacture Mean SD Min Max A Company 022 96 Cast 0.0221 0.0007 0.0208 0.0237 Ormco Damon Q 022 78 MIM 0.0229 0.0003 0.0221 0.0235 Unitek Victory 022 100 MIM 0.0229 0.0003 0.0213 0.0234 GAC In-Ovation 022 100 MIM 0.0224 0.0001 0.0221 0.0227 Opal Avex 022 100 Milled 0.0224 0.0003 0.0219 0.0231 SPEED 022 99 Milled 0.0222 0.0006 0.0211 0.0235 AO Mini Masters 022 100 MIM 0.0219 0.0006 0.0194 0.0232 AO Mini Masters 018 100 MIM 0.0184 0.0004 0.0175 0.0193 Opal Avex 018 100 Milled 0.0181 0.0003 0.0173 0.0186 Unitek Victory 018 100 MIM 0.0189 0.0003 0.0180 0.0196 exceeded the nominal value by as much as 16% and 8% (in the cases of nominal 0.018-inch and 0.022-inch slots, respectively). Finally, they underscored the need for the practitioner to know the exact dimension of the bracket in order to avoid compromising treatment mechanics. Bhalla et al. 6 measured five upper left central 0.022- inch self-ligating brackets from each of six different bracket series that represented four different manufacturers. They reported the brackets to be between 5% and 15% larger than the nominal values and that slot walls diverged from the base to the top of all the brackets. It was noted that brackets from the same manufacturer may vary in size as well. Major et al. 7 measured 30 upper right central 0.022- inch stainless steel self-ligating brackets from three manufacturers. They reported slot heights from 2% smaller to 3% over the nominal size. Parallel, divergent, and convergent walls were reported. The authors cautioned that one cannot assume a consistent slot shape. It is clear from these studies that there is considerable variation in the slot tolerances when measuring a single bracket from a bracket series. The purpose of this study is to measure the slot dimensions of entire series of metal orthodontic brackets and provide descriptive statistics of their measurements. This should give the clinician a much better appreciation of the capabilities and limitations of various bracket systems than a sample consisting of only a single bracket from the series. METHODS AND MATERIALS In this study, the following bracket series were evaluated: A-Company 0.022 inch (manufactured circa 1985), Ormco Damon Q 0.022 inch (Ormco Corp, Orange, Calif), Unitek Victory 0.022 inch and 0.018 inch (Unitek, Monrovia, Calif), GAC In-Ovation 0.022 inch (GAC, Bohemia, NY), Opal Avex 0.022 inch and 0.018 inch (Opal Orthodontics by Ultradent, South Jordan, Utah), SPEED 0.022 inch (Strite Industries, Cambridge, Ontario, Canada), and American Orthodontics Masters 0.022 inch and 0.018 inch (American Orthodontics, Sheboygan, Wis). The bracket series consisted of 100 brackets each, representing five cases of upper and lower second premolar to second premolar, except for Speed (n 5 99), A-Company (n 5 96), and Damon (n 5 78) due to only partial sets of lower incisors and second premolars (Table 1). The methods used to manufacture the various brackets were obtained verbally through company representatives or from published sales literature. A Clark Microhardness Tester-Model CM-700 and a Clark Instrument Automatic Reading System-Model CM-AR90 were used to image, scale and, measure each bracket. A Knoop/Vickers microhardness tester has two basic components: (1) a device that (in a very controlled fashion) forms small indents on the surface of specimens and (2) a very accurate measuring microscope. For this study, only the measuring microscope capability of the microhardness tester was used. The microhardness tester that was used has multiple objectives providing total magnifications between 503 and 5003 and filars (measuring lines) that can be set to provide greater than 1-mm resolution. The brackets were placed on a microscope slide using rope wax with the slots oriented vertically so that the line of view with the measuring microscope was parallel to the slot axis (Figure 1). The slide was then placed on the microscope table and the slot and light adjusted until a sharp, well-focused image was viewed on the screen and digitally captured (Figure 2). Figure 3 shows two screenshots from the Automatic Reading System showing a bracket with rounded internal line angles and another with a divergent slot. The digitally imaged bracket is automatically scaled and loaded into the measuring software. To compensate for both rounded and defective corners as well as excess material, the initial vertical filar was placed at the deepest portion of the bracket base approximating the middle of the slot. The second vertical filar was then moved and set at a distance of 100 mm down the slot. At 100 mm down the slot, the horizontal filars were moved until the first contact was made along this distance and the bracket wall. The vertical slot

680 BROWN, WAGNER, CHOI Figure 1. Orienting the bracket onto the microscope slide with rope wax. dimension was then recorded. Observers were able to measure the brackets with a correlation coefficient of.947. RESULTS Table 1 shows the bracket s characteristics and descriptive statistics of the measurements, while Figure 4 shows a graphic representation of the vertical slot dimensions. For 0.018-inch slots, the Opal Avex, a milled bracket, was on average closest to the accepted nominal value, being 0.0001 inch oversized. It also had the smallest range and similar standard deviation as the Unitek Victory. For 0.022-inch slots, the closest average sizes were American Orthodontics Mini Masters, a metal injection molded bracket, that was 0.0001 inch undersized, and Figure 3. Two screenshots showing the vertical filars set at the base of the bracket and 100 mm apart down the slot. The horizontal filars are set perpendicular to the second vertical filar where it intercepts the bracket walls. The top screenshot depicts a bracket with rounded internal line angles and the lower screenshot a bracket with a divergent slot. the A-Company bracket, a cast bracket, that was 0.0001 inch oversized from the accepted nominal value. The GAC In-Ovation, while oversized, had the smallest standard deviation and range of any bracket in the sample, while the A-Company had the largest standard deviation and the American Orthodontics Mini Masters the largest range of any bracket in the sample. DISCUSSION Figure 2. Digital view of the bracket slot properly oriented and focused on the microscope table. This sample of about five cases per bracket series indicates that torque issues as a function of vertical slot dimension remain problematic. Compounding the problem are the internal line angles, the orientation of the walls relative to the base, and presence of various size and topography of asperities.

ORTHODONTIC BRACKET SLOT DIMENSIONS 681 Figure 6. A right and left central incisor bracket from the same series. Not only would the same full-size wire not fit into the slot, a first-order discrepancy would also be introduced, due to a rounded lower corner on the right bracket. Figure 4. Graphic representation of the vertical slot dimensions. Alexander 8 has put the clinical torque problem in a simple perspective. It should be understood that for every 0.001 inch of freedom between the archwire and vertical bracket slot, approximately 5 degrees of effective torque is lost. This would obviously assume 90u interior line angles, parallel and smooth bracket walls, and accurate and geometrically correct arch wire sizes. This implies that a 0.019-inch thickness archwire in a 0.022-inch straight wire bracket with a typical 12u of torque in the prescription would produce no effective torque. Figure 5 is a cross section of a 0.019-3 0.025- inch archwire, scaled and superimposed into a 0.022- inch bracket from the sample. It should be emphasized that these are nominal values from the manufacturer. Figure 5. Depiction of a cross section of a polished 0.019-3 0.025- inch archwire that was imaged and scaled into the 0.022-inch slot of a bracket from our samples showing 620u of slot play. Similar examples could be found in 0.018-inch slots as well. Considering all 0.022-inch brackets that were measured in this study and assuming the walls were completely parallel and smooth, 36% of the slots would not be large enough to insert a full-size wire and 15% would be larger than 0.023 inches. For the 0.018-inch sample, 31% would not accommodate a full-size wire, and 20% would be larger than 0.019 inches. Brackets with single-digit torque values or singledigit differences in their respective prescriptions likely offer little if any advantage over a standard edgewise bracket or different bracket prescription, respectively. A recent study determined that the ability of clinicians to distinguish the posttreatment effects of brackets with 12u vs 17u of upper incisor torque was essentially no better than chance. 9 A direct comparison of manufacturing techniques is difficult as the actual processes and feed stocks remain proprietary. All bracket series contained some brackets manufactured at the nominally advertised values. In general, this study confirms the findings of the other cited studies that measured brackets of the same series and manufacturer. An examination of the images in our sample would also confirm a general slot shape (parallel walls, converging walls, or diverging walls) as well as quality of the bracket base to wall line angles as reported, with exceptions as noted. Cash et al. 4 reported the 0.022-inch slot Unitek Victory and Ormco Damon series to both be oversized (4.6% or 0.0221 inches and 16.86% or 0.0224 inches, respectively), with the Victory bracket diverging toward the top of the slot and the Damon bracket converging. 4 Both the Unitek Victory and Ormco Damon brackets in the current sample were oversized and with the same general slot shapes. Cash et al. 4 cautioned clinicians that the preadjusted bracket and wire systems may not produce the three-dimensional control required to produce an acceptable result. The 0.018-inch Unitek Victory was reported by Kusy and Whitley 5 to be oversized at 0.0183 inches (n 5 3) vs 0.0189 inches in the current study. 5 Both the In-Ovation and Speed brackets were considerably oversized, as reported by Bhalla et al. 6

682 BROWN, WAGNER, CHOI Figure 7. An ideal set of anterior straight wire brackets. (0.02385 and 0.02311 inches, respectively), with the In-Ovation bracket being the most consistent. In the current sample, the brackets were closer to nominal means. This study confirmed the speculation by Bhalla et al. 6 that brackets from the same manufacturer may vary (Figure 6). Major et al. 7 reported the Speed brackets to be slightly undersized at the base 0.556 mm (0.0219 inches) with pronounced rounding and large fillets where the base meets the wall and slight convergence. Such bases-to-wall issues were also present throughout this sample. For Major et al., the GAC In-Ovation measured 0.564 mm (0.0222 inches) at the base and most closely resembles a trapezoidal shape. The Damon Q s base was reported at 0.572 mm (0.0225 inches), with the highest measured manufacturing tolerance among the three tested and nearly 90-degree corners. While there are examples of Damon Q brackets in the current sample with 90u corners, the great majority have large rounded internal line angles. Manufacturing defects occurred both as a single anomaly as well as throughout the entire tooth-specific set of brackets in a series. While the correct bracket position is a given, continuously rebonding a straight wire bracket for more optimal tooth position may or may not achieve the desired result. In some cases, one would simply be rebonding a similarly defective bracket, and in others, the tooth position could potentially become even less optimal. Figure 7 shows one ideal set, although slightly oversized, of central, lateral, and canine brackets that have square corners, parallel walls, and are free from asperities. Most orthodontists prefer a particular bracket system. It is important to know the system used in treatment and why it was chosen. 10 The clinician is unlikely to have on hand complete sets (upper and lower 5-5) of ideal brackets that would consistently work as claimed in the brochure. It is necessary to avoid the mind-set that all brackets [bracket prescriptions, bracket series, and specific tooth brackets] are more-or-less the same. 11 One should both expect and be able to be accommodate tooth movement through wire bending in three planes of space to overcome any bracket deficiencies. CONCLUSIONS N The actual slot size and shape of an orthodontic bracket are likely to vary both larger and smaller from the advertised nominal value within a bracket series. N Using conventional wire sizes and a straight wire approach, some brackets marketed as preadjusted clearly are not able to produce a torqueing tooth movement without additional wire bending. N Manufacturing anomalies may occur in a single bracket, throughout the sets of specific tooth brackets, or generally throughout an entire bracket series. REFERENCES 1. Proffit W. Contemporary Orthodontics. 4th ed. St. Louis, Mo:Mosby Elsevier; 2007. 2. Bennett J. Fundamentals of Orthodontic Treatment Mechanics. 1st ed. London, UK:LeGrande Publishing; 2014. 3. King G. Biomedicine in orthodontics: from tooth movement to facial growth. Orthod Craniofac Res. 2009;12:53 58. 4. Cash A, Good S, Curtis R, McDonald F. An evaluation of slot size in orthodontic brackets are standards as expected? Angle Orthod. 2004;74:450 453. 5. Kusy R, Whitley J. Assessment of second-order clearances between orthodontic archwires and bracket slots via critical contact angle for binding. Angle Orthod. 1999;69:71 80. 6. Bhalla N, Good S, McDonald F, Sherriff M, Cash A. Assessment of slot sizes in self-ligating brackets using electron microscopy. Aust Orthod J. 2010;26:38 41. 7. Major T, Carey J, Nobes D, Major P. Orthodontic bracket manufacturing tolerances and dimensional differences between select self-ligating brackets. JDentBiomech. 2010:1 6. 8. Alexander R. The 20 Principles of the Alexander Discipline. Chicago, Ill:Quintessence Publishing; 2008. 9. Moesi B, Dyer F, Benson P. Roth versus MBT: does bracket prescription have an effect on the subjective outcome of preadjusted edgewise treatment? Eur J Orthod. 2013;35: 236 243. 10. English JD, Briss BS, Jamieson SA, et al. Common errors in preparing for and completing the American Board of Orthodontics clinical examination. Am J Orthod Dentofacial Orthop. 2011;139:136 137. 11. Bennett J. Fundamentals of Orthodontic Bracket Selection. 2nd ed. London, UK:LeGrande Publishing; 2010.