GORE EXCLUDER AAA Endoprosthesis demonstrates long-term durability. Michel Reijnen Rijnstate Hospital Arnhem, The Netherlands

Similar documents
Treatment options of late failures of EVAS. Michel Reijnen Rijnstate Arnhem The Netherlands

Durable outcomes. Proven performance.

Bilateral use of the Gore IBE device for bilateral CIA aneurysms and a first interim analysis of the prospective Iceberg registry

The Auckland Experience with the Nellix EVAS System. Andrew Holden, MBChB, FRANZCR

ENCORE, a Study to Investigate the Durability of Polymer EVAR with Ovation A Contemporary Review of 1296 Patients

My personal experience with INCRAFT in standard and challenging cases

Type 1a Endoleak in hostile neck anatomies: Endoanchor can fix it! D. Böckler University Hospital Heidelberg, Germany

Young-Guk Ko, M.D. Severance Cardiovascular Hospital, Yonsei University Health System,

A New EVAR Device for Infrarenal AAAs

Current Status of EVAR for Infrarenal AAA. 31 st Annual Florida Vascular Society. PENN Surgery

Durability of The Endurant Stent-Graft through 5 Years

Considerations for a Durable Repair

Feasibility of aortic neck anatomy for endovascular aneurysm repair in Korean patients with abdominal aortic aneurysm

Influence of patient selection and IFU compliance on outcomes following EVAS

Disclosures. EVAR follow-up: actual recommendation. EVAR follow-up: critical issues

Treating very short necks ( 4mm <10mm) using the Endurant stent graft + EndoAnchors: 1-year results and current insights

Standardization of the CHEVAR procedure: How a standard approach has improved outcomes. Prof Peter Holt St George s, London

Mid-term results from ANCHOR: How does this data influence the treatment algorithm for hostile EVAR anatomies

Insights from the PROTAGORAS/PERICLES Registries: impact on ChEVAR results

Aortic Neck Issues Associated Clinical Sequelae/Implications for Graft Choice

Chimney endovascular aneurysm sealing (ch-evas) for ruptured abdominal aortic aneurysms (AAA) due to type Ia endoleak following failed EVAS

Access More Patients. Customize Each Seal.

Predictors of abdominal aortic aneurysm sac enlargement after EVAR Longterm results from the ENGAGE Registry

Low profile TEVAR: is it an added value? Michel Bosiers, G. Torsello Münster

Why Nellix? Treating Concomitant Common Iliac Aneurysms

Lessons learned from Ch-EVAR for the treatment of. Miltos Matsagkas MD, PhD, FEBVS Professor of Vascular Surgery University of Thessaly

How to select FEVAR versus EVAR + endoanchors in short-necked AAAs

Conflicts of Interest. When and Why Complex EVAR in Tx of juxta/suprarenal AAA? Summary. Infrarenal EVAR for short necks 2y postop

Hostile Proximal Neck: A New Conformable EVAR Device

EVAS using Nellix in my practice Where are we today?

Anatomy-Driven Endograft Selection for Abdominal Aortic Aneurysm Repair S. Jay Mathews, MD, MS, FACC

LOWERING THE PROFILE RAISING THE BAR

COMBINED TECHNIQUE CHIMNEY + FENESTRATED ENDOGRAFT FOR COMPLEX ANEURYSMS ERIC DUCASSE - MD PHD FEVBS CHU BORDEAUX

Mid-term results of 300+ patients treated by endovascular aortic sealing (EVAS)

The Ventana Off-the-Shelf Graft for Pararenal AAA. Andrew Holden Associate Professor of Radiology Auckland Hospital

EVAR Revision Setting - How can Heli-FX EndoAnchors improve the outcomes?

EVAR replaced standard repair in most cases. Why?

Ovation. Sean Lyden, MD Department Chair, Vascular Surgery Cleveland Clinic

Clinical trial and real-world outcomes of an endovascular iliac aneurysm repair with the GORE Iliac Branch Endoprosthesis (IBE)

FEVAR FIFTEEN YEARS OF EFFICIENCY E.DUCASSE MD PHD FEBVS CHU DE BORDEAUX

When to use standard EVAR with EndoAnchors or CHEVAR in short-neck AAAs LINC ASIA 18

14F OD Ovation Abdominal Stent Graft System

EndoVascular Aneurysm Sealing (EVAS) with Nellix

Talent Abdominal Stent Graft

Anatomical challenges in EVAR

Is EVAS a proper choice in women?

Evolution of gender-related differences in outcome of EVAR

TriVascular Ovation Prime Abdominal Stent Graft System

AAA: DEBATE THERE ARE NO LIMITS USING EVAR FOR AAA. 2 nd -3 rd June 2016.

Challenges with Complex Anatomies Advancing Care in Endovascular Aortic Treatment

Improving Endograft Durability with EndoAnchors

EXPERIENCE AFTER 500 ENDOLUMINAL STENT GRAFTS. DEVICES SPECIFIC OUTCOME AND LESSONS LEARNED.

FROM THE EVERYDAY TO THE EXTRAORDINARY

Overcoming Challenging Aortic Anatomy: from Hostile Necks to Tortuous Iliac arteries

Analysis of Type IIIb Endoleaks Encountered with Endologix Endografts

INCRAFT system: Update from the Pivotal INSPIRATION Study

Optimizing Accuracy of Aortic Stent Grafts in Short Necks

When and where EVAR patients should be discharged?

Patient selection in Hostile Necks and how. to prevent endoleaks a word of caution

Challenging anatomies demand versatility.

Management of Endoleaks

What's on the Horizon for AAA: Unilateral & Percutaneous, "UP-EVAR" System Zoran Rancic M.D., Ph.D.

Hostile Neck During EVAR, The Role Of Endoanchores

History of the Powerlink System Design and Clinical Results. Edward B. Diethrich Arizona Heart Hospital Phoenix, AZ

Nellix Endovascular System: Clinical Outcomes and Device Overview

GORE EXCLUDER AAA Endoprosthesis ANNUAL CLINICAL UPDATE OCTOBER Section I Clinical experience. Section II Worldwide commercial experience

Taming The Aorta. David Minion, MD Program Director, Vascular Surgery University of Kentucky Medical Center Lexington, Kentucky, USA

Mid-term outcome of CERAB for aorto-iliac occlusive disease. Michel Reijnen Rijnstate Arnhem The Netherlands

How effective is preservation when viewed through a clinical and economic lens?

How to Categorize the Infrarenal Neck Properly? I Van Herzeele Dept. Thoracic and Vascular Surgery, Ghent University, Belgium

DIFFICULT ACCESS REMAINS A CONTRAINDICATION FOR EVAR APOSTOLOS K. TASSIOPOULOS, MD, FACS PROFESSOR AND CHIEF DIVISION OF VASCULAR SURGERY

CAUTION: The Nellix EndoVascular Aneurysm Sealing System is an investigational device. Limited by federal (or United States) law to investigational

Abdominal and thoracic aneurysm repair

Bifurcated system Proximal suprarenal stent Modular (aortic main body and two iliac legs) Full thickness woven polyester graft material Fully

Degeneration of the Neck Post Implementation - a New Era of AAA Stent

Endovascular aneurysm repair (EVAR) for the

Endovascular Repair o Abdominal. Aortic Aneurysms. Cesar E. Mendoza, M.D. Jackson Memorial Hospital Miami, Florida

EVAS is Associated with Lower All-Cause Mortality

THE ENDURANT STENT GRAFT IN HOSTILE ANEURYSM NECK ANATOMY

THE THE MORE MORE NATURAL APPROACH TO OPTIMAL FIT

Preliminary Data from the Gore EXCLUDER Thoracoabdominal Branch Endoprosthesis Early Feasibility Study

Endoanchor-assisted TEVAR

The Distal Seal Zone in AAA Repair A facet of EVAR that is not to be overlooked.

ChEVAR Vs. fevar for juxtarenal Aneurysm. E.Ducasse MD PhD FEVBS Unit of vascular surgery CHU bordeaux

Important Update to Field Safety Notice Nellix EndoVascular Aneurysm Sealing System Updated Instructions for Use (IFU)

CUSTOM-MADE SCALLOPED THORACIC ENDOGRAFTS IN DIFFERENT HOSTILE AORTIC ANATOMIES

Technique and Tips for Complicated AAA Cases with Stent Graft

No Neck Is Needed! Treat the Aneurysm Instead! Andrew Holden, MD Associate Professor of Radiology Auckland City Hospital

Challenges in the treatment of infrarenal aneurysm & considerations for a durable repair

Chungbuk Regional Cardiovascular Center, Division of Cardiology, Departments of Internal Medicine, Chungbuk National University Hospital Sangmin Kim

Use of Aptus Heli-FX EndoAnchor implants with standard endografts to strengthen seal in hostile anatomies:

All-In-One. Iliac Branch System. PER F ORM ANC E b y d e s i g n ILIAC BRANCH ENDOPROSTHESIS

Case Report Early and Late Endograft Limb Proximal Migration with Resulting Type 1b Endoleak following an EVAR for Ruptured AAA

Eight Year Experience with Type I Endoleaks at a Tertiary Care Center

Robert F. Cuff, MD FACS SHMG Vascular Surgery

symptomatic aneurysms or aneurysms that grow >1cm/yr

The evolution. AORFIX AAA Stent Graft now with the new AORFLEX Delivery System

The Fate of Unexpected Events Occurring in Standard EVAR

Right Choice for Right Angles

Reduction in cardiovascular related adverse events following active sac management with Nellix vs. EVAR: Are there biological advantages?

Transcription:

GORE EXCLUDER AAA Endoprosthesis demonstrates long-term durability Michel Reijnen Rijnstate Hospital Arnhem, The Netherlands

Disclosure Speaker name: Michel Reijnen I have the following potential conflicts of interest to report: Consulting Employment in industry Stockholder of a healthcare company Owner of a healthcare company Other(s) I do not have any potential conflict of interest

Long term outcome of EVAR Long-term data after EVAR are still limited Lack of device-specific reporting Bias of ongoing introduction of improvements and new devices EVAR-1 cohort with mean FU of 12,7± 1,5 years; beyond 8 years significantly higher mortality in EVAR group Increased AAA-related mortality mainly attributable to secondary rupture (7% vs 1%) Patel R, et al. Lancet 2016; 388: 2366 74

Long term outcome of EVAR Influence of Graft Type EVAR 1 and EVAR 2 Trial For the prespecified adjustment variables, there was some evidence to suggest that patients who had a GORE EXCLUDER Device experienced significantly lower rates of complications and reinterventions compared with the other graft types. 1 1 Brown LC, et al. British Journal of Surgery 2010;97(8):1207-1217

Gore Excluder Endoprosthesis 1997 Introduced in Europe 2004 Introduction of the low-permeability expanded PTFE to reduce endotension 2010 Introduction of the C3 delivery system Repositionable for - Precise proximal deployment - Cannulation of CL limb 2012 The profile is reduced while remaining the same devices; no compromise on durability Future Introduction of the conformable Excluder

Gore Excluder Endoprosthesis A low-permeability film layer, added in 2004, is the only modification to the GORE EXCLUDER AAA Device. Requiring so few changes is unprecedented compared to currently available endografts.

Gore concept and conformability eptfe & Nitinol Biocompatible, highly durable, long term performance MRI compatible Sutureless Construction Stent is heat bonded to graft Optimized for stent nesting Stent design Sinusoidal pattern Provides nitinol stent support with fewer unsupported gaps, better wall apposition, and limiting kinking and occlusion

Long term outcome Excluder 2 Dutch sites January 2000 and December 2015 248 patients treated with Gore Excluder out of 1643 EVAR procedures 87% male with mean age of 71.2 ± 8.2 years N=61 (25%) outside the IFU Anatomical characteristics Aneurysm diameter Infrarenal neck diameter Infrarenal neck length 59.1 ± 9.6 mm 22.9 ± 2.4 mm 32.3 ± 11.9 mm Neck angulation 23.6 ± 21.9 Concomitant iliac artery aneurysm n=30 (12%) Poublon C, J Vasc Intervent Radiol, in press

Long term outcome Excluder Median follow-up 26 months (range 1-190 months) 5 year n=63 (25.4%) 10 years n=20 (8.1%) Estimated re-intervention-free survival: 5 years; 85.2% 10 years 75.6% Independent risk factors for re-intervention: Technical success (P <.001) Endoleak type I (P <.001) Endoleak type II (P =.003) Poublon C, J Vasc Intervent Radiol, in press

Long term outcome Excluder Freedom from endoleak Reintervention for endoleak in 16 patients (6.5%) Type Ia n=6 Type Ib n=5 Type II n=5 Type Ia (95% CI) Type Ib (95% CI) Type II (95% CI) Type V (95% CI) 1 year 96.2% (93.8-98.6) 98.7% (97.3-100.1) 66.5% (60.4-72.6) 99.5% (98.5-100.5) 5 years 95.3% (92.4-98.2) 95.1% (91.2-99.0) 60.7% (53.8-67.6) 90.6% (84.9-96.3) 10 years 91.5% (83.7-99.3) 95.1% (91.2-99.0) 55.0% (45.0-65.0) 80.3% (69.1-91.5) Poublon C, J Vasc Intervent Radiol, in press

Long term outcome Excluder Re-interventions Aneurysm rupture 0.4% Limb occlusion 0.4% Stent migration 0.4% Aneurysm sac growth was more prevalent in: Original Excluder compared to the Low Permeability Excluder (P =.001) Presence of an endoleak type I (P =.002), II (P =.005) and V (P <.001) The overall survival at 5 and 10 years after EVAR was 68.4% and 49.0%, without reported aneurysm related deaths In two patients suspect to be AAA-related Poublon C, J Vasc Intervent Radiol, in press

Long term outcome Excluder Long term outcome Gore Excluder Poublon (n=248) Maleux (n=121) Bastos Gonçalves (n=144) Anatomy Aneurysm diameter 59.1 ± 9.6 mm 58.1 ± 11.8 60 (54-70) Infrarenal neck diameter 22.9 ± 2.4 mm 21.6 ± 2.2 24 (22.2-25) Infrarenal neck length 32.3 ± 11.9 mm 32.9 ± 13.3 28.5 (21-42) Follow-up (mean) 2.2 (0-15.8) 5.0 ± 3.1 5.0 (3.1-6.4) Survival 5-year 68% 75% 62% 10-year 49% 58% 39% Re-interventions 12% 15% 23% Intervention free survival 5-year 85% 90% NA 10-year 76% 78% NA Endoleak Ia 5% 2% 5% Ib 3% 2% 3% II 36% 32% 23% III 0% 0% 0% Endotension 7% 9% 11% Limb occlusion 0.4% 2% 1% Bastos Gonçalves et al. J Vasc Surg 2012;56:920-8 Maleux et al. Cardiovasc Intervent Radiol 2012;35(3):498-507 Poublon et al. J Vasc Intervent Radiol, in press

Long term outcome Excluder Re-interventions Long term outcome Gore Excluder Poublon (n=248) Maleux (n=121) Bastos Gonçalves (n=144) Overall (n=513) Number of reinterventions N=34 N=18 N=39 18% Indication for reintervention Insufficient proximal seal N=7 N=2 N=10 4% Insufficient distal seal N=7 N=2 N=9 4% Type II endoleak N=9 N=4 N=8 4% Infolding of the proximal graft N=1 N=2 N=0 0.6% Limb kinking N=0 N=1 N=0 0.2% Limb occlusion N=1 N=1 N=1 0.6% Endotension N=3 N=5 N=8 3% Conversion N=1 N=4 relining N=3 N=4 N=2 fenestration N=0 N=2 Endograft Infection N=0 N=1 N=1 0.4% Bastos Gonçalves et al. J Vasc Surg 2012;56:920-8 Maleux et al. Cardiovasc Intervent Radiol 2012;35(3):498-507 Poublon et al. J Vasc Intervent Radiol, in press

Endotension Maleux et al.: In all of the patients with endotension an original permeability stentgraft was implanted Bastos Gonçalves et al: The generation of implanted endoprosthesis (OD vs. LP) was significantly associated with the risk of sac growth Poublon et al.: Sac growth was more prevalent with the Original Gore Excluder compared to the Low Permeability Excluder (P=.001) Bastos Gonçalves et al. J Vasc Surg 2012;56:920-8 Maleux et al. Cardiovasc Intervent Radiol 2012;35(3):498-507 Poublon et al. J Vasc Intervent Radiol, in press

Maleux et al.: Mortality the majority died from cancer (19/35) No AAA-related mortality Bastos Gonçalves et al: The estimates for AAA-related death or rupture (including 30 days) were 2.4% at 5 years and 2.4% at 10 years One death related to raaa in conjunction with infection Poublon et al.: Malignancy most frequent cause of death (28%) No reported AAA-related mortality although two deaths were suspected to be aneurysm related Bastos Gonçalves et al. J Vasc Surg 2012;56:920-8 Maleux et al. Cardiovasc Intervent Radiol 2012;35(3):498-507 Poublon et al. J Vasc Intervent Radiol, in press

Gore Excluder and conformability Excluder is related to very low incidence of Limb occlusions 0-2% Kinking 0-1% Distal migration 0-4% Relation between aorto-iliac morphology and complications has been established with more graft related complications in; Higher infrarenal neck angulation Higher common iliac artery tortuosity

Gore Excluder and conformability Anatomical fit between patient and stent graft may be an important factor for success of treatment EVAR devices do reduce the aorto-iliac tortuosity; Conformability; Zenith < Endurant < Excluder Conformability ; Cook IBD < Gore IBE Coulston J et al. J Vasc Surg 2014 Taudorf M, et al. Eur J Vasc Endovasc Surg 2014 Wyss TR, et al. J Vasc Surg 2011 Della Schiava, et al. Ann Vasc Surg 2016

C3 delivery system Option for re-positioning August 2010 and December 2012 400 patients (86.8% male, mean age 73.9 ± 7.8 years) Technical success 99% Graft repositioning in 192/399 (48%) patients, most frequently for level readjustment with regard to the renal arteries Mean follow-up 15.9 ± 8.8 months Estimated freedom-from-reintervention 1 year 95% (95% CI 92-97%) 2 years 92% (95% CI 87-95%) Verhoeven EL, et al. Eur J Vasc Endovasc Surg. 2014 Aug;48(2):131-7

Conclusions The Gore Excluder has demonstrated its durability at the longterm with acceptable re-intervention rates and low AAArelated adverse events Late new-onset endoleaks are observed and emphasize the need for long-term surveillance The high conformability of the Gore Excluder may attribute to its low limb occlusion rates The introduction of the C3 delivery system could further reduce the re-intervention rates