The magic angle phenomenon in tendons: effect of varying the MR echo time

Similar documents
Magic angle artifact in MRI of the patellar ligament: preliminary comparison between conventional and weightbearing

Sensitivity and Specificity in Detection of Labral Tears with 3.0-T MRI of the Shoulder

Modified Oblique Sagittal Magnetic Resonance Imaging of Rotator Cuff Tears: Comparison with Standard Oblique Sagittal Images

RECENT ADVANCES IN CLINICAL MR OF ARTICULAR CARTILAGE

FieldStrength. Achieva 3.0T enables cutting-edge applications, best-in-class MSK images

Tissue-engineered medical products Evaluation of anisotropic structure of articular cartilage using DT (Diffusion Tensor)-MR Imaging

Knee Articular Cartilage in an Asymptomatic Population : Comparison of T1rho and T2 Mapping

Meniscus T2 Relaxation Time at Various Stages of Knee Joint Degeneration

Musculoskeletal MR Protocols

Usefulness of Unenhanced MRI and MR Arthrography of the Shoulder in Detection of Unstable Labral Tears

MR imaging of the knee in marathon runners before and after competition

Intramuscular Rotator Cuff Cysts: Association with Tendon Tears on MRI and Arthroscopy

Comparative Study of MRI and Operative Findings in Rotator Cuff Tears. University of Medical Sciences, Iran. Medical Sciences, Iran.

Musculoskeletal Imaging at 3T with Simultaneous Use of Multipurpose Loop Coils

Meniscal tears on 3T MR: Patterns, pearls and pitfalls

of the lumbar facet joints there

Why Talk About Technique? MRI of the Knee:

KNEE ALIGNMENT SYSTEM (KAS) MRI Protocol

Original Report. Sonography of Tears of the Distal Biceps Tendon. Theodore T. Miller 1,2 Ronald S. Adler 3

Articular cartilage and labral lesions of the glenohumeral joint: diagnostic performance of 3D water-excitation true FISP MR arthrography

A Cause of Increased Signal in the Normal Lateral Meniscus on Short-TE MR Images of the Knee

Prevalence of Meniscal Radial Tears of the Knee Revealed by MRI After Surgery

MR Advance Techniques. Vascular Imaging. Class II

The Magic Angle Effect: A Source of Artifact, Determinant of Image Contrast, and Technique for Imaging

Evaluation of Chondromalacia of the Patella with Axial Inversion Recovery Fast Spin-Echo Imaging

醫用磁振學 MRM 肌肉骨骼磁振造影簡介 肌肉骨骼磁振造影. 本週課程內容 General Technical Considerations 肌肉骨骼磁振造影簡介 盧家鋒助理教授國立陽明大學生物醫學影像暨放射科學系

The Low Sensitivity of Fluid-Attenuated Inversion-Recovery MR in the Detection of Multiple Sclerosis of the Spinal Cord

In vivo diffusion tensor imaging (DTI) of articular cartilage as a biomarker for osteoarthritis

Magnetic Resonance Angiography

MRI of Bucket-Handle Te a rs of the Meniscus of the Knee 1

The mandibular condyle fracture is a common mandibular

Added value of MR myelography using 3D COSMIC sequence in the diagnosis of lumbar canal stenosis: comparison with routine MR imaging

Arthroscopy / MRI Correlation Conference. Department of Radiology, Section of MSK Imaging Department of Orthopedic Surgery 7/19/16

This presentation is the intellectual property of the author. Contact them for permission to reprint and/or distribute.

Radiological Reasoning: Acutely Painful Swollen Finger. Musculoskeletal Imaging Chew and Richardson Benign-Appearing Bone Mass.

This presentation is the intellectual property of the author. Contact them at for permission to reprint and/or distribute.

Contrast between Scar and Recurrent Herniated Disk on Contrast-Enhanced MR Images

How Much Tesla Is Too Much?

Magic Angle Effects in MR Neurography

High-Sensitivity Coil Array for Head and Neck Imaging: Technical Note

ACR MRI Accreditation: Medical Physicist Role in the Application Process

Demystifying ABER (ABduction and External Rotation) sequence in shoulder MR arthrography

Post-injury painful and locked knee

Comparative study of imaging at 3.0 T versus 1.5 T of the knee

HSC Physics. Module 9.6. Medical Physics

Meniscal Tears with Fragments Displaced: What you need to know.

Frequency of Baker s Cysts in Cases of Meniscal Injury

9/18/18. Welcome- MSK Ultrasound Workshop. Introduction to Musculoskeletal Ultrasound. Acknowledgement of Country. The Workshop.

Original Research Article

Case Report: Knee MR Imaging of Haemarthrosis in a Case of Haemophilia A

Message of the Month for GPs June 2013

Radiologic Imaging Magnetic Resonance Imaging (MRI)

JMSCR Vol 05 Issue 01 Page January

Magnetic resonance imaging of femoral head development in roentgenographically normal patients

FILED: QUEENS COUNTY CLERK 12/15/ :11 PM INDEX NO /2015 NYSCEF DOC. NO. 54 RECEIVED NYSCEF: 12/15/2017

ACR MRI Accreditation Program. ACR MRI Accreditation Program Update. Educational Objectives. ACR accreditation. History. New Modular Program

Sonographic Findings of Adductor Insertion Avulsion Syndrome With Magnetic Resonance Imaging Correlation

CT ARTHROGRAPHY It s not Always About the

Lipoma Arborescens of Subacromial-subdeltoid Bursa: Ultrasonographic Findings

IMAGING TECHNIQUES CHAPTER 4. Imaging techniques

Table 2. Mean Scores of Articular Cartilage by Qualitative E v a l u a t i o n

High Field MR of the Spine

Cover Page. The handle holds various files of this Leiden University dissertation.

The Shoulder. Systematically scanning the shoulder provides extremely useful diagnostic information. The Shoulder

CLINICAL PRESENTATION AND RADIOLOGY QUIZ QUESTION

Musculoskeletal Ultrasound. Technical Guidelines SHOULDER

Value of Lipid- and Water- Suppression MR Images in Distinguishing Between Blood and Lipid Within Ovarian Masses

Meniscal Tears: Role of Axial MRI Alone and in Combination with Other Imaging Planes

Distribution of MR-detected cartilage defects of the patellofemoral joint in chronic knee pain

E. ÇAĞLAR, G. ŞAH N 1, T. OĞUR, E. AKTAŞ. Introduction. Abstract. OBJECTIVE: To identify changes in

Rotator Cable: MRI Study of Its Appearance in the Intact Rotator Cuff With Anatomic and Histologic Correlation

Oak foundation for donating the 3T Siemens Verio scanner. Board of directors BBH and Frh Hospitals for supporting the

Cervical Spondylosis: Three-dimensional Gradient-Echo MR with Magnetization Transfer

Clinics in diagnostic imaging (177)

Ultrasound assessment of most frequent shoulder disorders

Orthopedic Hardware Imaging Part II: MRI v. Metal

Ultra-High Field MR Diffusion Tensor Imaging Characterization of Rabbit Tendons and Ligaments

MR Arthrography of Shoulder

Personalized Solutions. MRI Protocol for PSI and Signature Guides

Sport Specific MRI. The symptoms of the majority, if not all sports injuries are experienced when upright, and weight-bearing

UPPER EXTREMITY

Research Article UTE-T2 Analysis of Diseased and Healthy Achilles Tendons and Correlation with Clinical Score: An In Vivo Preliminary Study

MRI of the long head of the biceps tendon: a pictorial review.

Mast Cell Tryptase Modifies The Contraction Of Human Joint Capsule Cells In Vitro

UCLA UCLA Previously Published Works

Syllabus References. Resources. Video: MRI Introduction

Rotator Cuff and Biceps Pathology

Role of Magnetic Resonance Imaging in Internal Derangement of Shoulder

Ultrashort TE Spectroscopic Imaging (UTESI): Application to the Imaging of Short T2 Relaxation Tissues in the Musculoskeletal System

Fatty Degeneration and Atrophy of Rotator Cuffs: Comparison of Immediate Postoperative MRI with Preoperative MRI

Viviane Khoury, MD. Assistant Professor Department of Radiology University of Pennsylvania

Unlocking the locked Knee

The posterolateral corner of the knee: the normal and the pathological

Muscle-Tendon Mechanics Dr. Ted Milner (KIN 416)

Basics of MRI Part I

A Method to Monitor Local Changes in MR Signal Intensity in Articular Cartilage: A Potential Marker for Cartilage Degeneration in Osteoarthritis

MRI PEDIATRIC PROTOCOLS (Updated 6/19/2018)

Achilles Tendon After Percutaneous Surgical Repair: Serial MRI Observation of Uncomplicated Healing

Accuracy of SPECT bone scintigraphy in diagnosis of meniscal tears ABSTRACT

Transcription:

T he British Journal of Radiology, 7 (998), 3 36 998 The British Institute of Radiology The magic angle phenomenon in tendons: effect of varying the MR echo time W C G PEH, FRCR and J H M CHAN, MSc Department of Diagnostic Radiology, The University of Hong Kong, Queen Mary Hospital, Hong Kong Abstract. Increased signal intensity on magnetic resonance (MR) imaging of tendons arising from the magic angle phenomenon is well recognized. This study aimed to evaluate the effect of varying the echo time (TE) upon tendon signal intensity, and to determine if a modified TE value produces acceptable T and proton density (PD) weighted images. Fresh bovine tendons were imaged in a.5 T MR scanner using spin echo (SE) T and PD weighted sequences and utilizing a number of different coils. For each set of sequences, the tendon was orientated at 55 to the main magnetic field (B ) and imaged using constant TR and incremental TE values. Signal intensity was measured on images at each TR/TE value and compared with the signal intensities of tendons orientated at to B, obtained using minimum TE values. This experiment was repeated with a. T MR scanner and utilizing a spine coil. The Achilles tendon of a human volunteer was similarly imaged using a general purpose flex coil. For bovine and human tendons orientated at 55 to B, the signal intensities decreased exponentially with increasing TE. A critical TE value exceeding 37 ms, for each sequence, reduced the signal intensities to the levels obtained with the tendons orientated at to B, such that the magic angle phenomenon could be avoided. Although there was variability of the signal intensities with different coils, the critical TE value remained constant and the anatomical clarity was not degraded. The critical TE value was unaltered using two MR scanners of different field strengths. In MRI, the magic angle phenomenon causes on both T and proton density (PD) weighted increased signal on short echo time (TE) images. images, above which the magic angle phenomenon It affects tissues with well-ordered collagen fibres should not be apparent. It was hoped that clinical in one direction, such as tendon or articular hyaline application of modified TE values might reduce cartilage, which both behave in an anisotropic the number of false-positive diagnoses of tendon manner in a magnetic field. The induced hyperintense disease. signal is dependent on the orientation of the tissue to the main magnetic field (B ), with maximum signal intensity observed at the magic Materials and methods angle of 54.74 (usually approximated to 55 ) to In vitro experiments were conducted using B [ 3]. bovine Achilles tendons. The tendons were fresh In clinical practice, it is important to distinguish and intact, with the distal portions of the gastroeffects of the magic angle phenomenon in normal cnemius and other contributing calf muscles still tissue from signal abnormalities due to degener- attached. A total of six tendons were used, each ation or partial tear. Suggested methods include for less than h at room temperature to ensure close comparison with T weighted images, looking freshness. The same tendon was used in each series for other features of tendonopathy and re- of experiments in order to achieve consistency in positioning the patient to avoid the magic angle comparison. MR imaging was performed using a phenomenon [4, 5]. Most of the previous studies.5 T clinical whole-body scanner with the 5.4 of this phenomenon have evaluated the influence version software (Signa Advantage, General of either tendon or articular cartilage orientation Electric Medical Systems, Milwaukee, WI, USA). upon their MR appearances [6 8]. The same settings of centre frequency, transmitter The purpose of this investigation was to study gain and receiver gain were used for each set of the effect of systematically varying the MR TE sequences. values upon the increased signal intensity observed In the first in vitro experiment, the tendon was with the magic angle phenomenon in tendons. We placed on a plastic sheet and covered by ultrasonic hypothesized that there was a critical TE value transmission gel (Aquasonic, Parker Laboratories Inc., Orange, NJ, USA). The purpose of the ultrasonic Received 5 July 997 and in revised form 9 September gel was to provide contrast with the anti- 997, accepted September 997. cipated low signal of the tendon and assist the T he British Journal of Radiology, January 998 3

W C G Peh and J H M Chan tuning procedure of the MR system. Imaging was varying by 3 ms intervals from 3 ms to 4 ms, and performed using a 5-inch general purpose (GP) SE PD weighted images with TE values varying circular coil, placed beneath the plastic sheet by 4 ms intervals from ms to 4 ms. (Figure ). Parameters used were cm fieldof-view The in vivo experiment was carried out on the (FOV), 56 9 matrix, one excitation, Achilles tendon of a healthy 33-year-old human 6 khz bandwidth and 4 mm interleaved section volunteer with no history of ankle or foot probthicknesses. lems. Imaging was performed using a GP flex coil The tendon was placed in the centre of the with the following parameters for SE T weighted gantry, with its long axis orientated at to B. images: cm FOV, 56 9 matrix, one Following a spin echo (SE) T weighted (3/3 excitation, 6 khz bandwidth and 3 mm inter- [TR/TE]) coronal localizer image, true axial SE leaved section thicknesses. These parameters were T weighted (5/3 [TR/TE]) images were the same for SE PD weighted images, except that obtained. The same tendon was then re-orientated a 56 6 matrix was used instead. SE T so that its long axis was 55 to B. The angle was weighted (5/3 [TR/TE]) and SE PD weighted checked initially using a protractor and then by (/ [TE/TR]) true axial images were measurement on the MR console monitor after obtained, following a SE T weighted (3/ obtaining the coronal localizer image. Oblique [TR/TE]) sagittal localizer, with the long axis of sections were performed to obtain a series of nine the Achilles tendon orientated at to B. Using true axial SE T weighted images of the tendon. the same method as the in vitro experiment, the The TR of these images were kept constant at patient s foot was then re-orientated at 55 to B. 5 ms, while the TE was varied by 3 ms intervals Ten true axial SE T weighted images were acquired for each series from 3 ms to 37 ms. at 3 ms intervals from 5/3 to 5/4 [TR/TE], Quantitative evaluation of tendon signal was while seven true axial SE PD weighted images measured by manually placing a region of interest were acquired at a constant TR value of ms (ROI) cursor in the centre of a selected axial and the following TE values:,, 8, 3, 36, 4 tendon image. Care was taken to ensure that the and 48 ms. ROI was of the same size and shape, and that the Objective evaluation of tendon signal was made, same site was measured on anatomically corresponding using the methods described for the in vitro experi- axial slices. Measurements were made ments. In addition, for each selected T and PD with the tendon orientated to to B (5/3 weighted image obtained at 55 orientation, [TR/TE]), and at 55 to B (5/3 5/37 muscle, bone marrow and subcutaneous fat were [TR/TE]). The in vitro experiments using T subjectively assessed for anatomical clarity on the weighting were repeated using GP flex coils, phase- MR console monitor. These were compared with array spine coils, head, pelvis and body coils. The corresponding T and PD weighted images, experiment was also performed, using GP 5-inch obtained with the tendon at orientation. Using and phase-array spine coils, to obtain conventional ROI cursors of the same size and shape, quantitat- SE PD-weighted images with TE values varying ive measurements of signal intensity of bone by 4 ms intervals from / to /5 marrow, subcutaneous fat, muscle and Achilles [TR/TE]. tendon were made, at the same sites on corresponding T weighted images (5/3 to 5/4 The in vitro experiment was repeated using a. T clinical whole-body scanner with the 5.5 [TR/TE]). version software (Signa Horizon, General Electric Medical Systems, Milwaukee, WI, USA). A spine Results coil was used to image the bovine tendon. SE T weighted images were obtained with TE values As expected, the signal intensities of both bovine and human Achilles tendons remained low when using minimum TE values on both SE T and PD weighted images with tendon orientation at to B. When the tendons were re-orientated at 55 to B, the same T and PD weighted TR/TE values produced markedly increased intratendon signal intensity, due to the magic angle phenomenon (Figures a and b). Signal intensity decreased in an exponential manner with increasing TE values on both T and PD weighted images of the bovine tendons (Figure 3). If the tendon signal intensity measured at orientation was taken as being Figure. Bovine tendon covered with ultrasonic trans- mission gel. The 5-inch general purpose circular coil is just visible beneath the plastic sheet. normal, then a critical TE value of about 37 ms or more was required to reduce the magic angle 3 T he British Journal of Radiology, January 998

Magic angle phenomenon in tendons (a) (b) (c) Figure. Axial SE T weighted images of bovine tendon. The same window level and width was used for all images. (a) Tendon at to B (5/3) shows normal tendon signal. (b) Tendon at 55 to B (5/3) shows increased signal (arrowed) due to the magic angle phenomenon. (c) The high intratendon signal is markedly reduced with alteration of the TE time (5/37) at the same 55 to B orientation. Table. Effect of different coils on bovine tendon signal intensity (TR/TE=5/3 ms) Coil Signal Signal Ratio of signal intensity at intensity at increase at to B 55 to B 55 to B Body 53 4.7 Spine 7 43 3.5 5 inch GP 7 85 3.9 Head 58 9 3.8 GP Flex 7 88 4. Pelvic 7 74 3.9 Figure 3. Relative bovine tendon signal intensities plotted weighted images of the human tendon orientated against different SE PD weighted TE values, with tendon orientated at 55 to B (bold curved line). TR is constant at 55 to B, obtained at TE values just exceeding at ms. The signal intensity obtained with the same the critical level of 37 ms, provided similar anatom- tendon orientated at to B, with a SE PD weighted ical information to SE T and PD weighted images (/) sequence, is shown by the dotted straight line. of the tendon orientated at to B acquired with The two lines intersect at a TE of about 37 ms. minimum TE values of 3 ms and ms, respectively. With adjustment of the window width and induced signal intensity near to that of a tendon levels to optimize anatomical clarity, the subcut- orientated at to B (Figure c). Similarly-shaped aneous fat, bone marrow and muscle had similar exponential decay curves were achieved with the appearances on corresponding SE T and PD various other coils, with a lower (. T) field weighted images, despite differences in TE values strength scanner and with the human in vivo experiment. Although the different coils produced differ- intensities of the SE T -weighted images of the (Figure 4). At the critical TE of 37 ms, signal ent signal intensities and ratios of signal increase, human tendon showed 4% reduction for fat (bone varying from.7 for the body coil to 4. for the marrow and subcutaneous fat), 56% reduction for GP flex coil (Table ), the critical TE value of muscle but an 8% reduction for tendon (Table ). 37 ms remained statistically constant with a stan- In both T and PD weighted images of tendon, dard deviation of. ms for all coils. signal-to-noise ratio (SNR) decreased by about With the appropriate windowing, SE T and PD 4% at the critical TE values (Table 3). T he British Journal of Radiology, January 998 33

W C G Peh and J H M Chan (a) (b) Figure 4. Axial MR images of the human volunteer, showing similar anatomical clarity with short and long TE. (a) SE PD weighted short TE (/) image with tendon orientated at to B. (b) SE PD weighted long TE (/4) image with tendon orientated at 55 to B. The Achilles tendon is arrowed. Table. Comparison of signal intensities of subcutaneous Table 3. Calculated decrease in signal-to-noise ratio fat, bone marrow, muscle and Achilles tendon with (SNR) with increasing TE values on T and PD weighted various echo times (TE) on SE T weighted (constant images of a normal human Achilles tendon TR of 5 ms) axial images of a normal human ankle TE SNR % DROP TE Signal intensity (ms) (a) T weighted images (T R=5 ms) Subcutaneous Bone Muscle Tendon 3 57 % fat marrow 6 55 4% 9 5 9% 3 6 56 59 4 48 6% 6 568 57 54 8 5 4 8% 9 537 54 33 68 8 39 3% 53 47 5 54 3 36 37% 5 48 436 8 45 34 35 39% 8 453 46 96 34 37 3 43% 3 44 373 85 3 4 9 49% 34 374 346 73 4 (b) PD weighted images (T R= ms) 37 357 33 7 87 % 4 337 3 67 6 7 9% 8 6 3% Discussion 3 57 35% 36 5 4% 4 48 45% Normal tendons produce little or no signal on 48 43 5% conventional MRI sequences. This is because the tendon consists of type I collagen fibres orientated in a parallel manner into highly ordered bundles. times in the order of 5 ms. The resultant rate of Tendons, therefore, display structural anisotropy, T decay is so rapid that a very low signal is creating a static local magnetic field which contrib- obtained irrespective of TE value [ 3]. utes further to spin spin interaction, causing The static local magnetic field component B is z normal tendons to have ultra-short T relaxation superimposed onto the static field B of the magnet, 34 T he British Journal of Radiology, January 998

Magic angle phenomenon in tendons causing the spins to precess at different Larmor constant while using different coils and with MR frequencies, and resulting in spin dephasing. The scanners of different field strengths. This critical rate of spin dephasing caused by the anisotropic TE value was independent of the differences in structure of tendons is thus proportional to a tendon signal intensity produced among the coils factor (3 cosw ), where w is the angle between used. It was considered unnecessary to scan a large the static field of the magnet and the longitudinal number of animal or human tendons as our axis of the tendon. This additional spin dephasing hypothesis of a critical TE value could be consistently vanishes if 3 cosw =, where w=54.74. In demonstrated on the generated exponential other words, when w=54.74, the spin spin inter- decay curves plotting signal intensity against TE action due to the static local field becomes zero in each series of experiments. and the T decay is governed by the time-varying Imaging the ankle of a normal human volunteer field only. Consequently, the T relaxation time showed that anatomical information was not lost rises to about ms and the rate of T decay is with increasing TE up to the critical TE value. If less rapid, so that the normal tendon appears increased signal from the magic angle phenomenon hyperintense at short TE values [ 3]. is suspected on routine short TE sequences, for On the other hand, the T relaxation time is example where a tendon alters direction across a independent of tendon orientation. The signal joint, repeating the T or PD weighted sequence intensity is governed by the same T relaxation with an increased TE value (of 37 ms or more) time decay curve regardless of whether there is T, should be helpful in distinguishing increased signal PD or T weighting. For each type of sequence, due to the magic angle phenomenon from degener- the TR (which is constant) and T relaxation time ation, tendinitis or partial tear. These and other (which is constant and independent of tendon abnormalities of tendon would be expected to orientation) values would only determine the initial remain visible, if not more prominent, with signal intensity, hence the pattern of T decay is increased and long TE values. The decrease in unaffected. Thus, regardless of whether the SNR with increased TE values can be compensated sequences are T or PD weighted, the point in the by doubling the number of excitations in the repeat graph where the T decay curve of a tendon at 55 scan. If the magic angle phenomenon is present, orientation intersects the horizontal line of a the need to reposition the patient or structure of tendon at orientation remains unchanged interest can also be avoided using this method. (Figure 3). Alternatively, one may possibly prospectively per- The clinical significance of the magic angle form T and PD weighted images with TE values (w=55 ) phenomenon lies in its ability to produce just exceeding 37 ms instead of minimum TE increased signal intensity in normal tendons. The values, if the magic angle phenomenon is anticipated overlap among MRI of the normal rotator cuff from the position of certain tendons and tendon, tendinosis and partial tear makes differen- cartilaginous structures. It would be useful to tiation difficult in some cases [9]. Increased signal conduct prospective studies comparing the appear- on short TE images has been attributed to early ances of various types of tendon lesions in the tendinosis or tendon degeneration []. This pattern clinical setting, employing both standard TE and of signal alteration is supported by Kjellin increased critical TE values. et al who showed that MR correlated well with the histological findings of tendon degeneration in Acknowledgment cadavers []. The magic angle signal artefacts have been recogassistance. We thank Ms Alice Lau for her secretarial nized at several sites, particularly in tendons of the wrist, ankle and rotator cuff, as well as cartilaginous structures such as the knee meniscus and References glenoid labrum. Besides evaluating secondary signs. Fullerton GD, Cameron IL, Ord VA. Orientation of of tendonopathy such as tendon thickening, shape tendons in the magnetic field and its effect on T changes and presence of tendon sheath fluid, most relaxation times. Radiology 985;55:433 5. authors have advocated repositioning the tendon. Erickson SJ, Cox IH, Hyde JS, Carrera GF, Strandt in question in order to eliminate the magic angle JA, Estkowski LD. Effect of tendon orientation on MR imaging signal intensity: a manifestation of effect [, 4, 5, 7, 8]. We have shown that the the magic angle phenomenon. Radiology tendon signal changes produced by the magic angle 99;8:389 9. phenomenon can be greatly reduced by increasing 3. Erickson SJ, Prost RW, Timins ME. The magic the echo time to above a certain critical value. A angle effect: background physics and clinical rel- critical TE value of about 37 ms was obtained in evance. Radiology 993;88:3 5. 4. Peterfy CG, Linares R, Steinbach LS. Recent the experiments with both fresh bovine and normal advances in magnetic resonance imaging of the human volunteer tendons, using T as well as PD musculoskeletal system. Radiol Clin North Am weighted sequences. The critical TE value remained 994;3:9 3. T he British Journal of Radiology, January 998 35

W C G Peh and J H M Chan 5. Hayes CW, Parellada JA. The magic angle effect 9. Robertson PL, Schweitzer ME, Mitchell DG, in musculoskeletal MR imaging. Topics MRI Schlesinger F, Epstein RE, Frieman BG, et al. 996;8:5 6. Rotator cuff disorders: interobserver and intraobserver 6. Rubenstein JD, Kim JK, Morava-Protzner I, variation in diagnosis with MR imaging. Stanchev PL, Henkelman RM. Effects of collagen Radiology 995;94:83 5. orientation on MR imaging characteristics of bovine. Weatherall PT, Crues JV III. Musculotendinous articular cartilage. Radiology 993;88:9 6. injury. MRI Clin North Am 995;3:753 7. 7. Peterfy CG, Janzen DL, Tirman PFJ, van Dijke CF,. Kjellin I, Ho CP, Cervilla V, Haghighi P, Kerr R, Pollack M, Genant HK. Magic-angle phenom- Vangness CT, et al. Alterations in the supraspinatus enon: a cause of increased signal in the normal tendon at MR imaging: correlation with histo- lateral meniscus on short-te MR images of the pathologic findings in cadavers. Radiology knee. AJR 994;63:49 54. 99;8:837 4. 8. Timins ME, Erickson SJ, Estkowski LD, Carrera GF, Komorowski RA. Increased signal in the normal supraspinatus tendon on MR imaging: diagnostic pitfall caused by the magic-angle effect. AJR 995;64:9 4. 36 T he British Journal of Radiology, January 998