Effect of Cryogenic Treatment on Fracture Resistance of Nickel Titanium Rotary Instruments a Systematic Review

Similar documents
Effect of cryogenic treatment on nickel-titanium endodontic instruments

Microstructural and mechanical characterizations of new Ni-Ti endodontic instruments

Glassman Webinar PPT 7/18/2014. Nickel Titanium Clinical Performance (2012) Depends on: three main factors :

The advent of nickel-titanium (NiTi) rotary instruments has revolutionized root canal

Effect of autoclave sterilization on the cyclic fatigue resistance of thermally treated Nickel Titanium instruments

Improve Your Root Canal Preparations with Adaptive Motion Technology

Race family Endodontic NiTi rotary files

A Review of the Various Surface Treatments of NiTi Instruments

The introduction of nickel-titanium (NiTi) to endodontics over 2 decades ago has

Over the past 2 decades, nickel-titanium (NiTi) instruments have become an important

Differential Scanning Calorimetric (DSC) Study of New and Sterilized Nickel-Titanium Rotary Endodontic Instruments

Comparison of the flexibility and torsional resistance of nickel-titanium rotary instruments.

Torsion Property and Cyclic Fatigue Fracture Behavior of. Nickel-Titanium Endodontic Instruments

thinking beyond cyclic fatigue Testing standards for endo files vary and there are other qualities to be concerned with, too

Hend M. Abou El Nasr * and Shehab El Din M. Saber ** ABSTRACT صور 4 لون

CIC Edizioni Internazionali. In vitro resistance to fracture of two nickel-titanium rotary instruments made with different thermal treatments

Evolution of Nickel titanium Alloys in Endodontics

Surface properties of nickel-titanium rotary instruments

Evaluation of physicochemical and mechanical properties of NiTi endodontic instruments made with conventional and controlled-memory alloys

Comparison of cyclic fatigue resistance of XP-endo Shaper, HyFlex CM, FlexMaster and Race instruments

Comparison of cyclic fatigue resistance of original and counterfeit rotary instruments

An In Vitro Comparison of Cyclic Fatigue of Profile Vortex and Endosequence Rotary Nickel- Titanium Files

Nickel-titanium (NiTi) endodontic files are particularly helpful for successful shaping

The nickel-titanium (NiTi) alloy, which consists of nickel and titanium in a nearly

Cyclic fatigue life of Tango-Endo, WaveOne GOLD, and Reciproc NiTi instruments

HyFlex nickel titanium rotary instruments after clinical use: metallurgical properties

Cyclic fatigue resistance and fractographic analysis of RaCe and ProTaper rotary NiTi instruments

Is the reciprocating movement per se able to improve the cyclic fatigue resistance of instruments?

Nickel titanium instruments in endodontics: a concise review of the state of the art

AN INVESTIGATION OF PHASE TRANSFORMATION MECHANISMS FOR NICKEL-TITANIUM ROTARY ENDODONTIC INSTRUMENTS DISSERTATION

Effective cleaning and shaping of the root

A numerical method for predicting the bending fatigue life of NiTi and stainless steel root canal instruments

This in-vitro investigation was designed to form a simulated clinical use using RaCe

Torsion and Bending Properties of EdgeEndo Files

Lifespan of One Shape Files used in Severely Curved Canals in Resin Blocks and Extracted Teeth

irace Quick, effective and safe

in simulated curved canals

Endodontic instrument fracture: causes and prevention

Effect From Surface Treatment of Nickel-Titanium Rotary Files on the Fracture Resistance

Rotary Fatigue Testing to Determine the Fatigue Life of NiTi alloy Wires: An Experimental and Numerical Analisys

ANATOMIC. XP-endo Finisher. optimal cleaning while preserving dentine

ORIGINAL PAPERS. Porównanie skuteczności opracowania sztucznych kanałów w kształcie litery L dwoma systemami maszynowymi: Hyflex CM i ProTaper Next

XP-endo Shaper. The One to Shape your Success

REPORT DOCUMENTATION PAGE

Torsional Testing of Race, K3 and Profile Nickel Titanium Endodontic Files

Microestructure alterations of rotary files after multiple simulated operative procedures

Safety and Simplicity!

UNITED STATES PATENT AND TRADEMARK OFFICE BEFORE THE PATENT TRIAL AND APPEAL BOARD. US ENDODONTICS, LLC, Petitioner,

Evaluation of selected mechanical properties of NiTi rotary glide path files manufactured from controlled memory wires

Comparison of the Effects of Three Different Nickel titanium Rotary Instruments on the Fracture Resistance of Obturated Roots: An in vitro Study

The Effect of Surface Defects on the Cyclic Fatigue Fracture of HEROShaper Ni-Ti rotary files in a Dynamic Model: A Fractographic Analysis

Comparative analysis of torsional and bending behavior through finiteelement models of five Nickel-Titanium endodontic instruments

The use of sodium hypochlorite (NaOCl) to irrigate root canals is currently the gold

Torque During Canal Instrumentation Using Rotary Nickel-Titanium Files

Management of fractured instruments during dental treatment by dental practitioners

Comparison of the shaping ability of Twisted Files with ProTaper and RevoS nickel-titanium instruments in simulated canals

A comparison of cyclic fatigue between used and new Mtwo Ni Ti rotary instruments

Comparison of the Different Techniques to Remove Fractured Endodontic Instruments from Root Canal Systems

Evaluation of canal preparation with rotary and hand Niti files in curved root canals using Cone Beam Computed Tomography: An In Vitro Study

Analysis of Surface Characteristics of ProTaper Universal and ProTaper Next Instruments by Scanning Electron Microscopy

Integra Miltex. Endodontic Files. Limit uncertainty with the brand you trust

Dynamic and Cyclic Fatigue of Engine-Driven Rotary Nickel-Titanium Endodontic Instruments

Evaluation of the level of safety of three instrumentation systems to prepare curved canals

Cutting Efficiency of ESX, EndoSequence, and ProFile Instruments Using the TrueTooth Replica

Comparison of the shaping ability of GT Series X, Twisted Files and AlphaKite rotary nickel-titanium systems in simulated canals

WHERE SAFETY MEETS EFFICIENCY

by Drs. Clifford J. Ruddle, Pierre Machtou and John D. West

Instructions for use: FlexMaster. the rotary NiTi-System for every case. Endo Easy Efficient

Flute and Shank Dimensions of Reciprocating Instruments before and after Simulated Root Canal shaping

COMPARATIVE STUDY OF APICALLY EXTRUDED DEBRIS AND IRRIGANT AFTER USING TWO ROTARY SYSTEMS (K3, RACE)

Research Article Structural Characterisation and Mechanical FE Analysis of Conventional and M-Wire Ni-Ti Alloys Used in Endodontic Rotary Instruments

Physical and mechanical properties of a thermomechanically treated NiTi wire used in the manufacture of rotary endodontic instruments

Fracture and deformation of ProTaper Next instruments after clinical use

1) INDICATIONS FOR USE These products have to be used only in hospital environments, clinics or dental offices by qualified dental personnel.

THE INFLUENCE OF GLIDE PATH PREPARATION ON THE FAILURE RATE OF NICKEL-TITANIUM RECIPROCATING INSTRUMENTS

IN VITRO EVALUATION OF ROOT CANAL PREPARATION WITH TWO ROTARY INSTRUMENT SYSTEMS PRO TAPER AND HERO SHAPER

CBCT EVALUATION OF CANAL TRANSPORTATION USINGNEWER ROTARY NI-TI FILE SYSTEMS-AN IN VITRO STUDY.

WaveoneandoneShapeFilesSurvivalinSeverelyCurvedArtificialCanals

Cyclic Fatigue Resistance of Three Single-use NiTi Instruments after Immersion in Sodium Hypochlorite

A WHOLE NEW DIMENSION OF TOTAL PERFORMANCE.

Case report for the new HyFlex CM NiTi File system of Coltene Endo

LegacyENDO. Create your own Legacy! Controlled Memory Cross Cut. Apex Access. Multi Taper. K-Files

UNICLIP DIRECTIONS FOR USE REF C226U - 0,8 MM / 1 MM FOR DENTAL USE ONLY. FISDR / F EN / 11 / updated 06/2016 1/7

ProUltra ULTRASONIC SURGICAL TIPS

Single File Reciprocating Technique Using Conventional Nickel Titanium Rotary Endodontic Files

Dental Research Journal

ProUltra Ultrasonic Non Surgical Endo Tips

Shaping Potential of Manual NiTi K-File and Rotary ProTaper and Analyzing the Final Outcome of Shaped Canals using CT

Shaping Goals. Endodontic Canal Preparation: New Developments. Clinical goals. Mechanical goals. Where are we standing with this?

Prevention and management of fractured instruments in endodontic treatment

Software and Hardware Implementation of a Microtorquemeter

In vitro Study of Apically Extruded Debris and Irrigant Following the Use of Conventional and Rotary Instrumentation Techniques

Rotary Systems: An Insight

Mechanical properties of WaveOne and ProTaper instruments in relation to torsion, hardness and tenacity

A survey on NiTi rotary instruments usage by endodontists and general dentist in Tehran

Influence of cervical preflaring on apical file size determination - An in vitro study

UNITED STATES PATENT AND TRADEMARK OFFICE BEFORE THE PATENT TRIAL AND APPEAL BOARD. US ENDODONTICS, LLC, Petitioner

Howard E. Strassler, DMD University of Maryland School of Dentistry

Failure of ProTaper rotary Ni-Ti instruments used by undergraduate students

Fracture Resistance of Wave-one Rotary File Used In Reciprocating Motion

Transcription:

IJCRR Section: Healthcare Sci. Journal Impact Factor 4.016 ICV: 71.54 Review Article Effect of Cryogenic Treatment on Fracture Resistance of Nickel Titanium Rotary Instruments a Systematic Review Kumar Ujjwal 1, Anamika C. Borkar 2, Piyush Oswal 3, Karan Bhargava 3 1 Post Graduate, Dr. D.Y. Patil Dental College & Hospital, Pune, India; 2 Professor & PG Guide, Dr. D.Y. Patil Dental College & Hospital, Pune, India; 3 Senior Lecturer, Dr. D.Y. Patil Dental College & Hospital, Pune, India. ABSTRACT Introduction: Endodontic instruments upon rotation are subjected to both tensile and compressive stresses in curved canals. This stress is localized at the point of curvature. It has been demonstrated that the continuous cycle of tensile and compressive forces in the area of curvature of the canal to which (Nickel Titanium) NiTi rotary instruments are subjected produce a very destructive form of loading, causing cyclic fatigue and eventually fracture of the instrument. 2 The aim of this study is to conduct a systematic review on the effect of cryogenic treatment on fracture resistance of nickel titanium rotary instruments. Methods: Two Internet sources of evidence were used in the search of appropriate papers satisfying the study purpose: the National Library of Medicine (MEDLINE PubMed) and Google Scholar, Google and manual search using DPU college library resources. All cross reference lists of the selected studies were screened for additional papers that could meet the eligibility criteria of the study. Results: Preliminary screening consisted total of 54 articles out of which 24 articles were selected. The papers were screened independently by two reviewers. At first the papers were screened by title and abstract. As a second step, full text papers were obtained when they fulfilled the criteria of the study aim. Any disagreement between the two reviewers was resolved after additional discussion. Finally 3 articles were included in this study. Conclusions: Within the limitations of this study we can conclude that the cryogenic treatment increases the fracture resistance of nickel titanium rotary instruments. Key Words: Treatment, Fracture Resistance, Nickel-Titanium Rotary Instruments INTRODUCTION Nickel-titanium alloy exhibits the unique properties of shape memory (SM) and pseudoelasticity. 1 It is well known that instrument separation occurs either by brittle fracture (cyclic fatigue) or by ductile fracture (torsional fatigue). Endodontic instruments upon rotation are subjected to both tensile and compressive stresses in curved canals 2. This stress is localized at the point of curvature. It has been demonstrated that the continuous cycle of tensile and compressive forces in the area of curvature of the canal to which NiTi rotary instruments are subjected produce a very destructive form of loading, causing cyclic fatigue and eventually fracture of the instrument 2. Despite the extreme flexibility, 4 these thermomechanically processed NiTi instruments undergo permanent plastic deformation, especially while reusing the smaller size instruments 5,6 due to inadequate martensite content. Cryogenic treatment is a supplementary procedure of subjecting stainless steel 7,8 and superelastic NiTi 3 to subzero temperatures that affects the entire bulk of the material rather than the accessible surface alone. One of the mechanisms proposed behind the improvement in properties following cryogenic treatment is the complete martensitic transformation from the austenite phase. 9 It has been proved that deep cryogenic treatment (DCT) at a soaking temperature of up to 185 C increases the martensite content of SM NiTi alloys by a similar mechanism. 10 Previous studies have been done evaluating the effect of cryogenic Corresponding Author: Kumar Ujjwal, Post Graduate, Dr. D.Y. Patil Dental College & Hospital, Pune, India. Mob: 9021451669; Email: guptaujjwal05@gmail.com ISSN: 2231-2196 (Print) ISSN: 0975-5241 (Online) DOI: 10.7324/IJCRR.2017.9156 Received: 19.05.2017 Revised: 16.06.2017 Accepted: 20.07.2017 Int J Cur Res Rev Vol 9 Issue 15 August 2017 32

treatment on the fracture resistance of nickel titanium rotary instruments. However, no systematic review comparing the effect of cryogenic treatment on fracture resistance of nickel titanium rotary instruments have been conducted. The aim of this study is to conduct a systematic review on the effect of cryogenic treatment on fracture resistance of nickel titanium rotary instruments. METHODS Inclusion criteria: i. Articles in English or those having detailed summary in English. ii. In vitro Studies published between 1 st January 1985 and 31 st June 2016. iii. New packet of Nickel-Titanium rotary files. iv. Cryogenic treatment modality. Exclusion criteria: i. Used nickel-titanium rotary file. ii. In vivo studies, reviews, case reports were excluded. Included Eligibility Screening Identification STUDY SELECTION PRISMA 2009 Flow Diagram Records Records identified identified through through database Additional Additional records records identified identified database searching searching through through other other sources sources (n = 54) (n (n =0) =0) Records after duplicates removal Records excluded (n = 24) (n = 20) Records screened according to to title and abstract (n =4) Full-text articles excluded, Full-text Full-text articles articles assessed with reasons assessed for eligibility for eligibility (n = 1) (n = 4) Based on different outcome = 1 Studies included in qualitative synthesis (n = 3) PICO: P Product: Nickel-titanium rotary files. I - Exposure: treatment modalities. C -Comparison: between cryogenically treated and non treated nickel titanium rotary instruments. O - Outcome: increase in the fracture resistance of treated nickel-titanium rotary files. Two Internet sources of evidence were used in the search of appropriate papers satisfying the study purpose: the National Library of Medicine (MEDLINE PubMed) and Google Scholar, Google and manual search using DPU college library resources. All cross reference lists of the selected studies were screened for additional papers that could meet the eligibility criteria of the study. The databases were searched up to and including 31 st June 2016 using the search strategy. RESULTS Preliminary screening consisted total of 54 articles out of which 24 articles were selected. The papers were screened independently by two reviewers. At first the papers were screened by title and abstract. As a second step, full text papers were obtained when they fulfilled the criteria of the study aim. Any disagreement between the two reviewers was resolved after additional discussion. Finally 3 articles were included in this study. (Fig no.1) Figure 1: Summarizing the article selection process. DISCUSSION Biomechanical preparation has been considered the most important step in root canal therapy. 15 Advent of rotary NiTi instruments has aided endodontists in cleaning and shaping of root canals. Civjan et al in 1975 suggested the use of NiTi alloy for fabrication of hand and rotary endodontic instruments. 15 The extraordinary characteristics of super elasticity and shape memory of the NiTi alloy have made it possible to manufacture rotary instruments.1 Endodontic instruments upon rotation are subjected to both tensile and compressive stresses in curved canals. 2 This stress is localized at the point of curvature. It has been demonstrated that the continuous cycle of tensile and compressive forces in the area of curvature of the canal to which NiTi rotary instruments are subjected produce a very destructive form of loading, causing cyclic fatigue and eventually fracture of the instrument. 2 It has been suggested that cyclic fatigue has accounted for 50 90% of mechanical failures. 2 Cryogenic methods have been used to increase the wear, abrasion, corrosion resistance, and to improve the strength of metals. Cryogenically treated NiTi instruments have shown increased microhardness. Deep cryogenic treatment involves suspending the metal over a super-cooled bath containing liquid nitrogen at 196 C or 320 F and then allowing the metal to slowly warm to room temperature. 16 33 Int J Cur Res Rev Vol 9 Issue 15 August 2017

Methods like boron ion implantation, 17 thermal nitridation, 18 physical vapor deposition of titanium nitride 19 and electropolishing 20 have been used for increasing the cutting efficiency of rotary NiTi instruments. Anderson et al in 2007 showed that electropolished instruments performed significantly better than non-electropolished instruments in both the cyclic fatigue testing and static torsional loading, which is likely to be caused by a reduction in surface irregularities that serve as points of stress concentration and crack initiation in nickel titanium. 21 Methods like boron ion implantation 17 thermal nitridation 18 physical vapor deposition of titanium nitride 19 and electropolishing 20 have been used for increasing the cutting efficiency of rotary NiTi instruments. Anderson et al in 2007 showed that electropolished instruments performed significantly better than non-electropolished instruments in both the cyclic fatigue testing and static torsional loading, which is likely to be caused by a reduction in surface irregularities that serve as points of stress concentration and crack initiation in nickel titanium. 21 Kim et al 2005 22 studied the effect of cryogenic treatment on nickel-titanium endodontic instruments.30 size 25 nickeltitanium K-files (NTO2525; Dentsply-Tulsa Dental) were embedded in epoxy resin (811-563-103 and 811-563-104; Leco, St Joseph, MI, USA) mixed according to the manufacturer s instructions. 15 control instruments and 15 cryogenically treated instruments were used. Each instrument was cut at the handle and then placed within a mounting ring (20-8161-010; Buehler, Lake Bluff, IL, USA). The mounting rings were brushed with releasing agent (20-8185-032; Buehler) and placed upon a flat surface. The resin and hardener were then mixed until clear in appearance and was then poured into each mounting ring. The epoxy resin was left to cure for 8 h. The resin blocks were then removed from the mounting rings and ground to reveal a cross-section of the instruments and polished flat using a grinder/polisher (Phoenix Beta; Buehler). Silicon carbide polishing papers (240, 320, 400, 600 and 1200) were used in succession followed by Al 2 O 3 powder/h 2 O suspensions (1.0, 0.3 and 0.05μm particle sizes) for final polishing. A Vicker s indenter was used to make two indentations adjacent to the edge of the instrument cross-section (FM-7; Future Tech, Tokyo, Japan). A 9.8 N indentation load was applied for a 15 s dwell time. Both indentation diagonals were measured, and the Vicker s microhardness (VHN), was calculated from the size of the indentation. According to the equation VHN = 0.1891F/ d 2 Where, F is the indentation load having unit (N) and d is the average diagonal length of indentation having unit (mm). It was concluded that there was an increase in the microhardness following cryogenic treatment. Non treated instruments had a mean VHN of 339.3 ± 23.0, and treated instruments had VHN of 346.7 ± 20.6 (Fig. 2). A Student s t-test showed this to be a statistically significant difference (P < 0.001; β > 0.999). Mahalaxmi Sekar et al 2011 16 studied the effect of deep dry cryotreatment on the cutting efficiency of three rotary nickel titanium instruments. 20 NiTi instruments of RaCe (FKG Dentaire, Switzerland), K3 (Sybron Endo, CA, USA) and Hero Shaper (Micro Mega, France) of size 25, 21mm length and of 0.06 taper, were used for this study. The files were randomly divided into two groups for evaluating the cyclic fatigue as follows: Group A: untreated NiTi rotary files and Group B: cryotreated NiTi rotary files. These were further divided as the follows. Group A1: 10 untreated RaCe rotary files : Group A2: 10 untreated K3 rotary files:group A3: 10 untreated Hero Shaper rotary files. Group B1: 10 cryotreated RaCe rotary files: Group B2: 10 cryotreated K3 rotary files: Group B3: 10 cryotreated Hero Shaper rotary files. Mean number of cycles required to fracture in Group B1 (405 ± 3) was significantly higher than that of Group A1 (352 ± 2) (P < 0.0001). Mean number of cycles to fracture in Group B2 (708 ± 4) was significantly higher than that in Group A2 (452 ± 3) (P < 0.0001). Mean number of cycles to fracture in Group B3 (359 ± 10) was significantly higher than that in Group A3 (268 ± 7) (P < 0.0001). Groups B1, B2, and B3 showed significantly higher number of cycles to fracture than Groups A1, A2 and A3 (P < 0.0001). The results of this study showed an increase in the fracture resistance of NiTi files which were treated as compared to untreated files. This could be attributed to the fact that the complete transformation of the austenitic phase of the alloy to martensitic phase, which could have occurred at 195 C, would have decreased the internal stress within the alloy due to plastic deformation. Amini et al [15] have explained in their study that the presence of residual austenite phase in an alloy decreases hardness and also reduces the wear resistance of the tool. Thus, increasing resistance to wear, reduction of internal stresses can be regarded as the most important benefits of using cryogenic treatment. It was observed that by decreasing the temperature of cryogenic environment, better properties were obtained. The deep dry cryogenic treatment (DCT) has been seen to affect the entire cross section of the instrument rather than just the surface, with no change in the elemental crystalline composition of the alloy. T.S.Vinothkumar et al 2016 14 studied mechanical behavior of deep cryogenically treated martensitic shape memory nickel titanium rotary endodontic instruments. The martensitic Int J Cur Res Rev Vol 9 Issue 15 August 2017 34

SM NiTi instruments made from controlled memory wire (HyFlex CM; ColteneWhaledent Inc., Cuyahoga Falls, OH, USA) of size 25, 0.06 taper were selected for this study. A total of 75 instruments were randomly divided into three groups of 25 each as follows: DCT 24: Soaking temperature 185 C, soaking time 24 h, DCT 6: Soaking temperature 185 C, soaking time of 6 h, number of cycles to failure (NCF) for each instrument. The NCF is directly proportional to the cyclic fatigue resistance of the instrument. They concluded that the maximum NCF of each HyFlex CM rotary endodontic instrument following rotation in artificial root canals with 30 angle of curvature was observed in DCT 24 group followed by DCT 6 and control. Despite the extreme flexibility of HyFlex CM SM NiTi instrument, supplementary cryogenic treatment would further increase the volume of martensite resulting in superadded benefits, especially for smaller size instruments. Cryogenic treatment can be classified based on the soaking temperature as deep ( 185 C) and shallow ( 80 C). [9,19,20] Deep soaking temperatures are effective in increasing the martensite content of SM NiTi alloys than the conventional shallow cold treatments. Consequently, DCT significantly reduces the hardness and increases the wear resistance of the SM NiTi alloy. Therefore, the objective of this study was only to evaluate the role of DCT soaking time with respect to the constant cooling and warming rate. The advantage of the dry DCT facility used in this study is to gradually increase or decrease the temperature in order to avoid thermal shock to the instrument that would make it brittle. CONCLUSION This systematic review summarizes that the cryogenic treatment of NITI Rotary endodontic instruments increases the fracture resistance of the instruments. It could be attributed to the complete transformation of the austenitic phase to martensitic phase. LIMITATIONS Search strategy may not be complete due to limited accessibility. IMPLICATIONS FOR FUTURE RESEARCH Cryogenic treatment appears to be a promising supplementary method in the manufacturing of nickel titanium endodontic instruments. Cryogenic treatment significantly increases the fracture resistance of nickel titanium alloy. Further investigations are required to evaluate the impact of cryogenic treatment on fracture resistance of nickel titanium rotary instruments. ACKNOWLEDGEMENTS The completion of this systematic review would not have been possible without the people who fill my life with positivity. It has been a privilege to work under my advisor, my mentor and guide Dr. Anamika C. Borkar. I thank her for her timely advice and constant encouragement. I would like to acknowledge all the authors and researchers for their contribution to the subject. Abbreviations: 1. NITI Nickel Titanium 2. SM - Shape Memory 3. DCT Deep Cryogenic Treatment 4. VHN Vicker s Hardness Number 5. NCF Number of Cycles to Failures REFERENCES 1. Walia H, Brantley WA, Gerstein H. An initial investigation of thee bending and torsional properties of Nitinol root canal files. J Endod 1988;21:346-51. 2. Pruet JP, Clmenat DJ, Carnes DL. Cyclic fatigue testing of NiTi endodontic instruments. J Endod 1997;23:77-85. 3. Vinothkumar TS, Miglani R, Lakshminarayananan L. Influence of deep dry cryogenic treatment on cutting efficiency and wear resistance of nickel-titanium rotary endodontic instruments. J Endod 2007;33:1355-8. 4. Testarelli L, Plotino G, Al-Sudani D, Vincenzi V, Giansiracusa A, Grande NM, et al. Bending properties of a new nickel-titanium alloy with a lower percent by weight of nickel. J Endod 2011;37:1293-5. 5. Haapasalo M, Shen Y. Evolution of nickel-titanium instruments: From past to future. Endod Topics 2013;29:3-17. 6. Peters OA, Gluskin AK, Weiss RA, Han JT. An in vitro assessment of the physical properties of novel Hyflex nickel-titanium rotary instruments. Int Endod J 2012;45:1027-34. 7. Lal DM, Renganarayanan S, Kalanidhi A. Cryogenic treatment to augment wear resistance of tool and die steels. Cryogenics 2001;41:149-55. 8. Bensely A, Senthilkumar D, Lal DM, Nagarajan G, Rajadurai A. Effect of cryogenic treatment on tensile behavior of case carburized steel-815m17. Mater Charact 2007;58:485-91. 9. Barron RF. Cryogenic treatment of metals to improve wear resistance. Cryogenics 1982;22:409-13. 10. Vinothkumar TS, Kandaswamy D, Prabhakaran G, Rajadurai A. Microstructure of cryogenically treated martensitic shape memory nickel-titanium alloy. J Conserv Dent 2015;18:292-6. 11. Fayyad DM, ElhakimElgendy AA. Cutting efficiency of twisted versus machined nickel-titanium endodontic files. J Endod 2011;37:1143-6. 12. Parashos P, Messer HH. Rotary NiTi instrument fracture and its consequences. J Endod 2006;32:1031-43. 13. Plotino G, Grande NM, Cordaro M, Testarelli L, Gambarini G. A review of cyclic fatigue testing of nickel-titanium rotary instruments. J Endod 2009;35:1469-76. 35 Int J Cur Res Rev Vol 9 Issue 15 August 2017

14. Vinothkumar TS, Kandaswamy D, Prabhakaran G, Rajadurai A. Mechanical behavior of deep cryogenically treated martensitic shape memory nickel titanium rotary endodontic instruments. Eur J Dent 2016;10:183-7. 15. Civjan S, Huget EF, DeSimon LB. Potential applications of certain nickel titanium alloys. J Dent Res 1975;54:89-96. 16. George GK, Sanjeev K, Sekar M. An in vitro evaluation of the effect of deep dry cryotreatment on the cutting efficiency of three rotary nickel titanium instruments. J Conserv Dent. 2011 Apr;14(2):169-72. 17. Lee DH, Park B, Saxeba A, Serene TP. Enhanced surface hardness by boron implantation in Nitinol alloy. J Endod 1996;22:543-6. 18. Rapisarda E, Bonaccorso A, Tripi TR, Fragalk I, Condorelli GG. The effect of surface treatments of nickel-titanium files on wear and cutting efficiency. Oral Surg Oral Med Oral Pathol Oral RadiolEndod 2000;89:363-8. 19. Schäfer E. Effect of physical vapour deposition on cutting efficiency of nickel titanium files. J Endod 2002;28:800-2. 20. Baumann MA. Reamer with alternating cutting edges concept and clinical application. Endod Top 2005;10:176-8. 21. Anderson EM, Price JW, Parashos P. Fracture resistance of electropolished rotary nickel-titanium endodontic instruments. J Endod 2007;33:1212-6. 22. Kim J. W., Griggs J. A., Regan J. D., Ellis R. A. andcai Z. Effect of cryogenic treatment on nickel-titanium endodontic instrument. IntEndod J. 2005 June ; 38(6): 364 371. Int J Cur Res Rev Vol 9 Issue 15 August 2017 36