Doses Received by Patients during Thorax X-Ray Examinations

Similar documents
PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS

Survey of dose area product received by patients undergoing common radiological examinations in four centers in Nigeria

Measurement of Kidneys Dose in Three Hospitals during Chest and Abdomen Radiographic Examinations

Measurement of Entrance Skin Dose and Calculation of Effective Dose for Common Diagnostic X-Ray Examinations in Kashan, Iran

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA

A Survey of Radiation Exposure Techniques Factors Used for Common Diagnostics X- Ray Examinations in Akwa Ibom State, Nigeria

Research Article. Corresponding author H. Osman

Radiation exposure of the Yazd population from medical conventional X-ray examinations

Entrance Surface Air Kerma for Skull, Pelvis and Abdomen X-ray Examinations in Some Diagnostic Radiology Facilities in Akwa Ibom State, Nigeria

Patient doses in radiographic examinations in Western and Eastern Azerbyjan provinces of Iran

International Journal of Radiation Research, January 2017

Entrance Surface Dose from Pediatric Diagnostic X-ray Examinations in a Developing World Setting: Are We ALARA Principle Compliant?

A Standardised Approach to Optimisation

Survey Effect of the Corrective Actions and Quality Control in Image Quality and Patient Dose in Radiography Examinations

THE LEVEL OF COMPLIANCE OF SELECTED NIGERIAN X-RAY DEPARTMENTS TO EUROPEAN GUIDELINES ON GOOD RADIOGRAPHIC TECHNIQUES

Iranian Journal of Medical Physics

Patient effective dose evaluation for chest X-ray examination in three digital radiography centers

Application of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology

Local reference dose evaluation in conventional radiography examinations in Iran

Radiation Doses to Patients Undergoing Selected Radiographic Examinations in Three Hospitals South-West, Nigeria

Measurement of Absorbed Doses in Anatomical Phantoms

Patient dose in routine X-ray examinations in Yazd state

Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology

Dosimetric Consideration in Diagnostic Radiology

P T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3

Impact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals

ABSTRACT I. INTRODUCTION

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Assessment of effective dose in paediatric CT examinations

Patient dose measurement in common medical X-ray examinations in Iran

Equivalent dose to organs and tissues in Hysterosalpingography calculated with the FAX (Female Adult voxel) phantom

Radiation Protection Dosimetry Advance Access published April 22, 2008

CALDose_X a software tool for absorbed dose calculations in diagnostic radiology

Assessing the standard dose to standard patients for x-ray investigations

Comparison of organ doses estimations in radiology with PCXMC application based on MIRD phantoms and CALDose-X application based on voxel phantoms

Quality Control in Mammography: Results from a study in Recife, Brazil. H. J. Khoury, V. S. de Barros, M. Sampaio

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 6, 2016

The impact of SID and collimation on backscatter in radiography

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Patient doses in general radiography examinations

OPTIMISATION OF PATIENT PROTECTION AND IMAGE QUALITY IN DIAGNOSTIC RADIOLOGY

Evaluation of radiation dose and image quality following changes to tube potential (kvp) in conventional paediatric chest radiography

Iranian Journal of Medical Physics

Entrance Surface Dose Measurement on the Thyroid Gland in Orthopantomography: The Need for Optimization

Evaluation of the doses delivered to premature babies in the Belgian Neonatal Intensive Care Units

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Survey of patients CT radiation dose in Jiangsu Province

Receptor dose and Patient Dose in Radiographic exposures - a 15 year review

Establishment of local diagnostic reference levels for quality control in intraoral radiography

Radiation Protection Dosimetry Advance Access published September 13, 2006

Diagnostic reference levels as a quality assurance tool

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

CALDose_X online: Web-based, real time Monte Carlo calculations for patient dosimetry in X-ray diagnosis

An Assessment of Organ and Effective Dose of Patients who Undertake CT Examinations in two Teaching Hospitals of Mashhad&Isfahan

Assessment of radiation risk to pediatric patients undergoing conventional X-ray examinations

Estimation of Radiation Absorbed Dose to Patients Undergoing Chest X-ray Examination in Four Government Own Hospitals in Nigeria

JMSCR Vol 06 Issue 12 Page December 2018

Mdiagnosing breast by using x-ray to both breasts. It

Managing Patient Dose in Computed Tomography (CT)

Assessment of Patient Dose from CT Examinations in Khorasan, Iran

Introduction and Background

Diagnostic reference levels in intraoral dental radiography in Korea

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

ESTIMATION OF EFFECTIVE DOSE FROM PEDIATRIC RADIOGRAPHY OF THE CHEST IN MEDELLÍN-COLOMBIA a

Doses to Patients arising from Dental X-ray Examinations in the UK, A Review of Dental X-ray Protection Service Data

THE STUDY OF PATIENT DOSES IN DIAGNOSTIC RADIOLOGY FOR ESTABLISHMENT OF NATIONAL DIAGNOSTIC REFERENCE LEVELS

Limited information related to: - Type and frequency of procedures; - Exposure levels involved.

SMM SUDAN MEDICAL MONITOR. National University. ISSN Vol 9 Issue 1 January 2014

Conventional and spiral CT dose indices in Yazd general hospitals, Iran

Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography

Debra Pennington, MD Director of Imaging Dell Children s Medical Center

Quality Control Status of Radiology Centers of Hospitals Associated with Mashhad University of Medical Sciences

Olowookere, C.J. Dept of Physics Ajayi Crowther University, Oyo

K E M Mohamadain 1,6,LARdaRosa 2, A C P Azevedo 1, M R N Guebel 3, M C B Boechat 4 and F Habani 5

Radiation Dose To Pediatric Patients in Computed Tomography in Sudan

Efficacy of Increasing Focus to Film Distance (FFD) for Patient s Dose and Image Quality in Pediatric Chest Radiography

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS

Pilot study of patient and phantom breast dose measurements in Bulgaria

Quality Control and Patient Dosimetry on line for Computed Tomography

A more accurate method to estimate patient dose during body CT examinations with tube current modulation

Dosimetry in digital mammography

Radiation Dose Assessment and Risk Estimation During Extracorporeal Shock Wave Lithotripsy

Radiation dose levels assessment in mammography

National collection of local diagnostic reference levels in Norway and their role in optimization of X-ray examinations

Ask EuroSafe Imaging Tips & Tricks. CT Working Group. On the use of Diagnostic Reference Levels in CT

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

PNRA-RG (Rev. 0) December, 2017 PROTECTION OF PATIENTS IN DIAGNOSTIC RADIOLOGY REGULATORY GUIDE PAKISTAN NUCLEAR REGULATORY AUTHORITY

IAEA TECDOC 1517 Quality Control in Mammography Software

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

The use of gonadal shielding in singular common diagnostic radiographic procedures

EVOLUTION OF THE EXCESS ABSOLUT RISK (EAR) IN THE VALENCIAN BREAST CANCER SCREENING PROGRAMME

Mean glandular dose values used for the mammography screening program in Poland according to the type of image registration system

Background Radiation in U.S. ~ msv/yr msv/yr ~0.02 ~0.02 msv msv/day /day (~2 m rem/day) mrem/day) NCRP 4

CT Radiation Risks and Dose Reduction

ICRP Symposium on the International System of Radiological Protection

Transcription:

Iranian Journal of Medical Physics Vol. 9, No. 4, Autumn 01, 45-51 Received: September 16, 01; Accepted: February 0, 013 Original Article Doses Received by Patients during Thorax X-Ray Examinations Nsikan U. Esen 1*, Rachel. I. Obed Abstract Introduction Radiation exposures from diagnostic medical examinations are generally low and are almost always justified by the benefits of accurate diagnosis of possible disease conditions. Therefore, entrance skin dose (ESD), body organ dose (BOD), and effective dose (ED) from adult patients undergoing routine thorax posterioranterior (PA) and thorax right lateral (RLAT) were estimated in University Hospital, Port Harcourt, Southern Nigeria. Materials and Methods Totally, 10 patients were considered in this work. Using software packages to carry out ESD, BOD, and ED is a recent resource in dosimetry and is being widely used in hospitals. The software used in this work was CALDose_X 5.0. The software makes use of the technical exposure parameters and the tube output of the X- ray machine. Results The estimated ESD median values were 0.96 and 1.85 mgy for thorax posterior anterior (PA) and right lateral (RLAT), respectively. The highest BOD was in the adrenals (70 µgy) for thorax PA and Liver (63 µgy) for thorax RLAT. Similarly, ED for thorax PA and RLAT examination were 0.068 and 0.107 mgy, respectively. Conclusion It could be observed that examinations that imparted the highest ESD were thorax PA when compared with the established dose level. Therefore, these results call for quality assurance program (QAP) in diagnostic X- ray units in Nigeria hospitals. Keywords: Effective Dose, Entrance Skin dose, Radiography 1- Department of Physics, Akwa Ibom State University, PMB 1167 Uyo, Nigeria *Corresponding author: Tel: +34(0)803763537; Email: nsikanjesus@yahoo.com - Department of Physics, University of Ibadan, Ibadan, Nigeria Iran J Med Phys, Vol. 9, No. 4, Autumn 01 45

Nsikan Esen et al. 1. Introduction The radiation protection system for patients subjected to medical exposures in diagnostic radiology is governed by principles of justification and optimization, including the consideration of diagnostic reference levels (DRLs). Therefore, a diagnostic radiological procedure is justified if the benefits to the individual patient from the radiological image balance the individual detriment the exposure may cause. Once a medical exposure has been justified, the principle of optimization is applied, that is, the radiological examination must be carried out with equipment and exposure parameters that ensure doses to patients as low as reasonably practicable and consistent with the intended diagnostic purpose [1]. National Radiation Protection Board (NRPB) [] in the United Kingdom published a national protocol for measuring radiation dose to patients in diagnostic radiology. It has also been recommended that the absorbed dose be included in the medical record of patient for certain radiographic procedure such as entrance skin dose (ESD) measurement [3]. The United Nations Scientific Committee on the Effect of Atomic Radiation (UNSCEAR) [4] and International Atomic Energy Agency (IAEA) [5] recommended ESD as dose descriptor for guidance levels in diagnostic radiography because it provides an indication of maximum skin dose and is useful for periodic checking of patient dose. Moreover, it serves as an indication of effective dose for particular radiography procedure. In Nigeria, most of the studies on patient dosimetry have been carried out on the measurement of entrance skin doses to patients in four common diagnostic examinations by thermoluminescence dosimetry (TLD) [6-9]. Moreover, the research on radiological parameters and radiation doses of patients undergoing X-ray examinations in three Nigerian hospitals by Ogundare et al., on TLDs have been used to measure ESDs of patients undergoing X-ray examinations [10]. Their findings revealed that their mean ESDs values were within the international level. The aim of this work is to estimate patients doses during diagnostic X-ray examination of thorax posterior anterior (PA) and thorax right lateral (RLAT) in University Hospital, Port Harcourt, Southern Nigeria (UPTH). The results of this study will serve as a useful baseline against which individual X-ray unit in Nigeria hospitals may compare their doses to reduce patients doses and will also be a useful review of dose evaluation of patients in Nigeria hospitals.. Materials and Methods In total, 10 patients were included throughout the study. Fifty-two patients for each of the projections were considered. Both males and females were included in the research. Patient s information (age and weight) and exposure parameters (tube potential and current-time product) for the routine X-ray examination considered are indicated in Table 1. The ESD received by 10 patients were included in this study. This dose survey was conducted between May 01 and September 01. The patients were randomly selected from adult patients of both sexes (male and female) attending medical investigations in UPTH. The two projections considered on thorax radiograph in this study included PA and Right Lateral (RLAT). For each patient, age, sex, weight, and height were recorded. X- ray factors used during each radiographic exposure such as tube voltage (kvp), charge (mas), focus-to-film distance (FFD), field size, and projection were recorded. The filmscreen speed was 00 (i.e sensitivity 0.64 mr). A Silhouette VR X-ray machine was used in this work. The equipment had a total filtration of.5 mmal with fixed grid configuration. The system had an AEC (X-ray generator) with Bucky grid ratios system of 10:1. Silhouette VR X-ray tube used recommended units (X-ray generator) of 3 and 35 kw. The operating potential used for each radiographic examination was determined based on the type of examination, patient s weight and thickness. 46 Iran J Med Phys, Vol. 9, No. 4, Autumn 01

Patient Dose in Thorax Radiology In order to facilitate measurement and optimization of patient dose, the National Radiological Protection Board (NRPB) introduced the national protocol for patient dose measurement [11]. Moreover, from their recommendation, ESD was directly measured on a sample of patient using TLD. Free-air measurement of tube s radiation output together with the calculation of ESD using standard factors may be employed in appropriate circumstances [1]. The application of software to evaluate patients doses in common routine X-ray examination is a modern resource in dosimetry and is widely used in hospitals [13-19]. For the present study, we employed windowbased software for calculation of patients dose due to unavailability of TLD chips and TLD readers in Nigeria. The result obtained in this study was compared with result obtained from TLD measurement in Southwest Nigeria [10] and Dosecal software for evaluation [6]. In order to increase the speed and efficiency of the patient dosimetry process, a windowsbased computer program, CALDose_X 5.0, has been developed by Kramer et al. CALDose_X is a software tool that provides the possibility to calculate incident air kerma (INAK) and entrance surface air kerma (ESAK), two important quantities used in X- ray diagnosis, based on the output of the X-ray equipment. Additionally, the software uses conversion coefficients (CCs) to assess absorbed dose to organs and tissues of the human body, the effective dose as well as the patients cancer risk for radiographic examinations [15]. The CCs, ratios between organ and tissues absorbed doses and measurable quantities, have been calculated with the FAX06 and the MAX06 phantoms for 34 projections of 10 commonly performed X-ray examinations[15]. A combinations of 40 tube potential and filtration ranging from 50 to 10 kvp and from.0 to 5.0 mm Al, respectively. Various field positions, for 9 selected organs and tissues had been simultaneously calculated for the measurable quantities(inak, ESAK, and kerma area product (KAP))[15]. Based on the X-ray irradiation parameters defined by the user, CALDose_X shows images of the phantom together with the position of the X- ray beam. By using true to nature voxel phantoms, CALDose_X improves earlier software tools, which were mostly based on mathematical MIRD5-type phantoms, i.e. poor representations of human anatomy [14]. For the CALDose _X 5.0 to work, it is necessary to furnish the output in mgy/mas, of all X-rays machines used in the evaluation of doses. Once the tube potential, the tube current, the exposure time, the FDD, and focus-to-skin distance (FSD) were known, ESD could be calculated by (1) kv ESD=Output 80 100 mas BSF (1) FSD The output is in mgy/mas of the X-ray tube at 80 kv at a distance 100 cm normalized to 10 mas. BSF is backscatter factor for a particular examination at the required potential and was taken from NRPB numerical simulations [13]. In order to determine the effective dose, CALDose _X5 provides separately calculated weighted MAX06 and FAX06 whole body absorbed doses [15], which represent the sexspecific contributions to the effective dose. According to equation 4.5 in International Commission on Radiological Protection (ICRP) publication 103 [0], the effective dose is then just the arithmetic mean of the two sexspecific weighted absorbed doses. According to ICRP 103 section (13), equation 4.5, the effective dose can be evaluated by this expression: E = [H (female) + H (male)] E = [ H (female) + H (male)] () CALDose_X 5.0 calculates a weighted female dose (F) and a weighted male dose (M) given at the end of the result tables [15]. The effective dose based on CALDose_ X5.0 is then the average of the sex-specific weighted doses as in equation 3 E = [F + M] (3) Iran J Med Phys, Vol. 9, No. 4, Autumn 01 47

Nsikan Esen et al. 3. Results The descriptive statistics of ESD, i.e., minimum (Min), maximum (Max), first (1st) and third (3rd) quartile, median, and mean are presented in Table. Table 3 presents the max/min ratio, inter-quartile range (ratio 3rd/1st), standard deviation (SD), and percentage coefficient of variation (CV%) of ESD of individual adult patient. All the ESD are below the dose reference level [-5] except for thorax PA projection. The max/min ratio of ESD for PA and RLAT are.538 and.068 mgy, respectively. Table 4 presents the distribution of mean value of Effective Dose (ED) with ESD median value. Table 5 presents the obtained values for tissue/organ dose for thorax PA and RLAT. The highest body organ dose (BOD) was in adrenals PA with 194 µgy and liver RLAT with 64 µgy for thorax radiography. Table 6 shows the comparison of calculated ESD with the international established reference dose level for SW (Southwest) Nigeria, IAEA, NRPB and UNSCEAR recommendations. The hospital considered used low tube potential and employed a total filtration of.5 mm Al. Nevertheless, the filtration values in this study were not measured, but given by the radiographers. Figure1 is the distribution of the ESD with the established reference level. Table 1. Exposure parameters and patient s information for UPTH with mean and range in bracket. Types of Examination THORAX Projection PA Patient Age (year) 35.6 (0 70) Patient Height (cm) 169.4 (163 176) Patient Weight (kg) 66.4 (60 73) Tube voltage (kvp) 71.04 (67 77) currenttime product (mas) 0.78 (1.5 5) FFD (cm) 186.5 (185 188) RLAT 47.40 (7 80) 169.6 (163 176) 66.6 (60 73) 76.19 (70 80) 30.70 (5 35) 187.4 (185 189) Table. Descriptive statistics of ESD (mgy) in UPTH. Radiograph Projection Min Max Median 1 st 3 rd Quartile Quartile Mean THORAX PA 0.504 1.79 0.911 0.849 1.151 0.958 RLAT 1.05.176 1.9 1.671 1.939 1.850 Table 3. Distribution of ESD with max/min, 3 rd /1 st quartile, standard deviation and coefficient of variation of individual patient in UPTH. Radiograph Projection Max/Min Median 3 rd /1 st Quartile SD CV% THORAX PA.538 0.911 1.355 0.548 57% RLAT.068 1.9 1.160 0.795 43% Table 4. Distribution of effective dose (msv) with doses reported in the literature. Radiograph Projection ED (Mean) Values reported in literature (msv) * THORAX PA 0.06±0.063 0.007 0.050 RLAT 0.6±0.318 0.05 0.4 *Mettler et al. [1] 48 Iran J Med Phys, Vol. 9, No. 4, Autumn 01

Patient Dose in Thorax Radiology Table 5. Distribution of tissue/organ dose (mgy) for thorax radiograph Organ/tissue Thorax PA SD. Thorax RLAT SD ESAK 0.946 0.537 1.83 0.778 ADRENALS 0.194 0.197 0.146 0.083 BREASTS, 0.03 0.035 0.15 0.078 KIDNEYS 0.171 0.179 0.103 0.057 LIVER 0.078 0.09 0.63 0.155 LUNGS 0.10 0.193 0.70 0.141 ESD 0.958 0.548 1.850 0.795 S. AVERAGE 0.188 0.158 0.06 0.111 MAX.RBM 0.31 0.04 0.508 0.55 MAX.BSC 0.30 0.60 0.60 0.318 ED 0.068 0.063 0.107 0.057 Table 6. Comparison of ESD (median value) with the international established reference dose level in (mgy) Radiograph Projection Present Study SW Nigeria 009 UNSCEAR 008 IAEA 1996 NRPB 000 THORAX PA 0.911 0.59 0.35 0.4 0. RLAT 1.9 0.61 0.81 1.5 1.0 1.8 1.6 1.4 1. 1 0.8 0.6 THORAX PA THORAX RLAT 0.4 0. 0 Present Study SW Nigeria 009 UNSCEAR 008 IAEA 1996 NRPB 000 Fig.1 Distribution of the ESD with the established reference level. 4. Discussion From the results obtained, there is a wide difference in patients doses in the hospital considered for individual patient and for each projection, though the mean dose does not really vary greatly from one projection to another. From Table 6, the highest ESD median value was from both thorax PA and RLAT examinations. The mean patient weight was 66.6 kg and the mean age was 35.6 kg which Iran J Med Phys, Vol. 9, No. 4, Autumn 01 49

Nsikan Esen et al. show that the study sample is younger than the UK (47 66) [13] and nearly the same as Malaysia (37-49) [19] surveys. Moreover, the range of the tube potential for the projections was (68 80) kvp which is similar to that of UK and Malaysia studies (66 89) kvp. Tables and 3 present the statistical analysis (minimum, maximum, standard deviation, ratio of max/min, and coefficient of variation) distributions of the patients while Table 4 shows the comparison of this work s mean effective dose with the ranges of effective doses reported in the literature for the different examinations and procedures. Comparing the mean effective dose in the studied hospital with that of the reported literature is greater than the doses obtained in the literature by a factor of 0.1. Table 5 present the distribution of tissue/organ dose for thorax examination. From the distribution, the highest value was recorded in the lungs with 70 µgy for PA and RLAT projections, respectively. Table 6 is a comparison of the obtained ESD in this work with established reference doses in the UK (NRPB), IAEA, and UNSCEAR recommendation. In this comparison, the values of the obtained ESD in the study area for thorax examination for PA and RLAT projection are higher than those of UK (NRPB), IAEA, South-West Nigeria (SWN), and UNSCEAR. The ESD for thorax PA was found to be higher than UK and IAEA by a factor of.5. The thorax RLAT projection had the lowest variation with the UK, varying with a factor of.1. In all, the use of low potentials and high mas values was common in the studied hospital and have been observed as the main causes of errors/differences in doses when compared with the reference level. Similarly, results have also been reported regarding the variations in ESD [6, 10, and 15]. They were no better reasons by the radiographer when asked than saying that it gives a better resolution, i.e., a clearer image. In addition, these variations in dose levels for the ESD may be due to many limitations amongst which the efficiency of the equipments, the processing systems, and the radiographic techniques used in each of the hospitals can be mentioned. These variations in doses suggest that investigations are required to be made to ensure the ALARA principle. This also suggests that there could be reduction in doses if radiographers adhere strictly to the guidelines that correct the operative modalities. 5. Conclusion An estimation of entrance surface dose and effective dose during diagnostic X-ray examination of adult patients in University of Port-Harcourt Teaching Hospitals, Rivers State, has been carried out. The patient s individual ESD and body organ dose values were observed to be consistent with the range of values of the existing knowledge. The ESD (median) values of the present study were compared with the reference levels of the existing knowledge and the values obtained on the present study are mostly comparable. The ESD in all the examinations were below the DRLs [3,5] except for thorax PA projection. The ranged factor obtained in this work was moderately low when compared with those reported from Southwest (SW) Nigeria. The mean effective doses when compared with those found in existing knowledge were lower. This implies that the radiation risk to an average patient in the hospitals included in this work is low and the risk to workers in the hospitals is generally low. The findings suggest that there is a serious need for quality assurance program and monitoring aimed towards reducing patients dose in Nigeria. This can be achieve by organizing regular workshops and conferences for radiographers, setting of guidelines for different exposure, and establishing of diagnostic reference levels with which individual hospitals may compare their doses. Acknowledgements The authors thank the entire radiologists and radiographers in the hospital participating in the survey. Also appreciate Prof. R. Kramer 50 Iran J Med Phys, Vol. 9, No. 4, Autumn 01

Patient Dose in Thorax Radiology for providing us with the CALDose software program. References 1. Wall BF. Radiation protection dosimetry for diagnostic radiology patients. Radiat Prot Dosimetry. 004;109(4):409-19.. National Radiological Protection Board (NRPB) report R63, 1993. Radiation Exposure of the UK population 000. 3. International Commission on Radiological Protection, 000. Avoidance of Radiation Injuries from Medical Interventional Procedures ICRP Publication 85 Anal of ICRP, 30: 7 57.ational Radiological Protection Board (NRPB) report R63, 1993. Radiation Exposure of the UK population 1993 Review (NRPB, Chilton, Didcot, UK). 4. United Nations Scientific Committee on the Effect of Atomic Radiation. UNSCEAR 008 Report: Vol. 1. 53 59. 5. IAEA, International Atomic Energy Agency. 1996. International Basic Safety Standard for Protection Against Ionizing Radiation and for the Safety of Radiation Sources. Safety Series No 115. IAEA: Vienna, Austria. 6. Obed, R.I., Ademola, A.K., Adewoyin, K.A and Okunade, O.A. Doses to Patients in Routine X- Ray Examinations of Chest, Skull, Abdomen and Pelvis in Nine Selected Hospitals in Nigeria. Research Journal of Medical Sciences. 007; 1( 4), 09-14. 7. Egbe NO, Inyang SO, Ibeagwa OB, Chiaghanam NO. Pediatric radiography entrance doses for some routine procedures in three hospitals within eastern Nigeria. J Med Phys. 008 Jan;33(1):9-34. 8. Ajayi I, Akinwumiju A. Measurement of entrance skin doses to patients in four common diagnostic examinations by thermoluminescence dosimetry in Nigeria. Radiat Prot Dosimetry. 000;87(3):17-0. 9. Ogunseyinde AO, Adeniran SA, Obed RI, Akinlade BI, Ogundare FO. Comparison of entrance surface doses of some X ray examinations with CEC reference doses. Radiat Prot Dosimetry. 00;98():31-4. 10. Ogundare FO, Ajibola CL, Balogun FA. Survey of radiological techniques and doses of children undergoing some common x-ray examinations in three hospitals in Nigeria. Med Phys. 004 Mar;31(3):51-4. 11. National Radiological Protection Board (NRPB) report R63, 1993. Radiation Exposure of the UK population 1993. 1. Toosi MTB, Asadinezhad M. Local diagnostic reference levels for some common diagnostic X-ray examinations in Tehran county of Iran. Radiation protection dosimetry. 007;14():137-44. 13. Davies M, McCallum H, Whiter G, Brown J, Helm M. Patient dose audit in diagnostic radiography using custom designed software. Radiography. 1997;3(1):17-5 14. Hart D, Hillier M, Wall B. Doses to patients from medical X-ray examinations in the UK-000 review: National Radiological Protection Board Chilton; 00. 15. Kramer R, Khoury H, Vieira J. CALDose_X a software tool for the assessment of organ and tissue absorbed doses, effective dose and cancer risks in diagnostic radiology. Physics in medicine and biology. 008;53():6437. 16. ICRP 007 Recommendations of the International Commission on Radiological Protection ICRP 000 Publication 85 Annal of ICRP 30 7-57 17. Biju K, Nagarajan PS. Normalized organ doses and effective doses to a reference Indian adult male in conventional medical diagnostic x-ray examinations. Health Phys. 006 Mar;90(3):17-. 18. Johnston DA, Brennan PC. Reference dose levels for patients undergoing common diagnostic X-ray examinations in Irish hospitals. Br J Radiol. 000 Apr;73(868):396-40. 19. Omrane LB, Verhaegen F, Chahed N, Mtimet S. An investigation of entrance surface dose calculations for diagnostic radiology using Monte Carlo simulations and radiotherapy dosimetry formalisms. Phys Med Biol. 003 Jun 1;48(1):1809-4. 0. Ng KH, Rassiah P, Wang HB, Hambali AS, Muthuvellu P, Lee HP. Doses to patients in routine X-ray examinations in Malaysia. Br J Radiol. 1998 Jun;71(846):654-60. 1. Mettler FA, Jr., Huda W, Yoshizumi TT, Mahesh M. Effective doses in radiology and diagnostic nuclear medicine: a catalog. Radiology. 008 Jul;48(1):54-63. Iran J Med Phys, Vol. 9, No. 4, Autumn 01 51