Poorly placed tracheostomy tubes: Effects on flow and resistance

Similar documents
Disclosures. Learning Objectives. Coeditor/author. Associate Science Editor, American Heart Association

Other methods for maintaining the airway (not definitive airway as still unprotected):

Foundations of Critical Care Nursing Course. Tracheostomy Workbook

1. When a patient fails to ventilate or oxygenate adequately, the problem is caused by pathophysiological factors such as hyperventilation.

Tracheostomy and laryngectomy airway emergencies: an overview for medical and nursing staff

Translaryngeal Tracheostomy - TLT Fantoni Method

I. Subject: Continuous Positive Airway Pressure CPAP by Continuous Flow Device

TRUSTED SOLUTIONS FOR SIMPLIFIED PATIENT CARE. Shiley Tracheostomy Product Guide GettyImages

Bergen Community College Division of Health Professions Department of Respiratory Care Fundamentals of Respiratory Critical Care

Emergency)tracheostomy)management)/)Patent)upper)airway)

Don t let your patients turn blue! Isn t it about time you used etco 2?

Tracheostomy management Kate Regan MRCP FRCA Katharine Hunt FRCA

Clearing the air.. How to assist and rescue neck breathing patients. Presented by: Don Hall MCD, CCC/SLP Sarah Markel RRT, MHA

Shiley Tracheostomy Products Quick Reference Guide

Section 4.1 Paediatric Tracheostomy Introduction

Teacher : Dorota Marczuk Krynicka, MD., PhD. Coll. Anatomicum, Święcicki Street no. 6, Dept. of Physiology

Translaryngeal tracheostomy

Comparison of patient spirometry and ventilator spirometry

Capnography Connections Guide

Basic Airway Management

Tracheo-innominate artery fistula (TIF) is an uncommon

Misty Max 10 nebulizer

CLINICAL SIGNIFICANCE OF PULMONARY FUNCTION TESTS

GE Healthcare. Non Invasive Ventilation (NIV) For the Engström Ventilator. Relief, Relax, Recovery

Cuff Diameter: A Quantitative

Pedi-Cap CO 2 detector

Ventilating the paediatric patient. Lizzie Barrett Nurse Educator November 2016

Continuous Aerosol Therapy

Changing tracheostomy tubes

NBRC Exam RPFT Registry Examination for Advanced Pulmonary Function Technologists Version: 6.0 [ Total Questions: 111 ]

The Influence of Altered Pulmonarv

Kevin K. Nunnink Extracorporeal Membrane Oxygenation Program

Policy x.xxx. Issued: Artificial Airways and Airway Care. ABC Home Medical Company Policy & Procedure Manual. A. Tracheostomy Tubes ( trach tubes)

Anatomy and Physiology

ASPIRUS WAUSAU HOSPITAL, INC. Passion for excellence. Compassion for people. SUBJECT: END TIDAL CARBON DIOXIDE MONITORING (CAPNOGRAPHY)

Biphasic Capnogram in a Single Lung Transplant Recipient A Case Report

A Protocol for the Analysis of Clinical Incidents September Incident Summary: failure to administer anaesthetic gas at start of operation

October Paediatric Respiratory Workbook APCP RESPIRATORY COMMITTEE

Capnography: The Most Vital of Vital Signs. Tom Ahrens, PhD, RN, FAAN Research Scientist, Barnes-Jewish Hospital, St. Louis, MO May, 2017

Rota-Trach Double Lumen Tracheostomy Tube VITALTEC

HeartCode PALS. PALS Actions Overview > Legend. Contents

Full Range of Tracheostomy Solutions

LEVITAN S FIBREOPTIC STYLET: BEYOND BARRIERS. - Our Perspective.

Bronchoscopes: Occurrence and Management

Audra Fuller MD, Mark Sigler MD, Shrinivas Kambali MD, Raed Alalawi MD

Pulmonary circulation. Lung Blood supply : lungs have a unique blood supply system :

Kapitex Healthcare. making things clearer for tracheostomy patients

Equipment: NRP algorithm, MRSOPA table, medication chart, SpO 2 table Warm

Airway Clearance Devices

OPEN ACCESS ATLAS OF OTOLARYNGOLOGY, HEAD & NECK OPERATIVE SURGERY

Weaning from Mechanical Ventilation. Dr Azmin Huda Abdul Rahim

The Pressure Losses in the Model of Human Lungs Michaela Chovancova, Pavel Niedoba

Test Bank Pilbeam's Mechanical Ventilation Physiological and Clinical Applications 6th Edition Cairo

Tracheostomy in pediatric. Tran Quoc Huy, MD ENT department

Welcome to the Specialized Medical Services respiratory training webinar series!

Tracheostomy management

Competency 1: General principles and equipment required to safely manage a patient with a tracheostomy tube.

Pediatric Cardiac Arrest General


I. Subject: Medication Delivery by Metered Dose Inhaler (MDI)

H: Respiratory Care. Saskatchewan Association of Licensed Practical Nurses, Competency Profile for LPNs, 3rd Ed. 79

TRACHEOSTOMY EMERGENCIES

Simulation 3: Post-term Baby in Labor and Delivery

PRODUCTS FOR THE DIFFICULT AIRWAY. Courtesy of Cook Critical Care

Small Volume Nebulizer Treatment (Hand-Held)

I. Subject: Therapeutic Bronchoscopy and Bronchoscope Assisted Intubation

June 2011 Bill Streett-Training Section Chief

I. Subject: Pressure Support Ventilation (PSV) with BiPAP Device/Nasal CPAP

² C Y E N G R E M E ssignac Cardiac Arrest Resuscitation Device uob

Transcutaneous Monitoring and Case Studies

Airway management problem during anaesthesia. Airway management problem in ICU / HDU. Airway management problem occurring in the Emergency Department

Tracheal Trauma: Management and Treatment. Kosmas Iliadis, MD, PhD, FECTS

Novatech Products for Interventional Pulmonology

Tracheostomy Tube Malposition in Patients Admitted to a Respiratory Acute Care Unit Following Prolonged Vent i la t i o n*

PANELISTS. Controversial Issues In Common Interventions In ORL 4/10/2014

Objectives. Module A2: Upper Airway Anatomy & Physiology. Function of the Lungs/Heart. The lung is for gas exchange. Failure of the Lungs/Heart

Articles. The Advantages of Nebulization in the Treatment of Mechanically Ventilated Neonates. Kristin Smith, RRT-NPS

CARING FOR THE TRACHEOSTOMISED PATIENT: WHAT TO LOOK OUT FOR

Handling Common Problems & Pitfalls During. Oxygen desaturation in patients receiving mechanical ventilation ACUTE SEVERE RESPIRATORY FAILURE

ENDOTRACHEAL INTUBATION POLICY

Chapter 3. Pulmonary Function Study Assessments. Mosby items and derived items 2011, 2006 by Mosby, Inc., an affiliate of Elsevier Inc.

Hemodynamic Monitoring

Capnography (ILS/ALS)

Cuffed Tracheal Tubes in Children - Myths and Facts. PD Dr. Markus Weiss Department of Anaesthesia University Children s Hospital Zurich Switzerland

Introducing Infant Flow Advance SIPAP. By Joanne Cookson March 2008

Waitin In The Wings. Esophageal/Tracheal Double Lumen Airway (Combitube ) Indications and Use for the Pre-Hospital Provider

You are caring for a patient who is intubated and. pressure control ventilation. The ventilator. up to see these scalars

New and Future Trends in EMS. Ron Brown, MD, FACEP Paramedic Lecture Series 2018

Risk Factors of Early Complications of Tracheostomy at Kenyatta National Hospital.

TRACHEOBRONCHIAL FOREIGN BODY REMOVAL ADVICE IN DOGS AND CATS

F: Respiratory Care. College of Licensed Practical Nurses of Alberta, Competency Profile for LPNs, 3rd Ed. 59

All bedside percutaneously placed tracheostomies

A VENTILATOR ADAPTER FOR FIBREOPTIC BRONCHOSCOPY. R.E. NEEDS, M.B., B.CH., r.xa.c.p.(c)

Cough assist T70 for the Tracheostomy Child

Hypoventilation? Obstructive Sleep Apnea? Different Tests, Different Treatment

Kingdom; 2 University of Cambridge, Cambridge, United Kingdom

ISPUB.COM. The Use of LMA in Newborn Resuscitation. R Vadhera INTRODUCTION VENTILATION

NON INVASIVE LIFE SAVERS. Non Invasive Ventilation (NIV)

Jay B. Brodsky, M.D. Professor Department of Anesthesia tel: (650) Stanford University School of Medicine fax: (650)

Transcription:

Original article Poorly placed tracheostomy tubes: Effects on flow and resistance Joshua Moorhouse 1, Tahir Ali 2, Tobias Moorhouse 3 and David Owens 3 Journal of the Intensive Care Society 2015, Vol. 16(4) 282 286! The Intensive Care Society 2015 Reprints and permissions: sagepub.co.uk/ journalspermissions.nav DOI: 10.1177/1751143715582040 jics.sagepub.com Abstract This study aimed to investigate the effects of an incorrectly positioned tracheostomy tube on flow resistance. A laboratory-based model of the trachea was used with both cuffed and uncuffed tracheostomy tubes inserted to variable depths. With a constant metered flow through the model, the pressure for given depths of insertion was recorded. The model was then re-run to test the effect of different flow rates on the system. A total of 468 individual results were grouped and statistically analysed. They showed that both over- and under-insertion increase the pressure within the circuit and that a cuffed tracheostomy tube offers a degree of protection against this. These results were statistically significant with P < 0.05 demonstrating that incorrect positioning has a greater resistance to flow. These results provide an essential scientific basis for further work to assess the clinical significance of incorrect positioning as well as suggesting the need to monitor tube position. Keywords Airway resistance, flow, malposition, tracheostomy Introduction Tracheostomy tube insertion is a common procedure with an estimate of over 12,000 undertaken in the UK each year. 1 Tracheostomy tubes are typically of a basic design with majority consisting of a rigid fixed 90 degree curvature enabling the tube to simultaneously lie perpendicular (and upon) the skin of a patient s neck yet also within and parallel to the patient s trachea. Tracheostomy tube selection has traditionally been based upon the ratio between tracheal diameter and the distance between the tracheal wall and the cutaneous surface of the skin. In clinical practice, however, it is recognised that there is a natural anatomical variation in the skin to trachea:trachea diameter ratio 2 and in the gross dimensions of patient s necks. 3,4 Additionally, the national trend of increasing body habitus 5 together with the natural variation of the population have affected the previously accepted ratio used in tracheostomy tube selection thus complicating sizing. This has led to the postulation; one size does not fit all. Considering the rigid fixed curvature of traditional tracheostomy tubes, the position of the external portion of the tube directly affects the position of the internal portion, specifically the terminal meatus and its parallel relationship to the tracheal wall. Loss of this parallel relationship risks complete or partial occlusion of the tube thus increasing airflow resistance and ventilatory effort. This concept was described in clinical practice with a case report postulating that intra-operative ventilation difficulties were encountered when a tracheostomy tube s internal position was altered and partially obstructed by the tracheal wall. 6 In addition, chronic complications have been described due to variation in tube position. For patients with tracheostomies being weaned off mechanical ventilation, almost 10% were malpositioned (of which 90% were partially occluded by the tracheal wall) and these patients experienced almost double the time to be weaned from mechanical ventilation. 7 While the article mainly examined the risks associated with surgically inserted tracheostomies, it could be argued that the current risk of malpositioning will be lessened by the trend towards routine percutaneous 1 Anaesthetics Department, Wycombe Hospital, High Wycombe, UK 2 Anaesthetics Department, John Radcliffe, Oxford, UK 3 ENT Department, University Hospital Wales, Cardiff, Wales, UK Corresponding author: Joshua Moorhouse, Anaesthetics Department, Wycombe Hospital, Queen Alexandra Road, High Wycombe, Bucks HP11 2TT, UK. Email: joshua.d.moorhouse@gmail.com

Moorhouse et al. 283 tracheostomies followed by bronchoscopic verification. This verification does allow for the early correction of any obstructive positioning of the tube but it does not account for a theory that the majority of malpositioning occurs days to weeks post insertion. Although the consequences of tracheostomy tube malpositioning have been postulated theoretically and described clinically, little evidence exists as to the objective consequences with regard to airflow and resistance. This study examines the effect tracheostomy tube malposition has on airflow and ventilatory effort, using flow resistance as objective evidence. By using a laboratory model of the airway within a controlled environment, quantifiable values could be obtained for this theorised circumstance providing objective data and evidence to further base clinical interpretation. Method The study design involved creating a model in which a tracheostomy tube could be placed into an artificial trachea and held static at varying degrees of insertion. A flow could then be introduced through the model, allowing the change in resistance with insertion depth to be recorded. The experiment could then be repeated for both cuffed and uncuffed tracheostomy tubes. This could then be followed by multiple repeats of the experiment and stratifying the data into groups allowing for statistical analysis. Study equipment Trachea/neck model. Standard bore anaesthetic ventilation tubing was used to act as a model trachea. A fenestration was cut within the ventilation tubing for the tracheostomy tube to be inserted (Figure 1). Once inserted to the required depth (measured using a straight rule) the tracheostomy tube was held static using a clamp. Size 10 Shiley tracheostomy tubes with internal diameter of 8.9 mm and a length of 81.0 mm were used throughout the study with both cuffed and uncuffed varieties tested. The cuffed tube was inflated with 10 ml of air when it was in use. Airflow model. Compressed air from the laboratory s regulated main supply was directed into the model via an adjustable tap which enabled the rate of flow to be varied from the constant output main supply. This supply was then directed through the first RT- 200 calibration analyser to record the flow before being channelled into the tracheostomy tube. A small T piece outlet after the calibration analyser was directed to a second RT-200 calibration analyser, which recorded the pressure within the circuit. All connections within the circuit were attained using universal connectors and standard bore anaesthetic tubing. RT-200 calibration analyser. The RT-200 (Figure 2) (Timeter Instrument Corporation; now Allied Healthcare Products, St Louis, Missouri, USA) is used to check and calibrate ventilation systems and is capable of measuring the flow, pressure and volume of both air and oxygen. The functions used were air high flow (L/min, accuracy 2%), airflow pressure (cm H 2 O, accuracy 1.6%). Two machines were used to enable simultaneous monitoring of both flow and pressure. The analysers are checked and calibrated annually. Study procedure Airflow and resistance through the tracheostomy/ neck model were assessed by first inserting and Figure 1. Neck model configuration showing the tracheostomy tube clamped and inserted through a fenestration within the anaesthetics tubing.

284 Journal of the Intensive Care Society 16(4) Figure 2. Pair of RT-200 calibration analysers. clamping the tracheostomy tube at a pre-determined and measured depth. The range of insertion depths used was between 2 and 8 cm, with 0.5 cm intervals. Each insertion depth was assessed at a range of flow rates to simulate the different stages of ventilation. The flow rates used were: 5, 10, 15, 20, 25 and 30 L/ min. For each flow and depth scenario, the pressure before the tracheostomy tube was recorded. This procedure was carried out for both the cuffed and uncuffed tracheostomy tubes with each measurement being repeated three times to give a mean result. Statistical analysis Statistical analysis was performed by grouping the results of each airway/airflow scenario into three categories of insertion depth to assess for changes in flow resistance. These categories were: (1) over-insertion (2) mid-insertion and (3) under-insertion. The null hypothesis states that there would be no statistical difference in resistance between the three groups. To test for statistical significance a Student s t-test was applied between each data set. Results A total of 468 individual data recordings were taken during the study and the means results were plotted graphically (Figure 3). Assessment of pressure in relation to insertion depth Mean results show that as the depth of tracheostomy tube insertion approaches either extreme the pressure within the system increases. These trends were similar for both the cuffed and uncuffed tracheostomy tubes. The data set was grouped into three different categories according to depth of insertion of the tracheostomy tube. The depth categories were based upon observations of pressure variations between different insertions and were defined as: (1) over-insertion (2 3.5 cm), (2) mid-insertion (4 5 cm) and (3) underinsertion (5.5 8 cm). At each flow rate, the raw data from the three insertion groups were compared for statistical significance. The results showed that, for flow rates above 10 L/min, a statistically significant difference existed between the pressure variations for the correct (mid-insertion) and incorrectly placed tracheostomy tubes (P < 0.05). The only instance where this was not true was for a cuffed tracheostomy tube comparing over-insertion to correct positioning; the results were significantly different (P < 0.05) for flow rates of 25 and 30 L/min. Analysis of the difference between cuffed and uncuffed tubes The data sets were then compared for differences between cuffed and uncuffed tubes. Analysis highlighted that, at high rates of flow, a cuffed tube provided less resistance when malpositioned compared to an uncuffed tube. Statistical analysis using the same categories for level of insertion demonstrated that this finding was significant (P < 0.05) at flow rates of 15 L/min and above for over-inserted tubes and at 20 L/min and above for under-inserted tubes. Discussion This study examined the relationship between resistance to flow and tracheostomy tube insertion depth within a trachea/neck model. Specifically, the study examined whether malpositioning of the tracheostomy tube had an effect on resistance to airflow.

Moorhouse et al. 285 Flow 5 L/min Flow 10 L/min 0.08 0.17 0.06 0.04 0.02 0.16 0.15 0.14 0.00 0.13 Flow 15 L/min Flow 20 L/min 0.32 0.53 0.30 0.28 0.26 0.51 0.49 0.47 0.24 0.45 Flow 25 L/min Flow 30 L/min 0.80 1.10 0.75 0.70 0.65 1.05 1.00 0.95 0.60 0.90 Cuffed tracheostomy tube Uncuffed tracheostomy tube Figure 3. Line graphs showing the depth of insertion of the tracheostomy tube and the resultant pressure variation while being segregated for flow rate and type of tracheostomy tube. The results collected show that the incorrect positioning (over or under insertion) of a tracheostomy tube increases the required pressure to maintain a constant airflow; incorrect positioning increases the resistance to flow (extrapolated from the principle of Poiseuille s law). These results were statistically significant when comparing incorrect against correct positioning. These findings have a direct clinical consequence as this resistance would require greater effort of selfrespiration or mechanical ventilation to maintain adequate oxygenation and carbon dioxide removal. It could be hypothesised that this difference in resistance to flow upon incorrect positioning is due to a change in the flow through the model verging towards turbulent from laminar. Further analysis of the data does appear to suggest this to be the case but further work would be required to make a valid extrapolation. Additionally, the results demonstrated that cuffed tracheostomy tubes were, to a degree, protective

286 Journal of the Intensive Care Society 16(4) against the effect of tube malposition with significantly lower resistance noted at both over- and under-insertion compared to uncuffed tubes. This again may have clinical consequence, specifically when there is a decision to change from a cuffed to an uncuffed tracheostomy tube during disease resolution as this may inadvertently affect patient outcome. Within the operative setting, case reports highlight that obstructions 8 or structural abnormalities 9 with the breathing circuits have significant clinical impact on the anaesthetised patient. For intensive care patients, malposition of a tracheostomy tube increases the average time for a patient to be weaned off ventilation. 7 In conjunction with the results from this study it could be suggested that the effects of tracheostomy tube malposition could lead to increased resistance of breathing and be a cause of a clinically observable effect on a patient. The results obtained in this study are from a laboratory bench model using a minimally compliant, open-ended corrugated tube to model the trachea. This does limit the total significance of the data as the absolute pressure values obtained would not be translatable to the in vivo clinical setting. However, an in vivo apparatus would be complex to establish and a more advanced in vitro model lacks the necessary compliance and characteristics to significantly increase the validity of the model. 10 Although lacking in the ability to extrapolate the absolute pressure values, this simple model addresses the main objective that we set out to investigate: determining whether positioning has an effect on the resistance to flow. Clinically, the effects of an increase in resistance could mean that tracheostomy tube malposition would reduce the positive pressure delivered to a mechanically ventilated patient, potentially reducing the treatment effects from this ventilation. It could also suggest that spontaneously breathing patients would have to increase their work of breathing to accommodate an incorrectly placed tracheostomy tube. This could potentially result in an increase in the time required to wean the patient off a ventilator, or render such attempts unsuccessful. A direct clinical application of these results could be the regular monitoring of tube positioning, especially in patients requiring higher than expected ventilation pressures. This could expose and correct any malpositioning before it is detrimental to the patients. Further studies introducing animal models or in vivo apparatus would help to add directly applicable pressure values in order to assess whether the findings are translatable to the theorised clinical consequence. In conclusion, incorrect positioning of a tracheostomy tube leads to an increased resistance to flow within a tracheal/neck model with this finding being present irrespective of a cuffed or uncuffed tube being used. Cuffed tracheostomy tubes do, however, convey some protection against this effect. Theoretically, this finding may have consequences on clinical care and the need to monitor tube position; however, further research is required to assess the actual clinical significance. Declaration of Conflicting Interests The authors declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article. Funding The authors received no financial support for the research, authorship, and/or publication of this article. References 1. Gwinnutt C. The national tracheostomy project. The emergency management of tracheostomy and emergency management of laryngectomy. Resuscitation Council Newsletter (UK). www.resus.org.uk/newsletr/ nletr12w.pdf (2012, accessed 28 March 2015). 2. Griscom NT, and Wohl ME. Dimensions of the growing trachea related to body height. Length, anteroposterior and transverse diameters, cross-sectional area, and volume in subjects younger than 20 years of age. Am Rev Respir Dis 1985; 131: 840 844. 3. Muhammad JK, Major E, and Patton DW. Evaluating the neck for percutaneous dilatational tracheostomy. J Craniomaxillofac Surg 2000; 28: 336 342. 4. Szeto C, Kost K, Hanley JA, et al. A simple method to predict pretracheal tissue thickness to prevent accidental decannulation in the obese. Otolaryngol Head Neck Surg 2010; 143: 223 229. 5. Finucane MM, Stevens GA, Ezzati M, et al. National, regional, and global trends in body-mass index since 1980: systematic analysis of health examination surveys and epidemiological studies with 960 country-years and 9.1 million participants. Lancet 2011; 377: 557 567. 6. Abola RE, Tan J, Wallach D, et al. Intraoperative airway obstruction related to tracheostomy tube malposition in a patient with achondroplasia and Jeune s syndrome. Middle East J Anesthesiol 2010; 20: 735 738. 7. Schmidt U, Hess D, Kwo J, et al. Tracheostomy tube malposition in patients admitted to a respiratory acute care unit following prolonged ventilation. Chest 2008; 134: 288 294. 8. Sprung J, Bourke DL, Harrison C, et al. Endotracheal tube and tracheobronchial obstruction as causes of hypoventilation with high inspiratory pressures. Chest 1994; 105: 550 552. 9. El-Khatib MF, Husari A, Jamaleddine GW, et al. Changes in resistances of endotracheal tubes with reductions in the cross-sectional area. Eur J Anaesthesiol 2008; 25: 275 279. 10. Walenga RL, Longest PW, and Sundaresan G. Creation of an in vitro biomechanical model of the trachea using rapid prototyping. J Biomech 2014; 47: 1861 1868.