OPTIMIZATION OF COLLIMATOR PARAMETERS TO REDUCE RECTAL DOSE IN INTENSITY-MODULATED PROSTATE TREATMENT PLANNING

Similar documents
New Technologies for the Radiotherapy of Prostate Cancer

Measurement of Dose to Critical Structures Surrounding the Prostate from. Intensity-Modulated Radiation Therapy (IMRT) and Three Dimensional

Feasibility of the partial-single arc technique in RapidArc planning for prostate cancer treatment

Comparison of multileaf collimator and customized blocks for 3-D conformal radiotherapy of prostate cancer with six-field technique

Treatment of exceptionally large prostate cancer patients with low-energy intensity-modulated photons

Department of Radiotherapy & Nuclear Medicine, National Cancer Institute, Cairo University, Cairo, Egypt.

IMRT/IGRT Patient Treatment: A Community Hospital Experience. Charles M. Able, Assistant Professor

Statistical Analysis and Volumetric Dose for Organ at Risk of Prostate Cancer

A PRACTICAL METHOD TO ACHIEVE PROSTATE GLAND IMMOBILIZATION AND TARGET VERIFICATION FOR DAILY TREATMENT

A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer

Development of a treatment planning protocol for prostate treatments using intensity modulated radiotherapy

A Dosimetric Comparison of Whole-Lung Treatment Techniques. in the Pediatric Population

Effect of bladder filling on doses to prostate and organs at risk: a treatment planning study

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

IMRT Planning Basics AAMD Student Webinar

CyberKnife Radiotherapy For Localized Prostate Cancer: Rationale And Technical Feasibility

Dosimetric Analysis of 3DCRT or IMRT with Vaginal-cuff Brachytherapy (VCB) for Gynaecological Cancer

Feasibility of 4D IMRT Delivery for Hypofractionated High Dose Partial Prostate Treatments

Evaluation of Monaco treatment planning system for hypofractionated stereotactic volumetric arc radiotherapy of multiple brain metastases

IMRT - the physician s eye-view. Cinzia Iotti Department of Radiation Oncology S.Maria Nuova Hospital Reggio Emilia

Knowledge-Based IMRT Treatment Planning for Prostate Cancer: Experience with 101. Cases from Duke Clinic. Deon Martina Dick

IMRT for Prostate Cancer

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015

Intrafractional prostate motion during external beam radiotherapy monitored by a real-time target localization system

Evaluation of Whole-Field and Split-Field Intensity Modulation Radiation Therapy (IMRT) Techniques in Head and Neck Cancer

THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS

3-Dimensional conformal radiotherapy versus intensity modulated radiotherapy for localized prostate cancer: Dosimetric and radiobiologic analysis

Chapters from Clinical Oncology

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 6, 2016

The Effects of DIBH on Liver Dose during Right-Breast Treatments: A Case Study Abstract: Introduction: Case Description: Conclusion: Introduction

Target localization for post-prostatectomy patients using CT and ultrasound image guidance

Efficient SIB-IMRT planning of head & neck patients with Pinnacle 3 -DMPO

INTRAFRACTION PROSTATE MOTION DURING IMRT FOR PROSTATE CANCER

Evaluation of Three-dimensional Conformal Radiotherapy and Intensity Modulated Radiotherapy Techniques in High-Grade Gliomas

3D ANATOMY-BASED PLANNING OPTIMIZATION FOR HDR BRACHYTHERAPY OF CERVIX CANCER

Dosimetric effects of prone and supine positions on post-implant assessments for prostate brachytherapy

Prostate immobilization using a rectal balloon

Intensity Modulated Radiation Therapy (IMRT)

IS SMALLER BETTER? COMPARISON OF 3-MM AND 5-MM LEAF SIZE FOR STEREOTACTIC RADIOSURGERY: A DOSIMETRIC STUDY

Evaluation of the Dynamic Arc-Therapy in Comparison to Conformal Radiation Therapy in Radiotherapy Patients

INTENSITY MODULATED RADIATION THERAPY: Next Generation 3-D CRT or Distinct Form of RT?

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 2, SPRING 2005

Measure the Errors of Treatment Set-Ups of Prostate Cancer Patient Using Electronic Portal Imaging Device (EPID)

Evaluation of three APBI techniques under NSABP B-39 guidelines

Original Article. Teyyiba Kanwal, Muhammad Khalid, Syed Ijaz Hussain Shah, Khawar Nadeem

Reena Phurailatpam. Intensity Modulated Radiation Therapy of Medulloblastoma using Helical TomoTherapy: Initial Experience from planning to delivery

WHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY

The British Journal of Radiology, 82 (2009),

Would SBRT Hypofractionated Approach Be as Good? Then Why Bother With Brachytherapy?

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

The objective of this lecture is to integrate our knowledge of the differences between 2D and 3D planning and apply the same to various clinical

Silvia Pella, PhD, DABR Brian Doozan, MS South Florida Radiation Oncology Florida Atlantic University Advanced Radiation Physics Boca Raton, Florida

Utilization of cone-beam CT for offline evaluation of target volume coverage during prostate image-guided radiotherapy based on bony anatomy alignment

Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)

Intensity Modulated Radiation Therapy (IMRT)

Dosimetric impacts of endorectal balloon in CyberKnife stereotactic body radiation therapy (SBRT) for early-stage prostate cancer

Corporate Medical Policy

IMRT FOR CRANIOSPINAL IRRADIATION: CHALLENGES AND RESULTS. A. Miller, L. Kasulaitytė Institute of Oncolygy, Vilnius University

Outline. Contour quality control. Dosimetric impact of contouring errors and variability in Intensity Modulated Radiation Therapy (IMRT)

Jean Pouliot, PhD Professor and Vice Chair, Department of Radiation Oncology, Director of Physics Division

Improving plan quality and consistency by standardization of dose constraints in prostate cancer patients treated with CyberKnife

Changing Paradigms in Radiotherapy

UNIVERSITY OF WISCONSIN-LA CROSSE Graduate Studies

The Effects of DIBH on Liver Dose during Right-Breast Treatments Introduction

Additional Questions for Review 2D & 3D

Automated Plan Quality Check with Scripting. Rajesh Gutti, Ph.D. Clinical Medical Physicist

Intensity modulated radiotherapy (IMRT) for treatment of post-operative high grade glioma in the right parietal region of brain

Dose prescription, reporting and recording in intensity-modulated radiation therapy: a digest of the ICRU Report 83

Completion of Treatment Planning. Eugene Lief, Ph.D. Christ Hospital Jersey City, New Jersey USA

I. Equipments for external beam radiotherapy

Intensity Modulated Radiation Therapy (IMRT)

To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence

ART for Cervical Cancer: Dosimetry and Technical Aspects

Ritu Raj Upreti, S. Dayananda, R. L. Bhalawat*, Girish N. Bedre*, D. D. Deshpande

Radiation Therapy for Prostate Cancer. Resident Dept of Urology General Surgery Grand Round November 24, 2008

Role of Belly Board Device in the Age of Intensity Modulated Radiotherapy for Pelvic Irradiation

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

EVALUATION OF THERAPEUTIC POTENTIAL OF HEAVY ION THERAPY FOR PATIENTS WITH LOCALLY ADVANCED PROSTATE CANCER

doi: /j.ijrobp

Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy

Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer?

A treatment planning study comparing Elekta VMAT and fixed field IMRT using the varian treatment planning system eclipse

HDR vs. LDR Is One Better Than The Other?

Leila E. A. Nichol Royal Surrey County Hospital

A Thesis. entitled. based on CBCT Data Dose Calculation. Sukhdeep Kaur Gill. Master of Science Degree in Biomedical Science

NSABP PROTOCOL B-39B RTOG PROTOCOL 0413

Margins in SBRT. Mischa Hoogeman

Advanced Technology Consortium (ATC) Credentialing Procedures for 3D Conformal Therapy Protocols 3D CRT Benchmark*

A quantitative study of IMRT delivery effects in commercial planning systems for the case of oesophagus and prostate tumours

BLADDER RADIOTHERAPY PLANNING DOCUMENT

Quantification of the margin required for treating intraprostatic lesions

IGRT Solution for the Living Patient and the Dynamic Treatment Problem

Radiation Planning Index for dose distribution evaluation in stereotactic radiotherapy

Application of asi-kvcbct for Volume Assessment and Dose Estimation: An Offline Adaptive Study for Prostate Radiotherapy

Asynchronization. (aka MLC interplay effect with tumor motion)

biij Initial experience in treating lung cancer with helical tomotherapy

Performance Evaluation of Calypso (R) 4D Localization and Kilovoltage Image Guidance Systems for Interfraction Motion Management of Prostate Patients

Trina Lynd, M.S. Medical Physicist Lifefirst Imaging & Oncology Cullman, AL Tri-State Alabama, Louisiana and Mississippi Spring 2016 Meeting April

REVISITING ICRU VOLUME DEFINITIONS. Eduardo Rosenblatt Vienna, Austria

Transcription:

Medical Dosimetry, Vol. 30, No. 4, pp. 205-212, 2005 Copyright 2005 American Association of Medical Dosimetrists Printed in the USA. All rights reserved 0958-3947/05/$ see front matter doi:10.1016/j.meddos.2005.06.002 OPTIMIZATION OF COLLIMATOR PARAMETERS TO REDUCE RECTAL DOSE IN INTENSITY-MODULATED PROSTATE TREATMENT PLANNING JULIE CHAPEK, C.M.D., MATT TOBLER, C.M.D., BEAU J. TOY, M.D., CHRISTOPHER M. LEE, M.D., DENNIS D. LEAVITT, PH.D. Huntsman Cancer Hospital, University of Utah Health Science Center, Salt Lake City, UT Radiation Oncology Centers of Las Vegas, Las Vegas, NV (Received 1 March 2005; accepted 1 June 2005) Abstract The inability to avoid rectal wall irradiation has been a limiting factor in prostate cancer treatment planning. Treatment planners must not only consider the maximum dose that the rectum receives throughout a course of treatment, but also the dose that any volume of the rectum receives. As treatment planning techniques have evolved and prescription doses have escalated, limitations of rectal dose have remained an area of focus. External pelvic immobilization devices have been incorporated to aid in daily reproducibility and lessen concern for daily patient motion. Internal immobilization devices (such as the intrarectal balloon) and visualization techniques (including daily ultrasound or placement of fiducial markers) have been utilized to reduce the uncertainty of intrafractional prostate positional variation, thus allowing for minimization of treatment volumes. Despite these efforts, prostate volumes continue to abut portions of the rectum, and the necessary volume expansions continue to include portions of the anterior rectal wall within high-dose regions. The addition of collimator parameter optimization (both collimator angle and primary jaw settings) to intensity-modulated radiotherapy (IMRT) allows greater rectal sparing compared to the use of IMRT alone. We use multiple patient examples to illustrate the positive effects seen when utilizing collimator parameter optimization in conjunction with IMRT to further reduce rectal doses. 2005 American Association of Medical Dosimetrists. Key Words: Optimization, Collimator parameters, Prostate radiation therapy, IMRT. INTRODUCTION Developments in treatment planning systems have allowed a shift from conventional, nonconformal radiation therapy to highly conformal 3-dimensional (3D) planning and intensity-modulated radiation therapy (IMRT). As dose is made to be more conformal to the planning target volume (PTV), there can be a reduction of dose delivered to normal tissues. Improved dose conformality allows for the consideration of prescription dose escalation. In turn, the ability to safely achieve higher doses has lead to important milestones in the treatment of prostate cancer, where data support that dose escalation in the treatment of prostate cancer results in significantly improved prostate-specific antigen (PSA) relapse-free survival in patients with intermediate and unfavorable prognosis. 1 Along with dose escalation, it is essential to minimize the volume of normal tissues that are treated to higher doses to be able to achieve a higher therapeutic ratio. In treating the prostate, the main limiting factor is dose delivered to the rectum. This dose must not only be considered as a maximum point dose, but the entire dose-volume histogram (DVH) must be evaluated. When treating patients to doses of 70.2 75.6 Gy, it has been reported that rectal bleeding correlated to significant Reprint requests to: J. Chapek, C.M.D., Department of Radiation Oncology, Huntsman Cancer Hospital, 1950 Circle of Hope, Salt Lake City, UT 84112-5560. E-mail: Julie.chapek@hci.utah.edu increase in irradiated rectal volume. 2 In a study by Wachter et al., it was reported that grade 2 late rectal complications occurred in patients who had more than 57% of the rectum irradiated to 60 Gy. 3 Boersma et al. also showed, through analysis of irradiated volumes, that there was a significant increase in the incidence of severe rectal bleeding in patients where more than 40 50% of the rectal wall received at least 65 Gy. 4 In their preliminary report of a randomized 3D-CRT (conformal radiation therapy) dose escalation trial, Storey et al. reported that patients with 25% of rectal volume irradiated to 70 Gy developed grade 2 or higher complications. 5 Over a longer follow-up period, Huang et al. found that dose and volume were continuously interrelated variables with a significant volume effect observed at rectal doses of 60, 70, 75.6, and 78 Gy, with the risk of rectal complications increasing exponentially with increased irradiated volume. 6 This dose-volume relationship becomes increasingly significant as dose escalation occurs. Arm 2 of the current Radiation Therapy Oncology Group (RTOG) protocol for prostate cancer (#0126) recommends prostate prescription doses of 79.2 Gy (with dose inhomogeneity 7%) and provides strict guidelines for the evaluation of the DVH. With these escalated prescription doses, the need to reduce dose to rectal volumes becomes more critical. In this RTOG protocol, it is recommended that less than 15% of the rectal volume receive 75 Gy, less than 25% receive 70 Gy, less than 205

206 Medical Dosimetry Volume 30, Number 4, 2005 35% receive 65 Gy, and that no more than 50% volume receive more than 60 Gy. In an effort to reduce the volume of rectal wall, irradiated innovative techniques have been employed to better localize or stabilize the target volume at each treatment. These include intrarectal balloons, daily ultrasound, and fiducial markers implanted in the prostate. Patel et al. have reported on the use of a rectal balloon catheter resulting in a significant decrease in the volume of rectal wall receiving high dose ( 60 Gy). 7 Localizing the prostate using transabdominal ultrasound on a daily basis provides a rapid way of adjusting fields to match the target volume prior to each treatment. 8,9 Many authors have evaluated the feasibility of gold-seed fiducial markers implanted into the prostate and utilized daily to localize the prostate prior to treatment. 10 13 Better volume stabilization and/or daily volume localization will lead to increased confidence in daily treatment reproducibility. As a result of this increased confidence, physicians may decrease the amount of expansion they require around the clinical target volumes (CTVs) to create the planning target volumes (PTVs). Despite gains made in daily localization and the possible reduction in size of PTVs, prostate volumes directly abut portions of the rectum and when these volumes are expanded to create PTVs, they may actually include portions of the rectum in the high-dose regions. It has been shown that utilizing 3D-conformal planning or IMRT reduces rectal and bladder doses and therefore improves the risk-to-benefit ratio. 14 These planning techniques have also led to clinically acceptable escalation of prescription doses for prostate cancer treatment. It is therefore essential for dosimetrists to continually investigate techniques that will result in the further reduction of doses to the rectum. In this study, we examined the effect of adjusting collimator parameters. Primary collimator settings and collimator angles were adjusted relative to the geometry of the rectum, making these parameters more conformal to the rectum. We evaluated whether the adjustments of these parameters, when used in conjunction with IMRT planning, can result in a reduction of dose delivered to the rectum. METHOD Patient selection Three patients were selected for this study. Patients were chosen based on the angle of the rectum compared to a neutral collimator angle (i.e., 0, 90, 180 or 270 ). It was felt that a large angle between the rectum and the neutral collimator angle would be the condition when the maximum amount of rectum would be unshielded. This is partially due to the stair-stepped appearance of the multileaf collimator (MLC) leaves along the length of the rectum as they create the field shape required to cover the PTV (Fig. 1a). This figure also shows that there is a large volume of rectum that is shielded only by the MLC Fig. 1. Currently, the standard IMRT collimator setup procedure utilizes no optimization of collimator parameters. (A) Collimator angle is set at a neutral angle (180 ) and the primary collimators are set to the pre-set default settings outside the maximum MLC aperture. (B) Full optimization of the collimator parameters utilizes optimization of both collimator angle to match the angle of the rectum and of the primary collimators to bring jaws in as close as possible to the edge of the MLC aperture. leaves and not the primary collimators. This underlying rectal volume would receive dose during treatment in the form of transmission through the MLC leaves (which is on the order of 2.4%). Definition of rectum and rectal wall For the purpose of our study, the rectum was contoured from the inferior portion of the ischial tuberosities up to the superior aspect of the rectosigmoid flexure. The rectal wall was defined as a 2-mm-thick wall extending from 2.5 mm inferior to the most inferior aspect of PTV to 2.5 mm superior to the most superior aspect of the PTV. The rectal wall was evaluated in this study, because it extended such a small distance beyond the PTV; therefore; this volume benefited most from the study, resulting in greater reduction to rectal wall dose.

Optimization of collimator parameters J. CHAPEK et al. 207 Fig. 2. DVHs for the CTV of (A) patient A, (B) patient B, and (C) patient C show that no optimization (diamond), primary collimator setting optimization only (triangle), collimator angle optimization only (circle), and the fully optimized collimator parameter (cross) plan delivered comparable results. Plans done with 1-cm MLC leaves. Definition of target volumes In each case, the CTV was defined as the prostate and proximal 1 cm of seminal vesicles without a margin. The PTV was an expansion of 1 cm beyond the CTV in all directions except posterior, where expansion was 6 mm. The dosimetric coverage of the target volumes was defined as the 100% isodose line to cover the entire CTV and the 95% isodose line to cover the entire PTV. Evaluated variables For each patient studied the following variables were applied: 1. Collimator angle set at a neutral angle of 180. Primary collimators were set to preset default of 3 mm (in Y direction) and 5 mm (in X direction) outside the MLC leaves. (Fig. 1a). This is currently the most commonly accepted collimator parameter setup procedure. 2. Primary collimator settings optimized to bring the jaws in as close as possible to the edge of MLCs without impinging on the treatment field. This will reduce the amount of extra dose delivered as a result of transmission through the MLC leaves, as the primary collimators shield as much of the MLC area as possible. 3. Collimator angle optimized so that the angle of the collimator followed a best fit of the angle of the rectum. Primary collimator settings were allowed to follow the preset default defined above in point no. 1. Collimator angle was determined by visual

208 Medical Dosimetry Volume 30, Number 4, 2005 4. Collimator angle and primary collimator settings were fully optimized as described in point nos. 2 and 3 above, to achieve the maximum amount of shielding on the rectum while still covering the PTV (Fig. 1b). Plans were evaluated using the Varian Eclipse treatment planning system. (Varian Medical Systems, Palo Alto, CA). It was essential to ensure that all plans were executed in a similar manner. Only the evaluated individual variables were altered for each plan. Due to the computer algorithm applied within the planning system as it performs an optimization on an IMRT plan, it is never possible to achieve an identical optimization on the same patient with the same plan. To reduce this variation to a minimum, an initial plan was run in which both collimator angle and primary collimator position were fully optimized. This plan was the first one to be executed because we felt that this situation of fully optimized collimator parameters would give the best end results. In the IMRT optimization process for this plan, the optimization parameters (dose constraints and priority settings) were determined. This developed optimization parameter set was saved as an optimal constraint template. Fluence patterns were then deleted from the initial plan and the plan was re-optimized and calculated using the saved optimal constraint template. The IMRT optimization process was allowed to continue until no further improvements were registered in the planning system with each additional iteration. In the same manner, this constraint template was then applied to all subsequent planning situations. Fig. 3. Patient A. (A) DVH for rectal wall shows that the non-optimized plan (diamond) delivers the highest dose to the rectal wall compared to the plans with primary collimators optimized (triangle), collimator angle optimized (circle), and the fully optimized collimator and primary collimator positions (cross). (B) DVH comparison for the rectal wall showing only the non-optimized results (diamond) and the fully optimized collimator and primary collimator positions results (cross) illustrate a dramatic improvement in rectal sparing with the fully optimized technique. Plans done with 1-cm MLC leaves. adjustment until it best followed the slope of the rectum. This resulted in a minimized stair-step effect of the MLCs, as the angle of incidence of the leaves was more conformal to the angle of the rectum. This more conformal collimator angle also resulted in the primary collimators shielding an increased area of the MLCs, therefore reducing transmission through the leaves to the patient. Fig. 4. Patient A. The DVH for the rectal wall showing results for the non-optimized collimator parameters (diamond) and the fully optimized collimator angle and primary collimator settings (cross) with the 0.5-cm MLC leaves. Results show an improvement in rectal sparing with the fully optimized plan.

Optimization of collimator parameters J. CHAPEK et al. 209 0.5-cm MLC leaves, and the resulting DVHs would be compared. RESULTS DVH evaluation of doses delivered to the CTV revealed that for all 3 patient cases, equivalent volume coverage was achieved for each planning situation. Hence, the application of the studied variables in no way compromised CTV coverage (Fig. 2a c). Evaluation of the DVH for the rectal wall, however, shows gains in rectal sparing as a result of the optimization of the collimator parameters. For patient A, Fig. 3a illustrates that the non-optimized technique (currently the most commonly used parameter setup procedure) achieved the worst result for dose delivered to the rectal wall. Optimization of the primary collimator setting alone or optimization of the collimator angle alone achieved results that were a slight improvement in rectal wall dose over no optimization. Combination of both primary collimator setting and collimator angle optimization achieved results that were a dramatic improvement over the other techniques evaluated. This improvement was also noted with the fully optimized 0.5-cm MLC leaves, although the improvement in rectal wall dose was not as dramatic as that achieved with the 1-cm leaves (Fig. 4). For both patients A and B, the angle of the rectum deviated from the neutral collimator angle by approximately 14. DVHs for patient B show results that are similar to those of patient A (Fig. 5a and 5b, Fig. 6). For the fully optimized technique, results show a greater reduction in rectal wall dose compared to the partially Fig. 5. Patient B. (A) DVH for the rectal wall again shows that the non-optimized plan (diamond) delivers the highest dose to the rectal wall compared to the plans with primary collimators optimized (triangle), collimator angle optimized (circle), and the fully optimized collimator and primary collimator positions (cross). (B) DVH comparison for the rectal wall showing only the non-optimized results (diamond) and the fully optimized collimator and primary collimator positions results (cross) illustrates, as for patient A, the reduction in rectal dose when utilizing the fully optimized plan. Plans done with 1-cm MLC leaves. Each variable was evaluated using geometrically identical 7 field plans. As per arm 2 of RTOG protocol for prostate cancer (#0126), a prescription dose of 79.2 Gy was used for all plans in this study. Within our department, we utilize linear accelerators equipped with either 1-cm leaves or 0.5-cm leaves. It was decided that all plans would be evaluated using both the 1- and Fig. 6. Patient B. The DVH for the rectal wall showing results for the non-optimized collimator parameters (diamond) and the fully optimized collimator angle and primary collimator settings (cross) with the 0.5-cm MLC leaves. Results show an improvement in rectal sparing with the fully optimized plan.

210 Medical Dosimetry Volume 30, Number 4, 2005 Fig. 8. Patient C. The DVH for the rectal wall showing results for the non-optimized collimator parameters (diamond) and the fully optimized collimator angle and primary collimator settings (cross) with the 0.5-cm leaves. Results show an improvement in rectal sparing with the fully optimized plan. Fig. 7. Patient C. (A) DVH for the rectal wall shows that the non-optimized (diamond) plan delivers the highest dose to the rectal wall compared to the plans with primary collimators optimized (triangle), collimator angle optimized (circle), and the fully optimized collimator and primary collimator (cross) positions. (B) DVH comparison for the rectal wall showing only the non-optimized results (diamond) and the fully optimized collimator and primary collimator positions results (cross) again show the benefit gained in reducing rectal dose in fully optimized plan. The separation of the DVH curves for this patient is greater than that for patients A or B. Reductions of rectal wall dose extend well into the high-dose regions for the curve. Plans done with 1-cm MLC leaves. optimized situations, even with this small amount of collimator angle introduced into the plan. For patient A, the 0.5-cm MLC leaves also show the greatest reduction in rectal wall dose when the fully optimized collimator parameters were applied to patient B. The reduction in rectal wall dose is most impressive for patient C. The anatomy of this patient was such that the rectal wall angle deviated from the neutral collimator angle by approximately 24. This large angle allowed for a large change in both shape and conformity (as a result of the collimator angle optimization) and in MLC transmission (due to primary collimator optimization). DVHs for patient C show reductions in rectal wall dose that extend well into the high-dose regions of the curve with the fully optimized collimator parameters for both the 1-cm MLC leaves (Fig. 7) and the 0.5-cm MLC leaves (Fig. 8). Table 1 shows that the dose to the rectal wall at the 50% volume point on the DVHs is reduced in the plans with fully optimized collimator parameters for all patients. The most commonly used collimator setup procedure (non-optimized situation) provided the Table 1. Comparison of dose to rectal wall at 50% volume point of DVH Patient Dose to 50% Volume of Rectal Wall (Gy) No Optimization Dose to 50% Volume of Rectal Wall (Gy) Full Optimization Change in Dose to 50% Volume of Rectal Wall A 47.7 44.0 8% less dose with optimization B 43.4 42.0 3.3% less dose with optimization C 46.5 44.8 3.8% less dose with optimization Plans done with 1-cm MLC leaves.

Optimization of collimator parameters J. CHAPEK et al. 211 Table 2. Comparison of volume of rectal wall receiving more than 60 Gy Patient Volume Receiving 60 Gy (cm 3 ) No Optimization Volume Receiving 60 Gy (cm 3 ) Full Optimization Change in Volume of Rectal Wall Receiving 60 Gy A 4.2 4.4 4.7% reduction in rectal wall irradiated to 60 Gy with optimization B 6.4 6.6 3.0% reduction in rectal wall irradiated to 60 Gy with optimization C 3.1 3.4 9.6% reduction in rectal wall irradiated to 60 Gy with optimization Plans done with 1-cm MLC leaves. worst result in all cases analyzed. Similarly, Table 2 shows that for the fully optimized scenario there is a reduction in the volume of rectal wall receiving doses greater than 60 Gy. This is most noticeable in patient C. As stated in these previously sited studies, these reductions in rectal dose would be predicted to lead to a decrease in the probability of both long- and shortterm rectal complications. DISCUSSION This study reveals that collimator optimization strategies such as utilizing either collimator angle optimization, primary collimator position optimization, or optimization of both of these parameters simultaneously resulted in a reduction of rectal wall volume dose over the non-optimized technique. The non-optimized technique is currently the most common technique employed in radiotherapy departments, although the fully optimized situation demonstrated the best results in its ability to reduce the rectal wall dose. The implementation of any one of these optimized techniques does not require the addition of any hardware or software to most linear accelerators or treatment planning systems. These are simple techniques that utilize currently available technology and are therefore techniques that can easily be implemented in any radiotherapy department. The application of these techniques need not be limited to IMRT plans, but can be applied to 3D planning, and should result in similar reductions of dose to the rectum or any other structure of interest. CONCLUSIONS This study has shown that through the optimization of collimator angle and primary collimator positions, in conjunction with IMRT planning for prostate cancer, it is possible to reduce the dose delivered to the volume of the rectal wall. Optimization of the collimator angle allows improvement of the conformality of the MLC leaves to the rectal volumes, therefore providing increased shielding to this normal structure. Optimizing the primary collimators so that they do not extend beyond the maximum extent of the MLC opening ensures that transmission through leaves to the patient is kept to a minimum. This MLC transmission reduction is especially important in IMRT planning, where commonly, an increased number of monitor units is delivered daily and therefore an increased dose is delivered to the patient through interleaf and intraleaf leakage. The application of collimator and jaw optimization is effective in reducing dose to the rectal wall volume and could be equally applied to other radiosensitive critical structures near target volumes in various anatomic sites. REFERENCES 1. Zelefsky, M.J.; Leibel, S.A.; Gaudin, P.B.; et al. Dose escalation with three-dimensional conformal radiation therapy affects the outcome in prostate cancer. Int. J. Radiat. Oncol. Biol. Phys. 41:491 500; 1998. 2. Jackson, A.; Skwarchuk, M.W.; Zelefsky, M.J.; et al. Late rectal bleeding after conformal radiotherapy of prostate cancer. II. Volume effects and dose-volume histograms. Int. J. Radiat. Oncol. Biol. Phys. 49:685 98; 2001. 3. Wachter, S.; Gerstner, N.; Goldner, G.; et al. Rectal sequelae after conformal radiotherapy of prostate cancer: Dose-volume histograms as predictive factors. Radiother. Oncol. 59:65 70; 2001. 4. Boersma, L.J.; van den Brink M., Bruce, A.M.; et al. Estimation of the incidence of late bladder and rectum complications after highdose (70 78 Gy) conformal radiotherapy for prostate cancer, using dose-volume histograms. Int. J. Radiat. Oncol. Biol. Phys. 41:83 92; 1998. 5. Storey, M.R.; Pollack, A.; Zagars G.; et al. Complications from radiotherapy dose escalation in prostate cancer: Preliminary results of a randomized trial. Int. J. Radiat. Oncol. Biol. Phys. 48:635 42; 2000. 6. Huang, E.H.; Pollack, A.; Levy, L.; et al. Late rectal toxicity: Dose-volume effects of conformal radiotherapy for prostate cancer. Int. J. Radiat. Oncol. Biol. Phys. 54:1314 21; 2002. 7. Patel, R.R.; Orton, N.; Tome, W.A.; et al. Rectal dose sparing with a balloon catheter and ultrasound localization in conformal radiation therapy for prostate cancer. Radiother. Oncol. 67:285 94; 2003. 8. Trichter, F.; Ennis, R.D. Prostate localization using transabdominal ultrasound imaging. Int. J. Radiat. Oncol. Biol. Phys. 56:1225 33; 2003. 9. Little, D.J.; Dong, L.; Levy, L.B.; et al. Use of portal images and BAT ultrasonography to measure setup error and organ motion for prostate IMRT: Implications for treatment margins. Int. J. Radiat. Oncol. Biol. Phys. 56:1218 24; 2003. 10. Dehnad, H.; Nederveen, A.J.; van der Heide, U.A.; et al. Clinical feasibility study for the use of implanted gold seeds in the prostate as reliable positioning markers during megavoltage irradiation. Radiother. Oncol. 67:295 302;2003. 11. Welsh, J.S.; Berta, C.; Borzillary, S.; et al. Fiducial markers implanted during prostate brachytherapy for guiding conformal

212 Medical Dosimetry Volume 30, Number 4, 2005 external beam radiation therapy. Technol. Cancer Res. Treat. 3:359 64;2004. 12. Poggi, M.M.; Gant, D.A.; Sewchand, W.; et al. Marker seed migration in prostate localization. Int. J. Radiat. Oncol. Biol. Phys. 56:1248 51;2003. 13. Kitamura, K.; Shirato, H.; Shimizu, S.; et al. Registration accuracy and possible migration of internal fiducial gold marker implanted in prostate and liver treated with real-time tumor-tracking radiation therapy (RTRT). Radiother. Oncol. 62:275 81;2002. 14. Zelefsky, M.J.; Fuks, Z.; Hunt, M.; et al. High-dose intensity modulated radiation therapy for prostate cancer: Early toxicity and biochemical outcome in 772 patients. Int. J. Radiat. Oncol. Biol. Phys. 53:1111 6; 2002.