Validity and Reliability of a Glucometer Against Industry Reference Standards

Similar documents
BMJ Open. A method-comparison study regarding the validity and reliability of the Lactate Plus hand-held lactate meter

Bern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D.

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood Glucose Meter

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System

A method-comparison study regarding the validity and reliability of the Lactate Plus analyzer

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)

GlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data

Report. Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013

Quality Control of Self-Monitoring of Blood Glucose: Why and How?

Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014

Performance of three portable blood glucose monitoring devices used in a veterinary application

Comparison of Three Whole Blood Creatinine Methods for Estimation of Glomerular Filtration Rate Before Radiographic Contrast Administration

ADAG Study Group Data Links A1C Levels with Empirically Measured Blood Glucose Values - New Treatment Guidelines Will Now be Needed

Clinical Evaluation for. Embrace No Code Blood Glucose Monitoring System

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

ISO INTERNATIONAL STANDARD

A model for managing and monitoring the quality of glucometers used in a high-volume clinical setting

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

Significant Insulin Dose Errors May Occur if Blood Glucose Results Are Obtained from Miscoded Meters

Accuracy and Precision Performance of the Accu-Chek Mobile System. Introduction. Method

BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS

VALIDATION OF CONTINUOUS GLUCOSE MONITORING IN CHILDREN AND ADOLESCENTS WITH CYSTIC FIBROSIS - A PROSPECTIVE COHORT STUDY

Ultra Blue. Test Strips UNISTRIP. Charlotte, NC V2-10/25/2018 TECHNOLOGIES,LLC

FreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in Glycemic Clamp Protocols

liquicolor (AMP Buffer, IFCC) Design Verification

OGCARE is a blood glucose monitoring device for checking glucose concentration in whole blood from fingertip pricking.

Accuracy and Reliability of the Nova StatStrip Glucose Meter for Real-Time Blood Glucose Determinations during Glucose Clamp Studies

Why do we need SD LipidoCare?

Determination of hemoglobin is one of the most commonly

Week 17 and 21 Comparing two assays and Measurement of Uncertainty Explain tools used to compare the performance of two assays, including

Evaluation of the Precision and Accuracy of the INRatio 2 Prothrombin Time (PT) Monitoring System

Performance Evaluation of B. Braun Omnitest 5 Blood Glucose Monitoring System for Self-Monitoring of Blood Glucose

Clinical Chemistry / INTENSIVE INSULIN THERAPY AND GLUCOSE VALUES

ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE

(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%)

Worksheet. Worksheet. Worksheet. Worksheet. Student Performance Guide. Student Performance Guide

Basic knowledge of POCT instuments. The Tide Resort Bangsan Beach 18 June 2009

Performance Characteristics of a New Single-Sample Freezing Point Depression Osmometer

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)

Method Comparison Report Semi-Annual 1/5/2018

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

Aina Blood Monitoring System

EVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:2003 ACCURACY AND USER EVALUATION REQUIREMENTS

GlucCell TM SYSTEM USER S GUIDE Ver 2.1 CELL CULTURE GLUCOSE METER. Important Information. Intended Use. Caution. About the System

SUMMARY... 2 PLANNING OF THE EVALUATION... 4 ANALYTICAL QUALITY SPECIFICATIONS... 5

PROCEDURE. TITLE: Bedside Glucose Monitoring PC Laboratory. Issuing Department: Clinical Director Signature: Departments Involved:

NICO /14 Nipro Diagnostics, Inc.

Point-of-Care glucose analysis in neonates using modified quinoprotein glucose dehydrogenase

Technical Evaluation of Five Glucose Meters With Data Management Capabilities

Point of Care (POC) Glucose Testing Revised on: December, 2018 Reference: Policy LAB 23-A; Point of Care Nova Stat Strip Glucose Meter

3. System Accuracy Evaluation LUX Hgb test 1. Samples

A meter designed for glucose self-measurement manufactured by LifeScan, Johnson & Johnson. Report from an evaluation organised by SKUP

GB/T Translated English of Chinese Standard: GB/T

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

METHOD VALIDATION: WHY, HOW AND WHEN?

SCIEX Vitamin D 200M Assay for the Topaz System

Analytical error of home glucose monitors: a comparison of 18 systems

C RITICAL C ARE 2. Angelo Manzoni, Susanna Zanardi, Pietro Zonca Jay Johnson. Glucose biosensor for whole blood analysis

GlucCell TM SYSTEM USER S GUIDE ver 2.3 CELL CULTURE GLUCOSE METER. Important Information. Intended Use. Caution. About the System

A meter designed for glucose self-measurement manufactured by Roche Diagnostics. Report from an evaluation organised by SKUP

Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007

Determination of Ethanol in Breath and Estimation of Blood Alcohol Concentration with Alcolmeter S-D2

Professional Procedure Guide (90821B 04/2008)

Determination of Hematocrit Interference in Blood Samples Derived from Patients with Different Blood Glucose Concentrations

Clinical Performance of the TRUE2go Blood Glucose Monitoring System Exceeds Minimum Criteria for Accuracy Using ISO 15197:2013

COBAS, ACCU-CHECK INFORM, MODULAR and LIFE NEEDS ANSWERS are trademarks of Roche Roche

Human Alpha 1 microglobulin ELISA Kit

Management of Central Venous Access Devices. Blood Glucose Monitoring

Biosensor and its electrochemical transducer

RayBio Human Granzyme B ELISA Kit

A New, Wireless-enabled Blood Glucose Monitoring System That Links to a Smart Mobile Device: Accuracy and User Performance Evaluation

CoQ10(Coenzyme Q10) ELISA Kit

LIPASE liquicolor. Design Verification. Multipurpose Reagent

Relationship between glucose meter error and glycemic control efficacy

The performance of Microdot Test Strips has been evaluated both in laboratory and in clinical tests.

Point-of-Care Blood Glucose Meter Accuracy in the Hospital Setting

EVALUATING THE UNCERTAINTY OF MEASUREMENT ON BLOOD'S GLUCOSE LEVEL

BLOOD GLUCOSE MONITORING: CLINICAL NEED AND TECHNOLOGY

1 PROTOCOL. Comparison Study Summary

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats

Analysis of Hemoglobin A1c from Dried Blood Spot Samples with the Tina-quant II Immunoturbidimetric Method

Selecting a Risk-Based SQC Procedure for a HbA1c Total QC Plan

TABLE OF CONTENTS GENERAL INFORMATION... 1

Statistical Considerations: Study Designs and Challenges in the Development and Validation of Cancer Biomarkers

Rat Hemoglobin A1c (HbA1c) Kit Instructions

Title: Abbott Optium Xceed Glucose Meter Effective date: 16/05/2011. Summary of Significant Changes at this Revision. SOP updated.

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000

Summary of Significant Changes at this Revision. Items Required

Quick Reference Guide

Evolution of Analytical Performance in Portable Glucose Meters in the Last Decade

Transcription:

514315DSTXXX10.1177/1932296813514315Journal of Diabetes Science and TechnologySalacinski et al research-article2014 Original Article Validity and Reliability of a Glucometer Against Industry Reference Standards Journal of Diabetes Science and Technology 2014, Vol. 8(1) 95 99 2014 Diabetes Technology Society Reprints and permissions: sagepub.com/journalspermissions.nav DOI: 10.1177/1932296813514315 dst.sagepub.com Amanda J. Salacinski, PhD 1, Micah Alford, BS 2, Kathryn Drevets, BS 2, Sarah Hart, BS 2, and Brian E. Hunt, PhD 2 Abstract As an appealing alternative to reference glucose analyzers, portable glucometers are recommended for self-monitoring at home, in the field, and in research settings. The purpose was to characterize the accuracy and precision, and bias of glucometers in biomedical research. Fifteen young (20-36 years; mean = 24.5), moderately to highly active men (n = 10) and women (n = 5), defined by exercising 2 to 3 times a week for the past 6 months, were given an oral glucose tolerance test (OGTT) after an overnight fast. Participants ingested 50, 75, or 150 grams of glucose over a 5-minute period. The glucometer was compared to a reference instrument. The glucometer had 39% of values within 15% of measurements made using the reference instrument ranging from 45.05 to 169.37 mg/dl. There was both a proportional ( 0.45 to 0.39) and small fixed (5.06 and 0.90 mg/dl) bias. Results of the present study suggest that the glucometer provided poor validity and reliability results compared to the results provided by the reference laboratory analyzer. The portable glucometers should be used for patient management, but not for diagnosis, treatment, or research purposes. Keywords accuracy, bias, oral glucose tolerance test, reproducibility Portable glucose analyzers or glucometers are recommended by the American Diabetes Association (ADA) for selfmonitoring at home, in the field, or in clinical settings. 1 The Food and Drug Administration requires that these devices meet or exceed the accuracy requirements set forth by the International Organization for Standardization (ISO), which requires glucometers to produce measurements within 20% of the reference value at glucose concentrations above 75.68 mg/dl. 1-3 However, the ADA recommends that these devices should produce measurements within 5% of reference values. 1,3 If the measurement error of these glucometers is less than 5%, they are an appealing alternative to reference glucose analyzers, particularly for small research programs. While there have been dramatic improvements in performance regarding sample blood volume and analysis time, only modest improvements in accuracy have been reported since 1989. 4,5 More recently, studies have compared the performance of these glucometers to the ISO standard or assessed the clinical significance of the measured differences. 2,3,5,6 Few studies have precisely quantified the accuracy, reliability, and bias of these glucometers if these instruments are to be used in the research laboratory. Therefore, to address these gaps in the literature, the purpose of this study was to more precisely quantify the validity, reliability, and bias of a glucometer for biomedical research. Methods Fifteen young (20-36 years; mean = 24.5 years), moderately to highly active men (n = 10) and women (n = 5) participated in the study. Highly active individuals were defined as those who participated in physical activity at least 3 times a week for the past 6 months and were moderately active if they participated 2 to 3 times a week for the past 6 months. 7 No participants were previously diagnosed with diabetes, and all had a fasting glucose level less than 86.5 (±12) mg/dl (range, 76.4-99.9 mg/dl), as determined by the reference instrument. All subjects read and signed an informed consent. The Institutional Review Boards at Wheaton College and Northern Illinois University approved all procedures, in compliance with the World Medical Association s Declaration of Helsinki. 1 Department of Kinesiology and Physical Education, Northern Illinois University, DeKalb, IL, USA 2 Department of Applied Health Science, Wheaton College, Wheaton, IL, USA Corresponding Author: Amanda J. Salacinski, PhD, Northern Illinois University, 204 Anderson Hall, DeKalb, IL 60115, USA. Email: asalacinski@niu.edu

96 Journal of Diabetes Science and Technology 8(1) Instruments As a reference instrument to determine the validity of the Bayer Contour glucometer we used the Yellow Springs Instruments (YSI) 2300 Stat Plus Glucose and Lactate analyzer (Yellow Springs, OH). This reference laboratory analyzer uses an electrochemical probe to measure glucose concentration in whole blood and uses a membrane-bound enzyme electrochemical methodology, where D-glucose is oxidized in the presence of glucose oxidase. A new membrane was used for each data collection session. The analyzer was initially calibrated using a 180 mg/dl and 900 mg/dl solution. In addition, an automated quality control was performed in triplicate every 45 minutes using a 180 mg/dl solution. Blood samples were collected from a finger stick into 2 or 3 heparinized capillary tubes. Blood was then mixed in a micro centrifuge tube. Two 25 µl samples were sequentially aspirated and measured by the analyzer for duplicate samples with the YSI. One probe on the YSI was set to analyze whole blood glucose. Therefore, 1 value was reported for each sample, and we took the average of the duplicate samples. All values reported are plasma glucose being corrected by multiplying the whole blood glucose value by 1.11 as recommended by the International Federation of Clinical Chemistry. 8 The glucometer uses an electrochemical glucose oxidase biosensor to measure glucose concentration in a 0.6 µl sample of whole blood. A quality control solution (106.31-147.75 mg/dl) was used to ensure proper operation of the meter as recommended by the manufacturer. Samples for the first 10 participants were taken directly from the finger as instructed in the device manual. Samples for the last 5 participants were not taken directly from the finger, but from the mixed blood sample used by the YSI 2300 reference instrument. Oral Glucose Tolerance Test Participants performed an oral glucose tolerance test (OGTT) after an overnight fast. Participants ingested 50, 75, or 150 grams of glucose over a 5-minute period (eg, Sun-Dex, Fisher Healthcare, Houston, TX, or glucose tablets, Walgreens, USA). Blood samples were obtained before ingestion (0 minutes) and 10, 20, 30, 60, and 90 minutes postprandial. For the first 10 participants all samples were taken specifically as follows: (1) handheld directly from finger, (2) capillary tubes for YSI, and (3) handheld directly from finger. For the last 5 participants, all blood was collected into capillary tubes. Two sequential samples using the glucometer were taken from the blood mixed in the micro centrifuge prior to measurement by the reference instrument. The duplicate measurements from each instrument were averaged and these mean values compared to determine the accuracy of the glucometer. Data Analysis Precision was determined from at least 3 serial measurements from each instrument using calibration solutions and expressed as coefficient of variation (CV). Three methods were used to assess validity. First, we determined the clinical accuracy using the ISO 15197:2013 performance criteria by calculating the percentage of individual glucometer readings falling within ±15% mg/dl of the corresponding mean reference values above 75 mg/dl and ±15 points for values below 75 mg/dl. 9 Second, a Bland Altman plot was constructed to afford a more direct comparison of the data to that of previous studies, since this approach is typically used to assess agreement between methods. However, because fixed and proportional biases cannot be determined independently from these plots, ordinary least products (OLP) regression analysis was employed. Validity was determined from the correlation coefficient in combination with the presence and degree of bias. The degree of fixed bias was determined by the 95% confidence intervals of the y-intercept. If the confidence interval for the intercept includes a value of 0, then there is no fixed bias. Proportional bias was determined from the 95% confidence interval for the slope. If the confidence interval for the slope includes the value of 1.0, then there is no proportional bias. OLP regression provides different slopes and y-intercepts than ordinary least squares (OLS) regression because error is assumed in both glucometer and reference analyzers. Reliability was determined using OLP regression to quantify the relationship between sequential measurements for both instruments. Results Instrument Precision The CV for the glucometer was 6.6% at a glucose concentration of 110 mg/dl and 5.3% at a glucose concentration of 445 mg/dl. The CV for the reference instrument was 1.2% at 180 mg/dl. Validity As measured by our reference instrument, blood glucose concentrations during the OGTT ranged from 50 to 188 mg/ dl. Of the 166 total measurements made using the glucometer, 136 (82%) met the ISO standard. When glucometer and reference instrument blood samples were taken directly from the finger the mean difference between instruments was 8.427 mg/dl (Figure 1), thus indicating the glucometer consistently provided values greater than our reference instrument. According to the OLP analysis, the fixed bias is 7.148 mg/dl (Figure 2), an underestimation of plasma glucose. However, the confidence intervals overlap with zero. Perhaps more important, there is a significant (+13%) proportional

Salacinski et al 97 Figure 1. Mean differences for glucose measurements between the glucometer and reference instrument. Figure 3. Proportional and fixed bias of the glucose measurements. Figure 2. Actual fixed bias in glucose measurements between glucometer and reference analyzer instruments. bias as indicated by the slope of the regression line being greater than 1.0. Reliability Serial measurements by our reference instrument were strongly related as depicted in Figure 3. There was no proportional bias as the confidence interval for the slope of the relation included the line of identity. However, there was a small fixed bias, 5.123 and 0.905 mg/dl. The glucometer showed no fixed bias, but a small proportional bias (Figure 3, bottom pane). Discussion The purpose of this study was to quantify the validity, reliability, and bias of the glucometer to the benchmark glucose analyzer for biomedical research. Our study found that the portable glucometers had poor validity and reliability compared to the benchmark laboratory analyzer. Furthermore, glucose readings from the glucometer showed fixed and proportional bias. It is important to note that we chose to use OLP regression to characterize the relation between the Bayer Contour glucometer and our reference analyzer. Most studies have employed a combination of Bland Altman plots, t tests, or least squares regression to determine the degree of agreement between various glucometers and a corresponding reference analyzer. The mean difference between analyzers is determined by the interaction of any fixed and proportional bias. Therefore, the mean difference between methods, as determined through Bland Altman plots or t tests, does not solely reflect the accuracy or fixed bias of the device, but in some cases, the presence of a proportional bias or loss of linearity. As reflected in the slope of the linear relation, the use of least squares regression to characterize the level of proportional bias is skewed because all error is assigned to the dependent variable, in this case the glucometer. The use of least products regression to compare methods avoids both of these issues, allowing independent and more accurate determination of any fixed or proportional bias. 10-12 Of the glucometer readings, 82% fell within the ISO criteria for clinical accuracy. While we compared 166 separate glucometer readings to the mean reference values ranging

98 Journal of Diabetes Science and Technology 8(1) from 50 to 190 mg/dl, it must be noted that we did not strictly follow the ISO protocol. The Bland Altman plots showed an 8.427 mg/dl mean difference within subjects, which is lower than the bias reported in plasma-calibrated glucometers. 13 This difference between analyzers is determined by the interaction of any fixed and proportional bias that may exist. As revealed by the OLP analysis, the fixed bias is 7.148 mg/dl. In addition there was approximately a +13% proportional bias. While the fixed bias was not different zero (±95% CI includes zero), there was a small proportional bias. However, this bias may be as small as 5%, having little impact on the accuracy of this instrument. This is reflected by the fact that 81.2% of the readings less than 75.0 mg/dl met the ISO criteria, compared to 82% across the whole range of glucose values. The total bias, or mean difference between the glucometer and reference instrument, is equal to approximately 7% of the average glucose value. This is greater than the 5% recommended by the ADA. 1,3 Only 31% of the measurements by our glucometer met this standard. These data indicate that even though these glucometers may be suitable for clinical monitoring and insulin management, the measurement error is greater than acceptable for use in biomedical research. The reference instrument contained a small fixed bias, amounting to approximately 3.6% of the average glucose value, while the glucometer displayed a fixed bias almost twice that of the reference instrument. Though it must be noted, the bias was not statistically significant, as the confidence intervals overlapped zero. We believed the larger bias may be due to the use of separate blood drops used for the duplicate glucometer samples. Therefore, for the last 5 study participants, we used the same blood sample as the reference analyzer; however, this did not reduce the fixed bias, indicating that the fixed bias was not due to sampling technique. While there was no significant fixed bias, and only a small proportional bias, the confidence intervals were 5-fold greater for the glucometer compared to the reference instrument. So while little or no bias exists in the glucometer, there was large trial-to-trial variability in readings, perhaps too large to use in biomedical research laboratory. Over the past few decades, there have been improvements in glucometer accuracy. 4 According to the results of this study, however, using portable glucometers in a lab-based study yielded lower reliability when compared to the reference analyzer. A review of portable glucometers claimed reliable glucose reading values within 10% of reference values were only made in one-half of the analyzers tested. 5 As such, our study is the latest to confirm that many glucometers do not meet either the ISO or ADA standard. Possible errors for portable glucometers are a result of many factors, such as: improper calibration, failure to clean, not an adequate amount of blood on the strip, improper storage of strips (such as temperature, humidity, and moisture), and sometimes-patient medications. 6 Difference in the testing strips used for glucometers or proper calibration may diminish the analytical quality. Studies have revealed the variability between the test strips alone for glucometer systems is poor. 14,15 To minimize any sources of error, the following study was conducted in a controlled laboratory environment where the temperature and humidity were controlled for both the data collection and storage of strips. All testing was done on healthy active participants to mitigate the chance of drugs interfering with the results. All equipment was properly calibrated prior to testing as recommended by the manufacturer. The reference instrument only performed 1 quality control solution (180 mg/dl) in triplicate every 45 minutes instead of 2 solutions, which could be a limitation. It should be noted that the use of a single glucose concentration (180 mg/dl in triplicate every 45 minutes) for quality control of the reference instrument may not have allowed us to verify that it was operating properly at lower glucose concentrations. As one can see in Figure 1, however, the difference between the glucometer and reference instrument was consistent across the entire range of glucose values. The strips used for the portable glucometer were newly ordered and within the expiration date provided by the manufacturer. However, whether the strips in the order were all from the same lot number could not be verified. All methods performed consisted of duplicate samples. However, most users of self-monitoring blood glucose monitors (SMBG) do not take duplicate measures for an accurate reading. SMBG has surged in popularity among patients with diabetes to help them achieve their goals and is recommend by the ADA for all diabetes patients for management of the disease. The monitoring of glycemic status is meant to access the patient s physiological status and nutritional therapy. 1 The accuracy of SMBG is user dependent and also important for health care providers to monitor the users technique. 1 Researchers have claimed that confidence and accuracy of the results is dependent on how the information obtained is used. 6 The data are consistent with those of previous studies that have assessed the clinical reliability of these glucometers. 2,3,5,6 However, our data continue to show significant bias in accuracy (Figure 2, 13% proportional bias) and a relatively large imprecision (Figure 2, 5-fold larger confidence intervals). Taken together, our data suggest that glucometers may be suitable for clinical monitoring and insulin management but are not currently acceptable for use as laboratory instruments. Due to the lack of reliability and validity compared to reference lab analyzers, the results of this study indicate SMBG may only be used for patient maintenance, but this should be done with great caution. To maximize the accuracy of these devices patients must use SMBG in accordance with manufacturer recommendation. Harrison and colleagues showed that under highly controlled conditions and supplies approximately 99% of all readings of the Bayer Contour were within 15% of the reference instrument. 16 The portable glucometer was not highly correlated to the reference analyzer, as the criteria from the ADA were not met in glycemic and hyperglycemic ranges. The SMBG or portable

Salacinski et al 99 glucometer can still be employed for patient management, but laboratory and clinical measures should be used for patient diagnosis and treatment purposes, as well as for research given the significant measurement errors. Abbreviations ADA, American Diabetes Association; ISO, International Organization for Standardization; OGTT, oral glucose tolerance test; SMBG, selfmonitoring of blood glucose; YSI, Yellow Springs Instruments. Declaration of Conflicting Interests The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article. Funding The author(s) received no financial support for the research, authorship, and/or publication of this article. References 1. American Diabetes Association. Tests of glycemia in diabetes. Diabetes Care. 2003;26(S1):S106-S108. 2. Kimberly MM, Vesper HW, Caudill SP, et al. Variability among five over-the-counter blood glucose monitors. Clin Chim Acta. 2006;364(1-2):292-297. 3. Strok ADM, Kemperman H, Erkelens DW, et al. Comparison of the accuracy of the hemocue glucose analyzer with the Yellow Springs Instrument glucose oxidase analyzer, particularly in hypoglycemia. Eur J Endocrinol. 2005;153:275-281. 4. Böhme P, Floriot M, Sirveaux MA, et al. Evolution of analytical performance in portable glucose meters in the last decade. Diabetes Care. 2003;26:1170-1175. 5. Alto WA, Meyer D, Schneid J, et al. Assuring the accuracy of home glucose monitoring. J Am Board Fam Pract. 2002;15:1-6. 6. Tonyushkina K, Nichols JH. Glucose meters: a review of the technical challenges to obtaining accurate results. J Diabetes Sci Technol. 2009;3(4):971-980. 7. Kraemer WJ, Ratamess NA. Fundamentals of resistance training: progression and exercise prescription. Med Sci Sports Exerc. 2004;36:647-688. 8. D Orazio P, Burnett RW, Fogh-Andersen N, et al. Approved IFCC recommendation on reporting results for blood glucose. J Clin Chem. 2005;51(9):1573-1576. 9. International Organization for Standardization. Determination of Performance Criteria for in Vitro Blood Glucose Monitoring Systems for Management of Human Diabetes Mellitus. ISO 15197. Geneva, Switzerland: International Organization for Standardization; 2002. 10. Ludbrook J. Comparing methods of measurements. Clin Exp Pharmacol Physiol. 1997;24:193-203. 11. Ludbrook J. Statistical techniques for comparing measurers and methods of measurement: a critical review. Clin Exp Pharmacol Physiol. 2002;29:527-536. 12. Brace RA. Fitting straight lines to experimental data. Am J Physiol. 1977;233(3):R94-R99. 13. Savoca R, Jaworek B, Huber AR. New plasma referenced POCT glucose monitoring systems are they suitable for glucose monitoring and diagnosis of diabetes? Clin Chim Act. 2006;372(1-2):199-201. 14. Kristensen GBB, Christensen NG, Thue G, et al. Between-lot variation in external quality assessment of glucose: clinical importance and effect on participant performance evaluation. Clin Chem. 2005;51(9):1632-1636. 15. Louie RF, Tang Z, Sutton DV, et al. Point-of-care glucose testing: effects of critical care variables, influence of reference instruments, and a modular glucose meter design. Arch Pathol Lab Med. 2000;124(2):257-266. 16. Harrison B, Leazenby C, Halldorsdottir S. Accuracy of the CONTOUR blood glucose monitoring system. J Diabetes Sci Technol. 2011;5(4):1009-1013.