CLINICAL SCIENCES. Effect of Central Corneal Thickness, Corneal Curvature, and Axial Length on Applanation Tonometry

Similar documents
Effect of corneal parameters on measurements using the pulsatile ocular blood flow tonograph and Goldmann applanation tonometer

CLINICAL SCIENCES. Clinical Significance of Central Corneal Thickness in the Management of Glaucoma

INTRODUCTION S. HERDENER, D. HAFIZOVIC, M. PACHE, S. LAUTEBACH, J. FUNK. University Eye Hospital, Freiburg - Germany

Correlating central corneal thickness and intraocular pressure in ocular hypertension and glaucoma

Central corneal thickness determined with optical coherence tomography in various types of glaucoma

Central corneal thickness and vascular risk factors in normal tension glaucoma

Dina H. Erickson, O.D., a Denise Goodwin, O.D., a Michael Rollins, O.D., b Amber Belaustegui, O.D., c and Chad Anderson a

Role of central corneal thickness measurement in management of open angle glaucoma and glaucoma suspects in Calabar, Nigeria

Jose M. Martinez-de-la-Casa, Julian Garcia-Feijoo, Ana Fernandez-Vidal, Carmen Mendez-Hernandez, and Julian Garcia-Sanchez METHODS

CLINICAL SCIENCES. tonometry (GAT) has become the gold standard against which other tonometers

The Association of Central corneal thickness with Intra-ocular Pressure and Refractive Error in a Nigerian Population

NIIOS. Cornea Specialist, Melles Cornea Clinic, NIIOS, Rotterdam, the Netherlands Naval Hospital, Athens, Greece VASILIS S.

MEDICAL POLICY SUBJECT: CORNEAL ULTRASOUND PACHYMETRY. POLICY NUMBER: CATEGORY: Technology Assessment

Central Corneal Thickness in Primary Angle Closure and Open Angle Glaucoma

Circadian variations in central corneal thickness and intraocular pressure in patients with glaucoma

Laser in situ keratomileusis (LASIK) has

Central corneal thickness among glaucoma patients attending Menelik II Hospital, Addis Ababa, Ethiopia

CORRELATION BETWEEN INTRAOCULAR PRESSURE, PAQUIMETRY AND KERATOMETRY IN A NORMAL POPULATION

Morphological Study of Corneal Endothelium and Corneal Thickness in Pseudoexfoliation Syndrome

True IOP No Doubt Facts and Figures Figures. Ziemer Ophthalmic Systems AG a Ziemer Group Company Allmendstrasse 11 CH-2562 Port, Switzerland

Recent evidence has stressed the clinical role of the assessment

Intraocular pressure (IOP) is of fundamental importance in

The effect of corneal thickness on intraocular pressure measurement in patients with corneal pathology

Central Cornea Thickness in Glaucoma and Non- Glaucoma African Population

Kobe University Repository : Kernel

Ever since the recognition that certain forms of blindness. Central corneal thickness, tonometry, and glaucoma risk a guide for the perplexed

Correspondence should be addressed to Fusako Fujimura;

Central Corneal Thickness and its Association with Ocular Parameters

The Egyptian Journal of Hospital Medicine (January 2018) Vol. 70 (8), Page

Medical Policy An independent licensee of the Blue Cross Blue Shield Association

Central Corneal Thickness-An important variable for prognostication in Primary Open Angle glaucoma; A Kolkata based study in Eastern India

Based on the studies by Goldmann and Schmidt 1 and Ehlers. Central Corneal Thickness Correlated with Glaucoma Damage and Rate of Progression

Intraocular Pressure Measurements Referring to the Corneal Thickness in Keratoconic Eyes After Corneal Crosslinking with Riboflavin and Ultraviolet A

Interest in central corneal thickness (CCT) and glaucoma was. Central Corneal Thickness and Glaucoma in East Asian People.

Inaccuracy of Intraocular Lens Power Prediction for Cataract Surgery in Angle-Closure Glaucoma

Comparison between Pascal dynamic contour tonometer and Goldmann applanation tonometer after different types of refractive surgery

Evaluation of the Effect of Diabetes Mellitus on Corneal Biomechanics

Intro to Glaucoma/2006

Article Central Corneal Thickness In Individuals With And Without Pseudo Exfoliation Syndrome A Prospective Study

Role of Central Corneal Thickness in Circadian Intraocular Pressure Fluctuations among Patients with Primary Open Angle Glaucoma

Goldmann applanation tonometry error relative to true intracameral intraocular pressure in vitro and in vivo

Peretyagin O.A., Cand. Med. Sc., Ass. Prof.; Dmitriev S.K., Dr. Med. Sc., Prof.; Lazar Yu.M., Cand. Med. Sc.; Tatarina Yu.A., Junior Research Fellow

Comparison of Intraocular Pressure Measurements by the Ocular Response Analyzer and Goldmann Applanation Tonometer after Penetrating Keratoplasty

Ocular Rigidity Evaluation After Photorefractive Keratectomy: An Experimental Study

Comparison of IOP measurement by ocular response analyzer, dynamic contour, Goldmann applanation, and noncontact tonometry

CLINICAL SCIENCES. Clinical Comparison of the Proview Eye Pressure Monitor With the Goldmann Applanation Tonometer and the TonoPen

Intraocular pressure changes and relationship. Corneal Biomechanics After SMILE and FS-LASIK

Central Corneal Thickness (CCT) among Glaucoma and Non Glaucoma Patients in a Hospital Based Population

Trabeculectomy is an effective method for lowering

NON-INVASIVE OCULAR RIGIDITY MEASUREMENT: A DIFFERENTIAL TONOMETRY APPROACH

STUDY OF EFFECTIVENESS OF LENS EXTRACTION AND PCIOL IMPLANTATION IN PRIMARY ANGLE CLOSURE GLAUCOMA Sudhakar Rao P 1, K. Revathy 2, T.

Glaucoma: Diagnostic Modalities

Comparison of Central Corneal Thickness Measurement with Pentacam and Ultrasound

Immersion Vs Contact Biometery for Axial Length Measurement before Phacoemulsification

CLINICAL SCIENCES. Comparison of Ocular Response Analyzer Parameters in Chinese Subjects With Primary Angle-Closure and Primary Open-Angle Glaucoma

1 Birmingham and Midland Eye Centre, City Hospital, Birmingham, 2 Ophthalmic Research Group, Life and Health Sciences, Aston

Diabetes, Hyperglycemia, and Central Corneal Thickness

Effect of brimonidine on intraocular pressure in normal tension glaucoma: A short term clinical trial

LARGE DISCS WITH LARGE CUPS A DIAGNOSTIC CHALLENGE IN AFRICAN PATIENTS. Darshana Soma

Glaucoma is the leading cause of irreversible blindness in

Accuracy and Repeatability of a New Portable Ultrasound Pachymeter

Goldmann applanation tonometry compared with corneal-compensated intraocular pressure in the evaluation of primary open-angle Glaucoma

A Case of Primary Congenital Glaucoma

Comparison of Central Corneal Thickness Measurements with Pentacam, Orbscan II, and Ultrasound Pachymeter

INTRODUCTION MIKLOS SCHNEIDER 1, GABOR BORGULYA 2, ANDRAS SERES 1, ZOLTAN Z. NAGY 1, JANOS NEMETH 1

Glaucoma. Glaucoma. Optic Disc Cupping

Training Checking Vision Tonometry

In Practice. Surgical Procedures Diagnosis New Drugs

Investigation of the relationship between central corneal thickness and retinal nerve fiber layer thickness in ocular hypertension

Clinical Study Evaluation of the New Digital Goldmann Applanation Tonometer for Measuring Intraocular Pressure

Comparison of icare Tonometer with Pulsair and Tonopen in Domiciliary Work

Intraocular Pressure Reconstitution and Maintenance for Cadaveric Porcine Eyes in Simulated Wet-Lab Experience

December 2016, Volume: 8, Issue: 12, Pages: , DOI:

INTRODUCTION J. DAWCZYNSKI, E. KOENIGSDOERFFER, R. AUGSTEN, J. STROBEL. Department of Ophthalmology, University Hospital Jena, Jena - Germany

The Effect of Phacoemulsification with Posterior Chamber Intraocular Lens Implantation on Intraocular Pressure and Anterior Chamber Depth

Partial Coherence Interferometry as a Technique to Measure the Axial Length of the Eye Archived Medical Policy

THE TAMILNADU DR.M.G.R. MEDICAL UNIVERSITY CHENNAI, TAMILNADU

Clinical Study Comprasion of ICare Rebound Tonometer and Goldmann Applanation Tonometer in High Myopia

Goldmann Tonometer Prism with an Optimized Error Correcting Applanation Surface

Comparison of a New Non-Contact Tonometer with Goldmann Applanation

Central corneal thickness and normal tension glaucoma: A cross-sectional study

1/31/2018. Course Objectives. Diagnostic Testing. Optic Nerve Damage ANATOMY AND PHYSIOLOGY OF A GLAUCOMA WORK-UP/TONOMETRY TECHNICIAN: -SDP

Changes in Central Corneal Thickness over Time

Accuracy of Biometry for Intraocular Lens Implantation Using the New Partial Coherence Interferometer, AL-scan

An Ultra-High-Speed Scheimpflug Camera for Evaluation of Corneal Deformation Response and Its Impact on IOP Measurement

Effects of Intracorneal Ring Segment on Corneal Biomechanics in Keratoconic Eyes. Abstract

The Relative Frequency and Risk Factors of Primary Open Angle Glaucoma and Angle Closure Glaucoma

Correlation between presence of primary irisand cilliary body cysts and intraocular pressure

Spontaneous Intraocular Pressure Reduction in Normal-Tension Glaucoma and Associated Clinical Factors

Saemah Nuzhat Zafar, Sorath Noorani Siddiqui, Ayesha Khan 1

9/25/2017 CASE. 67 years old On 2 topical meds since 3 years. Rx: +3.0 RE LE

IOL Power Calculation for Children

Author s Affiliation. Original Article. Comparison of Biometry in Phakic and Dense Modes. Muhammad Suhail Sarwar. Unaiza Mariam

Comparison of Intraocular Lens Power Calculation Using the Binkhorst and SRK Formulae: A Clinical Study

The Egyptian Journal of Hospital Medicine (October 2018) Vol. 73 (9), Page

Complication and Visual Outcome after Peadiatric Cataract Surgery with or Without Intra Ocular Lens Implantation

Glaucoma surgery with or without adjunctive antiproliferatives in normal tension glaucoma: 2 Visual field progression

Comparison between Pentacam HR and Orbscan II after Hyperopic Photorefractive Keratectomy

Transcription:

CLINICAL SCIENCES Effect of Central Corneal Thickness, Corneal Curvature, and Axial Length on Applanation Tonometry Markus Kohlhaas, MD; Andreas G. Boehm, MD; Eberhard Spoerl, PhD; Antje Pürsten, Dipl-Ing (FH); Hans J. Grein, MD; Lutz E. Pillunat, MD Objective: To evaluate the effect of central corneal thickness (CCT), corneal curvature, and axial length on applanation tonometry in an in vivo study. Methods: In a masked, prospective clinical trial, we examined 125 eyes of 125 patients scheduled for cataract surgery. Corneal curvature was measured by means of keratometry and axial length by A-scan ultrasonography. By cannulating the anterior chamber before surgery, intraocular pressure (IOP) was set to, 35, and 5 mm Hg in a closed system by means of a water column. After measuring thickness, the IOP was measured with an applanation tonometer. Pearson product moment correlations and multiple linear regression analyses were performed, and significance levels were evaluated by the paired, 2-tailed t test. Results: The difference between measured and real IOP was significantly dependent (P.1) on CCT. The associations between IOP and corneal curvature or IOP and axial length were not statistically significant (P=.31). The association between IOP reading and CCT is shown in the Dresdner correction table, which illustrates an approximately 1-mm Hg correction for every 25-µm deviation from a CCT of 55 µm. The correction values were positive as thickness decreased and negative as thickness increased. Conclusions: Central corneal thickness significantly affects IOP readings obtained by applanation tonometry according to the Goldmann principle. A correction of IOP readings by considering CCT according to the Dresdner correction table might be helpful for determining an accurate IOP value. Arch Ophthalmol. 6;124:471-476 Author Affiliations: Department of Ophthalmology, University Carl Gustav Carus, Dresden (Drs Kohlhaas, Boehm, Spoerl, and Pillunat), and Department of Optometry, University of Applied Sciences, Jena (Ms Pürsten and Dr Grein), Germany. Dr Kohlhass is now with the Augenklinik, St. Johannes-Hospital, Dortmund, Germany. INTRAOCULAR PRESSURE (IOP) REPresents a fundamental factor of ocular health and disease. Intraocular pressure is important not only in the diagnosis and management of glaucoma but also in the assessment of the postoperative course of all intraocular surgical interventions. For almost 5 years, applanation tonometry according to the Goldmann method has been the gold standard for the evaluation of IOP. 1,2 Intraocular pressure measurements obtained with the use of a Goldmann applanation tonometer are affected by central corneal thickness (CCT), corneal curvature, and axial length. 3,4 Normally hydrated, thicker corneas lead to higher readings and thinner corneas, to lower readings. The clinical significance of these observations has been demonstrated in studies 5-8 that showed a lower CCT in some cases of normaltension glaucoma and a higher CCT in cases of ocular hypertension and some cases of refractory glaucoma. Empirical studies 6,7,9-11 suggest a correction factor for applanation IOP readings of.19 to 1 mm Hg per -µm deviation from the average CCT. Those studies were based on comparisons of individuals and groups of patients with different CCTs without knowledge of the actual IOP. To our knowledge, only 1 study, published by Ehlers et al, 12 used a manometric, controlled closed system in which the actual IOP was known. However, that study included only a small series of eyes. Therefore, the aim of the present study was to examine the effect of CCT, corneal curvature, and axial length on Goldmann applanation tonometry in a manometric, controlled closed system in an in vivo model with a larger number of subjects. METHODS In a masked, prospective clinical trial, we examined 125 eyes of 125 consecutive patients scheduled for cataract surgery. Thirty-three of the patients were men and 92 were women, with a mean±sd age of 72.9±13.2 years (range, 18-91 years) (Figure 1). No eye had undergone previous intraocular surgery or had an abnormality other than cataract. All subjects received an ophthalmologic clinical examination, 471

4 7 6 5 4 19 21 22 23 24 25 26 27 28 29 31 32 33 Axial Length, mm Figure 4. Distribution of axial length in 125 patients. 25 35 4 45 5 55 6 65 7 75 8 85 9 Age, y Figure 1. Distribution of age in 125 patients. 46 48 5 5 54 56 58 6 6 64 66 68 7 7 Figure 2. Distribution of central corneal thickness (CCT) in 125 patients. 7. 7. 7. 7.4 7.5 7.6 7.7 7.8 7.9 8. 8. 8. 8. 8.4 Corneal Curvature, mm Figure 3. Distribution of corneal curvature in 125 patients. including slitlamp evaluation of the anterior segment. Patients with any corneal abnormality, such as scars or Fuchs dystrophy, were excluded. All included patients provided informed consent before entering the study, which was approved by the institutional ethics committee of the medical department at University Carl Gustav Carus, Dresden, Germany. Keratometry was performed preoperatively using a keratometer, and axial length was measured by A-scan ultrasonography (Sonomed 25; Technomed, Baesweiler, Germany). After the administration of peribulbar anesthesia, an eyelid retractor was placed and a cannula was passed into the anterior chamber at the temporal limbus. The CCT was measured by the first investigator (M.K.) using an ultrasonic pachymeter. The.8-mm opening of the cannula permitted access to the aqueous humor. The cannula was connected to an adjustable saline (balanced salt solution) reservoir. The IOP was directly calibrated by means of a water column at 3 different IOP levels (, 35, and 5 mm Hg). After each IOP level was reached, the cannula was closed after 1 minute to maintain a closed system during the actual IOP measurements. The precorneal film was stained with fluorescein solution, and the IOP was measured with a Perkins applanation tonometer (Haag-Streit USA, ) at each IOP level by the second investigator (A.G.B.). To maintain masking, the first investigator measured the CCT and the second measured the IOP without knowing the CCT. All variables (IOP, corneal curvature, axial length, and CCT) were measured 3 times and the means were calculated. Pearson product moment correlations were calculated between the IOP readings and the CCT, corneal curvature, and axial length. In addition, multiple linear regression analysis was performed for the IOP readings, implementing CCT, corneal curvature, axial length, patient age and sex, arterial hypertension, and diabetes mellitus into the model. Significance levels were evaluated by the paired, 2-tailed t test. P.5 was considered statistically significant. All statistical analyses were performed with SPSS software, version 11. (SPSS Inc, Chicago, Ill). All data are presented as mean±sd unless indicated otherwise. RESULTS Mean CCT was 569±44 µm (range, 462-75 µm) (Figure 2), corneal curvature was 7.72±.27 mm (range, 7.7-8.32 mm) (Figure 3), and axial length was 23.62±2.5 mm (range, 18.84-32.93 mm) (Figure 4). At mm Hg, the IOP readings varied from 16 to 27 mm Hg (mean,.86±1.99 mm Hg); at 35 mm Hg, these 472

ranged from 31 to 42 mm Hg (mean, 35.74±1.97 mm Hg); and at 5 mm Hg, from 45 to 55 mm Hg (mean, 5.5±2.1 mm Hg). We found a strong correlation between IOP and CCT at all IOP levels: at mm Hg (r 2 =.83; P.1), 35 mmhg(r 2 =.78;P.1),and5mmHg(r 2 =.75;P.1). We found no correlation between IOP and corneal curvature at mm Hg (r 2 =.12; P=.41), 35 mm Hg (r 2 =.18; P=.37), and 5 mm Hg (r 2 =.9; P=.87), and no correlation between IOP and axial length at mm Hg (r 2 =.9; P=.66), 35 mm Hg (r 2 =.9; P=.86), and 5 mm Hg (r 2 =.8; P=.35). We cannot rule out that there might be a difference in K values (from the keratometry readings) preoperatively and perioperatively. However, if such an effect exists, the difference would be small and most likely cause an offset but not affect the correlation significantly. Therefore, this effect would be negligible with regard to the multivariate analysis. At all IOP levels, patient age and sex, arterial hypertension, and diabetes mellitus had no significant effect on IOP readings. The correlations between IOP and CCT, corneal curvature, and axial length at 35 mm Hg are shown in Figure 5. The correlation between CCT and the difference between real IOP and IOP measured by applanation tonometry at 35 mm Hg is shown in the Figure 6. Multiple regression analysis showed similar equations describing the relationship between IOP and CCT for the 3 different IOP levels. Implementing the data of all 3 different IOP levels (, 35, and 5 mm Hg), multiple regression analysis revealed the equation IOP=(.423 CCT) 23.28. (The equations for all 4 regression lines are given in the legend to Figure 7.) Using this equation, we calculated correction values for applanation IOP readings for different CCTs (Table), which we termed the Dresdner correction table. Using our equation, IOP is approximately at a CCT of 55 µm. A residual analysis using Q- plots showed that the linear approach is appropriate. Nonlinear models were also calculated, which did not improve the fit (R 2 ) significantly. Therefore, the linear approach seems to be the most appropriate. A deviation of ± µm from a CCT of 55 µm resulted in a measurement error of ±.4 mm Hg. In other words, a deviation from 55 µm of ±25 µm in CCT was related to a difference of approximately ±1 mm Hg. When we used the equation IOP=(.423 CCT) 23.28, the corrected IOPs showed a better goodness of fit to the actual IOP compared with the uncorrected IOP readings. The correspondence between actual IOP and uncorrected IOP readings is shown in Figure 8.As shown, 49.4% of the measured values were in the clinically relevant range of ±1.5 mm Hg of the actual IOP, and 5.6% of the values showed a larger deviation. After correcting the IOP readings according to our formula, 92% of the corrected IOP values were in the clinically relevant range of ±1.5 mm Hg of the actual IOP, and only 8% of the values showed a larger deviation (Figure 9). COMMENT Intraocular pressure measurements by applanation tonometry were linearly correlated with CCT at different IOP levels. We found no correlation between IOP readings and cor- A B C IOP, mm Hg IOP, mm Hg IOP, mm Hg 44 42 4 38 36 34 32 4 5 6 7 8 44 42 4 38 36 34 32 7. 7.2 7.4 7.6 7.8 8. 8.2 8.4 Corneal Curvature, mm 44 42 4 38 36 34 32 18 22 24 26 28 32 34 Axial Length, mm Figure 5. Correlation between intraocular pressure and central corneal thickness (CCT) (A), corneal curvature (B), and axial length (C) in 125 patients. The regression line and 95% confidence intervals are shown. 473

6 Table. Dresdner Correction Table Showing the Dependence of the Applanation IOP Reading on CCT IOP Difference at 35 mm Hg, mm Hg 4 2 2 4 6 8 4 5 6 7 8 Figure 6. Correlation between corneal thickness (CCT) and the difference between real intraocular pressure (IOP) and measured IOP. The regression line and 95% confidence intervals are shown. Correction, mm Hg 5 4 3 2 1 mm Hg 35 mm Hg 5 mm Hg All 1 2 3 4 5 6 7 45 5 55 6 65 7 Figure 7. Regression lines describing the relationship between the correction of intraocular pressure (IOP) and central corneal thickness (CCT) for the 3 different IOP levels. At an IOP of mm Hg, the regression line was IOP=(.415 CCT) 22.721; at 35 mm Hg, IOP=(.399 CCT) 21.929; and IOP=(.474 CCT) 26.8) at 5 mm Hg. Implementing the data of all 3 IOP levels (, 35, and 5 mm Hg), the regression line was IOP=(.423 CCT) 23.28. Correction Value, mm Hg 475 3.19 5 2.13 525 1.7 55.2 575 1.4 6 2. 625 3.16 65 4.21 675 5.27 7 6.33 Abbreviations: CCT, central corneal thickness; IOP, intraocular pressure. neal curvature or axial length. To our knowledge, ours is the largest study to prove the correlation between IOP readings obtained by applanation tonometry and CCT using an in vivo model in which the real IOP was known and the actual deviation could be measured. Various studies 4,6,7,9-11 examining the correlation between IOP readings by applanation tonometry and CCT have reported a measurement error ranging from.19 to 1 mm Hg per -µm deviation from the average CCT. In the present study, the measurement error was.4 mm Hg per -µm deviation from 55 µm. Those studies, 4,6,7,9-11 in contrast to ours, used eyes previously classified as normal or as having ocular hypertension on the basis of IOP readings. A portion of the ocular hypertensive eyes actually may have had a normal IOP but a false high IOP reading owing to a thick cornea. This might lead to an underestimation of the effect of CCT on IOP measurement and might be responsible for the discrepancy between our results and those of population studies. In a study by Ehlers et al, 12 a manometric, controlled closed system was used to examine the correlation between CCT and IOP measured by applanation tonometry in 29 patients. Ehlers et al reported an error of ±.71 mm Hg between real IOP and IOP measured by applanation tonometry per -µm difference in CCT. We obtained a smaller correction factor of ±.4 mm Hg per -µm increment. This difference might be caused by the sample sizes (29 patients vs 125) and by the fact that, in contrast to our study, corneal curvature affected IOP readings in the study by Ehlers et al. In addition, real and measured IOPs were the same at a thickness of 5 µm in the study by Ehlers et al; in our study, they were the same at a thickness of 55 µm. This is an offset of µm between studies, which is most likely caused by the use of different techniques for measuring CCT. Ehlers et al used an optical pachymeter, which provides lower readings than does ultrasound pachymetry (our method). Orssengo and Pye 13 introduced a mathematical model that examined the influence of corneal properties on measurements of IOP obtained by applanation tonometry. Their calculations were based on a normal CCT of 5 µm. They reported good agreement with the data of Ehlers et al. However, at lower IOP levels, the calculated model showed lower correction values, suggesting that the correction factor obtained by Ehlers et al was too high. This is in good agreement with our results. Feltgen and coworkers 14 attempted in vivo measurements with direct intracameral IOP readings and observed no correlation between CCT and IOP. They concluded that no recalculation of IOP according to CCT was necessary. 14 However, in contrast to Ehlers et al and the present study, Feltgen et al used an open system rather than a closed one. Because an open system allows the aqueous humor to leave the eye, IOP is lowered during the measurement. This might explain the difference in their results. 474

7.5 4.5 6. 3. 4.5 3. 1.5 IOP, mm Hg 1.5. IOP, mm Hg. 1.5 1.5 3. 4.5 45 5 55 6 65 7 75 mm Hg 35 mm Hg 5 mm Hg Figure 8. Correspondence between actual intraocular pressure (IOP) and uncorrected IOP readings. Of the measured values, 49.4% were in the clinically relevant range of ±1.5 mm Hg of the actual IOP, and 5.6% of the values showed a larger deviation. CCT indicates central corneal thickness. In contrast to other studies, 5,6,11,15 we could not detect an effect of corneal curvature and axial length on applanation IOP readings. In addition, we examined 125 consecutive patients scheduled for cataract surgery; therefore, the distributions of corneal curvature, CCT, and axial length were random and not influenced by the investigators. Compared with the empirical studies by Mark et al 16 and Mark, 17 our study population did not show a wide scattering of corneal curvature (meancorneal curvature, 7.72±.27 mm; range, 7.7-8.32 mm) or axial length (meanaxial length, 23.62±2.5 mm; range, 18.84-32.93 mm). This distribution may have caused the lack of correlation between IOP readings, corneal curvature, and axial length. Therefore, a study with a similar design but with a wider scattering of these measurements seems necessary to determine whether corneal curvature or axial length has any effect on IOP measurements obtained by applanation tonometry. However, even if such effects exist, according to our data they would be minimal in comparison with the effect of CCT. The mean CCT in our study of 569±44 µm (range, 46-75 µm) was higher compared with earlierreported measurements in the general population. 9,15 Mean CCT has been reported to be 5±29 µm for optical pachymetry and 544±34 µm for ultrasound pachymetry 7 and may depend on the technique used. Ultrasound pachymetry, which gives higher readings than does optic pachymetry, was used to determine CCT in our study. In addition, our study population was older (mean age, 72.9±13.2 years), and CCT tends to be higher in older patients. 4 In conclusion, the present study demonstrates that thin corneas lead to an underestimation and thick corneas, to an overestimation of applanation IOP. Corneal curvature and axial length do not have a significant effect 3. 4.5 45 5 55 6 65 7 75 on IOP readings obtained by applanation tonometry. A deviation in CCT of 25 µm away from a measurement of 55 µm results in a difference of approximately 1 mm Hg between real IOP and IOP measured by applanation tonometry. Using the equation IOP=(.423 CCT) 23.28, we can provide evidence for goodness of fit. In the present study, 92% of the corrected IOP values were within the clinically relevant range of ±1.5 mm Hg, and only 8% showed larger deviations. Because CCT affects IOP readings obtained by applanation tonometry, we suggest that IOP readings be corrected by CCT according to the Dresdner correction table to get a good approximation of the real IOP in each patient. However, further studies are needed to validate the scale and to demonstrate its value for the management of glaucoma. Submitted for Publication: September 5, 3; final revision received July 15, 5; accepted September 12, 5. Correspondence: Markus Kohlhaas, MD, Augenklinik, St. Johannes-Hospital, Johannesstr 9-17, Dortmund, Germany (Markus.Kohlhaas@jojo-dortmund.de). Financial Disclosure: None. REFERENCES mm Hg 35 mm Hg 5 mm Hg Figure 9. After correcting the intraocular pressure (IOP) readings according to the formula ( IOP=[.423 CCT] 23.28), in which CCT indicates central corneal thickness, 92% of the corrected IOP values were in the clinically relevant range of ±1.5 mm Hg of the actual IOP, and only 8% of the values showed a larger deviation. 1. Goldmann H, Schmidt T. Über Applanationstonometrie. Ophthalmologica. 1957; 134:221-242. 2. Goldmann H, Schmidt T. Weiterer Beitrag zur Applanationstonometrie. Ophthalmologica. 1961;141:441-456. 3. Whitacre MM, Stein R. Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol. 1993;38:1-. 475

4. Whitacre MM, Stein RA, Hassanein K. The effect of corneal thickness on applanation tonometry. Am J Ophthalmol. 1993;115:592-596. 5. Bron AM, Creuzot-Garcher C, Goudeau-Boutillon S, d Athis P. Falsely elevated intraocular pressure due to increased central corneal thickness. Graefes Arch Clin Exp Ophthalmol. 1999;237:2-224. 6. Damji KF, Munger R. Influence of central corneal thickness on applanation intraocular pressure. J Glaucoma. ;9:5-7. 7. Doughty MJ, Zaman ML. Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol. ; 44:367-48. 8. Ehlers N, Hansen FK. Central corneal thickness in low-tension glaucoma. Acta Ophthalmol (Copenh). 1974;52:74-746. 9. Shah S, Chatterjee A, Mathai M, et al. Relationship between corneal thickness and measured intraocular pressure in a general ophthalmology clinic. Ophthalmology. 1999;6:2154-216.. Shah S. Accurate intraocular pressure measurement the myth of modern ophthalmology [editorial]? Ophthalmology. ;7:185-187. 11. Stodtmeister R. Applanation tonometry and correction according to corneal thickness. Acta Ophthalmol Scand. 1998;76:319-324. 12. Ehlers N, Bramsen T, Sperling S. Applanation tonometry and central corneal thickness. Acta Ophthalmol (Copenh). 1975;53:34-43. 13. Orssengo GJ, Pye DC. Determination of the true intraocular pressure and modulus of elasticity of the human cornea in vivo. Bull Math Biol. 1999;61: 551-572. 14. Feltgen N, Leifert D, Funk J. Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings. Br J Ophthalmol. 1; 85:85-87. 15. Wolfs RCW, Klaver CCV, Vingerling JR, Grobbee DE, Hofman A, de Jong PT. Distribution of central corneal thickness and its association with intraocular pressure: the Rotterdam Study. Am J Ophthalmol. 1997;123:767-772. 16. Mark HH, Robbins KP, Mark TL. Axial length in applanation tonometry. J Cataract Refract Surg. 2;28:54-56. 17. Mark HH. Corneal curvature in applanation tonometry. Am J Ophthalmol. 1973; 76:223-224. Call for Papers Archives of Ophthalmology will publish articles on men s health and ophthalmology in conjunction with a JAMA theme issue on the same topic in November 6. Manuscripts received by June 1, 6, will have the best chance for consideration for this theme issue. 476