APPLICATIONS TN µm Core-Shell Particle. (continued on page 2) Page 1 of 8. For additional technical notes, visit

Similar documents
APPLICATIONS TN Dramatically Improve Existing 5 µm and 3 µm Fully Porous Methods with Kinetex 5 µm Core-shell Technology.

Core-Shell Technology for Proteins and Peptides

APPLICATIONS TN A Comparison of Various Kinetex C18 Phases USP: L1

APPLICATIONS Improving Intact Biogeneric Protein Separations with Aeris WIDEPORE Core-Shell Columns

APPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology

Aeris. Precision Engineered Core- Shell Particles for Ultra-High Resolution BioSeparations. Aeris PEPTIDE. Aeris WIDEPORE

For Far. Greater. UHPLC Column Performance You Must Have Kinetex Core-Shell UHPLC Columns.

Enantiomeric and Diastereoisomeric Resolutions of Chiral Triazole Fungicides using Lux Polysaccharide-Based Chiral Stationary Phases

Columns The FIRST and ONLY sub-2 µm core-shell columns on the market

Preparative Chromatography Meets UHPLC

A Fast, Accurate, and Sensitive Quantitiation of Vitamin C in Plasma by Impact Protein Precipitation and Kinetex 5 µm Core-Shell XB-C18 HPLC Columns

APPLICATIONS TN Analytical Method

SIMPLIFIED CHIRAL SEPARATIONS. Lux Cellulose-1. Guaranteed Alternative. to CHIRALCEL OD-H. Exceptional Quality Sensibly Priced

Run 1 Run 2 TDP PLP TDP PLP. 3.0e5 2.8e5 2.2e5. 2.6e5 2.4e5. 2.2e5 2.0e5. Intensity, cps. 1.8e5 1.6e5 1.4e5 1.2e5 1.0e5 4.0e4. 2.0e e5 1.

APPLICATIONS TN Determination of Sugars in Animal Feed using HPLC-ELSD

Figure 1. Core-Shell Particle Technology

Protect Your Mass Spec

NEW. Kinetex. Biphenyl TMS TMS TMS TMS. Selectivity that a C18 just can t give you!

Finally, a Solid Phase Extraction product for Drugs of Abuse

APPLICATIONS TN Materials and Methods

Yes You Can! New! Amazing EVO C18. Core-Shell Technology

Core-Shell Technology. 11 HPLC Myths. Uncovered. & Why You Must Have Kinetex Core-Shell HPLC Columns

Labs Across the World Are Advancing Their Protein Precipitation

Phospholipid Removal Solutions

APPLICATIONS TN Rapid Analysis of Hop Acids in Beer using Strata -X Solid Phase Extraction and Kinetex 2.6 µm Core-Shell Technology Column

Rapid Analysis of Synthetic Cannabinoids and their Metabolites in Urine using Solid Phase Extraction and LC/MS/MS. JWH018-4-Hydroxypentyl metabolite

Am I getting the very best value from my UHPLC analyses?

Exceed your Purification Goals. Guaranteed. Phenomenex l WEB:

PAIN PANEL. Research and Solutions

CANNABIS TESTING ACCESSORIES

Extended Application Note

METHAMPHETAMINE & AMPHETAMINE

Ultimate Core Performance to Maximize Your Investment

Impurity Profiling of Carbamazepine by HPLC/UV

Maximizing Efficiency Using Agilent InfinityLab Poroshell 120 Columns

ACE. For increased polar retention and alternative selectivity. Alternative selectivity for method development

Awesome! Unconventionally. Kinetex EVO C18 HPLC/UHPLC Core-Shell Columns 2.6. Core-Shell Technology

APPLICATIONS TN A Sensitive Extraction Method for Nicotine and Metabolites from Urine using Simplified Liquid Extraction (SLE)

ACE C18-AR Use the Power!

APPLICATION TN Materials and Methods Reagents and Chemicals

ACE Excel UHPLC Columns

Shuguang Li, Jason Anspach, Sky Countryman, and Erica Pike Phenomenex, Inc., 411 Madrid Ave., Torrance, CA USA PO _W

Prep Columns. and Bulk Media. and Bulk Media

The high efficiency of sub-2 µm UHPLC columns combined with the low pressure and high speed of monolith columns.

Effective use of Pharmacopeia guidelines to reduce cost of chromatographic analysis for Fluticasone propionate

ACE C18-AR. A C18 bonded phase with unique selectivity. Guaranteed reproducibility. Exceptional bonded phase stability

SecurityGuard. Column Protection for UHPLC, HPLC to PREP. A Universal Guard Cartridge System. Increases HPLC Column Lifetime, Guaranteed!

Thermo Scientific Hypersil BDS Columns

Choosing Preparative Columns

A New Era in SFC Column Technology

Phenyl-Hexyl. UHPLC Columns. lternate, complementary selectivity to C18 and C8 bonded phases

Hyper-fast & Super-rugged

Comparison of a UPLC Method across Multiple UHPLC Systems

Lux. Five Distinct Polysaccharide Phases to Resolve 92 % of Your Enantiomers* Lux Cellulose-1. Lux Cellulose-3. Lux Cellulose-2.

Core-Shell / Solid core Particles: A Practical Substitute to Sub-2 Micron Particles: An overview

GlycanPac AXR-1 Columns

Keep It Clean. Solid Phase Extraction Solutions for Drugs of Abuse Analysis. Neutral Drugs of Abuse Basic Drugs of Abuse

NEW Kinetex F5. HPLC/UHPLC Core-Shell Columns

InertSustain Phenylhexyl. A New Second Choice Column to Change Chromatographic Behavior

COMBO ULTIMATE HPLC/UHPLC. Kinetex + Luna Omega NEW. Gain Incredible Performance, Versatile Selectivities and Upgrade Throughput

Ultra Columns. also available. ordering note

SIMPLIFIED CHIRAL SEPARATIONS. FIVE Complementary Chiral Phases. for Greater Selectivity. Resolve 92 %

2D-LC as an Automated Desalting Tool for MSD Analysis

Agilent Technologies Prep LC Columns

Increasing resolution using longer columns while maintaining analysis time Advantages of the wide power range of the Agilent 1290 Infinity LC System

Maximizing chromatographic peak capacity with the Agilent 1290 Infinity LC system

Ultra Columns HPLC COLUMNS

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

SUMMARY, CONCLUSION & RECOMMENDATIONS

Analytical Method Development for USP Related Compounds in Paclitaxel Using an Agilent Poroshell 120 PFP

ApplicationNOTE EXACT MASS MEASUREMENT OF ACTIVE COMPONENTS OF TRADITIONAL HERBAL MEDICINES BY ORTHOGONAL ACCELERATION TIME-OF-FLIGHT.

Ultra High Performance Liquid Chromatograph. Nexera X2 C196-E079

Analysis of Common Sweeteners and Additives in Beverages with the PerkinElmer Flexar FX-15 System Equipped with a PDA Detector

USP Method Transfer of Ziprasidone HCl from HPLC to UPLC

HPLC. Varian Dynamax Preparative HPLC Columns

LC-MS/MS Method for the Determination of Tenofovir from Plasma

Effective Surface Area (m 2 /g) Pore Size (Å)

A New HILIC/RP Mixed-Mode Column and Its Applications in Surfactant Analysis

Analytical and Preparative SFC Columns

Thermo Scientific. GlycanPac AXR-1. Column Product Manual. P/N: April, Part of Thermo Fisher Scientific

PAIN PANEL. Application Guide. Screen 41 compounds in under 4 minutes

High Throughput Extraction of Opiates from Urine and Analysis by GC/MS or LC/MS/MS)

YMC-Actus. Columns for Semi-preparative HPLC. Selectivity Durability Loadability

Dikma Technologies Inc.

Selectivity Comparison of Agilent Poroshell 120 Phases in the Separation of Butter Antioxidants

Gerry Hendrickx Regional Sales Manager Central Europe. Developments in Waters Column Chemistries : BEH Technology

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

Determination of Benzoate in Liquid Food Products by Reagent-Free Ion Chromatography

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine

TENOFOVIR TABLETS: Final text for addition to The International Pharmacopoeia (June 2010)

Sepax Technologies, Inc.

APPLICATIONS TN Rapid, High Resolution Analysis of Aflatoxin Extracts from Peanut Butter Using Kinetex Core-Shell Technology and Strata SPE

Application Note. Small Molecule Pharmaceuticals. 2-Ethyl-2-phenylmalonamide Ethosuximide Primidone Carbamazepine. Carbamazepine-

HPLC to UHPLC Transfer of USP Method for Amlodipine Besylate Using the Agilent 1290 Infinity II LC

Separation of Macrocyclic Lactones (Avermectins) on FLARE C18 MM & FLARE C18+ Columns

Doubling the throughput of long chromatographic methods by using a novel Dual LC workflow

Transferring a Method for Analysis of DNPH-Derivatized Aldehydes and Ketones from HPLC to UHPLC

Converting a CHP Method for Insulin to Agilent Poroshell 120 Columns

RITONAVIRI COMPRESSI RITONAVIR TABLETS. Final text for addition to The International Pharmacopoeia (July 2012)

Transcription:

Kinetex 5 µm Core-Shell Columns Deliver Significant Improvements in Chromatographic Efficiency and Resolution versus Traditional Fully Porous 5 µm Columns Philip J. Koerner and Jeff Layne Phenomenex, Inc., 411 Madrid Ave., Torrance, CA 951 USA Kinetex 5 µm core-shell technology columns provide chromatographers a simple solution for dramatically improving the performance of their methods developed on 5 µm fully porous columns. This newly introduced core-shell media delivers backpressures of a fully porous 5 µm particle at efficiencies equal to or better than a fully porous 3 µm particle. Without the need for extensive method development, replacing the fully porous 5 µm column with the Kinetex 5 µm core-shell column results in improved chromatographic resolution and sensitivity. In addition, the lower backpressure can provide many benefits such as longer column lifetime, higher throughput, and increased system compatibility. Introduction Over the last several years, UHPLC/HPLC columns that can provide previously unheard of column efficiencies, in excess of 25, plates/meter, have dominated the market. Smaller particle diameter (<2 µm) versions of the traditional 5 and 3 µm media based on fully porous silica are now widely available, but due to the inherent increase in pressure that accompanies this decrease in particle diameter, UHPLC systems capable of operating in the range of 6 to 12 bar (9 to 17,5 psi) must be used to fully realize the performance benefits of the columns packed with these sub-2 µm particles. More recently, 2.6 µm core-shell particles, which consist of a non-porous silica core surrounded by a thin layer of porous silica (Figure 1), have been introduced and widely adopted. These 2.6 µm core-shell particles can deliver UHPLC-like efficiencies and chromatographic performance, but in contrast to fully porous sub-2 µm silica particles, core-shell particle columns can be utilized on conventional HPLC systems. This can extend benefits of UHPLC performance to a wider audience without necessitating a major capital expenditure in new UHPLC systems. This technical note focuses on the most recently introduced 5 µm core-shell particle technology and the efficiency and resolution benefits it provides. Materials and Methods Unless stated otherwise, all chemicals and standards were obtained from Sigma Chemical Co. (St. Louis, Missouri). Standards for the analysis of paroxetine and its impurities were obtained from the European Directorate for the Quality of Medicines and Healthcare (EDQM). Solvents were purchased from EMD (San Diego, California). The Kinetex 5 μm XB-C18 and C18 columns were obtained from Phenomenex (Torrance, California). Non-Phenomenex brand columns were purchased from the original equipment manufactures as noted in the figures. The column dimensions and running conditions for the various assays are detailed in the figure captions. Chromatograms in Figure 3 were obtained using an Agilent 12SL HPLC system equipped with a micro flow cell. Chromatograms for Figures 4 and 6 were obtained using an Agilent 11 HPLC system equipped with a standard flow cell, and chromatograms in Figure 5 were obtained using an Agilent 11 HPLC system which had been optimized to reduce extra-column dead volume. Data was collected using ChemStation software (Agilent, Santa Clara, California). Results and Discussion We have recently released a core-shell media, Kinetex 5 μm, which generates the same backpressure as traditional fully porous 5 μm columns (Figure 2), while providing chromatographic efficiencies in the range of 18, plates/meter. This is about 2x greater than the typical efficiencies achieved with fully porous 5 µm columns (ca. 1, plates/meter, Figure 3) and comparable to the efficiencies attained with fully porous 3 µm columns (ca. 16, plates/meter, Figure 3). Thus, the new Kinetex 5 μm media is able to deliver efficiencies that are as good, or better, than fully porous 3 μm media at the typical operating pressures of a 5 μm column. The Kinetex 5 µm core-shell particle is scaled up directly from the original Kinetex 2.6 µm core-shell particle, maintaining the same ratio of core diameter to core-shell particle diameter (ρ =.73). Figure 1. Kinetex 2.6 µm Core-Shell Particle. Ratio (ρ) of solid core diameter to coreshell particle diameter remains the same for the new Kinetex 5 µm coreshell particle. 1.9 µm Solid Core.35 µm Porous Shell 2.6 µm Core-Shell Particle (continued on page 2) Page 1 of 8

Figure 2. Backpressures Delivered by 5 µm Core-shell Particle Pressure Comparison Keeping all of the conditions constant and simply replacing the fully porous 5 µm column with a Kinetex 5 µm C18 column, we see an immediate benefit of higher efficiency expressed by the Kinetex column. The resolution between paroxetine and the unidentified degradant has increased from.98 to 2.23, an increase of 127%. The increase in resolution cannot be fully attributed to the increase in column efficiency however, but also due to a small difference in selectivity. Note that while the overall selectivities of the core-shell and fully porous 5 µm columns are very similar, the retention on the Kinetex column is reduced slightly. This is due to the reduced surface area of the core-shell particle. F-P 5 µm A F-P 5 µm B Kinetex 5 µm F-P 5 µm C F-P 5 µm D F-P 3 µm E F-P 3.5 µm F F-P 3 µm G F-P 3 µm B Figure 3. Kinetex 5 µm Core-Shell Columns Deliver Efficiencies Fully Porous 3 µm Columns Efficiency on Agilent 12SL 2 18 16 It should also be noted that operating these columns under the same conditions on a conventional HPLC system (Agilent 11) generates the same pressure, so this increase in efficiency and resolution does not require changes to the system or investing in a higher pressure UHPLC system. Figure 4. Improved Resolution and Sensitivity for Paroxetine Stability-indicating Assay on Kinetex 5 µm Core-shell Versus Fully Porous 5 µm Column on standard HPLC system a. Agilent ZORBAX 5 µm XDB-C18 25 x 4.6 mm Plates per meter 14 12 1 8 6 4 2 F-P 5 µm H F-P 5 µm A F-P 5 µm F F-P 3 µm B F-P 3.5 µm A Kinetex 5 µm 16 14 12 1 8 6 4 a 2 1 2 min b c Height = 12 d R S =.98 b c App ID 2766 Figure 4 illustrates how the increased efficiency provided by Kinetex 5 µm core-shell columns can significantly improve chromatographic resolution for a complex separation. Here a stability-indicating method for paroxetine, an anti-depressant drug, was run using a 25 x 4.6 mm column packed with fully porous 5 µm C18 particles. In this method, one must be able to quantify four principle degradants, labeled a-d. In addition, there is also an unidentified degradant which elutes between degradant c and the paroxetine API peak that must be resolved. Zooming on the region around the base of the API peak, we can see that the resolution between paroxetine and the unidentified degradant was.98 using the fully porous 5 µm column. 16 14 12 1 b. Phenomenex Kinetex 5 µm XB-C18 25 x 4.6 mm 8 6 4 2 Conditons for both columns Column: Kinetex 5 µm C18 ZORBAX 5 µm XDB-C18 Dimensions: 25 x 4.6 mm Mobile Phase: A: TFA/Water/THF (5/9/1) B: TFA/Acetonitrile/THF (5/9/1 1 4 8 1.5 ml/min Flow Rate: Temperature: 4 C Detection: UV @ 295 nm Sample: 1. Impurity A a 4. Unidentified 2. Impurity B 5. Paroxetine 3. Impurity C 6. Impurity D b c d 1 2 min Height = 165 R S = 2.23 b c App ID 2767 Agilent and ZORBAX are registered trademarks of Agilent Technologies, Inc (Palo Alto, CA, USA). Phenomenex is not affiliated with Agilent Technologies. Comparative separations may not be representative of all applications. Page 2 of 8

In Figure 5 we illustrate another example of the benefits of the Kinetex 5 µm core-shell media. In this method, the resolution between the two degradant peaks is 1.29 on the Agilent ZORBAX 5 µm SB-C18 column. Under the identical operating conditions, and at the same backpressure, the Kinetex 5 µm C18 column provides a 22 % increase in resolution. Simply by replacing the fully porous 5 µm column with the Kinetex core-shell 5 µm column, and without any additional method development or optimization, the chromatographic separation has been improved. This can be invaluable from a validation perspective since in most cases, simply changing from one column to a similar column (same length, particle size, bonded phase), a full re-validation of the method is not necessary. You might just need to perform an equivalency study depending upon the regulations of your particular company and/or group. Figure 5. Improved Resolution and Sensitivity for Mupirocin Degradation Analysis with Kinetex 5 µm versus Fully-Porous 5 µm Column on optimized HPLC system a. Agilent ZORBAX 5 µm SB-C18 15 x 4.6 mm 2 1 BP = 12 bar R S = 1.29 6 8 1 12 14 min App ID 2785 For methods that have a critical pair (or several critical pairs) of compounds to resolve, it can be difficult to meet system suitability requirements for resolution. In such cases one may find that instrument-to-instrument or column batch-to-batch variability can cause the method to fail suitability, resulting in valuable analysis time spent troubleshooting to resolve the issue. In such cases, simply switching to a Kinetex 5 µm coreshell column may provide a sufficient increase in resolution to overcome the variability that may cause sensitive methods to sometimes fail suitability. In the example illustrated in Figure 5, if additional resolution was required (and assuming the increase in analysis time was not a barrier) a longer 25 x 4.6 mm Kinetex 5 µm core-shell column could be used. b. Phenomenex Kinetex 5 µm XB-C18 15 x 4.6 mm 3 2 1 BP = 12 bar R S = 1.57 22% Increase in R S App ID 2796 6 8 1 12 14 min Conditions for all columns: Columns: Kinetex 5 µm XB-C18 ZORBAX 5 µm SB-C18 Dimension: 15 x 4.6 mm Mobile Phase: A: Water with.1% TFA B: Acetonitrile with.1% TFA 2 1 7 Flow Rate: 1.2 ml/min Temperature: Ambient Detection: UV @ 21 nm Sample: Mupirocin degradants Agilent 11 was optimized with a Core-Shell Performance Enhancement Kit (AQ-8892). Agilent and ZORBAX are registered trademarks of Agilent Technologies, Inc (Palo Alto, CA, USA). Phenomenex is not affiliated with Agilent Technologies. Comparative separations may not be representative of all applications. Page 3 of 8

The final example of the improved chromatographic performance possible with Kinetex 5 µm is shown in Figure 6. This is a stabilityindicating assay for degradants of the proton-pump inhibitor, omeprazole. Here we ran the same sample on Kinetex 5 µm core-shell column and a column made with an older generation of fully porous silica, Thermo-Keystone HYPERSIL 5 µm ODS. What is immediately obvious in the chromatograms is the significant improvement in peak shape for the basic omeprazole peak on the core-shell column (Figure 6A). Zooming out to fullscale (Figure 6B), one can really appreciate the difference in performance between the two columns. With the y-axis scaled the same, the effect of the excessive peak tailing on the intensity of the omeprazole peak is apparent. The reason for this vast difference in performance is related to the nature of the silica that is used to make the particles. Older generations of HPLC columns were based on Type A silica that contains a higher concentration of metal ions (>1 ppm) such as iron or aluminum, which activate free silanols on the silica surface, making them more acidic. These silanols can then interact with basic analytes via ion-exchange, resulting in significant peak tailing. Kinetex core-shell particles are manufactured from high purity Type B silica, which is much lower in metal content (<< 5 ppm); this results in much less secondary silanol interactions with basic analytes and much sharper, more symmetrical peaks, which improves analyte quantitation. Figure 6. Improved Peak Shape for Basic Analytes on Kinetex 5 µm XB-C18 versus HYPERSIL 5 µm ODS 6A - Zoom view a. Phenomenex Kinetex 5 µm XB-C18 25 x 4.6 mm 6B - Full view a. Phenomenex Kinetex 5 µm XB-C18 25 x 4.6 mm 286 8 Tybe-B Base-Deactivated Silica 2 6 4 2 285 b. Thermo-Keystone HYPERSIL 5 µm ODS 25 x 4.6 mm 8 6 4 2 1 2 3 4 min Tybe-A Silica 1 2 3 min Conditons for both columns: Column: Kinetex 5 µm C18 Hypersil 5 µm ODS Dimensions: 25 x 4.6 mm Mobile Phase: A: 2 mm Potassium Phosphate, ph 7 B: 5/5 Metthanol Acetonitrile 5 4 7 1.2 ml/min Flow Rate: Temperature: Ambient Detection: UV @ 28 nm Sample: Omeprazole degradants Hypersil is a registered trademark of Thermo Hypersil-Keystone (Waltham, Ma, USA). Phenomenex is not affiliated with Thermo Hypersil-Keystone. Comparative separations may not be representative of all applications. App ID 2794 App ID 285 1 1 2 3 min b. Thermo-Keystone HYPERSIL 5 µm ODS 25 x 4.6 mm 287 2 1 1 2 3 min Conditons for both columns: Column: Kinetex 5 µm C18 HYPERSIL 5 μm ODS Dimensions: 25 x 4.6 mm Mobile Phase: A: 2 mm Potassium Phosphate, ph 7 B: 5/5 Methanol/Acetonitrile 5 4 7 1.2 ml/min Flow Rate: Temperature: Ambient Detection: UV @ 28 nm Sample: Omeprazole degradants App ID 287 App ID 286 Page 4 of 8

Conclusion The new Kinetex 5 μm core-shell columns operate at pressures that are consistent with typical fully porous 5 µm columns, but with efficiencies that are equal to, or greater than, fully porous 3 μm columns. For methods currently using fully-porous 5 μm columns, Kinetex 5 μm core-shell columns will provide improved chromatography, in terms of improved efficiency, resolution, and sensitivity, without the need for extensive method development. Ordering Information 5 µm Analytical Columns (mm) SecurityGuard ULTRA Cartridges 5 x 2.1 3/pk 5 x 4.6 1 x 4.6 15 x 4.6 25 x 4.6 3/pk SecurityGuard ULTRA Cartridges XB-C18 B-465-AN AJ-8782 B-465-E D-465-E F-465-E G-465-E AJ-8768 C18 B-461-AN AJ-8782 B-461-E D-461-E F-461-E G-461-E AJ-8768 PFP B-462-AN AJ-8787 B-462-E D-462-E F-462-E G-462-E AJ-8773 Phenyl-Hexyl B-463-AN AJ-8788 B-463-E D-463-E F-463-E G-463-E AJ-8774 for 2.1 mm ID for 4.6 mm ID SecurityGuard ULTRA cartridges require holder for Part No. AJ-9. SecurityGuard 5 µm Axia Packed Preparative Columns (mm) ULTRA Cartridges 5 x 21.2 1 x 21.2 15 x 21.2 25 x 21.2 ea XB-C18 B-465-P-AX D-465-P-AX F-465-P-AX G-465-P-AX AJ-9145 C18 B-461-P-AX D-461-P-AX F-461-P-AX G-461-P-AX AJ-9145 PFP B-462-P-AX D-462-P-AX F-462-P-AX G-462-P-AX AJ-9146 Phenyl-Hexyl B-463-P-AX D-463-P-AX F-463-P-AX G-463-P-AX AJ-9147 for 21.2 mm ID SecurityGuard PREP 21.2 mm ID cartridges require holder, Part No. AJ-8223. If Kinetex core-shell columns do not provide at least an equivalent separation as compared to a competing column of the same particle size, similar phase, and dimensions, return the column with comparative data within 45 days for a FULL REFUND. Terms and Conditions Subject to Phenomenex Standard Terms & Conditions, which may be viewed at www.phenomenex.com/termsandconditions. Trademarks SecurityGuard and Axia are trademarks of Phenomenex. Kinetex is a registered trademark of Phenomenex. Waters and Symmetry are registered trademarks of Waters Corporation. Agilent and ZORBAX are registered trademarks of Agilent Technologies, Inc. HYPERSIL is a registered trademark of Thermo Hypersil-Keystone. Phenomenex is in no way affiliated with Agilent Technologies or Thermo Hypersil-Keystone. Dimensions and chromatographic conditions are the same for all columns unless otherwise noted. Comparative separations may not be representative of all applications 212 Phenomenex, Inc. All rights reserved. Page 5 of 8

Australia t: 2-9428-6444 f: 2-9428-6445 au Austria t: 1-319-131 f: 1-319-13 anfrage@phenomenex.com Belgium t: 2 53 415 (French) t: 2 511 8666 (Dutch) f: +31 ()3-2383749 be Canada t: (8) 543-3681 f: (31) 328-7768 Denmark t: 4824 848 f: +45 481 6265 Finland t: 9 4789 63 f: +45 481 6265 France t: 1 3 9 21 1 f: 1 3 9 21 11 france Germany t: 621-5883- f: 621-5883-11 anfrage@phenomenex.com India t: 4-312 24 f: 4-312 2411 india Ireland t: 1 247 545 f: +44 1625-51796 eire Italy t: 51 6327511 f: 51 6327555 italia Luxembourg t: +31 ()3-24187 f: +31 ()3-2383749 nl Mexico t: 1-8-844-5226 f: 1-31-328-7768 tecnicomx@phenomenex.com The Netherlands t: 3-24187 f: 3-2383749 nl New Zealand t: 9-478951 f: 9-478952 nz Norway t: 81 2 5 f: +45 481 6265 Puerto Rico t: (8) 541-HPLC f: (31) 328-7768 Sweden t: 8 611 695 f: +45 481 6265 United Kingdom t: 1625-51367 f: 1625-51796 uk United States t: (31) 212-555 f: (31) 328-7768 All other countries: Corporate Office USA t: (31) 212-555 f: (31) 328-7768 www.phenomenex.com Phenomenex products are available worldwide. For the distributor in your country, contact Phenomenex USA, International Department at international@phenomenex.com TN9662812_W Page 6 of 8