Test-Retest Reliability of the CSV-1000 Contrast Test and Its Relationship to Glaucoma Therapy

Similar documents
Assessing the Reliability, Discriminative Ability, and Validity of Disability Glare Tests

T he retinal ganglion cells of different sizes have distinct

Perimetric testing is used clinically to detect visual field

Visual acuity versus letter contrast sensitivity in early cataract

Specific deficits of flicker sensitivity in glaucoma and ocular hypertension

Spontaneous Intraocular Pressure Reduction in Normal-Tension Glaucoma and Associated Clinical Factors

Early Treatment of Congenital Unilateral Cataract Minimizes Unequal Competition

TUMBLING E RESOLUTION PERIMETRY IN GLAUCOMA

STANDARD AUTOMATED PERIMETRY IS A GENERALLY

Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection

Test-retest variability in visual field testing using frequency doubling technology

SUPPLEMENTARY INFORMATION Perceptual learning in a non-human primate model of artificial vision

CLINICAL SCIENCES. Contrast Sensitivity and Glare Disability After Radial Keratotomy and Photorefractive Keratectomy

Fluctuation of Intraocular Pressure and Glaucoma Progression in the Early Manifest Glaucoma Trial

Characteristics of Frequency-of-Seeing Curves in Normal Subjects, Patients With Suspected Glaucoma, and Patients With Glaucoma

Contrast Sensitivity and Reading: Assessment and Reliability with the Reading Explorer (REX) Test

Detection of true change in clinical status is a fundamental

MODERN DIAGNOSTICS AND TREATMENT OF GLAUCOMA

53 year old woman attends your practice for routine exam. She has no past medical history or family history of note.

INTRODUCTION AND LITERATURE REVIEW. Trans Am Ophthalmol Soc 2007;105:

Effect of brimonidine on intraocular pressure in normal tension glaucoma: A short term clinical trial

Fluctuations on the Humphrey and Octopus Perimeters

CLINICAL SCIENCES. Screening for Glaucoma With Frequency-Doubling Technology and Damato Campimetry

Investigative Ophthalmology & Vision Sciences MSc Course. Glaucoma Module. Visual Field Reliability Indices. David Henson 2014.

Sensitivity and specificity of new GDx parameters Colen TP, Tang NEML, Mulder PGH and Lemij HG Submitted for publication CHAPTER 7

CORRELATING OF THE VISUAL FIELD INDEX WITH MEAN DEVIATION AND PATTERN STANDARD DEVIATION IN GLAUCOMA PATIENTS

Bilateral Refractive Amblyopia Treatment Study

Visit Type (Check one)

21st Century Visual Field Testing

Reimbursement for Contrast Sensitivity Testing. Prepared for

Correlation of Blue Chromatic Macular Sensitivity with Optic Disc Change in Early Glaucoma Patients

MOVE IT OR LOSE IT: THE ROLE OF KINETIC VISUAL FIELDS

ADVISORY OPINION OF THE CODE OF ETHICS

NEPTUNE RED BANK BRICK

Vanishing Optotype acuity: repeatability and effect of the number of alternatives

Glaucoma: diagnosis and management (large print version)

The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation

CLINICAL COMPARISON OF THREE CONTRAST SENSITIVITY DEVICES. Michelle T. Valella. I q ~~

Method for comparing visual field defects to local RNFL and RGC damage seen on frequency domain OCT in patients with glaucoma.

CAN SECOND EYE.CATARACT EXTRACTION BE JUSTIFIED?

Common Causes of Vision Loss

Influence of multiple sclerosis, age and degree of disability, in the position of the contrast sensitivity curve peak

Reproducibility of Nerve Fiber Layer Thickness Measurements by Use of Optical Coherence Tomography

Misleading Statistical Calculations in Faradvanced Glaucomatous Visual Field Loss

glaucoma and ocular hypertension

Spectral sensitivity and color discrimination changes in glaucoma and glaucoma-suspect patients

Elevated intraocular pressure (IOP) is a major risk factor for

Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma

Ganglion cell complex scan in the early prediction of glaucoma

ALTHOUGH EARLY TREATment

3/16/2018. Perimetry

eye as a camera Kandel, Schwartz & Jessel (KSJ), Fig 27-3

BETAGAN Allergan Levobunolol HCl Glaucoma Therapy

SUPPLEMENTARY INFORMATION

Factors Influencing Normal Perimetric Thresholds Obtained Using the Humphrey Field Analyzer

CHAPTER 10 NON-CONVENTIONAL PERIMETRY

Balance between pattern and flicker sensitivities in

Changing expectations about speed alters perceived motion direction

Science & Technologies

Random measurement error in visual acuity measurement in clinical settings

Comparison of Visual Field Measurement with Heidelberg Edge Perimeter and Humphrey Visual Field Analyzer in Patients with Ocular Hypertension

Eye care practitioners (ECPs) assess the vision of

STRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma. Module 3a GDx

A contrast paradox in stereopsis, motion detection and vernier acuity

Definitions. Indications for Perimetry. Indications for Perimetry. Purposes of Perimetry. Indications for Perimetry 3/4/2015

Assisting in Ophthalmology. Copyright 2011, 2007, 2003, 1999 by Saunders, an imprint of Elsevier Inc. All rights reserved.

Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography

VISUAL FIELDS. Visual Fields. Getting the Terminology Sorted Out 7/27/2018. Speaker: Michael Patrick Coleman, COT & ABOC

The Evolution of Fundus Perimetry

N umerous techniques have shown vascular abnormalities

PSYCHOPHYSICS OF READING: VI. THE ROLE OF CONTRAST IN LOW VISION 1

THE CHRONIC GLAUCOMAS

CHAPTER 13 CLINICAL CASES INTRODUCTION

The Role of the RNFL in the Diagnosis of Glaucoma

RETINAL NERVE FIBER LAYER

Meeting the UK driving vision standards with reduced contrast sensitivity

UNCORRECTED PROOF. Ana Barrio & Beatriz Antona & María Cinta Puell. Keywords ETDRS. Repeatability. Visual acuity. Mesopic. High contrast.

Evolving glaucoma management True diagnostic integration for the preservation of vision

Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma

This is an Open Access document downloaded from ORCA, Cardiff University's institutional repository:

Optic neuritis (ON) is a clinical syndrome characterized

LOW VISION VISD241. MODULE LEADER: DR G WALSH B.Sc. OPHTHALMIC DISPENSING

Role of Central Corneal Thickness in Circadian Intraocular Pressure Fluctuations among Patients with Primary Open Angle Glaucoma

Intro to Glaucoma/2006

A genetic basis has been postulated for

NICE guideline Published: 1 November 2017 nice.org.uk/guidance/ng81

Effect of corneal parameters on measurements using the pulsatile ocular blood flow tonograph and Goldmann applanation tonometer

ADVANCED DIAGNOSTIC TECHNIQUES

VERTEPORFIN IN PHOTODYNAMIC THERAPY STUDY GROUP

Variability of Automated Visual Fields in Clinically Stable Glaucoma Patients

Interobserver Reliability of the Teller Acuity Card Procedure in Pediatric Patients

Goals. Glaucoma PARA PEARL TO DO. Vision Loss with Glaucoma

Noel de Jesus Atienza, MD, MSc and Joseph Anthony Tumbocon, MD

Electrodiagnostics Alphabet Soup

Medicem Institute, Kamenné Žehrovice, Czech Republic 2. Medical Dpt., Medicem Prague, Czech Republic

T he lamina cribrosa is a sieve-like membrane in the

Glaucoma surgery with or without adjunctive antiproliferatives in normal tension glaucoma: 2 Visual field progression

Selective changes of sensitivity after adaptation to simple geometrical figures*

Reproducibility of Scanning Laser Doppler Flowmetry in the Rat Retina and Optic Nervehead

Transcription:

Test-Retest Reliability of the CSV-1000 Contrast Test and Its Relationship to Glaucoma Therapy Glenn N. Pomerance* and David W. Evansf Purpose. To determine the relationship between changes in contrast sensitivity, if any, after glaucoma therapy and the test-retest reliability of the CSV-1000 contrast sensitivity test. Methods. Patients with primary open angle glaucoma (N = 16) were retrospectively evaluated to determine changes in visual function, as measured by contrast sensitivity, after beta-blocker therapy. A control group (N = 24) of normally sighted patients was tested and retested on contrast sensitivity. For the control group, the coefficients of repeatability (95% confidence interval for test-retest variability [COR]) were calculated for each spatial frequency. The CORs were compared to the changes in vision found after therapy in the patients with glaucoma. Results. The group with glaucoma showed a significant improvement (P <.01) in contrast sensitivity at all spatial frequencies. The test-retest variance for normals, as measured by the COR, was smaller than the mean differences in contrast sensitivity before and after therapy at all spatial frequencies, expect 18 cyc/deg. Conclusions. Visual function in patients with glaucoma, as measured by contrast sensitivity, does improve after beta-blocker therapy. Further, the CSV-1000 is a clinically reliable tool for monitoring these changes. Invest Ophthalmol Vis Sci. 1994;35:3357 3361. V^ontrast sensitivity testing is becoming widely used as a routine clinical tool. 1 Considerable interest has recently been shown in using contrast sensitivity to track the progression of ocular pathology, particularly glaucoma. Interestingly, several studies have shown improvements in central vision after both medical and surgical glaucoma therapy. Nordmann et al conducted a 3- year study in which they used contrast sensitivity in conjunction with other measures to track the progression of glaucomatous vision loss. They concluded that contrast sensitivity was not useful for monitoring visual changes in patients proven to have glaucoma because the changes in visual fields did not correlate well with changes in contrast sensitivity. 2 Their data did, however, show improvements in contrast sensitivity in certain patients after laser trabeculoplasty. Others have also found improvements in central vision after pharmacologic therapy. Tytla and colleagues showed in- From the *Plaza Eye Center, Chattanooga, Tennessee, and the ^Department of Visual Sciences, Indiana University School of Optometry, Bloomington, Indiana. Submitted for publication June 9, 1993; revised January 19, 1994; accepted March 15, 1994. Proprietary interest category: N (GNP); I, C (DWE). Reprint requests: Dr. David W. Evans, Indiana University School of Optometry, 800East Atwater Avenue, Bloomington, IN 47405. creases in temporal sensitivity in a subgroup of patients with ocular hypertension after beta-blocker therapy. 3 Most recently, Piltz and colleagues demonstrated significant improvement in spatial contrast sensitivity in patients with normal tension glaucoma after therapy using a calcium-channel blocker. 4 These studies suggest that improvements in central vision do occur in conjunction with glaucoma therapy and that using contrast sensitivity testing to track patients over time could be helpful in providing proper therapy for patients with glaucoma. To determine the usefulness of a specific contrast test for tracking a disease, the test-retest reliability of the particular test must be examined. Bland and Altman 5 suggested that for normally distributed data, the coefficient of repeatability (COR: 1.96 times the standard deviation of the test-retest difference) provides a useful metric for test reliability because it describes the 95% confidence interval for the variability in testretest data. Reeves et al 6 further suggested that a decision criterion could be established for each test in which the COR could be used to determine the minimum change necessary to indicate a significant change in vision on a particular test. Here we determined changes in contrast sensitivlnvestigative Ophthalmology & Visual Science, August 1994, Vol. 35, No. 9 Copyright Association for Research in Vision and Ophthalmology

3358 Investigative Ophthalmology & Visual Science, August 1994, Vol. 35, No. 9 ity in patients with glaucoma after beta-blocker therapy. We also determined the test-retest reliability of the CSV-1000 contrast sensitivity instrument in terms of the COR. We compared the CORs for each spatial frequency to any changes in vision found before and after beta-blocker therapy in patients with glaucoma. SUBJECTS AND METHODS Subjects Research followed the tenets of the Declaration of Helsinki, and all subjects consented to participate only after the nature and possible consequences of the study were explained. A retrospective study of patients with primary open angle glaucoma was conducted at the Plaza Eye Center in Chattanooga, Tennessee. Before and after beta-blocker therapy contrast sensitivity scores, using either timolol, levobunolol, or betaxolol, were evaluated. For inclusion in the retrospective sample, patients had to have significant visual field loss as measured by the Humphrey Visual Field Testing Instrument (Model 630) and analyzed by STATPAC 30-2 (Humphrey Instruments, San Leandro, CA) program. Patients had to have been tested by contrast sensitivity at least twice between January 1, 1991 and September 1, 1992, have 20/40 or better visual acuity, and be free of significant progressive disease other than glaucoma. Further, there could be no change in patients' visual fields or cup-disc ratios between the before and after therapy visits, nor could there be any changes in patients' medications. Thirty-five eyes met these criteria. The sample was further refined to include only one eye per patient and to include only patients whose test-retest intervals matched those of the normally sighted group (i.e., between 1 and 5 months). (If both eyes from the same patient met the original criteria, then only the right eye was used in the final retrospective sample.) The mean age was 69.98 years old (7.18 years SD) and the mean testretest interval was 2.68 months (.901 months SD). The final sample included test-retest data on 16 eyes. Contrast sensitivity scores of the group undergoing glaucoma therapy were not evaluated until after the final sample was determined. Test-retest subjects were solicited from among normally sighted visitors at the Plaza Eye Center. (These normal subjects were solicited from the escorts of patients with cataract who were scheduled for examination before and after surgery.) All subjects were free of ocular disease as determined by one author (GNP). Twenty-four subjects were recruited. The mean age was 63.9 years (12.17 years SD), and the mean test-retest interval was 2.65 months (.588 months SD). Methods At the Plaza Eye Center, every patient who either has or is suspected of having glaucoma is tested for con- FIGURE 1. CSV-1000 contrast sensivity testing instrument, trast sensitivity upon entering the examination room before being seated in the examination chair and before dilation. The test is conducted by a technician who has no knowledge of the previous contrast sensitivity score, if there was any, or of the status of the patient's glaucoma medication. Each patient was led into one of four examination rooms by one of three ophthalmic technicians. Acuity and contrast sensitivity were tested monocularly. Acuity was assessed with an American Optical (Buffalo, NY) slide projection system. Contrast sensitivity was assessed with the CSV-1000E Contrast Testing Instrument (VectorVision Dayton, OH) shown in Figure 1. On the second test visit, the same testing procedure was used. No attempt was made to regulate which ophthalmic technician or which examination room was used for each subject retest. The same testing technicians and instructions were used for both the normally sighted group and the patients with glaucoma. Contrast Sensitivity Testing The CSV-1000 provides a fluorescent luminance source that retro-illuminates a translucent chart. The instrument houses a series of photocells that automatically monitor and calibrate the instrument light level to 85 candelas per square meter ± 0.1 log unit. The testing light levels were measured at the beginning and during the course of the study. The testing light levels were all within ±0.1 log unit on each spatial

Test-Retest Reliability of the CSV-1000 Contrast Test and Glaucoma 3359 TABLE l. Pretherapy, Posttherapy, and Difference Scores for Patients With Glaucoma (n = 16) Spatial Test-Retest Frequency Pretherapy Posttherapy Difference 3 cyc/deg 1.203 ± 0.269 1.389 ± 0.211 0.186 ± 0.246 6 cyc/deg 1.427 ± 0.381 1.594 ± 0.340 0.167 ± 0.298 12 cyc/deg 1.038 ± 0.458 1.333 ± 0.414 0.295 ± 0.382 18 cyc/deg 0.709 ± 0.374 0.901 ± 0.455 0.191 ± 0.350 Values are mean ± SD. frequency from instrument to instrument and from the start to the end of the study. At the testing distance of 8 feet, the translucent chart presents four spatial frequencies: 3, 6, 12, and 18 cyc/deg. Each spatial frequency is presented on a separate row of the test. Each row presents 17 circular patches 1.5 inches in diameter. The first patch in the row presents a very high contrast grating (sample patch) in the far left of the row. The remaining 16 patches appear in eight columns presented across the row. In each column, one patch presents a grating, the other patch is blank. The patches that present gratings decrease in contrast moving from left to right across the row. The patient is directed to observe the first sample patch and is told to look for the grating pattern in each column. While reading across the row, that patient indicates whether the grating appears in the top patch or the bottom patch for each column. If the grating is not visible in either patch, the patient responds "both blank." It is important to note that the patient is encouraged to guess if a grating is at least partially visible as the threshold is approached. However, the patient is cautioned that if no gratings are visible, then the response should be "both blank." The contrast level of the last correct response is taken as the contrast threshold. Contrast sensitivity levels in each row range from.70 to 2.08,.91 to 2.29,.61 to 1.99, and.17 to 1.55 log units for 3, 6, 12, and 18 cyc/deg, respectively. Contrast levels diminish in a uniform logarithmic fashion in steps of 0.15 log units for contrast levels 3 through 8 and 0.17 log units for steps 1 through 3. The contrast change between the sample patch and level 1 is 0.3 log units. RESULTS The before therapy, after therapy, and difference contrast sensitivity scores for the patients with glaucoma are shown in Table 1. A repeated measures analysis of variance showed a statistically significant improvement in contrast sensitivity after the initiation of glaucoma therapy (P <.0005). A post hoc test across all spatial frequencies using Scheffe's multiple comparison test showed that the change in vision was significant at all spatial frequencies (P <.01). A Mest showed that the drop in intraocular pressure from 20.5 mm Hg to 18.2 mm Hg was significant (P <.01). No significant change in acuity was found. Only the right eye of those in the normally sighted group was used for evaluation. All subjects had 20/20 or better acuity. The mean and standard deviation for each spatial frequency and the test-retest mean differences and standard deviations are shown in Table 2. A repeated measures analysis of variance showed that the slight improvement in contrast sensitivity found on the retest was not statistically significant (P <.05). For COR to be used to evaluate the difference scores, the scores must be normally distributed. The distributions of difference scores (test 2 test 1) for each spatial frequency were plotted to determine normality. Visual inspection showed that none of the distributions were significantly skewed from the normal. (This result was anticipated because, as noted by Reeves et al, 6 difference scores of test-retest data are typically normally distributed.) Table 3 shows the COR for each spatial frequency, the test-retest mean differences for patients with glaucoma, and their ratios. Figure 2 shows the improvements in contrast sensitivity at each spatial frequency for the group with glaucoma versus the normal test-retest changes and the 95% confidence interval for the test-retest variability (COR). DISCUSSION For the group with glaucoma, the improvements found with therapy agree with previous reports showing upward shifts in contrast sensitivity after the initiation of therapy. 2 ' 4 Comparison of the CORs to the mean test-retest differences show that the COR is less than the average change induced by therapy at 3, 6, and 12 cyc/deg. That is, the ratio calculated by dividing the test-retest differences by the COR is greater than 1. This result suggests that the CSV-1000 can be used as a clinically reliable tool for measuring the changes in vision induced by glaucoma therapy, except at 18 cyc/deg, where the ratio between COR and mean difference is less than 1.

3360 Investigative Ophthalmology & Visual Science, August 1994, Vol. 35, No. 9 TABLE 2. Test, Retest, and Difference Scores for Normally Sighted Group (n = 24) Spatial Frequency 3 cyc/deg 6 cyc/deg 12 cyc/deg 18 cyc/deg Values are mean ± SD. Test 1.55 ± 0.155 1.76 ± 0.178 1.49 ± 0.226 0.913 ± 0.300 Retest 1.57 ± 0.149 1.84 ± 0.158 1.50 ± 0.152 0.951 ± 0.231 Test-Retest Difference 0.018 ± 0.048 0.078 ± 0.074 0.013 ± 0.116 0.038 ± 0.151 The average COR for the CSV-1000 (.191) compares favorably with those published for the Pelli- Robson Contrast Test (.18) and is much lower than the CORs for Vistech FDT (.36) and Miller-Nadler Glare Tester (.36). 7 A comparison of the individual CORs for each frequency also provides interesting information. The Pelli-Robson test measures sensitivity to the peak of the contrast sensitivity function. 8 Combining the two CORs from spatial frequencies on the CSV-1000 that test the peak sensitivity (i.e., 3 and 6 cyc/deg) shows that the average COR for this range (.12) is below that of the Pelli-Robson (.18). The Vistech FDT tests 6 cyc/deg. A direct comparison of the COR for 6 cyc/deg on the CSV-1000 (.145) again shows the CSV-1000 test-retest variance to be well below that of the Vistech (.36). The COR for 18 eye/ deg is approximately.3 log units. At this high level of test-retest variance, 18 cyc/deg can only be reliably used to measure large changes in vision. It may seem contradictory that the CORs for the CSV-1000 are much lower than that for the Vistech test because both tests are similar. However, several important differences exist between the tests. The Vistech test incorporates an unequal contrast interval that ranges from.36 to.11 log units (129% to 28% change), averaging approximately.25 log units (78% change). 9 The CSV-1000 uses a uniform log step of.15 log units (40%) across contrast steps 3 to 8 and.17 log units (50%) from steps 1 to 3. As noted by Elliott and Bullimore, 7 a larger contrast step increases test-re test variance. The Vistech test incorporates an orientation task that poses two difficulties in terms of reliability. First, TABLE 3. CORs for Normals and Test-Retest Mean Differences for Glaucoma Patients and Ratios Spatial Normal Glaucoma Mean Ratio COR/Mean Frequency CORs Difference Difference 3 cyc/deg 6 cyc/deg 12 cyc/deg 18 cyc/deg 0.094 0.145 0.227 0.296 0.186 0.167 0.295 0.191 1.98 1.15 1.30 0.64 determining grating orientation is a two-part task in which the patient must not only detect the grating but must also identify its orientation. This two-part task injects an additional criterion level into the test, compared to a "true detection" task of inspecting two circles to determine the presence of a grating pattern. Second, oriented gratings can confound with astigmatism. A patient may be unable to identify a higher contrast grating in an orientation that coincides with his or her astigmatism and yet be able to identify a lower contrast grating of a different orientation. Although a test with only one grating orientation, such as the CSV-1000, is also affected differently by different axes of astigmatism, this has little impact on test reliability because, in a given patient, the axis of astigmatism is fixed and all contrast levels are affected equally. The CSV-1000 uses a criterion-dependent paradigm in which the patient has three choices: to deter- 0.4 0.3 0.2 0.1 0-0.1-0.2-0.3-0.4 Test-Retest Changes In Contrast Sensitivity Normal Test/Retest 1 Glaucoma Pre/Post Therapy ' 95% Confidence Interval (COR) * «. - " 1 0 Spatial Frequency (cyc/deg) FIGURE 2. Improvements in contrast sensitivity for patients with glaucoma after beta-blocker therapy versus changes in normal patients due to retest. The heavy solid line shows the improvement for patients with glaucoma, and the heavy dashed line shows the retest change for normal patients. The lighter dashed line shows the 95% confidence interval for the normal test-retest variability. Note that the improvement in contrast sensitivity for the patients with glaucoma is above the upper limit of the 95% confidence interval for normal test-retest variance, except at 18 cyc/deg.

Test-Retest Reliability of the CSV-1000 Contrast Test and Glaucoma 3361 mine whether the gratings are in the top circle or the bottom circle or whether both are blank. Criteriondependent tests have been shown to be less reliable than forced-choice tests. 10 Given the criterion dependence of the CSV-1000 test, it is tempting to attribute the improvements in sensitivity for the patients with glaucoma as the result of such testing factors as learning or familiarity. However, the control group, which was matched for age and test-retest interval, showed only minor improvement that was not close in magnitude when compared to the group with glaucoma and was not statistically significant. Further, Woo and Bohnsack 1 ' found a similar learning effect with the Vistech test, that again was much smaller in magnitude than the changes we found before and after therapy and that was not statistically significant at the.05 level. Test-retest reliability can be artificially improved by the inclusion of truncated data, i.e., data from patients who are unable to see a grating of a particular spatial frequency, even at the highest contrast level. Because all the patients who participated in this study had good acuity (20/40 or better), no truncation occurred. The mechanism for the improvement in vision with beta-blocker therapy, as measured by contrast sensitivity, remains unknown. An initial explanation could be a miotic effect of the pharmacologic agents. This is unlikely, however; Nordmann et al 2 found improvements after laser surgery that had no effect on pupil size. Perhaps ganglion cells gain sensitivity as the intraocular pressure is reduced. Because we also found a significant decrease in intraocular pressure after treatment, this is a possible explanation. However, the correlation between changes in intraocular pressure and improvements in contrast sensitivity was not significant at any spatial frequency. Another possible explanation is that contrast sensitivity improves with improved ocular blood flow. Studies have demonstrated increased retinal blood flow after beta-blocker therapy (timolol) in both animal 12 and human eyes. 13 Further, Sponsel and colleagues 14 demonstrated a high correlation between asymmetry in retinal leukocyte velocity and contrast sensitivity in patients with glaucoma and those with ocular hypertension. However, in a study using an acute dose of a calcium-channel blocker, 4 an association between improvements in contrast sensitivity and macular blood flow after therapy was not evident. We think the shifts in contrast sensitivity were related to increased ocular blood flow. As noted above, however, current data are inconclusive and further studies are needed to confirm this hypothesis. Our results indicate that contrast sensitivity does improve with beta-blocker therapy and that the CSV- 1000 contrast testing instrument has sufficient reliability to measure these improvements. Perhaps the evaluation of the short- and long-term effects of betablocker therapy on visual function in patients with glaucoma can be augmented by the use of contrast sensitivity testing. Key Words contrast sensitivity, test reliability, glaucoma therapy, betablocker, visual function References 1. Koch D, Lui J. Survey of the clinical use of glare and contrast testing. / Cataract Refract Sur. 1990; 16:707-711. 2. Nordmann J, Topouzis P, Laroche L, Saraux H. Three year follow-up of glaucoma patients using contrast sensitivity. ARVO Abstracts. Invest Ophthalmol Vis Sci. 1992; 33:3469. 3. Tytla M, Trope G, Buncic J. Flicker sensitivity in treated ocular hypertension. Ophthalmology. 1990; 97: 36-43. 4. Piltz J, Bose S, Grunwald J, Petrig L, Riva C. Effect of nimodipine on automated threshold perimetry, spatial contrast sensitivity and macular blood flow in normal tension glaucoma and controls. ARVO Abstracts. Invest Ophthalmol Vis Sci. 1993; 34:2881. 5. Bland J, Altman D. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986; 1:307-310. 6. Reeves B, Wood J, Hill A. Reliability of high- and lowcontrast letter charts. Ophthalmic Physiol Opt. 1993; 13:17-26. 7. Elliott D, Bullimore M. Assessing the reliability, discriminative ability and validity of disability glare tests. Invest Ophthalmol Vis Sci. 1993;34:108-119. 8. Pelli D, Robson J, Wilkins A. The design of a new letter chart for measuring contrast sensitivity. Clin Vision Sci. 1988;2:187-199. 9. Contrast Sensitivity Value Key. Vistech Consultants Inc. Dayton, OH. 10. Higgins K, Jaffe M, Coletta N, Caruso R, DeMonasterio F. Spatial contrast sensitivity: Importance of controlling the patient's visibility criterion. Arch Ophthalmol. 1984; 102:1035-1041. 11. Woo G, Bohnsack H. Comparison of the distance and near Vistech vision contrast test systems (VCTS). Can JOptom. 1986; 48:12-15. 12. Chiou G, Chen Y. Effects of D- and L-isomers of timolol on retinal and choroidal blood flow in ocular hypertensive rabbit eyes./ Ocul Pharm. 1992;8:183-190. 13. Grunwald J. Effect of timilol maleate on the retinal circulation of human eyes with ocular hypertension. Invest Ophthalmol Vis Sci. 1990;31:521-526. 14. Sponsel W, DePaul K, Kaufman P. Correlation of visual function and retinal leukocyte velocity in glaucoma. Am J Ophthalmol. 1990; 109:49-54.