Variability of Automated Visual Fields in Clinically Stable Glaucoma Patients

Similar documents
Quantification of Glaucomatous Visual Field Defects with Automated Perimetry

Fluctuations on the Humphrey and Octopus Perimeters

Characteristics of Frequency-of-Seeing Curves in Normal Subjects, Patients With Suspected Glaucoma, and Patients With Glaucoma

STANDARD AUTOMATED PERIMETRY IS A GENERALLY

T he retinal ganglion cells of different sizes have distinct

Variability in Patients With Glaucomatous Visual Field Damage Is Reduced Using Size V Stimuli

Misleading Statistical Calculations in Faradvanced Glaucomatous Visual Field Loss

LABORATORY SCIENCES. Spatial and Temporal Processing of Threshold Data for Detection of Progressive Glaucomatous Visual Field Loss

Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma

Baseline Visual Field Characteristics in the Ocular Hypertension Treatment Study

CLINICAL SCIENCES. Fellow Eye Prognosis in Patients With Severe Visual Field Loss in 1 Eye From Chronic Open-Angle Glaucoma

NERVE FIBER LAYER THICKNESS IN NORMALS AND GLAUCOMA PATIENTS

Science & Technologies

CLINICAL SCIENCES. Steven L. Mansberger, MD; Pamela A. Sample, PhD; Linda Zangwill, PhD; Robert N. Weinreb, MD

CORRELATING OF THE VISUAL FIELD INDEX WITH MEAN DEVIATION AND PATTERN STANDARD DEVIATION IN GLAUCOMA PATIENTS

Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection

Diurnal IOP fluctuation: not an independent risk factor for glaucomatous visual field loss in high-risk ocular hypertension.

Spontaneous Intraocular Pressure Reduction in Normal-Tension Glaucoma and Associated Clinical Factors

Present relevant clinical findings of four landmark glaucoma trials OHTS, EMGT, CNTGS and CIGTS.

Glaucoma: Diagnostic Modalities

The optic disc in glaucoma. IV: Optic disc

CLINICAL SCIENCES. Glaucoma Monitoring in a Clinical Setting

VISUAL FIELD TESTING OCCUpies

C linical visual field tests are designed to provide. Measurement error of visual field tests in glaucoma LABORATORY SCIENCE

Access to the published version may require journal subscription. Published with permission from: Elsevier

CHAPTER 6 SELECTING A TEST STRATEGY INTRODUCTION. accurate but shorter test may yield more useful visual

21st Century Visual Field Testing

Evaluating visual field over time is a key element in the

IDIOPATHIC INTRACRANIAL HYPERTENSION

Fluctuation of Intraocular Pressure and Glaucoma Progression in the Early Manifest Glaucoma Trial

3/16/2018. Perimetry

Noel de Jesus Atienza, MD, MSc and Joseph Anthony Tumbocon, MD

The Relationship between Standard Automated Perimetry and GDx VCC Measurements METHODS

Citation for published version (APA): Wesselink, C. (2017). Glaucoma care optimised in an ageing population [Groningen]: Rijksuniversiteit Groningen

Test-retest variability in visual field testing using frequency doubling technology

Glaucoma Progression. Murray Fingeret, OD

Citation for published version (APA): Stoutenbeek, R. (2010). Population based glaucoma screening Groningen: s.n.

PERIMETRY A STANDARD TEST IN OPHTHALMOLOGY

Surgery for normal tension glaucoma

Correlation of Blue Chromatic Macular Sensitivity with Optic Disc Change in Early Glaucoma Patients

glaucoma and ocular hypertension

Clinical Guidance and Monitoring for Change. Cecilia Fenerty MD FRCOphth Manchester Royal Eye Hospital

COMPARISON BETWEEN THRESHOLD PUPIL PERIMETRY AND SUPRATHRESHOLD PUPIL PERIMETRY

Simulations for FASTPAC and the Standard 4-2 db Full- Threshold Strategy of the Humphrey Field Analyzer

Rate and Amount of Visual Loss in 102 Patients with Open-Angle Glaucoma Followed Up for at Least 15 Years

Clinical Study Visual Field Loss Morphology in High- and Normal-Tension Glaucoma

Visual Fields Shawn L. Cohen, M.D. Part 2 of 4. Definitions / Tables (Part 2 of 2) Static Perimetry (Humphrey, Octopus)

Investigative Ophthalmology & Vision Sciences MSc Course. Glaucoma Module. Visual Field Reliability Indices. David Henson 2014.

Incidence of manifest glaucoma

Effect of brimonidine on intraocular pressure in normal tension glaucoma: A short term clinical trial

Intraocular pressure, systemic blood pressure, and

Role of Central Corneal Thickness in Circadian Intraocular Pressure Fluctuations among Patients with Primary Open Angle Glaucoma

Summary HTA HTA-Report Summary Validity and cost-effectiveness of methods for screening of primary open angle glau- coma

Ganglion cell complex scan in the early prediction of glaucoma

TUMBLING E RESOLUTION PERIMETRY IN GLAUCOMA

Criteria for early CLINICAL STUDY. N Fujimoto 1, N Saeki 2, O Miyauchi 1

The Role of the RNFL in the Diagnosis of Glaucoma

53 year old woman attends your practice for routine exam. She has no past medical history or family history of note.

AUTOMATIC STATIC FUNDUS PERIMETRY FOR PRECISE DETECTION OF EARLY GLAUCOMATOUS FUNCTION LOSS

OCT in the Diagnosis and Follow-up of Glaucoma

The determination of glaucoma progression is based on a. Refinement of Pointwise Linear Regression Criteria for Determining Glaucoma Progression

GLAUCOMA SUMMARY BENCHMARKS FOR PREFERRED PRACTICE PATTERN GUIDELINES

PRESCRIBING IN GLAUCOMA: GUIDELINES FOR NZ OPTOMETRISTS

International Journal Of Basic And Applied Physiology

Inadequate diagnostic value of the

The overall rate of decrease in visual field sensitivity in

Correlation of Blue-on-Yellow Visual Fields With Scanning Confocal Laser Optic Disc Measurements

Retrospective analysis of risk factors for late presentation of chronic glaucoma

Description of new EyeSuite visual field and trend analysis functions

Sensitivity and specificity of new GDx parameters Colen TP, Tang NEML, Mulder PGH and Lemij HG Submitted for publication CHAPTER 7

Ocular hypertension is present in approximately 8%

eye as a camera Kandel, Schwartz & Jessel (KSJ), Fig 27-3

Evaluation of Sensitivity and Specificity of Spatial Resolution and Humphrey Automated Perimetry in Pseudotumor Cerebri Patients and Normal Subjects

Specific deficits of flicker sensitivity in glaucoma and ocular hypertension

Collaboration in the care of glaucoma patients and glaucoma suspects. Barry Emara MD FRCS(C) Nico Ristorante November 29, 2012

The effect of pilocarpine on the glaucomatous visual field

A Visual Field Index for Calculation of Glaucoma Rate of Progression

Financial Disclosure. Visual Field Interpretation RELIABILITY VISUAL FIELD INTERPRETATION IN GLAUCOMA METHODS OF DATA PRESENTATION

CHAPTER 10 NON-CONVENTIONAL PERIMETRY

N umerous techniques have shown vascular abnormalities

Prevalence Of Primary Open Angle Glaucoma in Diabetic Patients

Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography

Patterns of Subsequent Progression of Localized Retinal Nerve Fiber Layer Defects on Red-free Fundus Photographs in Normal-tension Glaucoma

Supported by King's College Hospital Research Committee, International Glaucoma Association and M.S.D. Chibret.

CHAPTER 11 KINETIC PERIMETRY WHAT IS KINETIC PERIMETRY? LIMITATIONS OF STATIC PERIMETRY LOW SPATIAL RESOLUTION

Introduction. Paulo de Tarso Ponte Pierre-Filho, 1 Rui Barroso Schimiti, 1 Jose Paulo Cabral de Vasconcellos 1 and Vital Paulino Costa 1,2

Translating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects

Intro to Glaucoma/2006

Long-term perimetric fluctuation in patients with different stages of glaucoma

Perimetry Phobia: Don t fear the field Savory Turman, COMT, CPSS

FLICKER-TOP PERIMETRY IN NORMALS AND PATIENTS WITH OCULAR HYPERTENSION AND EARLY GLAUCOMA

Study of correlation of cup disc ratio with visual field loss in primary open angle glaucoma

Discrimination between normal and glaucomatous eyes with visual field and scanning laser polarimetry measurements

Test-Retest Reliability of the CSV-1000 Contrast Test and Its Relationship to Glaucoma Therapy

LUP. Lund University Publications Institutional Repository of Lund University

Glaucomatous Visual Field Damage

GLAUCOMA EVOLUTION IN PATIENTS WITH DIABETES

The Evolution of Fundus Perimetry

Transcription:

Investigative Ophthalmology & Visual Science, Vol. 30, No. 6, June 1989 Copyright Association for Research in Vision and Ophthalmology Variability of Automated Visual Fields in Clinically Stable Glaucoma Patients Elliot D. Werner,* Benno Perrig.f Theodore Krupin.f and Kim I. Bishopf The total variability of the visual field was measured in 20 patients with open-angle glaucoma who appeared to be clinically stable and well controlled on medical therapy. All patients had at least four visual fields performed on the Octopus 201 perimeter with at least 12 months follow-up since their first visual field. The four most recently performed visual fields were analyzed. Two different methods for calculating total variability were used. One was based on the variance of the threshold determinations and the other was based on the range. The average total variability per subject was 2.8 decibels (db) using the variance-based calculation and 5.1 db using the range-based calculation. Ninety-five percent of the test locations had a variability of less than 6 db by the variance-based calculation method and 13 db by the range-based calculation method. We discuss the possibility of using this type of data to develop criteria for detection of progressive visual field loss in glaucoma. Invest Ophthalmol Vis Sci 30:1083-1089,1989 It is well recognized that the visual field of an individual often fluctuates on repeated testing. Techniques for measuring the fluctuation of the visual field over time have been developed by several workers. 1 " 10 It has now been established that larger amounts of variability are found in the visual fields of patients with glaucoma and glaucoma suspects than in normal subjects. 6 ' 7 '"" 19 Clinicians are primarily interested in learning whether or not their patients are getting worse so that appropriate glaucoma therapy may be applied. In the presence of large amounts of variability, the separation of true change in the visual field from fluctuation becomes an important problem. Various statistical techniques have been applied to the visual field in an effort to define progressive visual field loss by detecting statistically significant change. 20 " 23 None of these techniques has been clinically validated. It has recently been shown that experienced clinical observers From the 'Department of Ophthalmology, Hahnemann University, and the tglaucoma Service, University of Pennsylvania School of Medicine and the Department of Ophthalmology, Scheie Eye Institute, Philadelphia, Pennsylvania. Presented in part at the annual meeting of the Association for Research in Ophthalmology, Sarasota, Florida, May 5, 1988, and at the Biannual meeting of the International Perimetric Society, Vancouver, British Columbia, May 11, 1988. Supported in part by the Elaine O. Weiner Teaching and Research Fund and an unrestricted grant from Research to Prevent Blindness, Inc., New York. Submitted for publication: February 11, 1988; accepted December 22, 1988. Reprint requests: Elliot B. Werner, MD, Hahnemann University Department of Ophthalmology, 216 North Broad Street, Mail Stop 209, Philadelphia, PA 19102. and standard statistical tests do not agree well with each other in detecting progressive visual field loss in glaucoma patients. 24 The purpose of this study is to determine the total amount of visual field fluctuation that may be expected over a long period of time in a sample of clinically stable glaucoma patients. Using this information we may be able to develop criteria for the amount of change in the visual field that must occur to be interpreted as significant when measured by automated perimetry. Materials and Methods The charts of all patients followed by either one of two of the authors (TK or EW) in the glaucoma service of the Scheie Eye Institute were reviewed. All subjects who met the following criteria were included in the study: 1. Diagnosis of primary open-angle glaucoma, pigmentary glaucoma or pseudoexfoliation with glaucoma; 2. At least four available automated visual field examinations with at least 12 months follow-up since the first visual field; 3. Presence of a definitive nerve fiber bundle visual field defect detected on manual (Goldmann) perimetry prior to the patient's first automated visual field examination; 4. Visual acuity of 20/40 or better with no change in visual acuity greater that one Snellen line during the follow-up period; 5. No intraocular pressure greater than 19 mm Hg at 1083

1084 INVESTIGATIVE OPHTHALMOLOGY 6 VISUAL SCIENCE / June 1989 Vol. 30 Fig. 1. The pattern of test locations within the central 30 of the visual field on program 32 of the Octopus perimeter with the ten visual field sectors superimposed. any time since the initiation of the patient's present medical therapy; 6. No change in the appearance of the optic nerve head which could be detected by the treating ophthalmologist (TK or EW) after reviewing serial stereoscopic disc photographs obtained at yearly intervals on each subject; 7. No change in medical therapy for glaucoma during the follow-up period; 8. No laser or surgical treatment for glaucoma during the follow-up period and the 12 preceding months. All subjects were phakic and had no other known ocular, neurologic or systemic disease likely to affect the visual field or other visual functions. All visual fields were performed on the Octopus 201 perimeter using program 32 with an undilated pupil. A mean, sample variance and range were calculated for the sensitivity values expressed in decibels (db) for each test location in the visual field of each subject. Only the four most recently available visual field examinations were included in the calculations. Each subject was followed for at least 12 months after their first automated visual field examination. It is possible, however, in some subjects that the four most recent examinations that were used for analysis were obtained over less than 12 months. The two test locations corresponding to the usual blind spot area were eliminated. There were, thus, a total of 74 test locations but 84 threshold values for each visual field because each of the two values for the doubly determined points were treated as separate test locations for purposes of statistical analysis. All visual field threshold data were entered into an electronic spreadsheet (1-2-3, Lotus Development Corporation, Cambridge, MA) on a personal computer for analysis. Two methods were used to calculate the total variability per test location. In the first method, total variability per test location was defined as the square root of the sample variance (standard deviation) of the sensitivities for that location. Using this method based on variance, the total fluctuation of the entire visual field for each subject was denned as the square root of the mean variance of all subject's tjst locations, excluding test locations for which the threshold was always zero db. Test locations for which the threshold was always zero were excluded because no variability can be measured if the threshold exceeds the maximal luminance available on the perimeter. Inclusion of test locations where the threshold was measurable during some of the examinations and zero at other times may have slightly underestimated the variability at those locations. We felt, however, that if the threshold was zero on some examinations but recordable on others, then the threshold for that location was probably varying around a mean greater than zero, and, therefore, the variability was measurable based on the available threshold determinations. In the second method, total variability was derived from the range of the sensitivity values for each test location. We call this range variability. Range variability for a single test location was denned as the absolute difference between the highest and lowest sensitivity readings for that test location. With this method, total fluctuation of the entire visual field for a single subject was defined as the mean of the range variabilities of all test locations, excluding test locations where the sensitivity was always zero db. Each visual field was divided into ten sectors roughly corresponding to the nerve fiber layer anatomy of the retina (Fig. 1). The number of threshold determinations in each sector varied between five and 13. A mean threshold was then calculated for each sector of each visual field for each individual subject. This value is called the sector threshold. Each sector threshold was treated as a single data point. The variance and range for each sector of each subject's visual field over time were calculated. As was done for individual test locations, total variability of the entire visual field was calculated in terms of either the square root of the mean variance or the mean range of all sector thresholds.

No. 6 VISUAL FIELD VARIABILITY IN STABLE GLAUCOMA PATIENTS / Werner er QI 1085 Results Twenty subjects were found who satisfied the inclusion criteria. Each subject was followed for at least 12 months after their first automated visual field examination; however, the average time interval encompassed by the four most recent visual fields analyzed on each subject was 22.5 months (range 8-36 months). If both eyes of a subject qualified for the study, one eye was selected at random. The average mean sensitivity of each subject's visual field for the entire sample was 15.8 db (range 5.4-24.7 db). The study thus included a sample of glaucoma patients ranging from minimal to severe damage (Fig. 2). Of the 1680 test locations in the sample, there were 176 whose sensitivity was always 0 db. These points were excluded from further calculations of variability. Of the 200 visual field sectors in the sample, there were nine sectors of four subjects whose sector threshold was always 0 db. These sectors were also excluded from further calculations of variability. When using the calculation of variability based on variance, the mean total fluctuation per subject was 2.8 db with a range of 1.4-4.7 db. By using the rangebased calculation, the mean total fluctuation per subject was 5.1 db with a range of 2.8-8.8 db. The variability per test location was also calculated for the entire sample. If the square root of the variance of the four values for each test location is used as the measure of variability, 95% of the test locations had a variability of less than 6 db while 99% had a variability less than 9 db (Fig. 3). If the range of the four values for each test location is used as the measure of variability, 95% of the test locations had a variability less than 13 db while 99% of the test locations had a fluctuation less than 19 db (Fig. 4). The total variability per sector threshold was also calculated. If the square root of the variance for each sector is used as the measure of variability, 95% of the sectors had a variability of less than 3 db while 99% of the sectors had a variability of less than 4 db (Fig. 5). If the range of the four values for each sector is used as the measure of variability, 95% of the sectors had a variability less than 7 db and 99% of the sectors had a variability of less than 8 db (Fig. 6). The size of each subject's pupil was measured prior to each automated visual field test. As might be expected, there was a bimodal distribution of pupil size because some patients were using miotics and some were not. The mean pupil size for the sample was 3.0 mm with a range of 1.0 to 6.0. Nine of the subjects had an average pupil size of less than 2.0 mm while eight of the subjects had a mean pupil size greater 10 r o 6 <8 8-12 12-16 16-20 20-24 24-28 Mean sensitivity (decibels) Fig. 2. Frequency distribution showing number of subjects against mean sensitivity. The sample includes patients ranging from minimal visual field defects (high sensitivity) to extensive visual field loss (low sensitivity). than 3.5 mm. The other three subjects were in between. We also measured the range over which the pupil size varied during the four visual field examinations. Three of the subjects showed no change in pupil size while another 12 showed less than 1.0 mm variation - 200 11 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Variance based variability (decibels) Fig. 3. Frequency distribution showing the number of test locations against the variability per test location calculated using the variance based method. Test locations where threshold was always zero db have been excluded; 95% of the test locations vary by less than 6 db.

1086 INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE / June 1989 Vol. 30 C\J\J 240-200 160 120 80 40 n -. I I : i j' ill 'lllllll. 2 4 6 8 10 12 14 16 18 20 22 24 26 28 Range based variability (decibels) Fig. 4. Frequency distribution showing the number of test locations against the variability per test location calculated using the range based method. Test locations where threshold was always zero db have been excluded; 95% of the test locations vary by less than 13 db. in the size of the pupil. The mean of the range of variation in pupil size for the sample was 0.8 mm. The largest change in pupil size in any of the subjects was 2.0 mm. There was no correlation between the mean size of the pupil and the total fluctuation of the visual field as measured by the range based calculation described 120 r 100 80 60 40 20 0 1 2 3 4 5 Variance based variability (decibels) Fig. 5. Frequency distribution showing the number of visual field sectors against the variability of the sector threshold calculated with the variance-based method; 95% of the sectors vary by less than 3 db. above (Pearson correlation coefficient =.223, P >0.10, Spearman rank coefficient = -.214, P > 0.10). There was also no correlation between the range over which the subjects' pupils varied and the total fluctuation of the field (Pearson correlation coefficient =.161, P> 0.10, Spearman rank coefficient =.082, P> 0.10). Discussion The ability of any measuring device to detect a true difference among several measurements depends on the reproducibility of its results. In the case of perimetry, reproducibility is confounded by the fact that the object being measured, the visual field, seems to be constantly changing. Furthermore, the magnitude of this change varies among individuals. This constant changing of the visual field has been termed fluctuation. Our calculation of fluctuation includes both long- and short-term components. We therefore use the term total fluctuation to describe the variability measured in this paper. A difference between examinations done at different times must be greater than the expected physiologic fluctuation to be recognized as a genuine alteration. 825 Thus, any technique that is designed to detect either improvement or progressive deterioration of the visual field in glaucoma must take into account the magnitude of the expected fluctuation in glaucoma patients, not in normal subjects. A detailed discussion of the sources of variability in visual field testing is beyond the scope of this paper. There are, however, two possible sources of variabil-

No. 6 VISUAL FIELD VARIABILITY IN STABLE GLAUCOMA PATIENTS / Werner er ol 1087 ity that we would like to address specifically. The first is pupil size. Visual fields were performed in subjects with an undilated pupil. The pupil size was recorded, but no attempt was made to control the pupil size. It is well recognized that a change in pupil size can have profound effects on the overall sensitivity of the visual field. In fact, the pupil size in our subjects did not vary much between examinations since the subjects' medical therapy was not altered during this study. Unless pupils are routinely dilated for perimetry, random changes in the size of a patient's pupil may be a source of variability. In our sample, however, neither the actual size of the pupil nor changes in nupil size between examinations seemed to correlate with the total fluctuation measured for the subjects. Another possible source of variability is the presence of a learning effect in automated perimetry. These subjects all had experience with manual perimetry prior to their first automated visual field. We have previously shown that in such patients the effect of patient experience is very small and can only be demonstrated on the short-term fluctuation. 26 Learning, therefore, probably did not contribute much to the variability measured in this study. With one exception, 8 previous studies of long-term fluctuation of glaucoma patients have generally evaluated multiple visual fields performed over a relatively short period of time. 1519 ' 27 Our study reproduces the usual clinical situation in which a patient is examined and perimetry is performed at intervals of several months over a period of 1 or more years. The magnitude of total fluctuation in such circumstances should provide useful guidance in developing criteria to evaluate true change in the visual field. In any sample of patients with chronic glaucoma followed for a long period of time, there will be some who are getting progressively worse. In order to evaluate true long-term fluctuation, these patients would have to be excluded. The patients included in this study were all clinically stable so far as we could determine. There was nothing about them that would have led any reasonable clinician to conclude otherwise. The Sample should, therefore, provide the best available approximation of the expected total fluctuation in patients with chronic glaucoma who are receiving medical treatment. It is possible, of course, that the visual fields of some of these subjects may be changing over time despite the clinical impression that their glaucoma is stable. It has recently been shown, however, that this determination is extremely difficult to make using standard statistical tools such as regression analysis, t-test or aaalysis of variance. 24 In general, calculations of fluctuation in automated perimetry have been based upon the variance " 40 E ll. 0 1 2 3 4 5 6 7 8 9 Range based variability (decibels) Fig. 6. Frequency distribution showing the number of visual field sectors against the variability of the sector threshold calculated with the range-based method; 95% of the sectors vary by less than 7 db. of repeated determinations of threshold. Some authors, however, have based variability calculations on the range. 5 ' 28 We used both types of calculations, and, as one would expect, the range-based calculation of fluctuation gave values about twice that of the variance-based calculation. In clinical practice there are certain advantages to using the range as the measure of variability. The range of a series of values is more readily obvious without calculation than the variance or standard deviation. If only a few visual fields are available on a patient, the range or difference between the first and last fields may be more useful for detecting progressive change in the field than a variance calculated from a small number of values. Based on our entire sample of 1504 threshold determinations, one can now define the upper limit of change in visual thresholds that might be considered a result of random fluctuation. For example, using the range-based calculation of variability, a change of 13 db or more in a single test location due to random fluctuation will be observed only 5% of the time. There are, however, 74 available test locations in the test pattern known as Octopus program 32. The probability that any number (r) of these 74 (n) test locations vary by at least 13 db due only to fluctuation and not true progressive visual field change is given by the following binomial formula 29 : n! P = (n - r)!r! X pr q n " r (1) where n = number of points in the visual field tested (74 for program 32); r = number of points in which the specified change occurs (1-73), p = probability

1088 INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE / June 1989 Vol. 30 Table 1. Number of visual field sectors whose sector threshold would have to change by various specified amounts to be considered outside the expected range of random variability at the 0.05 and 0.01 probability levels Number of sectors 1 2 3 4 * Range of variability (db). >0.05 8* 7 6 4 Probability level >0.01 point will vary by the specified amount or more (.05 or.01), and q = probability point will vary by less than the specified amount (.95 or.99). Thus for one test location (r = 1), this probability is.891. In other words, even though the probability of a single test location varying over a range of 13 db or more is 5%, the probability of observing any one of the 74 test locations varying by at least this amount due to random fluctuation alone is nearly 90%. It is not until we reach seven test locations varying simultaneously over a range of 13 db or more (r = 7) that the probability of this being due to random fluctuation becomes less than 5%. As a sample criterion of progressive change, one could say that if seven test locations anywhere in the visual field change by 13 db or more in the same direction, then one can be 95% certain that such a change is not due to random fluctuation. We do not, however, recommend using data from individual test locations to develop criteria of change for several reasons. One is that our calculations assume that the fluctuation at each test location is independent of adjacent locations, which is probably not true. Second, we assume that there is a measurable threshold at all 74 test locations which is not always true. Third, our technique does not take into consideration the position of the test location in the visual field. Heijl et al have shown that the variability of test locations in the field is position-dependent in normal subjects. 30 While this has not been studied in patients with glaucoma, it is likely that the variability of individual test locations is position-dependent in patients with glaucoma as well. To eliminate the problems of dealing with single test locations, we divided the visual field into sectors. By treating the mean threshold of each sector as a single data point, we deal with groups of test locations. The location and shape of the sectors (Fig. 1) was chosen to reflect the nerve fiber layer anatomy of the retina and the usual patterns of visual field loss in glaucoma. 9* 8 7 6 The sector threshold of 95% of the visual field sectors in our sample varied over a range of less than 7 db, and there are only ten sectors in each field. As was done for individual test locations, we can apply the same binomial formula for the sectors. In this case, however, n = 10 and r < 10. The probability of the sector threshold of any one of the ten sectors (r = 1) varying over a range of 7 db, or more is.264, but the probability of two sectors (r = 2) in the same visual field varying by at least 7 db is 0.046. Using our data, one can now develop rational criteria for progressive visual field loss in glaucoma based upon probabilities. Table 1 shows the number of visual field sectors that would have to change by differ ent specified amounts to be considered outside the range of random fluctuation at the 0.05 and 0.01 probability levels. For example, if one sector varies by 8 db, then this is expected to be due to physiologic fluctuation less than 5% of the time, or if three sectors vary by 7 db, this is expected to be due to fluctuation less than 1% of the time, etc. The magnitude of these sector-based variabilities are very similar to those reported by Hoskins et al. 31 These guidelines should be useful both for clinicians who follow patients with automated perimetry as well as for researchers using the visual field to evaluate risk factors or therapy in glaucoma. Key words: glaucoma, automated perimetry, visual fields, variability, long-term fluctuation References 1. Bebie H, Fankhauser F, and Spahr J: Static perimetry: Accuracy andfluctuations.acta Ophthalmol 54:339, 1976. 2. Koerner F, Fankhauser F, Bebie H, and Spahr J: Threshold noise and variability offielddefects in determinations by manual and automatic perimetry. Doc Ophthalmol Proc Ser 14:53, 1977. 3. Fankhauser F and Bebie H: Thresholdfluctuations,interpolations and spatial resolution in perimetry. Doc Ophthalmol Proc Ser 19:295, 1979. 4. Flammer J, Drance SM, and Schulzer M: The estimation and testing of the components of long-term fluctuation of the differential light threshold. Doc Ophthalmol Proc Ser 35:383, 1983. 5. Wilensky JT and Joondeph BC: Variation in visualfieldmeasurements with an automatic perimeter. Am J Ophthalmol 97:328, 1984. 6. Langerhorst CT, Van Den Berg TJTP, Van Spronsen R, and Greve EL: Results of afluctuationanalysis and defect volume program for automated static threshold perimetry with the scoperimeter. Doc Ophthalmol Proc Ser 42:1, 1984. 7. Lynn JR, Batson EP, and Fellman RL: Internal inconsistencies vs. root mean square as measures of threshold variability. Doc Ophthalmol Proc Ser 42:7, 1985. 8. Gloor BP and Vokt BA: Long-termfluctuationsversus actual field loss in glaucoma patients. Dev Ophthalmol 12:48, 1985. 9. Lewis RA, Johnson CA, Keltner JL, and Labermeier PK: Variability of quantitative automated perimetry in normal observers. Ophthalmology 93:878, 1986.

No. 6 VISUAL FIELD VARIABILITY IN STABLE GLAUCOMA PATIENTS / Werner er ol 1089 10. Katz J and Sommer A: A longitudinal study of the age-adjusted variability of automated visualfields.arch Ophthalmol 105:1083. 1987. 11. Werner EB and Drance SM: Early visualfielddisturbances in glaucoma. Arch Ophthalmol 95:1173, 1977. 12. Donovan HC, Weale RA, and Wheeler C: The perimeter as a monitor of glaucomatous changes. Br J Ophthalmol 62:705, 1978. 13. Werner EB, Saheb N, and Thomas D: Variability of static visual threshold responses in patients with elevated IOP's. Arch Ophthalmol 100:1627, 1982. 14. Flammer J, Drance SM, and Zulauf M: Differential light threshold short and long-term fluctuation in patients with glaucoma, normal controls and patients with suspected glaucoma. Arch Ophthalmol 102:704, 1984. 15. Flammer J, Drance SM, Fankhauser F, and Augistiny L: Differential light threshold in automatic static perimetry: Factors influencing short-termfluctuation.arch Ophthalmol 102:876, 1984. 16. Flammer J, Drance SM, and Schulzer M: Covariates of the long-term fluctuation of the differential light threshold. Arch Ophthalmol 102:880, 1984. 17. Crick RP, Crick JCP, and Ripley LG: Some aspects of perimetry in glaucoma. Glaucoma 7:27, 1985. 18. Flammer J and Zulauf M: The frequency distribution of the deviations in static perimetry. Doc Ophthalmol Proc Ser 42:17, 1985. 19. Brenton RS and Argus WA: Fluctuations on the Humphrey and Octopus perimeters. Invest Ophthalmol Vis Sci 28:767, 1987. 20. Heijl A: A simple routine for demonstrating increased threshold scatter by comparing stored computer fields. Doc Ophthalmol Proc Ser 42:35, 1985. 21. Bebie H and Fankhauser F: Statistical program for the analysis of perimetric data. Doc Ophthalmol Proc Ser 26:9, 1981. 22. Holmin C and Krakau CET: Regression analysis of the central visual field in chronic glaucoma cases. Acta Ophthalmol 60:267, 1982. 23. Hirsch J: Statistical analysis in computerized perimetry. In Computerized Visual Fields, Whalen WR and Spaeth GL, editors. Thorofare, NJ, Slack, Inc., 1985, pp. 307-344. 24. Werner EB, Bishop KI, Koelle J, Douglas GR, LeBlanc R, Mills RP, Schwartz B, Whalen WR, and Wilensky JT: A comparison of experienced clinical observers and statistical tests in the detection of progressive visualfieldloss in glaucoma using automated perimetry. Arch Ophthalmol 106:619, 1988. 25. Bebie H: Computerized techniques of visual field analysis. In Automatic Perimetry in Glaucoma, Drance SM and Anderson DA, editors. New York, Grune & Stratton, 1985, pp. 147-160. 26. Werner EB, Adelson A, and Krupin T: Effect of patient experience on the results of automated perimetry in clinically stable glaucoma patients. Ophthalmology 95:764, 1988. 27. Piltz JR, Starita RJ, Fechtner RD, and Twersky YD: Fluctuation of serial automated visual fields in glaucomatous and normal eyes. ARVO Abstracts. Invest Ophthalmol Vis Sci 27(Suppl):159, 1986. 28. Magee SD, Hoskins HD, and Kidd MN: Long-term fluctuation in glaucomatous visual fields. ARVO Abstracts. Invest Ophthalmol Vis Sci 28(Suppl):269, 1987. 29. Parsons R: Statistical Analysis: A Decision Making Approach. New York, Harper & Row, 1974. 30. Heijl A, Lindgren G, and Olsson J: Normal variability of static perimetric threshold values across the central visualfield.arch Ophthalmol 105:1544, 1987. 31. Hoskins HD, Magee SD, Drake MV, and Kidd MN: Confidence intervals for change in automated visual fields. Br J Ophthalmol 72:591, 1988.