Product Guide for LudgerSep TM R1 HPLC Column for DMB labelled Sialic Acid Analysis

Similar documents
Product Guide for LudgerSep TM ur2 UHPLC Column for DMB Sialic Acid Analysis

Product Guide for LudgerSep TM R1 HPLC Column for Glycan Analysis

Product Guide for LudgerSep TM C3 anion exchange HPLC Column for Glycan Analysis

Certificate of Analysis

Fetuin Glycoprotein Standard

EASIMIP TM PATULIN Product Code: P250 / P250B

Product Guide for LudgerLiberate TM Orela Glycan Release Kit

KE-SIALIQ Sialic Acid Quantitation Kit. SialiQuant Sialic Acid Quantitation Kit

Comparison of a UPLC Method across Multiple UHPLC Systems

Ludger Guide to Sialylation: II. Highly Sialylated Glycoproteins

Product Guide for LudgerTag TM 2-AA (2-aminobenzoic acid) Monosaccharide Release and Labeling Kit

Sialic Acid Determination in Glycoproteins: Comparison of Two Liquid Chromatography Methods. Experimental. Introduction. Method 1.

Lipidomic Analysis by UPLC-QTOF MS

Thermo Scientific. GlycanPac AXR-1. Column Product Manual. P/N: April, Part of Thermo Fisher Scientific

EASI-EXTRACT BIOTIN Product Code: P82 / P82B

GlycanPac AXH-1 Columns

METHOD 8316 ACRYLAMIDE, ACRYLONITRILE AND ACROLEIN BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC)

Pr oducts List Ludger Ltd Culham Science Centre Oxford OX14 3EB United Kingdom

EASI-EXTRACT MULTI-VIT B (LGE) Product Code: P183 / P183B

Title Revision n date

EASI-EXTRACT FOLIC ACID Product Code: P81 / P81B

Products Price List. US Dollars Ludger Ltd Culham Science Centre Oxford OX14 3EB United Kingdom

LC Columns - Exceed the limit. A premium inert range of LC columns delivering optimal peak shape. ProteCol -P PEEK lined

DELFIA Tb-N1 DTA Chelate & Terbium Standard

Products Price List. Euros Ludger Ltd Culham Science Centre Oxford OX14 3EB United Kingdom

Fibrlok II 2529 Universal Optical Fiber Splice

LudgerPure TM APTS Labelled IgG Glycan Library

Simultaneous Analysis of Active Pharmaceutical Ingredients and Their Counter-Ions Using a Mixed-Mode Column

Converting a CHP Method for Insulin to Agilent Poroshell 120 Columns

Rapid Gradient and Elevated Temperature UHPLC of Flavonoids in Citrus Fruit

Lipid Quantification Kit (Colorimetric)

CORESTA RECOMMENDED METHOD N 8

Analysis of Amino Acids Derived Online Using an Agilent AdvanceBio AAA Column

CYCLOSERINI CAPSULAE - CYCLOSERINE CAPSULES (AUGUST 2015)

Calibration Standards

ABIOpure TM Viral (version 2.0)

XCF TM COMPLETE Exosome and cfdna Isolation Kit (for Serum & Plasma)

Heparin Sodium ヘパリンナトリウム

TENOFOVIR TABLETS: Final text for addition to The International Pharmacopoeia (June 2010)

VITAMINES A / E IN PLASMA BY UV - FAST CODE Z18610

Ergovaline. [Methods listed in the Feed Analysis Standards] 1 Liquid chromatography Note 1, 2 [Feed Analysis Standards, Chapter 5, Section 2

ZIDOVUDINE, LAMIVUDINE AND ABACAVIR TABLETS Draft proposal for The International Pharmacopoeia (September 2006)

LEVONORGESTREL AND ETHINYLESTRADIOL TABLETS. (January 2012) DRAFT FOR COMMENT

Triglyceride Assay Kit-Bulk 5 Plate Kit Cat# TG-5-RB

INTERNATIONAL PHARMACOPOEIA MONOGRAPH ON LAMIVUDINE TABLETS

Analytical Method Development for USP Related Compounds in Paclitaxel Using an Agilent Poroshell 120 PFP

DELFIA Eu-DTPA ITC Chelate & Europium Standard

Doubling the throughput of long chromatographic methods by using a novel Dual LC workflow

Iodine HPLC Assay. Catalog Number: IOD34-H Tests For Research Use Only. Not for use in diagnostic procedures.

RITONAVIRI COMPRESSI RITONAVIR TABLETS. Final text for addition to The International Pharmacopoeia (July 2012)

A HIGH PERFORMANCE LIQUID CHROMATOGRAPHIC ASSAY FOR LERCANIDIPINE HYDROCHLORIDE

DRAFT MONOGRAPH FOR THE INTERNATIONAL PHARMACOPOEIA EFAVIRENZ, EMTRICITABINE AND TENOFOVIR TABLETS

Ultra Columns HPLC COLUMNS

DRAFT PROPOSAL FOR THE INTERNATIONAL PHARMACOPOEIA: CARBAMAZEPINI COMPRESSI - CARBAMAZEPINE TABLETS

EMTRICITABINE AND TENOFOVIR TABLETS

PAPRIKA EXTRACT SYNONYMS DEFINITION DESCRIPTION FUNCTIONAL USES CHARACTERISTICS

20S Proteasome Activity Assay Kit

LANCE Eu-W1024 ITC Chelate & Europium Standard AD0013 Development grade

retardation in infants. A wide variety of analytical methods for the analysis of

SeQuant ZIC -HILIC For all who expect more...

Carnitine / Acylcarnitines Dried Blood Spots LC-MS/MS Analysis Kit User Manual

Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products)

Caution: For Laboratory Use. A product for research purposes only. Eu-W1024 ITC Chelate & Europium Standard. Product Number: AD0013

GlycanPac AXR-1 Columns

YMC HILIC Columns for Polar Analytes

Rat Hemoglobin A1c (HbA1c) Kit Instructions

PROPOSAL FOR REVISION OF MONOGRAPH PUBLISHED IN The International Pharmacopoeia: REVISION OF ph test ABACAVIR ORAL SOLUTION (JULY 2012)

Fig.1. Denatonium benzoate (DB) chemical structure

DELFIA Tb-DTPA ITC Chelate & Terbium Standard

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract.

Ultra Columns. also available. ordering note

Rev. CO Packed Columns for Ion Chromatography

The Raptor HILIC-Si Column

ELEGANCE Chlamydia pneumoniae IgG ELISA KIT

Protect your expensive HPLC and UHPLC columns from premature failure with MAC-MOD Pre-Column Filters

RayBio DPP4 Inhibitor Screening Kit

Caution: For Laboratory Use. A product for research purposes only. Eu-W1284 Iodoacetamido Chelate & Europium Standard. Product Number: AD0014

PEEK SS Columns & Guard Columns

Application Note. Determination of Amino acids by UHPLC with automated OPA- Derivatization by the Autosampler. Summary. Fig. 1.

Sialic Acid Fluorescence Labeling Kit

Malachite Green Phosphate Assay Kit. Item No

TABLE OF CONTENTS GENERAL INFORMATION... 1

SeQuant ZIC -HILIC For all who expect more...

Application Note. Authors. Abstract. Food

Selectivity Comparison of Agilent Poroshell 120 Phases in the Separation of Butter Antioxidants

HYDROXYPROPYLCELLULOSE, LOW SUBSTITUTED Stage 4, Revision 1 CP: USP BRIEFING NOTE

Dienes Derivatization MaxSpec Kit

Morinaga Mouse C-peptide ELISA Kit

Human Glucagon-Like Peptide-1 (7-36 amide)(glp-1 (7-36 amide)) LANCE Ultra Detection Kit

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

TABLE OF CONTENTS 1.1. WARRANTY DAMAGED SHIPMENTS SERVICE INTRODUCTION UNPACKING AND INSTALLATION

Tetracycline Plate Kit (for Honey) PN 52254B

Separation of Macrocyclic Lactones (Avermectins) on FLARE C18 MM & FLARE C18+ Columns

Analysis of L- and D-Amino Acids Using UPLC Yuta Mutaguchi 1 and Toshihisa Ohshima 2*

Application Note. Abstract. Authors. Pharmaceutical

The Nitrofurantoin Capsules Revision Bulletin supersedes the currently official monograph.

Application Note. Separation of clindamycin phosphate and process impurities. Summary. Introduction. Category Pharmaceutical analysis Matrix

jetsi 10 mm sirna delivery reagent for the brain

Small Scale Preparative Isolation of Corticosteroid Degradation Products Using Mass-Based Fraction Collection Application

Transcription:

Product Guide for LudgerSep TM R1 HPLC Column for DMB labelled Sialic Acid Analysis Product # LS-R1-4.6x150 Ludger Document # LS-R1-DMB-Guide-v5.1 Ludger Ltd Culham Science Centre Oxford OX14 3EB United Kingdom Tel: +44 1865 408 554 Fax: +44 870 163 4620 Email: info@ludger.com www.ludger.com

Contents Page Contents... 2 Specifications for LudgerSep R1 Column... 3 HPLC System Requirements... 4 Installation of the Column... 4 Preconditioning of the Newly Installed Column... 5 Column Cleaning and Storage... 5 Analysis of DMB-labelled Sialic Acids... 5 References... 6 Warranties and Liabilities... 6 Troubleshooting Guide... 7 Document Revision Number... 7 Ludger Limited, 2017 Page 2

Specifications for LudgerSep R1 Column Applications Analysis of sialic acid variants labelled with DMB (1,2-diamino-4,5 methylenedioxybenzene.2hcl) Description The LudgerSep R1 HPLC column contains particles with an octadecylsilane coating optimized for hydrophobic chromatography. Particles 3 µm silica derivatized with octadecylsiliane coating. 120 Angstrom pore size. Column Size Cat # Diameter x Length Column Volume LS-R1-4.6x150 4.6 x 150 mm 2.49 ml Column Tube Stainless steel Flow Rates Typical flow rates = 0.3 to 2.0 ml/min. Pressure Pressure should not exceed 2000 psi. ph Range 2-8 Temperature Typical operating temperature, 30 C, but increasing the temperature may improve resolution for some samples. Temperature range = 15 to 50 C. Solvents A typical solvent system for DMB labelled sialic acid analysis uses 7:9:84 methanol:acetonitrile:water as solvent A and acetonitrile as solvent B. Solvents should be degassed. Avoid strong oxidants and anionic detergents. Column Protection It is good practice to install a 0.2 µm in-line filter in front of the column. Suitable Samples DMB labelled sialic acids Sample Filter samples, if required, through a 0.2 µm filter. Avoid exposure of DMB labelled samples to light. Preparation Dilute samples in water (see LT-KDMB-A1 guide). Sample Detection Fluorescence: Excitation: 373 nm, Emission: 448 nm Ludger Limited, 2017 Page 3

Handling: Ensure that any glass, plasticware or solvents used are free of glycosidases and environmental carbohydrates. Use powder-free gloves for all sample handling procedures and avoid contamination with environmental carbohydrate. Safety: Please read the Safety Data Sheets (SDS's) for all chemicals used. All processes involving labelling reagents should be performed using appropriate personal safety protection - eyeglasses, chemically resistant gloves (e.g. nitrile), etc. - and where appropriate in a laboratory fume cupboard For research use only. Not for human or drug use HPLC System Requirements LudgerSep R1 columns can be used with any HPLC pumping system capable of delivering accurate gradients at a flow rate of 0.3 to 2.0 ml/min. In general, systems that mix solvents at high pressure (after the pump head) have lower dead volumes and supply more accurate gradients that are appropriate at the flow rate needed for LudgerSep columns. We recommend keeping sample injection volumes at or below 25 µl, ideally use full loop injection. Although an example chromatogram is shown in this guide, retention times will vary dependent on the HPLC system used. Installation of the Column During column installation we recommend that: You should connect the LudgerSep R1 column to your HPLC system using standard 1/16 OD tubing and Valco compatible fittings in either stainless steel or PEEK (polyetheretherketone). Handtight PEEK fittings and tubing (0.17 mm / 0.007 ID) are recommended for ease of connection and to minimise damage to the column threads. Flow direction is indicated by an arrow on the column label. Keep the lengths of tubing between the injector to column and column to detector as short as possible to minimise dispersion effects. Install a 0.2 µm in-line filter with minimal dead volume either immediately before the injector or between the injector and the head of the LudgerSep R1 column to prevent damage to the column by particles. Before analysing any samples, the newly installed column should be conditioned using the protocol described below. Ludger Limited, 2017 Page 4

Preconditioning of the Newly Installed Column The following preconditioning step is recommended prior to use of the column: For DMB labelled sialic acid analysis, flush the column at a flow rate of 0.5 ml/min with 7:9:84 methanol:acetonitrile:water. Column Cleaning and Storage After heavy use, the LudgerSep R1 column may become contaminated with strongly adsorbed sample constituents that will lead to a loss in column performance. Peptide or other components in the sample may cause retention times to shift over a period of time. To keep the column in good condition we recommend cleaning with 10% 7:9:84 methanol:acetonitrile:water, 90% acetonitrile at the end of each run (see methods below). A more prolonged wash for 30 min may be required for heavy contamination. The LudgerSep R1 column should be stored in a low aqueous solvent. We recommend acetonitrile (minimum 50%). Analysis of DMB-labelled Sialic Acids For release of sialic acids from glycoproteins and labelling with 1,2-diamino-4,5-methylenedioxybenzene.2HCl (DMB), please follow the guide provided with the LudgerTag DMB Sialic Acid Labelling Kit (Cat. No. LT- KDMB-A1). Prepare the LC system. Ensure that the solvent lines are primed. Solvent A = Acetonitrile:Methanol:Water 9:7:84 Solvent B= Acetonitrile Fluorescence: Excitation: 373 nm, Emission: 448 nm Column temp = 30 C; Sample temp = 10 C. Time (min) Flow ml/min %A %B 0 0.5 100 0 19 0.5 100 0 19.5 0.5 10 90 23.5 0.5 10 90 24 0.5 100 0 30 0.5 100 0 Table 1. 30 minute running method for HPLC analysis using a LudgerSep-R1 column column (4.6 x 150 mm, 3 µm particles) LS-R1-4.6x150. Injection volume = 25 µl. Ludger Limited, 2017 Page 5

The DMB labelled sialic acid reference panel (a component of the DMB sialic acid labelling kit) is a good system suitability standard to run on the LudgerSep R1 column to ensure efficient performance of the column for DMB labelled sialic acid identification. An example DMB sialic acid reference panel chromatogram is shown (Figure 1). 1 2 6 Fluorescence 4 * * 3 5 7 0.00 2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 Minutes Figure 1: Chromatogram of DMB Labelled Sialic Acid Reference Panel (CM-SRP-01), run on the LudgerSep R1 HPLC column. Peaks: 1 = Neu5Gc; 2 = Neu5Ac; 3 = Neu5,7Ac 2; 4 = Neu5Gc,9Ac; 5 = Neu5,8Ac 2; 6 = Neu5,9Ac 2; 7 = Neu5,x,xAc 3 (where x is an unknown acetyl position); * = Reagent Note: This chromatogram is provided as an example only. Peak width, resolution and retention are dependent on the HPLC system setup in your laboratory. References 1. Ludger Document: DMB-kit-Validation-Report-GP-0057-v1.0. Validation of the DMB kit with Ludger Standards. 2. Ludger Document: S-GP-0048-WG-50381-Report-v1.0. Determination of the effect of freezing of DMB Labelled Sialic Acids. Warranties and Liabilities Ludger warrants that the above product conforms to the attached analytical documents. Should the product fail for reasons other than through misuse Ludger will, at its option, replace free of charge or refund the purchase price. This warranty is exclusive and Ludger makes no other warrants, expressed or implied, including any implied conditions or warranties of merchantability or fitness for any particular purpose. Ludger shall not be liable for any incidental, consequential or contingent damages. This product is intended for in vitro research only. Ludger Limited, 2017 Page 6

Troubleshooting Guide 1. Low signals on HPLC. Incomplete acid hydrolysis: We recommend using an oven rather than a heating block for the acid hydrolysis step. Some heating blocks cause evaporation and condensation of the acid in the sample vial lid causing incomplete acid hydrolysis. We also recommend the use of small sample vials, no greater than 0.5 ml in volume, for the acid hydrolysis. Salts in the sample interfering with labelling: Salts and buffers can interfere with the sialic acid labelling method. If you suspect salt interference with your sample, dialyse the sample into a salt free solvent before analysis. 2. High levels of free dye peaks in chromatograms. This can be caused by too much light exposure. Ensure that incubation steps are performed in the dark. Once the samples are labelled it is ideal to run them immediately on the LC to avoid degradation as prolonged exposure of samples to light and heat causes an increase in non-sialic acid specific chromatogram peaks. The issue may also be caused by contamination of the LC column over time, see below. The amounts of Neu5Ac and Neu5Gc have been shown to be stable when the DMB labelled samples are stored at in the dark at 10 C for up to 72 hours, provided that the calibration standards have been stored in the same conditions and are analysed at the same time [Ref 1]. If this is not possible then the DMB labelled samples can be frozen for up to 2 days [Ref 2]. 3. Variation in LC chromatogram peaks retention times; unstable baseline. Incorrect or old LC solvent. Always prepare the solvents in the same way (making a solvent up to one litre in a measuring cylinder, for example, by mixing two solvents together, is not the same as measuring out the two solvents separately and mixing in a bottle). Isocratic gradients are particularly sensitive to variations in solvent preparation. Solvent composition can change over time due to evaporation. Contamination of the column with excess free dye/ peptides etc can lead to retention time shifts and extra peaks on the chromatogram. This can be more of a problem for sample with low levels of sialylation where larger amounts of protein are injected onto the columns. Wash the column at the normal flow rate with a 10:90 mixture of normal running solvent and acetonitrile. 4. Problem: There is precipitate in the labelling solution Although rare, it is possible that a slight precipitate may form during the preparation of the DMB labelling solution (mixture of sodium dithionite, mercaptoethanol and DMB dye). We have also observed this occurrence, have tested this mixture for its labelling efficiency and can confirm that the precipitate does not impact the labelling reaction. Document Revision Number Document # LS-R1-DMB-Guide-v5.1.doc Ludger Limited, 2017 Page 7