Dialysate Composition: Accuracy of the Prescription

Similar documents
Is Your Patient Receiving the Treatment You Prescribed?

Fresenius Medical Care North America Corporate Headquarters Medical Department To: FMCNA Medical Directors 95 Hayden Ave

AK 200 S AK 200 ULTRA S. A better way to better care

Paul Smith CCNT Corporate Director of Biomedical Support Services.

MEDICAL EQUIPMENT II HEMODIALYSIS

Innovations for a better world. Mixing plants. For premix and ready mix formulas.

Solutions for Calibration and Maintenance

Crit-Line Monitor. Frequently Asked Questions

Setting up deviation, incident, nonconformance. Presented by Debbie Parker 4 July, 2016

What is renal failure?

28-Feb-10 Reza Sabagh

Advances in Papermaking Wet End Chemistry Application Technologies

Gambro Renal Products Technical Assistance Services

Commercial Processing Example: Pickled Herring

UW MEDICINE PATIENT EDUCATION. Peritoneal Dialysis. A treatment option for kidney disease. There are 2 types of PD: continuous ambulatory

Curing as a Single Special Process Regulatory Agency Jurisdiction NAME (fill in form)

JJF Translated English of Chinese Standard: JJF

Material balance calculations involved with dilution and mixing

Active UMMC Protocols

Intravenous Fluid and Drug Therapy

Package leaflet: Information for the user. multibic potassium-free solution for haemodialysis/haemofiltration

On-site production of a dialysis bath from dry salts. Results of solute concentration control by routine clinical chemistry

Know The Facts About Home Dialysis Choices

INFORMATION TOPIC: II-5 OR DEMONSTRATION: II-5. DOSAGE, MEASUREMENTS, AND DRUG FORMS (Lesson Title) OBJECTIVES THE STUDENT WILL BE ABLE TO:

No. GA Page 1 of 7 Effective Date: 03/04/ CFR (c), ANSI/ISEA Z TITLE: Chancellor Date: 03/04/16

Critical Care Services: Equipment and Procedures Information for Patients, Relatives and Carers

Functional Phosphates & Non Phosphates for Seafood

water & dialysate testing services

Implementing therapy-delivery, dose adjustments and fluid balance. Eileen Lischer MA, BSN, RN, CNN University of California San Diego March 6, 2018

HI Total Acidity Mini Titrator for Water Analysis

A VIDEO SERIES. living WELL. with kidney failure WHAT IS KIDNEY FAILURE?

040716_ANDI_VENTIS WEBSITE(new) Printout 1. HOME PAGE. Welcome to our world of Dry Powder Inhalers!

Hemocontrol. Management of Intradialytic Hypotension

Innovative method for Vitamin Analysis at point of production

Experience of paediatric formulations in Marketing Authorisatrion Applications

Sterile U Network. Quality Control for Table-top Steam Sterilizers. Introduction: T U T O R I A L S

Care Chemicals & Formulators. Librel BMX. Technical Information. = Registered trademark of BASF group Multi-Micronutrient EDTA Chelate.

IPEC Europe Suggested Alternative (if none then original text is clear and needs no alteration) Purpose and Scope

A reliable, practical and easy-to-use Determination of Soil Conductivity and Salt Content with TRIME Probes

ADVANCED ASSESSMENT Medical Math

APIC IC Challenges in Dialysis Danilo B. Concepcion,CBNT, CCHT-A Operations Manager, Renal Service

Within the proportion the two outside numbers are referred to as the extremes. The two inside numbers are referred to as the means.

Reagent Set DAS ELISA, Alkaline phosphatase label SRA 22001, SRA 23203, SRA 27703, SRA & SRA ToRSV, ArMV, GFLV, AnFBV and PDV

SureFlow. Oxygen Flow Station PROVIDER TECHNICAL MANUAL

Understanding Your Hemodialysis Access Options UNDERSTANDING YOUR HOME HEMODIALYSIS OPTIONS

D5. Feed Handling. D5.1 Medicated Feed

ph Technical Calibration Solutions solutions Technical Solutions (±0.01 ph) for Each Point of the ph Scale Easy to Use Single Dose Sachets

Measuring Sodium in dialysis patients. UCL Center for Nephrology

Modes of Extracorporeal Therapies For ESRD Patients

All About Aerosols. Jim Hammer Pharmasol Corporation

Eyewash and Safety Showers HM 20 Page 1

YY/T Translated English of Chinese Standard: YY/T PHARMACEUTICAL INDUSTRY STANDARD

Managing Allergens within a manufacturing context. Helen Brown CampdenBRI Chipping Campden

Hemodialysis AK 98 MONITORS Artis Dialysis System AK 98 Self-Care

Comparison of Oxalic Acid Cleaning Results at SRS and Hanford and the Impact on Enhanced Chemical Cleaning Deployment

Kidney Failure. Haemodialysis

Instructions for Performing In-Office Lab Tests

Ingredient Listing Qty. Unit NDC # Supplier g

STANDARD FOR PORCINE SEMEN QUALITY IN AI CENTRES

Chemistry 151 Last Updated: Dec Lab 2: Measurements & Chemical Changes

FAO Specifications for Plant Protection Products PROFENOFOS (AGP:CP/359)

FORTIFICATION OF MILLED MAIZE PRODUCTS

WHAT DO I NEED TO KNOW ABOUT MY VASCULAR ACCESS? FISTULA/GRAFT

CHAPTER 1 COMMUNITY PHARMACY M.ASHOKKUMAR DEPT OF PHARMACY PRACTICE SRM COLLEGE OF PHARMACY SRM UNIVERSITY

Calcium Management for Patients Receiving Extended Duration Hemodialysis

FOR PLANT PROTECTION PRODUCTS FOOD AND AGRICULTURE ORGANIZATION OF THE UNITED NATIONS

Excipient Considerations for Continuous Manufacturing Implementation

DETERMINATION OF TRACE ELEMENTS IN WATERS AND WASTES BY INDUCTIVELY COUPLED PLASMA - MASS SPECTROMETRY EPA Method (Revision 5.

Haemodialysis. The AutoFlow Function for the 5008 Therapy System Optimising the Dialysis Fluid Flow Rate

Drivers have GPS. Dialysis Patients have. Making possible personal.

There are no shortcuts to Dialysis

Nonparenteral medications

Manufacturing Facility Brownwood, TX

Blueberry scorch virus Reagent Set DAS ELISA, Alkaline phosphatase label for BlScV Catalog number: SRA 19100

HyProBlack Hexavalent chrome free RoHS and ELV compliant. HyPro System

Reference ID:

Sterilization of health care products - Radiation - Part 3: Guidance on dosimetric aspects

The Kidney & Its Role in Maintaining Homeostasis. A lesson on the importance of,, &

UNDERSTANDING THE CRRT MACHINE

Hemodialysis today has evolved

Non Destructive Testing Services

FAO SPECIFICATIONS FOR PLANT PROTECTION PRODUCTS DNOC. 4,6-dinitro-o-cresol WITH PETROLEUM OIL PRODUCTS

Reef Care Program - FAQ

CRRT Fundamentals Pre-Test. AKI & CRRT 2017 Practice Based Learning in CRRT

First steps for success.

Provläsningsexemplar / Preview INTERNATIONAL STANDARD. Rubber condoms - Part 1: Requirements

Allergens Assessing risk and mapping your facility

Partner with the Global Leader in Drug Delivery Systems

Ingredient Listing Qty. Unit NDC # Supplier. q.s. to ml

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE

Understanding Your Fluid Status During Dialysis and the Crit-Line. IV Monitor

Capturing lightning In A Bottle

Aquamimicry (2.0) Shrimp Farming Protocol

STANDARD FOR PORCINE SEMEN QUALITY IN AI CENTRES

YaraTera KRISTALON. Premium water soluble NPK fertilizer. Growth stage based formulas. Top grade Crop Nutrition

EASIMIP TM PATULIN Product Code: P250 / P250B

Dialysis System DBB-07

Reading and Analyzing your Fertilizer Bag. Dr. Cari Peters Vice President

Module 8: Practice Problems

BPG 03: Continuous Renal Replacement Therapy (CRRT)

Transcription:

Rx Tx Dialysate Composition: Accuracy of the Prescription NANT 2016 Annual Symposium March 23, 2016 Las Vegas, NV Joe Johnston Sr. Vice Pres. Biomedical Services Fresenius Medical Care

The Starting Point The role of the biomedical technician is to: a) protect patients, through proper practice in the repair and maintenance of the technology, b) bring transparency to the technology, so medical professionals can deliver the best care For many in the industry, technical is a black box.

The Voice of the Customer The VOC is an important element in product design. Question : If the manufacturers are going to ask you for what you want in new products and technology for dialysis, what do you say? a) Faster b) Cheaper c) Easier d) Better What does better mean?

Selected highlights of the paper: Small study (4 clinics) The range of difference (prescribed vs. measured) was -13 to + 6 meq/l of Na + On average, the sodium measured was 2-4 meq/l higher than the nominal (prescribed) Dialysate loop delivery pressures were not measured. Differences were smaller in units using one, predominantly prescribed dialysate Na + Multiple factors affecting delivered Na + are listed

Isn t the AAMI standard for dialysate sodium +/- 2.5 % from nominal value?

Dialysis Fluid Composition Standards & Regulations RD62 - ANSI/AAMI/ISO 13959:2009 Water for hemodialysis and related therapies RD61 ANSI/AAMI/ISO 13958:2009 Concentrates for hemodialysis and related therapies ANSI/AAMI/ISO 11663:2009 Quality of dialysis fluid for hemodialysis and related therapies RD5 - ANSI/AAMI/IEC 60601-2-16:2012 - Particular requirements for basic safety and essential performance of hemodialysis, hemodiafiltration and hemofiltration equipment RD52 - ANSI/AAMI/ISO 23500:2011 Guidance for the preparation and quality management of fluids for hemodialysis and related

Dialysis Fluid Composition ANSI/AAMI/ISO 13958:2009 Concentrates for hemodialysis 4.1.2.1 Liquid solute concentrations All electrolytes identified on the label shall be present within ± 5% or ± 0.1 meq/l (expressed as dialysis fluid concentrations), whichever is greater, of the stated concentration, with the exception of sodium, which shall be present within ± 2.5 % of the labeled concentration 4.1.2.2 Solute concentrations from powder When concentrate is packaged in dry form or a combination of dry and liquid and is mixed according to the manufacturer's instruction for use, the final concentrate shall meet the requirements of 4.1.2.1. 5.4 Systems for mixing concentrate at a dialysis facility * The manufacturer must only meet their stated specifications *

Dialysis Fluid Composition ANSI/AAMI/IEC 60601-2-16:2012 - Particular requirements for basic safety and essential performance of hemodialysis, hemodiafiltration and hemofiltration equipment Subclause 201.12.4.4.101 Composition of the DIALYSIS FLUID The requirement for a PROTECTIVE SYSTEM is also applicable to human errors (e.g. mistaking of DIALYSIS FLUID CONCENTRATEs) and also refers to Clause 15 (Construction of ME EQUIPMENT) and Clause 16 (ME SYSTEMS). In acetate treatment, it is considered to be appropriate if the PROTECTIVE SYSTEM is designed such that it prevents a deviation beyond the following limits: conductivity of final DIALYSIS FLUID 12 ms/cm 16 ms/cm sodium in DIALYSIS FLUID ±5 % from set point Additionally in bicarbonate treatment: bicarbonate in DIALYSIS FLUID ±25 % from set point If other components can be added individually, additionally: other electrolytes in DIALYSIS FLUID ±20 % from set point

Dialysis Fluid Composition ANSI/AAMI/ISO 23500:2011 - Guidance for the preparation and quality management of fluids for hemodialysis and related therapies 7.5.1 Monitoring of mixing systems Systems for preparing either bicarbonate or acid concentrate from powder or other highly concentrated media at a dialysis facility should be monitored according to the mixing system manufacturer's instructions to ensure appropriate dissolution. If a facility designs its own system, procedures should be developed and demonstrated to ensure proper mixing of the concentrate, including establishment of acceptable limits for tests of proper concentration. In this case, verification can be accomplished by testing a sample from each batch prepared over a 3 day period. Acid and bicarbonate concentrates may be tested by measuring conductivity, density with a density meter or specific gravity with a hydrometer according to the manufacturer's instructions. Although not required, some manufacturers of concentrate powder or other highly concentrated media may provide allowable ranges for either the conductivity, density or the specific gravity of concentrates prepared from their powder.

When it was really a bath (the batch of dialysate), the existing standard worked (only volume of water and volume of concentrate varied) Now, other sources of error have been added: Machine proportioning (pump calibration, intake resistance and backpressure, servo-sensor calibration, servo-control lag time) Bicarb dialysate (now 2 parts to the concentrate, each has sodium) On-site mixing of one or both concentrates from powder Central delivery of concentrates (esp. pressures) The goal is NOT a final dialysate conductivity, but correct chemical composition [GIGO]

Dialysis Fluid Composition Nominal 137 Na (100 from acid & 37 from bicarb) Water AAMI/ISO 3.0 meq Mfg 3.0 meq* Acid (liquid/dry) 2.5% (2.5mEq/L) 2.5% (2.5mEq/L) Bicarbonate Concentrate Distribution (liquid/dry) Mixers Systems 2.5% (0.9mEq/L) N/A N/A 2.5% (0.9mEq/L)?? Dialysis Machines 5.0% (6.9mEq/L) 5.0% (6.9mEq/L) What we believed 2.5% (3.4mEq/L) * RO alarms significantly lower Ask your Nephrologist, what is an allowable deviation from prescription for delivered Na? 1 meq/l 3 meq/l 5 meq/l 7 meq/l

AAMI TIR 72 RDN469:2013 (draft) Hemodialysis Fluid Chemical Composition The objective is to provide dialysis practitioners additional background information related to the dialysis processes described in ANSI/AAMI 23500, Guidance for the preparation and quality management of fluids for hemodialysis and related therapies, that affect the chemical composition of dialysis fluid.

Dialysis Fluid - Process Flow Liquid Hemodialysis Concentrates System Diagram Manufacturing (Concentrates) Clinic (Including Hemodialysis Machine) Laboratories (XYZ Labs) Start Concentrate Stored at Clinic AAMI Quality Water System Samples Received at Lab Incoming Raw Material Acid Conc. (Drm/Jug) Bicarb. Conc. (Drm/Jug) Water Quality Monitoring Sodium Analysis by Dilution & ISE Liquid Acid Concentrate Manufacture Release Testing Inspection * In-house Lab Warehouse Liquid Bicarb. Concentrate Manufacture Release Testing Acid Metering Pump Hemodialysis Machine Bicarb. Metering Pump Mixing Chamber Patient Water Metering Dialysate Sample Blood Sample Data Storage Color Key Acid Concentrate Path Bicarbonate Path Final Dialysate Solution Sodium Data Collection Points

Solid (Dry Powder) Hemodialysis Concentrates System Diagram Manufacturing (Concentrates) Clinic (Including Hemodialysis Machine) Laboratories (XYZ Labs) Start Acid Reconstitution Procedure Bicarbonate Reconstitution Procedure Samples Received at Lab Incoming Raw Material Concentration Verification Concentration Verification AAMI Quality Clinic Water System Sodium Analysis by Dilution & ISE Inspection Clinic-wide Acid Conc. Dist. System Clinic-wide Bicarb. Conc. Dist. System Water Quality Monitoring Data Storage Dry Acid Conc. Manufacture Dry Bicarbonate Powder (NaHCO 3 ) Acid Metering Pump Bicarb. Metering Pump Water Metering Release Testing * * In-house Lab Release Testing * Dialysate Mixing Chamber Color Key Warehouse Hemodialysis Machine Patient Dialysate Sample Blood Sample Acid Concentrate Path Bicarbonate Path Final Dialysate Solution Sodium Data Collection Points

Sources of Error Step through the process flow chart where are the sources of error! Correct formulas RO water Powdered concentrate mixing Concentrate delivery pressure Distribution loop feed pressure Level/height of concentrate jug Machine Calibration of pumps and sensors Random pumping errors

Right Formulation? Too many formulas available in the clinic? Storage conditions well controlled? No mix ups! Good control on jug labeling, cleaning? Connected to the right wall outlet? Wall box plumbing Is 2 K coming from the 3 K outlet? How do you know? Do you check at install/repair

RO water Electrolytes normally included in dialysis fluid Ca +2 Mg +2 K - Na - 0.05 mmol/l 0.15 mmol/l 0.2 mmol/l 3.0 mmol/l The Water? Do you monitor / trend the sodium? Materials used in water systems Everyone knows about aluminum pump heads. How about plastic colorant in tubing, micro-static plastics (imbedded silver salts), gaskets, o-rings and lubricants? Does your policy call for only OEM parts?

Powdered concentrate Powder to liquid? Dissolution Gallons vs. liters, 10 Tx vs 12 Tx bags Water volume check how is the tank calibrated? Powder fully dissolved? How is dilution during delivery prevented? Following IFU s for times, temperature, testing? What would your clinic do if a little powder is spilled? Would you dump the batch and start over? Would conductivity catch an error in powder addition? Mixing and storage Control procedure for the cases? Is 2 K being pumped to the 3 K tank? Do you use manifolds?

Dialysis Fluid Composition Answer Dialysis machine manufacturers are required by ANSI/AAMI/IEC standards to prevent deviations of bicarbonate concentration in delivered dialysate from exceeding: a) 25% b) 5.0% c) 2.5%

Delivery? Concentrate central delivery Distribution loop feed pressure Level/height of concentrate jug Fresenius machines: -1.3 to + 2 psi

The Machine? Machine Proportioning: Calibration of pumps and sensors Are there limits on the entered value for pump calibration and sensor settings? Intake resistance and backpressure Random pumping errors Machine Sensors: Calibration technique and method Servo-sensor calibration Servo-control lag time Machine Alarms and Alarm Resets: * Know the effect and consequence of each alarm *

The Lab? Lab testing errors can be 1-2%. Don t chase ghosts.

Dialysis Fluid Composition Things to Remember Current Standards / Regulations do not define parameters for every source of error in the preparation of dialysis fluid Users should understand the limitations of equipment, processes, and tests involved in the preparation and quality monitoring of dialysis fluid User may wish to consider evaluating variation even within current limits of protective systems to determine if composition of dialysis fluid meets expectations.