High performance graphite furnace tube for determination of lead in blood

Similar documents
Determination of Copper in Green Olives using ICP-OES

Analysis of Blood Serum on the Liberty Series II ICP OES with the Axially-Viewed Plasma

Analysis of Micronutrients in Milk by Flame Atomic Absorption Using FAST Flame Sample Automation for Increased Sample Throughput

Determination of Calcium in Serum Samples by AAS Using a Fuel Lean Flame

Analysis of Blood Serum on the Liberty Series II ICP OES with the Axially- Viewed Plasma

Trace Elemental Characterization of Edible Oils with Graphite Furnace Atomic Absorption Spectrophotometer

Guidelines for Troubleshooting and Maintenance of AA Systems. Presented by Eric Vanclay, Atomic Spectroscopy Consumables Product Manager

Evaluating the performance of the OneNeb Series 2 Nebulizer with PerkinElmer Optima 7/8x00 Series ICP-OES systems

Application note. Determination of metals in soil by microwave plasma - atomic emission spectrometry (MP-AES) using DTPA extraction.

Highly Repeatable Ultra Low Detection of Estradiol Using Triple Quadrupole GC/MS in NCI Mode

Determination of metals in industrial wastewaters by microwave plasmaatomic emission spectrometry

Application Note novaa 800 D. Determination of Macro and Trace Minerals as well as Toxic Trace Metals in Powdered Milk. Challenge.

V. LAB REPORT. PART I. ICP-AES (section IVA)

Baby Food and Juice Compendium BABY FOOD AND JUICE

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Determination of available nutrients in soil using the Agilent 4200 MP-AES

Determination of Tetracyclines in Chicken by Solid-Phase Extraction and High-Performance Liquid Chromatography

Analyze Barbiturates in Urine with Agilent 6430 LC/MS/MS and Poroshell 120 EC-C18

Estimation Of Copper, Zinc And Lead In Choornas By Flame Atomic Absorption Spectroscopy Krishna veni N 1 *, Meyyanathan S N 1 and Suresh B 2

Toxicology Blood Lead Comprehensive Testing

Performance Characteristics of the Agilent High Matrix Sample Introduction (HMI) Accessory for the 7500 Series ICP-MS. Product Overview.

Toxicology Blood Lead Comprehensive Testing

Selenium, atomic absorption spectrometric, hydride, automated

ACQUITY UPLC WITH PDA DETECTION: DETERMINING THE SENSITIVITY LIMITS OF OXYBUTYNIN AND RELATED COMPOUNDS

Analysis of Omega 3 and Omega 6 FAMEs in Fish Oil and Animal Fat Using an Agilent J&W DB-FATWAX Ultra Inert GC Column

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

Toxicology Blood Lead Comprehensive Testing

A significant dietary source of lead to Nunavik Inuit: Ptarmigan hunting

Elemental Scientific. seafast S2. Elution Profiles. Elemental Scientific

CORESTA RECOMMENDED METHOD N 8

Residue Monograph prepared by the meeting of the Joint FAO/WHO Expert Committee on Food Additives (JECFA), 82 nd meeting 2016.

Analysis of Cholesterol-Lowering Drugs (Statins) Using Dried Matrix Spot Technology

Application Note. Author. Abstract. Introduction. Food Safety

The certification of the mass concentrations of aluminium, selenium and zinc in human serum BCR-637 BCR-638 BCR-639

Pelagia Research Library

Measurement of copper, zinc and magnesium in serum and urine by DC plasma emission spectrometry

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

Analysis of Organic Fertilizers for Nutrients with AAnalyst 800 Atomic Absorption Spectrophotometer

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System

Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products)

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

Rat cholesterol ELISA Kit

EliKine Free Thyroxine (ft4) ELISA Kit

Determination of Benzodiazepines in Urine by CE-MS/MS

Analysis of trace elements in nutraceuticals in compliance with USP chapter <2232> Elemental Contaminants in Dietary Supplements

TOP 7 ACTIVITIES OF CEN AND ISO STANDARDIZATION OF METHODS FOR ANALYSIS OF ANIMAL FEEDS

SUPPLEMENTARY MATERIAL

Development and validation of stability indicating RP-LC method for estimation of calcium dobesilate in pharmaceutical formulations

Rapid Analysis of 37 FAMEs with the Agilent 8860 Gas Chromatograph

ESS Method 310.2: Phosphorus, Total, Low Level (Persulfate Digestion)

METHOD 8316 ACRYLAMIDE, ACRYLONITRILE AND ACROLEIN BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC)

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Supplementary materials for manuscript entitled Development and Analytical

World Journal of Pharmaceutical Research

LC-MS/MS Method for the Determination of Tenofovir from Plasma

Optimization of the Mark VI Flame Atomization System

MULTI-COMPONENT ANALYSIS OF HEAVY METALS

Analysis of Milk for Major and Trace Elements by ICP-MS

CORESTA Recommended Method No. 78

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

HEPARIN SODIUM. Heparinum natricum

Trace Metal Characterization of Soils Using the Optima 7300 DV ICP-OES

Simultaneous Analysis of Active Pharmaceutical Ingredients and Their Counter-Ions Using a Mixed-Mode Column

US EPA SW-846 Method 6010D using the Thermo Scientific icap 7400 ICP-OES Duo

Agilent 1220 Infinity LC System Performance Specifications

Quantitative Analysis of Opiates in Urine Using RRHT LC/MS/MS. Application. Authors. Introduction. Abstract. Forensics

Insulin ELISA. For the quantitative determination of insulin in serum and plasma

Quantitative Method for Amphetamines, Phentermine, and Designer Stimulants Using an Agilent 6430 LC/MS/MS

Analysis of Chloride, Phosphate, Malate, Sulfite, Tartrate, Sulfate, and Oxalate in Red and White Wine

Available Online through Research Article

New RP - HPLC Method for the Determination of Valproic acid in Human Plasma

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

The Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis

JPSBR: Volume 4, Issue 6: 2014 ( ) ISSN NO

Fluorescent Carbon Dots as Off-On Nanosensor for Ascorbic Acid

F. Al-Rimawi* Faculty of Science and Technology, Al-Quds University, P.O. Box 20002, East Jerusalem. Abstract

Matrix Interferences in ICP-MS: Causes, Effects, and Strategies to Reduce or Eliminate Them

GB Translated English of Chinese Standard: GB NATIONAL STANDARD

Available online Research Article

TSH Receptor Autoantibody ELISA

USP <232> and <233> Understanding Your Path to Compliance with the New Elemental Impurity Chapters. Steve Wall Agilent Technologies

Application Note. Abstract. Authors. Pharmaceutical

Forensic Analysis of Blood Alcohol Concentration

LC-MS/MS analysis of Chlorates in Milk and Whey Powder using the Agilent 6470 QQQ

Evaluation of the Mark-VI Spray Chamber for Flame Atomic Absorption Spectrometry

Application Note. Authors. Abstract. Food

Purity Tests for Modified Starches

Stability indicating RP-HPLC method development and validation of Etizolam and Propranolol hydrochloride in pharmaceutical dosage form

Insulin (Porcine/Canine) ELISA

Analysis of Organic Acids and Alcohols Using the Agilent J&W DB-624UI Ultra Inert GC Column

Determination of Amantadine Residues in Chicken by LCMS-8040

Mercury, Cadmium and Lead Levels in Three Commercially Important Marine Fish Species of in Sri Lanka

ARTICLE. Accuracy and Reproducibility of Blood Lead Testing in Commercial Laboratories

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE

LC Columns - Exceed the limit. A premium inert range of LC columns delivering optimal peak shape. ProteCol -P PEEK lined

HbA1c (Human) ELISA Kit

Determination of B-vitamins in Energy Drinks by CE/MS/MS

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000

Transcription:

High performance graphite furnace tube for determination of lead in blood Application note Clinical research Authors Kai Robinson, John Sanders, James Barker Agilent Technologies, Melbourne, Australia Introduction With the development of Graphite Furnace-Atomic Absorption Spectroscopy (GF-AAS) there has been continued development of graphite tubes and platforms. Early tubes used uncoated graphite, had poor lifetime, memory and low sensitivity. Development of pyrolytically coated tubes improved resistance to oxidation. Coated tubes produced superior lifetime, sensitivity and precision of analytes. The next advance in tube design was the introduction of a graphite platform ( L vov platform ) onto which the sample is injected for atomization. The use of a platform delays atomization until the gas environment has reached a stable temperature relative to the tube wall. This prevents condensation and recombination of the analytes in the gas phase. A variety of platform designs have been utilized over the years, some of these having limitations such as small sample volume and restricted atomization temperature limits.

This application note describes a graphite tube design that addresses these limitations, resulting in improved sensitivities and detection limits compared to previous tube designs. This design is called the Agilent Omega Platform Tube. Exposure to lead has adverse effects on human health. The Centers for Disease Control and Prevention (CDC) state that a blood lead level of greater than 10 µg/dl (equivalent to 0.48 µmol/l) is of concern and acknowledges studies indicating evidence of adverse effects in children at lower levels [1]. In this application note the Agilent Omega Platform Tubes are evaluated for the determination of lead in blood. Deuterium and Zeeman background correction are two commonly used techniques that are suitable for this application. Zeeman back- ground correction is used for background correction in this application. This complex matrix requires good control of the furnace temperature profile. Instrumentation The instrumentation used for the determination of lead in blood was an Agilent 280Z AA with Zeeman background correction. The Agilent 280Z AA is configured for 8 lamps in fixed positions. Agilent UltrAA lamps, which boost the emission intensity for reduced noise and enhanced sensitivity, can be used with this instrument. The spectrometer was equipped with an Agilent GTA 120 graphite tube atomizer and an Agilent PSD 120 Programmable Sample Dispenser. The Agilent PSD 120 provides the capacity for up to 135 solutions and allows automatic standard preparation and over-range dilution. The furnace viewing camera, Tube-CAM, is a standard feature of the Agilent 280Z AA. Tube-CAM provides real time viewing inside the graphite tube via SpectrAA software. Still images and videos can be recorded throughout an analysis. Tube-CAM enables the dispensing height to be set accurately, confirmation of furnace conditions during method development, and monitoring of potential matrix buildup inside the tube after the atomization stage. Table 1 lists the instrument settings and Table 2 lists the furnace settings used in the analysis. The inert gas used was high-purity argon. Table 1. Instrument Settings Instrument Parameters Settings Measurement Replicates 2 Mode Peak Area Optical Lamp Pb UltrAA Wavelength 283.3 nm Slit Width 0.5R Lamp Current 10.0 ma Standards Standard 1 0.0483 µmol/l Standard 2 0.0966 µmol/l Standard 3 0.1449 µmol/l Standard 4 0.2419 µmol/l Standard 5 0.3382 µmol/l Sampler Total Volume 10 µl Sample Volume 10 µl Number of Injections 1 Table 2. Furnace Settings Step Temp ( C) Time (s) Flow (L/min) Read 1 80 5 0.3 2 95 40 0.3 3 120 5 0.3 4 250 5 0.3 5 250 5 0.3 6 400 10 0.3 7 700 10 0.3 8 700 10 0.3 9 700 2 0 10 2200 0.7 0 Yes 11 2200 1.5 0 Yes 12 2600 0.4 0.3 13 2600 2 0.3 2

Materials and Reagents A non-ionic surfactant was used to hemolyze the blood samples. A phosphate chemical modifier was used to permit the use of a higher ashing temperature, reduce matrix interferences and to stabilize the lead signal [2]. The following reagents were used for preparation of calibration solutions and for sample preparation: Lead standard solution, 1000 mg/l (SpectrosoL, BDH Laboratory Supplies, Poole, BH15 1TD, England) Non-ionic Surfactant (Triton X-100, LABCHEM, AJAX Chemicals, Sydney, NSW, Australia) High-purity nitric acid (Ultrapur, Merck, Kilsyth, Victoria, Australia) Ammonium dihydrogen phosphate (Suprapur, Merck, Kilsyth, Victoria, Australia) Class 1 Water (MW/cm) Whole Blood Metals Control (Lyphochek Levels 1, 2 and 3, BIO-RAD Laboratories, Irvine, CA, USA) Porcine Blood (EDTA Anticoagulated, Gamma Irradiated, QCT Quality Control Technologies Pty Ltd, Newcastle, NSW, Australia) Sample Preparation A non-ionic surfactant/phosphate solution was prepared from 1.5 g ammonium dihydrogen phosphate and 5 ml of 1% (v/v) Triton X-100 solution made to 100 ml with Class 1 water. A 1 mg/l lead solution (4.83 µmol/l Pb) was prepared in 0.1% nitric acid. This solution was then used to prepare the calibration solutions and matrix spike sample. Five calibration solutions and a blank were prepared by diluting 1 ml of porcine blood and the required amount of 1 mg/l lead solution to 10 ml in non-ionic surfactant/phosphate solution. Lyphochek Whole Blood Metal Controls were reconstituted per included instructions. The Certified Reference Material (CRM) solutions were then prepared by diluting 0.5 ml of reconstituted Lyphochek to 5 ml in non-ionic surfactant/phosphate solution. A Lyphochek Level 1 spike was prepared by diluting 0.5 ml of Lyphochek Level 1 and 20 µl of 1 mg/l Pb solution to 5 ml in non-ionic surfactant/phosphate solution. All calibration solutions and CRM solutions were prepared fresh daily. The rinse solution for the PSD-120 programmable sample dispenser was 0.1% nitric acid and 0.1% nonionic surfactant in Class 1 water. An alternate rinse solution of 0.1% nitric acid and 10% ethanol in Class 1 water could be used. Results and Discussion Accurate results were obtained for the spiked sample and CRMs (refer to Table 3). The method exhibited excellent corre- lation coefficient R 2 = 0.9992 (refer to Figure 1). The Omega Platform Tube exhibited good performance for this difficult matrix with minimal carbon buildup (refer to Figure 2). Table 3. Spiked Sample and CRM Results Sample Pb conc (µmol/l) Recovery (%) Porcine blood 0 N/A Porcine blood spike (0.14 µmol/l Pb) 0.14 100 Lypocheck 1 (0.41 µmol/l Pb) 0.40 98 Lypocheck 2 (1.34 µmol/l Pb) 1.31 98 Lypocheck 3 (2.50 µmol/l Pb) 2.63 105 The temperature profile of the standard and sample show the high signal to noise ratios obtained with the Omega Platform Tube (refer to Figures 3 and 4). No carry-over was observed between samples. After each analytical sequence the dispenser capillary was soaked in aqua-regia for 1 minute to remove any protein and to ensure that the capillary remained hydrophobic. This is a recommend practice for this matrix. 3

Figure 4. Temperature Profile for Lypocheck 1 SRM. Conclusion Figure 1. Calibration curve for Pb in blood showing excellent correlation coefficient, R 2 = 0.9992. The high performance Agilent Omega Platform Tube has been successfully used for the determination of lead in blood by graphite furnace AAS using an Agilent 280Z AA. The complex sample matrix produced broad-band molecular absorption which was accurately corrected using Zeeman background correction. Excellent recoveries for spiked samples (100%) and standard reference materials (98% at 0.41 µmol/l) were obtained. The benefits of the Agilent Omega Platform Tube design are applicable to a variety of other analyses. Figure 2. Image of Omega Platform Tube from the Tube-CAM viewing camera at the completion of the work. Figure 3. Temperature Profile for 0.0483 µmol/l Pb in blood standard. 4

References 1. J. L. Gerberding, et al., Preventing Lead Poisoning in Young Children, A Statement by the Centers for Disease Control, Public Health Service, Atlanta, GA (5th Revision August 2005). U.S. Department of Health and Human Services, Centers for Disease Control and Prevention Web site. Available at: http://www.cdc.gov/nceh/lead/ ACCLPP/recom- mend.htm Accessed 20th June, 2008. 2. Hinderberger, E. J., Kaiser, M. L., and Koirtyohann, S. R. At. Spectrosc. 1981, 2 (1), 1-7. 5

www.agilent.com/chem For Research Use Only. Not for use in diagnostic procedures. This information is subject to change without notice. Agilent Technologies, Inc. 2015 Published August 12 2015 Publication number: SI-1586