The Treatment of Complex Central Sleep Apnea (CompCSA)

Similar documents
A simple solution for your complex patients

Titration protocol reference guide

Titration protocol reference guide

Clinical update of BiPAP autosv for treatment of Sleep Disordered Breathing

Auto Servo Ventilation Indications, Basics of Algorithm, and Titration

COMPLEX SLEEP APNEA IS IT A DISEASE? David Claman, MD UCSF Sleep Disorders Center

Mario Kinsella MD FAASM 10/5/2016

Complex Sleep Apnea. Can we do better? David Weed D.O.,FCCP,FAASM. September 8, 2016

BiPAPS/TVAPSCPAPASV???? Lori Davis, B.Sc., R.C.P.T.(P), RPSGT

Natural performance. Introducing the BiPAP A30 - because ease of use and therapy efficacy are key to patient well-being

Chronic NIV in heart failure patients: ASV, NIV and CPAP

One intelligent solution

Helpful hints for filing

One intelligent solution

Pain patient with sleep-disordered breathing

Index. Servo ventilation...page 2 Design elements...page 8 Titration...Page 13 Case studies... Page 20. White papers...page 41

The intelligent solution. System One Sleep Therapy System uses advanced intelligence to deliver excellent care while making patient management easy

Somnolyzer 24x7. Addressing cost pressures in the sleep lab

PAP Therapy Devices: Delivering the Right Therapy To The Right Patient. Ryan Schmidt, BS, RRT Clinical Specialist Philips Respironics

DATE: 07 August 2012 CONTEXT AND POLICY ISSUES

Advanced Algorithms for Sleep Disordered Breathing. 1.5 AARC and AAST CEU credits

Enclosed on Page 5 is an authorization form to release your health information.

In-Patient Sleep Testing/Management Boaz Markewitz, MD

Choosing the Appropriate Mode of PAP Therapy in the Perioperative Setting

Recognition and Management of High Loop Gain Sleep Apnea

Best Match. Non-Invasive Ventilation. WSS Fall Objectives. BiPAP AVAPS Patient Types 9/10/2018. Advanced Algorithms - Clinical Applications

Index. sleep.theclinics.com. Note: Page numbers of article titles are in boldface type.

Enhanced care for your patients. A holistic approach to therapy management reporting

DreamStation. Provider guide

Which device for your patient?

Policy Specific Section: October 1, 2010 January 21, 2013

A 74-year-old man with severe ischemic cardiomyopathy and atrial fibrillation

Data Management of the Sleep Disordered Breathing Patient

DIGITAL AUTO-TRAK + Bi-FLEX + AUTO Bi-LEVEL THE POWER OF. Sometimes when three come together the results can be quite extraordinary. Unique even.

The International Palestinian Congress in Sleep Medicine

(To be filled by the treating physician)

Bi-Level Therapy: Boosting Comfort & Compliance in Apnea Patients

All About Positive Airway Pressure (PAP) Therapy

Monitoring and Troubleshooting Adherence to PAP Devices and Understanding Device Downloads

a new beginning in managing CSA ResMed.com/AirSolutions

Pocket Guide. System One REMstar Pro and REMstar Auto

PVDOMICS. Sleep Core. Cleveland Clinic Cleveland, Ohio

Monitoring: gas exchange, poly(somno)graphy or device in-built software?

Enhancing patient comfort, flexibility, and compliance. Therapy and compliance solutions selection guide

Sleep disordered breathing (SDB), which includes. Bilevel Positive Airway Pressure Worsens Central Apneas During Sleep*

WHAT DO YOU WANT FROM A HOME VENTILATION SYSTEM? 8322_RS_HomeNIV_brochure_v14.ind1 1 4/7/06 12:57:35

Opioids Cause Central and Complex Sleep Apnea in Humans and Reversal With Discontinuation: A Plea for Detoxification

W J R. World Journal of Respirology. Effectiveness of adaptive servo-ventilation. Abstract REVIEW. Yasuhiro Tomita, Takatoshi Kasai

Average volume-assured pressure support

Identification and Treatment of the Patient with Sleep Related Hypoventilation

Positive Airway Pressure Therapy With Adaptive Servoventilation Part 1: Operational Algorithms

11/20/2015. Beyond CPAP. No relevant financial conflicts of interest. Kristie R Ross, M.D. November 12, Describe advanced ventilation options

Therapy guide. A simple guide to managing your treatment of Obstructive Sleep Apnea

Reasons Providers Use Bilevel

3/10/2014. Pearls to Remember. 1) Consequences of OSA related to both arousals and hypoxia. 2) Arousals provoke increased

How to write bipap settings

ResScan Report Interpretation Guide. FOR CLINICAL USE ONLY V1.0 June, 2011

Κλινικό Φροντιστήριο Αναγνώριση και καταγραφή αναπνευστικών επεισοδίων Λυκούργος Κολιλέκας Επιμελητής A ΕΣΥ 7η Πνευμονολογική Κλινική ΝΝΘΑ Η ΣΩΤΗΡΙΑ

Joel Reiter*, Bashar Zleik, Mihaela Bazalakova, Pankaj Mehta, Robert Joseph Thomas

Medicare C/D Medical Coverage Policy. Respiratory Assist Devices for Obstructive Sleep Apnea and Breathing Related Sleep Disorders

Corporate Medical Policy

2/13/2018 OBESITY HYPOVENTILATION SYNDROME

SLEEP APNEA IN THE ELDERLY SLEEP THAT KNITS UP THE RAVELED SLEEVE OF CARE

Edoardo Gronda UO cardiologia e Ricerca Dipartimento Cardiovascolare IRCCS MultiMedica

My CoughAssist. A patient guide to CoughAssist E70

Novel pathophysiological concepts for the development and impact of sleep apnea in CHF.

Medical Affairs Policy

New Government O2 Criteria and Expert Panel. Jennifer Despain, RPSGT, RST, AS

CAPNOGRAPHY in the SLEEP CENTER Julie DeWitte, RCP, RPSGT, RST Assistant Department Administrator Kaiser Permanente Fontana Sleep Center

Dr. Karan Madan Senior Resident

Challenging Cases in Pediatric Polysomnography. Fauziya Hassan, MBBS, MS Assistant Professor Pediatric Pulmonary and Sleep

Positive Airway Pressure and Oral Devices for the Treatment of Obstructive Sleep Apnea

PEDIATRIC PAP TITRATION PROTOCOL

In the last decade, there has been a major change in the management

Positive Airway Pressure and Oral Devices for the Treatment of Obstructive Sleep Apnea

Positive Airway Pressure and Oral Devices for the Treatment of Obstructive Sleep Apnea

Multiple studies have consistently demonstrated an association SCIENTIFIC INVESTIGATIONS METHODS

What is SDB? Obstructive sleep apnea-hypopnea syndrome (OSAHS)

Apnea-Hypopnea- Index The new old biomarker for Sleep-Disordered Breathing. Alan S Maisel MD

Martin R Cowie Professor of Cardiology, National Heart & Lung Institute Imperial College London (Royal Brompton Hospital)

A Deadly Combination: Central Sleep Apnea & Heart Failure

Just When You Think You Have Heard Everything About Sleep Disordered Breathing

Respiratory Assist Device E0470:

Sleep and the Heart. Physiologic Changes in Cardiovascular Parameters during Sleep

Sleep and the Heart. Rami N. Khayat, MD

NON-INVASIVE VENTILATION MADE RIDICULOUSLY SIMPLE

Name of Policy: Noninvasive Positive Pressure Ventilation

BPAP 25A Training A.Giudice,RPSGT Clinical Education Manager

PORTABLE OR HOME SLEEP STUDIES FOR ADULT PATIENTS:

Polysomnography (PSG) (Sleep Studies), Sleep Center

Financial Presentation

Medical Affairs Policy

11/13/2017. Jeremy Tabak MD, FAASM Medical Director Baptist Hospital Sleep Lab Medical Director Baptist Sleep Lab at Galloway

Portable Sleep Testing in Hospitalized Patients

OSA and COPD: What happens when the two OVERLAP?

Prevalence of Sleep Disordered Breathing in Congestive Heart Failure as Determined by ApneaLink, a Simplified Screening Device

RESPIRATORY ASSIST DEVICE E0470

Pap Settings. A review of fine tuning settings For patient comfort and compliance Wendy Cook BSRT Judy Salisbury RPGST

Sleep and the Heart Reversing the Effects of Sleep Apnea to Better Manage Heart Disease

Sleep 101. Kathleen Feeney RPSGT, RST, CSE Business Development Specialist

Transcription:

The Treatment of Complex Central Sleep Apnea (CompCSA) Including Cheyne-Stokes Breathing (CSB), with Respironics BiPAP autosv Advanced therapy system Authors: Shahrokh Javaheri, MD 1, Mark Goetting, MD 2, Rami Khayat, MD 3, Jamie Goodwin, PhD 4 and Paul Wylie, MD 5 Introduction: This study was conducted to evaluate the therapeutic performance of a new servo-ventilation device (BiPAP autosv Advanced, Philips Respironics) for the treatment of Complex Central Sleep Apnea (CompCSA), Periodic Breathing (i.e., Cheyne Stokes Breathing). Both interventions successfully treated sleep-disordered breathing events as compared to previous clinical diagnostic and CPAP titration nights. Values for apnea hypopnea index, obstructive apnea index, central apnea index, and mixed apnea index were significantly lower with the BiPAP autosv Advanced compared to the BiPAP autosv device. Study design: A prospective multicenter randomized controlled trial. Setting: Professional sleep laboratories five (5) sites within the United States (US). Conclusions: The results of this study indicate that recent improvements to Philips Respironics current autosv technology successfully reduce or eliminate both CompCSA and CSB in patients presenting with these conditions. Participants: Thirty-two (32) participants with CompCSA, including Cheyne-Stokes Breathing (CSB). Measurements and results: Qualifying subjects were randomly assigned to complete two full-night attended polysomnograms (PSG) while treated with either the currently marketed device (BiPAP autosv, Philips Respironics) or the BiPAP autosv Advanced device. The BiPAP autosv Advanced includes an automatic EPAP adjustment and a modified auto backup rate. Standard sleep and breathing endpoints were evaluated and analyzed. Abbreviations: PSG polysomnogram; PAP Positive Airway Pressure; EPAP Expiratory PAP; IPAP Inspiratory PAP; CPAP Continuous PAP; BiPAP Bi-level PAP; autosv Advanced BiPAP autosv Advanced; autosv BiPAP autosv; SDB Sleep-Disordered Breathing; OSA Obstructive Sleep Apnea; CSA - Central Sleep Apnea; CSR Cheyne-Stokes Respiration; CompCSA Complex Central Sleep Apnea (SDB inclusive of Central, Obstructive, Mixed Apneas and CSR); AHI Apnea Hypopnea Index; Complex Sleep Apnea Syndrome (CSAS). Key words: Bi-level Positive Pressure Ventilation, Servo-Ventilation, Auto EPAP, Pressure Support 1 Sleepcare Diagnostics, Mason, Ohio 2 Sleep Health, Portage, Michigan 3 Ohio State University, Columbus, Ohio 4 University of Arizona, Tucson, Arizona 5 Arkansas Center for Sleep Medicine, Little Rock, Arkansas

Introduction Complex Central Sleep Apnea (CompCSA), including: Cheyne-Stokes Respiration (CSR), idiopathic CSA, Periodic Breathing (PB or altitude induced CSA), narcoticinduced CSA and CPAP-emergent Central Sleep Apnea has been and continues to be of significant interest to the sleep community. Cheyne-Stokes Respiration (CSR) is well known and has long been characterized as a cyclic waxing and waning pattern of respiration lasting approximately 50-70 seconds, resulting from ventilatory instability related to increased loop gain. Similarly, idiopathic CSA and periodic breathing have also been reviewed and characterized in the literature for some time and may lack the crescendo / decrescendo pattern of CSR and are shorter in duration, typically 30-40 seconds. Narcoticinduced CSA has been suspected for some time, but until recently has received limited exposure in the literature. Recent publications by Javaheri et al. and Webster et al. highlight this condition and explore the prevalence, respectively. i,ii Characterization of narcotic-induced CSA is ongoing, but is typically highly variable and less cyclic. iii CPAP-emergent Central Sleep Apnea was first reported by Gilmartin et al. in 2005 iv as Complex Sleep Disordered Breathing, and then as Complex Sleep Apnea Syndrome (CSAS) in 2006 by Morganthaler et al. v Lehman et al. vi again noted this condition and the specific development and effects of treatment with CPAP. This area continues to receive significant attention in the literature and throughout the sleep community. Recently, Javaheri et al. published The Prevalence and Natural History of Complex Sleep Apnea, vii which concluded that approximately 6.5 percent of patients with a primary diagnosis of OSA have central sleep apnea during CPAP titration. Of the total population of patients prescribed CPAP for long-term use, approximately 1.5 percent showed persistent central sleep apnea, i.e., their SDB was not corrected with CPAP alone. Although bi-level therapy, with or without a backup rate, may be a recommended therapy, this too has been shown to be insufficient based on data from Morganthaler et al. viii As aptly pointed out by Eckert et al., While the precise precipitating mechanisms involved in the various types of CSA may vary considerably, unstable ventilatory drive during sleep is a principal underlying feature. ix Characterization of the physiology and potential phenotyping of patients exhibiting these types of SDB have not been completely elucidated and are not the focus of this paper. The intent of this current study was to evaluate two auto servo-ventilation devices on patients who demonstrated some form of persistent complex central sleep apnea, including Cheyne-Stokes Breathing (CSB), that was not successfully treated with standard CPAP therapy. This study included participants with the following medical diagnoses: Heart disease (including CHF, A-fib, s/p MI, pacemaker, CAD, HTN, and Dysrhythmias), CVA, pain (including neuropathic, arthritic, and cancer), COPD, GERD and diabetes. Frank CSR was identified in at least seven of the 32 participants and some indication of heart disease was present in nearly half of the evaluated population (15/32). 2

BiPAP autosv Advanced This study evaluated the performance of the BiPAP autosv Advanced device. Servo-ventilation devices provide a mode of pressure support to treat obstructive and complex central sleep apnea disorders. The main features of the BiPAP autosv include: Normalization of ventilation by automatically adjusting IPAP pressure to achieve a target peak flow. Response to central apneas and hypopneas, and periodic breathing by increasing the magnitude of ventilation. Timed, backup breaths during central apneas. The optimal backup rate is automatically determined by the device (based on the patient s sleep-disordered breathing presentation). An enhanced form of the BiPAP autosv device, the BiPAP autosv Advanced, was developed to provide the following enhancements: Improved timed backup breath delivery during central apnea. Automatic control of EPAP pressure to treat obstructive events. Methods Participants were enrolled after completing either an in-lab or home diagnostic PSG. Portable testing was permitted as part trial inclusion, however, all participants completed an in-lab PSG. Randomization was contingent upon the results of a CPAP titration night. Only participants that exhibited continued CompCSA and met all criteria for participation after their CPAP titration night were randomized into the study. All randomized subjects had prior PAP use for more than four weeks and continued to have persistent CSA. Five sites in the United States studied 32 participants with CompCSA. Subjects were recruited if they demonstrated an Apnea Hypopnea Index (AHI) 10 and a Central Apnea Index (CAI) 5 on their full-night diagnostic polysomnogram (PSG) and continued in the study if on their full-night CPAP titration they had a CAI 5. Qualifying subjects underwent two additional full-night in-lab PSGs, one with the BiPAP autosv, and the other with the BiPAP autosv Advanced device. The order of treatment was randomized and participants were blinded as to which device they received during their study nights. Table 1: Inclusion / exclusion criteria Inclusion criteria Pre-study inclusion criteria: Age 21-80 Ability to provide consent Documentation of medical stability by investigator Enrollment inclusion criteria: Participants who, during the ambulatory PSG study (Stardust) or in-lab diagnostic PSG, demonstrated an AHI 10 and CAI 5 or Participants who previously demonstrated CSA, with a CAI 5 on CPAP titration Exclusion criteria Participants who are acutely ill, medically complicated, or medically unstable. Pregnancy (confirmed absence of pregnancy with a urine or serum pregnancy test in women of child bearing potential). Participants in whom PAP therapy is otherwise medically contraindicated. Participants who are unwilling to wear CPAP. Participants who are currently prescribed oxygen therapy. Participants with previously diagnosed respiratory failure or respiratory insufficiency and who are known to have chronically elevated arterial carbon dioxide levels while awake (PaCO 2 45mmHg). Participants who have had surgery of the upper airway, nose, sinus, or middle ear within the previous 90 days. Participants with untreated, non-osa/csa sleep disorders, including but not limited to insomnia, periodic limb movement syndrome, or restless legs syndrome (PLM Arousal Index > 15). Participants who are unwilling to participate in the study. 3

Device settings for both groups in the investigation are listed in Table 2. The BiPAP autosv Advanced device automatically adjusted the level of Pressure Support (PS), EPAP, and respiratory rate. In the BiPAP autosv, EPAP was set to, and remained at, the best CPAP setting determined previously on the CPAP titration night. If indicated, the protocol allowed for additional EPAP titration within the first two hours of the test night with BiPAP autosv. Pressure Support (PS) and respiratory rate were automatically adjusted by the device. Table 2: Device settings Device setting BiPAP autosv BiPAP autosv Advanced EPAP Determined from CPAP titration Not available EPAPmax Not available 16 cm H 2 O EPAPmin Not available Determined from CPAP titration Set to EPAP - 2 IPAPmax 30 cm H 2 O Not available IPAPmin Set equal to EPAP Not available PSmax Not available 20 cm H 2 O PSmin Not available 0 cm H 2 O Backup rate Determined automatically Determined automatically Results demographic data Sixty-two patients were screened for participation, and 35 participants were randomized. Data are presented from the completed-cases population which consisted of 32 participants (5 females and 27 males) with an average age of 63.7 ± 11.2 (S.D.) years, and an average BMI of 31.2 ± 5.3. Three participants were excluded from the completed-cases analysis. One subject who completed both randomized nights failed to meet the required minimum Total Sleep Time (as defined in the protocol). The second participant also completed both randomized nights, but the BiPAP autosv device was set up in error with the min and max EPAP settings fixed on the BiPAP autosv Advanced night. Although data were available for this participant on both nights, the comparison to the predicate device was not appropriate due to this error and thus these data were excluded from the completedcases analysis. The third participant withdrew from the study prior to completing either randomized night. 4

Table 3 presents the summary statistics and significance levels for the completed-cases analysis comparing the Dx PSG, CPAP titration (CPAP), BiPAP autosv (autosv), and the BiPAP autosv Advanced (autosv Advanced). Figure 1 presents the AHI data graphically, with mean values indicated in bold. Compared to the diagnostic PSG and CPAP titration, treatment with the BiPAP autosv Advanced resulted in a statistically significant reduction in AHI and CAI (p <0.001). Table 3: Comparison across four PSG nights (Dx PSG vs. CPAP vs. autosv vs. autosv Advanced; N=32) Std. Overall Variable Mean deviation Median Min Max p-value* Dx PSG 45.29 20.15 43.28 17.29 89.81 AHI CPAP 29.30 16.90 23.42 9.47 71.39 0.0001 autosv 8.17 8.00 5.20 0.15 31.98 autosv Advanced 5.20 5.66 2.89 0.19 26.74 Dx PSG 9.94 11.20 8.22 0.00 54.39 CAI CPAP 15.14 15.76 9.77 4.98 66.08 0.0001 autosv 2.11 3.32 1.13 0.00 14.15 autosv Advanced 0.50 0.76 0.27 0.00 2.91 Dx PSG 10.71 16.50 5.78 0.00 73.31 OAI CPAP 0.73 1.16 0.41 0.00 5.52 0.0001 autosv 1.69 2.38 1.02 0.00 12.89 autosv Advanced 1.10 1.71 0.51 0.00 8.83 Dx PSG 4.57 9.94 0.51 0.00 48.81 MAI CPAP 0.51 1.23 0.00 0.00 6.21 0.0019 autosv 0.33 0.67 0.13 0.00 3.65 autosv Advanced 0.19 0.39 0.00 0.00 2.03 Dx PSG 20.08 13.59 18.61 1.35 55.24 HI CPAP 12.93 10.76 10.31 0.85 40.76 0.0001 autosv 4.00 4.41 2.19 0.00 15.35 autosv Advanced 3.41 4.58 1.72 0.00 20.87 *Overall p-values per the Friedman Test. Post-hoc comparisons showed a significant reduction in all respiratory indices with autosv Advanced compared to the diagnostic night, as well as a significant reduction in AHI, CAI, and HI compared to CPAP. 5

Figure 1: AHI comparative data across all nights 6 BiPAP autosv Advanced exhibited superior treatment to BiPAP autosv, with respect to the AHI, the CAI, the Obstructive Apnea Index (OAI), and the Hypopnea Index (HI) (p<0.05). Other key variables, including Arousal Index (AI), Mixed Apnea Index (MAI), Total Sleep Time (TST), Sleep Efficiency (SE%), Wake After Sleep Onset (WASO), and Baseline SpO 2 and Low SpO 2, did not differ significantly between devices. There were no significant differences between the two devices with respect to sleep stage measures. Discussion These data indicate that recent improvements made to the current auto servo-ventilation technology (automatically adjusted EPAP and enhanced auto backup rate) will likely lead to a more successful reduction or elimination of CompCSA in patients presenting with this condition. With both autosv devices there were clinically and statistically significant reductions in the AHI, CAI, OAI, MAI, and the HI from diagnostic and CPAP titration nights (p<0.002 for all). While both the current BiPAP autosv and the new BiPAP autosv Advanced provided clinically effective significant treatment, as revealed by an AHI < 10, the new autosv Advanced demonstrated statistical superiority in all of the key endpoints including AHI, CAI, OAI, and HI (p < 0.05 for all). Additional long-term studies are necessary to determine if the short-term benefits are maintained over time and whether the reduced apnea would result in less comorbidities or better quality of life. Although some patients were not completely treated with these devices (AHI < 15), an overwhelming majority of patients (84 percent) were. Recognizing the challenges in treating the comprehensive condition known as CompCSA, sleep clinicians and technicians can be confident that the new BiPAP autosv Advanced will aid caregivers in providing optimal therapy that is tailored to these specific patients needs. Conclusion This paper reports the findings of a recent randomized, controlled comparative clinical trial of the BiPAP autosv device versus the BiPAP autosv Advanced. These data indicate that both devices provide therapeutic benefit in patients diagnosed with Complex Central Sleep Apnea (CompCSA) including those with Cheyne-Stokes Breathing (CSB).

References: i Javaheri et al., Adaptive pressure support servoventilation: a novel treatment for sleep apnea associated with use of opioids. J Clin Sleep Med. 2008 Aug 15;4(4): 321-3. ii Webster et al., Sleep-disordered breathing and chronic opiod therapy. Pain Medicine 2008; 9(4): 425-432. iii Walker et al., Chronic opioid use is a risk factor for the development of central sleep apnea and ataxic breathing. J Clin Sleep Med. 2007 Aug 15;3(5): 455-61. iv Gilmartin, et al., Recognition and management of complex sleep-disordered breathing. Curr Opin Pulm Med 2005; 11: 485-93. v Morganthaler, et al., Complex sleep apnea syndrome: is it a unique clinical syndrome? Sleep 2006; 29(9): 1203-9. vi Lehman, et al., Central sleep apnea on commencement of continuous positive airway pressure in patients with a primary diagnosis of obstructive sleep apnea-hypopnea. J Clin Sleep Med 2007; 3(5): 462-6. vii Javaheri, et al., The prevalence and natural history of complex sleep apnea. J Clin Sleep Med 2009; 5(3): 205-11. viii Morganthaler, et al., Complex sleep apnea syndrome: is it a unique clinical syndrome? Sleep 2007; 30(4): 468-75. ix Eckert, et al., Central sleep apnea: pathophysiology and treatment. Chest 2007; 131: 595 607. 7

Respironics, autosv, BiPAP and Stardust are trademarks of Respironics and its affiliates. All rights reserved. 2014 Koninklijke Philips N.V. All rights are reserved. Philips Healthcare reserves the right to make changes in specifications and/or to discontinue any product at any time without notice or obligation and will not be liable for any consequences resulting from the use of this publication. CAUTION: US federal law restricts these devices to sale by or on the order of a physician. edoc SLU 5/5/14 MCI 4106199 Philips Healthcare is part of Royal Philips Philips Respironics Asia Pacific +65 6882 5282 Philips Respironics Australia +61 (2) 9947 0440 Philips Respironics China +86 400 828 6665 Philips Respironics United Kingdom +44 800 1300 845 Philips Respironics 1010 Murry Ridge Lane Murrysville, PA 15668 Customer Service +1 724 387 4000 800 345 6443 (toll free, US only) www.philips.com/respironics