Rotary Fatigue Testing to Determine the Fatigue Life of NiTi alloy Wires: An Experimental and Numerical Analisys

Similar documents
A numerical method for predicting the bending fatigue life of NiTi and stainless steel root canal instruments

Torsion Property and Cyclic Fatigue Fracture Behavior of. Nickel-Titanium Endodontic Instruments

Is the reciprocating movement per se able to improve the cyclic fatigue resistance of instruments?

thinking beyond cyclic fatigue Testing standards for endo files vary and there are other qualities to be concerned with, too

Microstructural and mechanical characterizations of new Ni-Ti endodontic instruments

Improve Your Root Canal Preparations with Adaptive Motion Technology

Comparative analysis of torsional and bending behavior through finiteelement models of five Nickel-Titanium endodontic instruments

Comparison of cyclic fatigue resistance of original and counterfeit rotary instruments

Abstract. Keywords: Fatigue Failure, Surface Roughness, Fatigue Strength Factor, Bio-Compatible Dental Implant, Electron Beam Melting (EBM)

Cyclic fatigue life of Tango-Endo, WaveOne GOLD, and Reciproc NiTi instruments

Glassman Webinar PPT 7/18/2014. Nickel Titanium Clinical Performance (2012) Depends on: three main factors :

Research Article Structural Characterisation and Mechanical FE Analysis of Conventional and M-Wire Ni-Ti Alloys Used in Endodontic Rotary Instruments

Instructions for use: FlexMaster. the rotary NiTi-System for every case. Endo Easy Efficient

An effort is made to analyse the stresses experienced by the human femur. In order

International Journal of Engineering Research and Development e A Mathematical Model to Calculate Contact Stresses In Artificial Human Hip Joint

Race family Endodontic NiTi rotary files

CIC Edizioni Internazionali. In vitro resistance to fracture of two nickel-titanium rotary instruments made with different thermal treatments

Comparison of the flexibility and torsional resistance of nickel-titanium rotary instruments.

Failure criteria for Adhesives. Sainath Kadam, 3mE 19 oktober 2014

ENDO-EZE TM GENIUS ENDODONTIC SYSTEM WHERE SAFETY MEETS EFFICIENCY

Buy Direct. Buy Brasseler USA. Systems for Endodontic Excellence. R a Ce T M. Rotary Endodontic System with Safety Memo Disc. RaCe TM.08. RaCe TM.

Professor Stephen D. Downing

Hend M. Abou El Nasr * and Shehab El Din M. Saber ** ABSTRACT صور 4 لون

ACCURATE NUMERICAL ANALYSIS OF GEAR STRENGTH BASED ON FINITE ELEMENT METHOD

Comparison of cyclic fatigue resistance of XP-endo Shaper, HyFlex CM, FlexMaster and Race instruments

Effect of autoclave sterilization on the cyclic fatigue resistance of thermally treated Nickel Titanium instruments

Finite Element Analysis of Radius and Ulna. Eli Pavlatos April 24, 2013

First Technology Safety Systems. Design Freeze Status. Flex-PLI-GTR Development

Effective cleaning and shaping of the root

The advent of nickel-titanium (NiTi) rotary instruments has revolutionized root canal

K3 Nickel-Titanium Files

Finite-element analysis of NiTi wire deflection during orthodontic levelling treatment

The Development and Analysis of an FEA Simulation on an Artificial Knee

ENDODONTOLOGY ABSTRACT INTRODUCTION

Design and Analysis of Dental Surgical Fixture

Biomechanics of Fractures and Fixation

Evaluation of physicochemical and mechanical properties of NiTi endodontic instruments made with conventional and controlled-memory alloys

History of NiTinol. Ni = Nickel Ti = Titanium N = Naval O = Ordnance L = Laboratory

Effect of Cryogenic Treatment on Fracture Resistance of Nickel Titanium Rotary Instruments a Systematic Review

MULTIPHYSICAL ASPECTS OF SHAPE MEMORY ALLOY ACTUATOR. Peter Drahoš, Vladimír Kutiš

Keywords: Angioplasty, Explicit finite elements method, Tube hidroforming, Stents.

ENDODONTICS ENDODONTICS

Experimental and Numerical Comparison of Stresses Level between Titanium and Novel Composite Single Tooth Implant

Finite Element Analysis of High Contact Ratio Gear

Topic 7b: Biological Membranes

APPLICATION AND DEPLOYMENT OF ADVANCED NDE TECHNIQUES IN HIGH PRESSURE VESSELS

A WHOLE NEW DIMENSION OF TOTAL PERFORMANCE.

Geometric and dimensional characteristics of simulated curved canals prepared with protaper instruments

Comparative analysis of load/deflection ratios of conventional and heat-activated rectangular NiTi wires

Effect From Surface Treatment of Nickel-Titanium Rotary Files on the Fracture Resistance

Single File Reciprocating Technique Using Conventional Nickel Titanium Rotary Endodontic Files

H. Caballero-Flores, C. K. Nabeshima, E. Binotto & M. E. L. Machado. Introduction

wiirre e a a n n d d w wiirre e ffo orrms ms

Lifespan of One Shape Files used in Severely Curved Canals in Resin Blocks and Extracted Teeth

Theoretical and Finite Element Modeling of Fine Kirschner Wires in Ilizarov External Fixator

Modal Analysis of Hip Joint Implant Used In the Human Body

The Effect of Surface Defects on the Cyclic Fatigue Fracture of HEROShaper Ni-Ti rotary files in a Dynamic Model: A Fractographic Analysis

ProUltra ULTRASONIC SURGICAL TIPS

XP-endo Shaper. The One to Shape your Success

EXTERNAL FIXATION SYSTEM

Femtosecond Laser Applications in NiTi

1) INDICATIONS FOR USE These products have to be used only in hospital environments, clinics or dental offices by qualified dental personnel.

WHERE SAFETY MEETS EFFICIENCY

Innovative revolutionary approach to root canal preparation

Gear Tooth Root Fillet Optimization Program User manual

Failure of ProTaper rotary Ni-Ti instruments used by undergraduate students

Tools for Endodontic Success

ARCHIVES OF CIVIL ENGINEERING, LX, 2, 2014

(c) B. Ravi, IIT Bombay 1

SE NiTi Heat Activated NiTi Ultra T Heat Activated NiTi Stainless Steel Beta Titanium. Ultra T Archwires

Software and Hardware Implementation of a Microtorquemeter

Test the Mechanical Properties of Articular Cartilage using Digital Image Correlation Technology

FINITE ELEMENT ANALYSIS OF KNEE IMPLANT AND PROTOTYPE FABRICATION USING RPT PROCESS

Push-out of root canal filling using the material testing stage (MTS) inside a Micro-CT: preliminary observations

Numerical predicting of contact and pressure sore of lower extremity parts caused by prosthetic and orthotic

FUNCTIONAL WARP-KNITTED FABRICS WITH INTEGRATED SUPERELASTIC NITI FILAMENTS

flexible performance Chemin du Verger 3 CH-1338 Ballaigues Switzerland

Topology optimisation of hip prosthesis to reduce stress shielding

Illustrative exercises for the lectures

Mechanical Behavior of Silver Alloys Used as Cast Post and Core in Maxillary Incisors

Fracture Resistance of Wave-one Rotary File Used In Reciprocating Motion

People have Priority. Endodontics

INTERNATIONAL JOURNAL OF PURE AND APPLIED RESEARCH IN ENGINEERING AND TECHNOLOGY

Vibration Analysis of Finger Using Non linear FEM to Understand HAV Syndrome

Evaluation of selected mechanical properties of NiTi rotary glide path files manufactured from controlled memory wires

Comparative Study of Fixation Devices for Intertrochanteric Fractures

Mechanical Aspects of an Interference Screw Placement in ACL Reconstruction

Microtubule Teardrop Patterns

Over the past 2 decades, nickel-titanium (NiTi) instruments have become an important

Computer Aided Surgery (CAS) for Abdominal Aortic Aneurysm (AAA)

A comparison of cyclic fatigue between used and new Mtwo Ni Ti rotary instruments

Endo Easy Efficient. Single File Root Canal Preparation.

ProUltra Ultrasonic Non Surgical Endo Tips

ENDODONTICS ENDODONTICS

EVALUATION ON THE SHEAR STRENGTHENING EFFECT OF RC COLUMNS WITH CARBON FIBER SHEETS

Effect of implant stiffness on spinal growth in the pig spine

Fatigue life prediction methodology of automotive rubber component. *Chang-Su Woo 1)

Development of a Wear Testing Machine for Dental Crown Application

Torque During Canal Instrumentation Using Rotary Nickel-Titanium Files

ARCHWIRE. An extensive selection of high quality wire types and arch forms WIRE PRODUCTS. Wires. TruFlex Nickel Titanium Archwire. TruFlex.

Transcription:

Available online at www.sciencedirect.com Procedia Engineering 00 (2016) 000 000 www.elsevier.com/locate/procedia 1st International Conference on Structural Integrity Rotary Fatigue Testing to Determine the Fatigue Life of NiTi alloy Wires: An Experimental and Numerical Analisys André Carvalho a, Manuel Freitas a,, Luis Reis a, Diogo Montalvão b, Manuel Fonte a,c a IDMEC - Instituto Superior Técnico, Universidade de Lisboa, Av. Rovisco Pais, 1, 1049-001 Lisboa, Portugal b School of Engineering and Technology, University of Hertfordshire, College Lane Campus, Hatfield, Herts AL10 9AB, UK c Escola Náutica Infante D. Henrique, Av. Eng. Bonneville Franco, 2770-058, Paço d Arcos, Portugal Abstract Endodontic rotary file instruments used to treat root canals in dentistry suffered breakthrough transformations in recent years when stainless steel was replaced by Nickel-Titanium (NiTi). NiTi alloys used in Endodontics possess superelastic properties at body temperature (37C) that bring many advantages on the overall performance of the root-canal treatment. They can follow curved root canals more easily than stainless steel instruments and have been reported to be more effective in the removal of the inflamed pulp tissue and protection of the tooth structure. However, these instruments eventually fracture under cyclic bending loading due to fatigue, without any visible signals of degradation to the practitioner. This problem brought new challenges on how new instruments should be tested, as NiTi alloys are highly non-linear and present a large hysteresis cycle in the Elastic domain. Current existing standards are only available for Stainless Steel testing. Thus, many authors have attempted to design systems that can test NiTi endodontic files under fatigue loads. However, no approach has been universally adopted by the community yet, as in most cases they are based on empirical set ups. Following a more systematic approach, this work presents the results of rotary fatigue tests for several NiTi wires from different manufacturers (Memry and Euroflex ).The formulation is presented, where the material strength reduction can be quantified from the determination of the strain and the number of cycles until failure, as well numerical FEM simulation to verify the analytical model predictions. 2016 The Authors. Published by Elsevier B.V. Peer-review under responsibility of the Scientific Committee of PCF 2016. Keywords: Fatigue; Files; NiTi wires, Life evaluation; 1. Introduction In dentistry, the root canal procedure is done using a rotary file that removes the existing nerve endings on a tooth. In the past, endodontic files used in this procedure were made from highly flexible steel alloys. However, steel alloy files, while being flexible, are still too rigid to avoid damaging the walls of the root canals. In order to minimize these adverse effects, Nickel-Titanium alloys are now being used in the design of endodontic rotary files instead of stainless steel alloys. NiTi alloys are superelastic metal alloys that are able to fully recover from large deformations (up to Corresponding author. E-mail address: mfreitas@dem.ist.utl.pt 2452-3216 2016 The Authors. Published by Elsevier B.V. Peer-review under responsibility of the Scientific Committee of PCF 2016.

2 Andre Carvalho et al. / Procedia Engineering 00 (2016) 000 000 strains of 10% Montalva o et al. (2014)). These alloys, however, have a drawback when compared to steel files: their fatigue life is relatively shorter than steel and, as seen in commercial endondontic files, they break without a previous mechanical warning, increasing the risk of the file failing inside the teeth. There are some studies to determine the fatigue life of NiTi alloys, through traditional uniaxial fatigue tests and rotary bending fatigue tests Plotino et al. (2009). Rotary bending tests are the tests that most accurately replicate the kind of loads and deformation a file is subjected to when inside a root canal. The great majority of the existing machines in the literature only perform the fatigue test with a predetermined set of shapes. However, most of the imposed deformations are far from the complex shapes of the root canals. Plotino et al. (2009). Of special interest is the rotary fatigue machine designed by Cheung and Darvell Cheung and Darvell (2007). This machine consists of three pins that can be positioned manually to deform the endondontic file. The file is then put into rotation using a contra-angle. This type of machine has an advantage of being more versatile than the more common machines with simulated canal carved in a stainless steel plate, where one can have only one predetermined curvature per plate De-Deus et al. (2010); Gambarini et al. (2012); Lopes et al. (2010); Plotino et al. (2010) In this work, an automated configurable rotary bending-testing machine was designed. This testing machine was designed to adapt and change the degree of bending from simple point bending to more complex multi-point bending. The machine consists in three pairs of pins positioned by servomotors, which deform the specimen into a desired complex shape. The specimen is then put into rotation until failure is detected. The machine design also enables rotary bending tests with constant curvature (constant strain) along a segment. With a constant strain, one can compare directly the result with the more common uniaxial fatigue tests. Also, one can perform tests in different regions of the superelastic stress-strain curve, enabling an estimation of the stress and the metallic phase of the alloy during the test. 2. Designing the Testing Machine The testing machine was designed to be versatile and to offer a wide range of possible bending configurations. Based in the machine designed by Cheung and Darvell Cheung and Darvell (2007), the testing machine has three pins that can be configured to make 1 to 4 point bending tests. In our machine, however, the pins are positioned at configurable locations using servo-motors. Configurable positioning enables a greater precision and repeatability, while providing a simple interface (Fig. 1). Fig. 1: Testing area with an endondontic file. For the rotation of the specimens, a brushless DC motor with variable speed is used instead of the more common contra-angle. With a standalone DC motor, one can control the velocity of the test more precisely and automatically. The motor is able to do tests from 100rpm (16.67Hz) to 3000rpm (50Hz), which covers the range of most NiTi endodontic files with a drive speed usually between 150 and 350rpm. To detect when the specimen fail (whether it is wire or an endondontic file), an electronic failure detection system was implemented. This system uses the natural conductivity of NiTi alloys to detect any failure by constantly monitoring the level of voltage between the specimen and each pin. When the circuit is open it means that the specimen failed and the test stops automatically.

Andre Carvalho et al. / Procedia Engineering 00 (2016) 000 000 3 3. Positioning control and Constant Curvature The most important aspect of this machine is its ability to generate bending configurations automatically. The configurations are set by positioning the pins (Fig. 1) in order to deform the specimen into a shape (Fig.2). The Fig. 2: Wire deformed during a test. machine software enables two positioning modes: individual pin positioning and constant curvature radius generation. 3.1. Constant Curvature Generation for Wire Specimens The main advantage of having a constant curvature in a uniform beam is that in that region the strain is also constant and proportional to the radius as in (1). x = z 1 ρ (1) where ρ is the radius of curvature, x is the strain along the length of the beam and z is the distance to the neutral surface. Using the Euler-Bernoulli beam theory to obtain the analytical model of the wire, we obtain the deflection for the wire. For an infinitesimal element (1) becomes: x (x) = z d2 w dx2 (2) However, we want a constant strain level for an entire segment of the specimen and, consequently, the position of the pins must be determined in to generate a constant strain section. Using a standard square minimization algorithm to find the pin displacements needed for the desired constant strain level in the specimens mid section, while minimizing the overall deformation, we obtain the three pin positions Carvalho et al. (2015). We also have to take into account the influence of thickness of the wire and the requirement that the wire must be tangent to the pins at all times. This leads to a variable distance between the contact point at the surface and the neutral surface Carvalho et al. (2015). Adding this requirement to the otimization algorithm, we obtain deformation profiles similar to the one in Fig. 3. For the case of Fig. 2, which corresponds to setting the desired strain to 2%, the curvature can be seen in Fig. 4. Knowing that the internal radius of the pins is 1mm and using (1), we obtain that the angle of the arc is 11.71 deg and the radius of curvature is 50mm. These two parameters can be used to compare results with testing methods such as the Pruet et al Pruett et al. (1997).

4 Andre Carvalho et al. / Procedia Engineering 00 (2016) 000 000 Fig. 3: Beam under a desired constant strain of 2%, with color-coded strain and tangent pins. Fig. 4: Deformed wired with a desired strain of 2% superimposed with the curvature radius. 4. Results for a NiTi Wire The first materials to be tested in the rotary fatigue machine were a Memry and a Euroflex wires, both with a diameter of 0.8mm. The wires, when under an uniaxial load, has the stress-strain relation in Fig. 5. 2500 Stress (MPa) 2000 1500 1000 500 0 0 Memry 0.8mm EuroFlex 0.8mm 0.02 0.04 0.06 0.08 0.1 0.12 0.14 Strain Fig. 5: Uniaxial tension test of the Memry 0.8mm diameter wire. Figure 5 has three distinct regions: a first austenitic elastic region, a horizontal elastic region with a mixture of austenitic and martensitic phases (also known as R-phase) and a final plastic martensitic region. Another characteristic of these Ni-Ti alloys is that they exhibit a large hysteresis when unloading occurs. For simplicity sake, this phenomenon was not considered in the study of the fatigue of the wire specimens. The fatigue tests were concentrated on the first two regions, imposing strains level from 0.8% to 6%. The results can be seen in Fig. 6. The specimens with strains smaller or equal to 1% (corresponding to the first zone in the stressstrain plot in Fig. 5) show a large fatigue life when compared with the rest of the points, with two specimens never failing. The remaining points show a decrease of the fatigue life as the imposed strain increases. The specimen with the largest strain showed a life of 34 cycles for the Memry wire and 16 cycle for the Euroflex wires. 5. NUMERICAL SIMULATIONS Numerical simulations using FEA (Finite Element Analysis) can contribute to a better understanding of the failure mechanisms of the Ni-Ti wires in the proposed testing machine. Also, they allow predicting the stress distribution on the wire more accurately than the Euler-Bernoulli beam theory approximation, since SMA (Shape Memory Alloy) or superelastic material properties are highly non-linear.

André Carvalho et al. / Procedia Engineering 00 (2016) 000 000 5 6% Memry 5% Surface Strain 4% 3% 2% 1% 0% 1E+00 1E+01 1E+02 1E+03 1E+04 1E+05 1E+06 Number of Cycles Fig. 6: Strain vs. number of cycles for a Memry 0.8mm diameter wire. 5.1. Model A static structural FEA was conducted in ANSYS 16.1 Workbench with the model shown in Fig. 7 to replicate the experiments conducted in the testing machine. The 0.8mm diameter wire was modelled with 1632 solid elements with superelastic material properties Carvalho et al. (2015); Yin et al. (2015) introduced in the projects Engineering Data cell as shown in table 1. The wire was clamped at one end and it was deformed under bending using three steel actuator pins that moved in the y direction based on the constant curvature formulation earlier presented. The movement of the actuators was simulated as being gradual, in an explicit iterative way. Contact properties were added between the actuator pins and the wire, with a considered frictional coefficient of 0.2. Fig. 7: Mesh, boundary conditions and loads (displacements) for the FEA model. 5.2. Results The assessment of the results will be based on the normal stress in the x direction, since the beam is in pure bending on the xy plane with the loads being applied in the y direction. The maximum stress in the x direction for both materials during loading and unloading is plotted in Fig. 8. Fig. 8 shows that the behaviour of the alloys is non-linear and follows what is expected for a superelastic alloy. However, the maximum values for the normal stress is being determined to be near the clamping region according to the FEA results. This should not be a surprise, because the clamping is modelled as being infinitely rigid. This may also lead to a stress concentration location, reason why the values can be quite significant. Since the true clamping region

6 André Carvalho et al. / Procedia Engineering 00 (2016) 000 000 Table 1: Material properties for the Memry and Euroflex alloys introduced in ANSYS Workbench projects Engineering Data project cell. Material Property Memry Euroflex Isotropic Elasticity Young Modulus (GPa) 70 70 Poisson s Ratio 0.33 0.33 Superelasticity Sigma SAS (MPa) 368 736 Sigma FAS (MPa) 484 798 Sigma SSA (MPa) 232 337 Sigma FSA (MPa) 100 135 Epsilon (mm/mm) 0.063 0.059 Alpha 0 0 Fig. 8: Maximum normal stress in the x direction S x. is very difficult to be modelled accurately and the wires failed near the 1 st and the 2 nd pins during the experiments (in an apparently random fashion), the following analysis neglects the clamping location. Fig. 9 shows a few examples of the normal stress distribution in the x direction on the Memry (Figs. 9a to 9c) and Euroflex (Fig. 9d to 9f) wires respectively for different loading conditions. It is observed that the normal stress (hence the normal strain) in the x direction is approximately constant in the constant curvature region between actuator pins 1 and 2, validating the theoretical formulation assumptions. Notwithstanding, what is noteworthy to mention is that the maximum compression stress (identified as Min in Fig. 9) was consistently determined to be near the first actuator pin and larger than the maximum traction stress (identified as Max in Fig. 9). This might be due to a frictional contact property that was added between the pins and the wire as an attempt to mimic the existing testing machine. The maximum traction stress seems to be randomly floating within the constant strain curvature region, i.e., it can either be near the first pin, near the second pin, or in between the two pins. Being a constant strain region, the floating of the maximum traction stress location can only be attributed to numerical error. Therefore, it is expected that the wires will fracture near either pin 1 or pin 2 due to friction, being more likely to fail near pin 1. 6. Fracture Surfaces The fracture surfaces of the specimens tested in the rotary testing bending machine were observed on a Scanning Electronic Microscopy, JEOL existent at MicroLab from IST. For comparison purposes, the fracture surface of the specimens tested in uniaxial tension was also observed. Fig. 10 show respectively, the general view of the fracture

André Carvalho et al. / Procedia Engineering 00 (2016) 000 000 7 (a) Memry, theoretical ex = 1.0%. (b) Memry, theoretical ex = 2.4%. (c) Memry, theoretical ex = 4.0%. (d) Euroflex, theoretical ex = 1.0%. (e) Euroflex, theoretical ex = 2.0%. (f) Euroflex, theoretical ex = 4.0%. Fig. 9: Maximum normal stress in the x direction S x surface of the uniaxial tension specimen Memry 0.8 mm and the amplification of the fracture surface whre a characteristic ductile facies can be observed with dimples developed during the failure process. (a) General view (b) Amplification of the fracture surface Fig. 10: Uniaxial tension The fracture surfaces observed in fatigued specimens are substantially different from the uniaxial testing ones and are characterized by flat fractures surfaces, as usually observed in fractures specimens under rotating bending conditions. For lower cycles to failure a very small fatigue zone is observed near the surface. As the number of cycles to failure increases, the fatigue zone on the specimens also increases. Fig. 11a shows the fractured surface of the specimen of Memry 0.8 mm failed by fatigue at 44551 cycles. A clear focus of crack initiation can be observed and is shown in detail on Fig. 11b which is significantly differet from Fig. 10b. Therefore these NiTi alloy fractured at rotating bending fatigue do not differ from fractured surfaces by fatigue in more classic isotropic metals.

8 André Carvalho et al. / Procedia Engineering 00 (2016) 000 000 (a) Failure at 44551 cycles (b) Crack initiation Fig. 11: Rotary bending fatigue 7. CONCLUSIONS The work presented on this paper presents a testing apparatus designed to perform rotary fatigue tests on NiTi endondontic files and wires. The testing machine was designed to be versatile, enabling a series of testing configurations and test types, using computer controlled automatic positioning of system that bends the specimens into user supplied shapes or into a bending configuration that has a constant strain section. A series of fatigue tests were done using a Memry and Euroflex 0.8mm diameter wires. The wires showed a very long fatigue life (most of the specimens did not fail) when under strain levels in the elastic austenitic phase. When imposing strain in the R-phase region, the fatigue life of the wire drastically reduced, with fatigue life ranging from 20,000 (lower strain levels) to 16 cycles (higher strain levels). For future work, other wires with different diameters will be tested, as well different endondontic files. Acknowledgements This work was supported by national funds through FCT - Fundação para a Ciência e a Tecnologia as part of the project PTDC/EME/AME/122795/2010. References Carvalho, A., Freitas, M., Reis, L., Montalvão, D., Fonte, M., 2015. Rotary fatigue testing machine to determine the fatigue life of niti alloy wires and endondontic files. Procedia Engineering 114, 500 505, {ICSI} 2015 The 1st International Conference on Structural Integrity Funchal, Madeira, Portugal 1st to 4th September, 2015. URL http://www.sciencedirect.com/science/article/pii/s1877705815017373 Cheung, G. S. P., Darvell, B. W., 2007. Fatigue testing of a niti rotary instrument. part 1: strain-life relationship. International endondontic Journal 40, 612 618. De-Deus, G., Moreira, E. J. L., Lopes, H. P., elias, C. N., 2010. Extended cyclic fatigue life of F2 ProTaper instruments used in reciprocating movement. International endondontic Journal 43, 1063 1068. Gambarini, G., Gergi, R., Naaman, A., Osta, N., Al Sudani, D., 2012. Cyclic fatigue analisys of twisted file rotary niti instruments used in reciprocating motion. International endondontic Journal 45, 802 806. Lopes, H. P., Britto, I. M. O., elias, C. N., de Oliveira, J. C. M., Neves, M. A. S., Moreira, E. J. L., Siqueira, J. F., 2010. Cyclic fatigue resistance of protaper universal instruments when subjected to static and dynamic tests. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 110, 401 404. Montalvão, D., Alçada, F., Fernandes, F., Correia, S., 2014. Structural characterisation and mechanical fe analysis of conventional and m-wire ni ti alloys used in endondontic rotary instruments. The Scientific World Journal: Materials Science 2014. Plotino, G., Grange, N. M., Cordaro, M., Testarelli, L., Gambarini, G., 2009. A review of cyclic fatigue testing of nickel-titanium rotary instruments. Cyclic Fatigue or Rotary Instruments 35 (1), 1469 1476. Plotino, G., Grange, N. M., Melo, M. C., Bahia, M. G., Testarelli, L., Gambarini, G., 2010. Cyclic fatigue of niti rotary instruments in a simulated apical abrupt curvature. International endondontic Journal 43, 226 230. Pruett, J., Clement, D., Carnes, D., 1997. Cyclic fatigue testing of nickel titanium endodontic instruments. Journal of Endodontics 23, 77 85. Yin, W., Sczerzenie, F., Long, M., Belden, C., Manjeri, R., Lafond, R., June 2015. Structure and properties of large diameter hot rolled niti bars for seismis applications. SAES Smart Materials, New Harthford, New York 2015.