Comparison of Central Corneal Thickness Measurement with Pentacam and Ultrasound

Similar documents
Comparison of Central Corneal Thickness Measurements with Pentacam, Orbscan II, and Ultrasound Pachymeter

Accuracy and Repeatability of a New Portable Ultrasound Pachymeter

MEDICAL POLICY SUBJECT: CORNEAL ULTRASOUND PACHYMETRY. POLICY NUMBER: CATEGORY: Technology Assessment

Medical Policy An independent licensee of the Blue Cross Blue Shield Association

Central corneal thickness (CCT) measurement is a critical

RECENT ADVANCES IN ANTERIOR SEGMENT IMAGING OF THE EYE. by Eszter Szalai, M.D. Supervisor: László Módis, M.D., Ph.D.

CLINICAL SCIENCES. Effect of Central Corneal Thickness, Corneal Curvature, and Axial Length on Applanation Tonometry

Cataract Surgery in the Patient with a History of LASIK or PRK

INTRODUCTION MIKLOS SCHNEIDER 1, GABOR BORGULYA 2, ANDRAS SERES 1, ZOLTAN Z. NAGY 1, JANOS NEMETH 1

Comparison between Pentacam HR and Orbscan II after Hyperopic Photorefractive Keratectomy

December 2016, Volume: 8, Issue: 12, Pages: , DOI:

FEP Medical Policy Manual

Jose M. Martinez-de-la-Casa, Julian Garcia-Feijoo, Ana Fernandez-Vidal, Carmen Mendez-Hernandez, and Julian Garcia-Sanchez METHODS

CENTRAL CORNEAL THICKNESS; COMPARISON OF CENTRAL CORNEAL THICKNESS MEASUREMENTS BY PENTACAM AND ULTRASOUND PACHYMETRY, IN MYOPIC AND KERATOCONUS EYES.

Comparison of Corneal Thickness with Online Optical Coherence Pachymetry and Ultrasound Pachymeter

Comparison of Corneal and Anterior Chamber Parameters following Myopic laser in situ keratomileusis and photorefractive keratectomy by

Keratoconus Clinic. Optometric Co-management Opportunities

Central corneal thickness among glaucoma patients attending Menelik II Hospital, Addis Ababa, Ethiopia

Central Corneal Thickness and its Association with Ocular Parameters

NIH Public Access Author Manuscript Cornea. Author manuscript; available in PMC 2014 May 29.

INTRODUCTION S. HERDENER, D. HAFIZOVIC, M. PACHE, S. LAUTEBACH, J. FUNK. University Eye Hospital, Freiburg - Germany

Evaluation of the Effect of Diabetes Mellitus on Corneal Biomechanics

Role of Central Corneal Thickness in Circadian Intraocular Pressure Fluctuations among Patients with Primary Open Angle Glaucoma

BY MICHAEL W. BELIN, MD

IOL Master. Photorefractive keratectomy

Relationship between central corneal thickness and parameters of optic nerve head topography in healthy subjects

Evaluation of corneal thickness and topography in normal eyes using the Orbscan corneal topography system

Nature and Science 2016;14(9)

Central Corneal Thickness-An important variable for prognostication in Primary Open Angle glaucoma; A Kolkata based study in Eastern India

L. Spadea, R. Ferrante, F. Romani, A. Di Gregorio

Correlating central corneal thickness and intraocular pressure in ocular hypertension and glaucoma

Laser in situ keratomileusis (LASIK) has

Corneal densitometry using Pentacam based scheimpflug imaging system: Indian rural population

Partial Coherence Interferometry as a Technique to Measure the Axial Length of the Eye Archived Medical Policy

CLINICAL SCIENCES. Clinical Significance of Central Corneal Thickness in the Management of Glaucoma

Predictability and accuracy of IOL formulas in high myopia

Effect of corneal parameters on measurements using the pulsatile ocular blood flow tonograph and Goldmann applanation tonometer

FUTURE DRAFT Local Coverage Determination (LCD) for Corneal Pachymetry (DL32410)

Interpretation of corneal tomography

Refractive Surgery Dilemma

Dr. Harvey Richman, OD, FAAO, FCOVD Diplomate American Board of Optometry Executive Committee AOA Third Party Center Founder Ask the AOA Coding

The Association of Central corneal thickness with Intra-ocular Pressure and Refractive Error in a Nigerian Population

Correspondence should be addressed to Fusako Fujimura;

Central corneal thickness and vascular risk factors in normal tension glaucoma

Central and Peripheral Changes in Anterior Corneal Topography after Orthokeratology and Laser in situ Keratomileusis

Anterior segment imaging

Morphological Study of Corneal Endothelium and Corneal Thickness in Pseudoexfoliation Syndrome

Number 80. Laser Eye Surgery in Myopia. Date of decision October 2017 Date of review October 2020 GUIDANCE

CLINICAL STUDY. AC-M Wong, C-C Wong, NS-Y Yuen and S-P Hui

Simultaneous Topography-guided Surface Ablation with Collagen Cross-linking for Keratoconus

Nature and Science 2017;15(11) Mohamed Elmoddather. MD

Corneal Topography / Computer- Assisted Corneal Topography/Photokeratoscopy

PHOTOREFRACTIVE KERATECTOMY (PRK) PATIENT INFORMATION BOOKLET

Comparison of Intraocular Lens Power Calculation Methods Following Myopic Laser Refractive Surgery: New Options Using a Rotating Scheimpflug Camera

LASIK has been the primary type of corneal refractive surgery

Abstracts. Edited by Dr. Tahir Mahmood

Management of Myopia in Adolescent. Non Invasive Kerato- Refractive Procedures. Early Orthokeratology. Sami El Hage, PhD, DSc, OD FAAO,FAAS,FAAOMC

Circadian variations in central corneal thickness and intraocular pressure in patients with glaucoma

3/23/2016. Diagnostic Services Taylor Pannell CRA, OCT-C. Services Available. Important info for the Tech to know. Visual Fields

Patient Selection IOL Power Calculation. Patient Selection. Biometry IOL-Power calculation using Vericalc 2.0. AC-Depth > 3.0 mm (FDA 3.

Central corneal thickness determined with optical coherence tomography in various types of glaucoma

Anterior chamber depth measurement: interchangeability between Pentacam and other devices

Laser in situ keratomileusis (LASIK) has proven to be

Investigation of the relationship between central corneal thickness and retinal nerve fiber layer thickness in ocular hypertension

Corporate Medical Policy

POST-LASIK ECTASIA MANAGEMENT

Assessing the Relationship between Central Corneal Thickness and Retinal Nerve Fiber Layer Thickness in Healthy Subjects

The pinnacle of refractive performance.

Pearls for the Refractive Technician Fadiah Alkhawaldeh, IMBA, COT, ROUB

Effect of Refractive Surgery on Corneal Properties and Visual Acuity

High-speed Optical Coherence Tomography of Corneal Opacities

The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation

Management of Unpredictable Post-PRK Corneal Ectasia with Intacs Implantation

Glaucoma: Diagnostic Modalities

Preserving the Cornea for the Future

Estimation accuracy of surgically induced astigmatism on the cornea when neglecting the posterior corneal surface measurement

Analysis of eye movements during myopic laser in situ keratomileusis

CENTRAL CORNEAL THICKNESS STUDY CCTS

Immersion Vs Contact Biometery for Axial Length Measurement before Phacoemulsification

Author s Affiliation. Original Article. Visual outcomes after LASIK (laser-assisted in-situ keratomileusis) for various refractive errors.

A Study to Reach the Proper Measurement of the White to White Diameter Using both the Caliper and IOL Master for ICL Suitable Size

Ocular Rigidity Evaluation After Photorefractive Keratectomy: An Experimental Study

Effect of refractive surgery on corneal properties and visual acuity

CORNEAL CONDITIONS CORNEAL TRANSPLANTATION

THE PENTACAM AXL. Improving Cataract Surgery Outcomes. Optical biometry and anterior segment tomography in one device

SMILE A Solution to Those who go in Harm s Way

In recent years, more and more studies have focused on. Characteristics of Straylight in Normal Young Myopic Eyes and Changes before and after LASIK

EUROTIMES ESCRS SUPPLEMENT FEBRUARY 2007

Effects of Intracorneal Ring Segment on Corneal Biomechanics in Keratoconic Eyes. Abstract

Author s Affiliation. Original Article. Comparison of Biometry in Phakic and Dense Modes. Muhammad Suhail Sarwar. Unaiza Mariam

Artiflex Toric Phakic Intraocular Lens Implantation in Congenital Nystagmus

Corneal haze is a condition resulting from several

Evaluation of The Accuracy of Sub Bowman s Keratomileusis (SBK) Microkeratome in Flap Creation during Lasik Surgery

Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography

Results INTRA PROTOCOL ANALYSIS

Introduction. Fuchs Endothelial Corneal Dystrophy (FECD) represents a non-inflammatory dystrophy of the corneal endothelial layer which

Three-year Follow-up of the Artisan Phakic Intraocular Lens for Hypermetropia

Comparison of Corneal Shape Changes and Aberrations Induced By FS-LASIK and SMILE for Myopia

Photorefractive Keratectomy as A Retreatment of Residual Myopia after Previous Laser in Situ Keratomileusis

CATARACT SURGERY AFTER RADIAL KERATOTOMY

Transcription:

Australian Journal of Basic and Applied Sciences, 4(12): 5744-5752, 2010 ISSN 1991-8178 Comparison of Central Corneal Thickness Measurement with Pentacam and Ultrasound Tamer Adel Refai, MD, FRCS. Department of Ophthalmology, Refractive surgry unit,research Institute of ophthalmology, Giza (Egypt)2010. Abstract: Purpose:Thre purpose of this study was to compare central corneal thickness (CCT) measurements obtained with a novel rotating Scheimpflug camera, ALLEGRO Oculyzer(Pentacam)Version 1074; Allergo,Germany with ultrasound pachymetry (Sonoscan model 4000AP,SONOMED,USA).Methods:CCT in 117 healthy eyes of (64 egyptian patients) seeking for refractive was measured with each modality by one observer.at least three measurements are taken for each eye and the average number for CCT is included in the study. For the rotating Scheimpflug camera,only images with Quality specification (QS):OK were includedin the study. Results:The mean central corneal thickness by ultrasound pachymetry was 552.1 μ ± 43.96μ standard deviation and by Pentacam 541.82 μ ± 39.87 μstandard deviation with correlation of 0.892 (P value < 0.01 ) implying a hightly correlation but with slightly lower values (by about 8.60μ ± 14.1 ) obtained by Pentacam with no difference attributed to age or sex of the patients. Conclusion:In the assessment of normal corneas, the Pentacam measured CCT values close to ultrasound pachymery with tendency for slightly lower values without differences attributed to age or sex therefore,it is considered as valid in determining central corneal thickness as ultrasound pachymetry with other advantages over the ultrasound pachymetry. Key words: Central corneal thickness, method comparison,scheimpflug imaging, ultrasound pachymetry. INTRODUCTION Accurate assessment of central corneal thickness (CCT)is important for many reasons: For example,refractive surgery is routinely planned according to preoperative measurement of central corneal thickness (American Academy, 1999; Price et al., 1999) to avoid complications such as corneal ectasia.also,cct can be regarded as a correlate of the physiological condition of the corneal endothelium and used in the diagnosis of certain corneal diseases such as keratoconus and Fuchs dystrophy (Gherghel et al., 2004). Also for accurate determination of intraocular pressure,a modification has to be made according to the central corneal thickness (Doughty M.J., 2000; Stodtmeister R., 1998; Gordon M.O., 2002; Wolfs R.C.,1997; Herndon L.W., 1997; Argus W.A. 1995; Medeiros F.A.,2003). Although ultrasound-based systems offer the advantages of relative ease of use, they experience relatively high errors in measurements, possibly as a result of difficulties in centration and alignment (McLaren et al., 2004; Doughty and Zaman, 2000; Thornton, 1984). The need for topical anesthesia and contact of the probe with the cornea has led to the search for noninvasivealternative solutions without the risk of epithelial lesions or transmission of infection. A novel apparatus capable of modeling the anterior chamber is the (Pentacam) Allergo, Germany.The system is based on a rotating Scheimpflug camera, which scans and measures thecomplete cornea and anterior chamber in approximately 2 seconds. Its reproducibility and accuracy in assessing CCT compared with traditional modalities have to be established. The purpose of the present study was to compare CCT measurementsobtained with the rotating Scheimpflug camera with ultrasound pachymetry to provideestimates for the average difference between measurements with the two different modalities and the validity of using the Pentacam as a good alternative for Ultrasound pachymetry prior to refractive surgery. Corresponding Author: TAMER ADEL REFAI,MD,FRCS., Departments of Ophthalmology, Refractive surgery unit, Research Institute of ophthalmology, Giza (Egypt). E-mail: tamerrefaey@hotmail.com. 5744

Aust. J. Basic & Appl. Sci., 4(12): 5744-5752, 2010 Methods: 117 eyes of (64 patients) seeking for refractive surgery; 72 female (F) eyes (of 38F patients) and 45 male (M) eyes (of 26 M patients ) with a mean age of 26.7 years ±5.6 years standard deviation were recruited for the study. All subjects had normal eyes without corneal abnormalities as verified by slit lamp examination. Softcontact lens wearers were included but were required not to wear contact lenses within 72 hours before the investigations. There fractive error was measured with an autorefractometer model Topcon 5000, Japan. The mean spherical equivalentwas -6.41D ± 3.76 standard deviation. Measurements were taken on one or both eyes according to patients demands for refractive surgery at least 3 hours after awakening. Central corneal thickness was determined with 2 different modalities: 1- Pentacam( ALLEGRO Oculyzer)Version1074;Allergo,Germany 2- ultrasound pachymetry (Sonoscan model 4000AP,SONOMED,USA). Each modality was carried out by the same observer. Ultrasound pachymetry was performed last to avoid influence of corneal microabrasions that might be induced by the ultrasound probe on the pentacam readings. Rotating Scheimpflug Imaging: Rotating Scheimpflug imaging was performed with the patient seated using a chinrest and forehead strap. The patient was asked tokeep both eyes open and to fixate on a blue fixation target. The system uses a rotating Scheimpflug camera and a monochromatic slit light source (blue LED at 475 nm) that rotate together aroundthe optical axis of the eye. Within 2 seconds, the system rotates180 and acquires 25 images that contain 500 measurement pointson the front and back corneal surfaces to draw a true elevation map. The software acquires the images as volume data, thus multiplan arreprojections allow the creation of sagittal and tangential maps (figure: 1). The thinnest value of corneal thickness was recorded as CCT. Only measurements with Quality Specification (QS) denoting OK were included in the study to avoid errors resulting from poor quality photos.at least three measurements are taken for each eye and the average number for CCT is included in the study. Fig. 1: Pentacam printout showing corneal thickness in addition to further multiple measurements. Ultrasound Pachymetry: CCT was measured using Sonoscan model 4000AP,SONOMED,USA) calibrated by the manufacturer.the cornea was anesthetized with topical 1%Bupivacaine. The patient was brought into a faceup position on the 5745

examination chair and asked to fixate a target on the ceiling.the pachymeter probe was brought in light contact with the cornea centrally and perpendicularly. The central reading of at least 9 pachymetry readings in the printout is taken as the CCT(Fig:2) At least 3 measurements are taken for the CCT and the average number is included in the study. Fig. 2: Ultrasound pachymetry printout showing the central and the other readings. Statistical Analysis: For the entire sample,central pachymetry for each patient by ultrasound and by Pentacam was subjected to statistical analysis including Mean value, Standard deviation, Minimum value, maximal value, Correlation of the measurements by both machines and Student t test for comparison of the measurements between both machines. To study the effect of sex on the measurements,the same statistical methods were applied to female and males groups under the sudy. To study the effect of age on the measurements,the same statistical methods were applied to two age groups (below 30y) &( above 30y) under the study. Results: The mean value for the central corneal thickness by ultrasound for the entire sample was 552.11μ ± 43.96 standard deviation with a minimum value of 471μ and a maximal value of 656μ while the mean value for 5746

the central corneal thickness by Pentacam for the entire sample was 541.82μ ± 39.87standard deviation with a minimum value of 451μ and a maximal value of 629μ. The student t test shows a value of 0.081(P=>0.05) which reveals a non difference between mean central corneal thickness measurement by both machines (table 1) & Pearson correlation shows a value of 0.892 (P value < 0.01 ) implying a hightly correlation between mean central corneal thickness measurement by both machines (table 2 ). The values for the the mean central corneal thickness by Pentacam were slightly lower than that obtained by ultrasound ( see Chart 1)by about 8.60μ ± 14.1. To study the effect of sex on the measurements,patients eyes were divided into two groups:a-female group: 72 female (F) eyes (of 38F patients) & B-Male group:45 male (M) eyes(of 26 M patients). For Females,the mean value for the central corneal thickness by ultrasound was 554.71μ± 46.14 standard deviation with a minimum value of 471μ and a maximal value of 656μ while the mean value for the central corneal thickness by Pentacam was 549.94μ±36.94 standard deviation with a minimum value of 462μ and a maximal value of 629μ. The student t test shows a value of 0.523 (P=>0.05) which reveals a non difference between mean central corneal thickness measurement by both machines for females (table 3) & Pearson correlation shows a value of 0.876 (P value < 0.01 ) implying a hightly correlation between mean central corneal thickness measurement by both machines for females (table 4). The values for the the mean central corneal thickness by Pentacam were slightly lower than that obtained by ultrasound for females (see Chart 2 and Graph 1). For males, the mean value for the central corneal thickness by ultrasound was 548.18μ±41.32standard deviation with a minimum value of 482μ and a maximal value of 646μ while the mean value for the central corneal thickness by Pentacam was 529.23μ±41.87standard deviation with a minimum value of 451μ and a maximal value of 629μ. The student t test shows a value of 0.042 (P=>0.05) which reveals a non difference between mean central corneal thickness measurement by both machines for males(table 3) & Pearson correlation shows a value of 0.931 (P value < 0.01) implying a hightly correlation between mean central corneal thickness measurement by both machines for males (table 4). The values for the the mean central corneal thickness by Pentacam were slightly lower than that obtained by ultrasound (see Chart 2&Graph 1). Inspite of slightly higher values by both machines for the female group(that is probably related to the larger size of the female group i:e 72 female (F) eyes compared to 45 male(m) eyes),yet no statistically difference between measurements by both machines was attributed to sex of the patient. To study the effect of age on the measurements,patients eyes were divided into two age groups: Age Group A (<30y):87eyes (of 49 patients) & Age Group B (>30y):30 eyes (of 17 patients). For age group below 30y,The mean value for the central corneal thickness by ultrasound was 558.36μ± 44.90 standard deviation with a minimum value of 489μ and a maximal value of 656 μ while the mean value for the central corneal thickness by Pentacam was 545.3μ± 37.26 standard deviation with a minimum value of 462μ and a maximal value of 609μ. Chart 1: Mean value for the central corneal thickness for the entire sample by ultrasound and pentacam 5747

Chart 2: Mean value for the central corneal thickness for females by ultrasound and pentacam. Graph.1: Mean value for the central corneal thickness for females and males by ultrasound and pentacam The student t test shows a value of 0.094(P=>0.05) which reveals a non difference between mean central corneal thickness measurement by both machines for patients below 30y(table 5) & Pearson correlation shows a value of 0.869 (P value < 0.01) implying a hightly correlation between mean central corneal thickness measurement by both machines. (table 6). The values for the the mean central corneal thickness by Pentacam were slightly lower than that obtained by ultrasound (see Chart 3 and Graph 2). 5748

Chart 3: Mean value for the central corneal thickness for according to age by ultrasound and pentacam. Graph.2: Mean value for the central corneal thickness for according to age by Ultrasound and pentacam For age group above 30y,The mean value for the central corneal thickness by ultrasound was 541.63μ± 32.80standard deviation with a minimum value of 490μ and a maximal value of 585 μ while the mean value for the central corneal thickness by Pentacam was 535.86μ± 24.71standard deviation with a minimum value of 505μ and a maximal value of 566μ. The student t test shows a value of 0.71(P=>0.05) which reveals a non difference between mean central corneal thickness measurement by both machines for patients above 30y(table 5) & Pearson correlation shows a value of 0.953 (P value < 0.01) implying a hightly correlation between mean central corneal thickness measurement by both machines. (table 6). The values for the the mean central corneal thickness by Pentacam were slightly lower than that obtained by ultrasound ( see Chart 3 and Graph 2). Inspite of slightly higher values by both machines for age group <30y(that is probably related to the larger size of age group <30y i:e 87eyes (of 49 patients )compared to 30 eyes (of 17 patients) for the age group >30y, yet no statistically difference between measurements by both machines was attributed to age of the patient. 5749

Table 1: Comparison of Mean value and standard deviation for the central corneal thickness for the entire sample for the etire sample by ultrasound and pentacam by t- test. Item CCT By ultrasound CCT By Pentacam t test Probability Significance Mean value for central 552.11± 43.96 541.82± 39.87 0.081 >0.05 Non corneal thickness (CCT) Table 2: Mean value and standard deviation for the centeral corneal thickness for the entire sample by ultrasound and pentacam and their correlation. Item CCT By ultrasound CCT By Pentacam Correlation Probability Significance Mean value for central 552.11± 43.96 541.82± 39.87 0.892498 <0.01 Highly corneal thickness (CCT) Table 3 : Comparison of Mean value and standard deviation for the central corneal thickness by ultrasound and pentacam by t-test for females and males. Item CCT By ultrasound CCT By Pentacam t test Probability Significance Mean value for central 554.71± 46.14 549.94± 36.94 0.523 >0.05 Non for females Mean value for central 548.18± 41.32 529.23± 41.87 0.042 >0.05 Non for females Table 4: Mean value and standard deviation for the central corneal thickness for females and males by ultrasound and pentacam and their correlation. Item CCT By ultrasound CCT By Pentacam Correlation Probability Significance Mean value for central 554.71± 46.14 549.94± 36.94 0.876 <0.01 Highly for females Mean value for central 548.18± 41.32 529.23± 41.87 0.931 <0.01 Highly for males Table 5: Comparison of Mean value and standard deviation for the central corneal thickness by ultrasound and pentacam by t-test for patients below and above 30 years. Item CCT By ultrasound CCT By Pentacam t test probability Significance Mean value for central 558.36±44.90 545.30±37.26 0.094 >0.05 Non for <30y Mean value for central 545.07±32.80 535.86±24.71 0.710 >0.05 Non for for >30y Table 6: Mean value and standard deviation for the central corneal thickness for patients below and above 30years by ultrasound and pentacam and their correlation. Item CCT By ultrasound CCT By Pentacam correlation probability Significance Mean value for central 558.36±44.90 545.30±37.26 0.869 <0.01 Highly for <30y Mean value for central 541.63± 32.80 535.86±24.71 0.953 <0.01 Highly for for >30y Discussion: As previously mentioned,accurate measurement of corneal thickness is an important issue for a variety of reasons ;including prior to refractive surgery(to decrease the incidence of ectasia),for proper measuement of intraocular pressure and as an indirect indication for the endothelial cell function.although ultrasound pachymetry is considered the traditional way for measurement of corneal thickness, yet nowadays Pentacam is increasingly used for this purpose being easy, fast,noncontact way in addition to the ability of performing further measurements like anterior, posterior corneal topography in addition to anterior chamber depth all are of importance in refractive surgery. In this study,i compared the central corneal thickness measurements by ultrasound pachymetry (Sonoscan Model 4000 AP,Sonamed, USA) and by Pentacam (ALLEGRO Oculyzer Version 1074; Allergo,Germany). I found that the The mean value for the central corneal thickness by ultrasound for the entire sample was 552.11 ± 43.96 standard deviation while the mean value for the central corneal thickness by Pentacam for the entire sample was 541.82± 39.87standard deviation. The student t test shows a value of 0.081(P=>0.05) which reveals a non difference between mean central corneal 5750

thickness measurement by both machines & Pearson correlation shows a value of 0.892 (P value < 0.01) implying a hightly correlation between mean central corneal thickness measurement by both machines. The values for the the mean central coneal thickness by Pentacam were slightly lower than that obtained by ultrasound by about 8.60μ ± 14.1μ. Yaniv Barkana, MDetal in 2005 had reported that central corneal thickness measurements obtained by the Pentacam Scheimpflug device were highly correlated with those of the US pachymeter. Pentacam and US measurement differed by a mean of only 6 mm. Therefore, it seems that for most practical purposes, measurements with these instruments can be used interchangeably (Yaniv Barkana,). While Birgit Lackner, MD etal 2005,reported that in the present collective of 60 eyes of healthy European volunteers, CCT determined with with rotating Scheimpflug imaging (Pentacam)was on average 13 μ (2.4%)thinner than with ultrasound pachymetry. They conclude that Pentacam can be regarded as valid in determining CCT in eyes with normal cornea as the other modalities (Birgit et al., 2005). To study the effect of sex on the measurements,patients eyes were divided into two groups:a-female group: 72 female (F) eyes (of 38F patients) & B-Male group:45 male(m) eyes (of 26 M patients ). For Females,the mean value for the central corneal thickness by ultrasound was 554.71μ± 46.14 standard deviation while the mean value for the central corneal thickness by Pentacam was 549.94μ±36.94 standard deviation. The student t test shows a value of 0.523 (P=>0.05) which reveals a non difference between mean central corneal thickness measurement by both machines for females & Pearson correlation shows a value of 0.876 (P value < 0.01) implying a hightly correlation between mean central corneal thickness measurement by both machines for females. The values for the the mean central corneal thickness by Pentacam were slightly lower than that obtained by ultrasound for females. For males, the mean value for the central corneal thickness by ultrasound was 548.18μ±41.32standard deviation while the mean value for the central corneal thickness by Pentacam was 529.23μ±41.87standard deviation. The student t test shows a value of 0.042 (P=>0.05) which reveals a non difference between mean central corneal thickness measurement by both machines for males& Pearson correlation shows a value of 0.931 (P value < 0.01) implying a hightly correlation between mean central corneal thickness measurement by both machines for males. The values for the the mean central corneal thickness by Pentacam were slightly lower than that obtained by ultrasound. Inspite of slightly higher values by both machines for the female group (that is probably related to the larger size of the female group i:e 72 female(f) eyes compared to 45 male(m) eyes), yet no statistically difference between measurements by both machines was attributed to sex of the patient. To study the effect of age on the measurements,patients eyes were divided into two age groups: Age Group A(<30y):87eyes (of 49 patients) & Age Group B(>30y):30 eyes (of 17 patients). For age group below 30y,The mean value for the central corneal thickness by ultrasound was 558.36μ± 44.90 standard deviation while the mean value for the central corneal thickness by Pentacam was 545.3μ± 37.26 standard deviation. The student t test shows a value of 0.094(P=>0.05) which reveals a non difference between mean central corneal thickness measurement by both machines for patients below 30y & Pearson correlation shows a value of 0.869 (P value < 0.01) implying a hightly correlation between mean central corneal thickness measurement by both machines.the values for the the mean central corneal thickness by Pentacam were slightly lower than that obtained by ultrasound. For age group above 30y, The mean value for the central corneal thickness by ultrasound was 541.63μ± 32.80standard deviation while the mean value for the central corneal thickness by Pentacam was 535.86μ± 24.71standard deviation. The student t test shows a value of 0.71(P=>0.05) which reveals a non difference between mean central corneal thickness measurement by both machines for patients above 30y& Pearson correlation shows a value of 0.953 (P value < 0.01) implying a hightly correlation between mean central corneal thickness measurement by both machines. The values for the the mean central corneal thickness by Pentacam were slightly lower than that obtained by ultrasound.inspite of slightly higher values by both machines for age group <30y(that is probably related to the larger size of age group <30y i:e 87eyes (of 49 patients) compared to 30 eyes (of 17 patients) for the age group >30y, yet no statistically difference between measurements by both machines was attributed to age of the patient. 5751

Conclusion: In the assessment of normal corneas, the Pentacam measured CCT values close to ultrasound pachymery with tendency for slightly lower values by about 8.60μ ± 14.1μ without differences attributed to age or sex therefore,it is considered as valid in determining central corneal thickness as ultrasound pachymetry with other advantages over the ultrasound pachymetry. REFERENCES American Academy of Ophthalmology, 1999. Excimer laser photorefractive keratectomy (PRK) for myopia and astigmatism.ophthalmic procedure preliminary assessment. Ophthalmology, 106: 422-437. Argus W.A., 1995. Ocular hypertension and central corneal thickness. Ophthalmology, 102: 1810-2. Birgit Lackner, M.D., Gerald M.D. Schmidinger, et. al, 2005. Optometry and Vision; 82(10): 892-899. Doughty M.J., Zaman ML. Human 2000. corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach.surv Ophthalmol., 44: 367-408. Doughty M.J., M.L. Zaman, 2000. Human corneal thickness and its impacton intraocular pressure measures: a review and meta-analysis approach.surv Ophthalmol., 44: 367-408. Gherghel D., S.L. Hosking, S. Mantry, S. Banerjee, S.A. Naroo, S.Shah, 2004. Corneal pachymetry in normal and keratoconic eyes: Orbscan II versusultrasound. J Cataract Refract Surg, 30: 1272-7. Gordon M.O.,.JA. Beiser, J.D. Brandt, et al. The Ocular Hypertension Treatment Study; baseline factors that predict the onset of primary open angleglaucoma; the Ocular Hypertension Treatment Study Group.Arch Ophthalmol., 120: 714-720. Herndon L.W., S.A. Choudhri, T. Cox, K.F. Damji, M.B. Shields, R.R. Allingham, 1997. Central corneal thickness in normal, glaucomatous, and ocular hypertensive eyes.arch Ophthalmol., 115: 1137-41. Medeiros F.A., P.A. Sample, L.N. Zangwill, C. Bowd, M. Aihara, R.N. Weinreb, 2003. Corneal thickness as a risk factor for visual field loss in patients with preperimetric glaucomatous optic neuropathy.am J Ophthalmol., 136: 805-13. Medeiros F.A., P.A. Sample, R.N. Weinreb, 2003. Corneal thickness measurement sand visual function abnormalities in ocular hypertensive patients.am J Ophthalmol., 135: 131-7. McLaren J.W., C.B. Nau, J.C. Erie, W.M. Bourne, 2004. Corneal thickness measurement by confocal microscopy, ultrasound, and scanning slitmethods.am J Ophthalmol., 137: 1011-20. Price F.W., Jr, D.L. Koller, M.O. Price, 1999 Central corneal pachymetry in patients undergoing laser in situ keratomileusis. Ophthalmology, 106: 2216-2220. Stodtmeister R., Applanation tonometry and correction according to corneal thickness. Acta Ophthalmol Scand., 76: 319-324 Thornton, S.P., 1984. A guide to pachymeters.ophthalmic Surg, 15: 993-5. Wolfs R.C., C.C. Klaver, J.R. Vingerling, D.E. Grobbee, A. Hofman, P.T. dejong Distribution of central corneal thickness and its association with intraocular pressure: the Rotterdam Study. Am J Ophthalmol., 123: 767-72. Yaniv Barkana., M.D., Yariv P.h.D. Gerber, et al, 2005. Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry. J Cataract Refract Surg, 31: 1729-1735. 5752