Talar Dome System Surgical Technique

Similar documents
Hip Resurfacing System

Technique Guide Hip Resurfacing System

WristMotion Wrist Hemiarthroplasty System Instructions for Use

Technique Guide Wrist Hemiarthroplasty System

Description Materials Indications Patient selection factors to be considered include: Contraindications Absolute contraindications include:

Technique Guide Toe Hemiarthroplasty

Technique Guide Toe Hemiarthroplasty

Lapidus Arthrodesis System Instructions for Use

Pinit Plate Small Bone Fusion System Bone Plate & Screw System

Surgical Technique. Poly UHMWPE. Bio Hyaluronic Acid. Advancing Materials. Advancing Outcomes.

Surgical Technique. Poly UHMWPE. Bio Hyaluronic Acid. Advancing Materials. Advancing Outcomes.

Technique Guide Small Bone Fusion System

Arthroscopically assisted Knee Resurfacing (AKR)

Glenoid Resurfacing Technique Guide

Surgical Technique. Poly UHMWPE. Bio Hyaluronic Acid. Advancing Materials. Advancing Outcomes.

Technique Guide. Primary Stemless Shoulder System

Technique Guide Lapidus Arthrodesis System

Positioning Sleeve Surgical Technique

Technique Guide Hammertoe Correction System

Technique Guide Patello-Femoral Arthroplasty

ACETABULAR CUP SURGICAL TECHNIQUE

SURGICAL TECHNIQUE GUIDE

Technique Guide KISSloc Suture System

BIOLOX delta Option Ceramic Femoral Head System. Product Features and Instructions for Use

BIOLOX delta Option Ceramic Femoral Head System

Cementless Tapered Femoral Stem Surgical technique

LOWER EXTREMITIES SINGLE-USE INSTRUMENTATION SURGICAL IMPLANTATION TECHNIQUE. First Metatarsal Phalangeal Joint FOOT THE DIFFERENCE IS MOVING.

Taperloc Complete Hip System. Surgical Technique

Cementless Tapered Femoral Stem Surgical technique

FUSEFORCE. Hand Fixation System SURGICAL TECHNIQUE

S H O U L D E R Solutions by Tornier. Aequalistm. Pyrocarbon Humeral Head. surgical technique

CAUTION Federal law (USA) restricts this device to sale, by or on the order of a physician.

CAUTION: Ceramic liners are not approved for use in the United States.

SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS

SURGICAL TECHNIQUE CHI. Cannulated Hemi Implants 1 ST MPJ ARTHROPLASTY

Salto Talaris Total Ankle Prosthesis

3. PATIENT POSITIONING & FRACTURE REDUCTION 3 8. DISTAL GUIDED LOCKING FOR PROXIMAL NAIL PROXIMAL LOCKING FOR LONG NAIL 13

Headless Compession Screw 2.5 / 3.0

Headless Compession Screw 4.5 / 6.5 SURGICAL TECHNIQUE. Titanium or Stainless Steel. Cannulated Headless Design. Multiple Thread Options.

Encina Taper Stem. Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA

INBONE II. Total Ankle System SURGICAL TECHNIQUE SUPPLEMENT TIBIAL REAMING SPECIFICATION

Surgical Technique. Cup System

Flexible Fragment Fixation. Surgical Technique

Surgical Technique. Poly UHMWPE. Bio Hyaluronic Acid. Advancing Materials. Advancing Outcomes.

For the Attention of the Operating Surgeon: IMPORTANT INFORMATION ON THE MATRIXRIB FIXATION SYSTEM

CMA TM Cemented Modular Augmentable. Baseplate. 2 Knee

Revision A Date:

TM TM Surgical Technique

DESIGN RATIONALE AND SURGICAL TECHNIQUE

CANNULINK. Intraossous Fixation System SURGICAL TECHNIQUE

Important notice: the device(s) can be prescribed and implanted only by a doctor legally authorized to perform this type of surgery.

Comprehensive Reverse Shoulder System Augmented Baseplate

TriboFit Hip System. Issue 3 - August 2013 English

SURGICAL TECHNIQUE GUIDE TRESTLE. Anterior Cervical Plating System

CARTIVA. Synthetic Cartilage Implant SURGICAL IMPLANTATION TECHNIQUE. First Metatarsal Phalangeal Joint THE DIFFERENCE IS MOVING.

Precautionary Statement ( )

System. Humeral Nail. Surgical Technique

Surgical procedure. METS Modular Proximal Femur

VECTRA-T SURGICAL TECHNIQUE. The Translational Anterior Cervical Palate System. This publication is not intended for distribution in the USA.

EVOLVE TRIAD BONE SCREWS

U2 PSA. Revision Knee. Surgical Protocol

Royal Oak Cervical Plate System

TC-PLUS Primary Knee IMPORTANT MEDICAL INFORMATION SPECIAL NOTE. This Package Insert is for product distributed in the US only.

nva Anterior Lumbar Interbody Fusion System

Progeny Hip Stem. Surgical Protocol and Product Specifications

Revision. Hip Stem. Surgical Protocol

nvt Transforaminal Lumbar Interbody Fusion System

Table of Contents.

U2 PSA. Revision Knee. Surgical Protocol

Optimizing function Maximizing survivorship Accelerating recovery

ESC. Enhanced Stability Liners. Design Rationale & Surgical Technique

Active Articulation. Surgical Technique. Dual Mobility Hip System. Knees Hips Extremities Cement and Accessories PMI Trauma Technology

Document No Rev. B Pg. 1 of 5 Approvals / Date Title: INSTRUCTIONS FOR USE DJO SURGICAL MODULAR REVISION HIP SYSTEM AND ACETABULAR CAGE

Technique Guide. VA-Locking Intercarpal Fusion System. Variable angle locking technology for mediocarpal partial arthrodesis.

Manza Cup HA SURGICAL TECHNIQUE.

TaperFill. Surgical Technique

LMH. Minimal Bone Resection. Lesser Metatarsal Head Implant. Thin, low-profile design for minimal bone resection

nvp Posterior Lumbar Interbody Fusion System

2.4 mm Variable Angle LCP Intercarpal Fusion System. Variable angle locking technology for mediocarpal partial arthrodesis.

Ceramic Femoral Heads and Acetabular Cup Liners

HELIOS h i p s y s t e m

Regenerex Primary Tibial Tray

Arcos Interlocking Distal Stem. Surgical Technique Addendum to the Arcos Modular Femoral Revision System

ZMR Over-the-Junction Instruments for Revision Hip Arthroplasty. Surgical Technique IMAGE TO COME

Metallic Internal Fixation Devices The following languages are included in this packet:

Surgical procedure. METS Coned Hemi-Pelvis

Zimmer NexGen Trabecular Metal Tibial Tray

Lesser MPJ Hemi Implant

HEMI IMPLANT ARTHROPLASTY FOR THE SECOND METATARSOPHALANGEAL JOINT

LBL-014. Rev

Enhanced Stability Constrained Liners. Design Rationale Surgical Technique

SURGICAL TECHNIQUE GUIDE

Cervical Spacer System surgical technique

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

DYNAMIC, TRANSVERSE COMPRESSION. Low Profile, Anatomic Design, Type II Anodized. Mechanical Compression Designed to Stimulate the Fusion Process

ACP1 CERVICAL PLATE SPINAL SYSTEM SURGICAL TECHNIQUE GUIDE II.

Duraloc CONSTRAINED LINER

ACP. Anterior Cervical Plate System SURGICAL TECHNIQUE

Trinica and Trinica Select Anterior Cervical Plate System

Transcription:

Talar Dome System Surgical Technique

CAP TALAR DOME RESURFACING HEMIARTHROPLASTY IMPLANT Surgical Technique Guide Description The HemiCAP Contoured Articular Prosthetic incorporates an articular resurfacing component and a cancellous taper post component that mates together via a taper interlock to provide stable and immobile fixation of the implant and stress bearing contact at the bone/prosthetic interface. Materials: Articular Resurfacing Component: Surface Coating: Taper Post Component: Cobalt-Chromium-Molybdenum alloy (Co-Cr-Mo) Titanium (CP Ti) Titanium alloy (Ti-6AI-4V)

Indications for Use: Partial resurfacing of the talar dome of the ankle for use in the treatment of patients with localized post-traumatic degenerative disease, necrosis associated with large unstable osteochondral fractures, or osteochondritis dessicans. Soft tissues and other structures contributing to joint stability should be intact or reconstructable. The intended use of the device is part of an interim clinical strategy for patients who have not responded to other treatments and who will likely receive a joint replacement or fusion in the future. The device is a single use implant. Patient selection factors to be considered include: 1) Patient has localized disease of the ankle and talar dome 2) Patient s need to obtain pain relief and improve function; 3) Patient s age as relative contraindication to an arthrodesis or joint replacement procedure; and 4) Patient s overall well-being, including ability and willingness to follow instructions and comply with activity restrictions.

Contraindications: Absolute contraindications include: 1) Extensive talar necrosis, significant bone demineralization or inadequate bone stock; 2) Inadequate skin, musculotendinus or neurovascular compromise; 3) Inflammatory or rheumatoid arthritis, infection, sepsis, and osteomyelitis; and 4) Patients that have a known sensitivity to Cobalt-Chrome alloys typically used in prosthetic devices. Relative contraindications include: 1) Uncooperative patient or patient incapable of following pre-operative and postoperative instructions; 2) Osteoporosis; 3) Metabolic disorders which may impair the formation or healing of bone; 4) Infections at remote sites which may spread to the implant site; rapid joint destruction or bone resection visible on roentgenogram 5) Loss of lateral ligaments, chronic instability or deficient soft tissues and other support structures; and 6) Vascular or muscular insufficiency. 7) Diffuse tibiotalar degeneration 8) Severe malalignment of the ankle Pre-operative Considerations and Assessments: 1) Extent of the lesion and radiographic alignment; 2) Neurovascular status and soft tissue coverage; 3) Location of the lesion to be resurfaced and approach that will be required for exposure

History and Challenges of Talar Why? Where the occur? Defects Challenge to reproduce anatomic fit Why HemiCAP is suitable to address these lesions Other thoughts?

Surgical Exposure as described by Prof. C. Niek van Dijk, Chair, AMC, Netherlands Exposure Pictures used here are representative and not final

1. Use Drill Guide to locate the axis normal to the articular surface and central to the defect. Place Guide Pin into a Cannulated Powered Drill and secure at the etch marking on the Guide Pin. Advance Guide Pin into bone making sure that it is central to the defect. (It is important to verify the Drill Guide is seated on the curved surface such that four points of contact are established. A normal axis is necessary for proper implant fit.)

2. Place Cannulated Drill over Guide Pin and drive until the proximal shoulder of the Drill is flush with the articular surface. (Use lavage during drilling to prevent possible tissue damage from heat effects.)

3. Tap hole to etched depth mark on Tap.

4. Before inserting the Taper Post, thoroughly cleanse the pilot hole of any debris and then inject the cement in a retrograde fashion from the end of the hole upwards.

5. Place the Driver into the Taper Post and advance the Taper Post until the line on the Driver is flush with the cartilage surface.

6. Clean taper in Taper Post with Taper Cleaner. 7. Place Trial Cap into Taper Post to confirm correct depth of Taper Post. The height of the Trial Cap must be flush or slightly below the existing articular cartilage surface to avoid the Articular Component from being placed proud or above the surface of the defect. Adjust depth if needed using the Driver to rotate the Taper Post (rotate clockwise to advance and counterclockwise to retract). Remove Trial Cap.

8. Place Centering Shaft into taper of Taper Post. Place Contact Probe over Centering Shaft and rotate around Centering Shaft. Read Contact Probe to obtain offsets at the 4 indexing points and mark each of the identified offsets on the appropriate Sizing Card. Select appropriate Articular Component using Sizing Card.

9. Remove Centering Shaft and replace with Guide Pin. Advance Circle Cutter onto the articular surface by twisting the Circle Cutter back and forth avoiding any bending of the Guide Pin.

10. Choose the appropriate Surface Reamer based on the offsets. Confirm selection by matching the color code on the Articular Component package with the colored band on the Surface Reamer shaft. Drive Surface Reamer over Guide Pin until it contacts the top surface on Taper Post. (Use lavage during drilling to prevent possible tissue damage from heat effects.) Make sure not to bend the Guide Pin during drilling as it may result in Articular Component malalignment.

6. Clean taper in Taper Post with Taper Cleaner.

11. Place the Sizing Trial into the defect that matches the offset profile of the chosen HemiCAP Articular Component. Confirm the fit of the Sizing Trial so that it is congruent with the edge of the surrounding articular surface or slightly recessed.

12. Before placing the Articular Component on the Implant Holder make sure that sufficient suction is present to hold the device on the distal suction cup. Align the Articular Component on the Implant Holder. Orient the etch marks on the back of the Articular Component with the etch mark on the handle of the Implant Holder. Align the Articular Component with the appropriate offsets. Insert into taper of Taper Post.

13. Use a slight tap on the Impactor to seat Articular Component. Progressively tap the Impactor until the Articular Component is firmly seated on the bone.

Warnings Improper selection, placement, positioning, alignment, and fixation of the implant components may reduce the service life of the prosthetic components. Inadequate preparation and cleaning of the implant components mating surfaces may result in improper fixation of the device. Improper handling of the implants can produce scratches, nicks or dents that may have adverse clinical effects on mating joint surfaces. Do not modify implants. The surgeon shall be thoroughly familiar with the implants, instruments, and surgical technique prior to performing surgery. When mapping articular surfaces, care should be taken to ensure that instrument taper surfaces are properly aligned and mated with taper in Taper Post. Visually confirm distal tip of contact probe is making contact on articular surfaces and free from any soft tissue structures to ensure accuracy. Use light pressure on contact probe to slightly indent articular surface at mapping points, ensuring that the selected implant will be flush or slightly recessed just below articular surface at margins of implant. Prior to placing implant, carefully trim articular cartilage debris around margin of implant. Remove bone particles and lavage thoroughly. To ensure mechanical interlock of the Taper Post and implant, carefully clean Taper Post taper with provided instruments. All drilling or reaming should be done at slowest speeds possible with vigorous lavage to minimize heat effects to adjacent bone and cartilage tissues. Accepted practices in post-operative care should be used. The patient is to be instructed and monitored to ensure a reasonable degree of compliance to post operative instructions and activity restrictions. Excessive activity, impact, and weight gain have been implicated in the reduction of the benefit and service life of prosthetic devices.

Precautions The HemiCAP Contoured Articular Prosthetic Talar Dome implant is intended to be fitted and installed with the HemiCAP Contoured Articular Prosthetic Talar Dome instrument set. Use of instruments from other systems may result in improper implant selection, fitting, and placement which could result in implant failure or poor clinical outcome. The instrument set should be regularly inspected for any signs of wear or damage. Do not reuse implants. Possible Adverse Effects Material sensitivity reactions. Implantation of foreign material in tissues can result in histological reactions. Particulate wear debris and mild tissue discoloration from metallic components have been noted in other prosthetic devices constructed of similar materials. Some types of wear debris have been associated with osteolysis and implant loosening. Infection or allergic reaction. Loosening, migration or loss of fixation of implant. Fretting and crevice corrosion can occur at the interface between the implant components. Fatigue fracture of the implants as a result of bone resorption around the implant components. Wear and damage to the implant articulating surface. Wear and damage to the adjacent and opposed articular cartilage surfaces or soft tissue support structures. Intraoperative or postoperative bone fracture Postoperative pain or incomplete resolution of preoperative symptoms. Sterility Prosthetic components are sterilized by exposure to gamma irradiation. Do not resterilize. Do not use components if packaging is opened or damaged. Do not use components if beyond expiration date.

T000-1500 Kit, Instrument, 15mm, Talus 9007-1300 2mm Guide Pin, Talus T103-0020 Taper Post, 10.3mm T152-0540 T152-0545 T152-1040 T152-1045 T152-1050 T152-1055 T152-1540 T152-1545 T152-1550 T152-1555 T152-2040 T152-2045 T152-2050 T152-2055 15mm Talus Art. Comp., 0.5 x 4.0mm 15mm Talus Art. Comp., 0.5 x 4.5mm 15mm Talus Art. Comp., 1.0 x 4.0mm 15mm Talus Art. Comp., 1.0 x 4.5mm 15mm Talus Art. Comp., 1.0 x 5.0mm 15mm Talus Art. Comp., 1.0 x 5.5mm 15mm Talus Art. Comp., 1.5 x 4.0mm 15mm Talus Art. Comp., 1.5 x 4.5mm 15mm Talus Art. Comp., 1.5 x 5.0mm 15mm Talus Art. Comp., 1.5 x 5.5mm 15mm Talus Art. Comp., 2.0 x 4.0mm 15mm Talus Art. Comp., 2.0 x 4.5mm 15mm Talus Art. Comp., 2.0 x 5.0mm 15mm Talus Art. Comp., 2.0 x 5.5mm