Low-Profile Wrist Fixator. For stabilization of fractures of the distal radius.

Similar documents
Medium External Fixator Humeral Shaft Frame. Modular frame for upper extremity use.

Medium External Fixator Pediatric Femoral Shaft Frame. Using medium multi-pin clamps.

Technique Guide. Adjustable Distal Radius Fixator. Part of the Synthes External Fixation Systems.

External Distal Radius Fixator. Supplement to the 8 mm rod fixator system

Large External Fixator Modular Knee Bridge. Using multi-pin clamps.

Sterile-Packaged Large External Fixator Kits. For treatment of long bone and pelvic fractures that require external fixation.

For Small-Statured Adults and Pediatric Patients. Medium External Fixator Basic Modular Frame

Large External Fixator Modular Knee Bridge

Medium External Fixator Delta Frame Ankle Bridge

Small External Fixator Wrist Spanning Frame. For the treatment of wrist fractures.

For the Treatment of Wrist Fractures. Small External Fixator Nonspanning Wrist Frame

Large External Fixator Delta Frame Ankle Bridge. For staged fixation of the distal tibia.

Small External Fixator Nonspanning Wrist Frame. For the treatment of wrist fractures.

Large External Fixator Delta Frame Ankle Bridge. Using pin clamps with outrigger posts.

Large External Fixator Traveling Traction

Mandible External Fixator II. Provides treatment for fractures of the maxillofacial area.

2.4 mm LCP Volar Column Distal Radius Plates. Part of the 2.4 mm LCP Distal Radius System.

Technique Guide. The Distraction Osteogenesis Ring System. Intra-articular distal tibia frame.

Technique Guide. 3.5 mm LCP Olecranon Plates. Part of the Synthes locking compression plate (LCP) system.

Mini External Fixator

2.7 mm/3.5 mm LCP Distal Fibula Plate System. Part of the Synthes locking compression plate (LCP) system.

Large External Fixator Pelvic Frame

Basic Trocar System. Basic transbuccal instrumentation system for simplified intraoral plate and screw application.

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

3.5 mm LCP Tibial Plateau Sets. Configuration options.

LCP Distal Femur Plates. Shape based on Distal Femur LISS Plate design.

Technique Guide. Self-Drilling Schanz Screws. Designed to optimize the bone/pin interface.

MEFiSTO. Monolateral External Fixation System for Trauma and Orthopaedics.

Small External Fixator

Hybrid Fixator Proximal Tibia Frame. Using rings with clamps.

Technique Guide. 2.7 mm/3.5 mm LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments. Part of the 3.5 mm LCP Percutaneous Instrument System.

Technique Guide. SureLock Distal Targeting Device. C-arm guided targeting for trochanteric fixation nail.

Mini External Fixator.

Femur Frame. The Distraction Osteogenesis Ring System

Pediatric LCP Plate System. For osteotomies and fracture fixation of the proximal and distal femur.

The Wrist Fusion Set. Stainless Steel and Titanium TECHNIQUE GUIDE. Instruments and implants approved by the AO Foundation

Small External Fixator, Radiolucent, Sterile. Specific design for distal radius fractures.

2.4 mm Variable Angle LCP Intercarpal Fusion System. Variable angle locking technology for mediocarpal partial arthrodesis.

designed to advance the treatment of hip fractures.

FastFrame External Fixation System. Damage Control Surgical Technique

Surgical Technique. Distal Radius and Foot

Spinal Correction FRA Spacer System. For use with the Small Stature FRA Spacer and the FRA Spacer.

Compact Distal Radius System. Consolidated solution for Distal Radius Plating Systems

Technique Guide. 2.4 mm Variable Angle LCP Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

Technique Guide. Midface Distractor System. For the temporary stabilization and gradual lengthening of the cranial or midfacial bones.

3.0/3.5/4.0/4.5/6.5/7.0/7.3. Cannulated Screws. Surgical Technique

LCP Metaphyseal Plates. For extra-articular fractures.

LCP Periprosthetic System. Part of the Synthes locking compression plate (LCP) system.

The Distraction Osteogenesis Ring System

The Calcaneal Plate. The Synthes non-locking solution for the Calcaneus.

The Distraction Osteogenesis Ring System

OBSOLETED. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

Technique Guide. Cranial Clamps with Application Instrument. Quick and stable fixation of cranial bone flaps following a craniotomy.

Lag Screw Device Intended for symphyseal fracture fixation of the mandible

Distal Radius Plate 2.4/2.7 dorsal and volar

Technique Guide. 3.5 mm LCP Periarticular Proximal Humerus Plate. Part of the Synthes locking compression plate (LCP) system.

Technique Guide. Epoca Revision Instrument Set. For extraction and revision of Epoca shoulder implants.

Visualize, stabilize, mobilize. Wristore * Distal Radius Fracture Fixator Abbreviated Surgical Technique

2.4 mm LCP Radial Head Plates. Part of the Synthes LCP Distal Radius Plate System.

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

LCP Anterior Ankle Arthrodesis Plates. Part of the Synthes Locking Compression Plate (LCP) System.

Simply Complete. VA-LCP Elbow Plating System.

Trephine System. Principle-based anterior lumbar interbody fusion.

Pediatric LCP Hip Plate. For osteotomy and trauma applications in the proximal femur.

The CerviFix System. Including the StarLock Components. CerviFix Clamp and Screw. StarLock Clamp and Screw

The Distraction Osteogenesis Ring System

Wrist Fusion Instrument and Implant Set.

Thoracolumbar Spine Locking Plate (TSLP) System. A low-profile plating system for anterior stabilization of the thoracic and lumbar spine.

Case Study. Micrognathia Secondary to Pierre Robin Sequence. Treated with distraction osteogenesis using an internal mandible distractor.

Monolateral External Fixation System for Trauma and Orthopaedics

*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Knee Implants

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

Elbow Hinge Fixator. Guided Flexion/Extension for Unstable Elbow Fractures.

DISTRACTION PRODUCT OVERVIEW. For a wide variety of facial applications

The NBX Non-Bridging External Fixator A Non-Bridging External Fixator/Locking Plate capturing a series of.062mm K-wires and 3mm half-pins that are

Technique Guide. IMF Screw Set. For intermaxillary fixation.

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

CSLP Quick Lock Screws. Preassembled expansionhead screw and locking screw for use with Cervical Spine Locking Plates (CSLP).

EasyStep. Operative technique

VA-Locking Intercarpal Fusion System. Variable angle locking technology for mediocarpal partial arthrodesis.

Mandible External Fixator II

RAPIDSORB RAPID ReSORBABLe CRANIAL CLAMP

MEFiSTO. Monolateral External Fixation System for Trauma and Orthopaedics.

Distal Radius Plate Instrument and Implant Set. Discontinued December 2017 DSUS/TRM/0916/1063(1)

MRI Guidelines for the Axonics Sacral Neuromodulation System

Technique Guide. SynPOR Porous Polyethylene Implants. For craniofacial and orbital augmentation and reconstruction.

Extron External Fixator

*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Shoulder Implants

LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

3. Insert Tocar Sleeves Insert the NCB tissue protection sleeve assembly 1.6 to 10mm through a skin incision (Fig. 38).

3.5 mm LCP Olecranon Plates

Hoffmann II Compact External Fixation System. Modular System for

QUICK REFERENCE GUIDE. The PreFix Fixator (92000 Series) ALWAYS INNOVATING

Mandible External Fixator II. Provides treatment for fractures of the maxillofacial area.

3.5 mm Locking Attachment Plate

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

pediatric orthopedic implants fixator

LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System

Transcription:

Low-Profile Wrist Fixator. For stabilization of fractures of the distal radius. Technique Guide Part of the External Fixation System

Low-Profile Wrist Fixator Indications Intended for stabilization of fractures of the distal radius. Low-Profile Wrist Fixator Clamp Independently locks to Schanz screws while allowing universal joint motion for reduction in all planes Allows secondary adjustment of length without loss of reduction Disposable Parallel Drill Guide Lightweight plastic handle with stainless steel drill sleeves 6.0 mm Carbon Fiber Rod Includes a choice of 200 mm or 220 mm length carbon fiber rod Lightweight for patient comfort Radiolucent for improved intra- and postoperative radiographic visualization Synthes 1

MRI Information Synthes Low-Profile Wrist Fixator devices are labeled MR Conditional according to the terminology specified in ASTM F2503-05, Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Enviroment. Nonclinical testing demonstrated that when used in the specific configurations stated in Synthes labeling, Synthes Low-Profile Wrist Fixator devices are MR Conditional. Representative Synthes Low-Profile Wrist Fixator devices used in a typical construct include clamps, rods and various attachments. A patient with a Synthes Low-Profile Wrist Fixator may be scanned safely after placement of the fixator under the following conditions. Static magnetic field of 1.5 Tesla when the fixation frame is positioned: 7 cm or less from within the outside edge of the bore of the MRI at Normal Operating Mode or; Completely outside of the MRI bore in First Level Controlled Mode Static magnetic field of 3.0 Tesla when the fixation frame is positioned: 7 cm or less from within the outside edge of the bore of the MRI at Normal Operating Mode or; Completely outside of the MRI bore in First Level Controlled Mode Highest spatial gradient magnetic field of 900 Gauss/cm or less Maximum MR system reported whole body averaged specific absorption rate (SAR) of 2 W/kg for the Normal Operatinig Mode and 4 W/kg for the First Level Controlled Mode for the 15 minutes of scanning Use only whole body RF transmit coil, no other transmit coils are allowed, local receive only coils are allowed Note: In nonclinical testing, the Synthes external fixation frame was tested in several different configurations. This testing was conducted with the construct positioned 7 cm from within the outside edge of the MRI bore. The results showed a maximum observed heating for a wrist fixation frame of 6ºC for the 1.5 T and less than 1ºC for 3.0 T with a machine reported whole body averaged SAR of 2 W/kg. Patients may be safely scanned in the MRI chamber at the above conditions. Under such conditions, the maximal expected temperature rise is less than 6ºC. Because higher in vivo heating cannot be excluded, close patient monitoring and communication with the patient during the scan is required. Immediately abort the scan if the patient reports burning sensation or pain. To minimize heating, the scan time should be as short as possible, the SAR as low as possible, and the device should be as far as possible from the edge of the bore. Temperature rise values obtained were based upon a scan time of 15 minutes. The above field conditions should be compared with those of the user s MR system to determine if the item can safely be brought into the user s MR enviroment. If placed in the bore of the MR scanner during scanning, Synthes MR Conditional external fixation devices may have the potential to cause artifact in the diagnostic imaging. All components of Synthes external fixation frames must be identified as MR Conditional before being placed in or near an MR enviroment. Artifact information MR image quality may be compromised if the area of interest is in the same area or relatively close to the position of the Synthes Low-Profile Wrist Fixator construct and it may be necessary to optimize MR imaging parameters, to compensate for the presence of the fixation frame. Representative devices used to assemble a typical Synthes Low-Profile Wrist Fixator frame have been evaluated in the MRI chamber and worst case artifact information is provided below. Overall, artifacts created by Synthes Low-Profile Wrist Fixator devices may present issues if the MR imaging area of interest is in or near the area where the fixation frame is located. For FFE sequence: Scan duration: 3 min, TR 100 ms, TE 15 ms, flip angle 15º, and SE sequence: Scan duration: 4 min, TR 500 ms, TE 20 ms, flip angle 70º radio echo sequence, worst-case artifact will extend approximately 10 cm from the device. Warning Do not place any radio frequency (RF) transmit coils over the external fixation frame. 2 Synthes Low-Profile Wrist Fixator Technique Guide

Low-Profile Wrist Fixator Technique Overview 1 Insert distal Schanz screws To avoid entrapping the extensor mechanism in extension, flex the second metacarpophalangeal joint to 90. Make a 25 mm longitudinal incision over the radial aspect and dissect the soft tissue. Using the parallel drill guide, insert 4.0 mm/3.0 mm selfdrilling Schanz screws or 4.0 mm/2.5 mm Schanz screws into the second metacarpal. Placement should be in the proximal and distal diaphyseal bone, 40 60 to the frontal plane. Note: The basic self-drilling Schanz screw insertion technique requires the tip to be embedded in the far cortex to resist cantilever forces. It is not necessary for the tip to penetrate through the far cortex. However, the screw may be inserted so that it protrudes slightly through the far cortex in some cases such as osteopenic bone. 2 Insert proximal Schanz screws Repeat Step 1 in the distal radius, taking care to avoid the sensory branch of the radial nerve. Synthes 3

Low-Profile Wrist Fixator Technique Overview 3 Apply frame With the clamps loosened, place the assembled frame over the Schanz screws. The clamps should always be positioned with the carbon fiber rod lying on the ulnar side of the Schanz screws to allow easy adjustment and clearance for the thumb. 4 Secure clamps on Schanz screws Tighten each of the Schanz screw locking screws using the offset wrench. 4 Synthes Low-Profile Wrist Fixator Technique Guide

5 Reduce fracture Reduce the fracture and tighten the remaining adjustment points. 6 Perform secondary adjustments Supination/pronation may be adjusted by loosening a clamp on the carbon fiber rod. Flexion/extension, and radial/ulnar deviation may be adjusted by loosening the ball joint screw. Synthes 5

Components Included in Sterile Package Fixation Material, MR Conditional 294.768 4.0 mm/2.5 mm Schanz Screw or 294.771 4.0 mm/ 3.0 mm Self-Drilling Schanz Screw Low-Profile Wrist Fixator Clamp 6.0 mm Carbon Fiber Rod Low Profile Wrist Fixator Instruments, MR Unsafe* 321.263 Offset Wrench, 3.5 mm hex Disposable Parallel Drill Guide General Instruments 395.769 Protective Cap, for 6.0 mm Carbon Fiber Rods *MR Unsafe: An item that is known to pose hazards in all MR environments 6 Synthes Low-Profile Wrist Fixator Technique Guide

Product Information 03.304.220S Low-Profile Wrist Fixator, with 4.0 mm/2.5 mm Schanz screws and 200 mm carbon fiber rod, sterile 03.304.222S Low-Profile Wrist Fixator, with 4.0 mm/2.5 mm Schanz screws and 220 mm carbon fiber rod, sterile 03.304.320S 03.304.322S Low-Profile Wrist Fixator, with 4.0 mm/ 3.0 mm self-drilling Schanz screws and 200 mm carbon fiber rod, sterile Low-Profile Wrist Fixator, with 4.0 mm/ 3.0 mm self-drilling Schanz screws and 220 mm carbon fiber rod, sterile 03.304.320S Note: For additional information, please refer to package insert. For detailed cleaning and sterilization instructions, please refer to http://us.synthes.com/medical+community/cleaning+and+sterilization.htm or to the below listed inserts, which will be included in the shipping container: Processing Synthes Reusable Medical Devices Instruments, Instrument Trays and Graphic Cases DJ1305 Processing Non-sterile Synthes Implants DJ1304 Synthes 7

Synthes (USA) 1302 Wrights Lane East West Chester, PA 19380 Telephone: (610) 719-5000 To order: (800) 523-0322 Fax: (610) 251-9056 Synthes (Canada) Ltd. 2566 Meadowpine Boulevard Mississauga, Ontario L5N 6P9 Telephone: (905) 567-0440 To order: (800) 668-1119 Fax: (905) 567-3185 www.synthes.com 2005 Synthes, Inc. or its affiliates. All rights reserved. Synthes is a trademark of Synthes, Inc. or its affiliates. Printed in U.S.A. 5/11 J6207-E