INSTRUCTIONS FOR THE DENTIST. synocta impression procedure, Screw-retained and Snap-on. Screw-retained (Open tray) Snap-on (Closed tray)

Similar documents
CROWN AND BRIDGE RESTORATIONS WITH THE SYNOCTA PROSTHETIC SYSTEM

Straumann synocta Prosthetic System Basic Information. Technical Information

synocta Meso abutment for cement-retained restorations PROSTHETICS Step-by-step instructions

ß For cement-retained crowns and bridges. Note: synocta cementable abutments. ß Can be especially helpful with large multi-unit cases Note:

STEP-BY-STEP INSTRUCTIONS ON TEMPORARY ABUTMENTS. Straumann Temporary Abutments, VITA CAD-Temp

CEMENT-RETAINED CROWNS AND BRIDGES WITH THE SOLID ABUTMENT SYSTEM

IMPRESSION-TAKING, BITE REGISTRATION, AND TEMPORARY RESTORATION ON CAMLOG IMPLANTS. a perfect fit

PROSTHETIC OPTIONS FOR NARROW NECK IMPLANTS

INSTRUCTIONS FOR USE STRAUMANN VARIOBASE ABUTMENT. Straumann Variobase Abutment

TITANIUM BASES CAD/CAM FOR CROWN AND BRIDGE RESTORATIONS CAD/CAM PROSTHETICS ON CAMLOG AND CONELOG IMPLANTS

OT EQUATOR CASTABLE Single Attachment for Overdentures

SS Implant System 2013 PROSTHETIC PROCEDURE

OCCLUSALLY SCREW-RETAINED

BASIC INFORMATION ON THE prosthetic procedures. Straumann Bone Level Implant Line

The LOCATOR concept. Simplicity and versatility for prosthesis fixation

NARROW DIAMETER implant

Straight Octagon Abutment for Tissue Level Implant

Blue Sky Bio Multi unit Abutments BIO Multi-Units are available for all our Bone Level Implants

BASIC INFORMATION ON THE prosthetic procedures. Straumann Bone Level

NEW. The SmartFix Concept Prosthetic solution on tilted implants. SmartFix

Very small abutment head easy and secure handling. Ankylos. The SmartFix concept. Prosthetic solution on angled implants

User Manual Complete Prosthetic Kit (CPK)

SimpleLine II Surgical / Prosthesis Manual

IMPRESSION PROCEDURES TRAINING MANUAL

cement-retained single crowns using cementable abutments

SwissPlant Prosthetic Catalog. SwissPlant

For the oral rehabilitation

Part 2: Internal Tri-Channel Connection. Nobel Biocare Compatible Components. Tri-Channel Internal Connection Implants

Prosthetic Guideline. Dental Implant System Tissue Level

SIC invent. SIC P2F Abutment Dr. Galip Gurel From the gingiva shaper to a provisional crown to the final aesthetic restoration

N-SERIES. compatible to: Straumann / SynOcta 32 NT-TRADING CATALOGUE PLATFORM 3,5 mm NN 4,8 mm RN 6,5 mm WN TITANIUM BASE SCAN BODY 3D-GUIDE

INTRODUCTION. Why Choose Cortex Prosthetics?

B&B DENTAL. implant company PROSTHETIC PROCEDURES BASIC INFO DURAVIT IMPLANT LINES 3P - EVOLUTION - WIDE

Restoring Straumann implants. with Locator abutments.

Hex-Lock Abutment System. Restorative Manual

Abutment level impression Ⅰ. Combi abutment/dual abutment Multi units

Step-by-Step Cemented Bridge Using Using CPK Abutments. Internal Hex. Implant System

Implant Phase Flow Chart

US Implant System 2013 PROSTHETIC PROCEDURE

PROSTHETIC EFFICIENCY. Straumann Angled Solutions (AS) Flexibility and precision from a different angle

Product Catalog 2017/2018. Straumann International Edition

Straumann Screw-retained Prosthetics

Straumann CARES Implant-borne prosthetics Basic Information. Technical Information

Legacy Prosthetic Catalog

Anchorage system. tomas / Sets Page 270 tomas / Pins Page 271 Instruments and accessories Page 273 Patient consultation material Page 282

Type 1. Type 2. Straumann Tissue Level. The N-Series abutments are compatible with» N-SERIES «

ESTHETIC LINE implant

Quick start Prosthetics

More than Abutments. catalogue

INDEX. Pag. 12. Pag. 05. Pag. 06. Pag. 13. Pag. 06. Pag. 14. Pag. 15. Pag. 15. Pag. 15. Pag. 15. Pag. 08. Pag. 16. Pag. 09. Pag.

Description: Prosthetic Components for dental restoration under CE-Series, CL- Series, CN-Series, CH-Series, CF-Series, CR-Series, CS-Series.

Ankylos. Acuris conometric concept. Manual and product catalog

PROSTHETIC PROCEDURES

UDELL DENTAL LABORATORY Instructions for Use PREAT Precision Attachments

Prosthetic procedures for the. Straumann Narrow Neck CrossFit Implant Line

Straumann Dental Implant System. Fax Order Form Collection.*

Multi-Unit Abutment System SIC Safe on Four. Optimum use of available bone by angled placement of implants

Type 1. Type 2. Straumann Bone Level. The L-Series abutments are compatible with» L-SERIES «

Leading material offering and application range. Straumann CARES CADCAM

PRODUCT INFORMATION FOR CORTEX ABUTMENTS AND OTHER RESTORATIVE COMPONENTS

TRINON. Index. Index. Overview Page 4. Catalogue

Guidance for. implant removal. Straumann Dental Implant System

Osteotome inserts kit, angled. Impression tube (open tray) Selection abutment kit / Selection measurement post. Bar base PMMA & Finisher

Stock Prosthetic Guide 2014

prosthetic technique manual

Straumann Dental System Indication-based prosthetic workflow. Technical Information

Ankylos. Product Catalog

PRODUCT CATALOG 2014 CONELOG IMPLANTATSYSTEM PRODUCT PRODUKTKATALOG CATALOG 2014

4766 Research Dr. San Antonio, TX insightdentalsystems.com

isy indications and techniques

Bridging the Gap Between Original and Multi-Platform Systems.

SwishPlant. SwishPlant System Catalog. 4.8mmD mmD. 4.8mmD mmD. 3.7mmD mmD. 6.5mmD

T-SERIES. compatible to: DENTSPLY Implants / Xive. PLATFORM 3,4 mm 3,8 mm 4,5 mm 5,5 mm TITANIUM BASE SCAN BODY 3D-GUIDE LAB ANALOG IMPLANT PICK-UP

Height* FC width Prep depth RC width 4.0mm+ 4.3mm N/A 6.3mm

R-SERIES. compatible to: Zimmer Dental / Tapered Screw-Vent 36 NT-TRADING CATALOGUE PLATFORM 3,5 mm 4,5 mm 5,7 mm TITANIUM BASE

STRAUMANN DENTAL IMPLANT SYSTEM RESToRATIVE SELECTION GUIDE

SuperLine & Implantium Surgical / Prosthesis Manual

BIO-SERIES. compatible to: BioHorizons / internal. PLATFORM 3,0 mm 3,5 mm 4,5 mm 5,7 mm TITANIUM BASE SCAN BODY 3D-GUIDE NT-MILL LAB ANALOG

cement-retained crown using the Laser-Lok Easy Ti temp abutment and PEEK plastic sleeves

for individual milled Zirconium Abutment incl. screws Titanium Grade 5 ELI CAM CAM CAM CAM

Titanium Base abutments plus screw and associated restorative and dental laboratory components.

» For multi-unit restorations -

PRODUCT CATALOG. Since 1985» Simple. Predictable. Profitable.

SuperLine & Implantium Surgical / Prosthesis Manual

Smarter Thinking. Simpler Design. TiLobe Prosthetic Manual for PrimaConnex & Genesis implants

S-SERIES. compatible to: DENTSPLY Implants / Astra Tech Implant System OsseoSpeed TX. PLATFORM 3,5 / 4,0 mm 4,5 / 5,0 mm TITANIUM BASE

E-SERIES. compatible to: Nobel Biocare / Replace Select. PLATFORM 3,5 mm NP 4,3 mm RP 5,0 mm WP 6,0 mm TITANIUM BASE SCAN BODY 3D-GUIDE LAB ANALOG

Personal Attention Accurate Dental Solutions. Precision Down to the Smallest Detail. Wide Range of Dental Solutions Engineerd for a Lifetime

BEG-SERIES. compatible to: BEGO Implant Systems / Semados S, RS-RX, RI-Line 12 NT-TRADING CATALOGUE

_ Instructions for multi-unit screw-retained restorations

Xive. Product Catalog

LOCATOR ROOT ATTACHMENT SYSTEM

ANKYLOS Planning Components 100. ANKYLOS Implants 102. ANKYLOS Surgical Kits 103. ANKYLOS Surgical Unit, Handpieces 106

NobelActive. Quick start page 5. Surgical procedures page 8. Prosthetic procedures page 28. Appendix IV cleaning and. sterilization page 66

The Multi-unit Abutments of the CE-Series are indicated for Nobel Biocare / NobelReplace Tapered. The current implant diameters have to be attended.

Basic Information on Straumann Guided Surgery. Straumann Dental Implant System

TS Implant System 2013 PROSTHETIC PROCEDURE

Guided surgery manual

ICX-PROSTHETICS. The future of dental implantology MANUAL 5.0. The FAIR Implant-System. customer service: +49 (0)

Transcription:

INSTRUCTIONS FOR THE DENTIST synocta impression procedure, Screw-retained and Snap-on Screw-retained (Open tray) Snap-on (Closed tray) click 1. Attach synocta Impression caps to implants. Hand-tighten screws. 2. Take the impression with a special tray with perforations for screws and elastomer impression matieral (vinyl polysiloxane or polyether rubber). 1. Attach impression caps to the implants. The cap has engaged when a click is heard. If the cap is correctly seated, it can be turned on the implant. 2. The octagon of the synocta Positioning cylinder must be aligned with the octagon of the implant and inserted into the impression cap up to the limit stop. 3. Undo the screws and remove the impression. 4. Sending the impression to the dental lab. 3. Take the impression with elastomer impression materials (vinyl polysiloxane or polyether rubber). 4. Sending the impression to the dental lab. 5. Screw the synocta Analog to the impression caps hand-tight. 5. Replace synocta Analogs into the impression. The shoulder must engage audibly. Make the working model from special hard plaster, Type 4 (DIN 13911). Note: The impression-taking method is identical for implant shoulder Ø 4.8 mm RN and implant shoulder Ø 6.5 mm WN.

Art. No. Article Dimensions Material Components for synocta impressions 048.017V4 RN impression cap, snap-on height 8.0 mm plastic 048.070V4 RN synocta positioning cylinder, red, snap-on height 12.0 mm plastic 048.010 RN synocta impression cap, screw-retained, red, with integral guide screw height 10.1 mm anodized aluminium/titanium 048.090 RN synocta impression cap, built-in handle, red, with integral guide screw height 21.0 mm anodized aluminium/titanium 048.013 WN impression cap, snap-on height 8.0 mm plastic 048.095 WN synocta positioning cylinder, white, snap-on height 12.0 mm plastic 048.091 WN synocta impression cap, screw-retained, with integral guide screw height 10.0 mm aluminium/titanium synocta Master cast fabrication 048.124 RN synocta analog, gray (with red stripe) length 12.0 mm stainless steel 048.171 WN synocta analog, gray length 12.0 mm stainless steel Prosthetic Instruments 046.400 SCS Screwdriver, extra short length 15.0 mm stainless steel 046.401 SCS Screwdriver, short length 21.0 mm stainless steel 046.402 SCS Screwdriver, long length 27.0 mm stainless steel 046.410 SCS Screwdriver, extra short, for handpiece adapter length 20.0 mm stainless steel 046.411 SCS Screwdriver, short, for handpiece adapter length 26.0 mm stainless steel 046.412 SCS Screwdriver, long, for handpiece adapter length 32.0 mm stainless steel Caution: The plastic parts are intended for single use only. They must not be sterilized. To avoid damage (loss of elasticity, embrittlement) to the plastic parts, they must be protected from exposure to strong light or heat. For detailed information, refer to our brochure PROSTHETICS, Crown and Bridge Restorations with the synocta Prosthetic System, Art. No. 152.255. RN = Regular Neck WN = Wide Neck 11/06 152.259/e B21106

INSTRUCTIONS FOR THE DENTIST RN synocta 1.5 Screw retained Placement of two RN synocta 1.5 Screw retained abutments in implants After fabrication, the restoration is given to the dentist on the master cast together with the abutments. Note: The method of placing the RN and WN synocta 1.5 Screw-retained abutments is identical. 1. Remove the superstructure and the abutments from the master cast with an SCS screwdriver. 2. Insert the abutments into the implants and tighten the basal screws to a torque of 35 Ncm. 3. Two methods for screw retention of the superstructure: A. Screw retention with SCS occlusal screw (Art. No. 048.350). In this method, the screw head is covered with some wax or guttapercha and subsequently the transocclusal screw channel is sealed off (e.g. with composite). Tightening torque = 15 Ncm! B. Screw retention with SCS guide screw (Art. No. 048.360/361/ 362/363/364). In this method, the SCS guide screw is shortened in the mouth to the level of the occlusal plane. Tightening torque = 15 Ncm!

Important: For tightening the basal screw, the ratchet (Art. No. 046.119) with attached torque control device (Art. No. 046.049) and an SCS screwdriver (Art. No. 046.400/401/402) are required. Important: The abutment must be positioned in the octagon before the screw is tightened. The basal screw in the synocta 1.5 Screw-retained abutment is tightened to a torque of 35 Ncm. The SCS occlusal and guide screws are tightened to a torque of 15 Ncm. Art. No. Article Dimensions Material SCS occlusal and guide screws (according to the type of restoration) 048.350 048.350V4 SCS Occlusal screw length 4.4 mm titanium 048.360V4 Guide screw for RN synocta coping bar, cannot be shortened length 6.0 mm titanium 048.361V4 Guide screw for RN synocta coping bar, can be shortened by 1.6 mm length 6.0 mm titanium 048.362V4 SCS Guide screw for milling cylinder, cannot be shortened length 8.0 mm titanium 048.363V4 Guide screw for milling cylinder, can be shortened by 2.0 mm length 8.0 mm titanium 048.364V4 Guide screw, can be shortened by 2.0 mm length 10.0 mm titanium Prosthetic instruments SCS screwdriver (extra short 046.400, short 046.401, long 046.402) Ratchet, incl. service instrument 046.119 Torque control device for ratchet 046.049 Holding key 046.064 For detailed information, refer to our brochure PROSTHETICS, Crown and Bridge Restorations with the synocta Prosthetic System, Art. No. 152.255. RN = Regular Neck WN = Wide Neck 11/06 152.259/e B21106

INSTRUCTIONS FOR THE DENTIST RN synocta gold abutment RN synocta gold abutment placement of the final restoration After fabrication, the restoration is given to the dentist on the master cast. 1. Align the octagon of the customized gold abutment on the octagon of the implant and insert the mesostructure into the implant. Tighten the basal screw with a force of 35 Ncm. 2. Then seal off the screw channel with gutta-percha, and finally cement the crown onto the mesostructure.

Important: For tightening the basal screw, the ratchet (Art. No. 046.119) with attached torque control device (Art. No. 046.049) and an SCS screwdriver (Art. No. 046.400/401/402) are required Important: The abutment must be positioned in the octagon before the screw is tightened. The basal screw in the RN synocta gold abutment is tightened to a torque of 35 Ncm. Prosthetic instruments SCS screwdriver (extra short 046.400, short 046.401, long 046.402) Ratchet, incl. service instrument 046.119 Torque control device for ratchet 046.049 Holding key 046.064 For detailed information, refer to our brochure PROSTHETICS, Crown and Bridge Restorations with the synocta Prosthetic System, Art. No.152.255. RN = Regular Neck 11/06 152.259/e B21106

INSTRUCTIONS FOR THE DENTIST RN and WN synocta Cement retained Placement of two RN synocta Cement retained abutments in implants After fabrication, the restoration is given to the dentist on the master cast together with the abutments. Note: The method of incorporating the RN and WN synocta Cement-retained abutments is identical. 1. Undo the basal screws with the SCS screwdriver and transfer the index, with the abutments, from the master cast to the implants. 2. Tighten the basal screws to a torque of 35 Ncm. Note: In order to be able to release the basal screws again if required, fill the screw head and the abutment with gutta-percha. Then cement the superstructure permanently.

Important: For tightening the basal screw, the ratchet (Art. No. 046.119) with attached torque control device (Art. No. 046.049) and an SCS screwdriver (Art. No. 046.400/401/402) are required. Important: The abutment must be positioned in the octagon before the screw is tightened. Tighten the basal screw in the synocta Cement retained abutment to a torque of 35 Ncm. Art. No. Article Dimensions Material Transfer of the synocta Cement retained abutments 048.059V4 Transfer aid for 048.605, RN height 6.5 mm plastic 048.054V4 Transfer aid for 048.606, WN height 6.5 mm plastic Prosthetic instruments SCS screwdriver (extra short 046.400, short 046.401, long 046.402) Ratchet, incl. service instrument 046.119 Torque control device for ratchet 046.049 Holding key 046.064 Caution: The plastic parts are intended for single use only. They must not be sterilized. To avoid damage (loss of elasticity, embrittlement) to the plastic parts, they must be protected from exposure to strong light or heat. For detailed information, refer to our brochure PROSTHETICS, Crown and Bridge Restorations with the synocta Prosthetic System, Art. No. 152.255. RN = Regular Neck WN = Wide Neck 11/06 152.259/e B21106

INSTRUCTIONS FOR THE DENTIST RN and WN synocta Angled Placement of an RN synocta Angled abutment into the implant After fabrication, the restoration is given to the dentist on the master cast together with the abutments. 1. Undo the basal screw with the SCS screwdriver and remove the index from the master cast. 2. Insert the abutment into the implant and tighten the basal screw to a torque of 35 Ncm. 3. Undo the occlusal screw and remove the index. Then cement or screw in the restoration. Note: If the restoration is cemented, the lateral and occlusal opening must be sealed off with gutta-percha. Placement of a WN synocta Angled abutment into the implant 1. Undo the basal screw with the SCS screwdriver and remove the index from the master cast. 2. Insert the abutment into the implant and tighten the basal screw to a torque of 35 Ncm. 3. Remove the index, then cement the restoration. Note: Before cementing the restoration, the lateral opening must be sealed off with gutta-percha.

Important: For tightening the basal screw, the ratchet (Art. No. 046.119) with attached torque control device (Art. No. 046.049) and an SCS screwdriver (Art. No. 046.400/401/402) are required. Important: The abutment must be positioned in the octagon before the screw is tightened. The basal screw in the RN and WN synocta Angled abutment is tightened to a torque of 35 Ncm. Important: If a screw-retained restoration is used, the SCS occlusal screw must be tightened to a torque of 15 Ncm. Art. No. Article Dimensions Material Transfer of the RN and WN synocta Angled abutments 048.000V4 Transfer aid for RN synocta Angled, short, 15 and 20 height 4.0 mm plastic 048.002V4 Transfer aid for RN synocta Angled, long, 15 and 20 height 4.0 mm plastic 048.032 Transfer aid for WN synocta Angled, 15 height 5.0 mm plastic 048.350 048.350V4 SCS occlusal screw length 4.4 mm titanium Prosthetic instruments SCS screwdriver (extra short 046.400, short 046.401, long 046.402) Ratchet, incl. service instrument 046.119 Torque control device for ratchet 046.049 Holding key 046.064 Caution: The plastic parts are intended for single use only. They must not be sterilized. To avoid damage (loss of elasticity, embrittlement) to the plastic parts, they must be protected from exposure to strong light or heat. For detailed information, refer to our brochure PROSTHETICS, Crown and Bridge Restorations with the synocta Prosthetic System, Art. No.152.255. RN = Regular Neck WN = Wide Neck 11/06 152.259/e B21106

INSTRUCTIONS FOR THE DENTIST RN synocta Transversal Placement of an RN synocta Transversal (TS) abutment into the implant After fabrication, the restoration is given to the dentist on the master cast together with the abutments. 1. Undo the basal screw with the SCS screwdriver and remove the index from the master cast. 2. Insert the abutment into the implant and tighten the basal screw to a torque of 35 Ncm. Then incorporate the superstructure. 3. Attach the restoration with the transversal screw, and carefully hand-tighten the transversal screw with the TS hexagonal screwdriver.

Important: For tightening the basal screw, the ratchet (Art. No. 046.119) with attached torque control device (Art. No. 046.049) and an SCS screwdriver (Art. No. 046.400/401/402) are required. Important: The abutment must be positioned in the octagon before the screw is tightened. Tighten the basal screw in the RN synocta Transversal (TS) abutment to a torque of 35 Ncm. Important: The transversal screw must only be hand-tightened with the TS screwdriver (Art. No. 046.420). Art. No. Article Dimensions Material Transfer of the RN synocta Transversal (TS) abutment 048.003V4 Transfer aid for RN synocta TS abutment height 5.0 mm plastic Prosthetic instruments SCS screwdriver (extra short 046.400, short 046.401, long 046.402) Ratchet, incl. service instrument 046.119 Torque control device for ratchet 046.049 Holding key 046.064 TS hexagonal screwdriver 046.420 Caution: The plastic parts are intended for single use only. They must not be sterilized. To avoid damage (loss of elasticity, embrittlement) to the plastic parts, they must be protected from exposure to strong light or heat. For detailed information, refer to our brochure PROSTHETICS, Crown and Bridge Restorations with the synocta Prosthetic System, Art. No. 152.255. RN = Regular Neck 11/06 152.259/e B21106

INSTRUCTIONS FOR THE DENTIST Screwing in an RN solid abutment with subsequent impression taking and temporary restoration Screwing in the abutment 1. Insert the RN solid abutment into the driver and screw it into the implant. Note: WN solid abutments must be screwed into the WN implant using the SCS screwdriver. 2. Position the ratchet with attached torque control device and stabilize with the holding key. 3. Tighten the abutment with a force of 35 Ncm and remove the instruments again. Taking the impression 1. Attach impression cap to implant (make sure it clicks into place). If it is correctly seated, the impression cap is easy to turn on the implant. 2. Insert the positioning cylinder into the impression cap. Check that it is seated without any gap. 3. Take the impression with elastomer impression material (vinyl polysiloxane or polyether rubber). Impression, e.g. for RN (gray, height 5.5 mm) and WN (green, height 4.0 mm) following removal from the mouth. Temporary restoration Option A: Temporary restoration with temporary copings (plastic). Temporary copings can be individually shortened and coated with plastic on the master cast or intraorally by the conventional technique. Option B: Temporary restoration with protective caps (PEEK). Protective caps are cemented on solid abutments with temporary cement. Note: The method of impression taking and the temporary restoration are identical for RN and WN solid abutments.

Art. no. Article Dimensions Material RN and WN solid abutments 048.540 RN Solid abutment, 6, yellow height 4.0 mm titanium 048.541 RN Solid abutment, 6, gray height 5.5 mm titanium 048.542 RN Solid abutment, 6, blue height 7.0 mm titanium 048.545 WN Solid abutment, 6, green height 4.0 mm titanium 048.546 WN Solid abutment, 6, brown height 5.5 mm titanium Color-coded components for RN and WN impressions 048.017V4 RN impression cap height 8.0 mm plastic 048.060V4 Positioning cylinder for 048.540, yellow height 10.2 mm plastic 048.061V4 Positioning cylinder for 048.541, gray height 10.2 mm plastic 048.062V4 Positioning cylinder for 048.542, blue height 10.2 mm plastic 048.013V4 WN Impression cap height 8.0 mm plastic 048.065V4 Positioning cylinder for 048.545, green height 10.0 mm plastic 048.066V4 Positioning cylinder for 048.546, brown height 10.0 mm plastic Components for RN and WN temporary restorations 048.654 RN Temporary coping for RN solid abutments, bridge height 8.5 mm plastic 048.655 RN Temporary coping for RN solid abutments, crown height 8.5 mm plastic 048.047V4 RN Protective cap, cemented, for 048.540 height 5.8 mm PEEK 048.048V4 RN Protective cap, cemented, for 048.541 height 7.3 mm PEEK 048.049V4 RN Protective cap, cemented, for 048.542 height 8.8 mm PEEK 048.656 WN Temporary coping for WN solid abutments, bridge height 7.3 mm plastic 048.657 WN Temporary coping for WN solid abutments, crown height 7.3 mm plastic 048.051 WN Protective cap, cemented, for 048.545 height 6.0 mm PEEK 048.052 WN Protective cap, cemented, for 048.546 height 7.5 mm PEEK Caution: The plastic parts are intended for single use only. They must not be sterilized. To avoid damage (loss of elasticity, embrittlement) to the plastic parts, they must be protected from exposure to strong light or heat. Inserting instruments for RN and WN Solid abutments For RN Solid abutments: solid abutment driver (short, 046.067; long, 046.068) For WN Solid abutments: SCS screwdriver (extra short 046.400, short 046.401, long 046.402 Ratchet 046.119 Torque control device 046.049 Holding key 046.064 For detailed information, refer to our brochures PROSTHETICS, fixed Crown and Bridge Restorations with the solid abutment system, Art. No. 152.254 and PROSTHETICS, temporary copings for solid abutments, Art. No. 152.282. RN = Regular Neck WN = Wide Neck 11/06 152.259/e B21106

INSTRUCTIONS FOR THE DENTAL TECHNICIAN Fabrication of a mesostructure with the RN synocta gold abutment and a cement-retained crown Fabrication of the mesostructure 1. In order to design the emergence profile optimally on the neck of the crown, a gingival mask should be made on the master cast. 2. Attach the RN synocta gold abutment to the analog, ensuring that the abutment is aligned in the octagon of the analog. Then handtighten the screw with an SCS screwdriver. 3. If required, the modeling aid can be shortened occlusally to suit the anatomical conditions. 4. Make a wax model of the mesostructure on the abutment. Minimum wax thickness 0.7 mm. 5. Check for correct spacing with the silicone index from the wax-up. 6. For reasons of hygiene, the mesostructure/crown cement gap must not be more than 2.0 mm below the gingiva. 7. Embed the mesostructure. Do not use a wetting agent! To prevent the cast-on alloy overflowing, it is essential to clean the slender edge of the coping and the internal configuration thoroughly with alcohol. 8. Cast the mesostructure in the conventional way. Do not use speed investment materials! Follow the instructions for use provided by the manufacturer of the investment material and cast-on alloy! 9. Carefully devest the casting using only ultrasound, water jet or pickling. Do not sandblast! 10. Finish the mesostructure. Polish the subgingival part.

Fabrication of the cement-retained crown 11. Block out the screw channel and model the crown. 12. Invest, cast and veneer the crown in the conventional way. 13. a The mesostructure with the finished crown. 13. b Art. No. Article Dimensions Material RN synocta gold abutment for transocclusal screw-retained or cement-retained crowns 048.642 RN synocta gold abutment for premounted modeling aid (including screw*) height 14.1 mm Ceramicor/burn-out plastic/titanium *also available as spare part Art. No. 048.356 For detailed information, refer to our brochure PROSTHETICS, Crown and Bridge Restorations with the synocta Prosthetic System, Art. No. 152.255. RN = Regular Neck 11/06 152.259/e B21106