Accuracy of Mass Spectrometry with the Ease of Immunoassay? LC-MS/MS Workflow improvements for Primary Aldosteronism Screening

Similar documents
Surviving Matrix Effects Experiments. Grace van der Gugten St. Paul s Hospital, Vancouver, BC, Canada

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

The Royal College of Pathologists Journal article evaluation questions

Vitamin D A game changer for Clinical Laboratories

Hypertension is a serious and often symptomless

Guideline PA. Renin. Renin From a Laboratory Perspective. Use of renin and aldosteron measurements. UMCG Dept. Laboratory Medicine

A High-throughput Mass Spectrometry Multiplexed Assay to Measure Insulin and C- peptide

Measurement of Renin Activity using Tandem Mass Spectrometry Ravinder J Singh, PhD, DABCC Mayo Clinic, Rochester, MN

High Throughput Extraction of Opiates from Urine and Analysis by GC/MS or LC/MS/MS)

Primary aldosteronism clinical practice guidelines: a re-appraisal The Management of Primary Aldosteronism

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No.

LC-MS/MS Method for the Determination of Tenofovir from Plasma

The analysis of Glucocorticoid Steroids in Plasma, Urine and Saliva by UPLC/MS/MS

Determination of Pharmaceutical Residues in Bovine Milk via LC MS/MS Following Solid Phase Extraction

A Comprehensive Screening of Illicit and Pain Management Drugs from Whole Blood Using SPE and LC/MS/MS

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Simultaneous Quantitative Analysis of Total Catecholamines and Metanephrines in Urine Using CLEAN UP CCX2 and LC-MS/MS

Comprehensive Study of SLE as a Sample. Preparation Tool for Bioanalysis

Clarification of hypertension Diagnosis of primary hyperaldosteronism

All stocks and calibration levels were prepared in water: methanol (50:50) v/v to cover range of all steroid concentrations (refer Table 1).

Automated Targeted screening of Benzodiazepines in urine using LDTD-MS/MS at 400 samples per hour rate

The Adiponectin Turbidimetric Immunoassay Reagent Kit

25 OH Vitamin D Rapid Test

Automated Chemiluminescence-Immunoassay for Aldosterone during Dynamic Testing: Comparison to Radioimmunoassays with and without Extraction Steps

Fast and simultaneous analysis of ethanol metabolites and barbiturates using the QTRAP 4500 LC-MS/MS system

Comparison of Biotage Extrahera vs. Manual Sample Processing Using a Vacuum Manifold

Simultaneous Quantitative Analysis of Total Catecholamines and Metanephrines in Urine Using 500 MG CLEAN UP CCX2 and LC-MS/MS

SCIEX Vitamin D 200M Assay for the Topaz System

Extraction of Synthetic and Naturally Occurring Cannabinoids in Urine Using SPE and LC-MS/MS

A RAPID AND SENSITIVE ANALYSIS METHOD OF SUDAN RED I, II, III & IV IN TOMATO SAUCE USING ULTRA PERFORMANCE LC MS/MS

Overcoming the Challenges of Sample Preparation and LC/MS/MS Method Development for Clinical Applications

Extraction of Multiple Mycotoxins From Nuts Using ISOLUTE Myco prior to LC-MS/MS Analysis

SWATH Acquisition Enables the Ultra-Fast and Accurate Determination of Novel Synthetic Opioids

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid

Fully automated sensitive determination of immunosuppressant drugs in whole blood, using high quality internal standardization

Improve the Hypertension Management

Diagnostic Role of Captopril Challenge Test in Korean Subjects with High Aldosterone-to-Renin Ratios

Updates in primary hyperaldosteronism and the rule

Determination of Tetracyclines in Chicken by Solid-Phase Extraction and High-Performance Liquid Chromatography

The Clinical Value of Salivary Aldosterone in Diagnosis and Follow-Up of Primary Aldosteronism

Extraction of Aflatoxin M1 From Infant Formula Using ISOLUTE Myco SPE Columns prior to LC-MS/MS Analysis

High-throughput lipidomic analysis of fatty acid derived eicosanoids and N-acylethanolamines

Updates in primary hyperaldosteronism and the rule

you-try-it-02.xlsx Step-by-Step Guide ver. 8/26/2009

Primary Aldosteronism: screening, diagnosis and therapy

Protein Precipitation for Biological Fluid Samples Using Agilent Captiva EMR Lipid 96-Well Plates

LONG-TERM EFFECTS OF SURGICAL MENAGEMENT OF PRIMARY ALDOSTERONISM ON THE CARDIOVASCULAR SISTEM

Integration of steroids analysis in serum using LC-MS/MS with full-automated sample preparation

Insulin (Porcine/Canine) ELISA

Simple Method (IS-MRM) to Monitor Lysophospholipids and Phospholipids During LC-MS Method Development via In-Source CID

Extraction of Multiple Mycotoxins From Grain Using ISOLUTE Myco prior to LC-MS/MS Analysis

Real-time PK Measurement of the Chemotherapeutic Drug Melphalan in Whole Blood by a Novel PaperSpray Mass Spectrometry

Confirmatory testing in normokalaemic primary aldosteronism: the value of the saline infusion test and urinary aldosterone metabolites

Upon completion, participants should be able to:

LDL (Human) ELISA Kit

Integrated Targeted Quantitation Method for Insulin and its Therapeutic Analogs

Fig. 1: Chemical structure of arachidonic acid COOH CH 3

Technical Information Guide

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

Dienes Derivatization MaxSpec Kit

HbA1c (Human) ELISA Kit

Supporting Information (SI)

Manual. Transfer step. Figure 1. Workflow comparison of manual phenylalanine assay and automated GSP Neonatal Phenylalanine assay

The Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis

Detection of Low Level of Chloramphenicol in Milk and Honey with MIP SPE and LC-MS-MS

Nutrition: obtaining reliable biomarker data to study the health status of populations

Improving Benzodiazepine Immunoassay Sensitivity by Rapid Glucuronide Hydrolysis Technology

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

Development and Validation of a Polysorbate 20 Assay in a Therapeutic Antibody Formulation by RP-HPLC and Charged Aerosol Detector (CAD)

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

Michael Blackburn HOW LOW CAN YOU GO: DRIVING DOWN LIMITS OF HYBRID IA-LC/MS. European Bioanalytical Forum. 19 th November 2015

Determination of Ractopamine and Clenbuterol in Beef Samples

Pelagia Research Library

Patients with primary aldosteronism (PA) are at a higher

Supporting Information

Keywords: albuminuria; albumin/creatinine ratio (ACR); measurements. Introduction

CASTLE HOWARD EVENT 2018

Primary Aldosteronism

Determination of Amantadine Residues in Chicken by LCMS-8040

25 OH Vitamin D Rapid test

Shuguang Li, Jason Anspach, Sky Countryman, and Erica Pike Phenomenex, Inc., 411 Madrid Ave., Torrance, CA USA PO _W

A Definitive Lipidomics Workflow for Human Plasma Utilizing Off-line Enrichment and Class Specific Separation of Phospholipids

Identification and Quantification of Psychedelic Phenethylamine 2C Series using SPE and LC-MS/MS

Performance Characteristics of Eight Estradiol Immunoassays

Application Note LCMS-108 Quantitation of benzodiazepines and Z-drugs in serum with the EVOQ TM LC triple quadrupole mass spectrometer

Mouse Adiponectin Immunoassay Kit

Clinical Chemistry. Your article (tracking # ) from Clinical Chemistry is available for download

A Simple and Accurate Method for the Rapid Quantitation of Drugs of Abuse in Urine Using Liquid Chromatography

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

Accuracy and Precision. Intra- and inter-assay accuracy and precision for both rifapentine

Comprehensive Forensic Toxicology Screening in Serum using On-Line SPE LC-MS/MS

Application Note. Author. Abstract. Introduction. Food Safety

Roles of Clinical Criteria, Computed Tomography Scan, and Adrenal Vein Sampling in Differential Diagnosis of Primary Aldosteronism Subtypes

A Prospective Study of the Prevalence of Primary Aldosteronism in 1,125 Hypertensive Patients

Renin Activity ELISA

Maria João Mendes Mendonça Barreira Department of Health Promotion and Chronic Deseases, INSA

DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS

Transcription:

Accuracy of Mass Spectrometry with the Ease of Immunoassay? LC-MS/MS Workflow improvements for Primary Aldosteronism Screening J Grace Van Der Gugten 1 and Daniel T. Holmes 1 1 Department of Pathology and Laboratory Medicine, St. Paul s Hospital and the University of British Columbia, 1081 Burrard St, Vancouver, BC, V6Z 1Y6, Canada 1 Background Primary aldosteronism (PA) is the most common form of secondary hypertension and is responsible for 10% of hypertension in allcomers [1]. The two most common forms of PA are idiopathic adrenal hyperplasia (IAH) and unilateral aldosterone producing adenoma (APA). IAH is amenable to treatment with aldosterone antagonists (spironolactone, eplelrenone) and unilateral adrenalectomy for APA often leads to complete cure of PA. In addition to the all of the expected sequelae of hypertension (increased risk of: atherosclerosis, stroke, chronic renal impairment, coronary heart disease), patients with PA have risk of cardiac [2] and renal disease [3] that is out of proportion to their hypertension, indicating the specific pathogenetic role that aldosterone plays in PA. For this reason, identification and treatment of PA is important from the standpoint of public health. The accepted manner of screening for primary aldosteronism is the collection of a concomitant plasma aldosterone and renin activity and to calculate their ratio. However, aldosterone is a difficult analyte owing to its low concentration and the presence of endogenous interferences [4]. Plasma renin is also technically challenging [5] and can be assessed either by the plasma renin activity (PRA) assay through the quantitation of angiotensin I (AngI), or by chemiluminescent sandwich assays directed at the renin molecule. However, the latter approach may not have significant correlation with PRA in the low renin state [6]. Both aldosterone and plasma renin activity determination have improved substantially from LC- MS/MS. For aldosterone, liquid-liquid [7, 8, 9] and supported liquid extraction methods [10, 11, 1

12] for aldosterone show much better intermethod comparison between one another [10, 12] than the comparison between IA methods (which may show 100 % bias) [13, 14] or the comparison between LC-MS/MS and IA [9, 12]. For plasma renin activity, both online and offline solid phase extraction (SPE) approaches have been developed and have been implemented into routine care [15, 16, 17]. These methods have better analytical sensitivity and specificity meaning that blanksubtraction is unnecessary [16, 18]. Despite the analytical advantages of LC-MS/MS determination of both aldosterone and renin over immunoassay, the attraction of walk-away automation using homogenous IA analyzers often wins out in the pragmatic and financially driven laboratory environments in which we exist. It is therefore our mandate to make the quality LC-MS/MS accessible by improving workflows. We present two proposed workflow improvements. First, we have developed a method for plasma renin activity and plasma aldosterone from the same sample preparation by sequential elutions off the same SPE plate and using mobile phase adjustments that have significantly enhanced aldosterone signal. Second, we present an exploration of using aldosterone:angi as an alternative screening tool to aldosterone:pra which could solve a number of standardization and technical problems with PA screening. 2 Methods 2.1 Aldosterone and PRA on the Same Sample Preparation PRA was performed by LC-MS/MS as previously described [19] but without blank subtraction as we have demonstrated it to be unnecessary. In brief, AngI is generated from endogenous renin and angiotensinogen using 250 μl of patient sample buffered with 50 μl of 1M tris acetate containing 0.2M EDTA, 1 mm PMSF, and 109 nmol/l d7-aldosterone at ph=5.5. The final ph of the buffered sample is 6. This buffered plasma is incubated for 3 h at 37 C and the reaction is quenched with the addition of 10% formic acid containing stable isotopically labelled AngI. Aliquots of the generate are added to a Phenomenex Strata-X (Part No. 8E-S100-UGB) conditioned SPE plate, washed and then eluted with 0.25 ml of 100% methanol. Eluted material is analyzed by LC-MS/MS as 2

described [19]. Following the AngI elution, the SPE plate undergoes an additional elution with 2 700 μl of methyltertbutylether (MTBE) into a 96-well deepwell plate. The MTBE is then dried down under house air at ambient temperature in a 96-well dry-down manifold, reconstituted with 125 μl of 20:80 MeOH:H 2 O analyzed as previously reported [10] but with lower concentration of ammonium acetate in mobile phases (0.25 nm). Comparisons of PRA and aldosterone was performed against our current production methods (n=36 for PRA and n=61 for Aldosterone). Aldosterone analysis using the presented method was performed on EDTA plasma while the production method was performed on concomitantly collected EDTA plasma. 2.2 Statistical evaluation of aldosterone:angi ratios In 665 sequential specimens, one aliquot buffered AngI solution was immediately acidified and analyzed without 37 C incubation while the second aliquot was generated as per normal. PRA was compared statistically with [AngI] and test characteristics of the aldosterone:angi ratio (AAR) was compared using aldosterone:pra (ARR) as the surrogate for a gold-standard. Calculations were performed using the ROCR package of the R statistical programming language. 3 Results 3.1 Aldosterone and PRA same sample preparation Aldosterone by second elution from SPE on EDTA plasma compared very well to the current SLE method, performed on concomitantly collected red top serum. Regression equation was [Aldo] SPE = 0.99[Aldo] SLE + 10.1 pmol/l 3

weighted Deming Regression, R 2 = 0.98, CI slope = [0.95, 1.02] and CI intercept = [7.2, 14.5] pmol/l, median difference = +7.5 pmol/l. Comparison is shown in figure 1. Figure 1: Regression comparison and difference plot between current production aldosterone method used for PA screening at St. Paul s Hospital vs exploratory method by second elution from the PRA SPE plate. Samples on the x-axis were performed on red top serum; those on the y-axis were performed on concomitantly collected EDTA plasma. PRA as generated in the presence of the aldosterone IS showed a regression comparison of P RA modified = 0.94 P RA current + 0.026 ng/ml/h. Slope and intercept did not differ statistically from 1 and 0 respectively, R 2 = 0.994. 3.2 Statistical evaluation of aldosterone:angi ratios Comparison between [AngI] and PRA showed a relationship of [AngI] = 0.04 P RA 0.01 ng/ml The correlation was strongly significant R 2 = 0.73, p < 10 10, for but demonstrated a large degree of scatter for the 620 specimens for which P RA < 10 ng/ml/h (figure 2). In the low renin speci- 4

mens (P RA < 1 ng/ml/h) wherein all PA is identified, this correlation deteriorated to R2 = 0.29 but remained highly significant p < 10 10. A large number (n=217) of the specimens had AngI concentrations below the limit of detection of the assay, in contrast to the 112 specimens that would have been reported as < 0.18 ng/ml/h by the PRA assay. Figure 2: Comparisons of [AngI] and PRA in specimens between 0-10 ng/ml/h and in low PRA specimens. The ability of the AAR to recapitulate the same screening information as the ARR was evaluated by ROC analysis. In the cohort of patients in whom aldosterone 200 pmol/l, i.e. those in whom PA was conceivably possible. The ROC area under curve was 0.89. The maximum diagnostic accuracy (i.e. maximum correct classification rate) was achieved at AAR of 17700 pmol/l:ng/ml but at this threshold sensitivity was only 70%. Greater than 95 % sensitivity (at 38% specificity) was achieved at 6370 pmol/l:ng/ml. See figure 3. 4 Conclusions Aldosterone analysis on the same sample preparation as AngI for PRA has been achieved and represents a significant improvement in workflow and an opportunity for substantial savings. Re5

Figure 3: ROC curve for [AngI] using the AAR result as gold-standard. gression characteristics against the current production method (SLE-LC-MS/MS) are excellent. AngI concentrations in samples for which there was no special handling had higher correlation with PRA than reported for DRC (R = 0.49, p < 10 10 vs R = 0.14, p > 0.05 [6]) in the low PRA range (< 1 ng/ml/h). However the AAR was unable to provide the same diagnostic information as the ARR when considering ARR the gold standard. We suggest that AAR and the aldosterone:angii ratio should be analyzed in a cohort of subjects who go on to full diagnostic workup with saline suppression or fludrocortisone suppression, preferably with final patient outcomes. Analytical sensitivity of AngI may require improvement and preanalytical issues with direct AngI analysis certainly represent a potential vulnerability. Correspondences to: dtholmes@mail.ubc.ca References [1] William F Young. Primary aldosteronism: renaissance of a syndrome. Clinical endocrinology, 66(5):607 618, 2007. 6

[2] Paul Milliez, Xavier Girerd, Pierre-François Plouin, Jacques Blacher, Michel E Safar, and Jean-Jacques Mourad. Evidence for an increased rate of cardiovascular events in patients with primary aldosteronism. Journal of the American College of Cardiology, 45(8):1243 1248, 2005. [3] Gian Paolo Rossi, Giampaolo Bernini, Giovambattista Desideri, Bruno Fabris, Claudio Ferri, Gilberta Giacchetti, Claudio Letizia, Mauro Maccario, Massimo Mannelli, Mee-Jung Matterello, et al. Renal damage in primary aldosteronism results of the papy study. Hypertension, 48(2):232 238, 2006. [4] Julia C Jones, Graham D Carter, and Graham A Macgregor. Interference by polar metabolites in a direct radioimmunoassay for plasma aldosterone. Annals of Clinical Biochemistry: An international journal of biochemistry in medicine, 18(1):54 59, 1981. [5] Duncan J Campbell, Juerg Nussberger, Michael Stowasser, AH Jan Danser, Alberto Morganti, Erik Frandsen, and Joël Ménard. Activity assays and immunoassays for plasma renin and prorenin: information provided and precautions necessary for accurate measurement. Clinical chemistry, 55(5):867 877, 2009. [6] CA Dorrian, BJ Toole, S Alvarez-Madrazo, A Kelly, JMC Connell, and AM Wallace. A screening procedure for primary aldosteronism based on the diasorin liaison R automated chemiluminescent immunoassay for direct renin. Annals of clinical biochemistry, 47(3):195 199, 2010. [7] Paul J Taylor, Donald P Cooper, Richard D Gordon, and Michael Stowasser. Measurement of aldosterone in human plasma by semiautomated hplc tandem mass spectrometry. Clinical chemistry, 55(6):1155 1162, 2009. [8] Edward Hinchliffe, Stephanie Carter, Laura J Owen, and Brian G Keevil. Quantitation of aldosterone in human plasma by ultra high performance liquid chromatography tandem mass spectrometry. Journal of Chromatography B, 913:19 23, 2013. [9] Jessica Grace Van Der Gugten, Joshua Dubland, Hua-Fen Liu, Alex Wang, Christine Joseph, and Daniel T Holmes. Determination of serum aldosterone by liquid chromatography and 7

tandem mass spectrometry: a liquid liquid extraction method for the absciex api-5000 mass spectrometry system. Journal of clinical pathology, 65(5):457 462, 2012. [10] Jessica Grace Van Der Gugten, Matthew Crawford, Russell P Grant, and Daniel T Holmes. Supported liquid extraction offers improved sample preparation for aldosterone analysis by liquid chromatography tandem mass spectrometry. Journal of clinical pathology, 65(11):1045 1048, 2012. [11] Laura J Owen and Brian G Keevil. Supported liquid extraction as an alternative to solid phase extraction for lc-ms/ms aldosterone analysis? Annals of Clinical Biochemistry: An international journal of biochemistry and laboratory medicine, page 0004563213480758, 2013. [12] Julie A Ray, Mark M Kushnir, John Palmer, Seyed Sadjadi, Alan L Rockwood, and A Wayne Meikle. Enhancement of specificity of aldosterone measurement in human serum and plasma using 2d-lc ms/ms and comparison with commercial immunoassays. Journal of Chromatography B, 970:102 107, 2014. [13] Caroline Schirpenbach, Lysann Seiler, Christiane Maser-Gluth, Felix Beuschlein, Martin Reincke, and Martin Bidlingmaier. Automated chemiluminescence-immunoassay for aldosterone during dynamic testing: comparison to radioimmunoassays with and without extraction steps. Clinical chemistry, 52(9):1749 1755, 2006. [14] Francesca Pizzolo, Angela Corgnati, Patrizia Guarini, Chiara Pavan, Antonella Bassi, Roberto Corrocher, and Oliviero Olivieri. Plasma aldosterone assays: comparison between chemiluminescence-based and ria methods. Clinical chemistry, 52(7):1431a 1432, 2006. [15] Victoria F Fredline, Eva M Kovacs, Paul J Taylor, and Anthony G Johnson. Measurement of plasma renin activity with use of hplc-electrospray-tandem mass spectrometry. Clinical chemistry, 45(5):659 664, 1999. [16] S Carter, LJ Owen, MN Kerstens, RPF Dullaart, and BG Keevil. A liquid chromatography tandem mass spectrometry assay for plasma renin activity using online solid-phase extraction. Annals of clinical biochemistry, 49(6):570 579, 2012. 8

[17] Cory E Bystrom, Wael Salameh, Richard Reitz, and Nigel J Clarke. Plasma renin activity by lc-ms/ms: development of a prototypical clinical assay reveals a subpopulation of human plasma samples with substantial peptidase activity. Clinical chemistry, 56(10):1561 1569, 2010. [18] Robert Popp, David Malmström, Andrew G Chambers, David Lin, Alexander G Camenzind, J Grace van der Gugten, Daniel T Holmes, Michael Pugia, Marta Jaremek, Shannon Cornett, et al. An automated assay for the clinical measurement of plasma renin activity by immunomaldi (imaldi). Biochimica et Biophysica Acta (BBA)-Proteins and Proteomics, 2014. [19] Alexander G Camenzind, Jessica Grace van der Gugten, Robert Popp, Daniel T Holmes, and Christoph H Borchers. Development and evaluation of an immuno-maldi (imaldi) assay for angiotensin i and the diagnosis of secondary hypertension. Clin Proteomics, 10:20, 2013. 9