Computational Analysis on Commercially Available Stent Designs

Similar documents
Computational Analysis on Commercially Available Stent Designs

Numerical Analysis of Coronary Stent for Diverse Materials

Current DES designs and expansion capacity testing

JADAVPUR UNIVERSITY & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1

Dynamic simulation of stent deployment effects of design, material and coating

FINITE ELEMENT ANALYSIS OF SHAPE MEMORY ALLOY NiTi STENT IN CORONARY ARTERY. Amatulraheem Al-abassi. A thesis presented to University of Guelph

Keywords: Angioplasty, Explicit finite elements method, Tube hidroforming, Stents.

Evaluation of Stent Performances using FEA considering a Realistic Balloon Expansion

In-stent Restenosis: the Achille's Heel of SFA Stenting

Stents selection and optimal implantation: sizes, design, deployment Abbott Vascular. All rights reserved.

Angiographic long-term results after implantation of the paclitaxel-eluting coronary stent coroflex please : Data under real-world conditions

FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY

A Novel Low Pressure Self Expanding Nitinol Coronary Stent (vprotect): Device Design and FIH Experience

HCS Working Group Seminars Macedonia Pallas Hotel, Friday 21 st February Drug-eluting stents Are they all equal?

Resolute Integrity. Independent Conformability Assessment

Modeling of stent implantation in a human stenotic artery

A Study of the Mechanical Properties of a Real Coronary Stent Dynamical Behaviour of an Optimal Stent

A Highly Flexible Slotted Tube Stent Design Coated with a-sic:h First Clinical Experiences

Nitinol biomedical devices: Design analysis

Development of Computational Models for Evaluation of Mechanical and Hemodynamic Behavior of an Intravascular Stent

Technical Characteristics of coronary prostheses(stents)

Experimental investigation of the stent artery interaction

In-stent Restenosis Diagnostic and Therapeutic Challenges. Kostis Raisakis General Hospital of Athens «G. Gennimatas»

THE COMPREHENSIVE FINITE ELEMENT MODEL FOR STENTING: THE INFLUENCE OF STENT DESIGN ON THE OUTCOME AFTER CORONARY STENT PLACEMENT

Lessons for technique and stent choice

Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery

Numerical Analysis of the Influence of Stent Parameters on the Fatigue Properties

Catch-up Phenomenon: Insights from Pathology

Le# main treatment with Stentys stent. Carlo Briguori, MD, PhD Clinica Mediterranea Naples, Italy

NC EMERGE TM PTCA Dilatation Catheter

Abstract. Introduction. Journal of Biomedical Engineering and Research

Intravascular Ultrasound

Finite Element Analysis of Stent Deployment: Understanding Stent Fracture in Percutaneous Pulmonary Valve Implantation

Sites of Atherosclerosis In order of Frequency

INDEX 1 INTRODUCTION DEVICE DESCRIPTION CLINICAL PROGRAM FIRST-IN-MAN CLINICAL INVESTIGATION OF THE AMAZONIA SIR STENT...

STENTYS for Le, Main Sten2ng. Carlo Briguori, MD, PhD Clinica Mediterranea Naples, Italy

Sirolimus Nanocrystal Balloon Based Delivery for Coronary DES ISR

Αγγειοπλαστική σε Eπαναστενωτικές Bλάβες

Computational analysis of mechanical stress-strain interaction of a bioresorbable scaffold with blood vessel

COMPLEX CASES: LEFT MAIN

Finally, the Control You Need to Deliver Accurate Treatment

Renal artery stenting- A pictorial review of stent designs, indications and outcomes

Investigation of metallic surface area of coronary stents

2010 Korean Society of Cardiology Spring Scientific Session Korea Japan Joint Symposium. Seoul National University Hospital Cardiovascular Center

A comparison of balloon- and self-expanding stents

Impact of overxpan/on strategies: Insights from bench Tes/ng

Longitudinal stent distortion in clinical practice. How common is the problem and does stent design matter?

Assessment of the Effects of Increasing Levels of Physiological Realism in the Computational Fluid Dynamics Analyses of Implanted Coronary Stents

Computational Structural Modelling of Coronary Stent Deployment: A Review

A Case of Recoiling of Everolimus - eluting Stent at the Ostium of the Right Coronary Artery by Out - stent Plaque Progression

Coroflex ISAR. Sirolimus Eluting Polymer-Free Coronary Stent System. Vascular Systems

PCI for Left Anterior Descending Artery Ostial Stenosis

Mechanisms and Predictors of Carotid Artery Stent Restenosis A Serial Intravascular Ultrasound Study

Intracoronary Stenting and Angiographic Results: Strut Thickness Effect on Restenosis Outcome (ISAR-STEREO-2) Trial

Guide Wires design and selection Abbott Vascular. All rights reserved.

PCI for Long Coronary Lesion

Biodegradable Stents An update and work-in

Branch Pulmonary Artery Stenosis Long-term Outcomes of Endovascular Stenting

Will Bioresorbable Scaffolds Change How We Think About Left Main PCI?

Davide Capodanno, MD, PhD Associate Professor, University of Catania, Italy

Design and Analysis of Dental Surgical Fixture

TRACKABILITY MEASUREMENT OF CORONARY STENT IN A CORONARY VESSEL MODELL

BIOFREEDOM: Polymer free Biolimus A9 eluting

DISRUPT CAD. Todd J. Brinton, MD Clinical Associate Professor of Medicine Adjunct Professor of Bioengineering Stanford University

Culprit Lesion Remodeling and Long-term (> 5years) Prognosis in Patients with Acute Coronary Syndrome

Drug Filled Stent Optical Coherence Tomography Results from RevElution Trial Stent Strut Coverage and Stent Apposition

ELUTING NOVEL BIODEGRADABLE VASCULAR STENTS FOR IMPLANTATION

Intravascular Imaging Insights into the Mechanism of Action of Focal Force Balloon Angioplasty

Clinical outcomes between different stent designs with the same polymer and drug: comparison between the Taxus Express and Taxus Liberte stents

Ruofei Jia, Zening Jin, Hong Li, Jing Han. Introduction

Drug Eluting Stents: Bifurcation and Left Main Approach

Comparison of Stent Designs using Computational Fluid Dynamics

CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries

The titanium nitride oxide stent an alternative to DES. Adam de Belder MD FRCP Sussex Cardiac Centre BSUH on behalf of Hexacath

Cover Page. Author: Wang, Ancong Title: Automatic quantification of intravascular optical coherence tomography Issue Date:

KEN-ICHIRO SASAKI, HIDETOSHI CHIBANA, TAKAFUMI UENO, NAOKI ITAYA, MASAHIRO SASAKI AND YOSHIHIRO FUKUMOTO

Prediction of stent endflare, arterial stresses and flow patterns in a stenotic artery

MULTIVESSEL PCI. IN DRUG-ELUTING STENT RESTENOSIS DUE TO STENT FRACTURE, TREATED WITH REPEAT DES IMPLANTATION

Complex PCI. Your partner in complex PCI: In-stent restenosis (ISR)

Bioabsorbable stents: early clinical results. Dr Angela Hoye MB ChB, PhD Senior Lecturer in Cardiology Hull & East Yorkshire Hospitals

ReZolve2 Bioresorbable Coronary Scaffold Clinical Program Update

For Personal Use. Copyright HMP 2013

Non-Destructive Analysis of Explanted Coronary Artery Stents

A computational study of stent performance by considering vessel anisotropy and residual stresses

Incidence and predictors of drug-eluting stent fractures in long coronary disease

Polymer-Free Stent CX - ISAR

Longitudinal compression of the cobalt-chromium. everolimus-eluting stent during Percutaneous Coronary

Angiographic and Intravascular Ultrasound Predictors of In-Stent Restenosis

Estimation and Comparison of T Graft Versus Conventional Graft for Coronary Arteries

Intravascular Ultrasound for Complex Cases

Design and Simulation of a Magnesium Based Biodegradable Stent for Hemodialysis Application

Stress and Displacement Analysis of Dental Implant Threads Using Three-Dimensional Finite Element Analysis

Declaration of conflict of interest. Nothing to disclose

LM stenting - Cypher

Biomed Environ Sci, 2014; 27(4):

Rationale and algorithm for below-the-knee acute gain optimization

Journal of the American College of Cardiology Vol. 46, No. 5, by the American College of Cardiology Foundation ISSN /05/$30.

Post-conditioning. P a g e 1. To my Thesis Committee,

Disclosure Eberhard Grube, M.D.

Transcription:

Computational Analysis on Commercially Available Stent Designs KRISTI BASU, PRANAB GHOSH, SUVRANGSHU DAS, SAUNAK BHATTACHARYA ABHIJIT CHANDA SCHOOL OF BIOSCIENCE AND ENGINEERING JADAVPUR UNIVERSITY

ATHEROSCLEROSIS- the problem

THE TREATMENT CABG ANGIOPLASTY ANGIOPLASTY followed by STENT

WHAT IS A STENT? A stent is an artificial 'tube' inserted into a natural passage/conduit in the body to prevent, or counteract, a disease-induced, localized flow constriction. The term may also refer to a tube used to temporarily hold such a natural conduit open to allow access for surgery.

Material requirements of stents Radiopacity Biocompatibility Haemocompatibility Corrosion-resistance Good fatigue properties

Design Requirements of Stents High Radial Strength Low Elastic Radial Recoil Good Flexibility Low Stent Profile Good Trackability Minimal Foreshortening Minimal Elastic Longitudinal Recoil Optimum Scaffolding

Restenosis : A MAJOR CONCERN Result of arterial damage with subsequent neointimal tissue proliferation Binary angiographic restenosis is defined as >= 50% luminal narrowing at follow up angiography 20% restenosis rate in 12 months restenosis occurring in 20-50% of vessels stented with bare-metal stents 3%-20% restenosis rate in DES DES have reduced restenosis rates significantly However, stent design is a still a major determinant of in-stent restenosis

Restenosis: CAUSES

Determination of cases where stent implantation is considered Identification of various materials and designs used in commercially used stents Identification of the stent deployment pressures by which a stent is inflated

Percentage Coronary Artery Total number of people Stenosis LAD LCA RCA in each category of stenosis 50 % 2 0 0 2 50-60% 4 3 4 11 60-70% 2 3 6 11 70-80% 0 2 3 5 80-90% 6 4 2 12 90-100% 4 3 8 15 Total number of people with stenosis in each artery 18 15 23 GRAND TOTAL 56

Measurement of Stenosis

Q: In a free fall, how long would it take for an object to reach the ground from the Leaning Tower of Pisa? Ans: I have never performed this experiment.

Stationary Analysis of Stent expansion: PS Stent Bx Velocity Stent

Stationary Analysis of Stent expansion: 1. COMSOL Multiphysics v4.3a 2. Structural Mechanics Module 3. Symmetry applied in all its arms (struts) 4. Boundary conditions: Deployment pressure of 14 atm, 15 atm and 16 atm applied at the inner wall of the stent A pressure of 780 KPa applied in the outer wall of the stent by atherosclerotic plaque build up

Stationary Analysis of Stent expansion: Material properties of stent materials: S. No. Material Density (kg/m 3 ) Modulus of Elasticity Poisson s Ratio UTS of material (MPa) (GPa) 1. 316L SS 7850 193 0.226 595 2. L 605 Co-Cr 9100 243 0.3 1020 3. Pt- Cr 9900 203 0.3 834 4. Ni- Ti 6478 83 0.3 1100-1200 5. Tantalum 1669 185 0.35 285

Meshing PS Stent Bx Velocity Stent

S. No. Material Von Mises Stress (MPa) 1. 316L Stainless Steel at 14 atm pressure at 15 atm pressure at 16 atm pressur e 878.53 1020.3 1162 2. L 605 Co-Cr 558.96 649.15 739.35 3. Pt- Cr 599.56 696.22 792.88 4. Nitinol 871.93 1012.60 1153.3 5. Tantalum 866.85 1006.7 1146.6

S. No. Material Will FAILURE occur? 1. 316L Stainless Steel at 14 atm pressure at 15 atm pressure at 16 atm pressur e YES YES YES 2. L 605 Co-Cr 3. Pt- Cr 4. Nitinol 5. Tantalum NO NO NO NO NO NO NO NO NO YES YES YES

S. No. Material Von Mises Stress (MPa) 1. 316L Stainless At 14 atm At 15 atm At 16 atm pressure pressure pressure 475.59 552.28 628.96 Steel 2. L 605 Co-Cr 470.66 546.56 622.45 3. Pt- Cr 469.0 544.88 620.66 4. Nitinol 470.62 546.51 622.39 5. Tantalum 466.79 542.06 617.32

S. No. Material Will Fracture Occur? 1. 316L Stainless Steel At 14 atm At 15 atm At 16 atm pressure pressure pressure NO NO YES 2. L 605 Co-Cr NO NO NO 3. Pt- Cr NO NO NO 4. Nitinol NO NO NO 5. Tantalum YES YES YES

Conclusion (1) 1.Stent deployment technique was an important factor that determined the success or failure of stents. 2.Stainless Steel stents might experience Mechanical failure under high deployment pressure. 3.L 605 Cobalt Chromium Alloy is highly acceptable biomaterial. 4.Platinum Chromium alloy also makes a very good metallic alloy for coronary stent design. 5.Nitinol can also be considered as a good biomaterial provided other aspects of the material like nickel release and corrosion gives satisfactory results. 6.Tantalum should be avoided when considering stent designs

Conclusion (contd.) The high values of Von Mises Stress in the stents lead to arterial injury which leads to neo intimal hyperplasia resulting in Restenosis. Even if the stresses exceed the UTS of the material, it doesn t necessarily lead to immediate breakage of the metallic stents. But as a result of this phenomenon micro cracks are likely to develop in the body of the stents which due to fatigue loading over time ultimately contributes to mechanical failure of the stents.

STRUT DESIGNS Rectangular struts Square struts Circular struts

Behavior of Rectangular struts (0.2mm X 0.1mm) in Newtonian Fluid Model 4 3.5 3 2.5 2 1.5 1 0.5 0 recirculation length (mm) during STRUT optimization 0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 1.25 recirculation length (mm) During FILLET optimization 1.2 1.15 1.1 1.05 1 0.95 0 0.01 0.02 0.03 0.04 0.05 0.06

5 4.5 4 3.5 3 2.5 2 1.5 1 0.5 0 Behavior of Rectangular struts (0.2 X 0.1mm) in Non Newtonian Fluid Model recirculation length (mm) during STRUT optimization 0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 0.9 recirculation length (mm) during FILLET optimization 0.8 0.7 0.6 0.5 0.4 0.3 0 0.01 0.02 0.03 0.04 0.05 0.06

Recirculation (mm) In Newtonian Models Recirculation (mm) In Non Newtonian Models 1 0.9 0.8 0.7 0.6 0.5 0.4 0.3 1 0.4 0.5 0.5 0.45 0.4 0.35 0.3 0.25 0.2 0.15 0.4 0.4 0.5 0.2 0.1 0.1 0.05 0 Rectangle Square Circle 0 Rectangle Square Circle

Conclusion (2) 1. In rectangular struts best results achieved when struts are placed at a distance of 1.5mm apart (Newtonian model) and 1.25mm apart (Non Newtonian Model). 2. Square and Circular struts delivered best results in terms of achieving minimum recirculation length if they were placed 0.7mm apart regardless of the Newtonian/ non Newtonian blood behavior 3. Presence of rectangular struts initiated larger recirculation lengths when compared with square or circular struts. 4. The rectangular struts initiated large vortex formations in the central region of fluid flow 5. The square shaped struts of stents gave the best results in our study. The recirculation lengths were found to be least in the case of square shaped struts

More clinical data analysis (n 500) required to find the exact trends in coronary artery disease where stent angioplasty is incorporated Intra Vascular Ultrasound images in this kind of clinical study would greatly enhance in finding the nature of stenosis and the exact three dimensional nature of atherosclerotic plaque radial design of the stents should be simulated so as to achieve exact mechanical behavioral patterns Radial displacement of stents due to expansion Detailed analysis of various factors counteracting the stent expansion Inlet flow pressure of coronary artery, Reynold s number or vorticity indices should be included

FUTURE SCOPE OF STUDY For a complete analysis of stents behavior, the research should be carried out under the following three radiuses: Extensive Clinical Analysis Complete Computational Analysis of stents incorporating all factors of structural mechanics and fluid dynamics Experimental analysis of various designs and fluid dynamics

References Ganong s Review of Medical Physiology, 23 rd Edition. McGraw Hill www.wikipedia.org www.ptca.org www.cornell.edu C. Lally, D. J. Kelly, P. J. Prendergast. Stents. 2006. Wiley Encyclopedia of Biomedical Engineering Metals Handbook, tenth edition, Volume 2: Properties and Selection: Nonferrous Alloys and Special-Purpose Materials, ASM International, Materials Park, Ohio, 1990, 1328 pages. O Brien B, Stinson JS, Larsen S, et al. A platinum chromium steel for cardiovascular stents. Biomaterials. 2010;31:3755-3761. Rainer Hoffman, Gary S Mintz, Philipp K Haager, Togul Bozoglu, Eberhard Grube, Michael Gross, Christian Beythien, Harald Mudra, Jurgen vom Dahl, Peter Hanrath. Relation of stent design and stent surface material to subsequent in- stent intimal hyperplasia in coronary arteries determined by intravascular ultrasound. Am J of Cardiol. 2002. 89(12): 1360-1364 Barbara Huibregtse, Juan F. Granada. New DES Platforms. Cardiac interventions today 2011 p: 35-39 Osterle, S. N., Whitbourn, R., Fitzgerald, P. J., Yeung, A.C., Stertzer, S. H., Dake, M. D., Yock, P. G., Virmani, R. (1998). The stent decade: 1987 to 1997. American Heart Journal, 136, 579-599 Kastrati A, Mehilli J, Dirschinger J, et al. Intracoronary Stenting and Angiographic Results: Strut Thickness Effect on REstenosis Outcome (ISAR-STEREO) trial. Circulation. 2001;103:2816-2821 Briguori C, Sarais C, Pagnotta P, et al. In-stent restenosis in small coronary arteries: impact of strut thickness. J Am Coll Cardiol. 2002;40:403-409.

References Frédérique Etave, Gérard Finet, Maurice Boivin, Jean-Claude Boyer, Gilles Rioufol, Gilbert Thollet, Mechanical properties of coronary stents determined by using finite element analysis. J Biomech.2001. vol 34 Issue8 pg 1065-75 Migliavacca F, Petrini L, Colombo M, Auricchio F, Pietrabissa R. Mechanical behavior of coronary stents investigated through the finite element method. J Biomech. 2002 Jun;35(6):803-11. Foin N, Alegria E, Sen S, Petraco R, Nijjer S, Di Mario C, Francis DP, Davies JE. Importance of knowing stent design threshold diameters and post-dilatation capacities to optimise stent selection and prevent stent overexpansion/incomplete apposition during PCI. Int J Cardiol. 2012 Oct 31. pii: S0167-5273(12)01291-0 Barbara M Johnston, Peter R Johnston, Stuart Corney, David Kilpatrick, Non-Newtonian blood flow in human right coronary arteries: steady state simulations. J Biomech 37(2004) 709-720 Mohammadi H, Bahramian F. Boundary conditions in simulation of stenosed coronary arteries. Cardiovasc Eng. 2009 Sep;9(3):83-91 Wu W, Qi M, Liu XP, Yang DZ, Wang WQ. Delivery and release of nitinol stent in carotid artery and their interactions: a finite element analysis. J Biomech. 2007;40(13):3034-40 Linxia Gu, Shijia Zhao, Ashwini K. Muttyam, James M. Hamel. The Relation between Arterial Stress and Restenosis Rate after coronary Stenting. Journal of Medical Devices 4 (2010), no. 031005.