The Hanau Modular Articulator System 194 Fixed 11 cm width, adjustable Radial-Shift Guidances. Illustrated Instruction Manual

Similar documents
The Hanau Modular Articulator System 190 Fixed 11 cm width, adjustable Bennett Angle and/or Adjustable Protrusion. Illustrated Instruction Manual

3000 Series Articulator System

Hanau Series H2 Articulator

The Condyle of the Articulator is a part of the Lower Member and functions as the condyle of the patient s mandible.

DENAR TRACK II SYSTEM instruction manual

Combi ArtiCulAtor SyStem instruction manual

Face-Bow Instructions

Articulator Instructions

8500 Series Articulator and QuickMount Face-Bow Instruction Manual

Prosthodontics. Dental articulator. 1-Allow most of the prosthetic work to be done in the absence of the patient.

Analog Selector Instructions

8500 Series Articulator and QuickMount Face-Bow Instruction Manual

Lab Procedure Prior to Try-in

529-A Treatment and Management of the Edentulous Patient. Upon completion of this course the student should be able to:

III Instructions CPI - Condylar Position Indicator. These instructions apply to the following items: h2500 Includes: 9611 D 9615 D

4000 Series Articulator and QuickMount Face-Bow Instruction Manual

Articulators. 5- Wax up and refining the occlusion for dental restorations.

MM01-Facebow And Bite Registration Procedure Checklist

Kois Dento-Facial Analyzer System Instructions

Fundamentals of Recording & Mounting. Workshop Instructions April 21, Michael J Melkers, DDS, FAGD & Jeanine M McDonald, DDS, FAGD

Unique Dental Supply. Catalog We Bring World Quality Products to Enhance your Business QUALITY PRODUCTS SINCE 1974

We have simplified the. for you! Just a few steps to the articulator system of your choice! choose the perfect articulator for you!

Horizontal jaw relations: The relationship of mandible to maxilla in a

Jaw relation (Maxillomandibular relationship): any one of the infinite

Arrangement of the artificial teeth:

Splint Appliance. Selection Guide. Great Lakes. See inside...

Concepts of occlusion Balanced occlusion. Monoplane occlusion. Lingualized occlusion. Figure (10-1)

Low Row. User manual E S S E N T I A L S T R E N G T H

Jaw relation registration in RPD

Horizontal Jaw Relation

Removable Prosthodontics. Complete Denture Procedures for a Team Approach in the Home

AD2 MEASURES CONDYLE DISPLACEMENT (MCD) MANUAL

1 Lecture hour per week; 3 credits. CUNY Certification in reading, writing and math.

Removable Prosthodontics. Complete Denture Procedures for a Team Approach

Bite-Tab Compound Disks 180/box Накусочные таблетки для фиксации прикусной вилки лицевой дуги. Features & Benefits. Bite-Tray регистраты 50/Bag

Multipower. User Guide E S S E N T I A L S T R E N G T H

KaVo PROTAR evo. Exciting design and function.

Exciting design and function.

E-type Straight Nose Cone

The Effect of the Central Bearing Plate Form on the Fischer Angle

Occlusal Splint Fabrication Technique

Linear Occlusion. By Dr. John P. Frush

Complete Denture Clinic Complete Removable Dental Prosthesis Procedures

Instructions for use for the medical professional English

Surveying. 3rd year / College of Dentistry/University of Baghdad ( ) Page 1

Instructions for Use SomnoDent Signature and Standard Line Fusion, Flex, Classic, Herbst Advance, Herbst Telescopic, Herbst Shim, AIR & AIR+ Devices

Implant and Tooth Supported Full-Mouth Rehabilitation with Hobo Twin-Stage Technique

LOCATOR F-Tx SEATING AND REMOVAL TOOL PROCEDURES ADDENDUM

WEIGHT BENCH EXERCISER User s Manual

DENAR Fully Adjustable Procedure Manual

USER S MANUAL QUESTIONS? CAUTION. Visit our website at. new products, prizes, fitness tips, and much more!

Jaw relations and jaw relation records

Symposium on Occlusal Articulation. Mandibular Movement Recordings and Articulator Adjustments Simplified. Harry C. Lundeen, D.D.S.

Swissfloat User Manual

Full-mouth rehabilitation of a patient with severe attrition using the Hobo twin-stage procedure

VISTA400 REMOVABLE FOOT PLATE. Instructions For Use. Step on yellow pedals to close and open legs to go around chairs & ambulate.

MX SELECT OWNER S MANUAL. Revolutionising Strength Training REGISTER YOUR PRODUCT AT MXSelect. MX_Select.

Articulators & Accessories

PROSTHODONTICS 525P CLINICAL PREPARATORY COURSE IN REMOVABLE PARTIAL DENTURES. Spring Quarter 2008

Element-Z Screw-Retained Hybrid

ESTHETIC AND FUNCTIONAL REHABILITATION OF THE PATIENT WITH SEVERELY WORN DENTITION USING TWIN STAGE PROCEDURE: A CASE REPORT

Introduction to the Panadent System

Arrangement of posterior artificial teeth Standardized parameters Curve of Wilson Curve of Spee

Restoring Severe Anterior Wear Cases; A Step by step Process

Occlusion in complete denture

LEARN FROM A LEADER. Intensive, hands-on training that will prepare you to confidently fabricate the appliances you want.

For over 95 years, Whip Mix has been a quality leader in offering innovative, well-crafted equipment and supplies to the dental industry.

Solid Zirconia Full-Arch Implant Prosthesis (Protocol C All-CAD with Multi-Unit Abutments) BruxZir. FIRST Appointment. The BruxZir

Adjustable Ab Board G

Exclusively distributed in the USA and Canada by: DMG America LLC Englewood, New Jersey 07631, U.S., Phone: (800)

ARAB AMERICAN UNIVERSITY. Lab. Manual. Prosthetic Dentistry1; Removable Prosthodontics. 3 rd year

Hex-Lock Abutment System. Restorative Manual

DLT 211 ADVANCED COMPLETE DENTURES

Treatment Options for Restoring Edentulous Jaws using One- and Two-Piece Implants from Implant Direct Int l

WEIGHT BENCH EXERCISER User s Manual

Distal Cut First Femoral Preparation

Vertical relation: It is the amount of separation between the maxilla and

PRIMARY POROUS PATELLA WITH TRABECULAR METAL. Surgical Technique

SelectTech 3.1 Bench Assembly / Owner s Manual


STEP-BY-STEP-ANLEITUNG

VIP Partner Laboratory (PL) Process Flow

Power-Pak 445 Adjustable Dumbbell System

AN EVALUATION OF MANDIBULAR BORDER MOVEMENTS THEIR CHARACTER AND SIGNIFICANCE

Midwest E Motor. Directions For Use. Doc. No

SelectTech 4.1 Bench. Assembly Manual

Try-in of the Trial Denture by Dr. Mahmoud Ramadan

weight.detach/bench ITEM NO: 76310

TOTAL KNEE ARTHROPLASTY SYSTEM

Occlusion and removable prosthodontics

Operating Instructions PROTAR evo. Always on the safe side.

Interim Denture Interim Complete Dental Prosthesis Clinical Steps

Real World Implant Prosthetics: Fixed and Removable Samuel M. Strong, DDS


INSIGHT & INNOVATION. Envelope of Parafunction: 7 Steps of Treatment Planning Many methods and theories have been

Full mouth rehabilitation of a patient with enamel hypoplasia using hobo s twin-tables technique for occlusal rehabilitation-a case Report

WEIGHT BENCH User s Manual

fusion treatment is powered by suresmile technology to provide 3D models and fully-customized wires based on your scans of the patient.

Thank you for purchasing the AB COASTER. Our customer service staff is ready to help and to answer any questions.

Transcription:

The Hanau Modular Articulator System 194 Fixed 11 cm width, adjustable Radial-Shift Guidances Illustrated Instruction Manual

A Study Case for Occlusal Analysis This instruction manual illustrates the preparation and adjustment of the Hanau Modular Articulator for complete denture prosthodontics. The Hanau Spring-Bow is used in this text to record and transfer the patient s condyle/maxillary arch relation to the Modular Articulator. Hanau Modular Articulator Hanau Spring-Bow facebow Master casts of the maxillary and mandibular edentulous arch are required. Stabilized baseplates with occlusal rims have also been prepared previously. Three interocclusal relation records are preferred. One centric relation; one right and one left lateral are required. If elected, an alternate straight protrusive may be substituted for the laterals. Centric Right & Left OR Lower Upper Master casts and occlusal rims. Centric & Protrusive Interocclusal relation records. 3

Illustrated Procedure 1 Heat forks of bitefork and pierce into the upper wax rim. Wax rim. Pierced and rigid. Stem at left. Do not distort occlusal surface. 2 Lower occlusal rim & upper occlusal rim/bitefork assembly seated in patient s mouth. 3 Spring-bow application. Earpiece. Orbitale. Bow. Bitefork stem. Transfer rod. 4

4 Remove the Spring-bow from patient. 5 Attach anterior elevator to the Spring-bow transfer rod. Transfer rod. Anterior elevator. 6 Disengage both centric latches and remove upper member. Centric latches. 5

7 Set both condylar inclinations at 0 degree. Loosen thumbscrews and rotate both guides to zero degree and re-tighten. Zero. 8 Set right and left radial-shifts at 3 mm and Bennett angles at 6. Radial-shift at 3 mm, Tighten thumbscrew. Bennett angle at 6, Tighten thumbscrew. 9 Set incisal pin at mid-line, chisel end down as shown, and attach a mounting plate. Lock. Mid-line. Incisal pin. Mounting plate. Chisel end. 6

10 Replace upper member and lock articulator in centric. Rotate and lock both latches for centric. Secure thumbscrews. Latches. 11 Align adjustable incisal guide. Chisel end. Zero indicating line. Zero flat. Slide Large platform lockscrew. Small locknut. 12 Attach accessory cast support. Cast support. 7

13 Remove earpieces and suspend Spring-bow from condylar shafts on articulator. Condylar shaft. Earpiece hole. Suspension hole. 12 mm 14 Elevate Spring-bow to align orbitale pointer at underside of mounting plate. Mounting plate. Underside. Orbitale pointer. Elevate. 15 Additional weight of stone cast is supported by the tee of cast support. Support wax rim. Lock. 8

16 Seat and lute upper cast in the baseplate. Upper cast. Baseplate. 17 Attach the upper cast to the mounting plate. Upper mounting plate. Stone or plaster. Upper cast. Incisal pin contact with guide. 18 Loosen bitefork clamp 3 and remove bow, cast support and carefully the bitefork. To prevent distortion of the occlusal surface, bitefork may remain until tooth set-up. Do not distort occlusal surface. 9

19 Centric record compensation = thickness at second molar x 3. Lower pin 3 mm. Centric record. 1mm at molar 2mm at central 3mm at incisal pin 20 Turn articulator over and attach a mounting plate. Mounting plate. 21 Seat and lute lower cast in baseplate. Position and lute upper/lower rims in centric. Lower cast. Lower baseplate. Centric record. Upper baseplate. 10

22 Mount the lower in centric using stone or plaster. Close. Contact. Centric record. Check latches. 23 Place articulator upright and remove centric relation record and dual-end pin. Centric record. Dual-end pin. 24 Release both latches. Loosen left thumbscrews for inclination and radial-shift. Thumbscrew for condylar inclination. Thumbscrew for radial-shift. Release both latches. 11

25 Place the right lateral relation record between the upper and lower rims. Upper bite rim. Right lateral. Lower bite rim. 26 Hold rims in right record. Rotate left guidance to contact the superior surface with the condylar element. Tighten thumbscrew last. Light pressure Rotate guide to contact Element. 27 The left guidance has been rotated to contact the superior surface with the condyle. Rotate Contact superior surface with condyle. In right lateral 12

28 Slide the radial-shift to contact the medial of condylar element. Slide radial-shift to contact. Tighten thumbscrew last. Element. Hold upper in right lateral. 29 The 6 degree left Bennett (radial-shift) has contacted the condylar element. Slide Radial-shift. Slide to contact 6 Bennett with condyle. 6 In right lateral 30 Loosen right thumbscrews for inclination and radial-shift, place left lateral record between bite rims. Thumbscrew for radial-shift. Thumbscrew for condylar inclination. Left lateral record. 13

31 Hold bite-rims in record. Rotate right guidance to contact the superior surface with the condylar element. Tighten thumbscrew last. Lightly hold bite rims in left lateral. Rotate Rotate guide to contact element. 32 Slide the radial-shift to contact the medial of condylar element. Tighten thumbscrew last. Slide Bite rims in left lateral record. 33 Remove left lateral record, lock in centric and replace incisal pin. Check alignment. Incisal pin. Correct if necessary Alignment of chisel end. Platform screw. Loosen small locknut. 14

34 Six upper and six lower anteriors arranged for esthetics and phonetics. 35 Loosen centric latches and protrude: lingual of upper to incisal of lower Protrude and lift off. 36 Rotate adjustable incisal guide to contact central table with chisel end pin. Chisel end contacting central table Then tighten small locknut. Rotate 15

37 Guide the upper cast to a right lateral, cusp-to-cusp relation and Thumb pressure Incisal pin. Lift-off. 38 Turn thumbscrew to elevate lateral wing to contact the corner of incisal pin. Contact incisal pin. Lateral wing. Elevate Thumbscrew. 39 Thumb guide the upper cast into a left lateral cuspid-to-cuspid relation and Thumb guide Incisal pin lifts off lateral wing. 16

40 Raise the lateral wing to contact incisal pin, secure both wings with locknuts. Elevate Contact. Lateral wing. Locknut. 41 Remaining teeth are set into occlusion and checked in working, balancing and protrusive. 17

Alternative Straight Protrusive Hanau Formula L = H/8 + 12 (substitute for steps 24 through 32) 42 Lock articulator in centric. Set both left and right radial-shift at 0 mm and lock with thumbscrew Radial-shift at 0 mm. Tighten thumbscrew. Locked in centric. 43 Set both left and right Bennett angles at 30 degree. Loosen centric latches and thumbscrews for condylar inclination. Loosen thumbscrew. Loosen latches. Set at 30 degree. 18

44 Place the protrusive relation record between the upper and lower rims. Upper bite rim. Protrusive relation record. Lower bite rim. 45 Hold upper rim in record. Rotate both guidances to contact superior surface with condylar element. Rotate right and left guides to contact element. Tighten thumbscrew last. Element. Upper rim. Protrusive record. 46 The right and left Bennett has been rotated to contact the superior surface with the condyle. Lock thumbscrews. Tighten both screws. Contact superior surface with condyle. 19

47 Adjust right and left Bennett angles to Hanau formula. Hanau Formula for Bennett angle: L = H/8 + 12 H is horizontal condylar inclination (protrusive angle). L is the calculated Bennett angle. 48 Example. The right condylar inclination is 44 degree and is divided by 8 and accepted as 6 to which 12 is added, totaling 18 degree. 44/8 + 12 = 18 44 condylar inclination. 49 The right Bennett angle is illustrated and is then adjusted to 18 and locked. Bennett angle at 18. Tighten thumbscrew. This completes the articulator adjustment. Record all of the settings for future reference. right and left Bennett angles, right and left radial-shift, right and left condylar inclinations, the articulator serial number, and patient name and date, etc. 20

Articulator Care and Maintenance Your Whip Mix articulator is a precision instrument and requires care and maintenance. Periodic cleaning and lubricating as described below will assure prolonged life and dependable service from the instrument. Failure to follow these instructions will void your warranty. Cleaning Use a mild soap and water solution with the aid of a brush to dissolve accumulations of wax and to wash away carborundum grit. Then air dry and lubricate. DO NOT use strong detergents, alkalies, gasoline or naphtha as cleaning agents! Lubrication Lubricate the working and bearing components with a thin film of sewing machine or high speed handpiece type oil. Wipe off excess oil to prevent accumulations of dust or grit. A thin coating of petroleum jelly must be applied to all articulator surfaces that will be contacted by the gypsum mounting material. Detail E Detail A Detail D Detail C Detail B 21

Lubrication points. Oil Oil C Oil Oil A B D Oil Oil E Oil Mounting Plate thumbscrew, clean as necessary. Pull Unscrew Remove Clean and Replace STORAGE Store the articulator in a clean, dry atmosphere free of plaster and carborundum dust; away from acids, alkalies or corrosive medicaments. Wait a full day after mounting casts before storing the articulator in a carrying case or corrugated carton. Moisture dissipation from the stone in an enclosed area causes alkalinity of the stone mixture which can damage the articulator surface. 22

Warranty Whip Mix Corporation warrants the articulator system to be free from defects in material and/or workmanship for a period of one year. In the event of a defect, please notify the factory in writing of the defect prior to returning the instrument. Whip Mix Corporation will, at its option, either repair, replace or issue credit for such defects. Because Whip Mix Corporation is continually advancing the design of its products and manufacturing methods, it reserves the right to improve, modify or discontinue products at any time, or to change specifications or prices without notice and without incurring obligations.

Whip Mix Corporation 361 Farmington Avenue Louisville, KY USA 40209 Toll-Free: 800-626-5651 Phone: 502-637-1451 Fax: 502-634-4512 www.whipmix.com Whip Mix and Hanau logo are registered trademarks and Hanau is a trademark of Whip Mix Corporation. 2008 Whip Mix Corporation FN: 339637-F AD R1008