Gamma3 TM Fragment Control Clip

Similar documents
GAMMA LOCKING NAIL INSTRUMENTS OPERATIVE TECHNIQUE

Distal Targeting System Gamma3 Long Nail R2.0 / T2 Recon Nail R2.0

Operative Technique For Gamma3 Long Nails R2.0 Hip Fracture

Hoffmann II MRI External Fixation Systems. Hoffmann II MRI Hoffmann II Compact MRI Modular Systems for Upper Extremity Lower Extremity Pelvis

Asnis. Micro Cannulated screw system. Xpress operative technique

Pre-Operative Planning. Positioning of the Patient

3. PATIENT POSITIONING & FRACTURE REDUCTION 3 8. DISTAL GUIDED LOCKING FOR PROXIMAL NAIL PROXIMAL LOCKING FOR LONG NAIL 13

Craniomaxillofacial. Universal Trocar System

EasyStep. Operative technique

QUICK REFERENCE GUIDE. The Orthofix Femoral Nailing System. By Prof. Dr. D. Pennig

Orthopedic Bone Nail System - Distal Femoral Nail Surgical Technique Manual

System. Humeral Nail. Surgical Technique

Hoffmann II Compact External Fixation System. Modular System for

Femur. Monoaxial Locking Plate System. Operative Technique. Distal Lateral Femur Universal Holes Targeting Instrumentation.

Aesculap Targon FN. Head Preserving Solution for Medial Femoral Neck Fractures. Aesculap Orthopaedics

Technique Guide. SureLock Distal Targeting Device. C-arm guided targeting for trochanteric fixation nail.

Sirus Antegrade Femoral Nail System Surgical Technique

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

AxSOS. Locking Plate System. Operative Technique. Small Fragment Basic Fragment

4Fusion. Shape Memory Implant. Operative Technique

VLIFT System Overview. Vertebral Body Replacement System

Surgical Technique. CONQUEST FN Femoral Neck Fracture System

Hoffmann II Compact External Fixation System

VariAx TM Distal Radius Locking Plate System

PROXIMAL FEMORAL NAIL REMOVAL SET

Technique Guide. DHS Blade. For osteoporotic bone.

Clavicle Hook Locking Plate

Biomet Large Cannulated Screw System

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

Zimmer ITST Intertrochanteric/ Subtrochanteric Fixation System. Abbreviated Surgical Technique

3. Insert Tocar Sleeves Insert the NCB tissue protection sleeve assembly 1.6 to 10mm through a skin incision (Fig. 38).

AxSOS Locking Plate System

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

Humerus Block. Discontinued December 2016 DSEM/TRM/0115/0296(1) Surgical Technique. This publication is not intended for distribution in the USA.

Hoffmann II Micro External Fixation System. Indications for the Hand

PROXIMAL HUMERAL NAILING SYSTEM OPERATIVE TECHNIQUE

Elbow Hinge Fixator. Guided Flexion/Extension for Unstable Elbow Fractures.

SMARTLockTM. Hybrid MMFTM

PAL Pelvic Alignment Level

Leibinger Next Generation Universal Trocar System

Collinear Reduction Clamp

Technique Guide. 3.5 mm LCP Olecranon Plates. Part of the Synthes locking compression plate (LCP) system.

Cannulated Pediatric Osteotomy System (CAPOS). A single system of osteotomy blade plates and cannulated instrumentation.

PediNail Pediatric Femoral Nail

Hoffmann II Micro External Fixation System

LCP Low Bend Medial Distal Tibia Plates 3.5 mm. Anatomic plates with low profile head for intra- and extraarticular fractures.

Asnis III Cannulated Screw System. Operative technique

Double Engine Orthopedic Bone Nail System Universal Humeral Nail

Open reduction; plate fixation 1 Principles

Periarticular Aiming Arm Instruments for LCP Proximal Tibial Plate 4.5/5.0. Part of the LCP Periarticular Aiming Arm Instrument System (large).

SWEMAC CHS. Compression Hip Screw System

Mini External Fixator.

CableFIX Xpress Carpometacarpal Fixation System. Operative technique

US Patent No. 6,309,396B1. G. David Ritland, M.D.

LCP Distal Tibia Plate

MetaFix Ludloff Plate

NeoGen Femoral Nail System

NCB Proximal Humerus Plating System

S2 Tibial Nail. Operative Technique

Surgical Technique. Targeter Systems Overview

T2 Arthrodesis Nailing System. Operative Technique

Monotube Triax. External Fixation System. External Fixation. Operative Technique

operative technique Kent Hip

Mandible Reconstruction Module Reference ence Guide

Cannulated Screw System. Product Overview

T2 Supracondylar Nailing System. Operative Technique

TRAUMATOLOGY. Trochanter

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

LCP Superior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

Foot & Ankle T2. Ankle Arthrodesis Nail. Foot & Ankle

Integra. DigiFuse Cannulated Intramedullary Fusion System SURGICAL TECHNIQUE

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments. Part of the 3.5 mm LCP Percutaneous Instrument System.

Zimmer MIS Periarticular 3.5mm Proximal Tibial Locking Plate

TABLE OF CONTENTS. 2 (8144 Rev 2)

Zimmer Natural Nail System

WINSTA-C. Clavicle Plating System

The NBX Non-Bridging External Fixator A Non-Bridging External Fixator/Locking Plate capturing a series of.062mm K-wires and 3mm half-pins that are

A locking plate system that expands a surgeon s options in trauma surgery. Zimmer NCB Plating System

OBSOLETED. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

LCP Distal Humerus Plates

PediLoc 3.5mm and 4.5mm Contour Femur Plate Surgical Technique

humerus InSafeLOCK Nail

VariAx Distal Radius Locking Plate System

Technique Guide. LCP Posterior Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment LCP system.

AxSOS 3 Titanium Proximal Lateral Tibia Locking Plate System. Operative technique Targeting instrumentation

A locking plate system that expands a surgeon s options in trauma surgery. Zimmer NCB Plating System

Surgical Technique. Cannulated Angled Blade Plate 3.5 and 4.5, 90

Hoffmann II External Fixation System

Technique Guide. PHILOS and PHILOS Long. The anatomic fixation system for the proximal humerus.

Asnis III. Cannulated Screw System. Operative Technique. 4.0mm 5.0mm 6.5/8.0mm. Shoulder. Elbow. Hand & Wrist. Hip. Pelvis. Femur.

Distal Ulnar Locking Plate

Lag Screw Device Intended for symphyseal fracture fixation of the mandible

TIBIAL NAIL Tx OPERATIVE TECHNIQUE

3.5 mm LCP Olecranon Plates

Cannulated Pediatric Osteotomy System (CAPOS). A single system of osteotomy blade plates and cannulated instrumentation.

Handling Technique. Collinear Reduction Clamp. For minimally invasive fracture reduction.

N-Force Fixation System. Surgical Technique

Technique Guide. 3.5 mm LCP Periarticular Proximal Humerus Plate. Part of the Synthes locking compression plate (LCP) system.

Transcription:

Osteosynthesis Gamma3 TM Fragment Control Clip Operative Technique Hip Fracture

Introduction Indication The Fragment Control Clip is designed to stabilize rotationally unstable femoral head-neck fragments during preparation of the Lag Screw canal and insertion of the Gamma3 Lag Screw. It can be used for Gamma3 Trochanteric and Long Nails. Lateral View Frontal View Instrument features The Fragment Control Clip is made from radiolucent carbon fibres as is the Gamma3 Targeting Arm Safety Clamp Lever arm Groove for the Lag Screw Guide Sleeve Note: Follow the Gamma3 Operative Technique up to the chapter entitled Lag Screw Insertion where the K-wire is in place and the Lag Screw length determination has been done. Situation prior to the Assembly of the Fragment Control Clip onto the Targeting Device At this point, continue with the Operative Technique. The instructions for the Fragment Control Clip should not interfere with or replace any chapters in the Gamma3 Operative Technique except for those detailing Lag Screw Insertion. Note: K-wires are not intended for re-use. They are single use only. K-wires may be damaged or bent during surgical procedures. If a K-wire is re-used, it may become lodged in the cannulated instruments and could be advanced into the pelvis, and may damage large blood vessels or cause other serious injuries. It is important to observe the K-wire tip during drilling with the image intensifier. 2 Figure 1

Operative Technique Assembling of the Fragment Control Clip on the Targeting Device Assemble the Fragment Control Clip, with the open Lever Arm onto the inserted Lag Screw Guide Sleeve and then on to the Targeting arm (Figure 2). A snap is felt, if the Fragment Control Clip is placed correctly onto the Lag Screw Guide Sleeve. Make sure that the Fragment Control Clip is positioned in close contact to the Targeting Arm (Figure 2). Close the Lever Arm completely to stabilize the Fragment Control Clip. Resistance is felt, if the Lever Arm is in its final position. Figure 2 Inserting of the Fragment Control Sleeve Advance the Fragment Control Sleeve through the anterior hole of the Safety Clamp in the Fragment Control Clip. The anterior Safety Clamp of the Fragment Control Clip has to be pressed down during this procedure to allow free sliding of the sleeve. The Safety Clamp is marked with anterior left, for the left femur and anterior right, for the right femur. If the Safety Clamp is released, it will keep the Fragment Control Sleeve in place and prevent it from falling out. It will also prevent the Fragment Control Sleeve from sliding during drilling and K-wire insertion. Figure 3a The Fragment Control Sleeve has to be pushed to the level of the skin. This indicates the site for a small incision down to the bone. The Fragment Control Sleeve is now advanced through the incision (Figure 3a). If the Fragment Control Sleeve catches the fascia lata, twisting will usually allow it to pass through to the bone. Alternatively, pass the K-wire with the blunt end in front through the Fragment Control Sleeve. Advance the Fragment Control Sleeve and the K-wire together to the bone, while pushing the anterior Safety Clamp. Remove the K-wire. In order to open the lateral cortex by drilling, the Fragment Control Sleeve must be in contact with the lateral cortex of the femur (Figure 3b). Figure 3b 3

Operative Technique Opening of the lateral cortex Using the white coded centre tipped drill 3.0mm 300mm, the lateral cortex should be carefully opened by using a power tool (Figure 4). K-wire insertion The white coded 3.0mm x 300mm drill is then replaced by the 3.2mm x 450mm K-wire. Using the Guide Wire Handle, the K-wire should be placed as close as possible to the subchondral bone of the femoral head, to allow maximum stabilization of the headneck fragment (Figure 5). Figure 4 Lag Screw Insertion and Lag Screw Fixation After proper fracture fragment stabilization, remove the Guide Wire Handle and follow the Gamma3 Operative Technique beginning at the chapter entitled Lag Screw Insertion, where the K-wire is in place and the Lag Screw length determination has been done. After the Lag Screw Insertion is complete, follow the Gamma3 Operative Technique until the chapter entitled Lag Screw Fixation (Figure 6). Figure 5 Make sure that the fixation of the Lag Screw and the insertion of the Set Screw is done correctly. Now disconnect the Lag Screwdriver by loosening the end thumbwheel, and remove the Lag Screwdriver and the K-wire of the Lag Screw. Removal of the Fragment Control Clip By using the Guide Wire Handle, the K-wire of the Fragment Control Clip must be removed after Lag Screw fixation. Remove the Fragment Control Sleeve by pressing the Safety Clamp down and then remove the Fragment Control Clip by opening the lever arm. After removal of the Fragment Control Clip, follow the Gamma3 Operative Technique beginning at the chapter entitled Distal Screw Locking, if distal locking is required. The Lag Screwdriver and the K-wire of the Lag Screw should already have been removed! Note: The Fragment Control Clip should only be used if the diameter of the femoral neck allows K-wire insertion without penetrating the cortex of the femoral neck, to avoid damaging the circumflex artery. It is strongly recommended to use the anterior approach of the Fragment Control Clip only. A posterior insertion of the K-wire could cause injury of the circumflex artery at the dorsal portion of the trochanteric region. If the fixation of the Fragment Control Clip on the Targeting Arm/ Lag Screw Guide Sleeve is not properly done, a dislocation of the K-wire may cause injuries. 4 Figure 6

Ordering Information - Instruments REF Description Standard Instruments 1320-0160 Fragment Control Clip 1320-0170 Fragment Control Sleeve 1320-3030S Drill 3.0mm x 300mm, AO small, sterile, white coded 1320-0160 Fragment Control Clip Lateral View 1320-0160 Fragment Control Clip Frontal View 1320-0170 Fragment Control Sleeve 1320-3030S Drill 3.0mm x 300mm, AO small, sterile, white coded This publication sets forth detailed recommended procedures for using Stryker Trauma devices and instruments. It offers guidance that you should heed, but, as with any such technical guide, each surgeon must consider the particular needs of each patient and make appropriate adjustments when and as required. A workshop training is required prior to first surgery. 5

Trauma, Extremities & Deformities Biologics Surgical Products Neuro & ENT Stryker Trauma GmbH Prof.-Küntscher-Straße 1-5 D-24232 Schönkirchen Germany www.osteosynthesis.stryker.com The information presented in this brochure is intended to demonstrate a Stryker product. Always refer to the package insert, product label and/or user instructions before using any Stryker product. Surgeons must always rely on their own clinical judgment when deciding which products and techniques to use with their patients. Products may not be available in all markets. Product availability is subject to the regulatory or medical practices that govern individual markets. Please contact your Stryker representative if you have questions about the availability of Stryker products in your area. Stryker Corporation or its subsidiary owns the registered trademark: Stryker Stryker Corporation or its subsidiary owns, uses or has applied for the following trademarks: Gamma3 Pentax owns the following trademark: Literature Number: B0300012 LOT B1806 Copyright 2006 Stryker Printed in Germany