MSP TM Metatarsal Shortening System. Surgical Technique

Similar documents
MSP TM Metatarsal Shortening System. Surgical Technique

Correction System. Surgical Technique

Correction System. Surgical Technique

Axi+LineTM Proximal Bunion CorrectionSystem. Surgical Technique

Bunion Correction System

Correction System. Surgical Technique

InCoreTM Lapidus System

Correction System. Surgical Technique

Flower Opening Wedge Plate

MetaFix Ludloff Plate

Osteotomy System. The reference lines on the plate help facilitate the creation of the osteotomy, when a free hand cut is preferred.

Osteotomy System. Surgical Technique

MEDICAL ADVANCED TECHNOLOGY FOOT SURGERY: COMPRESSION AND VARISATION STAPLE SYSTEM FOOT SURGERY STAPLE FIXATION

The Flower Medial Column Fusion Plate

SMV Scientific Bone Plate and Screw System Surgical Technique

The Flower Forefoot PROCEDURE GUIDE.

Flower Medium Headless & Cannulated Screws

CHARLOTTE. Snap-Off Screws SURGICAL TECHNIQUE

Merete BLP. Surgical Technique. - Bunion Locking Plate - Low Profile Locking Bone Plate System

Merete PlantarMAX Lapidus Plate Surgical Technique. Description of Plate

The Flower Four Corner Fusion Plate

PediLoc Tibia Surgical Technique

MetaCun DuoMetaCun. Midfoot Fusion Locking Plate System. Surgical Technique and Ordering Information

Ankle Fixation System. System Brochure

The Flower Straight Fibula Plate

Arcos Modular Femoral Revision System. Broach and Calcar Proximal Bodies Surgical Technique

DYNAMIC, TRANSVERSE COMPRESSION. Low Profile, Anatomic Design, Type II Anodized. Mechanical Compression Designed to Stimulate the Fusion Process

QUICK REFERENCE GUIDE. MiniRail System. Part B: Foot Applications. By Dr. B. Magnan, Dr. E. Rodriguez and Dr. G. Vito ALWAYS INNOVATING

DVR Crosslock Distal Radius Plating System. Product Brochure

Surgical Technique. Cannulated Angled Blade Plate 3.5 and 4.5, 90

Astrid Modular Forefoot System. Surgical Technique

The Flower Proximal Humerus Plate

MIS Cemented Tibial Component

Surgical Technique. CONQUEST FN Femoral Neck Fracture System

Cannulated Angled Blade Plate 3.5 and 4.5, 90.

MPJ Implant System. Patents Pending Worldwide

A New Benchmark in Tibial Fixation

PEDUS-L. Locking Plantar Lapidus Plate

LCP Anterior Ankle Arthrodesis Plates. Part of the Synthes Locking Compression Plate (LCP) System.

Humeral SuturePlate. Surgical Technique

DARCO. Bow 2 Plate SURGIC AL TECHNIQUE

LCP Distal Tibia Plate

1. Orthoapedic Associates of Michigan, PC, Grand Rapids, MI 2. Michigan State University College of Human Medicine, Grand Rapids, MI

Arcos Interlocking Distal Stem. Surgical Technique Addendum to the Arcos Modular Femoral Revision System

Locking Proximal Humerus Plate. For complex and unstable fractures.

S U R G I C A L T E C H N I Q U E David A. McQueen, MD Return to Menu

Integra. ADVANSYS Plating System SURGICAL TECHNIQUE

LCP Distal Humerus Plates

MICA. Minimally Invasive Foot Surgery CHEVRON OSTEOTOMY SURGIC AL TECHNIQUE

Cannulated Pediatric Osteotomy System (CAPOS). A single system of osteotomy blade plates and cannulated instrumentation.

NeoFit. Metatarsophal angeal Arthrodesis Pl ate

MiniRail System. Part B: Foot Applications. By Dr. B. Magnan, Dr. E. Rodriguez and Dr. G. Vito

OptiLock Periarticular Plating System For Distal Tibial Fractures. Surgical Technique

Hand Fracture System. Surgical Technique

Surgical Technique. Forearm Fracture Solutions

SALVATION 3Di. Plating System SURGICAL TECHNIQUE

HammerFUZE HAMMERTOE COMPRESSION SYSTEM SURGICAL TECHNIQUE

Accu-Cut Osteotomy Guide System

Forefoot/Midfoot Plating System. Surgical Technique

3.0 mm Cannulated Screw System. Surgical Technique

For Distal Femur Fractures. 95º Condylar Plate. Quick Reference Chart

Zimmer NexGen MIS Tibial Component. Cemented Surgical Technique IMAGE TO COME

FlexitSystem EFFICIENT MOBILITY. High Tibial Opening Wedge Osteotomy (HTO)

CHARLOTTE. 3.0 and 4.3 Multi-Use Compression Screws SURGIC A L T ECHNIQUE

DRACO FOOT & ANKLE SYSTEM PRODUCT INFORMATION

Surgical Technique 4.5/8.5MM BEAMING SYSTEM. Customer Service:

ToeMobile. Surgical Technique. Great Toe Endoprosthesis

Arthrodesis of the Ankle

System. Humeral Nail. Surgical Technique

NEXGEN COMPLETE KNEE SOLUTION S A. Tibial Stem Extension & Augmentation Surgical. ATechnique

RAYHACK Kienbock Radial Shortening System

Surgical Technique. Clavicle Locking Plate

Surgical Technique. Distal Humerus Locking Plate

3. PATIENT POSITIONING & FRACTURE REDUCTION 3 8. DISTAL GUIDED LOCKING FOR PROXIMAL NAIL PROXIMAL LOCKING FOR LONG NAIL 13

Locking Ankle Plating System. Surgical Technique

The Vilex FUZETM. Dual Thread Screw & Intramedullary Nail in One Implant. The Ultimate TTC Arthrodesis Internal Fixator

EasyStep. Operative technique

TABLE OF CONTENTS. 2 (8144 Rev 2)

LCP Ulna Osteotomy System 2.7. Low profile angular stable fixation for ulna shortening osteotomies.

Foot & Ankle. EasyClip. Osteosynthesis Compression Staples. Foot & Ankle

OSS Orthopedic Salvage System Compress Device. Surgical Technique

INnate. Surgical Technique Guide. Anatomic Fixation in Your Hands. The INnate Implant

Technique Guide. 2.4 mm Variable Angle LCP Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

lower extremity fixation system

Resurfacing Distal Femur. Orthopaedic Salvage System

CHARLOTTE Lisfranc. Recontruction System SURGICAL TECHNIQUE

Anchorage. Foot & Ankle. Plating System. Operative Technique

3. Insert Tocar Sleeves Insert the NCB tissue protection sleeve assembly 1.6 to 10mm through a skin incision (Fig. 38).

Surgical Technique. Distal Radius and Foot

Clinical. Solutions. Synthes Solutions. Foot and Ankle.

Surgical Technique. Elbow Plating System

PediNail Pediatric Femoral Nail

Olecranon Locking Plate II

Cannulated Pediatric Osteotomy System (CAPOS). A single system of osteotomy blade plates and cannulated instrumentation.

Lateral TTC Plate SURGICAL TECHNIQUE

Surgical technique. Angular Stable X-Plate and 2-Hole Plate. For osteotomies, arthrodeses and fractures of the foot.

FastFrame External Fixation System. Damage Control Surgical Technique

Transcription:

MSP TM Metatarsal Shortening System Surgical Technique

Metaphyseal/Diaphyseal osteotomy does not result in plantar displacement of metatarsal head Osteotomy guide integrated into implant allows for controlled shortening (up to 6mm) Keeled design aids in implant positioning and fixation strength T6 Driver 1.8mm AO Drill MSP Plate Drill Bushing Proximal Drill Guide DOG-LEG SLOT The MSP Metatarsal Shortening System is manufactured using Ti 6-4 ELI.

Surgical Technique 1. 2. DRILL BUSHING Select the left or right implant kit based on the operative foot. The keel is oriented medially, and the dog-leg slot is oriented proximally. The distal edge of the plate is positioned approximately 5mm proximal to the metatarsal head. Thread the MSP Drill Bushing into the distal hole of the MSP Plate. Do not overtighten. DOG-LEG SLOT DISTAL PROXIMAL KEEL Place the MSP Plate on the dorsal surface of the bone. Ensure the MSP Drill Bushing is in a vertical position. Apply dorsal and medial pressure to plate to ensure complete plate-tobone contact while drilling. NOTE: Clamp may be used (not provided) 3. Use the 1.8mm AO drill bit to drill the distal hole. Drill a hole through the MSP Drill Bushing, into the bone. Use the markings on the AO Drill to measure the screw length at the top of the MSP Drill Bushing and select the appropriate length locking screw. Remove the drill bushing. Place the Locking Screw in the distal hole and fully tighten. 16 14 12 10

4. Thread the drill bushing into the second distal hole and use the AO drill bit to drill the second distal hole. Use the markings on the AO Drill to measure the screw length at the top of the MSP Drill Bushing and select the appropriate length locking screw. Remove drill bushing. Place a Locking Screw in the second distal hole and fully tighten. 16 14 12 10 5. NOTE: Ensure the keel is in contact with the medial cortex. Using the MSP Proximal Drill Guide, drill a hole into the bone in the proximal end of the dog-leg slot. Drill as far proximal as possible. Measure the screw length at the top of the MSP Proximal Drill Guide and select the appropriate length Cortical Screw. Place the Cortical Screw, but do not fully tighten. Note: Use the end with the single groove and apply medial pressure to the plate while drilling to ensure complete bone-to-plate contact. Alternate Order for Screw Placement 1 Drill and insert screws in the following order: 1. Most distal hole 2. Dog-leg slot 3. Second distal hole 4. Proceed to step 6 as usual NOTE: The alternate screw placement technique may help ensure consistent, optimal lateralization of the implant. For both the standard and alternate screw placement technique, it is critical to apply medial pressure to the plate while drilling to confirm complete bone-to-plate contact. 2 3 6. Using an oscillating saw, cut the bone obliquely through the cutting guide in the MSP Plate. The saw will be perpendicular to the dorsal surface of the MSP Plate.

7. 8. 9. Shorten the metatarsal by moving the bone proximally to the desired length. A total correction of 6mm is achieved if the screw reaches the most distal edge of the dog-leg slot. Fully tighten the dog-leg screw. Using the double-groove side of the MSP Proximal Drill Guide, insert the AO drill bit into the second proximal hole and drill. Use the markings on the AO Drill to measure the screw length at the top of the MSP Drill Bushing and select the appropriate length Cortical Screw. Place the Cortical Screw in the remaining proximal hole and fully tighten. NOTE: Drill guide offsets AO drill to create 0.5mm compression across osteotomy. 9.1 For supplemental compression of the osteotomy prior to drilling the final screw hole, a clamp may be used. The MSP Plate has a divot on the keel for this purpose. NOTE: Clamp not provided.

36.5 mm 7 mm 15 degrees 1.1 mm wide cutting slot 0.5 mm compression hole 6 mm Part No. MS-SSR MS-SSL ORDERING INFORMATION Description MSP TM Metatarsal Shortening System Sterile Procedure Kit - Right 36.5 MSP TM Plate - Right, 1.8mm AO Drill, MSP TM Driver, MSP TM Drill Bushing, MSP TM Proximal Drill Guide MSP TM Metatarsal Shortening System Sterile Procedure Kit - Left 36.5 MSP TM Plate - Left, 1.8mm AO Drill, MSP TM Driver, MSP TM Drill Bushing, MSP TM Proximal Drill Guide ORDERING INFORMATION - LOCKING SCREWS Part # Description MS-2408L 2.4mm Diameter Locking Screw - 8mm Length MS-2410L 2.4mm Diameter Locking Screw - 10mm Length MS-2412L 2.4mm Diameter Locking Screw - 12mm Length MS-2414L 2.4mm Diameter Locking Screw - 14mm Length MS-2416L 2.4mm Diameter Locking Screw - 16mm Length MS-2418L 2.4mm Diameter Locking Screw - 18mm Length ORDERING INFORMATION - NON-LOCKING SCREWS Part # Description MS-2408 2.4mm Diameter Cortical Headed Screw - 8mm Length MS-2410 2.4mm Diameter Cortical Headed Screw - 10mm Length MS-2412 2.4mm Diameter Cortical Headed Screw - 12mm Length MS-2414 2.4mm Diameter Cortical Headed Screw - 14mm Length MS-2416 2.4mm Diameter Cortical Headed Screw - 16mm Length MS-2418 2.4mm Diameter Cortical Headed Screw - 18mm Length This surgical technique guide provides a recommended procedure for using MSP Metatarsal Shortening System. The content provided puts forth technique guidance, however, the surgeon must consider the individual needs of the patient making appropriate adjustments when and as required. Nextremity Solutions and MSP are trademarks of Nextremity Solutions, Inc. Zimmer Biomet is the exclusive distributor of the MSP Metatarsal Shortening System. The MSP Metatarsal Shortening System is manufactured using Ti 6-4 ELI.. Patented. 2017 Nextremity Solutions LIT-1502 Rev.04 Manufactured by: 210 N. Buffalo St. Warsaw, IN 46580 U.S.A. (732) 383-7901 nextremitysolutions.com Distributed by: 1800 West Center St. Warsaw, Indiana 46580 USA (800) 613-6131 contactus@zimmerbiomet.com