TONTARRA Medizintechnik GmbH. Surgical Instruments in Demand. Laminectomy Punches and Rongeurs

Similar documents
Endoscopic Technology. Operating Manual. Cannula system

Preparation Instructions for the CAMLOG /CONELOG Implant System

Manufacturer s information

Hygienic Reprocessing

SI-BONE ifuse Implant System Instruments Hospital Cleaning and Sterilization Instructions 0344

M6-C Artificial Cervical Disc Surgical Instruments

Reprocessing Guide. Autoclavable Arthroscope and Hardware Sterilization Tray

THIN FOOTPLATE KERRISON RONGEURS

ATHLET VBR. The real winner. VBR in PEEK-OPTIMA PRODUCT INFORMATION

INSTRUCTIONS FOR USE AND CARE

Straumann ProClean Cassette. Basic Information on the

THE. Curve THE CURVE STRIPPING TOOL

Aesculap Implant Systems

CURVES. Anatomically Adaptable IPR strips. Brochure Instructions for use Processing instructions

a perfect fit INSTRUCTION MANUAL CAMLOG TORQUE WRENCH WITH CONTINUOUS TORQUE ADJUSTMENT Handling Sterilization Care

INSTRUCTIONS FOR CLEANING, DISINFECTION AND STERILIZATION OF DFS INSTRUMENTS

INFORMATION TO THE PRODUCT RANGES. Instructions for Cleaning, Disinfection, Sterilization, Inspection and Maintenance of Medartis Products APTUS MODUS

UNICLIP DIRECTIONS FOR USE REF C226U - 0,8 MM / 1 MM FOR DENTAL USE ONLY. FISDR / F EN / 11 / updated 06/2016 1/7

Rolimeter 50A. Operator s Manual for measuring anterior/posterior knee laxity

How clean are your instruments?

ProUltra ULTRASONIC SURGICAL TIPS

1) INDICATIONS FOR USE These products have to be used only in hospital environments, clinics or dental offices by qualified dental personnel.

Minimally Invasive Posterior Instruments

TJF-160F/VF Cleaning and Disinfection Checklist

Velocity Orthopedics Instrument Cleaning and Sterilization

B&H SURGICAL. Laparoscopic Reference Guide. P.O. Box Santa Clarita, CA Toll Free: Ph: Fax:

Recommendations for cleaning, disinfection and sterilization of orthopaedic instruments and devices from Swemac.

The Food and Drug Administration (FDA) is providing a detailed list of supplemental duodenoscope reprocessing measures that emerged from an agency-led

OBSOLETE. Visit for the latest version. Ankle Plating System 3 PKGI-79-B EFFECTIVE

LBL-014. Rev

ÜBERCLEAN! Introducing the new STATMATIC PLUS.

Sterile U Network. Quality Control for Table-top Steam Sterilizers. Introduction: T U T O R I A L S

ProUltra Ultrasonic Non Surgical Endo Tips

Validation of manual cleaning and disinfection procedures

care NCDC-Cleanser

Disinfection and Decontamination Core Subject. The Process of Instrument Decontamination

BS EN ISO 14971: Risk Analysis and Risk Management Report

ISO CERTIFIED AMALGAM SEPARATION. Chairside, inside the dental unit Central, sedimentation-based With complete recycling package

Dentium Surgery Kit Dental Implant Surgical Instrument Set

The Multi-unit Abutments of the CE-Series are indicated for Nobel Biocare / NobelReplace Tapered. The current implant diameters have to be attended.

MATERIAL SAFETY DATA SHEET (MSDS) Information about stainless steel for Surgical/Dental/Orthodontic Instruments.

Before & After your MATRx Study

Scope This policy applies to all personnel and departments that clean, prepare and/or sterilize items intended for patient care use.

Surgical Technique Guide

Operating Instructions SaHoMa -II mobile ultrasonic nebulizer

MetaFix Ludloff Plate

Dear Marie Consultation on a proposed Infection Prevention and Control Practice Standard

Lab-Specific Standard Operating Procedure (LSOP)- Picric Acid (2,4,6-Trinitrophenol) and Solutions Containing Picric Acid

Good oral hygiene today

X-spine Systems, Inc. Axle Interspinous Fusion System

Cleaning and sterilization instructions for Astra Tech Implant System EV

ORTHODONTIC INSTRUMENTS

SPECIALIST SPINAL RETRACTORS

SPECIALIST SPINAL RETRACTORS

TABLE OF CONTENTS. PAL-650 Instrument System Instruction Manual. Applicable Part Numbers Introduction General Warnings...

New as of: CEREC SpeedGlaze. Operating Instructions (valid for USA) English

RibLoc U Plus CHEST WALL PLATING SYSTEM FOR THE PERSONAL ATTENTION OF THE OPERATING SURGEON

Educator s Teaching Teeth Guidebook

QuickPro. Instructions for use

ATHLET VBR. The real winner. VBR in PEEK-OPTIMA PRODUCT INFORMATION

BS EN ISO 14971: Risk Analysis and Risk Management Report

Straumann ProClean Cassette. Uncompromised hygiene. Ready for action.

SURGICAL TECHNIQUE GUIDE BRECKENRIDGE. Intervertebral Body/VBR Fusion System. ACDF Anterior Cervical Discectomy and Fusion

S5 Rotary Files. only. Manufacturer. Packing unit. Batch code. For professional use only (0)

3M Fast Release Varnish

ORAL HYGIENE SESSION 2

CONTENTS. Introduction to DASK. Crestal Approach. Lateral Approach

BS EN ISO 14971: Risk Analysis and Risk Management Report

Workshop on the reprocessing of medical devices. 5. December Brussels. Nikou Ghassemieh Managing Director

SURGICAL TECHNIQUE GUIDE BRECKENRIDGE. Intervertebral Body/VBR Fusion System. Oblique Lumbar Interbody Fusion

Instructions for use RONDOflex plus REF


Aesculap Implant Systems

Implant Locator. User Manual

Instructions for Use. Contra-angle handpieces. Endea Endo Cursor EB-62 Endea EB-75 / EB-79. Contra-angle handpieces. Endo NiTi WD-73 M / WD-74 M

Instructions for use. PROPHYflex perio tip Starter Set Refill PROPHYflex perio tip

`bob`=å~ãéê~=ó=å~êéi=åäé~åáåöi= ÇáëáåÑÉÅíáçå=~åÇ=ëíÉêáäáò~íáçå

Chapter 5: Checking and Maintaining Ultrasound Equipment

English. Fealite Nasal Pillows System. User Manual

X-spine Systems, Inc. Silex Sacroiliac Joint Fusion System

Zimmer Biomet Instrument Kit System

The fitting process of SomnoGuard AP. Hints to complement the user instructions

Altus Product description / Instructions for Use IFU 6 Rev. 1

X-spine Aranax Anterior Cervical Plating System Instructions for Use / Package Insert

How To Use White With Style Whitening Kit

Compression Screw 3.2 Bioabsorbable. Product information

C O N T E N T S. Introduction of DASK. Crestal Approach. Lateral Approach. Dentium - For Dentists By Dentists

Approval Signature: Original signed by B. Postl. June 2007

Simple. Stress-free. The Newtron Ultrasonics Series

with Ø Ø 1.5 mm Titanium Screws distributed by ARCS Alveolar Reconstruction System E

USS II ILIO-SACRAL Modular System for Stable Fixation in the Sacrum and Illium

Cleaning and disinfection Tonometer measuring prisms, contact lenses and Desinset

Instructions for use 0483 Permanent and temporary YASARGIL Aneurysm and Vessel Clips

Instructions for use EXPERTmatic handpiece E10 C - REF

URGENT FIELD SAFETY NOTICE

R.A.P.T.O.R. SYSTEM. Bisco. Instructions for Use. Composite/Porcelain Finishing Burs. U.S. Patent: 5,779,476

X-spine Systems, Inc. Calix PC Spinal Implant System

Smarter Thinking. Simpler Design. Prima Plus. Surgical Manual

Transcription:

TONTARRA Medizintechnik GmbH Surgical Instruments in Demand Laminectomy Punches and Rongeurs

KERRISON; Laminectomy Punches; clean wave KERRISON with Standard footplate Art.Nr. Size FL 240-072-02A-CW 2 mm 200 mm 240-072-03A-CW 3 mm 200 mm 240-072-04A-CW 4 mm 200 mm 240-072-05A-CW 5 mm 200 mm Advantage of the clean wave system: Especially the wavelike shape of the separating plane allows finding impurities and stains on the sliding shaft instrument which can be removed and cleaned when being detected. The effectiveness of a standardized cleaning procedure was evaluated and documented by an independent hygiene laboratory. The excellent cleaning results can be attributed to the fact that any type of cleaning utensil reaches very easily the area between upper and lower part of the sliding shaft instrument. Beyond the cleaning aspect thanks to the wavelike shape there is a clear reduction of friction of the sliding parts. FL = function length

KERRISON; Laminectomy Punches; clean wave KERRISON thin footplate Art.Nr. size FL 240-072-02AT-CW 2 mm 200 mm 240-072-03AT-CW 3 mm 200 mm 240-072-04AT-CW 4 mm 200 mm 240-072-05AT-CW 5 mm 200 mm

Laminectomy Rongeurs; clean wave CUSHING, LOVE-GRUENWALD, SPURLING Art.Nr. Spez. Jaw Size FL CUSHING 240-552-02-CW straight 2,0x 10,0 mm 180 mm 240-602-02-CW angled up 2,0x 10,0 mm 180 mm 240-652-02-CW angled down 2,0x 10,0 mm 180 mm LOVE-GRUENWALD 240-552-03-CW straight 3,0x 10,0 mm 180 mm 240-602-03-CW angled up 3,0x 10,0 mm 180 mm 240-652-03-CW angled down 3,0x 10,0 mm 180 mm SPURLING 240-552-04-CW straight 4,0x 10,0 mm 180 mm 240-602-04-CW angled up 4,0x 10,0 mm 180 mm 240-652-04-CW angled down 4,0x 10,0 mm 180 mm

Laminectomy Rongeurs; clean wave Casper Laminectomy Rongeurs; clean wave; toothed Art.Nr. Spez. Jaw Size FL 240-511-02-CW straight 2x12 185 mm 240-511-03-CW straight 3x12 185 mm 240-511-04-CW straight 4x14 185 mm 240-515-02-CW angled up 2x12 185 mm 240-515-03-CW angled up 3x12 185 mm 240-515-04-CW angled up 4x14 185 mm 240-518-02-CW angled down 2x12 185 mm 240-518-03-CW angled down 3x12 185 mm 240-518-04-CW angled down 4x14 185 mm

Manual Cleaning Process: 1. Immerse the instrument for 10 min into an ultrasonic bath at a minimum frequency of 35 khz. For ultrasonic cleaning use an enzyme cleaning agent (MediZym 1% (v/v); Dr. Weigert GmbH) in demineralised water. Cleaning solution temperature at the beginning of ultrasonic treatment: 20 +/- 2 C. 2. Use a soft non-metallic brush (e. g. medium-hard toothbrush). The entire surface, especially the hardly accessible parts of the instrument must be brushed until no residues were visible (for at least 1 min) under running tap water (minimum quality: drinking water; temperature of 30 +/- 5 C). 3. All sliding surfaces, recesses and openings must be flushed under running tap water (minimum quality: drinking water; temperature of 30 +/- 5 C) for 1 min. Move at the same time the articulations and sliding surfaces by opening up and closing the instrument for 20 times. 4. Afterwards rinse the instrument for 5 seconds with demineralised water. 5. Finally rub down the instrument by using a lint free cloth. Insufflate cavities of the instrument by using sterile compressed air. Automated Cleaning Process: Transfer the instruments with non-contact into the washer/disinfector Miele G7882 (or into an adequate washer/disinfector) equipped with a standard cabinet with two levels. Steps of the automated cleaning procedure: 1. Cold water rinse 2. Cleaning at 55 C (-0 C + 3 C) for 5 min with cleaning agent Sekumatic Multiclean 0.7 % v/v (Ecolab # 5109TE062). 3. Neutralization with cold water and neutralizer Sekumatic FNP 0.1 % v/v (Ecolab # 5436TE015). 4. Cold water rinse. Sterilisation: Sterilization of instruments by applying a fractionated pre-vacuum process (according. ISO 13060 / ISO17665) under consideration of the respective country requirements. Parameters for the pre-vacuum cycle: 3 prevacuum phases Heat up to a sterilization temperature of 132 +/- 1 C Minimum Holding time: 4 min Drying time: minimum 10 min Additional Instructions: It is the duty of the user to ensure that the reprocessing processes including resources, materials and personnel are capable to reach the required results. State of the art and often national law requiring these processes and included resources to be validated and maintained properly. Attention! The customer should be made aware, that prior to use always a visual test as well as a functional test should take place to assure that the device is safe and effective for its intended use. Instruments that are not in proper condition and not properly cleaned should be immediately removed from the tray.

TONTARRA Medizintechnik GmbH Daimlerstr 15 78573 Wurmlingen Germany Tel: +49(0) 7461 / 965 76-0 Fax: +49(0) 7461 / 965 76-26 info@tontarra.de www.tontarra.de