Comparison of Three Whole Blood Creatinine Methods for Estimation of Glomerular Filtration Rate Before Radiographic Contrast Administration

Similar documents
ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE

UvA-DARE (Digital Academic Repository) Point-of-care diagnostic tools Kosack, C.S. Link to publication

The intended place in therapy would be as an alternative to current laboratory-based creatinine testing in people needing contrast-enhanced imaging.

Papers in Press. Published September 15, 2011 as doi: /clinchem

Standardization of. S.M.Boutorabi DCLS,PhD

Can modifications of the MDRD formula improve the estimation of glomerular filtration rate in renal allograft recipients?

Clinical Chemistry / INTENSIVE INSULIN THERAPY AND GLUCOSE VALUES

Creatinine (serum, plasma)

Carboplatin Time to Drop the Curtain on the Dosing Debate

The Incidence Of Contrast-Induced Nephropathy Or Radiocontrast Nephropathy

Evaluation of the BiliChek Noninvasive Bilirubin Analyzer for Prediction of Serum Bilirubin and Risk of Hyperbilirubinemia

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

Validity and Reliability of a Glucometer Against Industry Reference Standards

Renal function vs chemotherapy dosing

Acute renal failure Definition and detection

9. GFR - WHERE ARE WE NOW?

NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE. Diagnostics Assessment Programme

Estimates of glomerular filtration rate (GFR) from serum

A New Approach for Evaluating Renal Function and Predicting Risk. William McClellan, MD, MPH Emory University Atlanta

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

Is the new Mayo Clinic Quadratic (MCQ) equation useful for the estimation of glomerular filtration rate in type 2 diabetic patients?

Comparison of Modified Jaffe s Kinetic Method and Enzymatic Method of Serum Creatinine Estimation for Precision, Linearity and Effect of Interferent

Correspondence should be addressed to I. P. L. Houben;

S150 KEEP Analytical Methods. American Journal of Kidney Diseases, Vol 55, No 3, Suppl 2, 2010:pp S150-S153

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

2017/3/7. Evaluation of GFR. Chronic Kidney Disease (CKD) Serum creatinine(scr) Learning Objectives

Evaluation of the Cockroft Gault, Jelliffe and Wright formulae in estimating renal function in elderly cancer patients

Multiple studies are available concerning the use of

Assessment of glomerular filtration rate in healthy subjects and normoalbuminuric diabetic patients: validity of a new (MDRD) prediction equation

Deciphering an Unexpected Serum Creatinine Result in a Critically Ill Patient

Correspondence should be addressed to Maisarah Jalalonmuhali;

מסקנות מיישום סטנדרטיזציה של בדיקת קראטינין : שימוש בנוסחאות לחישוב egfr

Development of a reference MSMS method for plasma creatinine

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

Nephrologists. from. All physicians Patients Health care system Policy makers World health leaders Researchers. Clinical laboratories

Relationship between glucose meter error and glycemic control efficacy

GlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data

Analytical and Clinical Performance of the epoc Blood Analysis System. Experience at a Large Tertiary Academic Medical Center

Acute Coronary Syndrome (ACS) Patients with Chronic Kidney Disease being considered for Cardiac Catheterization. PROVINCIAL PROTOCOL March 2015

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014

Εκηίμηζη ηης μεθρικής λειηοσργίας Ε. Μωραλίδης

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard

IDEXX Catalyst SDMA Test for in-house measurement of SDMA concentration in serum from dogs and cats

Keywords: albuminuria; albumin/creatinine ratio (ACR); measurements. Introduction

Page 1. Disclosures. Main Points of My Talk. Enlightened Views of Serum Creatinine, egfr, Measured GFR, and the Concept of Clearance

Chronic kidney disease (CKD) has received

Performance of MDRD study and CKD-EPI equations for long-term follow-up of nondiabetic patients with chronic kidney disease

Mortality Associated With Nephropathy After Radiographic Contrast Exposure

Applicable To Employees of Gundersen Boscobel Area Hospital Laboratory and Gundersen Palmer Lutheran Hospital and Clinics Laboratory.

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)

University of Bristol - Explore Bristol Research. Peer reviewed version. Link to published version (if available): / X

Measurement and Estimation of renal function. Professeur Pierre Delanaye Université de Liège CHU Sart Tilman BELGIQUE

Chronic Kidney Disease Prevalence and Rate of Diagnosis

Glomerular Filtration Rate. Hui Li, PhD, FCACB, DABCC

Glomerular Filtration Rate Estimations and Measurements. Pierre Delanaye, MD, PhD University of Liège CHU Sart Tilman BELGIUM

13. POINT-OF-CARE CREATININE TESTING IN HIGH- RISK PATIENTS

USRDS UNITED STATES RENAL DATA SYSTEM

SCIEX Vitamin D 200M Assay for the Topaz System

ACCESS hstni SCIENTIFIC LITERATURE

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System

Cardiac Troponin Testing and Chest Pain Patients: Exploring the Shades of Gray

Creatinine & egfr A Clinical Perspective. Suheir Assady MD, PhD Dept. of Nephrology & Hypertension RHCC

Glomerular filtration rate estimated by cystatin C among different clinical presentations

Determination of hemoglobin is one of the most commonly

Analytical performance of the i-stat cardiac troponin I assay

Seung Hyeok Han, MD, PhD Department of Internal Medicine Yonsei University College of Medicine

Measurement and Estimation of renal function. Professeur Pierre Delanaye Université de Liège CHU Sart Tilman BELGIQUE

Introduction. Study Objective. Methodology. Results. Data Analysis. Interpretation

POTASSIUM/K. 3.6 mmol/l

Supplementary Online Content

Validation of El-Minia Equation for Estimation of Glomerular Filtration Rate in Different Stages of Chronic Kidney Disease

Validity of the use of Schwartz formula against creatinine clearance in the assessment of renal functions in children

Introduction. Study Objective. Methodology. Results. Data Analysis. Interpretation

Performance of three portable blood glucose monitoring devices used in a veterinary application

Objectives. Pre-dialysis CKD: The Problem. Pre-dialysis CKD: The Problem. Objectives

Calibration of High-Density Lipoprotein Cholesterol Values From the Korea National Health and Nutrition Examination Survey Data, 2008 to 2015

Addressing error in laboratory biomarker studies

Research Article Changes in Renal Function in Elderly Patients Following Intravenous Iodinated Contrast Administration: A Retrospective Study

ORIGINAL ARTICLE Estimating the glomerular filtration rate using serum cystatin C levels in patients with spinal cord injuries

Creatinine Test Code: CRE

Evaluation of Renal Profile in Liver Cirrhosis Patients: A Clinical Study

CCQM-K6: Key Comparison on the Determination of Cholesterol In Serum* Final Report December 2001

How and why to measure renal function in patients with liver disease?

Two: Chronic kidney disease identified in the claims data. Chapter

Screening for chronic kidney disease racial implications. Not everybody that pees has healthy kidneys!

Urea Nitrogen (BUN) detection Kit

Recently, the National Kidney Foundation endorsed a

HHS Public Access Author manuscript Am J Kidney Dis. Author manuscript; available in PMC 2017 July 05.

CKD EVIDENCE TABLES - ALL CHAPTERS

British Columbia is the first

Assessing Renal Function: What you Didn t Know You Didn t Know

Egfr non african american vs african american

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers

Assessment of estimated GFR and clinical predictors of contrast induced nephropathy among diabetic patients undergoing cardiac catheterization

Accuracy of Three Dry-Chemistry Methods for Lipid Profiling and Risk Factor Classification

From the 1 Department of Transplantation, Mayo Clinic, Jacksonville, FL; 2 Baylor Regional Transplant Institute, Dallas, TX; 3 Division of

Contrast Induced Nephropathy

Early Detection of Chronic Kidney Disease (CKD)

Validation of the Korean coefficient for the modification of diet in renal disease study equation

Chapter 2: Identification and Care of Patients With CKD

Transcription:

Clinical Chemistry / Whole Blood Creatinine for egfr Comparison of Three Whole Blood Creatinine Methods for Estimation of Glomerular Filtration Rate Before Radiographic Contrast Administration Nichole L. Korpi-Steiner, PhD, 1 Eric E. Williamson, MD, 2 and Brad S. Karon, MD, PhD 1 Key Words: Creatinine; Estimated glomerular filtration rate; Point of care; Amperometric biosensor Abstract We compared the clinical concordance of estimated glomerular filtration rate (egfr) based on 3 whole blood creatinine assays with the egfr calculated from a reference plasma creatinine assay. Whole blood creatinine on the Radiometer ABL800 FLEX (Radiometer A/S, Bronshoj, Denmark) demonstrated the best correlation and concordance to plasma creatinine/egfr compared with the i-stat (i-stat, East Windsor, NJ) and StatSensor (Nova Biomedical, Waltham, MA). The i-stat had better sensitivity (compared with Radiometer) but poorer specificity for prediction of plasma egfr less than 60 ml/min/1.73 m 2. The StatSensor demonstrated lower concordance of whole blood to plasma egfr but offered a slope and an intercept offset feature that partially compensates for this effect. The optimal device for use in rapid determination of egfr from whole blood creatinine may depend on whether it is more important in a given practice to optimize sensitivity, specificity, or overall concordance for determining plasma egfr less than 60 ml/min/1.73 m 2. Chronic kidney disease is a major health problem affecting about 19 million persons in the United States. 1 Owing to its increasing incidence, national recommendations now exist for estimation of glomerular filtration rate based on serum creatinine values, using the Modification of Diet in Renal Disease (MDRD) study equation. 2,3 Estimation of the glomerular filtration rate from serum creatinine levels is particularly useful in situations requiring rapid evaluation and recognition of renal impairment, such as the case of patients undergoing procedures requiring administration of radiographic contrast media. 4 In these cases, the estimated glomerular filtration rate (egfr) is valuable in identifying patients at risk of developing contrast-induced nephropathy (CIN), a major cause of hospital-acquired renal failure and a significant cause of morbidity and mortality among hospitalized patients. 4 Measurement of the serum creatinine level, with calculation of egfr based on the MDRD equation, is strongly recommended as a means to identify patients at risk for CIN. 4 Strategies such as alternative contrast agents, preventive therapies, and reduced contrast agent dose can be used if patients with renal impairment are identified before administration of a contrast agent. Some of these therapies are effective in reducing the incidence of CIN. 4,5 For this reason, interest in rapid, whole blood creatinine assays for use at the point of care has increased, especially in institutions with a high volume of procedures requiring radiographic contrast media. However, calculation of the egfr from serum creatinine levels is not without limitations. Analytic bias and interferences in many commercially available serum or plasma creatinine assays is a source of major concern. 3 The 920 Am J Clin Pathol 2009;132:920-926 Downloaded 920 from https://academic.oup.com/ajcp/article-abstract/132/6/920/1761213

Clinical Chemistry / Original Article effect of bias can be partially overcome by using different MDRD equations depending on the calibration traceability of the laboratory creatinine assay and reporting higher egfr values as more than 60 ml/min/1.73 m 2 because creatinine assay bias impacts high (>60 ml/min/1.73 m 2 ) egfr estimations to a greater extent than it does lower values. 2,3 Patients with egfr less than 60 ml/min/1.73 m 2 are at higher risk of CIN, and for this reason, published guidelines recommend using the egfr to identify at-risk patients. 4 Whole blood creatinine assays for use at the point of care would be ideal for rapid estimation of the egfr in patients needing radiographic contrast media and for other situations in which rapid determination of the egfr would be useful. However, some whole blood creatinine methods have increased bias relative to central laboratory methods, 6,7 which could make estimation of egfr at the clinical decision point of 60 ml/min/1.73 m 2 even more problematic. To lower the risk of CIN, patients scheduled for computed tomography (CT) examinations requiring contrast administration in our practice must have a documented creatinine/egfr value available if they are older than 70 years, have a history of diabetes, or have a history of renal disease or renal transplantation. For patients scheduled for a CT examination who meet these criteria and do not have a current (within 30 days) creatinine/egfr value available and patients who have a creatinine/egfr value pending at the time of CT, a blood sample is sent to the stat laboratory for creatinine/egfr measurement. In this study, we compared 3 whole blood creatinine methods with a laboratory plasma creatinine assay used as a reference method in 266 patients scheduled for CT examination who required a stat creatinine/egfr before contrast administration. Clinical concordance of the egfr calculation was determined by using the 3 whole blood methods compared with the reference plasma creatinine assay. The results can be used to define and differentiate the ability of current whole blood creatinine assays to correctly identify patients at risk of CIN. Materials and Methods Sample Selection For the study, 266 excess lithium heparin whole blood samples submitted for stat evaluation of creatinine/ egfr before contrast administration for CT procedures were used. Institutional protocol requires creatinine/egfr measurement for patients older than 70 years, patients with a history of diabetes, and patients with a history of renal disease or renal transplantation. For patients meeting these criteria without a current (within 30 days) creatinine/ egfr value available and patients with a pending creatinine/egfr value at the time of CT, samples are sent to the stat laboratory for creatinine/egfr measurement. Sample selection was not consecutive because staff was available only during selected hours to perform the whole blood creatinine testing. The study was conducted between November 1 and December 31, 2008. Creatinine values, patient age, sex, and ethnicity (African American or not African American) were recorded for each sample. The study design was approved by the Mayo Clinic Institutional Review Board (Rochester, MN). Creatinine Methods The reference method was the Roche enzymatic creatinine method performed on lithium heparin plasma on a Roche Cobas Integra 400 analyzer (Roche Diagnostics, Indianapolis, IN). This assay is calibrated to the recommended calibration standard based on isotope dilution mass spectrometry (IDMS). 2,3 National Institute of Standards and Technology (NIST) Standard Reference Material 967- Creatinine in Human Frozen Serum (SRM 967) was used to validate the reference assay. Replicate (n = 5) measurements of NIST SRM 967 with assigned values of 0.8 and 3.9 mg/dl (71 and 345 μmol/l) resulted in mean (± SD) Roche enzymatic creatinine measurements of 0.8 ± 0.0 and 4.08 ± 0.04. Daily quality control at a target level of 0.8 mg/ dl (71 μmol/l) creatinine demonstrated an interassay coefficient of variation of less than 4% during the study period. Thus, the Roche IDMS-calibrated plasma assay used for the study was a suitable reference method for plasma creatinine and egfr determination, especially given that most samples tested in the stat laboratory have values between 1.0 and 2.0 mg/dl (88-177 μmol/l) creatinine. Whole blood creatinine methods tested included the Radiometer ABL800 FLEX blood gas analyzer (Radiometer Medical A/S, Bronshoj, Denmark), which uses an amperometric biosensor (electrode) based on the enzymatic conversion of creatinine to creatine, then sarcosine, followed by a reaction that generates hydrogen peroxide from sarcosine. 8 The i-stat (i-stat, East Windsor, NJ) and StatSensor (Nova Biomedical, Waltham, MA) handheld portable creatinine analyzers, which use similar chemical reactions to measure creatinine electrochemically, 7 were also evaluated. For each whole blood method, linearity and precision were assessed by replicate analysis of linearity material and commercial control samples produced by each manufacturer. For the 266 patient samples evaluated, excess lithium heparin whole blood was removed after sample mixing to run on the i-stat, StatSensor, and Radiometer methods (in that order) followed by centrifugation of the sample for 2 minutes at 4,500g for analysis of plasma creatinine on the Integra 400. Downloaded from https://academic.oup.com/ajcp/article-abstract/132/6/920/1761213 Am J Clin Pathol 2009;132:920-926 921 921 921

Korpi-Steiner et al / Whole Blood Creatinine for egfr Calculation of egfr For Roche Integra 400 and Radiometer ABL 800 FLEX measurements, the egfr was calculated by using the MDRD equation adjusted for methods calibrated to IDMS traceability 2 : egfr (ml/min) = 175 Cr 1.154 Age 0.203 ( 0.742 if female) ( 1.212 if African American) For i-stat and StatSensor creatinine measurements, the egfr was calculated by using the MDRD equation originally validated for use with conventional creatinine calibrations 2,3 : egfr (ml/min) = 186 Cr 1.154 Age 0.203 ( 0.742 if female) ( 1.212 if African American) Data Analysis Microsoft Excel (Microsoft, Redmond, WA) was used to calculate the egfr for each patient sample based on whole blood or plasma creatinine levels and using the equations defined in the preceding sections. Microsoft Excel was also used for linear regression analysis of whole blood vs plasma (reference) creatinine levels and to calculate mean (± SD) bias for whole blood vs plasma creatinine levels. Bland-Altman plots were constructed to visually compare the relationship between whole blood and plasma creatinine for the range measured. Sensitivity was defined as the number (percentage) of whole blood egfr values that correctly predicted a plasma egfr value of less than 60 ml/min/1.73 m 2. Specificity was defined as the number (percentage) of whole blood egfr values that correctly predicted a plasma egfr value of 60 ml/min/1.73 m 2 or more. This cutoff was chosen because published national guidelines suggest that patients with an egfr of less than 60 ml/min/1.73 m 2 are at higher risk of developing CIN and should be evaluated differently from patients with higher egfr. 4 Overall concordance was calculated as the number (percentage) of whole blood egfr values that fell into the same category ( 60 or <60 ml/min/1.73 m 2 ) as the corresponding plasma-based egfr value. The McNemar test was used to assess the statistical significance of differences in sensitivity, specificity, and overall concordance for the 3 whole blood methods considered. A P value of less than.02 was considered statistically significant to account for multiple comparisons. The StatSensor device offers users the ability to enter a slope and an intercept offset to match whole blood to plasma creatinine values. The manufacturer of the device recommends a simple fixed offset to match whole blood to plasma values within a narrow range of values. For this reason, we calculated egfr concordance for StatSensor values with and without application of a fixed offset of 0.28 mg/ dl (25 μmol/l) creatinine, which was the offset that maximized overall concordance between StatSensor whole blood and plasma egfr values for the 266 sample data set. Results Linearity and Precision Replicate analysis (n = 5) of linearity/calibration check material provided by each manufacturer demonstrated that all whole blood devices had a tendency to slightly overestimate creatinine levels at low concentrations and slightly underestimate creatinine levels at high concentrations. The StatSensor device demonstrated a mean recovery of 134% of the expected value when a commercial sample with an assigned value of 0.75 mg/dl (66 μmol/l) of creatinine was used. The i-stat and Radiometer methods recovered 106% and 120%, respectively, of expected levels at similar creatinine values. All of the devices recovered 95% to 100% of expected values when linearity material with assigned values of 2 to 10 mg/dl (177-884 μmol/l) was used. Daily quality control material provided by each manufacturer was used to assess interassay precision during the study period. All 3 devices demonstrated coefficients of variation between 3% and 7% when control materials with values between 1 and 5 mg/dl (88-442 μmol/l) of creatinine were used daily for 20 days. Patient Demographics Among the 266 patient samples collected, severe renal failure (plasma egfr < 30 ml/min/1.73 m 2 ) was present in 4 patients, whereas 25 patients had plasma egfr between 30 and 50, and 48 patients had plasma egfr between 51 and 60 ml/min/1.73 m 2. The mean (± SD) age of the patients was 68 ± 14 years (range, 22-92 years). There were 103 women and 163 men; 264 patients were not African American, and 2 patients were African American. Relationship Between Whole Blood and Plasma Creatinine Values The 266 patient samples were analyzed on all 3 whole blood devices and compared with plasma creatinine levels on the Roche Cobas Integra 400 used as the reference method. The mean (± SD) bias between StatSensor whole blood and plasma creatinine was 0.23 ± 0.18 mg/dl, indicating that the StatSensor whole blood method systematically underestimated plasma creatinine values. The mean bias between i-stat whole blood and plasma creatinine levels was 0.13 ± 0.08 mg/dl, indicating slight overestimation of plasma creatinine by the i-stat method. The Radiometer demonstrated a mean bias vs plasma creatinine level of 0.05 ± 0.09 mg/dl. Linear regression between whole blood and plasma creatinine levels resulted in the following relationships: Y(StatSensor whole blood) = 0.59 (plasma creatinine) + 0.18, r 2 = 0.61; Y(i-STAT whole blood) = 0.95 (plasma creatinine) + 0.18, r 2 = 0.93; and Y(Radiometer whole blood) = 0.95 (plasma creatinine) + 0.00, r 2 = 0.89. 922 Am J Clin Pathol 2009;132:920-926 Downloaded 922 from https://academic.oup.com/ajcp/article-abstract/132/6/920/1761213

Clinical Chemistry / Original Article Bland-Altman plots of whole blood vs plasma creatinine are shown in Figure 1 and Figure 2. The StatSensor demonstrated a negative bias vs plasma creatinine for the range of creatinine values measured (Figure 1) and also demonstrated more variability in the relationship between whole blood creatinine and plasma creatinine levels compared with the i-stat or Radiometer. The i-stat had a constant positive bias for the range of creatinine values measured (Figure 1), whereas the Radiometer method had the smallest absolute bias relative to plasma creatinine for the range of values measured (Figure 2). Clinical Concordance Between Whole Blood and Plasma egfr Values Whole blood and plasma creatinine values, along with patient age, sex, and ethnicity, were used to calculate the egfr as defined in the Materials and Methods section. Overall concordance was calculated as the number (percentage) of all whole blood creatinine values falling into the correct plasma creatinine category (ie, <60 or 60 ml/min/1.73 m 2 ). Because the StatSensor device offers users the ability to enter an offset to match whole blood to plasma creatinine values, we calculated egfr concordance for StatSensor with and without a fixed offset of 0.28 mg/dl (see the Materials and Methods section). The Radiometer analyzer displayed the best overall concordance with plasma creatinine (93%) Table 1. The sensitivity for the detection of plasma egfr less than 60 ml/min/1.73 m 2 was 81%, whereas specificity was 97% (Table 1). All 13 patients with plasma egfr less than 60 ml/ min/1.73 m 2 misclassified by the Radiometer (whole blood egfr reported as 60 ml/min/1.73 m 2 ) had plasma egfr values more than 45 ml/min/1.73 m 2 and plasma creatinine levels of 1.5 mg/dl (133 μmol/l) or less. The i-stat demonstrated a trend toward slightly lower overall concordance (87%) but had the best sensitivity (97%) for the detection of plasma egfr less than 60 ml/min/1.73 m 2. The increased sensitivity of the i-stat is a consequence of systematic overestimation of plasma creatinine owing to the inverse relationship between creatinine and egfr values. Whole Blood Plasma Creatinine (mg/dl) 0.6 0.4 0.2 0 0.2 0.4 0.6 0.8 1.0 0 0.5 1.0 1.5 2.0 StatSensor i-stat 2.5 3.0 3.5 Whole Blood Plasma Creatinine (mg/dl) 0.6 0.4 0.2 0 0.2 0.4 0.6 0.8 1.0 0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 Whole Blood and Plasma Mean Creatinine (mg/dl) Whole Blood and Plasma Mean Creatinine (mg/dl) Figure 1 Bland-Altman plots for StatSensor and i-stat whole blood creatinine values compared with reference plasma creatinine values (difference between whole blood and plasma creatinine values vs mean of whole blood and plasma value). Values are given in conventional units. To convert to Système International units (μmol/l), multiply by 88.4. Figure 2 Bland-Altman plot for Radiometer ABL800 FLEX whole blood creatinine values compared with reference plasma creatinine values (difference between whole blood and plasma creatinine values vs mean of whole blood and plasma value). Values are given in conventional units. To convert to Système International units (μmol/l), multiply by 88.4. Table 1 Clinical Concordance Between Whole Blood and Plasma egfr * Radiometer StatSensor StatSensor Offset i-stat Plasma egfr <60 (n = 68) 55 (81) 11 (16) 40 (59) 66 (97) Plasma egfr 60 (n = 198) 192 (97) 198 (100) 174 (88) 166 (84) Overall concordance (n = 266) 247 (93) 209 (79) 214 (80) 232 (87) egfr, estimated glomerular filtration rate. * Data are given as number (percentage). Downloaded from https://academic.oup.com/ajcp/article-abstract/132/6/920/1761213 Am J Clin Pathol 2009;132:920-926 923 923 923

Korpi-Steiner et al / Whole Blood Creatinine for egfr Devices that overestimate plasma creatinine levels will have increased sensitivity for the detection of plasma egfr less than 60 ml/min/1.73 m 2 at the expense of specificity (Table 1). The StatSensor demonstrated lower sensitivity for detection of plasma samples with egfr less than 60 ml/min/1.73 m 2 when the offset was not used, as a result of systematic underestimation of plasma creatinine levels. Use of the offset (Table 1) improved sensitivity for detection of egfr less than 60 ml/min/1.73 m 2. All 28 patients with plasma egfr less than 60 ml/min/1.73 m 2 misclassified by the StatSensor with offset values (reported as 60 ml/min/1.73 m 2 ) had plasma egfr more than 43 ml/min/1.73 m 2 and plasma creatinine levels of 1.5 mg/dl (133 μmol/l) or less. Differences in sensitivity (for detection of plasma egfr <60 ml/min/1.73 m 2 ) between whole blood devices were statistically significant (P <.02) for all comparisons. Differences in specificity were statistically significant for all comparisons except StatSensor offset vs i-stat (P =.29). Differences in overall concordance between devices were statistically significant except for borderline significance of the Radiometer vs i-stat (P =.06) and no significant difference between the StatSensor and the StatSensor offset (P =.28). Discussion Chronic renal disease is a major health problem in the United States. 1 Estimation of the glomerular filtration rate from serum creatinine levels is particularly useful in clinical situations requiring rapid decision making based on renal function, such as the decision to administer radiographic contrast media before radiographic or cardiology procedures. 4 The availability of a rapid serum creatinine assay may allow differential treatment decisions to be made for patients at risk of developing contrast-induced nephropathy. 4,5 However, significant concern remains about the impact of analytic bias on egfr. 3 Clinical studies are needed to demonstrate whether whole blood creatinine assays are accurate enough to provide reliable estimates of the glomerular filtration rate at the point of care. At least 2 previous studies have found that the i-stat whole blood creatinine assay overestimates plasma creatinine levels as measured by 2 different laboratory methods. 6,7 One of those studies compared concordance of i-stat whole blood egfr with plasma egfr calculated from a Roche IDMS-calibrated creatinine method on a small group of patients (n = 50). This study found that sensitivity for detecting plasma egfr less than 60 ml/min/1.73 m 2 was 100%, whereas specificity was only 85%. These numbers are very similar to the 96% sensitivity (<60 ml/min/1.73 m 2 ) and 87% specificity ( 60 ml/min/1.73 m 2 ) observed in our larger study. Thus, multiple studies have now demonstrated that systematic overestimation of plasma creatinine levels on the i-stat leads to very sensitive detection of plasma egfr less than 60 ml/min/1.73 m 2 at the expense of specificity. The usefulness of i-stat screening may, therefore, depend on the adverse consequences of misclassifying patients with plasma egfr 60 ml/min/1.73 m 2 or more, which likely differ by procedure and practice environment. No similar study has been done using the Radiometer ABL800 FLEX, a recently available whole blood creatinine assay. The Radiometer analyzer is the only whole blood creatinine method with calibration traceability to an IDMS standard and has previously demonstrated excellent agreement with laboratory creatinine assays using IDMS calibration. 8 We found that of the 3 devices, whole blood creatinine levels on the Radiometer correlated best with plasma creatinine levels on the Roche Cobas Integra 400. The Radiometer had a small negative bias compared with plasma creatinine, resulting in decreased sensitivity for the detection of plasma egfr less than 60 ml/min/1.73 m 2 compared with the i-stat method that overestimated plasma creatinine levels. However, specificity was improved on the Radiometer (compared with i-stat), and there was a trend toward improved overall concordance. Although an egfr less than 60 ml/min/1.73 m 2 is clearly a risk factor for contrast-induced nephropathy, 4,9 the risk of CIN increases in a graded manner as the egfr decreases from 60 to 20 ml/min/1.73 m 2. 9 One large study of CIN risk factors found that a serum creatinine level of more than 1.5 mg/dl (133 μmol/l) was a very strong risk factor for development of CIN. Use of the egfr instead of the serum creatinine level was also effective in identifying at-risk patients. However, patients with egfr values between 40 and 60 ml/min/1.73 m 2 were determined to be at lower risk for CIN compared with patients with lower egfr or serum creatinine levels of more than 1.5 mg/dl (133 μmol/l). 9 In fact, 1 large randomized trial that evaluated treatment options to prevent CIN enrolled only patients with serum creatinine levels of 2.0 mg/ dl (177 μmol/l) or more or egfr less than 40 ml/min/1.73 m 2. 5 Thus, it is clear that patients with egfr between 40 and 60 ml/min/1.73 m 2 and plasma or serum creatinine levels of 1.5 mg/dl (133 μmol/l) or less make up a relatively lower risk population of patients with renal impairment. In our study, all 13 patients with plasma egfr less than 60 ml/min/1.73 m 2 who were misclassified by the Radiometer whole blood assay had plasma egfr more than 45 ml/min/1.73 m 2 and plasma creatinine values of 1.5 mg/ dl (133 μmol/l) or less. Although other clinical risk factors for CIN were not identified in this study, the 13 misclassified patients with moderate renal insufficiency likely had relatively low risk of CIN. All patients with plasma egfr 45 ml/min/1.73 m 2 or less or plasma creatinine levels of more than 1.5 mg/dl (133 μmol/l) were correctly identified by the Radiometer whole blood assay. 924 Am J Clin Pathol 2009;132:920-926 Downloaded 924 from https://academic.oup.com/ajcp/article-abstract/132/6/920/1761213

Clinical Chemistry / Original Article Institutions that use an IDMS-calibrated laboratory creatinine assay may find that the Radiometer method is suitable for whole blood egfr determination before making contrast dosing decisions. Practices that want to maximize the sensitivity for detection of plasma egfr less than 60 ml/ min/1.73 m 2 might choose a device that slightly overestimates (as opposed to underestimates) plasma creatinine values. In weighing the use of the Radiometer device compared with the i-stat, the consequences of misclassifying a few patients with plasma egfr between 45 and 60 ml/min/1.73 m 2 and plasma creatinine levels of 1.5 mg/dl (133 μmol/l) or less (on the Radiometer) will need to be compared with the consequences of reducing contrast dose or changing treatment plans based on misclassification of patients with plasma egfr 60 ml/min/1.73 m 2 or more (on the i-stat). In our practice, patients with egfr less than 30 ml/min/1.73 m 2 will have contrast withheld, limiting the quality of the diagnostic image and potentially preventing a critical diagnosis from being made. Methods that systematically overestimate serum and plasma creatinine values may be less optimal for use in radiology practices that see many patients with severe renal disease because the frequency of falsely withholding contrast would be expected to increase. However, our study did not allow for direct comparison of methods in this patient population owing to small numbers of patients with egfr less than 30 ml/min/1.73 m 2. Institutions that use an IDMS-calibrated serum or plasma laboratory method and whole blood creatinine measurements on the same population of patients will need to determine the amount of confusion and inefficiency that might be caused by discordant egfr values. For example, if both laboratory and whole blood egfr results will remain in the medical record, will physicians who view these results after the radiologic procedure understand the reason for discrepancies in egfr classification? Physicians who view whole blood creatinine and egfr results after the radiology procedure will most likely consider the creatinine value and egfr before considering follow-up actions. For this reason, whole blood methods that minimize absolute bias (as opposed to just egfr concordance) may be more efficient for the medical institution as a whole. These practice issues may determine the relative importance of sensitivity, specificity, and overall concordance for a given hospital or laboratory. The StatSensor, another new device for measurement of whole blood creatinine levels, demonstrated lower concordance with the plasma creatinine reference method. However, the StatSensor offers users the option of entering an offset to match StatSensor whole blood to local laboratory creatinine values. In theory, this offset feature could be used to optimize the sensitivity or the specificity for detection of plasma egfr less than 60 ml/min/1.73 m 2. Similar to the Radiometer method, all patients with plasma egfr less than 60 ml/ min/1.73 m 2 misclassified by the StatSensor (with offset) fell into a relatively low-risk group for CIN (plasma egfr >40 ml/min/1.73 m 2 and plasma creatinine level 1.5 mg/dl [133 μmol/l]). In addition, 1 study found that concordance between egfr determined from whole blood and plasma creatinine values was better when capillary blood was used to dose the StatSensor compared with heparinized whole blood. 10 This device may be better suited for use with capillary whole blood. The cost-benefit of using whole blood creatinine assays for risk stratification of patients before contrast administration will likely vary with practice volume and environment. In our practice, the whole blood method decreases turnaround time for creatinine results by approximately 20 minutes. Assuming average patient throughput for our practice and the average number of patients per day who might wait for creatinine results before CT, we estimate that reducing turnaround time by 20 minutes would allow for 1 to 3 additional patients to have CT scans performed per day. Reagent cost is increased by $1 to $5 per test using whole blood creatinine methods. Given our current volume of stat creatinine testing, this would increase laboratory costs by approximately $650 to $3,250 per month. However, this would be offset by an expected revenue increase of approximately $1,000 to $3,000 per day, assuming that decreased creatinine turnaround time resulted in 1 to 3 additional patients receiving CT scans. Thus, use of whole blood creatinine assays may be appropriate in institutions with a high volume of CT scans or procedures requiring contrast administration. Limitations There were few patients with severe (plasma egfr <30 ml/min/1.73 m 2 ) renal insufficiency in our study population, and this likely adversely affected the observed sensitivity of the Radiometer and StatSensor methods that underestimated plasma creatinine values. The patient population was also older (mean age, 68 years), and this may limit the usefulness of egfr estimation regardless of method. In addition, clinical risk factors for CIN such as diabetes or history of previous renal disease or transplantation were not known for our study population; thus, the clinical risk of CIN could not be fully evaluated for each patient in the study. Conclusion We compared the i-stat, StatSensor, and Radiometer ABL800 FLEX whole blood creatinine assays with an IDMS-calibrated plasma creatinine method. The Radiometer device demonstrated the best overall correlation to plasma creatinine levels and best clinical concordance when creatinine values were used to calculate egfr. The i-stat Downloaded from https://academic.oup.com/ajcp/article-abstract/132/6/920/1761213 Am J Clin Pathol 2009;132:920-926 925 925 925

Korpi-Steiner et al / Whole Blood Creatinine for egfr systematically overestimated plasma creatinine levels and, thus, has improved sensitivity but decreased specificity for determination of egfr less than 60 ml/min/1.73 m 2 compared with the Radiometer. The StatSensor significantly underestimated plasma creatinine levels but offers users the ability to enter a slope and an intercept offset to match whole blood to plasma creatinine levels, and this feature significantly improved sensitivity for the detection of plasma egfr less than 60 ml/min/1.73 m 2. From the Departments of 1 Laboratory Medicine and Pathology and 2 Radiology, Mayo Clinic, Rochester, MN. Address reprint requests to Dr Karon: Dept of Laboratory Medicine and Pathology, Mayo Clinic, 200 First St SW, Rochester, MN 55905. Acknowledgments: We acknowledge Nova Biomedical for use of the StatSensor and supplies and Radiometer for use of the ABL800 FLEX and supplies. We thank Amy Wockenfus, Chris Koch, and Jan Lakin for technical assistance. References 1. Coresh J, Astor B, Greene T, et al. Prevalence of chronic kidney disease and decreased kidney function in the adult US population: Third National Health and Nutrition Examination Survey. Am J Kidney Dis. 2003;41:1-12. 2. Levey A, Coresh J, Greene T, et al. Expressing the Modification of Diet in Renal Disease study equation for estimating glomerular filtration rate with standardized serum creatinine values. Clin Chem. 2007;53:766-772. 3. Myers G, Miller W, Coresh J, et al. Recommendations for improving serum creatinine measurement: a report from the Laboratory Working Group of the National Kidney Disease Education Program. Clin Chem. 2006;52:5-18. 4. Schweiger M, Chambers C, Davidson C, et al. Prevention of contrast induced nephropathy: recommendations for the high risk patient undergoing cardiovascular procedures. Catheter Cardiovasc Interv. 2007;69:135-140. 5. Briguori C, Airoldi F, D Andrea D, et al. Renal insufficiency following contrast media administration trial (REMEDIAL). Circulation. 2007;115:1211-1217. 6. Nichols J, Bartholomew C, Bonzagi A, et al. Evaluation of the IRMA TRUpoint and i-stat creatinine assays. Clin Chim Acta. 2007;377:201-205. 7. Gault M, Seymour M, Howell W. Evaluation of the i-stat creatinine assay. Nephron. 2001;88:178-182. 8. Skurup A, Kristensen T, Wennecke G. New creatinine sensor for point-of-care testing creatinine meets the National Kidney Disease Education Program guidelines. Clin Chem Lab Med. 2008;46:3-8. 9. Mehran R, Aymong E, Nikolsky E, et al. A simple risk score for prediction of contrast-induced nephropathy after percutaneous coronary intervention. J Am Coll Cardiol. 2004;44:1393-1399. 10. Becker C, Michaely HJ, Reiser M, et al. Evaluation of a handheld-creatinine measurement device for realtime bedside determination of whole blood creatinine in radiology. Point Care. 2008;7:125. doi:10.1097/01. POC.0000335918.60912.39. 926 Am J Clin Pathol 2009;132:920-926 Downloaded 926 from https://academic.oup.com/ajcp/article-abstract/132/6/920/1761213