AppNote 11/2008. Analysis of Drugs and Metabolites in Blood and Urine using Automated Disposable Pipette Extraction KEYWORDS ABSTRACT

Similar documents
AppNote 3/2013. Automated Determination of Total Fat, Saturated Fat, Monounsaturated Fat and Trans Fat Content in Food Samples KEYWORDS ABSTRACT

Quantitative Determination of Terpenes in Cannabis Using Headspace Solid Phase Microextraction and GC/MS

Automated Hydrolysis, Extraction and Determination of Opioids in Urine using a Novel Robotic Autosampler and LC-MS/MS Platform

AppNote 5/2010. A Novel Automated Blood Analyzer for the Determination of Immunosuppressants in Whole Blood INTRODUCTION

AppNote 3/2008. Determination of Aldehydes and Ketones in Oily Matrices using a Novel Dynamic Headspace Sampler Coupled to GC/MS INTRODUCTION

AppNote 1/2005. Blood Alcohol Analysis Using an Automated Static Headspace Method ABSTRACT

AppNote 8/2007. Rapid Multidimensional GC Analysis of Trace Drugs in Complex Matrices KEYWORDS ABSTRACT

Direct Thermal Extraction Analysis of Food Packaging Material

AppNote 4/2015. Automated Desorption, SPE Extraction, and LC/MS/MS Analysis of Dried Blood Spots KEYWORDS ABSTRACT

AppNote 4/2002. Stir Bar Sorptive Extraction: Capacity and Competition Effects KEYWORDS ABSTRACT. SBSE, SPME, capacity

Drugs of Abuse in Oral Fluids: Automated SPE Extraction and LC/MS/MS Determination using a Robotic Autosampler

AppNote 1/2015. Automated Sample Preparation for Metabolomics Studies Using the Gerstel MPS Dual Head WorkStation

KEYWORDS ABSTRACT. 3-MCPD, Glycidol, edible oil, lab automation, ISO , AOCS Cd 29c-13, DGF C-VI 18 (10)

KEYWORDS ABSTRACT AppNote 3/2002

AppNote 10/2002. Detection of Spoilage Markers in Food Products using a Mass-Spectrometry Based Chemical Sensor KEYWORDS ABSTRACT

Rapid and Robust Detection of THC and Its Metabolites in Blood

A Simple and Accurate Method for the Rapid Quantitation of Drugs of Abuse in Urine Using Liquid Chromatography

AppNote 2/2014. Fully Automated SPE-GC/MS Determination of Δ9-Tetrahydrocannabinol (THC) and its Metabolites in Serum Samples KEYWORDS ABSTRACT

AppNote 3/2001. Elimination of Polar Matrix Components Prior to GC Analysis using Stir Bar Sorptive Extraction (SBSE) KEYWORDS ABSTRACT

E XCEL LENCE JUST GOT BETTER UCT FORENSICS CLEAN SCREEN XCEL SPE COLUMNS

High Throughput Extraction of Opiates from Urine and Analysis by GC/MS or LC/MS/MS)

Analyze Drugs of Abuse with Agilent J&W Ultimate Plus Tubing in an Inert Flow Path

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Extraction of 11-nor-9-carboxy-tetrahydrocannabinol from Hydrolyzed Urine by ISOLUTE. SLE+ Prior to GC/MS Analysis

Application Note. Abstract. Authors. Pharmaceutical

A Comprehensive Screening of Illicit and Pain Management Drugs from Whole Blood Using SPE and LC/MS/MS

4.5 Minute Analysis of Benzodiazepines in Urine and Whole Blood Using LC/MS/MS and an Ultra Biphenyl Column

AppNote 5/2012. Qualitative Analysis of Coconut Water Products using Stir Bar Sorptive Extraction Combined with Thermal Desorption-GC/MS KEYWORDS

Shuguang Li, Jason Anspach, Sky Countryman, and Erica Pike Phenomenex, Inc., 411 Madrid Ave., Torrance, CA USA PO _W

Comparison of Different Whole Blood Sample Pretreatment Methods for Targeted Analysis of Basic Drugs

Simultaneous Determination of Prescription and Designer Benzodiazepines in Urine and Blood by SPE and LC-MS/MS

Quantitative Determination of Drugs of Abuse in Human Plasma and Serum by LC/MS/MS Using Agilent Captiva EMR Lipid Cleanup

A Rapid Method for Detection of Drugs of Abuse in Blood Samples Using the Thermal Separation Probe and the 5975T LTM GC/MS

High-Throughput GC/MS Confirmation and Quantitation of Codeine and Morphine in Urine Using the DSQ II

The Development of LC/MS Methods for Determination of MDMA (Ecstasy) and Metabolites in Biological Samples

Rapid Hydrolysis of Benzodiazepines in Urine. Alicia Zook 1 and Crystal Xander B.S. 2. Cedar Crest College, Allentown, PA 1

LC Application Note. Dangerous driver?

Life Science and Chemical Analysis Centre. John Heim

Fast and easy separation of 23 drugs of abuse. including high, stable resolution of isobaric opioids from human urine by UHPLC-MS/MS

Detection of Cannabinoids in Oral Fluid with the Agilent 7010 GC-MS/MS System

Automated Sample Preparation for Profiling Fatty Acids in Blood and Plasma using the Agilent 7693 ALS

AppNote 9/2000. Flavor Profi ling of Different Olive Oils with Rancidity-Monitoring by Thermal Extraction GC/MS KEY WORDS ABSTRACT

using the Agilent 7696A Sample Prep

Protein Precipitation for Biological Fluid Samples Using Agilent Captiva EMR Lipid 96-Well Plates

2-Step Multi-Volatile Method (2-Step MVM) for Characterization of Aroma Compounds in Bread

BENZODIAZEPINE FINDINGS IN BLOOD AND URINE BY GAS CHROMATOGRAPHY AND IMMUNOASSAY

Keep It Clean. Solid Phase Extraction Solutions for Drugs of Abuse Analysis. Neutral Drugs of Abuse Basic Drugs of Abuse

AppNote 4/1994. Identification of Furan Fatty Acids as Minor Components of Human Lipids by Multidimensional GC-MSD

Determination of Bath Salts (Pyrovalerone Analogs) in Biological Samples

Ultrafast Analysis of Benzodiazepines in Urine by the Agilent RapidFire High-Throughput Triple Quadrupole Mass Spectrometry System

Determination of Gamma-Hydroxy-Butyrate (GHB) in Biological Samples

Innovations in Large Volume Injection

Validation of an Automated Method to Remove

Automated Targeted screening of Benzodiazepines in urine using LDTD-MS/MS at 400 samples per hour rate

Analysis of short chain organic acids in tobacco and in some flavored e-liquids

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No.

Offline and online sample extraction for the quantification of benzodiazepines in human plasma or serum for clinical research

Application Note. Author. Abstract. Introduction. Food Safety

Modified QuEChERS Procedure for Analysis of Bisphenol A in Canned Food Products

Application. Detection of Cannabinoids in Oral Fluid Using Inert Source GC/MS. Introduction. Authors. Abstract. Forensic Toxicology

Quantitative Determination of Drugs of Abuse in Human Whole Blood by LC/MS/MS Using Agilent Captiva EMR Lipid Cleanup

Application. Authors. Abstract. Introduction. Forensics

Quantitative Analysis of Opiates in Urine Using RRHT LC/MS/MS. Application. Authors. Introduction. Abstract. Forensics

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

LC-MS/MS Method for the Determination of Tenofovir from Plasma

Proof of Concept for Automated SPE/HPLC/MS/MS Methods to Replace Traditional Immunoassay with MS Confirmation of Driving Under the Influence Samples

Author. Introduction. Abstract

Screening by immunoassay and confirmation & quantitation by GC-MS of buprenorphine and norbuprenorphine in urine, whole blood and serum

Introduction. Abstract

Improving Benzodiazepine Immunoassay Sensitivity by Rapid Glucuronide Hydrolysis Technology

Analyze Barbiturates in Urine with Agilent 6430 LC/MS/MS and Poroshell 120 EC-C18

Overview. Introduction. Experimental. Cliquid Software for Routine LC/MS/MS Analysis

Validation Report for the Neogen Fentanyl Kit for ELISA Screening of Whole Blood and Urine Specimens

GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants

Application Note LCMS-108 Quantitation of benzodiazepines and Z-drugs in serum with the EVOQ TM LC triple quadrupole mass spectrometer

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Validation of a Benzodiazepine and Z-Drug Method Using an Agilent 6430 LC/MS/MS

Strata -X Polymeric SPE

AppNote 5/2013. Thermochemolysis A Simple and Rapid Methylation Method Based on TMAH for Gas Chromatographic Analysis of Linseed Oil and Amber

Analysis of HMF by HPLC

Overcoming the Challenges of Sample Preparation and LC/MS/MS Method Development for Clinical Applications

Rapid Analysis of 37 FAMEs with the Agilent 8860 Gas Chromatograph

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid

Enrichment of Phospholipids from Biological Matrices with Zirconium Oxide-Modified Silica Sorbents

AS Application Note 1602

Determination of red blood cell fatty acid profiles in clinical research

BETA-GLUCURONIDASE PRODUCT LINE INNOVATION THROUGH CHEMISTRY. FORENSICS

Toxicology Screening of Whole Blood Extracts Using GC/Triple Quadrupole/MS

25-Hydroxyvitamin D2-D3 Serum VD-200 UHPLC Analysis Kit User Manual

Automated Sample Preparation for FAME Analysis in Edible Oils Using an Agilent 7696A Sample Prep WorkBench

Liquid Liquid Extraction of Gamma Hydroxybutyric Acid (GHB) in Blood and Urine for GC-MS analysis

The challenging extraction of non-polar contaminants out of a non-polar vegetable oil sample

25-Hydroxyvitamin D2-D3 Serum LC-MS/MS APCI Analysis Kit User Manual

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine

Simultaneous Analysis of Active Pharmaceutical Ingredients and Their Counter-Ions Using a Mixed-Mode Column

Determination of Patulin in Apple Juice Using SPE and UHPLC-MS/MS Analysis

Evaluation of an LC-MS/MS Research Method for the Analysis of 33 Benzodiazepines and their Metabolites

Evaluation of the Impact of Expanding ELISA Screening in DUID Investigations. Aileen Lu*, Karen S. Scott, Aya Chan-Hosokawa, and Barry K.

Transcription:

AppNote 11/2008 Analysis of Drugs and Metabolites in Blood and Urine using Automated Disposable Pipette Extraction Fred Foster, Edward E. Pfannkoch, John R. Stuff, Jacqueline A. Whitecavage Gerstel, Inc., 701 Digital Drive, Suite J, Linthicum, MD 21090, USA Sparkle T. Ellison, Stephen L. Morgan, William E. Brewer Department of Chemistry and Biochemistry, University of South Carolina, 631 Sumter Street, Columbia, SC 29208, USA KEYWORDS Disposable pipette extraction (DPX), solid-phase extraction (SPE), large volume injection (LVI), multi-purpose sampler (MPS), sample preparation ABSTRACT In this work an automated DPX procedure is presented that enables the determination of acidic, basic, and neutral drugs in small blood or urine samples. The DPX method is based on a mixed-mode cation exchange sorbent (CX) with reversed phase characteristics. The automated extraction is shown to require only minimal manual labor when performed using the GERSTEL MPS 2 autosampler under MAESTRO software control. GC-MS data is provided showing that comprehensive, rapid and sensitive analysis can be performed using automated DPX combined with large volume injection. Automated sample preparation steps are performed during the GC/MS run of the preceding sample using the MAESTRO PrepAhead function. PrepAhead sample preparation enables high throughput analysis, improved productivity and just in time preparation of each sample.

INTRODUCTION The analysis of blood or urine for the presence of drugs by chromatographic methods typically requires sample preparation prior to analysis. Solid Phase Extraction (SPE) methods with various adsorbents in traditional cartridge formats are often used for this sample cleanup, however traditional SPE requires relatively large volumes of solvents and is performed in multiple steps. After sample cleanup, a solvent evaporation step is often required to achieve the necessary detection limits. Finally, many analyte types require derivatization prior to GC analysis. This sample preparation is typically time-consuming and is widely considered to be the bottle neck for laboratory analysis throughput. Disposable Pipette Extraction (DPX) was developed as an efficient and rapid extraction method alternative to conventional solid-phase extraction. The patented DPX tips comprise loose solid phase sorbent contained inside a pipette tip fitted with a lower screen and upper barrier. These tips do not require conditioning because the solution is mixed with the sorbent; channeling and flow rates have no impact on extraction efficiency for this type of dispersive solid-phase extraction technology. Therefore, extractions are much faster and require much less solvent. The barrier mounted in the upper opening of the DPX tip additionally acts as a transport adaptor, enabling the GERSTEL MPS 2 to move the tip around (patent pending). After DPX sample cleanup, the time-consuming solvent evaporation step can be eliminated by using Large Volume Injection (LVI) to evaporate the solvent and concentrate the analytes directly in the GC inlet liner. For compounds that do not require derivatization, LVI provides the most efficient means to achieve high sample throughput combined with low detection limits. Some compounds, such as benzodiazepines and opiates, must be derivatized prior to GC analysis. Two different methods were used to perform this task: Analytes were either derivatized in the sample vial prior to injection or directly in the inlet by delivering reagent into the GC inlet immediately following the sample using a sandwich injection. Both derivatization methods are performed automatically by the MPS 2 autosampler. This research was performed with the aim to develop a faster, better and less labor intensive method for analysis of drugs and metabolites in blood and urine by GC/MS. Using the LVI technique, limits of detection comparable to those found in typical HPLC/MS/MS applications were achieved. One of the AN/2008/11-2 primary goals of this research was to provide just in time sample preparation. Every sample is prepared at the exact same point in time, just prior to analysis. This reduces sample to sample variation by reducing the risk of analyte decomposition since derivatized samples are not waiting for a long period of time in the autosampler prior to injection. Furthermore, each sample is extracted and prepared for injection while the previous sample is being chromatographically analyzed, providing the highest throughput possible. EXPERIMENTAL Materials. BSTFA and MTBSTFA and ß-Glucuronidase (Type H-2, from Helix pomatin) were purchased from Sigma Aldrich (St. Louis, MO). DPX-CX 1 ml tips were obtained from DPX Labs, LLC (Columbia, SC). Acetonitrile, ethyl acetate, hexane, methylene chloride, isopropanol, ammonium hydroxide, and hydrochloric acid (all analytical grade) were purchased from Fisher Scientific (Pittsburg, PA). A drug-free urine sample was obtained from a lab volunteer. Ampoules of 6-monoacetylmorphine (6-MAM, 0.1 mg/ml), 6- MAM-d 3 (0.1 mg/ml), morphine (1.0 mg/ml), morphine-d 3 (0.1 mg/ml), codeine (0.1 mg/ml), codeine-d 3 (0.1 mg/ml), oxycodone (1.0 mg/ml), oxycodone-d 3 (0.1 mg/ml), PCP-d 5 (0.1 mg/ml), and pentobarbital-d 5 (0.1 mg/ml) were purchased from Cerilliant (Round Rock, TX). Quick-check drug solutions (0.5 mg/ml) containing 20 acidic, neutral and basic general drugs were purchased from Alltech (State College, PA). Instrumentation. Analyses were performed using a 6890N GC equipped with a 5975 (inert XL) mass selective detector (Agilent Technologies), PTV type inlet Cooled Injection System (CIS 4, Gerstel) and MPS 2 Prepstation with DPX option under MAESTRO software control (Gerstel). Analysis conditions. PTV: 1 min solvent vent (150 ml/min) splitless 80 C (1 min); 12 C/min; 300 C (3 min) Column: 30 m HP 5 (Agilent) d i = 0.25 mm d f = 0.25 μm Pneumatics: He, constant flow = 1.5 ml/min Oven: 100 C (0.5 min); 20 C/min; 300 C (12.5 min) MSD: Full Scan or SIM mode

SIM parameters for the determination of benzodiazepines. Group Analyte Target Ions (m/z) Qualifi er Ions (m/z) 1 Diazepam 256 283, 284 1 Nordiazepam-TBDMS 327 328, 329, 332* 2 Flunitrazepam 312 286, 294 2 7-aminofl unitrazepam 255 283, 254 2 Oxazepam-2TBDMS 457 513, 459 3 Nitrazepam-TBDMS 338 292, 394 3 Temazepam-TBDMS 357 283, 359 3 Lorazepam-TBDMS 491 513, 493 4 Clonazepam-TBDMS 372 374, 326 4 Alprazolam 279 204, 308 5 α-oh-alprazolam- TBDMS SIM parameters for the determination of opiates. Group Analyte Target Ions (m/z) 381 382, 383, 386* Qualifi er Ions (m/z) 1 Hydrocodone 299 242, 302* 2 Codeine-TMS 371 234, 178, 374* 3 Morphine-2TMS 429 236, 432* 4 6-MAM-TMS 399 340, 402* 5 Oxycodone-TMS 387 390* Initial Sample Preparation. Deuterated internal standards were added to the sample solutions. D 5 -PCP was added to the basic blood samples, bringing the final concentration to 0.2 ppm. For the benzodiazepine determination, d 5 -nordiazepam and d 5 -OH-alprazolam were added bringing the final concentration to 0.2 ppm. For opiates, deuterated analogues of each drug at a final concentration of 0.1 ppm were added. Preparing blood samples: 0.25 ml of whole blood sample was vortex mixed with 0.5 ml acetonitrile. The samples were subsequently centrifuged and the supernatant transferred to clean labeled tubes, each containing 0.1 ml of 0.1 M HCl. Preparing urine samples for benzodiazepine determination: Hydrolysis was performed by adding 10 μl ß-glucuronidase and 50 μl of 0.1 M sodium phosphate buffer ph 4 to 0.2 ml urine and then heating samples at 55 C for 2 hours. The samples were then cooled to room temperature and 0.25 ml acetonitrile was added to precipitate the enzyme. AN/2008/11-3 The solutions were centrifuged, and the resulting supernatants transferred to clean, labeled sample tubes. 200 μl of 0.1 M HCl was then added to the supernatant in each sample. All tubes were then placed onto the GERSTEL MPS 2 sample tray. Automated DPX extraction procedure. Automated DPX was performed using 1 ml DPX-CX (mixedmode) tips from DPX Labs, LLC. 250 μl of 30 % acetonitrile in water was dispensed from above onto the DPX sorbent. The DPX tip was inserted into the sample solution, which was aspirated and mixed with the sorbent. After equilibrating for 30 seconds, the solution was dispensed to waste. The sorbent was washed with 0.5 ml of 30 % acetonitrile in water and then with 0.5 ml of 100 % acetonitrile. Both wash solvents were dispensed to waste. Elution of the analytes was accomplished using 0.7 ml of elution solvent (2 % NH 4 OH : 78 % CH 2 Cl 2 : 20 % isopropanol). The eluent was dispensed directly into the pre-labeled GC vial. The total time required for extraction and liquid handling was approximately 6 minutes per sample. Automated injection with concentration and derivatization steps. The derivatizing reagent contained 50 % MTBSTFA in acetonitrile for the analysis of benzodiazepines. A sandwich injection of derivatizing reagent and eluent was performed by the left tower of the MPS 2 by first aspirating 20 μl of the derivatizing reagent, then aspirating 20 μl of air, then aspirating 50 μl of the eluent from the DPX extraction. The injection was performed relatively slowly at 1.32 μl/s. This allowed the solvent to evaporate from the inlet through the split vent at approximately the speed of injection, eliminating any risk of flooding. The solvent was almost entirely kept outside the GC column and analytes were concentrated inside the injection port. Derivatization in the sample vial. The eluent from the DPX extraction was dried with nitrogen, then 50 μl of MTBSTFA or BSTFA and 50 μl ethyl acetate were added. The solutions were heated at 70 C for 20 minutes. After cooling, each solution was introduced directly into the inlet using large volume injection (50 μl).

RESULTS AND DISCUSSION The DPX method was chosen for automated extractions because it can be performed in just a few minutes (6 min). Rather than extract and prepare a large number of samples in one batch and letting them wait for analysis with the associated risk of decomposition, only one sample is extracted at a time. Sample extraction is performed just in time during the GC or LC run of the previous sample. When the system becomes ready for the next injection, the next sample has been prepared and is ready to be introduced. This provides the highest possible throughput and system utilization while ensuring that all samples are treated exactly the same. The GC/MS or LC/MS system is never waiting idly for the next sample. Figure 1 shows a picture of the dual rail GERSTEL MPS 2 PrepStation that was used for the automated DPX and large volume injection (LVI). Figure 1. A picture of the dual rail GERSTEL MPS 2 with DPX automation platform. The left MPS is equipped with a 100 ul syringe for injecting eluent and derivatizing reagents, and the right MPS is equipped with a 2.5 ml syringe for performing DPX extractions. The automated extractions were performed using DPX-CX tips, which have previously been shown [1] to provide high recoveries of acidic, neutral and basic drugs in spiked urine (Table 1). It should be noted that the spiked urine samples in the previous study were not hydrolyzed. Some urine samples require hydrolysis in Table 1. Statistical data of basic, acidic, and neutral drugs extracted by DPX-CX with just 0.2 ml of nonhydrolyzed urine spiked at 0.5 ppm. Compound Ion [m/z] Internal Standard Recovery AN/2008/11-4 RSD R2 *LOD* [µg/ml] Butabarbital 156 Pentobarbital-d 5 96.3 2.63 0.9988 0.0120 Amobarbarbital 156 Pentobarbital-d 5 99.6 2.57 0.9988 0.0106 Pentobarbital 156 Pentobarbital-d 5 98.6 1.58 0.9985 0.0152 Secobarbital 168 Pentobarbital-d 5 97.6 1.41 0.9995 0.0261 Phenobarbital 204 Pentobarbital-d 5 95.3 2.54 0.9988 0.0129 Glutethimide 189 Pentobarbital-d 5 99.4 2.74 0.9980 0.0216 Merpridine 247 PCP-d 5 93.2 1.43 0.9931 0.0099 Methadone 72 PCP-d 5 96.4 2.65 0.9992 0.0051 Methaqualone 235 PCP-d 5 99.4 1.85 0.9902 0.0134 Amitriptyline 58 PCP-d 5 95.6 1.99 0.9990 0.0038 Cocaine 182 PCP-d 5 103 1.31 0.9961 0.0146 cis-doxepin 58 PCP-d 5 98.5 1.18 0.9988 0.0200 Imipramine 234 PCP-d 5 96.0 3.72 0.9984 0.0099 trans-doxepin 58 PCP-d 5 98.5 2.30 0.9987 0.0146 Pentazocine 217 PCP-d 5 95.7 4.37 0.9946 0.0104 Codeine 299 PCP-d 5 105 6.85 0.9864 0.0520 Oxycodone 315 PCP-d 5 111 8.80 0.9761 0.0638 Desipramine 234 PCP-d 5 91.5 9.60 0.9522 0.0742 PCP 200 PCP-d 5 96.8 1.04 0.9992 0.009

order to enable detection of various metabolites. We found that using high buffer concentrations (such as 1 M) for the hydrolysis resulted in poor analyte recovery. We therefore used 0.1 M buffer, which provided ample buffer capacity for hydrolysis of urine specimens. Whole blood samples are too viscous to be aspirated through the screens of the DPX tips. These samples were first protein precipitated using acetonitrile and then centrifuged, after which the samples could readily be extracted using DPX technology. When relying on a conventional split/splitless inlet, only a small volume of eluent can be introduced to the GC without impacting the quality of the analysis. This means that a concentration step is often needed prior to injection in order to achieve the required low detection limits. Performing evaporative concentration can be time consuming, even when only small elution volumes are involved.. Instead, we chose to perform Large Volume Injection (LVI) using the GERSTEL Cooled Injection System (CIS). A large volume (50 μl) of the DPX eluent was introduced directly into the CIS and the solvent was evaporated through the split vent, leaving the concentrated analytes in the inlet. To use LVI successfully for high throughput analysis, the eluent must be fairly clean. Otherwise, matrix residue will accumulate in the inlet leading to reduced analyte recovery due to adsorption or breakdown, which will have a negative effect on sensitivity and reproducibility If significant matrix build-up in the inlet liner is experienced, automated liner exchange can be performed after a user defined number of injections. This is possible when using the MPS 2 autosampler equipped with Automated Liner EXchange (ALEX) option. In this case, fortunately, the extracts from the DPX-CX method were very clean; multiple injections could be made with good reproducibility without replacing the inlet liner. Figure 2 shows total ion chromatograms from the 5th and the 20th injection respectively in a series of injections of whole blood extracts produced using automated DPX. Resulting peak intensities were shown to be comparable. Minimal decline in performance was experienced after approximately 50 such injections. 1) meperidine 2) PCP-d 5 (ion 205) 3) PCP 4) methadone 5) methaqualone 6) amitriptyline 7) cocaine 8) cis-doxepin 9) imipramine 10) trans-doxepin 11) desipramine 12) pentazocine 13) codeine Figure 2. Total ion chromatograms of basic drugs extracted from 0.25 ml of whole blood spiked at 0.5 ppm. The top chromatogram represents the 20th large volume injection of 50 μl of whole blood extracts, and the bottom chromatogram represented the 5th injection. *-Denotes septum bleed from the vial cap from multiple injections. AN/2008/11-5

It should be noted that the chromatograms shown in Figure 2 are obtained from extracts of just 0.25 ml whole blood spiked at 0.5 ppm. Furthermore, the data was acquired in FULL SCAN mode using a GC/MS instrument (Table 2). Based on this small sample volume, limits of detection below 0.1 ppm were obtained for all determined drugs. SIM analysis would allow the limits of detection to be reduced even further. The limited amount of time required per sample makes it possible to perform high throughput analysis one sample at a time, which again means that samples are not prepared until they can be injected into the analysis system, reducing the wait-time for prepared samples as well as the risk of analyte decomposition. Many analytes require chemical derivatization prior to GC analysis. The CIS inlet provides an inert and heated environment that is well suited for chemical derivatization. Additionally, it has the capability to purge and remove volatile compounds such as the solvent eluent and reagent to ensure that these are not introduced to the GC column. We decided to perform both solvent evaporation and analyte derivatization in the CIS inlet. To derivatize benzodiazepines, 20 μl of 50 % MTBSTFA in acetonitrile was aspirated into the the injection syringe followed by 20 μl of air and finally 50 μl of eluent from the automated DPX sample preparation. The resulting sandwich injection was performed slowly in a programmed stopflow procedure to ensure that the solvent was purged Table 2. Statistical data of basic drugs extracted from whole blood using FULL SCAN GC/MS analysis.. Compound Ion [m/z] Internal Standard Recovery *C.V.* Meperidine 247 PCP-d 5 94.1 9.20 PCP 200 PCP-d 5 72.5 2.20 Methadone 72 PCP-d 5 69.7 6.30 Methaqualone 235 PCP-d 5 78.9 7.70 Amitriptyline 58 PCP-d 5 70.3 7.20 Cocaine 182 PCP-d 5 67.9 7.70 cis-doxepin 58 PCP-d 5 78.2 8.80 Imipramine 234 PCP-d 5 64.3 6.30 trans-doxepin 58 PCP-d 5 110 3.20 Desipramine 234 PCP-d 5 52.0 14.1 Pentazocine 217 PCP-d 5 86.9 16.8 Codeine 299 PCP-d 5 67.0 9.60 through the split vent prior to introduction of the derivatizing reagent. After the derivatizing reagent had been introduced a further purge followed. The split vent was then closed and the temperature ramped rapidly to 300 C to derivatize and transfer analytes to the GC column in the splitless mode. Derivatization in the inlet was successful for several benzodiazepines in blood, but a few benzodiazepines were not easily derivatized using this method and derivatization of these was therefore performed in the sample vial. Figure 3 shows results for benzodiazepines extracted from urine using the automated DPX method followed by derivatization in the sample vial. It is important to note that only 0.2 ml of urine sample was required for analysis at 0.2 ppm. LODs 1) diazepam 2) nordiazepam 3) d 5 -Nor TBDMS 4) fl unitrazepam 5) 7-aminofl unitrazepam 6) oxazepam-2tbdms 7) temazepam-tbdms 8) lorazepam-2tbdms 9) nitrazepam-tbdms 10) clonazepam-tbdms 11) alprazolam 12) OH-alprazolam 13) d 5 -OH-alprazolam- TBDMS Figure 3. Extracted ion chromatograms of benzodiazepines extracted from 0.2 ml of hydrolyzed urine sample at 0.2 ppm. This derivatization was performed with MTBSTFA. AN/2008/11-6

for most of the benzodiazepines were approximately 1 ng/ml and LODs were below 20 ng/ml for all 22 drugs determined (Table 3). Extraction of opiates using the automated DPX method was successful and resulted in high recoveries for all compounds in the test. For opiates, inlet derivatization using BSTFA was attempted, but proved unsuccessful. Figure 4 shows chromatograms of opiates from whole blood samples, and Table 4 shows the statistical results from the opiate analysis. The majority of the resulting limits of detection were calculated to be less than 1 ng/ml based on analysis of just 0.25 ml of whole blood sample. Derivatization of opiates was performed in the sample vial rather than in the GC inlet, the process was automated using the MPS 2. Table 3. Statistical data of benzodiazepines extracted from hydrolyzed urine samples and derivatized with MTBSTFA. Compound Recovery Diazepam 96.7 Nordiazepam 82.1 Flunitrazepam 35.6 Oxazepam 54.4 Temazepam 52.0 Lorazepam 15.8 Alprazolam 62.1 OH-Alprazolam 71.3 7-Amino-fl unitrazepam 38.3 Nitrazepam 79.1 Clonazepam 52.3 1) Codeine-d 3 -TMS 2) Codeine-TMS 3) Morphine-d 3-2TMS 4) Morphine-2TMS 5) 6 MAM-d 3 TMS 6) 6 MAM-TMS 7) Oxycodone-d 3 TMS 8) Oxycodone-TMS Figure 4. Extracted ion chromatograms of opiates extracted from 0.25 ml of whole blood at 0.1 ppm. This derivatization was performed with BSTFA. Table 4. Statistical data of opiates extracted from whole blood using just 0.25 ml with extracts derivatized with BSTFA. Compound Ion [m/z] External Standard Recovery Hydrocodone 299 Hydrocodone-d 3 99.9 Codeine-TMS 371 Codeine-d 3 89.3 Morphine-2TMS 429 Morphine-d 3 85.6 6-MAM-TMS 399 6-MAM-d 3 87.7 Oxycodone-TMS 387 Oxycodone-d 5 86.2 Minimal sample pre-treatment was required prior to automated DPX. Whole blood specimens required protein precipitation with acetonitrile, while urine specimens required hydrolysis. The rest of the analysis was completely automated using a dual rail GERSTEL MPS 2 PrepStation with a cooled injection system (CIS) used for large volume injection (LVI). The automated extractions were performed using 1 ml DPX-CX (mixed-mode) tips. The time required per sample for complete DPX extraction was approximately 6 minutes. AN/2008/11-7

CONCLUSIONS The DPX technique enables rapid sample preparation and it is readily automated using the GERSTEL MPS 2; the MAESTRO software makes it easy to set up the DPX methods for Ready for Analysis sample preparation. DPX was used successfully to determine numerous basic drugs that don t require derivatization, such as cocaine, methadone, PCP, TCAs, and meperidine. Most benzodiazepines could readily be determined using derivatization with MTBSTFA in the GC inlet. These include diazepam, nordiazepam, oxazepam, temazepam, alprazolam, and α-oh-alprazolam. Automated DPX combined with GC/MS was shown to provide excellent results for the determination of 11 benzodiazepines in urine. High recoveries and high sensitivity was shown for the determination of opiates; for most of these, limits of detection were below 1 ng/ml in whole blood. The GERSTEL CIS inlet was shown to provide excellent results, when used for combined concentration and derivatization steps during the injection process. Automated DPX and sample introduction using the dual rail GERSTEL MPS 2 PrepStation provides the highest throughput available. REFERENCES [1] S. T. Ellison, W. E. Brewer, and S. L. Morgan, Comprehensive Analysis of Drugs and Metabolites in Urine with Disposable Pipette Extraction, PittCon 2008, New Orleans, LA. ACKNOWLEDGEMENTS The authors would like to acknowledge the support of the following organizations: Dept. of Chemistry and Biochemistry, University of South Carolina, Columbia, SC DPX Labs, LLC, Columbia, SC GERSTEL GmbH & Co. KG, Mülheim an der Ruhr, Germany AN/2008/11-8 For research use only. Not for use in diagnostic procedures. The information provided for this product is intended for reference and research purposes only. GERSTEL offers no guarantee as to the quality and suitability of this data for your specific application. Information, descriptions and specifications in this publication are subject to change without notice.

AN/2008/11-9

GERSTEL GmbH & Co. KG Eberhard-Gerstel-Platz 1 45473 Mülheim an der Ruhr Germany +49 (0) 208-7 65 03-0 +49 (0) 208-7 65 03 33 gerstel@gerstel.com www.gerstel.com GERSTEL Worldwide GERSTEL, Inc. 701 Digital Drive, Suite J Linthicum, MD 21090 USA +1 (410) 247 5885 +1 (410) 247 5887 sales@gerstelus.com www.gerstelus.com GERSTEL AG Wassergrabe 27 CH-6210 Sursee Switzerland +41 (41) 9 21 97 23 gerstelag@ch.gerstel.com www.gerstel.ch GERSTEL K.K. 1-3-1 Nakane, Meguro-ku Tokyo 152-0031 SMBC Toritsudai Ekimae Bldg 4F Japan +81 3 5731 5321 +81 3 5731 5322 info@gerstel.co.jp www.gerstel.co.jp GERSTEL LLP Level 25, North Tower One Raffles Quay Singapore 048583 +65 6622 5486 +65 6622 5999 SEA@gerstel.com www.gerstel.com GERSTEL Brasil Av. Pascoal da Rocha Falcão, 367 04785-000 São Paulo - SP Brasil +55 (11)5665-8931 +55 (11)5666-9084 gerstel-brasil@gerstel.com www.gerstel.com.br Information, descriptions and specifications in this Publication are subject to change without notice. GERSTEL, GRAPHPACK and TWISTER are registered trademarks of GERSTEL GmbH & Co. KG. Copyright by GERSTEL GmbH & Co. KG Awarded for the active pursuit of environmental sustainability