Impurity Profiling of Carbamazepine by HPLC/UV

Similar documents
MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

The Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS

LC-MS/MS Method for the Determination of Tenofovir from Plasma

Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine

Quantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation

LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine

Dayana Argoti, Kerry Hassell, Sarah J. Fair, and Joseph Herman Thermo Fisher Scientific, Franklin, MA, USA

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Rapid and sensitive UHPLC screening of additives in carbonated beverages with a robust organic acid column

F. Al-Rimawi* Faculty of Science and Technology, Al-Quds University, P.O. Box 20002, East Jerusalem. Abstract

USP purity analysis of pravastatin sodium using the Agilent 1120 Compact LC

GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants

Analytical Method Development for USP Related Compounds in Paclitaxel Using an Agilent Poroshell 120 PFP

Enhanced LC-MS Sensitivity of Vitamin D Assay by Selection of Appropriate Mobile Phase

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract.

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

Rapid and sensitive UHPLC screening for water soluble vitamins in sports beverages

Determination of Clinically Relevant Compounds using HPLC and Electrochemical Detection with a Boron-Doped Diamond Electrode

Effective use of Pharmacopeia guidelines to reduce cost of chromatographic analysis for Fluticasone propionate

Evaluation of an LC-MS/MS Research Method for the Analysis of 33 Benzodiazepines and their Metabolites

ACQUITY UPLC WITH PDA DETECTION: DETERMINING THE SENSITIVITY LIMITS OF OXYBUTYNIN AND RELATED COMPOUNDS

Converting a CHP Method for Insulin to Agilent Poroshell 120 Columns

Journal of Chemical and Pharmaceutical Research

HPLC to UHPLC Transfer of USP Method for Amlodipine Besylate Using the Agilent 1290 Infinity II LC

A New HILIC/RP Mixed-Mode Column and Its Applications in Surfactant Analysis

Nitrogen/Protein Determination in Starch by Flash Combustion using Large Sample Weight as an Alternative to the Kjeldahl Method

Measuring Phytosterols in Health Supplements by LC/MS. Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA

Identification and Quantitation of Microcystins by Targeted Full-Scan LC-MS/MS

Selectivity Comparison of Agilent Poroshell 120 Phases in the Separation of Butter Antioxidants

Integrated Targeted Quantitation Method for Insulin and its Therapeutic Analogs

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

Transferring a Method for Analysis of DNPH-Derivatized Aldehydes and Ketones from HPLC to UHPLC

Pelagia Research Library

Rapid and Direct Analysis of Free Phytosterols by Reversed Phase HPLC with Electrochemical Detection

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR ESTIMATION OF LACOSAMIDE IN BULK AND ITS PHARMACEUTICAL FORMULATION

Sanjog Ramdharane 1, Dr. Vinay Gaitonde 2

SUMMARY, CONCLUSION & RECOMMENDATIONS

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR QUANTITATIVE ANALYSIS TOLBUTAMIDE IN PURE AND PHARMACEUTICAL FORMULATIONS

Analysis of Isoflavones with the PerkinElmer Flexar FX-15 UHPLC System Equipped with a PDA Detector

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

Development and validation of stability indicating RP-LC method for estimation of calcium dobesilate in pharmaceutical formulations

A New Era in SFC Column Technology

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD ESTIMATION OF TOLVAPTAN IN BULK PHARMACEUTICAL FORMULATION

Increased Identification Coverage and Throughput for Complex Lipidomes

Analysis of Cannabinoids in Cannabis by UHPLC Using PDA Detection

J Pharm Sci Bioscientific Res (4): ISSN NO

METHOD DEVELOPMENT AND VALIDATION BY RP-HPLC FOR ESTIMATION OF ZOLPIDEM TARTARATE

CHAPTER 2 SIMULTANEOUS DETRMINATION OF ANASTROZOLE AND TEMOZOLOMIDE TEMOZOLOMIDE CAPSULES 20 MG AND ANASTROZOLE TABLETS 1 MG

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

Extended Application Note

RITONAVIRI COMPRESSI RITONAVIR TABLETS. Final text for addition to The International Pharmacopoeia (July 2012)

Doubling the throughput of long chromatographic methods by using a novel Dual LC workflow

APPLICATIONS TN µm Core-Shell Particle. (continued on page 2) Page 1 of 8. For additional technical notes, visit

Application Note. Small Molecule Pharmaceuticals. 2-Ethyl-2-phenylmalonamide Ethosuximide Primidone Carbamazepine. Carbamazepine-

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

Maximizing Efficiency Using Agilent InfinityLab Poroshell 120 Columns

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

Mercury Speciation Determinations in Asian Dietary Supplements

A Rapid UHPLC Method for the Analysis of Biogenic Amines and Metabolites in Microdialysis Samples

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System

Journal of Chemical and Pharmaceutical Research

APPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology

ASSAY AND IMPURITY METHOD FOR DURACOR TABLETS BY HPLC

ISSN (Print)

Detection and Quantification of Inorganic Arsenic in Fruit Juices by Capillary Ion Chromatography with Suppressed Conductivity Detection

The Raptor HILIC-Si Column

Application Note. Agilent Application Solution Analysis of fat-soluble vitamins from food matrix for nutrition labeling. Abstract.

Development and validation of related substances method for Varenicline and its impurities

Pelagia Research Library

Maximizing chromatographic peak capacity with the Agilent 1290 Infinity LC system

Carl Fisher, Terri Christison, Hua Yang, Monika Verma, and Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA

Development and Validation of a Simultaneous HPLC Method for Quantification of Amlodipine Besylate and Metoprolol Tartrate in Tablets

Rapid Gradient and Elevated Temperature UHPLC of Flavonoids in Citrus Fruit

Scholars Research Library

Probing for Packaging Migrants in a Pharmaceutical Impurities Assay Using UHPLC with UV and Mass Detection INTRODUCTION

Pankti M. Shah et al, Asian Journal of Pharmaceutical Technology & Innovation, 04 (17); 2016; 07-16

Analysis of Common Sweeteners and Additives in Beverages with the PerkinElmer Flexar FX-15 System Equipped with a PDA Detector

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR QUANTITATIVE ANALYSIS OF TRAMADOL IN PURE AND PHARMACEUTICAL FORMULATIONS

Ankit et al Journal of Drug Delivery & Therapeutics; 2013, 3(2), Available online at RESEARCH ARTICLE

International Journal of Research in Pharmaceutical and Nano Sciences Journal homepage:

Determination of Water- and Fat-Soluble Vitamins in Nutritional Supplements by HPLC with UV Detection

Analysis of Counterfeit Antidiabetic Drugs by UHPLC with the Agilent 1220 Infinity Mobile LC

International Journal of Pharma and Bio Sciences DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR THE ESTIMATION OF STRONTIUM RANELATE IN SACHET

DRAFT PROPOSAL FOR THE INTERNATIONAL PHARMACOPOEIA: CARBAMAZEPINI COMPRESSI - CARBAMAZEPINE TABLETS

Stability indicating RP-HPLC method development and validation of Etizolam and Propranolol hydrochloride in pharmaceutical dosage form

TENOFOVIR TABLETS: Final text for addition to The International Pharmacopoeia (June 2010)

PIF: Precursor Ion Fingerprinting Searching for a Structurally Diagnostic Fragment Using Combined Targeted and Data Dependent MS n

Analysis of bisphenol A leaching from baby feeding bottles

CYCLOSERINI CAPSULAE - CYCLOSERINE CAPSULES (AUGUST 2015)

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

Analysis of Emulsifiers in Foods by High Pressure Liquid Chromatography and Corona Charged Aerosol Detection

Chromatogram Search Tool Search by compound name, synonym, CAS # or keyword Pinnacle II Columns

Forensic and clinical products and services

Scholars Research Library. Der Pharmacia Lettre, 2016, 8 (6): (

Asian Journal of Chemistry Vol. 21, No. 1 (2009),

Transcription:

Application Note: 52049 Impurity Profiling of Carbamazepine by HPLC/UV Terry Zhang, Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Key Words Accela Hypersil GOLD Carbamazepine Drug Analysis Goal Develop a simple, accurate and robust HPLC/UV method for quantitative impurity analysis of carbamazepine. Introduction Carbamazepine is a medication indicated for use in epilepsy, trigeminal neuralgia and bipolar disorder. 1 This widely prescribed tricyclic anticonvulsant is administered orally in usual doses of 400 1200 mg/day and is available in several commercial tablet, capsule and suspension forms. 1 Due to its narrow therapeutic range, tight control and routine monitoring of plasma carbamazepine concentrations are required to optimize treatment while preventing adverse drug effects. 1 Safe and effective therapy necessitates reliable and accurate methods for assessing all aspects of drug quality, including purity, strength and stability. Synthesis-related organic impurities are usually present in bulk forms as well as in pharmaceutical formulations of carbamazepine. Pharmacopoeias set strict standards for the purity of carbamazepine to ensure drug efficacy and safety. The United States Pharmacopeia (USP) establishes 0.2% as the maximum limit for any individual impurity and 0.5% as the total of all impurities relative to the active pharmaceutical ingredient. 2 The International Conference on Harmonization (ICH) guidelines recommend the reporting of any impurity above a level of 0.05% and the identification of all impurities above a level of 0.10% in new drug substances that are administered at <2 g/day doses. 3 The USP describes a quantitative HPLC/UV procedure for impurity profiling of carbamazepine that utilizes a 4.6 mm 250 mm column with L10 packing (5 10 µm silica particles with cyano bonded phases) and a mobile phase consisting of water, methanol, tetrahydrofuran, formic acid and triethylamine. However, the utility of this method for routine quality control analysis is limited by the complexity of the mobile phase, which could lead to poor reproducibility, as well as by the difficulty in achieving the required resolution (Rs 1.7) between carbamazepine and the impurity 10,11-dihydroxycarbamazepine (carbamazepine related compound A), which have comparable polarities and differ widely in their concentrations. Routine drug purity analysis requires simple and robust analytical techniques that deliver exceptional resolution, sensitivity, and accuracy. The Thermo Scientific Accela UHPLC system and Thermo Scientific Hypersil GOLD columns together enable sensitive, accurate and robust LC analysis for pharmacopeia methods and maximize flexibility for method development. The Accela 1250 Pump delivers precise flows and accurate gradients at an expansive range of flow rates (up to 2 ml/min) and pressures (up to 1250 bar), and the Accela PDA detector offers up to five times the sensitivity of conventional photodiode array detectors. Hypersil GOLD columns enhance peak shapes, resolution and reproducibility through high purity silica technology and robust bonding and endcapping procedures. In this note, we describe a simple HPLC/UV assay for accurate and robust impurity analysis of carbamazepine using the Accela UHPLC system and a high performance Hypersil GOLD CN (cyano) column, and demonstrate that the method meets USP requirements and ICH guidelines.

Materials and Methods Sample Preparation Carbamazepine and related impurities A, B, D and F were purchased from Sigma Aldrich. Stock solutions of 2 mg/ml were prepared in methanol. Solutions of 100 µg/ml concentration were prepared by diluting in methanol and were used for method development. Calibration solutions were prepared by serial dilution of the stock solutions in methanol, at concentrations of 250 ng/ml 400 µg/ml for each impurity and 5 µg/ml 1.5 mg/ml for carbamazepine. A mixture containing 1 mg/ml of carbamazepine and 500 ng/ml of each impurity was prepared to examine method suitability at a 0.05% reporting threshold. For assessments of system suitability and quantification, a mixture of 1.5 mg/ml of carbamazepine with 3 µg/ml of each impurity was used. HPLC/UV Analysis Instrumentation HPLC separations were performed on an Accela 1250 UHPLC system and an Accela autosampler (Thermo Fisher Scientific, San Jose, CA, USA). UV absorbance was monitored at 211 nm at 20 Hz using an Accela PDA detector (Thermo Fisher Scientific, San Jose, CA, USA). LC Parameters: Columns: Mobile Phase: Thermo Fisher Scientific Hypersil GOLD aq column (2.1 100 mm, 1.9 µm particle size) and Hypersil GOLD CN column (2.1 150 mm, 3.0 µm particle size) A: Water B: Acetonitrile Column Temperature: 30 C Sample Injection Volume: 1 µl Needle Wash: 80:20 (v/v) acetonitrile:water Gradient: Time A % B % µl/min 0.00 88.0 12.0 400.0 10.00 88.0 12.0 400.0 14.00 75.0 25.0 400.0 28.00 70.0 30.0 400.0 19.00 15.0 85.0 400.0 30.00 15.0 85.0 400.0 31.00 88.0 12.0 400.0 Results and Discussion HPLC Separation of Carbamazepine and Related Impurities The major organic impurities that arise from the synthesis of carbamazepine are 10,11-dihydrocarbamazepine (impurity A, also referred to as USP carbamazepine related compound A), 9-methylacridine (impurity B), N-carbamoylcarbamazepine (impurity C), iminostilbene (impurity D), iminodibenzyl (impurity E), and 5-chlorocarbonyliminostilbene (impurity F). An HPLC/UV assay that can accurately and reproducibly detect, identify and quantitate these compounds at trace levels (as low as 0.05% 0.20% relative to the parent drug carbamazepine) is required to comply with USP requirements and ICH guidelines. The USP chromatographic method for determining carbamazepine purity specifies the use of a 4.6 mm 250 mm L10 column (cyano column with 5 10 µm silica particles), a mobile phase composed of 1000 ml water/methanol/tetrahydrofuran (85:12:3) with 0.22 ml formic acid and 0.5 ml triethylamine, a flow rate of approximately 1.5 ml/min, and detection at 230 nm. Drawbacks of this assay for QA/QC applications include the use of a complex mobile phase with an ion-pairing agent, which could lead to operational difficulties and poor reproducibility, and the difficulty in meeting the resolution requirement (Rs 1.7) for carbamazepine and impurity A. Resolution and method performance may be improved by using high efficiency HPLC columns with smaller ( 3 µm) particles and by modifying the mobile phase. Using the Accela 1250 UHPLC system, a Hypersil GOLD aq polar endcapped C18 column (1.9 µm, 2.1 100 mm) and a simple acetonitrile/water gradient, a standard mixture containing carbamazepine and impurities A, B, D and F at 100 µg/ml concentrations was separated and detected within 23 minutes (Figure 1a). Detection at 211 nm was determined to provide the optimum response for this analysis; acetonitrile exhibits little absorption at this wavelength. All analytes were baseline-resolved and eluted in the order of carbamazepine, impurity A, impurity D, impurity B and impurity F. Symmetric chromatographic peaks were obtained and a resolution of 2.02 between carbamazepine and impurity A was achieved. However, resolution between this critical pair was considerably reduced when the concentration of impurity A was 0.1% of the parent drug (Figure 1b). Enhanced resolution is required to ensure compliance with USP requirements.

a b Figure 1a: HPLC separation of a mixture of carbamazepine and its related impurities (100 µg/ml) using a 1.9 µm Hypersil GOLD aq column, 2.1 mm 100 mm. Mobile phase: A Water; B Acetonitrile. UV detection: 211 nm. Figure 1b: Impurity A at a level of 0.1% is partially masked by the parent drug. Higher resolution (>1.7) between this critical pair is required to meet USP requirements.

The Hypersil GOLD CN (cyano) column offers alternative selectivity in reversed phase chromatography with lower hydrophobicity compared to C18 alkyl chain phases. The optimal wavelength of 211 nm was determined based on the best signal-to-noise ratio (S/N) for all analytes (Figure 2). Baseline separation of carbamazepine and the impurities at 100 µg/ml concentrations was achieved in 22 minutes using a Hypersil CN column (3 µm, 2.1 150 mm) and a simple acetonitrile/water gradient (Figure 3). Carbamazepine, impurity A and impurity D were less retained on the cyano phase and the elution order of carbamazepine and impurity A was reversed with this column. Excellent UV responses of similar magnitude were obtained for all analytes (Figure 3). The impurity responses were calculated relative to carbamazepine and were applied later for impurity analysis and quantification of a standard carbamazepine solution. The inset table of Figure 3 shows the responses of the impurities relative to Carbamazepine. Maximizing peak heights leads to sensitivity gains and is especially beneficial in trace analysis. Compared to Hypersil aq column, the Hypersil CN column notably improved the peak shapes of carbamazepine and impurity A to enable enhanced resolution and sensitivity at ICH reporting and identification thresholds (0.05% and 0.10%, respectively). Figure 2: Selection of the wavelength for quantitation was based on the best S/N for all analytes Figure 3: HPLC separation of a mixture of carbamazepine and its related impurities (100 µg/ml) using a 3 µm Hypersil GOLD CN column, 2.1 mm 150 mm. Mobile phase: A Water; B Acetonitrile. UV detection: 211 nm.

A standard solution of 1 mg/ml carbamazepine was analyzed and 600 ng/ml of impurity A (0.06%) was detected (Figure 4). The resolution between the carbamazepine and impurity A peaks was 2.28, which exceeds the minimum resolution specified by USP (Figure 4). Figure 5 shows the chromatogram of the standard 1 mg/ml carbamazepine solution spiked with 0.05% of impurities A, B, D and F. All impurities were easily detected at these trace amounts, and the stronger UV response exhibited by impurity A was attributed to the 0.06% that was already present in the carbamazepine standard. As the majority of aqueous and matrix contaminants usually elute early at void volume, elution of the first analyte of interest after the sixth minute ensures a robust quantitation method. Method Validation System suitability was investigated by analyzing six replicate injections of a solution containing 1500 µg/ml of carbamazepine and 3 µg/ml of each impurity, corresponding to 0.2% of the parent drug. Retention time RSDs ranged from 0.01 0.02%, while peak area RSDs ranged from 1 2% (Table 1), indicating excellent method reproducibility, particularly of the LC pump. According to Equation 1, signal-to-noise ratios of 25 are required to comply with the 2% RSD criterion specified by the USP method: Figure 4: 1.0 mg/ml of carbamazepine standard was analyzed and 0.06% impurity A (600 ng/ml) was detected. Excellent resolution is achieved between carbamazepine and impurity A on the Hypersil GOLD CN column. The resolution between this critical pair is 2.28, which exceeds the USP criterion of >1.7. % RSD 50/(S/N) (Equation 1) Figure 5: Impurities are baseline-separated and easily detected at the ICH reporting threshold (0.05%). Inters is the zoomed chromatogram. Mixture contains 1.0 mg/ml of carbamazepine and 500 ng/ml each of the impurities A, B, D, and F. Column: 3 µm Hypersil GOLD CN, 2.1mm 150 mm. Mobile phase: A Water; B Acetonitrile. UV detection: 211 nm.

Compound Concentration µg/ml n RT RSD% Area RSD% Mean Rs Mean S/N CAA* Impurity_A 3 6 0.01 2 NA 73.4 111 Carbamazepine 1500 6 0.01 1 2.2 19720.8 NA Impurity_D 3 6 0.01 2 9.1 44.1 108 Impurity_B 3 6 0.02 2 11.0 28.5 113 Impurity_F 3 6 0.01 1 2.2 34.6 102 Table 1: Method validation parameters were assessed by analyzing six replicate injections of each compound. Column: 3 µm Hypersil GOLD CN, 2.1 mm 150 mm. Mobile phase: A Water; B Acetonitrile. UV detection: 211 nm. CAA*: corrected area % accuracy Mean signal-to-noise ratios of the impurities at 0.2% levels relative to the parent drug ranged from 28.5 to 73.4 (Table 1), ensuring method reliability for quantification and compliance with USP requirements. Mean resolutions ranged from 2.2 to 11.0, and the mean resolution of 2.2 between carbamazepine and impurity A exceeded the USP resolution requirement of 1.7 (Table 1). Quantitative accuracy for all impurities was excellent, ranging from 102 to 111% (Table 1). Table 2 shows that excellent linearity in detector response was observed over the range of 0.25 400 µg/ml (ppm) for all impurities and 5 1500 µg/ml (ppm) for carbamazepine, with correlation coefficients 0.997 for all analytes. Conclusion A simple, accurate and robust quantitative HPLC/UV assay for carbamazepine purity assessment was developed. Separation and detection of carbamazepine and four of its related synthetic impurities was achieved within 22 minutes using the Accela 1250 UHPLC system, a Hypersil GOLD CN column and a simple water/acetonitrile gradient. The highly efficient and selective Hypersil GOLD CN column enhanced resolution between carbamazepine and impurity A, provided sensitivity gains through improved peak shapes, and eliminated the need for ion-pairing agents. Trace levels of impurities (0.05% of parent drug) were easily detected. Impurities at levels of 0.2% of carbamazepine were quantifiable with 2% RSD. This simple HPLC/UV assay meets USP criteria and ICH guidelines and provides a powerful alternative to the HPLC-based procedures currently recommended by regulatory agencies for QA/QC of carbamazepine purity. Compound Concentration µg/ml Correlation Coefficients Impurity_A 0.25 400 1.0000 Carbamazepine 5 1500 0.9973 Impurity_D 0.25 400 0.9995 Impurity_B 0.25 400 0.9999 Impurity_F 0.25 400 0.9998 Table 2: Excellent linearity of detector response was achieved, with correlation coefficients 0.997 for all compounds References 1. http://www.merckmanuals.com/professional/lexicomp/carbamazepine.html 2. U.S. Pharmacopeia Monograph for Carbamazepine, USP32-NF27, page 1784. 3. International Conference on Harmonization of Technical Requirements for Registration of Pharmaceuticals for Human Use, Impurities in New Drug Substances Q3A(R2), 2006. In addition to these offices, Thermo Fisher Scientific maintains a network of represen - tative organizations throughout the world. Africa-Other +27 11 570 1840 Australia +61 3 9757 4300 Austria Belgium +32 53 73 42 41 Canada +1 800 530 8447 China +86 10 8419 3588 Denmark +45 70 23 62 60 Europe-Other Finland/Norway/ Sweden +46 8 556 468 00 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9434 Italy +39 02 950 591 Japan +81 45 453 9100 Latin America +1 561 688 8700 Middle East Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Russia/CIS South Africa +27 11 570 1840 Spain +34 914 845 965 Switzerland +41 61 716 77 00 UK +44 1442 233555 USA +1 800 532 4752 www.thermoscientific.com Legal Notices: 2011 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Fisher Scientific, San Jose, CA USA is ISO Certified. AN52049_E 03/11M Part of Thermo Fisher Scientific