i-stat CHEM8 + Cartridge Quick Reference Guide

Similar documents
Procedure Manual for the i-stat System

Procedure Manual for the i-stat System

Abbott PXP Glucose Meters Resource: Lin Gustafson, MT(ASCP) Point of Care Coordinator, Laboratory

CoaguChek S System. Quick Reference Guide. This is a CLIA-waived System. Coagulation Testing with Fresh Whole Blood

Point of Care Testing. Bedside Glucose Meter

PHSA Laboratories CW Site - Point of Care Title: CWPC_BGE_0105 istat1 Quick Reference Guide

Step-by-Step Instructions For OraQuick HCV Rapid Antibody Test

LeadCare BLOOD LEAD ANALYZER. Quick Reference Guide

PROCEDURE. TITLE: Bedside Glucose Monitoring PC Laboratory. Issuing Department: Clinical Director Signature: Departments Involved:

QUICK REFERENCE INSTRUCTIONS. THYROCHEK TSH Cassette

User Manual Important: First read the Medication Guide that comes inside your FORTEO carton.

Items in the package:

To provide you with necessary knowledge and skills to accurately perform 3 HIV rapid tests and to determine HIV status.

Tips for Training New Customers

Fraser Health Authority Accu-chek Inform II: Training Script

BLOOD COLLECTION GUIDELINES

Instructions For Use PRALUENT (PRAHL-u-ent) (alirocumab) Injection, for Subcutaneous Injection Single-Dose Pre-Filled Syringe (75 mg/ml)

CentriVet GK Blood Glucose & Ketone Monitoring System

Instructions for Use HEMLIBRA (hem-lee-bruh) (emicizumab-kxwh) injection, for subcutaneous use

PLEASE READ THIS USER MANUAL BEFORE USE

Written By: Joann O Connor Date: 03/05/2017 Effective Date: 03/28/2017 1

Blood Glucose & Ketone Monitoring System

DETERMINE SPECIMEN REQUIREMENTS HIV-1/2 Ag/Ab Combo TESTING PROCEDURE

Accu-Chek Inform II: Point of Care Glucose Testing. Sharp Healthcare 2014

Instructions for Use. HUMALOG KwikPen. insulin lispro injection (rdna origin) 100 units/ml, 3 ml pen

Management of Central Venous Access Devices. Blood Glucose Monitoring

Precision ceedpro. Online Training for. Glucometer & Capillary Collection

Instructions for Use. BASAGLAR KwikPen. insulin glargine injection (100 units/ml, 3 ml pen)

Precision Xtra Training

Patient Information Publications Warren Grant Magnuson Clinical Center National Institutes of Health

HumaPen LUXURA HD INSULIN DELIVERY DEVICE INSTRUCTIONS FOR USE

Cap Clip Rubber Seal Plunger Pen Body Dose Window

NovoPen Echo User guide

HUMULIN 70/30 KwikPen

PLEASE READ THESE INSTRUCTIONS BEFORE USE

Point of Care testing refers to all laboratory testing that is done outside of the walls of the clinical laboratory in the proximity of the patient.

HumaPen SAVVIO INSULIN DELIVERY DEVICE INSTRUCTIONS FOR USE

Worksheet. Worksheet. Worksheet. Worksheet. Student Performance Guide. Student Performance Guide

Title/Description: Point of Care Glucose Testing-Waived Department: Organization Wide Personnel: All Effective Date: November 29, 2016 Revised:

MICAFUNGIN MIXING

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

HemoCue Hb METHOD AND SAMPLE COLLECTION

ANTIBODY SCREENING by Uni-Gold Recombigen HIV

Blood Glucose Monitoring System. User Guide

DONGBAO PEN (ELECTRONIC) INSTRUCTION MANUAL

ALPROLIX Coagulation Factor IX (Recombinant), Fc Fusion Protein INSTRUCTIONS FOR USE Do not Do not YOUR KIT CONTAINS:

Standard Operating Procedure for cannulation

Instructions for Use Enbrel (en-brel) (etanercept) for injection, for subcutaneous use Multiple-dose Vial

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6

INSTRUCTIONS FOR USE TYMLOS (tim lows ) (abaloparatide) injection, for subcutaneous use

Seven Simple Steps to Start. Your Dexcom G4 PLATINUM System

For use only with Lilly 3 ml insulin cartridges (100 units/ml). Do not use other brands of insulin cartridges.

For use only with Lilly 3 ml insulin cartridges (100 units/ml). Do not use other brands of insulin cartridges.

INSTRUCTIONS FOR USE. Norditropin NordiFlex Somatropin (rdna origin) injection 30 mg/3 ml Prefilled Pen

Current Status: Active PolicyStat ID: Accu-Chek Whole Blood Glucose Testing Policy PURPOSE

How to Set Up and Infuse Your TPN

Glucose Meter. User Guide. Veterinary Monitoring System. For dog and cat use only

Lilly. Disposable Insulin Delivery Device User Manual

How to Use ENBREL : Vial Adapter Method

TITLE: Whole Blood Glucose Monitoring Using the Nova StatStrip Meter

DO NOT TRANSFER TO A SYRINGE SEVERE OVERDOSE CAN RESULT

PROCEDURE NO. POC LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version.

SALEM HOSPITAL SALEM, OREGON 97309

Regulatory Operations. Insert size: ,5x dev Current 2.0. Colour: PMS 280C + PMS 648C + PMS 151C + PMS Green C + CMYK

BLINDED MODE USERS GUIDE SUPPLEMENT

Dear HighQ Check System Owner :

Align Gluco-Monitoring System (BG1) Quick Start Guide

Procine sphingomyelin ELISA Kit

GUIDELINES FOR THE USE OF INVICORP INJECTION FOR MEN WITH ERECTILE DYSFUNCTION.

HOW TO USE... 5mg. Pocket Guide

Point of Care (POC) Glucose Testing Revised on: December, 2018 Reference: Policy LAB 23-A; Point of Care Nova Stat Strip Glucose Meter

INSTRUCTIONS FOR USE. EMGALITY TM (em-gal-it-ē) (galcanezumab-gnlm)

Operator s Manual. Blood Glucose and β-ketone Monitoring System. Abbott Diabetes Care Inc South Loop Road Alameda, CA USA

Inspection Window. Syringe Body

Acknowledgement goes to WHO for use of their hand washing technique

Before you use the TALTZ autoinjector, read and carefully follow all the step-by-step instructions.

Instructions for Use Enbrel (en-brel) (etanercept) injection, for subcutaneous use Single-dose Prefilled Syringe

For use only with INSTRUCTIONS FOR USE

The Enbrel SureClick autoinjector is a single-dose prefilled autoinjector. It contains one 50 mg dose of Enbrel.

SymlinPen (SĬM-lĭnPen) 120 (pramlintide acetate) Pen-Injector

QuickVet Diagnostic System

Align Gluco-Monitoring System (BG1) Quick Start Guide

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

MYALEPT (MAI-uh-lept) (metreleptin) for injection for subcutaneous use

Instructions for Use Neulasta (nu-las-tah) (pegfilgrastim) Injection, for subcutaneous use Single-Dose Prefilled Syringe. Plunger rod Used plunger rod

ATI Skills Modules Checklist for Central Venous Access Devices

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

Instructions for Use. For use with. 10 mg vial

W IMPORTANT INFORMATION FOR PERFORMING A BLOOD GLUCOSE TEST

Instructions for Use. Welcome!

CIMZIA (certolizumab pegol)

QUICK REFERENCE INSTRUCTIONS

CHEMSTRIP 10 Urine Test Strips * READ POLICY PRIOR TO STARTING TUTORIAL

This is a CLIA-waived system. Policies and Procedures Manual

Charging the AccuChek Inform II on the Base Unit. Highlights of New AccuChek Inform II Whole Blood Glucose System

User Manual second language

A step-by-step preparation guide

IMPORTANT: PLEASE READ

A step-by-step preparation guide

I. Subject: Ionized Calcium (Ca++) Analysis Whole Blood

Transcription:

i-stat CHEM8 + Cartridge Quick Reference Guide For use with a Certificate of Waiver The CHEM8+ cartridge has tests for sodium, potassium, chloride, total carbon dioxide, ionized calcium, glucose, urea nitrogen, creatinine and hematocrit. Regulatory Requirements To use the CHEM8+ cartridge and i-stat 1 Handheld Analyzer (Handheld) for patient testing, your facility must have a CLIA Certificate of Waiver, and meet all applicable state and local laboratory testing laws. Further, the instructions provided in this guide must be followed, or your facility may be in non-compliance with the CLIA Certificate of Waiver program. Training Requirements Read this whole guide before testing patient samples. Follow the instructions exactly. Carefully read the section on Cartridge Storage. This guide contains information needed to get accurate results with the CHEM8+ cartridge and Handheld. View the video companion CD to the Quick Reference Guide that comes with this guide. It shows how to fill a cartridge and insert it into the Handheld. Additionally, training can be done over the phone. Contact your local sales person to schedule training as needed. Critical Values Ask your clinician to record critical values on the Test Range table in this guide. Before you begin... See the Start-up Section of the i-stat System Manual for Waived Tests for how to prepare a new or replacement Handheld for use. For assistance, contact Technical Support: 800.284.0702 or email:techsvc @i-stat.com 06F20-06

Precautions Potential Sources of Error in Patient Results Potential Sources of Harm to the Operator Potential Sources of Damage to the Handheld Incorrect cartridge storage Incorrect sample collection and/or sample handling: Test only fresh whole blood samples collected in tubes with lithium heparin anticoagulant. Fill tubes to capacity. Do not use short fill tubes. Test samples within 10 minutes of collection. Any deviations will cause inaccurate results. Use of expired cartridges To protect yourself and others from infection: Do not perform blood or control fluid testing in areas where food and drink are stored or consumed. Use gloves and wash hands after handling blood or blood soiled items. Do not use a cartridge if blood is spilled on it. Discard contaminated (blood soiled) items in a biohazard waste container. Decontaminate Handheld if blood is spilled on it. See Start-up section of Manual for instructions. Since blood spots may not be noticeable on the Handheld and since a cartridge could contaminate the inside of the Handheld, treat the Handheld as capable of transmitting infection. A falling or dropped Handheld. Place Handheld and peripherals on a stable surface. Barcode scanner: Do not look into laser beam coming from scanner, or point into eyes of someone else. Needles: take care to prevent needle sticks. Use safety tips when transferring sample from a blood collection tube to a cartridge. Handheld and peripherals not suitable for use in oxygen enriched atmosphere. Trying to pull a cartridge out of the Handheld while Cartridge Locked message is displayed. Dropping the Handheld. Quality Control Requirements Quality checks are automatically performed during each test. However, a CHEM8+ Level 1 control sample must be tested with each new shipment of cartridges, each new lot of cartridges, and monthly to check cartridges in storage. Follow the procedures in this guide including recommended recordkeeping to ensure your facility is in compliance with the CLIA Certificate of Waiver program.

Cartridge Shipments and Storage Required procedure for handling new cartridge shipments: 1. Open box marked Refrigerate Upon Arrival. Find card with temperature strip attached. Read strip immediately as it will change once it is exposed to room temperature. Follow instructions on card. 2. Fill out information in Received section at bottom of card. Mark an x in each of the spaces corresponding to a red window. If the 4 window is red, do not use cartridges. Contact Technical Support. 3. Record temperature reading on Receipt of New Cartridges log found in the System Resources section of this Manual. 4. If window 4 is not red, take one cartridge from each CHEM8+ lot in the shipment and test with i-stat CHEM8+ Level 1 control (See Test a Control Sample for instructions). If any result is outside its range in the Value Assignment sheet, repeat the test. If any result is still outside its range, do not use the cartridges. Contact Technical Support. Required procedures for cartridge storage: Refrigerated Storage Store cartridges at 2 to 8 C (35 to 46 F). Refrigerated cartridges may be used until date shown on cartridge box and pouch. It is recommended (but not required) that refrigerated storage be equipped with a 24-hour temperature monitor, and that the temperature record be reviewed each day. Room Temperature Storage Store cartridges at room temperature up to 30 C (86 F) for no more than 14 days. To avoid waste, take only a few days worth of cartridges out of refrigerator. Read date on pouch label. Do not use cartridge if this date has passed. If cartridge is not used on day it is removed from refrigerator, use a soft felt pen to mark 14-day expiration date on pouch. Check Storage Conditions Monthly 1. Remove one CHEM8+ cartridge from refrigerated storage. 2. Test cartridge with i-stat CHEM8+ Level 1 control as described in Test a Control Sample. If any result is outside its range in the Value Assignment sheet, repeat the test. If any result is still outside its range, do not use the cartridges. Contact Technical Support. 3. Record results on the Monthly Cartridge Check log found in the System Resources section of this Manual.

Perform a Patient Test To perform a patient test you will need: i-stat 1 Handheld CHEM8+ cartridge Blood sample in green top tube 1 cc syringe with safety tip Disposable gloves Gauze Biohazard container Collect the Sample Correct sample collection & handling are important for accurate results! Use a blood collection tube containing lithium heparin (green top tube). Use a 20 to 23 gauge needle. Do not apply a tourniquet for more than a minute while looking for a vein. Release it and reapply after two to three minutes. Do not allow the patient to clench and unclench fist repeatedly. If alcohol is used to clean the site, allow it to air dry before performing venipuncture. Fill blood collection tube to capacity and gently invert tube about 10 times. Perform a new venipuncture if blood does not flow freely into collection tube. Perform CHEM8+ test within 10 minutes of sample collection. If the sample is not to be tested immediately, label tube with the patient s name and another identifier, such as date of birth. Prepare the Cartridge After removal from refrigerator, let single cartridge stand in its pouch for 5 minutes at room temperature. An entire box of cartridges should stand for 1 hour before use. Tear open cartridge pouch at notch. Remove cartridge from pouch. Always hold by sides. Place on flat surface.

Fill the Cartridge Cartridge Features Tab for Snap Closure Fill Mark Sample Well (Blood goes here) 1. Gently invert tube of blood 2 3 times. Proceed to next step without delay. 2. Invert the tube and push safety tip on syringe through stopper into blood sample. Slowly pull back on plunger to draw blood into syringe until it is about half full. Note: An air bubble near plunger of syringe will not affect results. An air bubble near tip of syringe must be removed - see step 3. If air bubbles are mixed with the blood sample, use another syringe. DO NOT put blood back into tube. 3. Hold syringe so tip is pointing upward. Hold a gauze pad at tip to absorb blood. Press syringe plunger slowly until air is removed. If air bubble is stuck, tap syringe gently. If an air bubble was trapped at top of syringe it may start to move toward tip. In this case, wait until bubble reaches tip and remove it. 4. Direct tip of syringe into sample well of cartridge. Gently press syringe plunger to deliver blood into cartridge. When blood reaches fill mark on cartridge, stop adding blood. Some blood must be in sample well add blood to well if necessary. 5. Pressing on tab, fold snap closure over sample well until it clicks into place. DO NOT press over sample well.

Insert the Cartridge 1. Push cartridge into Handheld port as shown until cartridge clicks into place. The Handheld will first display Identifying Cartridge, then a time-to-result bar. Results will be shown in 2 to 3 minutes. DO NOT remove cartridge until the Cartridge Locked message is removed and results are displayed on screen. 2. Following prompts on screen, use number keys to enter an Operator and Patient ID. Press ENT when each ID number is complete. If ID numbers are not required, just press ENT to continue. Note: Press to clear a wrong number. Note: ID numbers can also be scanned into the Handheld. (See Start-up section in Manual) 3. When results are shown, the cartridge may be removed and discarded with other test materials: Discard cartridge, syringe, tube of blood and gloves in biohazard container. Wash hands. Scan or Enter Operator ID Scan or Enter Patient ID Cartridge Locked Identifying Cartridge Please Wait Pt: Cartridge 12345 Locked i-stat CHEM8+ Calibrating Cartridge Locked 0 120 Time To Results Page Cartridge Locked Review Results The Handheld shows the numerical values and units with the results. It also shows bar graphs with tic marks for reference ranges. See Test Ranges on next page for a list of reportable (measurement) ranges and reference (normal) ranges. If any result is shown as ***, <, >, or < > see Flagged Results. If a message and code are displayed, see Troubleshooting. Pt: 12345 16:10 25Jan07 i-stat CHEM8+ Na mmol/l 141 K mmol/l 4.2 Cl mmol/l 104 ica mmol/l 1.22 TCO2 mmol/l 26 Page Pt: 12345 16:10 25Jan07 i-stat CHEM8+ Glu mg/dl 90 BUN mg/dl 17 Crea mg/dl 1.0 Hct %PCV 42 Hb g/dl 14.3 AnGap mmol/l 16 Page Not all results are displayed on the first screen. Press to page through all screens. Note: If Handheld turns off before review of results is complete, press turn it on, then press 1 for Last Result. to

Test Ranges: Reportable Range is the lowest to highest values the test system will report. Reference Range is the normal values for an adult population. Reference ranges may vary according to age, gender and heritage. Critical Values indicate that a patient may need treatment right away if results are at or below the low value, or at or above the high value. Test Reportable Reference Critical Values Test Symbol Units Range Range Low High Sodium Na mmol/l 100 180 138 146 120 160 Potassium K mmol/l 2.0 9.0 3.5 4.9 2.5 6.5 Chloride Cl mmol/l 65 140 98 109 70 120 Total Carbon Dioxide TCO 2 mmol/l 5 50 24 29 15 50 Ionized Calcium ica mmol/l 0.25 2.50 1.2 1.32 0.70 1.60 Glucose Glu mg/dl 20 700 70 105 45 400 Urea Nitrogen BUN mg/dl 3 140 8 26 50 Creatinine Crea mg/dl 0.2 20.0 0.6 1.3 2.0 Hematocrit Hct % PCV 10 75 38 51 27 70 Hemoglobin* Hb g/dl 3.4 25.5 12 17 Anion Gap* AnGap mmol/l -10 99 10 20 * Calculated Values Your clinician may wish to use different Reference Ranges and Critical Values for tests performed with the CHEM8+ cartridge. Use the table below to replace the values shown above (if applicable). Test Reportable Reference Critical Values Test Symbol Units Range Range Low High Sodium Na mmol/l 100 180 Potassium K mmol/l 2.0 9.0 Chloride Cl mmol/l 65 140 Total Carbon Dioxide TCO 2 mmol/l 5 50 Ionized Calcium ica mmol/l 0.25 2.50 Glucose Glu mg/dl 20 700 Urea Nitrogen BUN mg/dl 3 140 Creatinine Crea mg/dl 0.2 20.0 Hematocrit Hct % PCV 10 75 Hemoglobin* Hb g/dl 3.4 25.5 Anion Gap* AnGap mmol/l -10 99 * Calculated Values

Record and Report Results 1. Record results on patient record. Note: If using the portable printer, be aware that the printout will fade after about 5 years. If a more permanent record is required, photocopy the printout. 2. Report results to physician. If physician questions a result, request a fresh sample and repeat test. If a result is still questioned, check Factors That Affect Results on the back page of this guide. If an interfering substance is suspected, a sample should be sent to a commercial or hospital laboratory. Flagged results: If stars (***) are displayed instead of a result, it means that a test failed internal quality checks. All reported results are accurate. Remix tube of blood and repeat test using a fresh cartridge. If result is not displayed again, draw a fresh blood sample and repeat test. If result is still not displayed, call Technical Support. < is shown in front of the lowest reportable value when the result is lower than this value. See Test Ranges for reportable ranges. > is shown in front of the highest reportable value when the result is higher than this value. < > is shown in place of a result if the result is dependent on another result that is flagged with either the < or > symbol. Quality Checks: Quality checks are automatically performed during each test. If a quality check fails, the Handheld stops the test and shows a cause and action to be followed. Some examples and an explanation include: SAMPLE POSITIONED BEYOND FILL MARK USE ANOTHER CARTRIDGE Too much fluid was put into cartridge. SAMPLE POSITIONED SHORT OF FILL MARK USE ANOTHER CARTRIDGE Not enough blood was put into cartridge. Make sure blood reaches the fill mark. INSUFFICIENT SAMPLE USE ANOTHER CARTRIDGE Not enough blood was put into cartridge. Make sure some blood is in the sample well and that there are no air bubbles in the cartridge. UNABLE TO POSITION SAMPLE USE ANOTHER CARTRIDGE Cartridge not closed properly, sample was outside sample well or sample clotted. CLEW EXPIRING, UPDATE REQUIRED Software will expire 15 days after message is first displayed. Update software. CARTRIDGE ERROR, USE ANOTHER CARTRIDGE (Code 79 or 80) If code occurs a second time, Handheld has been damaged and needs to be replaced. A complete list of Quality Checks is in the System Resources section of the i STAT System Manual for Waived Tests.

Test a Control Sample To perform a control test you will need: i-stat 1 Handheld CHEM8+ cartridge Ampule of CHEM8+ Level One Control An ampule breaker, gauze or paper towel A 1cc plain syringe Value Assignment sheet Prepare the Control 1. Control fluid is stored in the refrigerator. Allow an ampule of control fluid to stand at room temperature for 30 minutes before use. Do not use a control fluid past the expiration date on the box and ampule label. Prepare the Handheld 1. Press to turn on Handheld. 2. Press MENU to change screen to Administration Menu. 3. Press 3 for the Quality Tests menu. 4. Press 1 for Control. 15:26 30SEP05 Administration Menu 1 - Analyzer Status 2 - Data Review 3 - Quality Tests 4 - Customization Quality 5 - Set Clock Tests 1 - Control 6 - Transmit Data 2 - Proficiency 7 - Utility 3 - Cal Ver 4 - Simulator 5. Following prompt on screen, use number keys to enter an Operator ID, then press ENT. If ID number is not required, just press ENT to continue. Note: Press to clear a wrong number. 6. Enter control ampule lot number and press ENT. Note: Ignore any letters in the lot number 7. Enter cartridge lot number and press ENT. Note: Ignore any letters in lot number 8. Proceed to next step. The Handheld will wait 15 minutes for a cartridge to be inserted before it turns off.

Fill the Syringe 1. Hold the ampule between the index finger and thumb and shake hard for 10 seconds. 2. Tap or flick the top part of the ampule to send all the fluid to the bottom part of the ampule. 3. Place an ampule breaker over the top part of the ampule or protect your fingers with a gauze or paper towel, then break off top with a quick snap. 4. Tilt opened ampule so that fluid flows close to opening. 5. Position syringe tip into the fluid. 6. Slowly pull back on syringe plunger to draw fluid into syringe until it is about half full. 7. Hold syringe tip over gauze pad or waste container. Press syringe plunger just enough to discard three drops of fluid. Fill, Close and Insert the Cartridge 1. Place the syringe tip into the sample well of the cartridge. 2. Gently press syringe plunger until fluid reaches fill mark on cartridge. The fluid is colorless so watch fluid carefully as it moves toward the Fill mark. 3. Pressing on tab, fold snap closure over sample well until it clicks into place. DO NOT press over sample well. 4. Push cartridge into Handheld port until cartridge clicks into place. The Handheld will first display Identifying Cartridge, then a time-to-result bar. Results will be shown in 2 to 3 minutes. DO NOT remove cartridge until the Cartridge Locked message is removed and results are displayed on screen.

5. When results are shown, the cartridge may be removed and discarded with other test materials. Control fluid is water-based and does not contain human blood product, so it is not necessary to discard used materials in a biohazard container. However, it should be disposed of in a container for broken glass. Review and Log Results The Handheld shows the numerical values for each test result. Cartridges may be used if all results are inside the ranges on the Value Assignment sheet. Results should be recorded in a log such as the one in the System Resources section of the i-stat System Manual for Waived Tests. Not all results are displayed on the first screen. Press to page through all screens. Control 02467 15:26 30SEP05 i-stat CHEM8+ Na mmol/l 121 K mmol/l 2.9 Cl mmol/l 78 ica mmol/l 1.46 TCO2 mmol/l 25 Page If any result is outside the range on the Value Assignment sheet, repeat the test following instructions carefully. If one or more results are still outside ranges on the Value Assignment sheet, do not use cartridges. Contact Technical Support for further instructions. Note: If Handheld turns off before review of results is complete, press turn it on, then press 1 for Last Result. to Troubleshooting In addition to the quality checks that are automatically performed during each test, other situations may also occur which require operator action: Problem Action Battery Low message and battery Change batteries. icon flashing. Handheld will not turn on when a cartridge is inserted. FAIL is displayed when a cartridge is inserted. Change the batteries This is a failure of the internal Electronic Simulator. Use external Electronic Simulator to verify failure. See Start-up Section for procedure. If FAIL occurs again, contact Technical Support.

Factors that Affect Results Interfering substances or other events may be encountered which can affect results. Refer to summary below, or see information about Cartridge and Test Information sheets in the System Resources section of the i STAT System Manual for Waived Tests. Factors that may Factors that may Test Increase Results Decrease Results Sodium (Na) Bromide ß-hydroxybutyrate Hemodilution 1 Lactate Potassium (K) Hemolysis Delay in testing Chloride (Cl) ß-hydroxybutyrate Hemodilution 1 Bromide Lactate Salicylate Thiocyanate Total Carbon Dioxide Delay in testing Exposure to air (TCO 2 ) Incomplete fill of collection tube Ionized Calcium (ica) Magnesium ß-hydroxybutyrate Hemodilution 1 Lactate Salicylate Incomplete fill of collection tube Delay in testing Glucose (Glu) ph above 7.4 Bromide Hydroxyurea ph below 7.4 Oxygen less than 20mmHg Thiocyanate Delay in testing Urea Nitrogen (BUN) Thiocyanate Creatinine (Crea) Acetaminophen PCO 2 Ascorbate Bromide Hydroxyurea Creatine N-acetylcysteine PCO 2 Hematocrit (Hct) White Blood Count Total Protein <6g/dL Total Protein >8g/dL Lipids 1 Hemodilution greater than 20% with normal saline or Ringer s Lactate Abbott Point of Care Inc. 104 Windsor Center Drive East Windsor, NJ 08520 www.abbottpointofcare.com 2007 Abbott Art: 716942-00A, Rev. Date:09OCT2007 i-stat is a registered trademark of Abbott Laboratories.