Reprocessing Guide. Autoclavable Arthroscope and Hardware Sterilization Tray

Similar documents
M6-C Artificial Cervical Disc Surgical Instruments

SI-BONE ifuse Implant System Instruments Hospital Cleaning and Sterilization Instructions 0344

Velocity Orthopedics Instrument Cleaning and Sterilization

Endoscopic Technology. Operating Manual. Cannula system

Scope This policy applies to all personnel and departments that clean, prepare and/or sterilize items intended for patient care use.

M Manufacturer by: Guide for Cleaning, Sterilization and Storage of Instruments for Caldera Medical Desara Product Family of Implants

Manufacturer s information

Aesculap Implant Systems

Sterile U Network. Quality Control for Table-top Steam Sterilizers. Introduction: T U T O R I A L S

Preparation Instructions for the CAMLOG /CONELOG Implant System

Surgical Technique Guide

OBSOLETE. Visit for the latest version. Ankle Plating System 3 PKGI-79-B EFFECTIVE

Hygienic Reprocessing

Straumann ProClean Cassette. Basic Information on the

TJF-160F/VF Cleaning and Disinfection Checklist

LBL-014. Rev

MATERIAL SAFETY DATA SHEET (MSDS) Information about stainless steel for Surgical/Dental/Orthodontic Instruments.

INSTRUCTIONS FOR USE AND CARE

How clean are your instruments?

INSTRUCTIONS FOR CLEANING, DISINFECTION AND STERILIZATION OF DFS INSTRUMENTS

Integra Total Foot System 2

TABLE OF CONTENTS. PAL-650 Instrument System Instruction Manual. Applicable Part Numbers Introduction General Warnings...

For use by an Accredited Orthopaedic Surgeon only

EVOLVE TRIAD BONE SCREWS

DISCLAIMER OF WARRANTY AND LIMITATION OF REMEDY

RibLoc U Plus CHEST WALL PLATING SYSTEM FOR THE PERSONAL ATTENTION OF THE OPERATING SURGEON

SIDEKICK EZ FRAME EXTERNAL FIXATION SYSTEM The following languages are included in this packet:

Recommendations for cleaning, disinfection and sterilization of orthopaedic instruments and devices from Swemac.

REDUCT Headless Compression Screw System

Bryan-Dumon Series II Rigid Bronchoscope and Stent Placement Kit USER MANUAL

TMJ Fossa-Eminence Prosthesis System. Instructions for Use. For partial reconstruction (hemiarthroplasty) of the temporomandibular joint

DARCO MIS FOREFOOT SCREW

The following languages are included in this packet:

ProUltra ULTRASONIC SURGICAL TIPS

EVOLVE TRIAD SYSTEM

Larry! 28/06/2016. House Keeping. Welcome! From the GoToWebinar page:

Disinfection and Decontamination Core Subject. The Process of Instrument Decontamination

X-spine Systems, Inc. Silex Sacroiliac Joint Fusion System

EXTERNAL FIXATION SYSTEMS The following languages are included in this packet:

DISCLAIMER OF WARRANTY AND LIMITATION OF REMEDY

Instruction Manual Aerogen, Inc. Part No. AG-AL1010 Rev. C

BIOFOAM BONE WEDGE

MICA SCREWS For additional information and translations please contact the manufacturer or local distributor.

Aesculap Long and Short Ventriculoscope Systems Instructions for Use

B&H SURGICAL. Laparoscopic Reference Guide. P.O. Box Santa Clarita, CA Toll Free: Ph: Fax:

CANNULATED SCREW SYSTEM The following languages are included in this packet:

SURGICAL TECHNIQUE GUIDE BRECKENRIDGE. Intervertebral Body/VBR Fusion System. Oblique Lumbar Interbody Fusion

Reprocessing of Implants: What are the Issues?

TMJ Fossa-Eminence and Condylar Prosthesis System. Instructions for Use. For total reconstruction (arthroplasty) of the temporomandibular joint

Dentium Surgery Kit Dental Implant Surgical Instrument Set

Table of Contents. Introduction Indications For Use Contraindications Warnings Precautions...5

ORTHOLOC 3Di PLANTAR LAPIDUS PLATE

Cleaning and disinfection Tonometer measuring prisms, contact lenses and Desinset

SelectTech 4.1 Bench Assembly / Owner s Manual

ProUltra Ultrasonic Non Surgical Endo Tips

English G EMINU S Select Distal Radius S ystem INSTRUCTIONS FOR USE

Instructions for Use AccuFit Lateral Plate System Caution: Federal (USA) law restricts this device to sale by or on the order of a physician

Medescan Nebuliser Med - S600A Instructions Manual

X-spine Systems, Inc. Axle Interspinous Fusion System

CLAW II POLYAXIAL COMPRESSION PLATING SYSTEM

Rolimeter 50A. Operator s Manual for measuring anterior/posterior knee laxity

SelectTech 4.1 Bench. Assembly Manual

X-spine Systems, Inc. Calix PC Spinal Implant System

INFECTION PREVENTION AND CONTROL POLICY AND PROCEDURES Sussex Partnership NHS Foundation Trust (The Trust)

Chapter 5: Checking and Maintaining Ultrasound Equipment

SURGICAL TECHNIQUE GUIDE BRECKENRIDGE. Intervertebral Body/VBR Fusion System. ACDF Anterior Cervical Discectomy and Fusion

SelectTech 3.1 Bench Assembly / Owner s Manual

A82-LIFE-Dog Nebulizer NBB02

ATEC Porous Ti System

PORTABLE COMPRESSOR NEBULIZER Guidebook & Manual Reorder No. 5606

TONTARRA Medizintechnik GmbH. Surgical Instruments in Demand. Laminectomy Punches and Rongeurs

INSTALLATION MANUAL. VIDEO Camera, Probe and Lightsource OTOSCOPES.

Reusable Instrument Lifespan Manual

Zeus Intervertebral Body Fusion Devices PACKAGE INSERT

Compressor Nebulizer System

Zavation Z-Link Cervical

Cleaning and sterilization. Guidelines for Nobel Biocare products including MRI information

Compressor Nebulizer. Guidebook

UNICLIP DIRECTIONS FOR USE REF C226U - 0,8 MM / 1 MM FOR DENTAL USE ONLY. FISDR / F EN / 11 / updated 06/2016 1/7

ULTR A SERIES G7-S75

Yasargil Permanent Aneurysm Clips

Failure to follow instructions may lead to patient injury.

ASTM Workshop on Medical Device Cleanliness

Reusable Instrument Lifespan Manual

For more information visit or contact hearx:

COMPRESSOR NEBULIZER MODEL: JB USER S MANUAL

METATARSAL DECOMPRESSION IMPLANT (MDI) SYSTEM

Approval Signature: Original signed by B. Postl. June 2007

How Do You Know If Your Sterilized Instruments Remain Sterile?

It is intended that the implants be removed after successful fusion.

Page 1 of 7 Doc. No.: IU/15 Rev. No.: 05 Rev. Date:03MAY2018. Instructions for use for ATLAS Tibial Fracture(TF) Nail

1) INDICATIONS FOR USE These products have to be used only in hospital environments, clinics or dental offices by qualified dental personnel.

OWNER S MANUAL and INSTALLATION INSTRUCTIONS

X-spine Systems, Inc. Certex Spinal Fixation System

Spartan S 3 Facet Screw System PACKAGE INSERT

OWNER S MANUAL and INSTALLATION INSTRUCTIONS

Sample Enhanced Visual Inspection of medical devices. SUBJECT: Enhanced Visual Inspection of medical devices

Amendia Interbody Fusion Devices PACKAGE INSERT

X-spine Aranax Anterior Cervical Plating System Instructions for Use / Package Insert

CATHETER ACCESS KIT. For use with Prometra Programmable Infusion Systems

Transcription:

Reprocessing Guide Autoclavable Arthroscope and Hardware Sterilization Tray 0233032116

Contents Introduction...2 Intended Use of Sterilization Trays...4 Warnings...5 Cautions...5 Instructions...6 Sterilization Tray Setup...11 References...12 1

Introduction This reprocessing guide provides instructions for the proper cleaning and sterilization of the following sterilization tray: 233-032-116 Autoclavable Arthroscope and Hardware Sterilization Tray This sterilization tray is intended for use only with the devices listed below. For tray/ device configuration, see the Sterilization Tray Setup section of this guide. Combination of any two of the instruments below. Stryker autoclavable arthroscopes: HD or Standard, C-mount, or with eye-piece on proximal end. Any six of the instruments below Part Number Description 502-344-530 3.2mm cannula with 1 rotating stopcock 502-344-550 3.2mm cannula with 1 fixed stopcock 502-344-560 3.2mm cannula with 2 rotating stopcocks 502-344-540 3.2mm cannula with 2 fixed stopcocks 502-344-500 Pencil tip obturator for 3.2mm cannula 502-344-510 Blunt tip obturator for 3.2mm cannula 502-344-520 Trocar for 3.2mm cannula 502-244-530 4.0mm cannula with 1 rotating stopcock 502-244-550 4.0mm cannula with 1 fixed stopcock 502-244-560 4.0mm cannula with 2 rotating stopcocks 502-244-540 4.0mm cannula with 2 fixed stopcocks 502-244-500 Pencil tip obturator for 4.0mm cannula 502-244-510 Blunt tip obturator for 4.0mm cannula 502-244-520 Trocar for 4.0mm cannula 502-144-530 2.8mm cannula with 1 rotating stopcock 502-144-550 2.8mm cannula with 1 fixed stopcock 502-144-560 2.8mm cannula with 2 rotating stopcocks 502-144-540 2.8mm cannula with 2 fixed stopcocks 502-144-500 Pencil tip obturator for 2.8mm cannula 502-144-510 Blunt tip obturator for 2.8mm cannula 502-144-520 Trocar for 2.8mm cannula 502-623-530 2.9mm cannula with 1 rotating stopcock 502-623-500 Pencil tip obturator for 2.9mm cannula 502-623-510 Blunt tip obturator for 2.9mm cannula 502-623-520 Trocar for 2.9mm cannula 502-714-530 4.0mm cannula with 1 rotating stopcock 2

Any six of the instruments below Part Number Description 502-714-550 4.0mm cannula with 1 fixed stopcock 502-714-560 4.0mm cannula with 2 rotating stopcocks 502-714-540 4.0mm cannula with 2 fixed stopcocks 502-714-500 Pencil tip obturator for 4.0mm cannula 502-714-510 Blunt tip obturator for 4.0mm cannula 502-714-520 Trocar for 4.0mm cannula 502-713-530 3.2mm cannula with 1 rotating stopcock 502-713-550 3.2mm cannula with 1 fixed stopcock 502-713-560 3.2mm cannula with 2 rotating stopcocks 502-713-540 3.2mm cannula with 2 fixed stopcocks 502-713-500 Pencil tip obturator for 3.2mm cannula 502-713-510 Blunt tip obturator for 3.2mm cannula 502-713-520 Trocar for 3.2mm cannula 502-712-530 2.9mm cannula with 1 rotating stopcock 502-712-500 Pencil tip obturator for 2.9mm cannula 502-712-510 Blunt tip obturator for 2.9mm cannula 502-712-520 Trocar for 2.9mm cannula 502-711-530 2.8mm cannula with 1 rotating stopcock 502-711-550 2.8mm cannula with 1 fixed stopcock 502-711-560 2.8mm cannula with 2 rotating stop cocks 502-711-540 2.8mm cannula with 2 fixed stopcocks 502-711-500 Pencil tip obturator for 2.8mm cannula 502-711-510 Blunt tip obturator for 2.8mm cannula 502-711-520 Trocar for 2.8mm cannula 502-577-550 5.8mm cannula with 2 rotating stopcock 502-577-515 Pencil tip obturator for 5.8mm cannula 502-577-510 Blunt tip obturator for 5.8mm cannula 502-577-520 Trocar for 5.8mm cannula 377-031-144 5.0mm cannula with 2 rotating stopcocks 377-031-145 5.0mm cannula with 1 rotating stopcock 377-031-146 5.0mm cannula with 1 fixed stopcock 377-031-148 5.0mm cannula with 2 fixed stopcocks 377-031-129 Pencil tip obturator for 5.0mm cannula 377-031-229 Trocar for 5.0mm cannula 343-331-430 4.0mm cannula with 1 rotating stopcock 343-331-450 4.0mm cannula with 2 rotating stopcocks 343-331-410 Pencil tip obturator for 4.0mm cannula 3

Any six of the instruments below Part Number Description 343-331-411 Blunt tip obturator for 4.0mm cannula 343-331-420 Trocar for 4.0mm cannula 377-031-650 6.5mm cannula with 2 rotating stopcocks 377-031-630 6.5mm cannula with 1 rotating stopcock 377-031-660 6.5mm cannula with 2 fixed stopcocks 377-031-640 6.5mm cannula with 1 fixed stopcock 377-031-610 Pencil tip obturator for 6.5mm cannula 377-031-611 Blunt tip obturator for 6.5mm cannula 377-031-620 Trocar for 6.5mm cannula 747-031-560 5.8mm cannula with 2 fixed stopcocks 747-031-550 5.8mm cannula with 2 rotating stopcocks 747-031-540 5.8mm cannula with 1 fixed stopcock 747-031-530 5.8mm cannula with 1 rotating stopcock 747-031-510 Pencil tip obturator for 5.8mm cannula 747-031-511 Blunt tip obturator for 5.8mm cannula 747-031-520 Trocar for 5.8mm cannula 747-031-650 Inflow/Outflow Cannula 2 rotating stop cocks 377-727-540 Bridge 2 rotating stopcocks 377-727-530 Bridge 1 rotating stopcock 377-727-000 Bridge 2 fixed stopcocks 377-727-520 Bridge 1 fixed stopcock While these instructions have been validated by the manufacturer as being capable of preparing the tray for re-use, it remains the responsibility of the processor to ensure that the reprocessing (as actually performed, using equipment, materials, and personnel in the reprocessing facility) achieves the desired result. This normally requires validation and routine monitoring of the process. Intended Use of Sterilization Trays Sterilization trays are plastic and/or metal containers used to hold and protect surgical devices during the sterilization process. They consist of an interlocking tray and lid, which are both perforated to allow the passage of sterilizing agent from outside the tray to the devices placed inside. Sterilization trays typically feature a silicon finger mat or group of device holders that secure devices during the sterilization process. Some models feature stacking internal trays to allow the segregation of devices. 4

Warnings These instructions are validated only for sterilization of the tray and device(s) identified herein. Using combinations or parameters not described in this manual may result in incomplete sterilization. These instructions do not replace the cleaning instructions provided with individual devices. Prior to sterilization, clean all devices as specified in their respective user manuals. Wear appropriate protective equipment (gloves, eye protection, etc.) when reprocessing any medical device. Both the device and the tray must be cleaned prior to sterilization, or incomplete sterilization will result. The sterilization tray, its lid, and any internal components have been designed and validated for use as a single system. Do not separate components from the system, for use individually or in combination, or incomplete sterilization may result. Inspect the tray and its components for visible damage, such as cracking or chipping, prior to use. Do not use the tray if it is damaged. Cautions Before lifting the tray assembly, verify that the latches connecting the lid to the tray are secure. The tray is not designed for use as a shipping container. To avoid damage, remove all devices and pack them separately. Limitations on Reprocessing Proper processing has a minimal effect on the tray. End of life is normally determined by wear and damage due to use. Do not leave the tray in solutions longer than necessary. This may accelerate normal product aging. Damage incurred by improper processing will not be covered by the warranty. 5

Instructions Instructions apply to the sterilization tray only. For instructions on how to reprocess the devices, consult their respective instructions for use. Point of Use Containment and Transportation Preparation for Manual Cleaning 1 Cleaning was validated using Enzol at 1 oz/gal. at 35-40 C. Wipe excess soil from the tray using disposable paper towels. If an automated cleaning method will be used, rinse the tray in sterile distilled water immediately after use. Reprocess device(s) / tray as soon as reasonably practical following use. 1 Disassemble the tray into its individual components: base, lid, and (if applicable) silicon mat. 2 Prepare an enzymatic detergent1 according to the manufacturer s recommendations. 3 Wipe the entire tray with a clean cloth dipped in detergent. 4 Immerse the tray in the detergent. 5 Soak the tray in the detergent for 15 minutes. Manual Cleaning 1 Brush Thoroughly brush the exterior of the tray with a soft-bristled brush, focusing on any mated or rough surfaces. Using a pipe cleaner and brush, brush and flush the hard-to-reach areas of the tray. Actuate the tray, brushing any movable parts in their extreme open and closed positions. 6

2 Rinse Rinse the tray with reverse osmosis/deionized (RO/DI) water at ambient temperature to remove all detergent residues. After all detergent residues have been removed, continue to rinse for a minimum of 30 seconds. Drain excess water from the tray and dry it using a clean cloth and filtered pressurized air (40 psi). Visually inspect the tray for cleanliness, paying close attention to hard-to-reach areas. If visible soil remains, repeat steps 1 and 2. 3 Soak Prepare a non-enzymatic detergent2 according to the manufacturer s recommendations. Fully immerse the tray and inject any mated surfaces with at least 50mL of the detergent. Soak the tray for a minimum of 15 minutes. 4 Brush Thoroughly brush the exterior of the tray using a soft-bristled brush. Inject the prepared detergent into any mated surfaces a minimum of five times. Actuate the tray, brushing any movable parts in their extreme open and closed positions. Using a pipe cleaner and syringe, brush and flush the hard-to-reach areas of the tray. 2 Cleaning was validated using Renu-Klenz at ¼ oz/gal. at 35-40 C. 5 Rinse Thoroughly rinse the tray with ambient RO/DI water until all detergent residue is removed. Flush any crevices five times with a syringe. After the detergent residue is removed, continue rinsing for a minimum of 30 seconds. Drain the excess water from the tray and dry it with a clean cloth and filtered pressurized air (40 psi). Visually inspect the tray for cleanliness, paying close attention to hard-to-reach areas. 7

Automated Cleaning 1 Rinse Disassemble the tray into its individual components: the base, lid, and (if applicable) silicon mat. Rinse the tray with RO/DI water at ambient temperature until there is no visible detergent residue. Use a syringe to assist in rinsing. After all detergent residues have been removed, continue to rinse for a minimum of 30 seconds. Place the tray in the washer on an incline to facilitate drainage. 2 Automated wash Program the washer using the settings below. Recirculation Time Water Temperature Detergent (type and concentration) Pre Wash 2 min. Cold tap water Enzyme Wash 2 min. Hot tap water Enzymatic Detergent3 Wash 1 2 min. Set point 60 C (140 F) Non-enzymatic Detergent4 Rinse 1 2 min. Heated 60 C (140 F) Dry Phase 7 min. Drying Temperature 115 C (239 F) 3 Drying 3 Cleaning was validated using Enzol at 1 oz/gal. 4 Cleaning was validated using Renu-Klenz at ¼ oz/gal. Remove the tray from the washer after the completion of the dry phase. Dry the tray using a clean soft cloth. Use filtered pressurized air (40 psi) to assist in drying. Visually inspect the tray for cleanliness, paying close attention to hard-to-reach areas. 8

Maintenance Inspection and Testing Packaging for Sterilization No routine maintenance is required for the sterilization tray. Inspect the tray for dents and cracks. If a problem is observed or suspected, the tray should be returned for repair. Inspect all components for cleanliness. If fluid or tissue buildup is present, repeat the above cleaning and disinfection procedures. Double-wrap the tray unless otherwise noted below. Sterilization Warning: Place the stopcock of the devices (where applicable) in the open position prior to sterilization else non sterile product may result. Caution: Only devices marked autoclavable are compatible with steam sterilization methods. Using steam sterilization on devices that do not bear this marking can cause permanent product damage. Prepare the device(s) and the sterilization tray as indicated in the Sterilization Tray Setup section below. Sterilize the device(s) inside the tray according to the parameters indicated below. Autoclave (Steam) Prevacuum Flash Prevacuum Wrapping double Temperature 132 C (270 F) 132 C (270 F) Time 4 minutes 4 minutes Dry Time 40 minutes 9

Warning: Drying time depends on several variables, including: altitude, humidity, type of wrap, preconditioning, size of chamber, mass of load, material of load, and placement in chamber. Users must verify that drying time set in their autoclave yields dry surgical equipment. Warning: These instructions were developed using the guidance from AAMI TIR 12, ISO 17665 and AAMI ST79 and Stryker recommends users observe these standards. Storage Additional Information Trays with non-absorbent liners (such as plastic or silicone-fingered organizing mats) can cause condensate to pool. To preserve sterility, remove devices from standing solution. N/A Manufacturer Contact Stryker Endoscopy 5900 Optical Court San Jose, CA 95138 +1.800.624.2244, +1.408.754.2000 www.stryker.com 10

Sterilization Tray Setup Maximum Weight Load (devices and tray combined) Internal Stacking External Stacking Accessories Device Distribution Chemical Indicator* Biological Indicator* 1.31 kg (2.9 lbs) No internal stacking is permitted with this tray. Do not stack other trays or devices on or below this tray. There are no accessories available for use with this tray. Devices should be placed in the tray as illustrated below. Place the chemical indicator ( ) in the upper-right corner of the tray as illustrated in the following picture. Place the biological indicator ( ) on any scope tip as illustrated in the following picture. *Note: Placement of the biological or chemical indicator is not required during sterilization of this tray. If, per hospital procedure, placement of an indicator is desired, the recommended placement sites are as follows. 11

References ISO 17664 Sterilization of medical devices Information to be provided by the manufacturer for the processing of resterilizable medical devices ISO 17665-1 Sterilization of health care products Moist heat Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices ANSI/AAMI ST77 Containment devices for reusable medical device sterilization ANSI/AAMI ST79 A comprehensive guide to steam sterilization and sterility assurances in health care facilities ANSI/AAMI ST81 Sterilization of medical devices Information to be provided by the manufacturer for the processing of resterilizable medical devices AAMI TIR12 Designing, testing, and labeling reusable medical devices for reprocessing in health care facilities: a guide for device manufacturers 12

Produced for Stryker Endoscopy. 5900 Optical Court. San Jose, CA 95138 USA. +1.800.624.4422. +1.408.754.2000. www.stryker.com European Representative. Regulatory Manager, Stryker France. ZAC Satolas Green Pusignan. Ave. De Satolas Green. 69881 MEYZIEU Cedex, France P10298 B. 2012/05 Stryker is a registered trademark of Stryker Corporation. ENZOL Enzymatic Detergent is a registered trademark of Advanced Sterilization Products, Division of Ethicon Inc., a Johnson and Johnson company. Renu-Klenz is a trademark of STERIS corporation.