GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants

Similar documents
High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine

Quantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation

Evaluation of an LC-MS/MS Research Method for the Analysis of 33 Benzodiazepines and their Metabolites

The Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis

Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D

Nitrogen/Protein Determination in Starch by Flash Combustion using Large Sample Weight as an Alternative to the Kjeldahl Method

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

Dayana Argoti, Kerry Hassell, Sarah J. Fair, and Joseph Herman Thermo Fisher Scientific, Franklin, MA, USA

LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine

Enhanced LC-MS Sensitivity of Vitamin D Assay by Selection of Appropriate Mobile Phase

LC-MS/MS Method for the Determination of Tenofovir from Plasma

Determination of Clinically Relevant Compounds using HPLC and Electrochemical Detection with a Boron-Doped Diamond Electrode

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Integrated Targeted Quantitation Method for Insulin and its Therapeutic Analogs

Results of Using Re-washed Vials and Closures

PIF: Precursor Ion Fingerprinting Searching for a Structurally Diagnostic Fragment Using Combined Targeted and Data Dependent MS n

Impurity Profiling of Carbamazepine by HPLC/UV

Use of a Tandem Mass Spectrometry Research Method for the Analysis of Amino Acids and Acylcarnitines in Dried Blood Spots

Fast and easy separation of 23 drugs of abuse. including high, stable resolution of isobaric opioids from human urine by UHPLC-MS/MS

Thermo Scientific LipidSearch Software for Lipidomics Workflows. Automated Identification and Relative. Quantitation of Lipids by LC/MS

Dynamic Analysis of HIV-Human Protein-Protein Interactions During Infection

FT-Raman Surface Mapping of Remineralized Artificial Dental Caries

Enhancing Sequence Coverage in Proteomics Studies by Using a Combination of Proteolytic Enzymes

Rapid and Direct Analysis of Free Phytosterols by Reversed Phase HPLC with Electrochemical Detection

Measuring Phytosterols in Health Supplements by LC/MS. Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA

Detection and Quantification of Inorganic Arsenic in Fruit Juices by Capillary Ion Chromatography with Suppressed Conductivity Detection

Providing a Universal, One-step Alternative to Liquid-Liquid Extraction in Bioanalysis

Offline and online sample extraction for the quantification of benzodiazepines in human plasma or serum for clinical research

Increased Identification Coverage and Throughput for Complex Lipidomes

Testing robustness: Immunosuppressant drugs in blood with a TSQ Quantis MS for clinical research

Analysis and Quantitation of Cocaine on Currency Using GC-MS/MS. No. GCMS No. SSI-GCMS-1501

A Rapid UHPLC Method for the Analysis of Biogenic Amines and Metabolites in Microdialysis Samples

Thermo Fisher Scientific, Sunnyvale, CA, USA; 2 Thermo Fisher Scientific, San Jose, CA, USA

Secrets to Successful GC-MS/MS Pesticide Method Development

Overview. Introduction. Experimental. Cliquid Software for Routine LC/MS/MS Analysis

Carl Fisher, Terri Christison, Hua Yang, Monika Verma, and Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA

Robust and Fast Analysis of Tobacco-Specific Nitrosamines by LC-MS/MS

Comprehensive Forensic Toxicology Screening in Serum using On-Line SPE LC-MS/MS

Mercury Speciation Determinations in Asian Dietary Supplements

PosterReprint OVERVIEW

Application Note LCMS-108 Quantitation of benzodiazepines and Z-drugs in serum with the EVOQ TM LC triple quadrupole mass spectrometer

Identification and Quantitation of Microcystins by Targeted Full-Scan LC-MS/MS

Solving One of Chromatography s Biggest Dilemmas Proper Sealing of Chromatography Autosampler Vials

Analyze Drugs of Abuse with Agilent J&W Ultimate Plus Tubing in an Inert Flow Path

4.5 Minute Analysis of Benzodiazepines in Urine and Whole Blood Using LC/MS/MS and an Ultra Biphenyl Column

Internal Calibration System of Thermo Scientific Varioskan Flash with Improved Sensitivity, Accuracy and Dynamic Range

Highly Repeatable Ultra Low Detection of Estradiol Using Triple Quadrupole GC/MS in NCI Mode

Improving Selectivity in Quantitative Analysis Using MS 3 on a Hybrid Quadrupole-Linear Ion Trap Mass Spectrometer

Determination of Benzodiazepines in Urine by CE-MS/MS

Rapid and Robust Detection of THC and Its Metabolites in Blood

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

Increasing Extraction Efficiency of Wet Samples Using a Novel New Polymer During Accelerated Solvent Extraction

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

Determination of Carbohydrates in Kombucha Using HPAE-PAD

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

Analysis of Phospholipids in Natural Samples by Normal Phase HPLC and Corona Charged Aerosol Detection

Determination of red blood cell fatty acid profiles in clinical research

[application note] Simultaneous detection and quantification of D 9 THC, 11-OH-D 9 T H C and D 9 THC-COOH in whole blood by GC tandem quadrupole MS

Improved Extraction and Analysis of Hexavalent Chromium from Soil and Water

Determination of Amantadine Residues in Chicken by LCMS-8040

ApplicationNOTE A MULTI-RESIDUE LC/MS/MS METHOD FOR THE DETERMINATION OF 81 PESTICIDE RESIDUES IN FRUIT AND VEGETABLES: PART 1, METHOD OVERVIEW

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

Metabolomics Strategies

LC-MS/MS Method for the Determination of Raloxifene and its Glucuronide Metabolites from Human Plasma Using SPE Micro Elution

Analysis of Acrylamide in French Fries using Agilent Bond Elut QuEChERS AOAC kit and LC/MS/MS

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Determination of Selected Illegal Drugs and its Important Metabolites in Waste Water by Large Volume Direct Injection HPLC-MS/MS

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

Analysis of Emulsifiers in Foods by High Pressure Liquid Chromatography and Corona Charged Aerosol Detection

Using Multiple Mass Defect Filters and Higher Energy Collisional Dissociation on an LTQ Orbitrap XL for Fast, Sensitive and Accurate Metabolite ID

Toxicology Screening of Whole Blood Extracts Using GC/Triple Quadrupole/MS

Meeting Challenging Requirements for the Quantitation of Regulated Growth Promoters Dexamethasone and Betamethasone in Liver and Milk

Rapid and sensitive UHPLC screening of additives in carbonated beverages with a robust organic acid column

Uptake and Metabolism of Phthalate Esters by Edible Plants

MALAYSIAN MALAYSIAN COCOA WORKSHOP ON THE SAFE USE OF PESTICIDES IN COCOA AND HARMONIZED

Key Advantages of Comprehensive Cannabis Analysis

Essential Lipidomics Experiments using the LTQ Orbitrap Hybrid Mass Spectrometer

Thermo Scientific Dionex CarboPac MA1 Column

PosterREPRINT INTRODUCTION. 2-D PAGE of Mouse Liver Samples. 2-D PAGE of E.coli Samples. Digestion / Cleanup. EXPERIMENTAL 1-D PAGE of BSA Samples

Extraction of a Comprehensive Steroid Panel from Human Serum Using ISOLUTE. SLE+ Prior to LC/MS-MS Analysis

Join the mass movement towards mass spectrometry

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

Application of LC/Electrospray Ion Trap Mass Spectrometry for Identification and Quantification of Pesticides in Complex Matrices

Detection of oxygenated polycyclic aromatic hydrocarbons (oxy-pahs) in APCI mode with a single quadrupole mass spectrometer

Determination of Gamma-Hydroxy-Butyrate (GHB) in Biological Samples

A NOVEL METHOD OF M/Z DRIFT CORRECTION FOR OA-TOF MASS SPECTROMETERS BASED ON CONSTRUCTION OF LIBRARIES OF MATRIX COMPONENTS.

A Fully Integrated Workflow for LC-MS/MS Analysis of Labeled and Native N-Linked Glycans Released From Proteins

SWATH Acquisition Enables the Ultra-Fast and Accurate Determination of Novel Synthetic Opioids

Validation Report 23 B

ETHYL GLUCURONIDE DETERMINATION IN HAIR AS AN INDICATOR OF CHRONIC ALCOHOL ABUSE: A FULLY VALIDATED METHOD BY GC-EI-MS/MS

Determination of Polymerized Triglycerides by High Pressure Liquid Chromatography and Corona Veo Charged Aerosol Detector

DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS

Gas Chromatography Syringe Selection Guide

Rapid, Simple Impurity Characterization with the Xevo TQ Mass Spectrometer

Determination of Bath Salts (Pyrovalerone Analogs) in Biological Samples

Method: Schematic of Method. 2. Introduction. 3. Scope

Transcription:

GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants Jeremy Matthews, 1 Alex Chen, 2 and Flavio Bedini 1 1 Thermo Fisher Scientific, Singapore; 2 Alpha Analytical Pte Ltd, Singapore

Overview Purpose: Assess the feasibility of using analyte protectants to improve the analysis of benzodiazepines. Methods: Benzodiazepines were analyzed using triple-quadrupole GC-MS/MS. The mass spectrometer was run in EI mode using 3 SRM transitions per analyte. Results: The calibration linearity of the benzodiazepines varied significantly between compounds; diazepam showed excellent linearity while lorazepam, lectopam, clonazepam and nitrazepam showed very poor responses. Sorbitol was used as analyte protectant to significantly improve the calibration behavior of the more challenging analytes. Introduction Benzodiazepines drugs are psychoactives whose structure contains a benzene ring fused with a diazepine ring. One of the most well known benzodiazepines diazepam has been marketed under the name Valium since 1963. These drugs are effective tranquilizers and as such are commonly used in medication to treat anxiety and sleep disorders amongst other conditions. The relative availability of these drugs combined with their sedative effect has led to their illicit use as either recreational drug and sometimes in suicides. Consequently, it is common practice to analyze for benzodiazepines in forensic and toxicology laboratories. Benzodiazepines compounds are bases and readily react with active sites in eg. the GC inlet liner causing problems in analysis at low levels and resulting in poor linearity and reproducibility. The use of analyte protectants reduces liner activity and often enables the detection of such active compounds at much lower levels. The use of analyte protectants in the GC-MS analysis of a group of benzodiazepines was investigated and the results are discussed below. 2 GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants

Methods Analytes The following benzodiazepines were analyzed in this work: lorazepam, diazepam, lectopam, nitrazepam, clonazepam. Standards of these analytes were spiked into a matrix. GC-MS System All measurements were carried out using the Thermo Scientific TSQ 8000 triple quadrupole GC-MS/MS system equipped with the Thermo Scientific TRACE 1310 GC with SSL Instant Connect SSL module and Thermo Scientific TriPlus RSH autosampler. The method details are given in Table 1 below. TABLE 1. Instrument method for benzodiazepine analysis TRACE 1310 GC Injection mode splitless Splitless Time 1.0 min GC Column Trace Gold TG-5SilMS, 15 m 0.25 mm 0.25 μm Carrier gas He(99.999 %) Flow 1.2 ml/min, constant flow Temperature program 50 C, 2 min 20 C/min to 300 C, 2 min Transfer line temperature 280 C Total analysis time 14.6 min TriPlus RSH Autosampler Injection volume 1 μl TSQ 8000 MS/MS Ionization mode EI, 70 ev Ion source temperature 250 C Scan mode SRM using timed SRM SRM transition setup automatically build-up by AutoSRM software, transitions see Table 2 Thermo Scientific Poster Note PN10385_ISCC_2014_E_05/14S 3

Results AutoSRM study Benzodiazepine standards were purchased, and calibration solutions were prepared between 10 and 10000 ppb. The 100 ppb solution was then used for an AutoSRM study to determine the optimal SRM transitions and for each compound, and thereby create the MS/MS method. AutoSRM is a unique MS/MS method development tool included in the TSQ 8000 GC-MS/MS software suite which uses a vial containing a standard solution of the compounds required for the method, in the TriPlus RSH autosampler and automatically optimizes retention time, precursor and product masses and collision energy for quan and confirming ions. The AutoSRM study was completed with only minimal user interaction required, and the optimized SRM transitions are shown in Table 2 below. TABLE 2. Instrument method for benzodiazepine analysis Compound CAS Number RT Precursor Mass Product Mass Collision Energy (min) (m/z) (m/z) (ev) Lorazepam 846-49-1 14.32 239.0 176.7 25 Lorazepam 846-49-1 14.32 274.0 239.1 10 Diazepam 439-14-5 14.45 256.1 165.1 30 Diazepam 439-14-5 14.45 256.1 221.1 10 Lectopam 1812-30-2 15.32 316.7 288.0 20 Lectopam 1812-30-2 15.32 316.7 289.1 10 Nitrazepam 146-22-5 16.02 280.3 205.1 30 Nitrazepam 146-22-5 16.02 280.3 234.1 10 Clonazepam 1622-61-3 16.51 280.0 234.1 10 Clonazepam 1622-61-3 16.51 314.0 268.1 15 Benzodiazepine Calibration Curves The optimized MRM method created using AutoSRM (described above) was used to record calibration data for the benzodiazepines at up to 10 000 ppb. The linearity of the calibration curves varied widely with the particular analyte as shown in Figures 1 (a) (d) below. 4 GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants

FIGURE 1. Calibration curves for (a) diazepam, (b) nitrazepam, (c) lorazepam and (d) lectopam at up to 10,000 ppb. (a) (b) (c) (d) The calibration curves for diazepam was very linear (R2 > 0.99) between 5 and 10,000 ppb also exhibited excellent peak shape. The other benzodiazepines, however, strongly deviated from linearity, and a quadratic was required to achieve a reasonable fit of the data (also for clonazepam, not shown). The peak areas for these compounds were also fairly low and showed poor signal-to-noise at relatively high concentrations. The difficulties encountered in analysis of such polar compounds have been previously reported in the analysis of pesticides 1, 2, 3. The high activity of polar functional groups has been observed to lead to poor recoveries and significant loss of analytes at low concentration due to trapping and degradation at active sites in the gas flow path. The degradation of trapped compounds from previous injections can then form further active sites (increased surface area), which may rapidly lead to poor chromatographic response and performance. Thermo Scientific Poster Note PN10385_ISCC_2014_E_05/14S 5

FIGURE 2. Chemical structure of the benzodiazepine compounds analyzed in this work. Diazepam Nitrazepam Lectopam Lorazepam Chlonazepam It has been previously shown that the use of analyte protectants, such as sorbitol is very effective in correcting this behavior for eg. pesticide analysis, whereby the protectant reacts with active sites in the GC flowpath effectively deactivating the system and enhancing analyte recovery, especially at low levels. This has several advantages including improving peak shape (less tailing) and hence repeatability (RSDs), as well as keeping the inlet liner clean for longer durations as analyte and matrix are no longer trapped and subsequently degraded. 6 GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants

FIGURE 3. Calibration curves for (a) clonazepam, (b) nitrazepam, (c) lorazepam and (d) lectopam between 10 and 200 ppb using 0.2 % sorbitol as analyte protectant. (a) (a) (b) (b) (c) (d) Analyte protectant (0.2% sorbitol) was added to the benzodiazepine calibration standards, and the calibration curves were measured as shown above in Figure 3. The calibration linearity improved dramatically for lorazepam, lectopam, clonazepam and nitrazepam with the sorbitol addition. The response factors for lectopam and lorazepam increased by roughly 16 and 7 times respectively indicating a dramatic improvement in sensitivity. Thermo Scientific Poster Note PN10385_ISCC_2014_E_05/14S 7

Conclusions Benzodiazepines are challenging to analyze by GC- MS/MS due to high activity as a result of polar functional groups. Activity of inlet liner and GC column result in adsorption/degradation of some benzodiazepine analytes leading poor linearity of calibration and sensitivity. The use of sorbitol as analyte protectant dramatically improved the linearity of the response, as well as signal intensity for these compounds. All benzodiazepines could be quantified at 10 ppb using analyte protectant. References 1. Anastassiades, M., Maštovská, K., Lehotay, S.J., Evaluation of analyte protectants to improve gas chromatographic analysis of pesticides, J. Chromatogr. A 1015 (2003) 163-184. 2. Mastovska, K., Lehotay, S.J., Anastassiades, M., Combination of Analyte Protectants To Overcome Matrix Effects in Routine GC Analysis of Pesticide Residues in Food Matrixes, Anal. Chem. 77(24) (2005) 8129-8137. 3. Li, Y., Chen, X., Fan, C., Pang, G., Compensation for matrix effects in the gas chromatography mass spectrometry analysis of 186 pesticides in tea matrices using analyte protectants, J. Chrom. A 1266 (2012), 131-142. www.thermoscientific.com 2014 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Africa +43 1 333 50 34 0 Australia +61 3 9757 4300 Austria +43 810 282 206 Belgium +32 53 73 42 41 Canada +1 800 530 8447 China 800 810 5118 (free call domestic) 400 650 5118 Denmark +45 70 23 62 60 Europe-Other +43 1 333 50 34 0 Finland +358 9 3291 0200 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9494 Italy +39 02 950 591 Japan +81 45 453 9100 Latin America +1 561 688 8700 Middle East +43 1 333 50 34 0 Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Norway +46 8 556 468 00 Russia/CIS +43 1 333 50 34 0 Thermo Fisher Scientific, Austin, TX USA is ISO 9001:2008 Certified. Singapore +65 6289 1190 Spain +34 914 845 965 Sweden +46 8 556 468 00 Switzerland +41 61 716 77 00 UK +44 1442 233555 USA +1 800 532 4752 PN10385_E 05/14S