How to Retain Polar and Nonpolar Compounds on the Same HPLC Stationary Phase with an Isocratic Mobile Phase

Similar documents
Maximizing chromatographic peak capacity with the Agilent 1290 Infinity LC system

Using Hydrophilic Interaction Chromatography (HILIC) for the Retention of Highly Polar Analytes

Increasing resolution using longer columns while maintaining analysis time Advantages of the wide power range of the Agilent 1290 Infinity LC System

Ultra Performance Liquid Chromatography Coupled to Orthogonal Quadrupole TOF MS(MS) for Metabolite Identification

LC/MS/MS Separation of Cholesterol and Related Sterols in Plasma on an Agilent InfinityLab Poroshell 120 EC C18 Column

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System

Profiling of Endogenous Metabolites Using Time-of-Flight LC/MS with Ion Pair Reverse Phase Chromatography

Identification of Steroids in Water by Ion Trap LC/MS/MS Application

Impact of Reversed-Phase Chiral Chromatography on the LC-MS Analysis of Drugs in Biological Fluids

John Haselden 1, Gordon Dear 1, Jennifer H. Granger 2, and Robert S. Plumb 2. 1GlaxoSmithKline, Ware, UK; 2 Waters Corporation, Milford, MA, USA

Effective use of Pharmacopeia guidelines to reduce cost of chromatographic analysis for Fluticasone propionate

Quantitative Analysis of Underivatized Amino Acids in Plant Matrix by Hydrophilic Interaction Chromatography (HILIC) with LC/MS Detection

Application Note. Abstract. Authors. Pharmaceutical

Extended Application Note

Chromatography Vacuum Ultraviolet Spectroscopy

Automated Purification and Analytical Reinjection of a Small Molecule Drug, Probenecid, on a Gilson LC/MS Dual Function System

SeQuant ZIC -HILIC For all who expect more...

Authors. Abstract. Introduction. Environmental

A New HILIC/RP Mixed-Mode Column and Its Applications in Surfactant Analysis

IAM Chromatography. Introduction

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

Polymer Additive Analysis by EI and APCI

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract.

The Raptor HILIC-Si Column

Qualitative and quantitative determination of phenolic antioxidant compounds in red wine and fruit juice with the Agilent 1290 Infinity 2D-LC Solution

Analysis of Amino Acids Derived Online Using an Agilent AdvanceBio AAA Column

High Resolution Glycopeptide Mapping of EPO Using an Agilent AdvanceBio Peptide Mapping Column

LC/MS Method for Comprehensive Analysis of Plasma Lipids

Separation of Macrocyclic Lactones (Avermectins) on FLARE C18 MM & FLARE C18+ Columns

Choosing the metabolomics platform

2D-LC as an Automated Desalting Tool for MSD Analysis

Facile Isolation of Carotenoid Antioxidants from Solanum lycopersicum using Flash Chromatography

Simultaneous Analysis of Active Pharmaceutical Ingredients and Their Counter-Ions Using a Mixed-Mode Column

Evaluation of a Silica Hydride Based Diol Stationary Phase For High Pressue Liquid Chromatogrphy

Pelagia Research Library

Unique Retention and Selectivity of Pentafluorophenylpropyl Phases for High-Throughput LC/MS Analysis

Analysis of several common. organic acids in tobacco leaf, snus, and moist snuff

IJPAR Vol.3 Issue 4 Oct-Dec-2014 Journal Home page:

Applying a Novel Glycan Tagging Reagent, RapiFluor-MS, and an Integrated UPLC-FLR/QTof MS System for Low Abundant N-Glycan Analysis

Vitamin D3 and related compounds by ESI and APCI

Impurity Profiling of Carbamazepine by HPLC/UV

Probing for Packaging Migrants in a Pharmaceutical Impurities Assay Using UHPLC with UV and Mass Detection INTRODUCTION

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

Quantitative Analysis of Carbohydrates and Artificial Sweeteners in Food Samples Using GFC- MS with APCI Interface and Post-column Reagent Addition

GlycanPac AXH-1 Columns

Measuring Lipid Composition LC-MS/MS

Separation of Vitamin D and Vitamin D Metabolites on FLARE C18 MM (Mixed Mode) HPLC Column

Application Note. Small Molecule Pharmaceuticals. 2-Ethyl-2-phenylmalonamide Ethosuximide Primidone Carbamazepine. Carbamazepine-

The HPLC Preparative Scale-Up of Soybean Phospholipids Application

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

Measuring Phytosterols in Health Supplements by LC/MS. Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA

Performance Characteristics of the Agilent 1290 Infinity II Multicolumn Thermostat

Agilent 1260 Infinity Analytical SFC System

Detection of oxygenated polycyclic aromatic hydrocarbons (oxy-pahs) in APCI mode with a single quadrupole mass spectrometer

Journal of Chemical and Pharmaceutical Research

Determination of Aflatoxins in Food by LC/MS/MS. Application. Authors. Abstract. Experimental. Introduction. Food Safety

Performance characteristics of the Agilent 1100 Series preparative pump. Technical Note. Abstract

Steviol Glycosides from Stevia rebaudiana Bertoni

Neosolaniol. [Methods listed in the Feed Analysis Standards]

Edgar Naegele. Abstract

SeQuant ZIC -HILIC For all who expect more...

Separation of 15 Underivatized Saccharide and Sialic Acid USP Standards

Evaluating the Isolation and Quantification of Sterols in Seed Oils by Solid-Phase Extraction and Capillary Gas Liquid Chromatography

Qualitative and quantitative determination of cannabinoid profiles and potency in CBD hemp oil using LC/UV and Mass Selective Detection

Analysis of Counterfeit Antidiabetic Drugs by UHPLC with the Agilent 1220 Infinity Mobile LC

Transferring a Method for Analysis of DNPH-Derivatized Aldehydes and Ketones from HPLC to UHPLC

Application Note. Authors: C. Ledesma, M. Gibert, J.R. Gibert Ingenieria Analitica S.L. Extracts from various food products

Studies on Stationary Phase Selectivity for Solid-Core Particles

Chapter 22: Gas and Liquid Chromatography. Liquid Chromatography: Stationary Phases and Mobile Phases: Normal Phase: Reversed Phase:

Separation of Polyphenols by Comprehensive 2D-LC and Molecular Formula Determination by Coupling to Accurate Mass Measurement

Comprehensive Two-Dimensional HPLC and Informative Data Processing for Pharmaceuticals and Lipids

Converting a CHP Method for Insulin to Agilent Poroshell 120 Columns

Rapid Gradient and Elevated Temperature UHPLC of Flavonoids in Citrus Fruit

Fig. 1: Chemical structure of arachidonic acid COOH CH 3

F. Al-Rimawi* Faculty of Science and Technology, Al-Quds University, P.O. Box 20002, East Jerusalem. Abstract

APPLICATIONS TN µm Core-Shell Particle. (continued on page 2) Page 1 of 8. For additional technical notes, visit

Key Words: Brassica oleraceae, glucosinolate, liquid chromatography mass spectrometry, FNH-I-003

ASSAY AND IMPURITY METHOD FOR DURACOR TABLETS BY HPLC

Application of LC/Electrospray Ion Trap Mass Spectrometry for Identification and Quantification of Pesticides in Complex Matrices

Introduction. and aryl-α-amino ketones, 11

Utility of Pentafluorophenylpropyl Stationary Phases for RP-HPLC Analyses: Biogenic Amines on Discovery HS F5

Doubling the throughput of long chromatographic methods by using a novel Dual LC workflow

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

14 Separation of the Phosphatidylcholines Using Reverse Phase HPLC

O O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions

Comparison of a UPLC Method across Multiple UHPLC Systems

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

InertSustain Phenylhexyl. A New Second Choice Column to Change Chromatographic Behavior

LC/MS Analysis of Various Hydrophilic Compounds Using a Polymer-Based Amino Column - Shodex TM HILICpak TM VG-50 2D

Single Quadrupole Compact Mass Spectrometer for Liquid Sample

Gerry Hendrickx Regional Sales Manager Central Europe. Developments in Waters Column Chemistries : BEH Technology

Simple Method (IS-MRM) to Monitor Lysophospholipids and Phospholipids During LC-MS Method Development via In-Source CID

Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

Lipidomic Analysis by UPLC-QTOF MS

Rapid Lipid Profiling of Serum by Reverse Phase UPLC-Tandem Quadrupole MS

Chapter 6 IDENTIFICATION AND CHARACTERIZATION OF FLAVONOIDS BY HPLC AND LC-MS/MS ANALYSIS

Analysis of HMF by HPLC

Fast Separation of Triacylglycerols in Oils using UltraPerformance Convergence Chromatography (UPC 2 )

Transcription:

296 LCGC NORTH AMERICA VOLUME 24 NUMBER 3 MARCH 26 www.chromatographyonline.com How to Retain Polar and Nonpolar Compounds on the Same HPLC Stationary Phase with an Isocratic Mobile Phase Separation and retention of both polar and nonpolar compounds by the same stationary phase can be a useful approach for analyses of complex samples with a broad range of chemical properties. Typical stationary phases are designed for retention of either polar or nonpolar compounds so that multiple steps are required when designing a separation strategy. Hydride silica based stationary phases are new materials with properties that allow for the simultaneous retention of both polar and nonpolar compounds over a range of aqueous organic mobile phase compositions. Adjustment of the aqueous organic ratio will determine whether polar or nonpolar compounds have greater retention. Joseph J. Pesek, Ph.D., and Maria T. Matyska, Ph.D. Department of Chemistry, San Jose State University Please direct correpsondence to Joseph Pesek at pesek@sjsu.edu. High performance liquid chromatography (HPLC) stationary phases can be segregated by their ability to separate either polar on nonpolar compounds, that is, reversedphase materials (C18, C8) strongly retain nonpolar solutes with polar solutes eluting at or near the void volume, and hydrophilic interaction chromatography (HILIC) and normal phase columns strongly retain polar analytes with nonpolar compounds being essentially nonretained (1). Increasingly, many analyses such as those encountered in drug discovery, proteomics, and metabolomics can be more complex with solutes encompassing a broad range of polarities. To overcome these deficiencies in column performance, more complex schemes of analysis might have to be devised to provide successfully qualitative and quantitative information about solutes with differing hydrophobicities and hydrophilicities. These approaches can include various types of sample preparation or two-dimensional chromatographic methods (2) as well as using chromatographic extremes of ph (3) and temperature (4). In most cases, such methodology can be cumbersome, time consuming, and damaging to your instrument and the HPLC column. In many instances, it would be desirable to have a stationary phase that can retain both polar and nonpolar compounds in an isocratic run so that a single separation strategy can be devised to analyze samples with a broad range of polarities. Recently, a third type of chromatographic strategy has been developed on stationary phases utilizing siliconhydride-based particles with a bonded phase (5 15). It is referred to as aqueous normal phase (ANP) chromatography. The principle of ANP chromatography is simple: retention behavior is analogous to that found in normal phase chromatography but the mobile phase has some water as part of the binary solvent. Normal phase implies that retention is

298 LCGC NORTH AMERICA VOLUME 24 NUMBER 3 MARCH 26 www.chromatographyonline.com 25 Retention time (min) 2 15 1 5 Metformin - methanol Metformin - acetonitrile Glyburide - methanol Glyburide - acetonitrile 4 5 6 7 8 9 1 organic in DI water +.5 FA Figure 1: Retention time versus organic solvent percentage in the mobile phase (retention map) for metformin and glyburide on a siliconhydride-based Cogent Bidentate C18 column using methanol and acetonitrile. Flow rate:.5 ml/min. greatest for polar solutes such as acids and bases. In addition, retention must increase as the amount of the nonpolar solvent in the mobile phase increases. So if the mobile phase consists of water and acetonitrile, retention increases as the

www.chromatographyonline.com (a) (b) 1 1 Glyburide m/z 369 2.99 2.37 Metformin m/z 13 2 4 6 8 1 12 14 16 Time (min) Figure 2: Chromatograms for metformin and glyburide obtained using a 75 mm 4.6 mm silicon-hydride-based Cogent Bidentate C18 column in (a) 8:2 acetonitrile water and (b) 5:5 acetonitrile water. Flow rate:.5 ml/min. Detection in the single ion monitoring mode by APCI. amount of acetonitrile increases. Typically, in ANP chromatography, the amount of the nonpolar component in the mobile phase must be 6 or greater with the exact point of increased retention depending upon the solute and the organic component of the mobile phase. If a stationary phase had the retention properties previously described and could only separate polar solutes, then it would be similar to a HILIC material (16 18). The term ANP is useful in distinguishing the hydride-based phases from typical HILIC phases. The hydride stationary phases also can retain nonpolar compounds by a traditional reversed-phase mechanism during the same isocratic run as described previously. Therefore, it is this dual retention capability that distinguishes the silicon-hydride material from other silica-based HPLC stationary phases. ANP chromatography is a useful term for indicating retention of ionizable polar compounds on a siliconhydride stationary phase also possessing reversed-phase capabilities as opposed 6.26 Metformin m/z 13 Glyburide m/z 369 13.41 to a HILIC process that can separate only polar solutes. Therefore, it is this dual retention capability that distinguishes the silicon-hydride material from other silica-based HPLC stationary phases. MARCH 26 LCGC NORTH AMERICA VOLUME 24 NUMBER 3 299 Experimental Instrumentation: All chromatographic experiments utilized a model 15 HPLC system with a diode-array detector (Agilent Technologies, Wilmington, Delaware) and interfaced to a MicroMass mass spectrometer (Waters Corp., Milford, Massachusetts). The Micromass Platform LC system was equipped with an Edwards model E2M3 rotary vacuum pump (Chell Instruments, Norfolk, UK) a Micromass Platform LC model M9415DC1 atmospheric pressure chemical ionization (APCI) probe, and a computer-based data acquisition system with MassLynx (version 3.4) software (Waters). The instrument was purged with high-pressure liquid nitrogen gas (1 psi). The acquisition parameters used for all the separations were as follows: APCI pin 3.2, cone 25, skimmer 2., source heater 14 C, APCI probe temperature 6 C, gas flow 25 ml/min..5 formic acid was added to the mobile phase for ionizing the test samples. All liquid chromatography mass spectrometry (LC MS) separations were performed using the APCI probe in the positive ion mode. Chemicals: Metformin and glyburide were obtained from Eon Pharmaceutical (Wilson, North Carolina). HPLC-grade acetonitrile (Spectrum Chemical, Gardena, California) and methanol (VWR, West Chester, Pennsylvania) were used as organic modifiers in the mobile phase. Formic acid (Sigma-Aldrich, St. Louis, Missouri) was added to the mobile phase as the APCI reagent. Deionized water for HPLC system was obtained from a Milli- Q water purification system (Millipore, Bedford, Massachusetts). HPLC columns: Three columns were used in the chromatographic testing: a Cogent Bidentate C18 column (Microsolv Technology, Eatontown, New Jersey), a Luna (2) C18 column (Phenomenex, Torrance, California), and a Zorbax Extend C18 column (Agilent Technologies). All columns were obtained from the manufacturer. Results and Discussion Example of simultaneous reversed- and normal-phase retention: A comparison of a compound that exhibits only reversed-phase behavior on a siliconhydride phase (the nonpolar molecule glyburide) to one with ANP retention (the highly polar compound metformin) is shown by the retention maps in Figure 1 along with their structures. Each compound is run on the bidentate C18 Type column in two solvents (methanol and acetonitrile). For glyburide ( and ), retention decreases as the amount of either methanol or acetonitrile increases in the mobile phase; typical reversed-

3 LCGC NORTH AMERICA VOLUME 24 NUMBER 3 MARCH 26 www.chromatographyonline.com (a) 3 Retention time (min) 25 2 15 1 5 Metformin Glyburide 4 5 6 7 8 9 1 Acetonitrile in DI water +.5 formic acid (b) Retention time (min) 2 18 16 14 12 1 8 6 4 2 Metformin Glyburide 4 5 6 7 8 9 1 Acetonitrile in DI water +.5 formic acid Figure 3: Retention time for metformin and glyburide versus acetonitrile percentage in the mobile phase on commercial columns: (a) 25 mm 4.6 mm Phenomenex Luna (2) C18 and (b) 15 mm 4.6 mm Agilent Zorbax Extend C18. The conditions were the same as in Figure 2. phase behavior. For metformin ( and ), retention increases as the amount of acetonitrile is added to the mobile phase; typical normal-phase behavior. For methanol, only a slight increase is observed above 8 organic in the mobile phase. An interesting feature of this plot is that for acetonitrile, it is possible for the highly polar compound metformin and the nonpolar compound glyburide to be coeluted with both having reasonably good retention. This coelution would occur at 7 acetonitrile with a retention time of over 4 min. Thus, it is possible to retain both polar and nonpolar compounds on the same stationary phase by utilizing the unique properties of the silicon-hydride stationary phase. The polar compounds are retained by the ANP mechanism while the nonpolar solutes undergo typical reversed-phase behavior. It can be said that under these circumstances, the use of a silicon-hydride-based stationary phase allows for separation in two dimensions on a single column. At present, a comprehensive understanding of the simultaneous reversed-phase aqueous normal retention mechanisms of the silicon-hydride phases has not been developed. Both chromatographic and spectroscopic investigations are underway to determine the sources of this behavior. This phenomenon could be due to the hydride surface alone, the absence of substantial amounts of adsorbed water on the surface, a magnified effect of the few remaining silanols, or some combination of these three possible properties. The stronger retention in acetonitrile is likely a polarity effect. It is less polar than methanol and as expected in a normal-phase mechanism, the least polar solvent will result in the greatest retention. The actual chromatograms for retention reversal are shown in Figure 2. In Figure 2a, the conditions are such (8 acetonitrile) that the more polar solute (metformin) is eluted after the less or nonpolar compound glyburide; in this case, ANP is the dominant mechanism. By decreasing the amount of acetonitrile to 5, it can be seen (Figure 2b) that

www.chromatographyonline.com (a) 1 (b) 1 (c) 1 2.88 3.55 4.52 2 4 6 8 1 Time (min) Figure 4: Comparison of the peak shape for metformin on three stationary phases: (a) Luna (2) C18, (b) Zorbax Extend C18, and (c) Cogent Bidentate C18 (MicroSolv). Mobile phase: 7:3 acetonitrile water. Other conditions were the same as in Figure 2. MARCH 26 LCGC NORTH AMERICA VOLUME 24 NUMBER 3 31 the two compounds now are eluted in the opposite order with glyburide being more retained than metformin; that is, the dominant mechanism is typical reversed-phase interaction. These retention properties can be contrasted to the retention maps for typical reversed-phase materials that are shown in Figure 3. Two well-established C18 phases are tested under the same conditions used for the bidentate C18 column that is based upon silicon-hydride as previously described. In each case, the nonpolar compound glyburide is well retained at low acetonitrile composition in the mobile phase and not retained (at or near the void volume) as the amount of acetonitrile exceeds 8. In contrast, the polar compound metformin is poorly retained or not retained over the whole range of acetonitrile compositions in the mobile phase. Figure 4 shows a comparison of peak shapes for the highly basic compound metformin on these three columns at constant mobile phase composition. The chromatogram on the first column (Figure 4a) demonstrates tailing that is common on many typical Circle 48

32 LCGC NORTH AMERICA VOLUME 24 NUMBER 3 MARCH 26 www.chromatographyonline.com reversed-phase materials indicating the presence of residual silanols on the surface. In contrast, the chromatogram of metformin on the extensively endcapped second C18 column (Figure 4b) results in a highly symmetric peak. The same result is obtained on the silicon-hydridebased C18 column (Figure 4c). However, in contrast to the second column, where the peak is eluted very close to the void volume, the bidentate C18 siliconhydride-based stationary phase has measurable retention for metformin that can Circle 49 be increased by using a mobile phase with a higher amount of acetonitrile. Other properties of silicon-hydridebased phases: Retention of glyburide in the previous example demonstrates the reversed-phase properties of the siliconhydride-based stationary phases. To compare its properties with those of other reversed-phase materials, a standard test based upon the retention of naphthalene and acenaphthene was conducted (19). The hydrophobicity parameter for the bidentate C18 based upon this test was.397. This value was within experimental error the same as the first and second C18 columns tested previously (19). Therefore, the silicon-hydride-based stationary phases have excellent reversedphase properties along with the ANP ability not available in other hydrophobic materials. Another crucial aspect of the ANP mechanism is its reproducibility. The retention of metformin in a 7:3 acetonitrile water mobile phase was measured. For 1 consecutive injections, the retention time was measured as 4.8.1 min with a RSD of.25. Therefore, excellent reproducibility can be obtained in the ANP mode with these columns. Similar ANP retention also has been demonstrated for other polar compounds (choline, acetylcholine, tobramycin, nortriptyline, imipramine, amytripyline, acetylsalicylic acid, 2-hydroxybenzoic acid, and 4-hydroxybenzoic acid) on the silicon-hydride-based stationary phases. Therefore, for many analyses, it might not be necessary to use a high-ph mobile phase to analyze polar compounds, as it would be with many bases. In the previous example, the aqueous component of the mobile phase contained.5 formic acid for compatibility with mass spectral analysis. The ANP at normal ph and formic acid buffered conditions are conducive to both the long-term stability of the stationary phase as well as leading to fewer instrumental problems such as pump repairs, filter replacements, and seal deterioration. In addition, the two mechanism retention characteristics of the siliconhydride materials allow for separation of mixtures of polar and nonpolar compounds where both types of solutes can be retained and separated. Such behavior is not possible on other stationary phases unless some additives are used in the mobile phase to make polar compounds nonpolar and, hence, retained on a reversed-phase material (2) or to make nonpolar species more polar to be retained on a HILIC type stationary phase. These procedures often are not successful for every compound in the sample and usually provide only limited improvement in identifying all of the analytes. In addition, most additives are not amenable to LC MS methods

www.chromatographyonline.com MARCH 26 LCGC NORTH AMERICA VOLUME 24 NUMBER 3 33 because often they are not volatile enough to be introduced into a mass spectrometer. The processes also can be accomplished by precolumn derivatization, but this adds another step to the analysis, which can be time-consuming, cumbersome, and also not successful for all compounds in the sample. Conclusions In summary, the example here can be duplicated with many polar nonpolar solute pairs and illustrates one unique property of silicon-hydride-based HPLC stationary phases. Both peak shape and reproducibility of reversed-phase retained solutes and ANP retained compounds is excellent. These stationary phases are a new tool that can be used to help solve the increasingly complex analytical problems encountered in pharmaceutical, biological, and environmental samples. Acknowledgments The authors would like to thank Micro- Solv Technology Corporation, Eatontown, New Jersey (http://www.mtcusa.com) for donation of the columns used in this study. References (1) L.S. Snyder, J.J. Kirkland, and J.L. Glajch, Practical HPLC Method Development, 2nd ed. (John Wiley & Sons, Hoboken, New Jersey, 1997). (2) P.G. Righetti, A. Castagna, B. Herbert, F. Reymond, and J.S. Rossier, Proteomics 3, 1397 (23). (3) J.J. Kirkland, M.A. van Straten, and H.A. Claessens, J. Chromatogr., A 797, 111 (1998). (4) D.R. Stoll and P.W. Carr, J. Amer. Chem. Soc. 127, 534 (25). (5) J.E. Sandoval and J.J. Pesek, Anal. Chem. 63, 2634 (1991). (6) C.H. Chu, E. Jonsson, M. Auvinen, J.J. Pesek, and J.E. Sandoval, Anal. Chem. 65, 88 (1993). (7) M.C. Montes, C. van Amen, J.J. Pesek, and J.E. Sandoval, J. Chromatogr. 688, 31 (1994). (8) J.J. Pesek, M.T. Matyska, J.E. Sandoval, and E. Williamsen, J. Liq. Chromatogr. 19, 2843 (1996). (9) J.J. Pesek, M.T. Matyska, E.J. Williamsen, M. Evanchic, V. Hazari, K. Konjuh, S. Takhar, and R. Tranchina, J. Chromatogr., A 786, 219 (1997). (1) J.J. Pesek, M.T. Matyska, M. Oliva, and M. Evanchic, J. Chromatogr., A 818, 145 (1998). (11) J.Pesek, M.T. Matyska, and P.F. Fu, Chromatographia 53, 635 (21). (12) J.J. Pesek, M.T. Matyska, and R.J. Yu, J. Chromatogr., A 97, 195 (22). (13) J.J. Pesek, M.T. Matyska, G.B. Dawson, A. Wilsdorf, P. Marc, and M. Padki, J. Chromatogr., A 986, 253 (23). (14) M.T. Matyska, J.J. Pesek, and X. Pan, J. Chromatogr., A 992, 57 (23). (15) L. Brown, B. Ciccone, J.J. Pesek, and M.T. Matyska, Am. Lab. 35 (24), 23 (23). (16) Y. Guo and S. Gaiki, J. Chromatogr., A 174, 71 (25). (17) T. Yoshida, J. Biochem. Biophys. Methods 6, 265 (24). (18) B.A. Olsen, J. Chromatogr., A 913, 113 (21). (19) U.D. Neue, K. Van Tran, P.C. Iraneta, and B.A. Aldon, J. Sep. Sci. 26, 174 (23). (2) J.J. Kirkland, LCGC 14, 486 (1996). Circle 5