AN EXPERIMENTAL STUDY ON TRANSDUCER FOR THE DDS TECHNOLOGY DEMONSTRATOR ANDRZEJ ELMINOWICZ, WALDEMAR LIS*, WŁADYSŁAW MĘCIŃSKI

Similar documents
Descriptions of NDT Projects Fall 2004 October 31, 2004

Multielement ultrasonic probes for projection imaging of biological media

Annular Array Transducer and Matched Amplifier for Therapeutic Ultrasound

Ultrasonic arrays are now widely used in underwater sonar

Application of Phased Array Radar Theory to Ultrasonic Linear Array Medical Imaging System

Ultrasonic Testing Level I:

Development of innovative transducer designs for NDT applications: From 1-3 piezocomposite definition to 2D array probe manufacture

Evaluation of the Quality of Thick Fibre Composites Using Immersion and Air- Coupled Ultrasonic Techniques

Linear Ultrasonic Wave Propagation in Biological Tissues

Pulse-Echo Ultrasound Imaging. Resolution in Ultrasound Imaging. Doppler Ultrasound. Resolution vs Penetration. Medical Imaging (EL582/BE620/GA4426)

High power density prototype for high precision transcranial therapy

Diploma of Medical Ultrasonography (DMU) Physical Principles of Ultrasound and Instrumentation Syllabus

ACOUSTORAPHY Acousto-Optical (AO) Ultrasonics & Its Applications

ULTRASOUND. OB/Gyn (Core) Ultrasound PIEZOELECTRIC EFFECT. Principles of Ultrasound Physics and Instrumentation. Nathan Pinkney, BS, CDOS

Ultrasonic Transducer. Piezoelectric ceramics / Piezoelectric polymer film / Piezoelectric thin film

OIML R 122 Annex C RECOMMENDATION. Edition 1999 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

w. D. Jolly, F. A. Bruton, and C. Fedor

Non-Destructive Inspection of Composite Wrapped Thick-Wall Cylinders

HiFFUT A New Class of Transducer

85/B:43 the Exploration of the Sea Fish Capture Committee. E. Ona and G. Vestnes Institute of Marine Research Bergen, Norway

Hollow Axle Ultrasonic Crack Detection for Conventional Railway Vehicles. Hiromichi ISHIDUKA Assistant Senior Researcher,

Physical Principles of Ultrasound

GUIDANCE NOTE. Application of MEMS Ultrasonic Transducers to Flow Measurement.

Available online at ScienceDirect. Physics Procedia 87 (2016 ) 35 41

Routine Quality Assurance Cookbook

sensors ISSN

Two Modified IEC Ear Simulators for Extended Dynamic Range

Developments in Ultrasonic Inspection II

Application of ultrasonic phased array in acoustic logging

NON-CONTACT ULTRASOUND (NCU)

Ultrasound Physics & Terminology

Principles of Ultrasound. Cara C. Prideaux, M.D. University of Utah PM&R Sports Medicine Fellow March 14, 2012

AN ULTRASONIC MESH SENSOR FOR TWO-PHASE FLOW VISUALISATION

Flaw Assessment Using Shear wave Phased array Ultrasonic Transducer

Lesson 03: Sound Wave Propagation and Reflection. This lesson contains 15 slides plus 14 multiple-choice questions.

PIEZOTRANSDUCERS WITH ACOUSTIC LENSES FORMING NARROW WEAKLY DIVERGING ULTRASONIC BEAMS

PVDF transducer for SAFT imaging of concrete structures.

Dr Emma Chung. Safety first - Physical principles for excellent imaging

10 Years Experience in Industrial Phased Array Testing of Rolled Bars

Preamble (disclaimer)

TG-128: Quality Assurance for Prostate Brachytherapy Ultrasound

STUDY OF PHASED ARRAY TECHNIQUES FOR CONCRETE INSPECTION

Ultrasonic Transducer Optimization Can Increase Sensor Reliability - Even in Caustic Environments

Supplement (videos)

INSPECTION THROUGH AN OVERLAY REPAIR WITH A SMART FLEXIBLE ARRAY PROBE.

What is Ultrasound? Resolution Image production Attenuation Imaging modes Ultrasound artifacts... 7

1 Fundamentals. Basic Definitions and Physics Principles. Fundamentals

CONTENTS. Test Number cpd Tanya Reynolds (Nat. Dip. Diag. Rad., B. Tech. Diag. Rad., B. Tech. Ultrasound)

ULTRASONIC TRANSDUCERS WITH VIBRATING PIEZOELECTRIC PLATES

Basic Physics of Ultrasound and Knobology

Ultrasound Measurements and Non-destructive Testing Educational Laboratory

Ultrasound. Principles of Medical Imaging. Contents. Prof. Dr. Philippe Cattin. MIAC, University of Basel. Oct 17th, 2016

A micro ultrasonic motor using a micro-machined cylindrical bulk PZT transducer

DEVELOPMENT OF ULTRASONIC TESTING TECHNIQUE TO INSPECT CONTAINMENT LINERS EMBEDDED IN CONCRETE ON NUCLEAR POWER PLANTS

Introduction to Biomedical Imaging

ULTRASOUND QA SOLUTIONS. Ensure Accurate Screening, Diagnosis & Monitoring DOPPLER FLOW PHANTOMS MULTI-PURPOSE PHANTOMS TRANSDUCER TEST PHANTOMS

Research on Digital Testing System of Evaluating Characteristics for Ultrasonic Transducer

ULTRASOUND QA SOLUTIONS. Ensure Accurate Screening, Diagnosis & Monitoring DOPPLER FLOW PHANTOMS MULTI-PURPOSE PHANTOMS TRANSDUCER TEST PHANTOMS

Ultrasound Principles cycle Frequency Wavelength Period Velocity

Optimization of Phased-Array Transducers for Ultrasonic Inspection in Composite Materials Using Sliding Probes

Underwater Acoustic Measurements in Megahertz Frequency Range.

Augmentation of the In Vivo Elastic Properties Measurement System to include Bulk Properties

Modelling and Fabrication of High Frequency Ultrasound Transducer Arrays for Medical Applications

ULTRASOUND QA SOLUTIONS. Ensure Accurate Screening, Diagnosis and Monitoring DOPPLER FLOW PHANTOMS MULTI-PURPOSE PHANTOMS TRAINING PHANTOMS

1. SCOPE ELIGIBILITY EXAMINATION CONTENT RENEWAL & RECERTIFICATION PROCEDURE ESSENTIAL READING...

Research on a Transmit-Receive Method of Ultrasonic Array for Planar Defects

Pipeline Technology Conference 2007

Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion

Oil Transmission Pipelines Condition Monitoring Using Wavelet Analysis and Ultrasonic Techniques

UUV and Sensor Health Monitoring

Feng Xiujuan National Institute of Metrology (NIM),China

The Value. Ultrasound System. Professional Concept for. Women s. Health

ULTRASOUND IMAGING EE 472 F2018. Prof. Yasser Mostafa Kadah

HiFFUT A New Class of Transducer

Confident and Conclusive Diagnosis with Flexible Solutions

Point-of-Care Ultrasound: An Introduction

Focused Ultrasound for Tactile Feeling Display

warwick.ac.uk/lib-publications

US - Ultrasonic Systems. Instrumentation and Applications

Diagnostic Ultrasound. Sutiporn Khampunnip, M.D.

A PRELIMINARY ANALYSIS OF POSSIBILITIES OF COMPENSATING FAULTS OF LASER AND ULTRASONIC TECHNOLOGIES IN SURGERY. Tadeusz GUDRA, Sylwia MUC

The Evolution and Benefits of Phased Array Technology for the Every Day Inspector

The Do s and Don ts of Pressure Transducers

DEVELOPMENT OF ULTRASONIC METHOD OF RADIAL CLEARANCE ASSESSMENT FOR BOLTED JOINTS AT RESTRICTED ACCESS

ULTRASONIC ARRAY APPROACH FOR THE EVALUATION OF ELECTROFUSION JOINTS OF POLYETHYLENE GAS PIPING

TASK: S-2 DISTORTION MEASUREMENT AND BIOMECHANICAL ANALYSIS OF IN VIVO LOAD BEARING SOFT TISSUES

HEARING AND PSYCHOACOUSTICS

Performance of phased array and conventional ultrasonic probes on the new ISO reference block

Ultrasonic Transducer for the Hydrothermal Method

FORMING SCREEN EFFECT ON ULTRASONIC BEAM FIELD

An Overview of Ultrasound Testing For Lesion Detection in Human Kidney

: Biomedical Signal Processing

PZT/ZnO EXPERIMENT MODELLING

MATCHING LAYER DESIGN OF AN ULTRASONIC TRANS- DUCER FOR WIRELESS POWER TRANSFER SYSTEM

Technical Discussion HUSHCORE Acoustical Products & Systems

Numerical Modelling of Ultrasonic Phased Array Transducers and Their Application

LION. Application Note PRECISION. Linear Position Measurement with Eddy-Current Sensors. LA March, Applicable Equipment: Applications:

A new method of sonograph lateral resolution measurement using PSF analysis of received signal

Manual Ultrasonic Inspection of Thin Metal Welds

APPLICATION AND DEPLOYMENT OF ADVANCED NDE TECHNIQUES IN HIGH PRESSURE VESSELS

Transcription:

AN EXPERIMENTAL STUDY ON TRANSDUCER FOR THE DDS TECHNOLOGY DEMONSTRATOR ANDRZEJ ELMINOWICZ, WALDEMAR LIS*, WŁADYSŁAW MĘCIŃSKI R&D Marine Technology Centre Dickmana 62, 81-109 Gdynia, Poland, *University of Technology *Gabriela Narutowicza 11/12, 80-233 Gdańsk, Poland, andrzeje@ctm.gdynia.pl, *wall@eti.pg.gda.pl R&D MTC has began designing and building the next generation wideband Diver Detection Sonar (DDS, with cylindrical multielement piezocomposite transducer. The paper will focus on key features of DDS technology demonstrator, with special attention on technologies of the transducer. The new transducer is designed at the cooperation with the producer of the transducer Materials Systems Inc. (Littleton, USA), remainder hardware and software solutions are designed in R&D MTC. In this article specific theoretical and technological principles of the wideband cylindrical piezocomposite transducer will be presented. The implementation of the transducer to the wideband Diver Detection Sonar (DDS) will be described. The results of transducer measurements, transmitter modules, matching circuits modeling and examples of application will be also presented. INTRODUCTION Piezocomposite 1 3 technology using shading technique shapes the transducer response (beampattern) without resort to complex electronics it is possible due to ceramic rod is electrically isolated from its neighbour as well as due to ceramics rods are embedded in polymer matrix which significantly reduces coupling between the different modes in the ceramic that effectively provides a uni modal response without unwanted resonances. Composites materials are inherently wideband because of damping provided by the polymer matrix and furthermore the introduction of specially designed matching layer between the outer electrode and the sea water.

The composite structures reduces the acoustic impedance closing it to impedance of the sea water it means the higher power is transferred, so efficiency increases. The electro-acoustic performance of these piezoelectric materials embedded in polymer matrix, sophisticated data processing and software solutions have enabled improvement in operational parameters of sonars, adaptation of sonar to specified protected object, propagation and environment condition. The designed technology demonstrator of DDS is a first attempt of our firm leading to next DDS generation intended to operate in shallow water. The new array has been created at the cooperation with Materials Systems Inc. (Littleton, USA), remainder hardware and software solutions are being formed by R&D MTC. Technology demonstrator (TD) of DDS has been designed as two modules: underwater and above water [4]. Both modules connect by cable: power + data (ETHERNET). Underwater parts will be in cylinder form with possibility to be placed on the sea bottom or suspended in column water. In the further parts of the paper, we described in details only one of the most important sub assembly: transducer. The following assumptions for TD s transducer have been established: tranducer will be made as curved piezocomposite (piezoelectric ceramic rods in a polymer matrix) array cylindrical array with 180 o aperture and 64 elements, tranducer will operate in shallow water (survival depth 100 m), frequency range: 60 khz 80 khz, source level: 205 db, vertical side lobe level: -21 db*, diameter: 350 mm, transducer height: 120 mm, elements spacing of transducer: < ½ λ, * vertical side lobe level without shaping is 13.4 db. 1. DESIGN OF TRANSDUCER Technology demonstrator (TD) of DDS utilizes piezoelectric materials in the transducer to generate and receive the acoustic signals. The transducer determines the performance limits of the sonar system. The use of piezocomposite improves transducer s performance. Advantages of piezocomposite transducers for sonars: increased sensitivity, better resolution by broader bandwidth, improved image contrast by reduced side lobes, better efficiency (increased signal to noise ratio), by improved impedance matched to water, low interelement cross talk, greater element to element phase and amplitude uniformity, low cost construction. Piezocomposite design goals: maximization of capacitance, electrical impedance match to system, acoustic impedance match to water,

maximization of electromechanical coupling, electrical loss tangent minimization, mechanical loss (1/Qm) minimization. Any composite design is a compromise among these parameters. For environment condition of Baltic Sea: temperature 10 o C and salinity = 7 o / oo sound speed is approx. 1450 m/s and element spacing of the transducer for 70 khz is 10.36 mm and for 80 khz is 9.06 mm. Calculated, at assumed transducer dimensions, spacing is: d = π * 175/64 8.6 mm We assumed that developed transducer will have 64 elements with width = 7.8 mm and gap = 0.8 mm; assumed spacing (see Fig. 1) is less then theoretical spacing for 80 khz by 5.1 %, which protects against spatial aliasing at lowest frequencies. Average 3 db beam width is: Θ h 50.7 *λ/l 9 o where: λ is wave length and L = 120 mm (see Fig. 1). Beams spacing Θ s 3 o. Fig.1. Elements shape and spacing Transducer element is a piezocomposite block with ceramic rods embedded in polymer matrix; it should be noted, that for ceramic rods with height compared to their lateral dimensions prevail length mode of vibration resulting in improved electroacoustic efficiency, then sensor sensitivity and signal/noise ratio increased. Other benefits, related to applying piezocomposite transducer to sonars, are: better resolution due to broader bandwidth, reduced side lobes and moreover low cost construction [2]. 2. FACTORY ACCEPTANCE TESTS Materials Systems Inc. invited CTM representative to visit MSI facility and laboratory to carry out the factory acceptance test of the sonar transducer. Tests conducted for following parameters: TVR (all elements), RVS (all elements), Impedance in water (all elements), Horizontal directivity with all other elements electrically shorted (all elements).

Tests were conducted in MSI tank for the transducer connected with all elements open circuited except for the element under test. It was determined that capacitive cross coupling between element s wire pairs within 5 meters cable adversely affected element s TVR, RVS and directivity. To eliminate the cross coupling and better simulate performance under actual operating condition (preamps and power amps located within 40 cm of the array), every element was retested for RVS and TVR with all other elements short circuited. This eliminated the performance irregularities. Test results: One element s TVR (ref. 1µP/V for 1 m), predicted 125 db and measured 144 db, One element s RVS (ref. 1V/1µP), predicted 188 db and measured 176 db, Transmit bandwidth (-3 db), predicted 60 khz to 80 khz and measured 61 khz to 82 khz, Receive bandwidth (-3 db), predicted 60 khz to 80 khz and measured 65 khz to 81 khz, One element s horizontal directivity, measured 80º, 24 elements segment beamwidth, measured 50º. The summary impedance in water of piezocomposite transducer elements are presented in Fig. 2 and 3. 100-0599 CTM Poland In-Water Measured Impedance 1000000 100000 Z 10000 1000 100 0 20 40 60 80 100 120 140 160 180 200 Frequency (khz) Fig.2. In water measured impedance module for 64 elements 100-0599 CTM Poland In-Water Measured Phase 90 70 50 30 Phase (degrees) 10-10 -30-50 -70-90 0 20 40 60 80 100 120 140 160 180 200 Frequency (khz) Fig.3. In water measured impedance phase for 64 elements

3. ARRAY VERTICAL BEAM PATTERN During the design of the transducer, vertical beam pattern shading have been included, which theoretically provides -30 db side lobes. Electrical and acoustic performance predictions are based on software models which have generally shown good correlation to typically measured performance, but do not predict secondary effects such as mounting and housing effects of manufacturing variation. Since prototypes are the first units built, according to new design, performance can not be predicted with complete accuracy. Therefore electrical and acoustic performance specifications for prototypes have been met on a best effort basis. However, because the piezocomposite is comprised of discrete piezoceramic pins, the effective pattern is determined by both the screen printed electrode pattern, which is quite precise, and the piezocomposite ceramic pin locations, which are more random. It is therefore not possible to implement a perfect pattern and side lobes should be higher than theoretical. We expect to come close to -21 db side lobes, but can only commit to meeting the specification on a best effort basis. MSI have been conducted additional modelling with 3 elements model of the transducer. Measurements of this model can give better information about vertical beam pattern side lobes. Vertical beam pattern tests of model were conducted in Technical University tank for selected frequencies. Selected test results of vertical beam pattern are presented in Fig 4. Fig.4. Single element vertical beam pattern for frequency 66 khz

Side lobes for narrow band system with one selected frequency are on the level -18 db and -20 db. This result should be accepted because the distance between the composite pins is greater than the pattern resolution required to achieve -30 db shading. We suspect that the results for wideband system will be -21 db. 4. MEASUREMENT AND CALCULATION The model of the transducer consisting of three composite elements was examined in detail. His electric impedance in the function of the frequency and the directivity pattern was measured. Results of the measurement of the impedance are confirming the advantages of composite technology. As a result of lowering the acoustic impedance, the transducer is better matched to the working environment. We can see as the influence of the radiation medium (the water) on the wheel shape of the impedance of the transducer -fig.5. Fig.5. Electric admitance of transducer in: a) air, b) water Through the application of composite technology, the dielectric coefficient of ceramic material falls down. Because of that, the static capacity is smaller and it is easier to compensate the imaginary part of the transducer. Next pictures fig.6 and 7 are showing the shape of real and imaginary part of admittance of the transducer. On this basis it is possible to conclude, that irregularities in shape s characteristic of not-loaded transducer becomes regular after immersing the transducer in water. In this case, when the transducer is loaded, it has one mode of vibration. It is possible to explain it with the fact that such a phenomenon is appearing when the transducer is heavy loaded, which in the case of composite technology is taking place here. Different modes of vibrations disappear.

Fig.6. Conductance of transducer in: a) - air, b) - water Fig.7. Susceptance of transducer in: a) - air, b) - water

5. CONCLUSIONS 1. Technology demonstrator of DDS is the good opportunity for check new technologies in piezocomposite transducers leading to new generation of sonars intended for detection of smaller targets in difficult littoral condition. 2. We anticipate the transducer which is being built using Piezocomposite 1 3 technology will fulfil our requirements and we would gather experience which will enable us to design transducers with other technical parameters (sector, number of elements, source level, etc.). 3. Performance achieved during FAT was generally very good and in the case of RVS and TVR levels much better than predicted. 4. For measured transducer there are a few elements with atypical response. Element-to-element variability should be reduced in future transducers by using more refined and extensive fixture s production. ACKNOWLEDGMENT This work is financed by fund dedicated for research and development project realized during 2007-2010 years. Research and development project No R 00 000 11/3. REFERENCES [1] A. Elminowicz, M. Okuniewski, E. Porosińska, L. Zajączkowski, Diver Detection Active Sonar, TEHOSS, Gdańsk 2005. [2] Brochure: Piezocomposite_Transducer_Design, www.matsysinc.com. [3] T.Janowski, A.Elminowicz, K.Szworski, Data Processing of Broad Band Signals Received by Multielement Cylindrical Transducer, 7 th EEA International Symposium on Hydroacoustic, Jastrzębia Góra, 2008. [4] A. Elminowicz, L. Zajączkowski DDS technology demonstrator, 55 OSA Wrocław Piechowice 2008.